Introduction

Pulmonary vein isolation (PVI) has become the cornerstone of catheter-based treatment for paroxysmal and early persistent atrial fibrillation [1, 2]. Since the introduction of PVI, several ablation technologies have been developed including point-by-point radiofrequency (RF) ablation, multi-electrode radiofrequency ablation, and cryoballoon ablation [1, 3, 4]. In conventional RF PVI, the RF energy is delivered to the tissue in a point-by-point force-controlled manner resulting in cellular necrosis by heating [2]. In cryoballoon (CB) ablation, PVI is achieved by freezing tissue in a “single shot” manner resulting in circumferential cell necrosis [3]. RF PVI guided by 3D electro-anatomic mapping techniques is an effective treatment modality and requires only limited use of fluoroscopy. However, the technique is also considered technically challenging, especially for less experienced operators [2, 5]. CB PVI can be performed without 3D navigation systems and seems to be more accessible, however is associated with a higher risk of phrenic nerve injury [3, 6]. In a recent randomized trial, CB and RF ablation were shown to be equally effective for the treatment of paroxysmal atrial fibrillation. Furthermore, there were no significant differences in overall safety [6]. Importantly, different generation cryoballoons and RF-ablation catheters with and without contact force were used in that study. Although a post-hoc analysis was performed for newer generation catheters, only a small proportion constituted contact-force RF ablation, and the study was not powered to detect differences. We therefore aimed to report single center ablation success, incidence of repeat ablation, extent and rate of pulmonary vein (PV) reconnection, and safety outcomes after second-generation CB and contact-force-guided RF ablation for drug-refractory paroxysmal atrial fibrillation.

Methods

Patient population

Patients who were suitable for primo PVI for drug-refractory paroxysmal or early persistent atrial fibrillation were randomly allocated to cryoballoon or contact-force radiofrequency catheter ablation and included in the Isala Heart Centre AF database. Procedures in which additional ablation strategies [e.g., complex fractionated atrial electrograms (CFAE), linear ablations, right atrium ablations] were applied, were excluded from the present study.

Pre-ablation protocol

All patients underwent cardiac computed tomography (CT) to assess left atrial and pulmonary vein (PV) anatomy. Transthoracic echocardiography was routinely performed to assess left ventricular function, valve abnormalities, and atrial dimensions. Patients were admitted 1 day before the ablation procedure. During hospitalization, cardiac rhythm was continuously monitored in all patients. On admission and after ablation, routine blood samples were taken for screening purposes. Pre-ablation transoesophageal cardiac echocardiography was performed to assess the interatrial septum and rule out left atrial thrombus.

Ablation protocol

The PVI ablation procedures were performed under conscious sedation or under general anaesthesia supervised by a cardiovascular anesthesiologist, in accordance with HRS/EHRA/ECAS expert consensus on AF ablation. The ablation catheter was introduced in the left atrium via a transseptal puncture under pressure and fluoroscopic guidance, using a Brockenbrough needle. During the procedure, anticoagulation was checked at intervals of 30 min, with a targeted activated clotting time between 300 and 350 s.

Catheter ablation technique

Cryoablation

Cryoballoon ablations were performed using the second-generation Artic Front cryoballoon catheter (Medtronic, Inc., Minneapolis, MN, USA). This catheter has a bidirectional steering mechanism and an over-the-wire design. The 28-mm diameter balloon was used to create “single shot” circular antral lesions at all four PV ostia [left superior PV (LSPV), left inferior PV (LIPV), right superior PV (RSPV), right inferior PV (RIPV)] in order to create electrical PV isolation. Balloon selection was based on the PV size assessed by pre-ablation cardiac CT. Catheter placement and PV occlusion were checked by fluoroscopy. Cryo-energy applications were performed up to 240 s, with an additional number of Cryo applications at the operator’s discretion. After all ablations were performed, electrical PV isolation of PVs was checked with a standard circular mapping catheter. Adenosine testing was not performed. Ablation of the RSPV and RIPV was performed under continuous phrenic nerve stimulation with palpation of diaphragmatic excursions and direct fluoroscopic visualization of the diaphragm in case of any diminished activity. Cryoablation was immediately terminated on any signs of phrenic nerve palsy.

Radiofrequency ablation

A steerable circular multipolar catheter (Lasso™, Biosense Webster Inc., Diamond Bar, CA, USA) and an irrigated RF contact-force ablation catheter with a 3.5-mm tip (Smarttouch Thermocool™, Biosense Webster Inc., Diamond Bar, CA, USA) were introduced in the left atrium through two transseptal sheaths. Electro-anatomical mapping of the left atrium and PV signal recordings were then performed, using 3D navigation system (Carto3™, Biosense Webster Inc., Diamond Bar, CA, USA). RF energy was applied in a point-by-point manner to create continuous circular lesions in order to achieve electrical PV isolation. Power setting was adjusted between 30 and 40 W with a continuous 0.9% NaCl flow rate of 2 ml/min and 20–30 ml/min during RF energy application. RF energy was applied with a targeted force–time integral (FTI) of > 400 g with an RF application duration of 20 to 40 s, targeting an impedance drop of at least 10 Ohm, or until elimination of the local electrogram. The endpoint of the ablation procedure was defined as the absence or dissociation of PV potentials within the PVs ≥ 30 min post-ablation.

Post-ablation protocol

Patients were hospitalized for at least 24 h with continuous telemetry monitoring. Anticoagulation was continued after the procedure with a target international normalization ratio (INR) of 2.0–3.0 for at least 3 months after ablation. In patients using non-vitamin K antagonist oral anticoagulation (NOAC), this was resumed after the ablation procedure. After 3 months, anticoagulation therapy was (dis)continued at the discretion of the referring physician and in accordance with atrial fibrillation guidelines. Complications were defined in accordance with HRS/EHRA/ECAS expert consensus on AF ablation [2].

Follow-up

A blanking period of 90 days was defined after PVI. Outpatient clinic visits were planned at 3, 6, and 12 months after PVI. Follow-up included physical examination, electrocardiography (all visits), event monitoring and 24–72 h Holter ECG (at 6 and 12 months). In case of symptoms, patients were immediately referred to the emergency department. Additional Holter ECG monitoring was performed if no arrhythmia could be detected. AF occurrence in the blanking period was not considered a relapse. After the blanking period antiarrhythmic drugs were discontinued. The need for a repeat ablation for recurrent symptomatic atrial arrhythmias was left at the discretion of the physician.

Redo procedures

Patients with recurrences, not sufficiently suppressed by AAD or not willing to use AAD, and with sufficient symptoms, underwent a redo procedure. The redo procedure was performed with Carto 3D electro-anatomical mapping using a circular decapolar mapping catheter (Lasso™, Biosense Webster Inc., Diamond Bar, CA, USA) and RF ablation with an irrigated contact-force ablation catheter with a 3.5-mm tip (Smarttouch Thermocool™, Biosense Webster Inc., Diamond Bar, CA, USA). After the creation of a 3D electro-anatomical map and voltage map of the left atrium and pulmonary veins, the antra were ablated with the primary goal of pulmonary vein isolation. Additional ablation was left to the discretion of the operator.

Study endpoints

The primary effectiveness endpoint was single procedure arrhythmia-free survival, defined as patients without AF/atrial flutter (AFl)/atrial tachycardia (AT) recurrence, after the blanking period of 90 days after PVI. Arrhythmia recurrence was defined as AF/AFl/AT recorded on an ECG or a 30 s telemetry strip, in accordance with HRS/EHRA/ECAS expert consensus statement on AF ablation.

Statistical analysis

Data are mentioned as mean ± standard deviation (SD), median with interquartile range (IQR), or percentage where appropriate. Statistical significance between groups was calculated using Student’s t test for continuous variables and Chi-square test for categorical variables. Arrhythmia-free survival differences between catheter groups were assessed with a log-rank test. Univariate and multivariate analysis was performed using Cox regression. A P value of ≤ 0.05 was considered statistically significant. Statistical analysis was conducted using IBM Statistics version 22.0 (IBM SPSS Statistics for Macinthosh, 2013: IBM Corp., Armonk, NY, USA).

Results

Patient population

Altogether, 269 consecutive patients with drug-refractory paroxysmal or early persistent AF undergoing PVI were included and randomly allocated to cryoballoon or radiofrequency ablation. Mean age was 58.9 ± 10.4 years, 191 (71%) were men; 136 patients underwent CB and 133 patients underwent RF ablation. General patients’ characteristics were similar in both treatment groups. Baseline characteristics are presented in Table 1.

Table 1 Baseline characteristics

Acute PVI

Acute PVI was achieved in all patients. A total number of 1103 PVs were ablated [1068 standard PVs, 4 left common PVs, 2 left middle PVs (LMPV), 29 right middle PVs (RMPV)]. Acute PVI results are described in Table 2.

Table 2 Procedural characteristics

Procedural characteristics

Mean procedure duration was significantly shorter in CB ablation as compared to RF ablation (166.5 ± 36.8 vs 184.1 ± 40.0 min, P = 0.016). Procedural characteristics are shown in Table 2. Fluoroscopy times of the ablation groups were not significantly different (11.5 ± 8.5 vs 9.1 ± 14.3 min, P = 0.26).

Complications

There were no significant differences with regard to overall complication rate between CB and RF ablation (6.0 vs 6.7%, P = 1.00). Temporary phrenic nerve palsy did occur more often in the CB group as compared to RF, although this did not reach statistical significance (2.9 vs 0%, P = 0.058). All complications were of a temporary nature. Complication characteristics are shown in Table 3.

Table 3 Complications

Arrhythmia-free survival

Median follow-up was 389 days (IQR 219–599) and did not differ in the CB and RF groups, respectively (P = 0.98). In the CB group, arrhythmia-free survival after a single PVI procedure was 75.2% with a median follow-up duration of 381 days (IQR 292–563). In the RF group, arrhythmia-free survival after a single PVI procedure was 57.4% with a median follow-up duration of 394 days (IQR 189–622). There was a significant difference in arrhythmia-free survival between catheter groups (log-rank test, P = 0.013) favoring CB. Arrhythmia-free survival among catheter groups is shown in Fig. 1. In univariate and multivariate analysis, RF ablation, persistent AF, and AF duration were significantly associated with decreased arrhythmia-free survival. Univariate and multivariate analysis results are displayed in Table 4. Analysis in paroxysmal AF patients only shows arrhythmia-free survival of 76.9% in CB group vs 60.7% in RF group, log-rank P = 0.019. There was no significant difference in type of atrial arrhythmia (AF/AFl/AT) recurrence between treatment groups (Chi-square test: P = 0.90), as described in Table 5.

Fig. 1
figure 1

Kaplan–Meier curve showing atrial arrhythmia-free survival after first pulmonary vein isolation procedure for cryoballoon and radiofrequency ablation. Log rank: P = 0.013

Table 4 Univariate and multivariate analysis
Table 5 Type of atrial arrhythmia recurrence

Repeat ablation and PV reconnection

In total, 53 patients (19.7%) required a repeat ablation for recurrent atrial arrhythmias. Incidence of repeat ablation procedures was significantly lower after CB ablation as compared to RF ablation (14.7 vs 24.8%, P = 0.045). As mentioned, in all patients, all PVs were isolated in the index procedure. A total number of 211 PVs were assessed for electrical PV reconnection. In 49 of 53 patients (92.5%), PV reconnection was found, without a significant difference between CB vs RF (85.0 vs 97.0%, P = 0.15).

However, the number of reconnected PVs at repeat ablations was lower after CB ablation as compared to RF (36.8 vs 58.1%, P < 0.01), as shown in Figs. 2 and 3. The reconnection rate of the right-sided PVs was significantly higher after RF ablation as compared to CB (90.9 vs 50.0%, P < 0.01). PV reconnection is shown in Table 6.

Fig. 2
figure 2

Percentage of PVs that exhibited reconnection at repeat PVI procedure. PV pulmonary vein, PVI pulmonary vein isolation. Chi-square test: P = 0.03

Fig. 3
figure 3

Number of PVs per patient that exhibited reconnection for each ablation catheter. PV pulmonary vein

Table 6 Pulmonary vein reconnection patterns

Sensitivity analysis

We performed Cox regression analysis and plotted a Kaplan–Meier curve (Fig. 4) in patients with paroxysmal AF only, and similar results were observed, i.e., a higher arrhythmia-free survival rate in patients undergoing cryoablation. Additionally, our multivariate analysis provided insight into the independent association of the sort of ablation (cryoablation/RF ablation) and the type of AF (paroxysmal/persistent) and with arrhythmia-free survival after ablation.

Fig. 4
figure 4

Kaplan–Meier curve showing atrial arrhythmia-free survival after first pulmonary vein isolation procedure for cryoballoon and radiofrequency ablation in paroxysmal AF patients. Log rank: P = 0.019

Discussion

This study reports single center data on single ablation success, safety outcomes, and pulmonary vein reconnection patterns after PVI, comparing second-generation cryoballoon ablation to point-by-point contact-force radiofrequency ablation. Acute PVI was always reached using both ablation strategies. However, higher arrhythmia-free survival, lower rate of repeat ablation and a lower rate of reconnected PVs per patient were found after CB as compared to RF ablation. Both techniques had a high safety profile, and all observed complications were of a temporary nature.

Procedural characteristics

High, acute ablation success (100%) was achieved using both ablation strategies. Procedure duration was significantly shorter in CB ablation as compared to RF, in accordance with a recently reported randomized multicenter trial [6]. This might be explained by easier catheter handling and the “single shot” ablation strategy of the CB. However, adequate CB placement was checked by using fluoroscopy after contrast injections, whereas RF ablations were guided by electro-anatomic and force-controlled mapping and RF ablation. In our study, fluoroscopy time in CB ablation was comparable to RF ablation. This is in contrast with the results of a recent trial that showed a higher fluoroscopy time in the CB arm [6]. Although a trend to lower radiation dose was observed in RF ablation, we did not find a statistical significant difference as compared to CB, possibly due to a lower number of patients that were involved in our study.

Complications

The reported overall complication rate of 6.3% is in line with previous reports [6, 7]. We found no significant difference in complication rate between both ablation strategies. Furthermore, all observed complications were of temporary nature. The most common complication in the CB group was temporary phrenic nerve palsy, with a rate of 2.9%, which is considerably lower than reported in the sustained treatment of paroxysmal atrial fibrillation (STOP AF) trial (13.5%) and in line with the Fire&Ice trial (2.7%) [3, 6]. The most reported complication in the RF group was vascular access complication (2.2%). Although not different from CB ablation (2.2%), the groin complication rate was lower than the reported 4.3% in the Fire&Ice trial. In this study, there was one (0.8%) transient ischemic attack (TIA), which is comparable to other PVI techniques.

Arrhythmia-free survival

Our main finding of improved success rate after a single procedure CB ablation compared with point-by-point RF, is in line with earlier reports [8, 9]. However, in a recent multi-center randomized trial (Fire&Ice), CB ablation was shown to be non-inferior to point-by-point RF ablation regarding efficacy [6]. Success of the CB might have been compromised in this trial, due to inclusion of first-generation CB procedures [10, 11]. The design of the second-generation CB with improved circumferential cooling, results in more extensive circumferential lesions, and might have contributed to improved lesion formation [12]. In addition, previous studies demonstrated that Cryo ablation creates more extensive lesions such as, at least partial, posterior wall ablation or ablation of ganglionated plexi, which might have incremental benefits, resulting in a higher arrhythmia-free survival [13, 14]. Indeed, results of the FIRE&ICE study show a trend of higher success rate after second-generation CB as compared to the first-generation; however, the trial was not powered to test superiority of one of the catheters [6]. Of note, the proportion of patients treated with the nowadays commonly used contact-force RF catheter was small in the Fire&Ice.

In multivariate analysis, persistent AF was an independent predictor of reduced arrhythmia-free survival (hazard ratio 1.667; 95% CI 1.006–2.761; P = 0.047). Catheter ablation of persistent AF remains challenging and is associated with less favorable outcome [7, 15]. There is a need for further improvement of ablation techniques, and knowledge of substrate modification to improve AF ablation outcome in this subgroup of patients. In our present analysis, however, we performed sensitivity analyses in paroxysmal AF patients only, and observed similar results in this subgroup. Moreover, multivariate analysis showed an independent association of cryoablation with arrhythmia-free survival, with the type of AF in the multivariate model, suggesting that possibly also in persistent AF patients, cryoablation could be superior. The number of patients with persistent AF was too little, however, to perform separate analysis. The proportion of patients with persistent AF was not significantly different between treatment groups, although relatively small cohorts are always at risk for selection bias. We realize that this could be a potential limitation.

PV reconnection

Significantly less patients required a repeat ablation after CB as compared to RF (14.7 vs 24.8% P = 0.045), which was in line with the lower recurrence rate. At repeat intervention, all PVs were assessed using a circular mapping catheter. We found a lower number of reconnected PVs after CB ablation as compared to point-by-point RF, as shown in Fig. 2. This finding is in line with previous studies that showed improved lesion durability following PVI with the second-generation cryoballoon [16, 17]. Moreover, the right-sided PVs were more likely to exhibit electrical reconnection after RF ablation as compared to CB ablation. This finding might be explained by a more challenging catheter approach to the right-sided PVs, resulting in reduced durability of the RF ablation lesions. Compared to the RF catheter, the circular design of CB allows a relative, simple positioning and alignment of the catheter, which might lead to improved lesion quality. In contrast to prior reports, we did not find higher reconnection rates of the right inferior PVs and left common PVs after CB ablation [17]. Although a waiting period is constituted before assessing acute PV isolation, when starting ablation in the left pulmonary veins, the waiting period is always shorter for the right-sided PVs. Furthermore, a steerable sheath was not systematically used in the RF-ablation group, but only when deemed necessary in order to reach adequate catheter contact (> 10 g and FTI > 400 g). We did not systematically test for durable PV isolation with adenosine, which might possibly have influenced these results. It could be that cryoablation is less prone to edema-induced acute PV isolation that reconnects in the days to weeks after the procedure. Pre-procedural assessment of PV anatomy may aid in utilization of CB or RF catheter. Limitation of the CB size should be considered in case of increased PV diameter of left common PVs.

Future perspectives

AF recurrences after PVI still remain an important clinical problem. Acute PV isolation does not assure long-term electrical PV isolation, and this might partially explain AF recurrence. Future research is needed to give more insight into mechanisms of conduction recovery. Furthermore, continuous improvement of catheter design and characteristics might improve catheter handling and lesion quality. This might result in better lesion durability and improved clinical outcome.

Limitations

There are several limitations that should be considered for interpretation of the results of this study. This is a single center, prospective, non-randomized study reviewing results of primo pulmonary vein isolation procedures using second-generation CB vs contact-force radiofrequency RF in patients with atrial fibrillation. Unadjusted confounding factors might be present in the study, i.e., no data on PV reconnection in patients without recurrence of arrhythmia are available. Pulmonary vein reconnection was only assessed at repeat ablation procedures, and therefore, by definition in symptomatic patients with AF recurrence.

Conclusions

In this single center, prospective study, improved arrhythmia-free survival and more durable PV isolation is seen after PVI using second-generation CB as compared to contact-force RF, in patients with drug-refractory paroxysmal AF. Complication rates for both ablation techniques are low, and all occurred complications were of a temporary nature.