Introduction

Cardiac amyloidosis is an underdiagnosed condition due to the deposition of non-soluble protein aggregates. Amyloid transthyretin (ATTR) cardiac amyloidosis is a rapidly progressive heart disease, with a median survival (if not treated) of 3.5 years [1].

Scintigraphy with bone-seeking radiopharmaceuticals has been demonstrated to be a sensitive non-invasive technique in diagnosing ATTR cardiac amyloidosis even in an early phase of disease [2].

Planar scintigraphic images obtained after intravenous injection of [99mTc]Tc-hydroxymethylene diphosphonate (HMDP) or [99mTc]Tc-3, 3-diphosphono-1, 2-propanodicarboxylic acid (DPD) are typically evaluated using qualitative (visual) analysis that requires late phase images, obtained 2 or 3 h after radiotracer injection [3]. Recently, Galat et al [4] reported that early phase images, obtained 10 min after [99mTc]Tc-HMDP injection, provide enough information to diagnose cardiac amyloidosis and proposed to use them to differentiate ATTR amyloidosis from other variants of cardiac amyloidosis and from other causes of left ventricle hypertrophy, especially in frail patients. On the other hand, it has been suggested that radiotracer staying in the blood pool may produce false positive findings and that tomographic imaging may be useful to distinguish myocardial radiotracer uptake from tracer accumulation in the cardiac blood pool [3, 5].

The aim of our study is to evaluate the diagnostic performance of early phase whole body scan using [99mTc]Tc-DPD with respect to late phase imaging.

Materials and methods

We retrospectively reviewed [99mTc]Tc-DPD imaging performed at our Institution in the last 15 years in patients referred for suspect cardiac amyloidosis. Inclusion criteria were as follows: a diagnosis of ATTR amyloidosis was reached; [99mTc]Tc-DPD was used for nuclear medicine imaging; both early and late phase whole body images were obtained; thorax SPECT imaging was obtained in the same imaging session. When a patient had been studied at different time points during the course of his/her disease, only one series of scans and SPECT was randomly chosen. The study was approved by our institutional review committee. Informed consent was obtained from all subjects.

Patients

We included in the study 53 patients (24 males; 29 females; median age: 53 years; age range 27–88 years); 48 had hereditary ATTR (ATTRv) amyloidosis and 5 wild type ATTR (ATTRwt) amyloidosis. Gene mutations in ATTRv amyloidosis patients were as follows: 25 E89Q (p.E109Q), 17 F64L (p.F84L), 3 T49A (p.T69A), 2 V122I (p.V142I), 1 E89Q + G6S (p.E109Q + p.G26S).

Among 48 patients with ATTRv amyloidosis, 30 patients showed no limitation of physical activity (NYHA class 1), 14 had slight limitation of physical activity (NYHA class 2), and 4 had marked limitation of physical activity (NYHA class 3). Biventricular systolic function was normal in most subjects — median left ventricular ejection fraction: 64.5% (range 54–71%); median tricuspid annular plane systolic excursion: 19 mm (range 12–23 mm) — whereas diastolic function was abnormal in many patients (median E/e’ = 8; range 3–18). Interventricular septum thickness varied greatly among patients (median: 12 mm; range 8–20 mm).

Ten out of 48 patients with ATTRv amyloidosis had no neuropathy symptoms, 14 had sensory disturbances but preserved walking capability (polyneuropathy disability score (PND) [6] stage I), 11 showed impaired walking ability without need for a stick (PND stage II), 4 could walk only with the help of one stick (PND stage IIIA), 4 could walk only with the help of 2 sticks (PND stage IIIB), and 5 were confined to a wheelchair or bedridden (PND stage IV).

Seventeen of these 48 patients had no autonomic nervous system dysfunction, 15 showed mild autonomic dysfunction (compound autonomic dysfunction test [7] between 16 and 19 for males and 12–15 for females), 10 moderate autonomic dysfunctions (compound autonomic dysfunction test 12–15 for males and 9–12 for females), and 6 severe autonomic dysfunctions.

Among 5 patients with ATTRwt amyloidosis, 1 patient showed no limitation of physical activity (NYHA class 1) and 4 had slight limitation of physical activity (NYHA class 2). Left ventricular ejection fraction ranged between 51 and 66% (median: 58%), tricuspid annular plane systolic excursion between 16 and 23 mm (median: 18 mm), E/eo between 5 and 21 (median: 15), and interventricular septum thickness between 16 and 23 mm (median: 20 mm).

[99mTc]Tc-DPD imaging acquisition and analysis

Whole body images (anterior and posterior views) were obtained 5–10 min (early phase images) and 180 min ± 15 min (late phase images) after intravenous injection of 740 MBq of [99mTc]Tc-DPD by using a dual headed gamma camera. Sweep whole body scans were obtained in cranio-caudal direction with a 11 cm/min scan speed. Thus, heart was imaged about 3–5 min after starting (8–15 min after radiotracer injection).

SPECT images were obtained immediately after late phase whole body imaging using a circular orbit, a 360° rotation, a 3° step-and-shoot technique, and an acquisition time of 25 s per frame with a 128 × 128 matrix.

Early phase images (anterior and posterior views) were randomly evaluated by two independent nuclear medicine physicians experienced in cardiac amyloidosis imaging. Early images were visually assessed for cardiac uptake and categorized as negative (no cardiac uptake) or positive (presence of cardiac uptake) (Fig. 1). A third nuclear medicine physician reviewed data and identified cases that have received different categorization by the two readers; cases were resubmitted to the two readers who evaluated images in consensus to provide a definitive categorization. One month later, late phase images (anterior and posterior views) were randomly evaluated by (the same) two independent nuclear medicine physicians according to Perugini’s score (0 = absent cardiac uptake and normal bone uptake; 1 = mild cardiac uptake inferior to bone uptake; 2 = moderate cardiac uptake with attenuated bone uptake; 3 = strong cardiac uptake with mild/absent bone uptake) [8] and categorized as negative (score 0) or positive (score 1 or 2 or 3). A third nuclear medicine physician reviewed data and identified cases that have received different scores by the two readers; cases were resubmitted to the two readers who evaluated images in consensus to provide a definitive score. One month later, thorax SPECT images were randomly evaluated, in consensus, by the same two nuclear medicine physicians and categorized as negative (no myocardial uptake) or positive (presence of myocardial uptake). Finally, case images showing different results across the three different sets of images were reviewed to find possible causes for the discrepancies, considering SPECT results as the standard of reference.

Fig. 1
figure 1

Representative images for visual assessment of early (soft tissue) phase scans showing no cardiac radiotracer accumulation (negative scan; a) or cardiac radiotracer uptake (positive scan; b)

Moreover, a semiquantitative analysis was performed on both early and late phase imaging. Two different indexes were obtained from early phase images: (1) heart to mediastinum ratio (H/M) and (2) heart to contralateral ratio (H/CL). H/M was obtained by dividing counts in the left ventricle region of interest (ROI) by counts in the mediastinum ROI (this ROI was obtained by copying ROI and pasting it on the large thoracic vessels, in the center of the thorax, just above left ventricle ROI) [4]. H/CL was obtained by dividing counts in the left ventricle ROI to counts in a mirror contralateral ROI. All these ROIs were placed on the anterior view of the early phase whole body scan [9]. Similarly, two different indexes — (1) heart retention (HR) and (2) heart/whole body ratio (H/WB) — were obtained from late phase images, as previously described [8, 10]. Briefly, HR was obtained as the ratio between decay-corrected counts of the heart in late phase images and counts in early phase whole-body images whereas H/WB was determined by dividing counts in the heart by whole-body images counts in late scintigraphy; heart and whole body counts were obtained as geometric means of respective counts in the anterior and posterior projections.

Statistical analysis

Cohen’s k statistics has been used to evaluate interobserver agreement for categorization of each phase imaging and to evaluate agreement between early and late phase categorizations for each observer and for consensus reading.

Kolmogorov–Smirnov test has been used to examine if continue variables were normally distributed.

The Mann–Whitney test or the Kruskal–Wallis test with Dunn’s post hoc test, as appropriate, has been used to assess the differences in quantitative parameters between groups. The chi-squared test for independence or the Fisher’s exact test, as appropriate, was used to assess the differences in categorical parameters between groups, after adequate pooling. In particular, Fisher’s exact test has been used to compare, for each observer and for consensus readings, findings of early and late phase images with SPECT results; sensitivity, specificity, and positive and negative predictive values of early and late phase images were also determined.

Receiver operating characteristics curves analysis and Youden’s test were used to compare performances and to determine the best threshold to predict SPECT findings for each semiquantitative index.

A two-sided p value < 0.05 has been considered to be statistically significant.

Statistical analyses were performed using MedCalc for Windows, version 19.4.1 (MedCalc Software).

Results

SPECT imaging was negative for cardiac accumulation in 25 patients and positive in 28 patients. No statistically significant difference in the left ventricular ejection fraction was found between the two groups (p = 0.28). There was no correlation between Perugini’s score and the left ventricular ejection fraction (p = 0.36). On the other hand, a statistically significant correlation was found between Perugini’s score and the following indicators: severity of cardiac symptoms (NYHA class) (p < 0.0001), TAPSE (p = 0.018), diastolic function (E/e’) (p = 0.0016), interventricular septum thickness (p = 0.0003), PND (p = 0.0002), and degree of autonomic nervous system dysfunction (p = 0.0015).

Results of visual analysis for both readers and consensus readings are reported in Table 1.

Table 1 Results of visual analysis of both early and late phase whole body images for two readers and consensus readings

An almost perfect agreement between the two observers was found for categorization of both early phase images (94.34% of agreement; Cohen’s K: 0.8867) and late phase images (96.23% of agreement; Cohen’s K: 0.9244); in all patients who received different evaluations from observers, Perugini’s score for late phase images never exceeded 1.

An almost perfect agreement was found between categorization of early and late phase images (Table 2); in all cases of disagreement, Perugini’s score for late phase images did not exceed 1.

Table 2 Agreement between categorization of early and late phase images for two readers and consensus readings

The capability of early phase images to predict SPECT findings (standard of reference) was extremely significant, but did not reach complete agreement in any case (Table 3); the capability of late phase images to predict SPECT findings was extremely significant, but complete agreement was reached only by consensus reading (Table 4). In all cases of disagreement, Perugini’s score for late phase images did not exceed 1.

Table 3 Sensitivity, specificity, positive and negative predictive values of early phase whole body images in predicting SPECT findings
Table 4 Sensitivity, specificity, positive and negative predictive values of late phase whole body images in predicting SPECT findings

Both readers reported evidence of cardiac blood pool as a cause of false positive results for both early and late phase images (Fig. 2); on the other hand, absence of clear radiotracer accumulation in myocardial wall represented a cause of false negative results for both early and late phase images (Fig. 3). Increased background activity (soft tissue radiotracer accumulation) and superimposition of radiotracer accumulating structures were additional causes of incorrect categorization of planar images by visual analysis.

Fig. 2
figure 2

Early (soft tissue) phase scan (a) and late (bone) phase scan (b) images in a patient harboring ATTRv amyloidosis (F64L mutation) with no evidence of cardiac radiotracer accumulation on SPECT images (c). Early phase scan was judged to be positive by one reader

Fig. 3
figure 3

Early (soft tissue) phase scan (a) and late (bone) phase scan (b) images in a patient harboring ATTRv amyloidosis (E89Q mutation) with evidence of cardiac radiotracer accumulation on SPECT images (c). Early phase scan failed to reveal cardiac radiotracer accumulation whereas late phase scan was judged to be negative by one reader

At late phase whole body images, we found (non-osseous) extracardiac lesions in 26/53 patients (49.1%); most frequently involved sites were soft tissues (muscles, fat) (49.1%) and pleuro-pulmonary (27.7%). At early phase whole body images, extracardiac radiotracer accumulation was found in only 7/53 patients (13.2%); most frequently involved sites were soft tissues (muscles, fat) (9.4%) and pleuro-pulmonary (5.7%). All patients showing extracardiac radiotracer accumulation at early phase imaging also showed it at late phase imaging. Moreover, all patients showing extracardiac radiotracer accumulation also presented cardiac radiotracer accumulation.

As far as semiquantitative indexes are regarded, receiver operating characteristics curves analysis revealed a better performance of H/M for early phase images evaluation and of HR index for late phase images evaluation; interestingly, semiquantitative analysis of early phase images performed better than semiquantitative analysis of late phase images (Table 5; Fig. 4). Namely, a threshold of 1.13 for H/M reached the best results showing 96.43% sensitivity and 96% specificity in predicting SPECT findings. Best thresholds for each index and related performances in predicting SPECT findings are reported in Table 5.

Table 5 Best thresholds and performances of semiquantitative indexes in predicting SPECT findings
Fig. 4
figure 4

Receiver operating characteristics curves showing capability of semiquantitative indexes to predict SPECT findings

Discussion

Amyloidosis is a systemic disease characterized by the deposition of non-soluble protein fibrils in several tissues and organs. ATTR amyloidosis may be due to the deposition of wild type ATTR (usually involving the heart of elderly patients) or mutated ATTR. ATTRv amyloidosis is a rare form of genetic disease due to different TTR gene mutations that may be responsible for different locations of amyloid fibrils deposition. Accordingly, the clinical picture may be characterized by a sensitive and/or vegetative neuropathy and/or a cardiomyopathy.

Cardiac amyloidosis is a rapidly progressive heart disease carrying a poor prognosis [1]. Nowadays, several therapeutic options for ATTR amyloidosis are available [11,12,13,14,15]; thus, an early diagnosis is strongly desirable. Although the standard of reference to diagnose cardiac amyloidosis is represented by endomyocardial biopsy with immunohistochemical analysis or mass spectroscopy, non-invasive techniques may largely contribute to assess cardiac amyloidosis. While it has been reported that even a combination of first-line techniques as echocardiography and electrocardiography can permit early identification of cardiac amyloidosis in patients with TTR gene mutations [16], scintigraphy using bone-seeking radiopharmaceuticals has been demonstrated to reveal cardiac involvement earlier; furthermore, cardiac [99mTc]Tc-DPD accumulation may precede neurological symptoms in patients carrying gene mutations associated with a mixed neurologic/cardiologic phenotype [17]. Moreover, nuclear medicine imaging using bone seeking radiopharmaceuticals may contribute to differentiate between amyloid light chain amyloidosis and ATTR amyloidosis, thus replacing the need for biopsy in several cases [18].

Nuclear medicine imaging using bone seeking radiopharmaceuticals is typically evaluated by visual grading on late phase images; indeed, the most largely used scoring system is based on visual comparison between cardiac and bone uptake [3, 8]. Similarly, several different semiquantitative indexes, for whose determination late phase images are needed, have been defined and studied [2]. Namely, it has been reported that, in ATTRv amyloidosis patients, cardiac retention of bone seeking radiotracers on late phase images correlates with septal thickness, diastolic and strain impairments [19] whereas whole body retention of radiotracer correlates with neurological impairment [20].

Recently, Galat et al reported that early phase [99mTc]Tc-HMDP scintigraphy (whole-body scan obtained 10 min after radiotracer administration) perfectly predicts late-phase findings; moreover, they reported that an early H/M ≥ 1.115 predicts late cardiac accumulation with a sensitivity of 100% and a specificity of 97%, while an early H/M ratio ≥ 1.210 discriminates amyloid light chain amyloidosis from ATTR cardiac amyloidosis with a perfect accuracy (100% sensibility and specificity) [4]. In another study of the same Group, Abulizi et al compared myocardial uptake intensity of [99mTc]Tc-HMDP and [99mTc]Tc-DPD in a small series of patients affected by cardiac ATTRv amyloidosis. However, the comparison that showed comparable radiotracers uptake involved only six patients and was based only on early phase images [21]. More recently, Santarelli et al demonstrated that it is possible to determine the presence of ATTR cardiac amyloidosis, by using [99mTc]Tc-HMDP, evaluating the time-activity curve “heart/bone(t)” and the retention index “heart–bone(t)”, for t > 6 min [22]. On the other hand, it has been reported that radiotracer staying in the blood pool may be responsible for false positives results on planar imaging and that SPECT/CT may be useful in solving this issue [3, 5].

In the present study, we tried to answer the following questions: can [99mTc]Tc-DPD early phase whole body scan images predict late phase whole body scan findings? What is the performance of early and late phase images with respect to SPECT imaging?

Our series included patients with different TTR gene mutations, characterized by different phenotypes and rates of cardiac involvement [15, 23, 24], and different degrees of cardiac uptake [25]. Moreover, differently from the study by Galat et al [4], our series included many ATTRv amyloidosis patients in early phase of disease.

We demonstrated that visual analysis of early phase whole body scan using [99mTc]Tc-DPD almost perfectly predicts late phase whole body findings and SPECT results even in patients in an early stage of disease; visual analysis of late phase images performs only slightly better than early images in predicting SPECT results; semiquantitative analysis does not ameliorate visual analysis performance.

Our data confirm that early phase planar images can correctly reveal cardiac radiotracer accumulation in a large majority of cases also using different bone seeking radiotracers. Indeed, our results using [99mTc]Tc-DPD are similar to that of Galat et al [4] and are in line with that of Santarelli et al [22] who used [99mTc]Tc-HMDP. On the other hand, the lack of studies evaluating the diagnostic capability of [99mTc]Tc-pyrophosphate (PYP) imaging at early time points (less than 1 h) makes it difficult to say if our results are applicable also for this radiotracer. Since similar blood pool clearance for [99mTc]Tc-DPD and [99mTc]Tc- PYP has been reported in a rat model [26], we can only speculate that similar results could be obtained.

Visual evaluation of early phase images may be difficult and inaccurate only when subtle and/or segmental myocardial radiotracer accumulation is appreciable on SPECT imaging. Nevertheless, the evaluation of planar late phase images does not perform significantly better. Furthermore, interobserver agreement is not absolute and the same observer may differently categorize early and late phase images from the same patient. Blood pool activity may be responsible for false positive results on both early and late phase images (Fig. 2); on the other hand, blood pool activity may obscure mild myocardial radiotracer accumulation resulting in false negative results (Fig. 3). Additionally, superimposition of radiotracer-accumulating structures (e.g., calcifying costal cartilage, sternum and spine superimposition) may determine false positive or false negative results on late phase imaging.

These assumptions could be confirmed by a multicenter study by Castano et al who reported that visual analysis of [99mTc]Tc-PYP scintigraphy obtained 1 h after radiotracer administration is more sensitive but less specific than visual analysis of [99mTc]Tc-PYP scintigraphy obtained at 3 h (95% vs 58%, and 79% vs 100%, respectively) because of reduction of myocardial uptake and increase of bone uptake over time. Furthermore, they suggested that the issue of increased bone activity at 3 h planar imaging can be overcome by SPECT imaging [27].

Semiquantitative analysis in our study did not show better performances than visual analysis. Moreover, semiquantitative indexes extracted from late phase images to determine cardiac amyloid load seem to be largely influenced by radiotracer accumulation in other soft tissues and bone [27,28,29], being unable to monitor cardiac amyloid deposition progression. Minutoli et al reported no statistically significant differences for HR and H/WB between patients with visual score 0 and patients with visual score 1 and no statistically significant modifications of such indexes have been found over 4 years follow-up [17]. Similarly, no significant modifications of H/CL have been recognized over 1.5 years follow-up in patients with advanced TTR-related amyloidosis despite clear clinical progression [30]. Moreover, a large overlap of semiquantitative indexes between patients with Perugini’s score 2 and 3 has been reported [31] and radiotracer uptake in skeletal muscle (and not in the myocardium) has been suggested to be responsible for different appearance in patients with Perugini’s grade 2 and grade 3 cardiac uptake [29].

Among the different semiquantitative indexes that we tested, H/M ratio performed better than H/CL ratio probably because it is less influenced by the presence of extracardiac uptake (namely lung uptake) which frequently occurs after [99mTc]Tc-DPD or [99mTc]Tc-HMDP administration [32, 33] and is far less common when [99mTc]Tc-PYP is used [34, 35]. Moreover, on early phase images, whatever bone seeking radiotracer is used, H/M ratio might exert a better “comparison” of radiotracer activity in the cardiac area (blood pool activity plus myocardial uptake) with blood pool activity in the mediastinum.

The main sign that permits to differentiate between cardiac blood pool activity and myocardial radiotracer accumulation due to amyloidosis is evidence of hyperactive myocardial wall (Fig. 5). Nevertheless, this “myocardial wall sign” can be not recognizable in those cases characterized by mild cardiac amyloid load or may be obscured, in late phase imaging, by bone uptake. In such cases, SPECT imaging may correctly identify myocardial uptake.

Fig. 5
figure 5

Early (soft tissue) phase scan (a) and late (bone) phase scan (b) images in a patient harboring ATTRv amyloidosis (E89Q mutation) with evidence of cardiac radiotracer accumulation on SPECT images (c). Both early and late phase scans clearly show myocardial radiotracer accumulation (myocardial wall sign)

A limitation of our study may be represented by its retrospective nature; however, a relatively large monocentric cohort of patients affected by a rare disease has been included and only one series of scans and SPECT has been evaluated when a patient had been studied at different time points during the course of disease. Moreover, acquisition of early and late phase scintigraphic images plus late phase thorax SPECT images was the standard methodology for suspect cardiac amyloidosis in our Centre. Another limitation of our study is the lack of early phase SPECT imaging that could help in obtaining a full comparison between early phase and late phase imaging performances; indeed, it is expected to have a better performance of tomographic findings [3, 5, 25] even in early phase imaging. Nevertheless, we clearly demonstrated that visual analysis of early phase whole body scan using [99mTc]Tc-DPD is accurate to diagnose cardiac amyloidosis.

In conclusion, although further studies are needed to fully compare early phase and late phase imaging, visual analysis of early phase whole body imaging using [99mTc]Tc-DPD is promising in diagnosing cardiac amyloidosis and may be considered satisfactory at least in frail patients with high cardiac burden of amyloid fibrils.