Abstract
Purpose
Loss of fixation and seal represent a key problem when undertaking endovascular repair of abdominal aortic aneurysms (AAA) with hyperangulated necks (HAN). This study assesses the outcomes following the use of adjunct endostapling to supplement proximal aorto-prosthetic fixation in patients who have AAAs with HAN.
Methods
A retrospective review of a prospective database of 42 patients with HAN (> 60°) who underwent endovascular aneurysm repair (EVAR) with supplementary endostapling was undertaken. Primary outcomes assessed were: change in post-EVAR neck angulation at first post-procedure scan, freedom from type 1 endoleaks, migration and reintervention for proximal seal complications. Secondary parameters included assessment for neck dilatation, sac size changes and EndoAnchor distribution patterns.
Results
In total, 42 patients underwent EVAR between 2013 and 2019. There was one 30-day mortality resulting in 41 patients (34 male, 7 females aged 76.8 ± 8.9 years)) being analysed; 251 EndoAnchors were deployed in total, averaging 6 ± 2 per patient; 38 such cases were primary deployments. Neck angulation was 76.9 ± 14 degrees pre-EVAR and 50.2 ± 14.5 degrees post-procedure (p < .001, paired T test). Mean follow-up time was 18.5 (95% CI 13.3–23.9) months. One patient had persistent type Ia endoleak, successfully banded. There was 6.8 ± 10.2 mm sac size reduction (p < .001, paired T test). There were no other neck-related reinterventions, despite continued neck dilatation (3.2 ± 3.7 mm, p < .001, paired T test).
Conclusion
This study suggests successful EVAR with adjunct endostapling for AAA with hyperangulated necks, with significant sac shrinkage and low rates of endoleaks, migration and reinterventions. More data are needed to consider influencing current instructions for use.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
Hostile anatomical conditions at endovascular aneurysm repair (EVAR) for abdominal aortic aneurysms (AAAs) can result in unsuitability for conventional infrarenal EVAR or compromise against device instructions for use (IFU), resulting in a non-durable result. Endostapling is a recognised technique for reinforcing fixation and seal [1], with recent terminology such as EndoSuture(d) Aneurysm Repair (ESAR) [2, 3] now added to the lexicon. This collaborative study specifically examines hostile neck anatomy in the context of hyperangulated necks, typically defined in the literature as a neck angle exceeding 60° [4], and whether using targeted adjunct fixation with the Heli-FX EndoAnchor (EA) system (Medtronic Ltd., Minneapolis, USA) at infrarenal EVAR for AAAs with hyperangulated necks can achieve acceptable sealing augmentation and prevent complications such as type Ia endoleaks and migration, the recognised typical complications in this scenario.
Materials and Methods
A retrospective analysis of a prospective database of 42 patients who underwent planned EVAR with supplementary endostapling for AAAs with hyperangulated necks (herein specifically referring to the β-angle) [5], defined as > 60°, was undertaken in two academic vascular centres with endostapling experience at EVAR and TEVAR. This includes one patient with Marfan syndrome. Patients underwent endostapling in a radial clockface fashion as typical (Fig. 1), but also supplementary columnar endostapling along the outer curve of the neck, to counter the pulling away forces that would contribute to loss of seal. These were done either on a lateral basis for a laterally curving outer curve (Fig. 2A), or anteriorly for necks that were hyperangulated along a sagittal plane (Fig. 2B). Data collected included device details, patient demographics and neck anatomical characteristics which are presented in Table 1. Primary outcomes assessed were change in neck angulation at first post-procedure scan, freedom from type Ia endoleakage and migration (defined as > 5 mm caudal displacement), and also reinterventions for endoleak-related complications. Other parameters assessed include neck diameter changes, sac size changes and EndoAnchor deployment patterns. Follow-up imaging was undertaken on a standardised chronological protocol in the early post-procedure phase (< 6 weeks), at 6 months and then annually thereafter. Neck diameter was assessed as the average inner-to-inner aortic diameter from the first postoperative scan as in other such analyses [6] and then for comparison against similar parameters in the most recent CTA or, where abdominal radiographs were used, against the outer-to-outer diameter of the aortic endoprosthesis as these two designated measurement parameters would be the closest, i.e. as the device is deployed into the inner aortic wall. Measurements were all done by a single assessor at each centre (AC, ARV) in order to minimise inter-operator variability, using local Picture Archiving and Communications System (PACS) software.
Data were populated in Microsoft Excel for statistical analysis within Minitab 19 for Windows. Categorical variables are presented as counts and percentages. Continuous variables are presented as mean/median ± standard deviation/range. Matched parametric data were analysed using paired t tests. The threshold of statistical significance was p < 0.05. Relationships between independent and linked outcome variables were compared using linear regression modelling. A distribution analysis of the variable follow-up using the Kaplan–Meier method was undertaken to identify the numbers at risk at each annual follow-up interval, generating a life table-based time series plot to present freedom from type I endoleak and migration.
Results
In total, 42 patients underwent EVAR between 2013 and 2019 for infrarenal AAAs that had a neck angle > 60°. Mean follow-up period was 18.5 (95% CI 13.0–23.9) months. There was one 30-day mortality from graft infection, resulting in 41 patients (34 male, 7 females aged 76.8 ± 8.9 years) being available for analysis. Devices used included Zenith Flex (Cook Aortic Interventions, Bloomington, USA; n = 9), Zenith LP (Cook Aortic Interventions; n = 2) Alpha (Cook Aortic Interventions; n = 5), Endurant (Medtronic; n = 19), Excluder C3 (WL Gore & Associates, AZ, USA; n = 6). Cuffs were used for proximal extension in 7 patients, primarily when the main aortic body was felt to have dropped (n = 3), or secondarily to treat a type Ia endoleak (n = 4). In the primary group, there was no residual type Ia endoleak after cuff extension, whilst there was one failure to achieve proximal seal in the latter, described below. The mean length of stay for the entire group was 5 ± 4 days.
A total of 251 EndoAnchors were implanted at a mean 6 ± 2 per patient. Choice of numbers deployed was arbitrary (though deployment positions were planned) based on operator choice, with regression modelling showing no correlation with neck angulation even when analysed in hindsight (p = 0.99, R2 = 0). Most of the EAs were positioned (according to clock face in descending order) at 3 o’clock (n = 47), 9 o’clock (n = 46), 2 (n = 32), 4 (n = 24), 8 (n = 26), 10 (n = 26) 1:30 (n = 5), 4:30 (n = 9), 10:30 (n = 19), 12 and 6 o’clock positions (n = 8), 7:30 (n = 1), each (Fig. 3). A total of 35 such cases were primary and 6 for secondary indications, namely type Ia endoleak. Of the latter, endostapling was successful in 5, with 1 patient needing supplementary open aortic neck banding. Mean neck length was 19.18 ± 11.99 mm. There were no EndoAnchor-related complications. Pre-EVAR neck angles were 76.9 ± 14 degrees (typically classed as ‘severe’ [7]), reducing to 50.2 ± 14.5 degrees post-procedure (p < 0.001, paired T test; Fig. 4). Resumption of the hyperangulated state was not noted to occur throughout the follow-up period in any patient.
One patient had persistent type Ia endoleak despite an endostapled cuff that required open surgical banding which was successful. Freedom from type 1a endoleak and migration at any time are therefore 97.5% and 100%, respectively (Fig. 5). Anatomical trends in the AAAs are represented in Fig. 6A, B; sac size shrinkage occurred in 31 (75.6%) of patients, with a mean overall reduction of 6.8 ± 10 mm (p = 0.001; paired T test). Neck diameters indicated a trend towards continued dilatation (mean increase 3.2 ± 3.7 mm, p = 0.001; paired T test), representing a median neck dilatation rate of 0.11 mm/month (IQR 0.3) equating to around 1.3 mm per year (Fig. 6).
The salient results including complications are summarised in Table 2.
Discussion
Most aortic prostheses are limited to deployment into an AAA with an infrarenal neck angle of < 60°, with a high risk of seal and migration-related complications beyond this [8]. Suprarenal hyperangulation is usually relevant in terms of device selection but is not really germane to sealing aspects as such, particularly where EndoAnchors are used. First-generation devices were felt to be at higher risk of migration [8], but even with modern devices licensed for use up to 90° a 3% risk of migration has been identified [9]. Currently, the only devices that have an IFU that accepts a 60°–90° neck angulation are the Aorfix device (Lombard Medical, Didcot, UK) [10] and the more recent Conformable C3 device (WL Gore & Associates) [11].
The Heli-FX EndoAnchor System has been devised for usage with several endograft systems [12, 13]. The initial promise of this adjunct system [14] seems to be borne out by the good midterm results of the ongoing ANCHOR registry, including in the context of HAN [15]. Other endovascular anchoring devices that were in development include the EndoRefix system (Lombard Medical) that is not available for use as clinical trials have been suspended [16]. There are no device-related contraindications for EndoAnchor usage, except with fragile PTFE-based devices such as the Powerlink system (Endologix, Irvine, USA). Furthermore, there are no neck angulation-related contraindications, and thus, adjunct endostapling for AAAs with HAN constitutes standard practice in our institutions. The number of EndoAnchors employed per patient is in keeping with recommendations for optimal fixation [17, 18].
IFUs for most EVAR devices have remained both ill-defined and static [13]—with fleeting references to endostapling in the recent European Guidelines for management of AAAs [19]. The only recent radical changes to IFU for EVAR are where the acceptable neck length for EVAR with primary endostapling has been changed to 4–10 mm for the Endurant device [2, 20]. It therefore seems device manufacturers have not been responsive overall to the presence of adjunct fixation technology even though it is firstly not very new [1] and secondly has proven fixation characteristics in terms of high resistive pullout forces [17] with corroborative data affirming the scope for reducing migration and endoleakage.
High sac regression rates with endostapling as indicated in other studies [21, 22] have been reflected in our results. A higher number of lateral EndoAnchors have been implanted (i.e. at 9 and 3 o’clock), linking to the higher incidence of lateral angulation in this series. Mechanical fixation of aortic neck to the endograft may resist post-EVAR effects such as longitudinal shrinkage, endograft shortening and stent-graft displacement secondary to proximal external compression [8]. Though all devices using radial force for proximal fixation seem to cause neck dilatation [23], the suggestion that endostapling has a protective effect is also borne out by our own results, which suggests some neck expansion albeit slower than published studies [24]. Primary endostapling also reduces the need for adjunct techniques such as bending stiff wires prior to deployment (given that the effect is lost immediately with removal of such wires) [8] or use of reinforcing balloon-expandable stents, such as the Palmaz XL (Cordis Corporation, a Johnson & Johnson company, FL, USA) [8].
Endoleak and/or migration rate in EVAR for AAAs with hyperangulated necks has been described [7]. Meta-analyses do not specifically analyse HAN-related adjunct procedures, but combine it under an umbrella of hostile neck conditions [25]. More specific allusions come from the EUROSTAR registry [4] where HAN is linked to neck dilatation, proximal type I endoleakage and reintervention rates. Historical series report complications as high as 70% where neck angle is > 60 degrees [26], but these are for older generations of devices and at that time without considering the possible role of EndoAnchors. Other criteria that have typically affected device migration—where this has been specifically studied—include short neck length (beside device choice) but not neck diameter, with migration rates as high as 70% at 4 years [6]. Relative to this study, we used the stricter threshold for migration i.e. ≥ 5 mm as opposed to the more relaxed threshold of 10 mm [6, 8].
Limitations
This study thus for the first time analyses the scope of adjunct endostapling at EVAR in AAAs with HAN, though this is limited by being a small series, with only isolated reports published prior [27,28,29]. Other limitations also include the lack of long-term follow-up for all patients beyond five years. However, these include the first usage in this scenario in a patient with Marfan syndrome who has remained free from both type Ia endoleakage and migrations at > 5 years. A perceived limitation may be the potential confounding effect of mixed devices using either supra- or infrarenal fixation, but this has not been shown to make a difference [30], and was thus not analysed. In addition, as all EndoAnchor deployment positions were planned, these would be in areas lacking in thrombus or calcification as per IFU, and we therefore feel these neck aspects do not confound the outcomes. In fact, it has been suggested that calcification is protective against neck dilatation [23]. We could not compare this to a control group without EndoAnchors, as most operators will choose now to either excessively oversize devices out of IFU (potentially accelerating the risk of late neck dilatation [23], an approach that we discourage), or opt for super-complex options like fenestrated/ branched/ chimney (F/B/Ch)EVAR; however, endostapled EVAR may offer equivalent safety in this context to ChEVAR at least [31]. This is despite the misconception that FEVAR is a ‘solution’ for HAN anatomy, whereas the IFU limits use of fenestrated devices such as the Zenith ZFEN (Cook Aortic Interventions) to a neck angle < 45° [32]. Also, given the nature of Ch-EVAR, there is no related IFU.
We accept the combined effect of the individual contribution of endografts ± cuffs, which have their own inherent stiffness and unpredictable tissue incorporation [7], but EndoAnchors certainly constrain the aortic tissue [33] on to the endograft fabric and prevent loss of apposition along the outer curve, which is the most prone area for failure to seal. The fact that there were 6 secondary interventions for type Ia endoleak who have remained endoleak-free is a possible indicator that primary usage could have been undertaken [34], but these patients were referred from elsewhere.
Conclusion
In conclusion, this study indicates the scope for a multiplanar approach to endostapling when dealing with hyperangulated neck anatomy at EVAR, specifically additional linear endostapling along a second line of fixation along the outer curve of the deployed device. Such adjunct techniques at EVAR to treat AAAs with hyperangulated necks may prevent migration and proximal endoleakage in the midterm at least and reduce the severe neck angulation to a more moderate one. Larger studies with longer duration of follow-up are needed to gauge the robustness of this approach, with a view to considering IFU changes to include primary endostapling in this scenario. This then also needs more uptake of primary endostapling by endovascular operators.
References
Deaton DH. Improving proximal fixation and seal with the HeliFx Aortic EndoAnchor. Semin Vasc Surg. 2012;25(4):187–92.
Arko FR 3rd, Stanley GA, Pearce BJ, Henretta JP, Fugate MW, Mehta M, et al. Endosuture aneurysm repair in patients treated with Endurant II/IIs in conjunction with Heli-FX EndoAnchor implants for short-neck abdominal aortic aneurysm. J Vasc Surg. 2019;70(3):732–40.
Reyes Valdivia A, Chaudhuri A. In Search of the optimal endosutured aneurysm repair. J Endovasc Ther. 2019;26(6):888–9.
Hobo R, Kievit J, Leurs LJ, Buth J, Collaborators E. Influence of severe infrarenal aortic neck angulation on complications at the proximal neck following endovascular AAA repair: a EUROSTAR study. J Endovasc Ther. 2007;14(1):1–11.
Lee KM, Choi SY, Kim MU, Lee DY, Kim KA, Park S. Effects of anatomical characteristics as factors in abdominal aortic aneurysm rupture: CT aortography analysis. Medicine (Baltimore). 2017;96(25):e7236.
Tonnessen BH, Sternbergh WC III, Money SR. Mid- and long-term device migration after endovascular abdominal aortic aneurysm repair: a comparison of AneuRx and Zenith endografts. J Vasc Surg. 2005;42(3):392–401.
de Almeida M, Yoshida WB, Hafner L, dos Santos JH, Souza BF, Bueno FF, et al. Factors involved in the migration of endoprosthesis in patients undergoing endovascular aneurysm repair. J Vasc Bras. 2010;9(2):61–71.
Ghouri M, Krajcer Z. Endoluminal abdominal aortic aneurysm repair: the latest advances in prevention of distal endograft migration and type 1 endoleak. Tex Heart Inst J. 2010;37(1):19–24.
Malas MB, Hicks CW, Jordan WD Jr, Hodgson KJ, Mills JL Sr, Makaroun MS, et al. Five-year outcomes of the PYTHAGORAS U.S. clinical trial of the Aorfix endograft for endovascular aneurysm repair in patients with highly angulated aortic necks. J Vasc Surg. 2017;65(6):1598–607.
Malas MB, Jordan WD, Cooper MA, Qazi U, Beck AW, Belkin M, et al. Performance of the Aorfix endograft in severely angulated proximal necks in the PYTHAGORAS United States clinical trial. J Vasc Surg. 2015;62(5):1108–17.
Rhee R, Peterson B, Moore E, Lepore M, Oderich G. Initial human experience with the GORE EXCLUDER Conformable AAA Endoprosthesis. J Vasc Surg Cases Innov Tech. 2019;5(3):319–22.
https://global.medtronic.com/xg-en/healthcare-professionals/products/cardiovascular/aortic-stent-grafts/heli-fx-endoanchor.html. Medtronic; [updated 29.05.2019].
Chaudhuri A. Commentary: is an ounce of EndoAnchors worth more than many pounds for reintervention? J Endovasc Ther. 2019;26(1):101–4.
Chen J, Stavropoulos SW. Management of Endoleaks. Semin Intervent Radiol. 2015;32(3):259–64.
Jordan WD Jr, de Vries JP, Ouriel K, Mehta M, Varnagy D, Moore WM Jr, et al. Midterm outcome of EndoAnchors for the prevention of endoleak and stent-graft migration in patients with challenging proximal aortic neck anatomy. J Endovasc Ther. 2015;22(2):163–70.
Evaluation of EndoRefix Endovascular Delivery System and Staple (EndoRefix) 2012. https://clinicaltrials.gov/ct2/show/NCT00668681.
Melas N, Perdikides T, Saratzis A, Saratzis N, Kiskinis D, Deaton DH. Helical EndoStaples enhance endograft fixation in an experimental model using human cadaveric aortas. J Vasc Surg. 2012;55(6):1726–33.
Goudeketting SR, Vermeulen JJM, van Noort K, Te Riet OGSG, Kuipers H, Slump CH, et al. Effect of different EndoAnchor configurations on aortic endograft displacement resistance: an experimental study. J Endovasc Ther. 2019;26:1–10.
Wanhainen A, Verzini F, Van Herzeele I, Allaire E, Bown M, Cohnert T, et al. Editor's choice - European Society for Vascular Surgery (ESVS) 2019 clinical practice guidelines on the management of abdominal Aorto-iliac artery aneurysms. Eur J Vasc Endovasc Surg. 2019;57(1):8–93.
https://global.medtronic.com/xg-en/healthcare-professionals/products/cardiovascular/aortic-stent-grafts/endurantii/indications-safety-warnings.html. [Updated 29.05.2019].
Jordan WD Jr, Mehta M, Ouriel K, Arko FR, Varnagy D, Joye J, et al. One-year results of the ANCHOR trial of EndoAnchors for the prevention and treatment of aortic neck complications after endovascular aneurysm repair. Vascular. 2016;24(2):177–86.
Muhs BE, Jordan W, Ouriel K, Rajaee S, de Vries JP. Matched cohort comparison of endovascular abdominal aortic aneurysm repair with and without EndoAnchors. J Vasc Surg. 2018;67(6):1699–707.
Oberhuber A, Schwarz A, Hoffmann M, Klass O, Schelzig H, Orend KH, et al. Influence of fixation mechanism on changes of the supra- and infrarenal segment of the aorta after endovascular treatment of infrarenal aortic aneurysm. Zentralbl Chir. 2010;135(5):433–7.
Tassiopoulos AK, Monastiriotis S, Jordan WD, Muhs BE, Ouriel K, De Vries JP. Predictors of early aortic neck dilatation after endovascular aneurysm repair with EndoAnchors. J Vasc Surg. 2017;66(1):45–52.
Antoniou GA, Georgiadis GS, Antoniou SA, Kuhan G, Murray D. A meta-analysis of outcomes of endovascular abdominal aortic aneurysm repair in patients with hostile and friendly neck anatomy. J Vasc Surg. 2013;57(2):527–38.
Sternbergh WC 3rd, Carter G, York JW, Yoselevitz M, Money SR. Aortic neck angulation predicts adverse outcome with endovascular abdominal aortic aneurysm repair. J Vasc Surg. 2002;35(3):482–6.
Chaudhuri A. EVAR for AAA in Marfan’s syndrome: the first endostapled case. Cardiovasc Intervent Radiol. 2015;38(S239):P142.
Chaudhuri A. Endostapling can constrain a hyperangulated neck and successfully treat a proximal type I endoleak after EVAR. Eur J Vasc Endovasc Surg. 2016;51(5):681.
Morgan-Bates K, Chaudhuri A. Deliberate stent-graft extension and endotacking straightens the hyperangulated aortic neck during endovascular aneurysm repair. Cardiovasc Intervent Radiol. 2014;S294:P-135.
Hager ES, Cho JS, Makaroun MS, Park SC, Chaer R, Marone L, et al. Endografts with suprarenal fixation do not perform better than those with infrarenal fixation in the treatment of patients with short straight proximal aortic necks. J Vasc Surg. 2012;55(5):1242–6.
Calarese AW, Stanley GA, Ballast JK, Briggs CS, Yammine H, Nussbaum T, et al. A comparison of endosuture aneurysm repair versus chimney stent grafts for treatment of short-neck infrarenal abdominal aortic aneurysms. J Vasc Surg. 2019;70:E147.
Zenith Fenestrated AAA Endovascular Graft. Instructions for Use. https://www.cookmedical.com/data/IFU_PDF/IFU-FU_V3.PDF.
Saratzis A, Modarai B. Commentary: understanding the role of EndoAnchors in infrarenal endovascular aneurysm repair. J Endovasc Ther. 2019;26(5):714–6.
Schlosser FJV, de Vries J, Chaudhuri A. Is it time to insert EndoAnchors into routine EVAR? Eur J Vasc Endovasc Surg. 2017;53(4):458–9.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors declare that they have no conflict of interest.
Ethical Approval
For this type of study, formal consent is not required.
Informed Consent
For this type of study, informed consent is not required.
Consent for Publication
For this type of study, consent for publication is not required (there is no patient-identifying data).
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
About this article
Cite this article
Chaudhuri, A., Kim, HK. & Valdivia, A.R. Improved Midterm Outcomes Using Standard Devices and EndoAnchors for Endovascular Repair of Abdominal Aortic Aneurysms with Hyperangulated Necks. Cardiovasc Intervent Radiol 43, 971–980 (2020). https://doi.org/10.1007/s00270-020-02488-4
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00270-020-02488-4