Introduction

Myocardial blood flow (MBF) measurements obtained using positron emission tomography (PET) have been documented to add prognostic value over the assessment of regional perfusion defects [1,2,3,4] and have recently been shown to predict benefit from revascularization [5]. 82Rb is the most widely used PET perfusion tracer as the tracer permits myocardial perfusion imaging in centers without a cyclotron [6,7,8]. Three recent joint statements from North America (ASNC and SNMMI Cardiovascular Council [1, 8]) and Europe (EANM and ESC [9]) suggest injections of 1100–1500 MBq 82Rb for cardiac PET. The recommendations on injection doses were based on experiences obtained for the 2D-PET systems, where the use of physical septa’s significantly reduced the scatter and limited the prompt-gamma effect [10, 11]. Injection of high doses of 82Rb, however, poses a risk of saturation for 3D PET systems during the bolus injection, where highly concentrated tracer activities might be observed locally with consequential increased scatter and deadtime in the PET crystals [12, 13]. The increased deadtime in the crystals translates into apparently reduced activity distributions in the PET sinograms [14] and, thus, reduced arterial input activities observed in the input functions as a consequence [14]. The changes in the arterial input functions affect the area under the input function curve and consequently leads to incorrect MBF estimations. The frequency and effects introduced by saturation of the PET system following high dose injections of 82Rb and its impact on the quantification of MBF are not yet well established.

The aims of this study were to evaluate the frequency and relative impact of saturation of 3D PET systems using a full-dose (FD) and half-dose (HfD) imaging protocol and to evaluate the differences due to using a HfD protocol versus FD for clinical assessment of MBF.

Methods

Study subjects

Twenty patients between the ages of 45 and 75 years old (11 male) underwent clinically indicated 82Rb PET/CT myocardial perfusion imaging. All patients were recruited between February and August 2016, thereby covering different timepoints in the generator lifespan. The scans were acquired using current recommendations provided by the North American [1, 8] and European societies [9], respectively; employing target doses of 1500 MBq or equivalent to the delivery capabilities of the dose-cart for the FD scans, while the HfD scans targeted injections of 750 MBq. The inclusion criteria were patients aged ≥ 50 and ≤ 75 and no known cardiac disease, while exclusion criteria included known pregnancies and contraindications to 82Rb. This study was performed in line with the principles of the Declaration of Helsinki. Approval was granted by the Cedars-Sinai Medical Center Institutional Review Board and written informed consent was obtained from all patients.

Full-dose and half-dose 82Rb PET/CT imaging protocol

Heart rate and blood pressure were recorded before the initial 82Rb injection in resting state. All datasets were acquired using the Bracco dose cart, with target deliveries of 1500 MBq (FD) and 750 MBq (HfD) in bolus form. PET imaging was performed using a whole-body 3D PET/CT scanner (Siemens Biograph-64 TruePoint PET/CT with the True V) [15]. For attenuation correction purposes, a low-dose helical CT (100 kV, pitch 1.5, 11 mAs, duration 3.4 s) was acquired in end-expiration breath-hold before each rest 82Rb PET acquisition. Alignment of PET emission data and the CT attenuation correction maps were confirmed before PET-image reconstruction and corrected when necessary by the imaging technologist [15]. Figure 1 shows the imaging protocol, which consisted of two 82Rb PET scans (6 min) acquired in the list-mode format. Data from both acquisitions were reconstructed into dynamic image-series consisting of 16 frames (12 × 10 s, 2 × 30 s, 1 × 60 s, and 1 × 120 s) using the vendor iterative method (Fourier rebinning + 2-dimensional attenuation-weighted ordered-subsets expectation-maximization) with 2 iterations, 8 subsets, and an 8-mm Gaussian postprocessing filter [16].

Fig. 1
figure 1

Study protocol. Myocardial blood flow 82Rb PET acquired with a full-dose and half-dose injection were compared. CT attenuation correction scans (CTAC) were obtained before each PET scan

Motion correction

All reconstructed dynamic datasets were evaluated using motion correction, given the growing evidence that this is necessary to ensure high reproducibility for both rest and stress scans [17, 18]. In the current study, we applied motion correction as previously described [18], relying on manual adjustments (rigid translations) of the myocardium relative to the contour derived from the last 2-min frame. Motion correction was performed on all dynamic frames where realignment of the myocardial segmentations was judged to be necessary. Frames with identified shifts were corrected using rigid translations. All corrections were performed by an experienced nuclear medicine specialist (OM).

Quantification of MBF

Analyses of global and regional MBF assessments were obtained from the two dynamic rest imaging series—full-dose (FD) and half-dose (HfD)—with the dedicated software (QPET, Cedars-Sinai Medical Center, Los Angeles, California) [16, 19] using transaxial PET images.

MBF measurements used a standard single tissue-compartment model, as described by Lortie et al. [6], employing input functions extracted from a 2-cm3 volume of interest located in the left ventricular lumen at the center of the mitral valve plane [20]. The global myocardium (global) as well as regional assessments for the left anterior descending (LAD), left circumflex (LCX), and right coronary artery (RCA) were reported.

Evaluation of PET system saturation

Saturation effects were evaluated visually from the input functions obtained in the left ventricle and sinogram data by two experienced readers. PET system saturation effects were suspected upon identification of plateau-like peaks occurring at the first-pass of the tracer-bolus (Fig. 2). Scans displaying these input-function features underwent further evaluation of the general image quality of the dynamic frames under suspicion of saturation. Identification of increased noise (speckle) (Fig. 2A and B ), led to the evaluation of single slice rebinned sinograms (compressed 3D-sinograms) [21]. Saturation was declared after visually detecting changes in the count rates (obtained in the central slices of the single-slice rebinned sinogram), corresponding to the location of the myocardium (Fig. 2A) and only when a consensus was obtained by two experienced readers (ML and OM).

Fig. 2
figure 2

Identification of PET system saturation. Identification of plateau of the aortic input function during the first-pass bolus initiated further inspections of the data. Four different cases are presented (A, B, C, D, one with PET system saturation (A), 2 where PET system saturation was suspected but ruled out through evaluations of PET images and sinograms (B and C) and finally a scan without any indication of saturation which serves a reference (D). Speckle noise in the reconstructions suggests potential saturation of the PET system (A, B), while homogenous uptake is common for non-saturated datasets. Evaluations of single-slice rebinned sinograms (3D compressed sinograms) were initiated upon identification of both: plateau findings for the input function, and increased speckle noise. PET system saturation was declared upon visually reduced uptake patterns in the sinograms (A), while normal sinogram uptake patterns excluded the saturation of the PET system (B). Case B suggests issues with count statistics for image reconstruction, while case C suggests that the first-pass arrival of the bolus happens in between two consecutive frames. For reference, case D shows an acquisition at the first-pass of the bolus (input function peak) with no signs of PET system saturation. Red x indicates PET systems saturation, whereas green checkmarks indicate images without PET system saturation

Reproducibility

Test–retest reproducibility of the MBF assessments was compared using the coefficient of variation (CV) using the root mean square method [22]:

$$ CV=100\times \sqrt{\frac{\sum {\left(\frac{d}{m}\right)}^2}{2n.}} $$

In the equation, d represents the difference between the two paired measurements, m the mean of the paired measurements, and n the sample size.

The data were subdivided into two analyses: one included all datasets and the other excluded datasets where saturation was identified in either rest scans (FD or HfD, respectively).

Half-dose to full-dose correction

To provide a method for serial comparison of MBFFD and MBFHfD values a bias correction was obtained using datasets without saturation effects.

Statistical analysis

Continuous variables are presented as mean and standard deviations. Categorical variables are presented as absolute numbers with percentages. The correlation between MBF values was shown as correlation plots with Pearson’s R correlation coefficients and Bland–Altman plots [23]. A comparison of average MBF values was tested using a paired t test. Evaluations of the within protocol (FD/HfD) variances were validated using Pitman–Morgan analyses. p values < 0.05 were considered statistically significant. Statistical calculations were carried out using MedCalc (Version 19.0.7).

Results

Global and regional MBF from the two rest acquisitions were successfully processed for all 20 patients and 60 vessels. The clinical characteristics of the study population are shown in Table 1.

Table 1 Participant Characteristics

Comparison between MBF-FD and MBF-HfD

Injected doses of 82Rb were 1247 ± 196 MBq and 662 ± 115 MBq for FD and HfD acquisitions, respectively, resulting in an average 47% dose reduction for the HfD scans (Supplementary Fig. 1). Of note, the bodyweight-corrected dose was equivalent to 14.2 ± 2.4 MBq/kg and 7.5 ± 1.2 MBq/kg for the FD and HfD scans, respectively. To establish accurate MBF estimates, motion correction was applied to one of the scans (N = 7, of which 5 had corrections for FD) or both scans (N = 13). Evaluation of the MBF obtained from the FD (MBFFD) and HfD (MBFHfD) protocols revealed a bias of 0.09 ml/g/min for the two imaging protocols (global MBFFD = 1.03 ± 0.29, global MBFHfD = 0.94 ± 0.22, p = 0.003) (Table 2). Evaluations of the within-group variances between the FD and HfD protocols revealed significantly increased variation for both global and regional MBF assessments using the FD protocol (Pitman–Morgan analyses, Table 2). Strong correlations (Pearson’s R) ( 0.88) were reported between the MBFFD and MBFHfD measurements for global and regional assessments (Table 2).

Table 2 MBFs obtained from full-dose and half-dose scans without removal of saturated datasets (N = 20)

Saturation effect

A total of 4 (20%) FD scans compared to 0 (0%) HfD scans were found to be affected by PET system saturation effects. The patients in which PET system saturation was identified had 82Rb injections of 1006, 1099, 1217, and 1491 MBq, respectively, thus, within the current joint statement guidelines [1, 8, 9]. The weight of the four patients were 80, 107, 92, and 107 kg with corresponding body mass indexes (BMI) of 25.8, 35.7, 30.7, and 36.8, respectively. The bodyweight-corrected doses for the 4 saturated scans were 12.7, 10.3, 13.3, and 14.0 MBq/kg and, thus, below the average weight–corrected dose observed for the FD protocol. Two case examples of saturation effects in the FD scans are shown in Fig. 3 and Supplementary Fig. 2, respectively. For all 4 patients showing signs of saturation effects, the MBF obtained using the FD protocol was increased when compared with the MBF obtained for the HfD protocol (22% difference, MBFFD = 1.25 ± 0.22, MBFHfD = 1.02 ± 0.18, p = 0.001). In comparison, the bias observed for the FD and HfD protocols was remarkably lower for the remaining 16 patients not showing evidence of PET system saturation, though still significantly different (5% difference, MBFFD = 0.97 ± 0.28, MBFHfD = 0.92 ± 0.24, p = 0.019).

Fig. 3
figure 3

Case example from our study for a 70-year-old female. Sinograms (A, D), dynamic curves of input function (B, E) and polar maps (C, F) derived from the full-dose (FD) and half-dose (HfD) injection are displayed. Compressed sinograms (summed in the axial direction) present the count profiles for a 5 s duration at the time of the peak (B, E). Sinogram representation for FD scan shows reduced counts in the center part indicating saturation of the scanner (A; yellow arrows). On the contrary, the HfD scan does not have the same saturation effect in the central part of the sinogram (D). For the FD injection, the dynamic curve shows the sign of the plateau (B; red arrow) which increases the MBF for the FD scan with more than 0.3 ml/g/min compared to the HfD scan (C, F)

Although PET system saturation effects were identified in 4 patients, there was no correlation between injected dose and MBF (Supplementary Fig. 3).

Following the exclusion of patients with evidence of PET system saturation, there were excellent correlations between MBF obtained from FD and HfD protocols, with Pearson’s R ≥ 0.93 (Table 3). The removal of the saturated datasets ensured the harmonization of the MBF assessments in the LAD and LCX regions (Table 3, paired t test). In addition, harmonization of the within-group variation (variances observed for MBF assessments using the FD/HfD protocols) were observed for LCX and RCA (Pitman–Morgan analyses, Table 3). Despite a general harmonization in MBF assessments, some differences were observed for global measurements and in the LAD territory (Table 3).

Table 3 MBFFD and MBFHfD using only datasets without saturation for the full-dose scans (N = 16)

Half-dose to full-dose serial scan correction

A bias between the full-dose MBF (MBFFD) and the half-dose MBF (MBFHfD) assessments was observed. From the MBF assessments obtained using the two protocols, the following linear correction of the data was obtained:

$$ \mathrm{Corrected}\ {\mathrm{MBF}}_{\mathrm{HfD}}:1.09\ast {\mathrm{MBF}}_{\mathrm{HfD}}-0.03 $$

Reproducibility

Bland–Altman plots of global and regional MBF revealed good agreement between the MBF for the FD and the HfD imaging protocols following the exclusion of 4 saturated datasets (Fig. 4). Relative reductions in a bias of up to 37.5% were observed between the MBFFD and MBFHfD following removal of patients showing signs of PET system saturation (Bias all datasets; global = − 0.08 ml/g/min, regional − 0.09 ml/g/min; bias non-saturated datasets only; global = − 0.05 ml/g/min, regional = − 0.07 ml/g/min) (Fig. 4). Removal of the saturated datasets led to minor improvements in the test–retest reproducibility, which were found excellent for all datasets with coefficient of variation (CV) ≤ 10.4% (Table 3, Fig. 4).

Fig. 4
figure 4

Bland–Altman plots of global and regional MBF before and after the exclusion of saturated datasets. Reduced variations are observed following the removal of the saturated datasets. Mean differences are represented with the green line, while the yellow lines mark the 95% confidence limits. MBF, myocardial blood flow [ml/g/min], w/o = without; FD, full-dose; HfD, half-dose; CV, coefficient of variation

Comparison of HfD/FD MBF assessments following bias correction (serial scan correction)

Applying the linear correction (Corrected MBFHfD: 1.09*MBFHfD-0.03) to the MBFHfD assessments, harmonized both global and regional flows to the ones obtained using the FD imaging protocol (Global MBF; MBFFD = 0.97 ± 0.28, MBFHfD = 0.92 ± 0.23 (paired t test p = 0.02), corrected MBFHfD: 0.98 ± 0.25, (paired t test: p = 0.77)) (Table 4). Additionally, no significant differences were observed for variance in the respective regional MBF values obtained from FD and HfD protocols (all p values ≥ 0.17, Pitman–Morgan analyses) (Table 4). Furthermore, test–retest variability following corrections (CV: global = 6.4%, regional (combined) = 9.1%, LAD = 11.5%, LCX = 7.7%, RCA = 7.4%, all p ≥ 0.46) were reported concordant with the non-corrected datasets (Table 3).

Table 4 MBF assessments before and after bias correction of MBFHfD data

Discussion

This study evaluated the frequency and impact of PET system saturation using the current clinical guidelines for 82Rb MBF evaluations with 3D PET systems [1, 8], as well as the potential use of a half-dose imaging protocol. The main finding of this study was a relatively high frequency of saturation in PET systems when using the current guideline proposed injection doses (20% of the scans), which was reduced to 0% when halving the injected dose. The significant differences observed for the MBF assessments using the FD and HfD protocol are likely due to generally increased crystal deadtimes observed for the FD acquisitions.

In the current study, three important observations were noted following dose reduction. First, the saturation of the PET system led to a systematic increase in the MBF values for the affected scans, resulting in an increased discrepancy of the MBFFD and MBFHfD when compared with patients without saturation effects. Secondly, MBFFD and MBFHfD obtained for datasets without saturation effects demonstrated high linearity with Pearson’s R ≥ 0.93, although bias was observed for the two protocols. The third observation was high test–retest repeatability measures with no statistical differences of either MBFs obtained for the FD and HfD protocols following a simple linear correction (test for MBF measurements: all p ≥ 0.77 (paired t test), variances (per territory and global MBF assessments for the FD and HfD protocols, respectively) all p ≥ 0.17 (Pitman–Morgan test).

The FD acquisition protocol was based on the most recent joint positions in North America (between ASNC and SNMMI Cardiovascular Council [1, 8]) and in Europe (between EANM and ESC [9]), which both suggest target doses of 1100–1500 MBq when using lutetium orthosilicate injected (LSO) 3D PET systems. These recommendations for relatively high doses of 82Rb are in part due to the technical drawbacks of 82Rb (short half-life and a low extraction fraction), while still facilitate reliable MBF measurements [24]. Although these recommendations are proposed for the 3D systems, the use of high-dose injections arose from the 2D PET systems, where increased doses were necessary to obtain sufficient image-quality. The high-dose protocols were originally required because of the use of physical septa, which significantly reduced the sensitivity of the systems when compared with the modern 3D systems [12, 25]. The increased sensitivity of the modern 3D PET systems, however, increases the risks of saturating when high tracer-activities are injected in bolus-forms such as in 82Rb scans [13, 14].

To avoid the risk of PET system saturation, alternative protocols using a slow-infusion of the tracer activity over 30 s have been proposed [26,27,28]. Unfortunately, the use of the slow-infusion protocols is not possible in the commercial products, where the commercial dose-carts target a constant flow through the generator, with tracer administration in bolus form as a consequence [26]. Another approach, focusing more on optimizing the image-quality, proposed to employ weight-specific injection profiles for a range of different PET systems [12]. The study showed the feasibility of employing patient-weight specific injection profiles, thus, avoiding saturation effects of the PET system. However, using such protocols increase the time spent on scan preparations for each scan where the dose-delivery cart needs calibration between the PET system and delivery carts. This can introduce nonconformities of the dose-deliveries with a bias of up to 68% being reported [12]. In a study by Tout et al. [14], the authors recommended using injection-profiles (bolus-injection) of 1110 MBq for the assessments of MBF to avoid PET system saturation. Although this injection profile might reduce the risks of PET system saturation, two of the four scans in the current study where PET system saturation was reported had injected doses of 1006 and 1099 MBq and, thus, below the recommended 1110 MBq injection profiles.

In the current study, we focused on the potential use of a fixed dose across all patients (HfD scan protocol) and its impact on MBF evaluations. Using the HfD protocol, no dependencies between injected activities and MBF estimates were observed (Supplementary Fig. 3). Furthermore, the test–retest repeatability measures obtained for the FD and HfD protocols were comparable to a recent study evaluating short-term reproducibility of MBF obtained using an FD/FD protocol from our center [18]). A general bias between the FD and HfD protocol for the MBFFD and MBFHfD was observed (0.05 ml/g/min), suggesting a small reduction for the MBFHfD evaluations with lower dose. Of note, while the 0.05 ml/g/min bias in the test–retest study appears somewhat increased in comparison to a recent study from our center evaluating short-term test–retest reproducibility using an FD protocol (0.02 ml/g/min bias) [18], both biases are not significantly > 0. This is potentially due to the relatively small sample in both studies (20 patients in the current cohort and 21 patients in our FD/FD test–retest reproducibility study [18]). However, it is also possible that these differences are due to the normal variability within one patient. Nevertheless, to facilitate serial comparisons between MBF evaluations obtained using the HfD and the standard HD protocols, we developed a bias correction factor compensating for the general reduction in MBF observed for HfD protocol. The correction factor was established using only datasets without any visible saturation effects to exclude potential bias in the corrections from the saturated data. Following the corrections, no significant differences between the MBFFD and corrected MBFHfD were observed. Based on these findings, no detectable differences in the bias for the two studies, and the FD scan to HfD repeatability coefficient being comparable with FD/FD scan–rescan repeatability [18], we propose the use of HfD protocol instead of the FD protocols for MBF evaluations to minimize the risks of PET system saturation.

One limitation of this study is the saturation effect was not evaluated during the stress scan. It was not possible to perform a second stress scan in the same imaging session with repeated pharmacological stress agent injection within this time. Additionally, the physiological variations due to the different stress responses would be large and potentially biased during repeated stress scan. Nevertheless, the injected doses are similar at stress and rest so similar saturation effects could be expected. Our study has several additional limitations. The limitation of the proposed HfD to FD correction technique is that it was tested and validated within the same cohort. A prospective trial will be required to fully validate this technique. Another limitation is the number of participants evaluated in this study, which is relatively small, rendering it difficult to determine whether the small bias between FD and HfD MBF estimates being caused by a potential type 2 error or by differences in the count-rates caused by the differences in the two injection profiles. However, despite the sample size, a significant correlation between the MBFHfD and MBFFD was identified. Another limitation of the study was that blood pressure and heart rate was only obtained before the initial scan (FD) such that MBF estimates cannot be corrected for changes in the RPP. The PET/CT system used for the clinical evaluations were based on older technology. The use of solid-state PET systems might reduce the frequency and risk of saturation because of the generally reduced dead-time and reduced bias in the detected activities [29]. However, most current PET systems still rely on photomultiplier tube technology where saturation effects still pose a risk of PET system saturation, with frequencies similar to the ones presented in the current study [14]. Finally, a limiting factor of this study is that all saturation effects were evaluated retrospectively using sinograms, without the access to the saturation profiles of the individual detectors. We also did not study the quality of static or gated perfusion images; the quality of static perfusion image, however, is not affected by saturation due to the fast decay of the tracer and distribution in the body [30]. Nevertheless, HfD static or gated perfusion images are created from the same PET acquisition and may potentially suffer from increased noise. The effect of dose reduction on static and gated perfusion images will need to be separately evaluated in comparison to full-dose studies before half-dose 82Rb are deployed clinically; however, these effects could be evaluated retrospectively from list-mode data in larger populations with external reference standards.

Conclusions

In this study, we report that up to 20% (4 of 20 scans) of all rest 82Rb cardiac PET scans with injected doses of 925–1480 MBq suffered from saturation effects in the first-generation 3D PET/CT systems; whereas repeated acquisitions using a dose of 480–900 MBq in the same patients did not result in saturation.