Abstract
Transcatheter patent ductus arteriosus (PDA) closure (TCPC) utilizing transthoracic echocardiogram (TTE) as the sole imaging guide could simplify care. This single-center study compares PDA dimensions obtained from the TTE and angiogram images of patients who underwent attempted TCPC at Stead Family Children’s Hospital from 10/01/2019 to 10/31/2020. Blinded investigators measured these dimensions solely for this study and had no impact on clinical care. Also, a hypothetical Piccolo device size was chosen based on the TTE dimensions and another on the angiographic dimensions, and then the correlation was analyzed. Sixty-two patients underwent TCPC attempts. TTE tends to overestimate the PDA narrowest dimension and underestimate the PDA length and aortic end dimension. Linear regression analysis revealed a weak correlation between the length and aortic diameter (R = 0.37 and 0.21, respectively). A modest correlation was observed for the smallest dimension without color Doppler (R = 0.57) and with color Doppler, which was utilized when needed (R = 0.6). Bland–Altman analysis revealed a smaller mean difference between the TTE and angiogram measurements of the narrowest diameter without color Doppler (0.4 mm) and with color Doppler (used as needed) (0.4 mm). However, the mean difference is larger for the aortic end (− 1.64 mm) and the length (− 1.73 mm). TTE accurately predicted the Piccolo device size in 43 (72%) patients and overestimated the size in 17 (28%) patients to the next size. Our findings should be verified with further studies, and additional development of protocols is needed to use TTE to guide TCPC without fluoroscopy.
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Avoid common mistakes on your manuscript.
Introduction
Patent ductus arteriosus (PDA) is commonly encountered in premature infants [1]. When medical therapy fails to close a hemodynamically significant PDA, surgical or transcatheter PDA closure (TCPC) becomes indicated [2,3,4,5,6,7,8,9,10]. Surgical closure was favored over TCPC for definitive PDA closure [2]. The FDA approved the first device designed for PDA closure in premature infants in 2019 [Amplatzer Piccolo Occluder (Abbott, Plymouth, MN)]. Since then, the experience with TCPC has rapidly accumulated. Many centers have adopted TCPC in premature infants as the primary definitive PDA closure option [11, 12]. Fluoroscopy and angiograms are utilized to size the PDA and guide catheter and wire manipulation. An intraoperative transthoracic echocardiogram (TTE) is employed to judge the adequacy of closure and device position. There is growing interest in performing TCPC under exclusive TTE guidance, hoping to minimize the risks involved with patient transport and exposure to radiation and contrast [12,13,14,15,16]. Yet studies regarding whether TTE can be reliably trusted to choose the appropriate device size are lacking.
In this study, we aim to investigate whether the PDA dimensions measured via TTE correlate with those measured angiographically.
Methods
This is a single-center study utilizing the TTE and angiographic images of premature infants who underwent attempted TCPC at the Stead Family Children’s Hospital/ University of Iowa from October 1, 2019, to October 31, 2021. The study was approved by the University of Iowa institutional review board. Patients were included when intraprocedural TTE and angiogram were performed in premature infants with hemodynamically significant PDA. Patients were excluded when TTE and angiogram images were unavailable or measurements could not be performed. Our center’s TTE criteria for TCPC closure have been previously described [11]. Blinded investigators measured essential PDA dimensions, outlined below, which determine Piccolo device size for closure solely for this study and did not impact clinical decisions. Piccolo device selection was based on a sizing chart shown in Table 1. We utilized this table to predict which device size would be chosen based on TTE and compared those predictions to the device chosen based on angiogram measurements for each patient. The angiographic measurements are considered the true ductal measurements. The Piccolo device sizes placed in most patients were 2 mm long and 3 or 4 wide. For this study, the chosen hypothetical devices were based on the patient's weight and the PDA’s narrowest measured dimension by each modality for each patient’s dataset based on the PDA sizing chart (Table 1).
Measuring the PDA Dimensions by TTE
All intraprocedural TTE images were obtained via either Philips iE33 or EPIQ 7 (Koninklijke Philips, Amsterdam, Netherlands). PDA dimensions were measured from the intraprocedural TTE obtained just before the start of the cardiac catheterization using the high left parasternal short-axis window, known as the “ductal view.” Two-dimensional (2D) images without color Doppler were utilized to obtain the measurements when feasible. The aortic end (ampulla) (ED1) width was measured at the insertion of the PDA with the aorta. The narrowest PDA dimension (ED2) was defined as the narrowest PDA segment width, which is usually located at or close to the connection of the PDA with the main pulmonary artery (MPA). The PDA length (EL) was measured as a straight line extending from ED1 to the pulmonic end of the PDA (Fig. 1). All measurements were obtained during systole.
As the smallest PDA dimension is a key in choosing the PDA device size, dimensions were measured on a 2D image with color Doppler when the quality of the 2D image without color Doppler did not allow satisfactory quantification.
TTE dimensions were measured by two investigators (UG) and (AB) from the same clip. We aimed to exclude poor image quality as a confounding factor in weakening the interclass correlation (ICC). As such, the second investigator (AB) measured the PDA dimensions only when the quality of the images allowed the first investigator to measure all dimensions. Both investigators were blinded to each other’s measurements.
Measuring the PDA Dimension by Angiogram
All angiograms were performed using the Siemens Artis Zee Biplane fluoroscopy system (Siemens Healthcare AG, Erlangen, Germany).
Patent ductus arteriosus dimensions were measured using the lateral projection angiogram by two blinded investigators (KC and BMN), who worked separately and measured the dimensions from the same clip. They were blinded to patient identification, TTE measurements, and each other's measurements.
The measured angiographic dimensions were as follows: AD1 was defined as the aortic end dimension, AD2 as the narrowest PDA dimension (again usually at or near the insertion into the MPA), and AL was the PDA length measured as a straight line from AD1 to the pulmonic end of the PDA (Fig. 2a). All measurements were obtained in systole when the PDA was distended.
Data Collection
Demographic data collected included procedure age, corrected age and weight, gestational age, and birth weight. The TTE data collected included PDA dimensions (ED1, ED2, and EL), aortic and LPA flow velocity, and the presence and severity of tricuspid valve regurgitation. One-hour post-procedural and last follow-up TTE data included the adequacy of device position, presence, and severity of residual shunt, aortic arch and LPA flow velocities, and severity of tricuspid valve regurgitation.
Angiogram data included PDA type and PDA dimensions (AD1, AD2, and AL).
All investigators were blinded to the measurements of the others as well as the size of the device and the TCPC outcomes.
Outcomes
The primary outcomes were comparing the angiographic and TTE PDA dimensions and whether TTE dimensions would predict the device selection based on the angiographic dimensions using the sizing chart (Table 1).
Secondary outcomes were demographics, PDA type, procedure time, fluoroscopy time, and radiation dose.
Statistical Analysis
Data were analyzed statistically using IBM SPSS Statistics ([ver. 28.0.11(14)], IBM, Armonk, NY). Categorical data were reported as counts and percentages. Continuous data were reported as either mean and standard deviation (SD) or median with interquartile range (IQR). The normality of the data was tested using Shapiro–Wilk test. Paired t, Pearson correlation, and Bland–Altman plot tests were utilized to compare and assess the agreement between the TTE and angiogram PDA dimensions. The interclass correlation coefficient (ICC) was utilized to assess the interobserver variability between blinded investigators. The ICC was calculated using Kappa Statistics (two-way mixed effect model). p value of < 0.05 was chosen as a cut-off of the statistically significant results.
A Brief TCPC Procedure Description
An abbreviated baseline TTE is performed in the cardiac catheterization lab. The following are assessed: the presence and the size of PDA (Fig. 1), flow in the LPA and the descending aorta, cardiac systolic function, and the presence and degree of tricuspid valve regurgitation.
An angiogram is obtained inside the PDA using 1 ml of contrast. Angiographic PDA dimensions are measured (aortic end, narrowest diameter, and length) to aid in choosing the PDA device size (Fig. 2a). The device type [Piccolo vs KA micro plug (KA Medical, Minneapolis, MN)] is chosen based on the discretion of the interventional cardiologist. The device is then deployed inside the PDA. Angiogram in the MPA (Fig. 2b and c) and TTE are performed. If the device is not entirely intraductal or obstructing the aorta or the LPA, the device is recaptured and redeployed. After confirming adequate positioning and no more than a trivial residual shunt, the device is released, and the delivery system is removed. A detailed TTE is performed after releasing the device.
Results
Demographics
Sixty-two patients were referred for TCPC. Two patients were excluded due to the inability to obtain an angiogram: one patient had occluded bilateral femoral veins, and the other developed inferior vena cava dissection at the beginning of the procedure. Both patients underwent uneventful surgical closure.
The median birth weight and gestational age were 687 (576–915) g and 25.1 (23.7–26.6) weeks, respectively. The median age, corrected age, and weight at the time of TCPC were 4.3 (3.4–6) weeks, 29 (28.1–30.9) weeks, and 1200 (1000–1600) g, respectively.
Primary Outcomes (Comparison of the TTE and Angiogram Measurements)
-
TTE findings
The quality of intraprocedural TTE without color allowed for satisfactory PDA measurement in 48 (80%) patients for the aortic end (ED1), 45 (75%) patients for the narrowest dimension (ED2), and 46 (77%) patients for the length (EL). Using color Dopler images, the narrowest PDA dimension was adequately measured in all patients.
The interclass correlation coefficient (ICC), evaluating the agreement between blinded investigators, revealed at best modest agreement for the narrowest (ED2) and the aortic end (ED1) dimensions with an ICC of 0.63 and 0.69, respectively. There was weak agreement for the PDA length (EL) with an ICC of 0.19.
-
Angiogram findings
The quality of the obtained angiogram allowed for measuring all PDA dimensions in all patients.
The interclass correlation coefficient, gauging the agreement between blinded investigators, revealed a strong correlation for all PDA dimensions with an ICC of 0.86 for the aortic end (AD1), 0.89 for the length (AL), and 0.98 for the narrowest diameter (AD2).
-
TTE and angiogram correlation
Using the paired sample t test, the PDA dimensions measured by TTE and angiogram were statistically different. It was noticed that the TTE tends to overestimate the smallest PDA dimension (ED2 vs. AD2) and underestimate the PDA length (EL vs. AL) and aortic end (ED1 vs. AD1) dimensions (Table 2).
Using a linear regression analysis (Pearson correlation), it was noticed that there was a modest correlation with the narrowest PDA dimension (ED2 vs. AD2) measured without color Doppler (R = 0.57, P < 0.001) as well with color Doppler when needed (R = 0.6, P < 0.001). The degree of correlation was weak for the length (EL vs. AL) (R = 0.21, p = 0.017) and aortic end diameter (ED1 vs. AD1) (R = 0.37, P = 0.01) measurements (Fig. 3).
Bland–Altman analysis revealed a smaller mean difference between the TTE and angiogram measurements of the pulmonary end (ED2 vs. AD2) without color Doppler (0.4 mm) as well as with color Doppler (when used as needed) (0.4 mm). However, the mean difference is larger for the aortic end (ED1 vs. AD1) (− 1.64 mm) and the length (EL vs. AL) (− 1.73 mm) (Fig. 4 and Table 3).
When Table 1 was used to test the ability of TTE to predict the accurate Piccolo device size, it was noticed that the TTE was able to predict the Piccolo device in 43 (72%) patients and overestimated the device to the next size in 17 (28%) patients. TTE did not underestimate the device size in any patient. Figure 5 summarizes device selections based on TTE and angiogram measurements when Table 1 is utilized.
Secondary Outcomes
-
PDA type and used devices
The type of PDA was as follows. Fetal type (hockey stick appearance) in 39 (65%), conical (type A) in 10 (16.7%), tubular (type C) in 8 (13.3%), and complex (type D) in 3 (5%).
The majority of the used devices were Piccolo devices in 58 (97%) patients. KA microplug was used in 2 patients (3%). The devices were 3/2 (width/length) mm Picclo device in 27 (45%), 4/2 mm Piccolo in 26 (43.3%), 4/4 mm Piccolo in 5 (8.3%), 5/4 mm Piccolo in 1 (1.6%), 5 mm KA micro plug in 1 (1.6%), and 3 mm KA microplug in 1 (1.6%) patients.
Discussion
Since the Amplatzer Piccolo Occluder (Abbott, Plymouth, MN) has gained Food and Drug Administration (FDA) approval for use in premature infants weighing > 700 g, experience with transcatheter PDA device closure has been rapidly increasing [3, 17]. In many centers, including ours, TCPC was adopted as the primary method for definitive PDA closure. When compared with surgical ligation, TCPC has many advantages. It is less invasive, with high success and low complication rates [9, 11, 12, 18].
To mitigate the potential risks associated with transportation, contrast, and radiation exposure, it has been an ambition to perform the procedure solely under TTE guidance at the bedside in the NICU [12,13,14,15,16]. Nevertheless, the ability of the TTE to predict the angiogram PDA dimensions remains uncertain as the evidence is confined to a single-center study [12].
Contrary to Paudel and colleagues’ findings which we cannot explain [12], this study did not find a strong correlation between TTE and angiogram PDA dimension measurements. Rather, the correlation was moderate for the pulmonary end measured on 2D without color Doppler images (R = 0.57) and on 2D with color Doppler images when needed (R = 0.6) (ED2 vs. AD2). In addition, the correlation was weak for the length (EL vs. AL) (R = 0.21) and aortic end (ED1 vs. AD1) (R = 0.37) measurements. The aortic end (ED1 and AD1) is not considered when choosing the device size. Furthermore, the length is not an important factor if a short device (2 mm in length) is intended to be placed [9, 19, 20]. It is worth mentioning that each Piccolo device size (width) is available in 3 different lengths (2 mm, 4 mm, and 6 mm) [17]. It is widely accepted that using long Picclo devices, i.e., 4 and 6 mm, should be restricted to cases where the length of the PDA allows for full intraductal placement of the device [16, 19]. In contrast, the short, i.e., 2 mm in length piccolo device can be utilized in short as well as long PDAs as well [12, 19]. The authors prefer to use mainly 2 mm long piccolo devices regardless of the PDA length. We believe that the short devices are easier to place entirely intraductal, minimizing the risk of iatrogenic aortic and/or LPA obstruction [9, 19, 20]. Therefore, the narrowest PDA diameter is pivotal in choosing the device size for a given weight. Although a short device could theoretically increase the risk of embolization, this was not observed in this study, nor in studies where only short devices were used [9, 20].
Traditionally, a 2D TTE image without color is used to measure the PDA dimensions [12]. However, due to the poor acoustic window in some patients, the PDA dimensions could not be adequately measured in all patients without utilizing color Doppler. The poor acoustic windows in some premature infants could be attributed to hyperinflated lungs and chest vibration, as most of our patients undergo the procedure on high-frequency jet ventilation [11]. Measuring the narrowest PDA dimension using colored images allowed assessing this dimension in all patients without decreasing the correlation strength with angiogram measurement (R = 0.60). The authors speculate that the TTE ICC correlations are not strong, at least partially due to two factors. First, despite measuring the PDA dimensions is performed during systole, it is difficult to measure at the exact timing within systole. Therefore, although the investigators measured the dimensions from the same clip, the measurements at the same point in the systole could not be guaranteed. Second, as the absolute values of the PDA dimensions are small in premature infants, even a minor measurement error could result in statistically significant different numbers.
The results of this study suggest that the TTE tends to overestimate the narrowest PDA dimension (ED2 vs. AD2) and underestimate the PDA length (EL vs. AL) and aortic end (ED1 vs.AD1) measurements (Tables 2 and 3). Although the TTE tends to overestimate the narrowest dimension, this is likely clinically insignificant as the TTE could predict the accurate size in most patients [43 (72%) patients] and overestimated the device's size in the remainder of the patients (17 patients = 28%). TTE did not underestimate the size of the narrowest segment of any patient. Theoretically, this oversizing will not compromise the device's stability but could increase the risk of LPA and aortic obstruction. TTE is useful to assess LPA and aortic flows after positioning but before release. This allows for device position adjustment if needed.
The discrepancy between TTE and catheterization measurements could be attributed to catheter-induced ductal spasm. The TTE’s overestimation of the narrowest PDA dimension could be related to the timing of TTE performance before any catheter-induced vasospasm. Furthermore, angiograms could underestimate this dimension due to the potential of catheter-induced ductal spasms. It is noteworthy that in one patient, severe ductal spasm was noticed on angiogram after the PDA was crossed uneventfully (Fig. 6). Moreover, it was observed that in 13 patients, the smallest diameter of the PDA according to the angiogram was less than 1.33 mm (which is the size of the 4-Fr AG catheter). At the same time, the TTE showed the narrowest diameter was at least 1.5 mm. In older children, the PDA is susceptible to spasm if crossed in an antegrade fashion. Therefore, a baseline aortic angiogram is obtained at the beginning of the procedure to measure the true (not spasmed) PDA dimensions. Additionally, after deploying the device, residual flow across the PDA and the device's position are assessed by repeating aortic angiogram through the arterial access [21]. However, in premature infants, it is widely accepted to perform the TCPC via a single venous sheath and avoid obtaining arterial access due to the high incidence of access-related complications [1, 3, 4, 8, 19, 22, 23]. Hence, the PDA is crossed from the pulmonic end (prograde fashion) before obtaining the angiogram. This could induce ductal spasm, which may result in underestimating the true PDA dimension and lead to a smaller device selection. The incidence of catheter-induced ductal spasms is unknown, especially in small premature infants. Nevertheless, older children who were born prematurely are specifically at risk of ductal spasms during TCPC [24]. When the authors suspect ductal spasm, they rely on the TTE smallest dimension to choose the device.
We speculate that TTE can be relied on to choose the Piccolo device and assess the adequacy of positioning without needing an angiogram. One of its strengths was that the narrowest PDA dimension could be measured in all patients without under-sizing device selection. Also, TTE was able to assess the arch and LPA flow in all patients. However, whether the TTE can guide catheter manipulation and device deployment is yet to be answered. The procedure might still be done at the bedside using a portable fluoroscopy machine without needing an angiogram with the help of TTE.
Limitations
This is a single-center study with a limited cadre of patients. Larger multicenter studies might help confirm our findings across a wider population of patients. TTE requires an experienced operator to optimize each acoustic window to measure the PDA dimensions, especially when the windows are poor due to lung hyperinflation and chest wall vibration. This might be reflected in the weak to modest ICC statistics between the blinded investigators. This wider variability in measurements could require more than one clinician to take measurements during these procedures.
Catheter-induced ductal spasm could not be ruled out as a possible cause for the discrepancy in measuring the smallest PDA dimension between TTE and angiography. This question could be answered by conducting studies measuring TTE PDA dimensions before and after obtaining the angiogram. Moreover, this study was not designed to investigate whether TTE can be exclusively utilized to guide TCPC in premature infants.
Conclusion
PDA dimensions measured by TTE do not match those obtained by angiogram. Although the difference was statistically significant, the device selection was either matched or slightly oversized compared to the one chosen based on the angiogram. The safety of relying solely on TTE to choose the device has yet to be studied.
Data Availability
No datasets were generated or analysed during the current study.
References
Hamrick SEG, Sallmon H, Rose AT, Porras D, Shelton EL, Reese J et al (2020) Patent ductus arteriosus of the preterm infant. Pediatrics 146(5).
Gray DT, Fyler DC, Walker AM, Weinstein MC, Chalmers TC (1993) Clinical outcomes and costs of transcatheter as compared with surgical closure of patent ductus arteriosus. The Patient Ductus Arteriosus Closure Comparative Study Group. N Engl J Med 329(21):1517–1523
Sathanandam SK, Gutfinger D, O’Brien L, Forbes TJ, Gillespie MJ, Berman DP et al (2020) Amplatzer Piccolo occluder clinical trial for percutaneous closure of the patent ductus arteriosus in patients ≥700 grams. Catheter Cardiovasc Interv 96(6):1266–1276
Backes CH, Cheatham SL, Deyo GM, Leopold S, Ball MK, Smith CV et al (2016) Percutaneous patent ductus arteriosus (PDA) closure in very preterm infants: feasibility and complications. J Am Heart Assoc 5(2)
Agrawal H, Waller BR, Surendan S, Sathanandam S (2019) New patent ductus arteriosus closure devices and techniques. Interv Cardiol Clin 8(1):23–32
Baspinar O, Sahin DA, Sulu A, Irdem A, Gokaslan G, Sivasli E et al (2015) Transcatheter closure of patent ductus arteriosus in under 6 kg and premature infants. J Interv Cardiol 28(2):180–189
Roberts P, Adwani S, Archer N, Wilson N (2007) Catheter closure of the arterial duct in preterm infants. Arch Dis Child Fetal Neonatal Ed 92(4):F248–F250
Zahn EM, Peck D, Phillips A, Nevin P, Basaker K, Simmons C et al (2016) Transcatheter closure of patent ductus arteriosus in extremely premature newborns: early results and midterm follow-up. JACC Cardiovasc Interv 9(23):2429–2437
Heyden CM, El-Said HG, Moore JW, Guyon PW, Katheria AC, Ratnayaka K (2020) Early experience with the Micro Plug Set for preterm patent ductus arteriosus closure. Catheter Cardiovasc Interv 96(7):1439–1444
Francis E, Singhi AK, Lakshmivenkateshaiah S, Kumar RK (2010) Transcatheter occlusion of patent ductus arteriosus in pre-term infants. JACC Cardiovasc Interv 3(5):550–555
Shibbani K, Mohammad Nijres B, McLennan D, Bischoff AR, Giesinger R, McNamara PJ et al (2022) Feasibility, safety, and short-term outcomes of transcatheter patent ductus arteriosus closure in premature infants on high-frequency jet ventilation. J Am Heart Assoc 11(10):e025343
Paudel G, Johnson JN, Philip R, Tailor N, Fahnhorst S, Briceno-Medina M et al (2021) Echocardiographic versus angiographic measurement of the patent ductus arteriosus in extremely low birth weight infants and the utility of echo guidance for transcatheter closure. J Am Soc Echocardiogr 34(10):1086–1094
Pouldar TM, Wong R, Almeida-Jones M, Zahn E, Lubin L (2021) Bedside transcatheter patent ductus arteriosus device occlusion in an extremely low birth weight neonate: a novel approach in a high-risk population. Case Rep Anesthesiol 2021:4716997
Deraz S, Tawfik M, Kotit S, Elafifi A (2022) Transcatheter device closure of patent ductus arteriosus by exclusive venous access under echocardiographic guidance without angiography. Cardiol Young 32(1):55–59
Georgiev S, Tanase D, Eicken A, Peters J, Hörer J, Ewert P (2021) Transvenous, echocardiographically guided closure of persistent ductus arteriosus in 11 premature infants: a pilot study. JACC Cardiovasc Interv 14(7):814–816
Aw TC, Chan B, Singh Y (2023) Transport and anaesthesia consideration for transcatheter patent ductus arteriosus closure in premature infants. J Cardiovasc Dev Dis 10(9).
Rahde Bischoff A, Backes C, Weisz D, Mohammad Nijres B, McNamara P (2023) Interventional management of the patent ductus arteriosus. In: Kluckow M (ed) Neonatology questions and controversies: neonatal hemodynamics, 4th edn. Elsevier, Philadelphia, pp 346–364
Barry OM, Gudausky TM, Balzer DT, Bocks ML, Boe BA, Callahan R et al (2023) Safety and short-term outcomes for infants < 2.5 kg undergoing PDA device closure: a C3PO registry study. Pediatr Cardiol 44(6):1406–1413
Sathanandam S, Gutfinger D, Morray B, Berman D, Gillespie M, Forbes T et al (2021) Consensus guidelines for the prevention and management of periprocedural complications of transcatheter patent ductus arteriosus closure with the amplatzer piccolo occluder in extremely low birth weight infants. Pediatr Cardiol 42(6):1258–1274
Guyon P, Duster N, Katheria A, Heyden C, Griffin D, Steinbergs R et al (2022) Institutional trend in device selection for transcatheter PDA closure in premature infants. Pediatr Cardiol 43(8):1716–1722
Masura J, Walsh KP, Thanopoulous B, Chan C, Bass J, Goussous Y et al (1998) Catheter closure of moderate- to large-sized patent ductus arteriosus using the new Amplatzer duct occluder: immediate and short-term results. J Am Coll Cardiol 31(4):878–882
Alexander J, Yohannan T, Abutineh I, Agrawal V, Lloyd H, Zurakowski D et al (2016) Ultrasound-guided femoral arterial access in pediatric cardiac catheterizations: a prospective evaluation of the prevalence, risk factors, and mechanism for acute loss of arterial pulse. Catheter Cardiovasc Interv 88(7):1098–1107
Tadphale S, Yohannan T, Kauffmann T, Maller V, Agrawal V, Lloyd H et al (2020) Accessing femoral arteries less than 3 mm in diameter is associated with increased incidence of loss of pulse following cardiac catheterization in infants. Pediatr Cardiol 41(5):1058–1066
Batlivala SP, Glatz AC, Gillespie MJ, Dori Y, Rome JJ (2014) Ductal spasm during performance of transcatheter ductal occlusion. Catheter Cardiovasc Interv 83(5):762–767
Author information
Authors and Affiliations
Contributions
Bassel Mohammad Nijres measured the PDA dimensions, analyzed the data, created the figures and tables, and wrote the main manuscript. Mohammad Khallaf collected the data. Kaitlin Carr, Umang Gupta, and Adrianne Rahde Bischoff measured the PDA dimensions. Jimmy Windsor, Osamah Aldoss, and Patrick J McNamara critically revised the manuscripts. All authors reviewed the final version.
Corresponding author
Ethics declarations
Conflict of interests
The authors declare no competing interests.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Springer Nature or its licensor (e.g. a society or other partner) holds exclusive rights to this article under a publishing agreement with the author(s) or other rightsholder(s); author self-archiving of the accepted manuscript version of this article is solely governed by the terms of such publishing agreement and applicable law.
About this article
Cite this article
Mohammad Nijres, B., Khallaf, M., Rahde Bischoff, A. et al. Transcatheter Patent Ductus Arteriosus Closure in Premature Infants: Comparison of Echocardiogram and Angiogram Measurements. Pediatr Cardiol (2024). https://doi.org/10.1007/s00246-024-03620-8
Received:
Accepted:
Published:
DOI: https://doi.org/10.1007/s00246-024-03620-8