Abstract
Direct Fick method is considered a standard reference method for estimation of cardiac output. It relies on indirect calorimetry to measure oxygen consumption. This is important as only a minor measurement error in oxygen consumption can result in false estimation of cardiac output. A number of studies have shown that indirect calorimetry overestimates oxygen consumption in adults. The aim of this prospective single center observational method comparison study was to compare the determination of oxygen consumption by indirect calorimetry and reverse Fick method in pediatric patients. Forty-two children mean age 352 days (range 30 to 1303 days) and mean weight 7.1 kg (range 2.7–13.6 kg) undergoing corrective cardiac surgery were included in the study. The mean (standard deviation) oxygen consumption by reverse Fick method was 43.5 (16.2) ml/min and by indirect calorimetry 49.9 (18.8) ml/min (p < 0.001). Indirect calorimetry overestimated the reverse Fick oxygen consumption by 14.7%. Bias between methods was 6.5 (11.3) ml/min, limits of agreement (LOA) − 15.7 and 28.7 ml/min and percentage error of 47.7%. A significant bias and large percentage error indicates that the methods are not interchangeable. Indirect calorimetry and the direct Fick method should be used with caution as a reference method in cardiac output comparison studies in young children.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
More than a century ago the German physician Adolf Eugen Fick (1829–1901) hypothesized that it should be possible to estimate cardiac output if the arterio-mixed venous oxygen content difference and systemic oxygen consumption are known [1, 2]. Time and progress in medical science would prove Fick right, as the 1956 Nobel prize was awarded to the pioneers in cardiac catheterization who confirmed the Fick principle in humans [3, 4].
The so-called direct Fick method has since then been considered a standard reference method for the estimation of cardiac output [5, 6]. Direct Fick method determines cardiac output by dividing the oxygen consumption (\({\text{mV}}_{{{\text{O}}_{2} }}\)), with the difference between arterial (\({\text{Ca}}_{{{\text{O}}_{2} }}\)) and mixed venous oxygen content (\({\text{Cv}}_{{{\text{O}}_{2} }}\)) (Eq. 1).
Oxygen consumption (\({\text{mV}}_{{{\text{O}}_{2} }}\)) is measured with indirect calorimetry, but in recent years with increasing demand on precision and accuracy in comparison studies, there has been a growing concern regarding its technical limitations and possible errors.
In the reverse Fick method, oxygen consumption (\({\text{cV}}_{{{\text{O}}_{2} }}\)) can be calculated if cardiac output (Q) has been measured by some other means and both arterial (\({\text{Ca}}_{{{\text{O}}_{2} }}\)) and mixed venous blood gas (\({\text{Cv}}_{{{\text{O}}_{2} }}\)) values are known.
A number of studies have shown discrepancies between measured \({\text{mV}}_{{{\text{O}}_{2} }}\) by indirect calorimetry and calculated \({\text{cV}}_{{{\text{O}}_{2} }}\) by reverse Fick method in adults [7,8,9,10]. As the direct Fick method relies on measured oxygen consumption by indirect calorimetry, measurement errors could cause false estimations of cardiac output.
The hypothesis of the study is that there is not a significant difference in the estimation of oxygen consumption using indirect calorimetry versus the reverse Fick method. The aim of this study was to investigate if oxygen consumption was overestimated in a cohort of young children undergoing corrective cardiac surgery.
Methods
Patient and Data Collection
Forty-two patients undergoing congenital cardiac surgery were included in this study which was a part of a larger ongoing single center prospective method comparison study [11]. The study was approved and registered by the Ethics Committee of Lund University, Sweden (Dnr 2,013,636). Inclusion criteria included parental consent, elective open heart surgery, weight less than 15 kg, pulse oximetry saturation > 93%, inspired oxygen fraction (\({\text{Fi}}_{{{\text{O}}_{2} }}\)) 0.21–0.40, endotracheal intubation with a cuffed tube without leakage, tidal volumes of 7–8 ml/kg, sinus rhythm and echocardiography not showing signs of a significant intracardiac shunt after coming off the heart–lung bypass machine. Exclusion criteria included arrythmias and active bleeding after surgery.
Oxygen Consumption Measured by Indirect Calorimetry (\({\text{mV}}_{{{\text{O}}_{2} }}\))
Measured oxygen consumption (\({\text{mV}}_{{{\text{O}}_{2} }}\)) was obtained by indirect calorimetry using GE Healthcare Datex-Ohmeda S/5 Compact Anaesthesia Monitor (Datex Ohmeda, Inc., Madison, WI, USA) with a E-CAiOVX module, specific Pedilite + paediatric flow sensor and adjusted to paediatric mode [12, 13]. Measurements can be collected after a 5-min warm-up period, including an automatic calibration, after which the monitor can yield a continuous \(V_{{{\text{O}}_{2} }}\) reading in 5-min intervals. The S/5 monitor was calibrated annually by biomedical engineering staff against a Datex-Ohmeda calibration syringe (containing specific proportion of gases) as recommended by the manufacturer [14].
Oxygen Consumption Calculated by Reverse Fick Method (\({\text{cV}}_{{{\text{O}}_{2} }}\))
Calculated oxygen consumption (\({\text{cV}}_{{{\text{O}}_{2} }}\)) was calculated using the reverse Fick method according to Eq. 2.
Arterial and Mixed Venous Oxygen Content \({\text{(Ca}}_{{{\text{O}}_{2} }} ,\;{\text{Cv}}_{{{\text{O}}_{2} }} )\)
Blood gases were taken simultaneously. Arterial blood gas was taken from the radial artery. Mixed venous blood gas was taken from the pulmonary truncus by direct puncture by the surgeon. Blood gas analysis was done with the ABL800 Flex Radiometer (Radiometer AS, Brønshøj, Denmark).
Arterial oxygen content (\({\text{Ca}}_{{{\text{O}}_{2} }}\)) was calculated as:
\({\text{Sa}}_{{{\text{O}}_{2} }}\) is arterial oxygen saturation, Hba is arterial hemoglobin, \({\text{Pa}}_{{{\text{O}}_{2} }}\) is arterial partial pressure of oxygen.
Mixed venous oxygen content (\({\text{Cv}}_{{{\text{O}}_{2} }}\)) was calculated as:
\({\text{Smv}}_{{{\text{O}}_{2} }}\) is mixed venous oxygen saturation, Hbv is venous hemoglobin, \({\text{Pmv}}_{{{\text{O}}_{2} }}\) is mixed venous partial pressure of oxygen.
Cardiac Output Measurement
Cardiac output (Q) was obtained by the COstatus monitor device (Transonic Systems, Inc., Ithaca, NY, USA) which uses a specific single-use extracorporeal arteriovenous circuit (AV loop) connected to arterial and central venous catheters already in place and external ultrasound senors (attaced to the AV loop) to detect transcardiopulmonary blood dilution [15]. This dilution is estimated by the change in ultrasound velocity of blood after the transcardiopulmonary passage. The ultrasound velocity of blood is defined in a certain range (1.560–1.585 m/s) as there are fluctuations in total blood protein concentration (sum of proteins in the plasma and red cells) that influence velocity. As a small bolus (0.5–1.0 ml/kg) of normothermic saline (which has an ultrasound velocity of 1.530 m/s) is injected into the venous side of the AV loop, there will be a transient decrease in ultrasound velocity of the blood. This decrease is detected by an ultrasound sensor on the arterial side of AV loop. The COstatus device, creates a dilution curve based on this information, that is then displayed on a specific monitor and used to estimate the cardiac output using the Stewart–Hamilton indicator dilution principle [16,17,18]. The COstatus monitor device is considered minimally invasive as it only relies on available central venous and arterial catheters and has been shown to be accurate, precise and safe to use in number of pediatric studies [19,20,21]. Annual calibration was done by the manifacturer for the duration of the study. The device automatically performs a baseline determination of the ultrasound blood velocity in the AV loop before each measurement session.
Experimental Protocol
Anesthesia was achieved in all patients in the same manner, fentanyl (5 mcg/kg) and penthothal (5 mg/kg) for induction, pancuronium (0.2 mg/kg) to facilitate endotracheal intubation with a cuffed tube and isoflurane (0.5–1.0%) for maintainance. Stable ventilation was maintained with the Dräger Apollo anesthesia machine (Drägerwerk AG and Co., Lübeck, Germany) using tidal volumes of 7–8 ml/kg and \({\text{Fi}}_{{{\text{O}}_{2} }} < 0.{4}0.\) It was ensured that no leakage was present in the breathing circuit, by carefully monitoring that inspiratory and expiratory tidal volumes and that the volume curve during measurements. All children received peripheral arterial catheters in the radial artery (Neoflow 24 G < 5 kg patient and Venflow 22 G > 5 kg patient, BD Ltd., Wokingham, UK) and central venous catheters (Multicath triple lumen 6 cm, 4.5 F, Vygon Ltd., Swindon, UK) in the right internal jugular vein. The COstatus monitor was then connected to arterial- and central venous catheters in preparation for cardiac output measurements. After the surgical correction, a postoperative transesophageal echocardiography confirmed a biventricular circulation without shunts. The measurement session consisted of five consecutive cardiac output measurements with a simultaneous measurement of \({\text{mV}}_{{{\text{O}}_{2} }}\) by indirect calorimetry by S/5 Compact Anaesthesia Monitor. Blood gases were taken immediately after the last cardiac output measurement. \({\text{cV}}_{{{\text{O}}_{2} }}\) was then calculated using reverse Fick method as described previously.
Statistical Analysis
Data was registered in Windows Excel (Microsoft Corporation, Redmond, WA, USA) and the statistical analysis was done with Statistica version 13 (Dell, Inc., Tulsa, OK, USA).
All data is expressed as the mean (standard deviation) in the paper unless indicated otherwise.
A priory statistical power analysis was performed for sample size estimation based on data from an earlier study comparing oxygen consumption in adults [8]. In that study, the bias was 19 ml/min and the SD 20 ml/min resulting in an effect size of 0.95. Using the G-Power 3.1.9.2 software, with paired t test for difference of means, an α error value of 0.05 and power value of 0.90, it was estimated that a sample size of at least 11 subjects were needed. However, as we had available data from an ongoing cardiac output comparison study, we could include 42 patients in this study.
The differences between the methods were analyzed using the Student’s t test for paired samples.
Bland–Altman analysis was used to estimate the mean difference (bias) between \({\text{cV}}_{{{\text{O}}_{2} }}\) and \({\text{mV}}_{{{\text{O}}_{2} }}\) and plotted against the average of the comparison \(\left( {{\text{cV}}_{{{\text{O}}_{2} }} + {\text{mV}}_{{{\text{O}}_{2} }} } \right)/2.\) The 95% LOA were calculated as mean bias ± 1.96 * SD (standard deviation of the bias). LOA analysis was performed to determine if the two methods agreed sufficiently to be interchangeble. The Shapiro–Wilks test was used to confirm normal distribution of the difference between different methods. The percentage error was calculated according to Critchley and Critchley as 1.96 × SD of the bias/mean \(V_{{{\text{O}}_{2} }}\) of both methods × 100% [22].
A percentage error of < 30% for comparisons was defined as the criterion for interchangeability. A p value of < 0.05 was regarded as statistically significant.
Results
This study involved 42 children with a mean age of 352 (357) days (range 30 to 1303 days), a mean weight of 7.1 (3.0) kg (range 2.7 to 13.6 kg) and a mean body surface area (BSA) of 0.3 (0.11) m2 (range 0.10 to 0.50 m2).
The mean cardiac output values from 5 consecutive measurements were calculated from a total of 210 cardiac output measurements in 42 children and converted to \({\text{cV}}_{{{\text{O}}_{2} }}\) using the reverse Fick method. Indirect calorimetry measurements were simultaneously done in the same 42 children. Oxygen consumption values for both methods are presented in Table 1.
The mean (SD) \({\text{cV}}_{{{\text{O}}_{2} }}\) was 43.5 (16.2) ml/min and \({\text{mV}}_{{{\text{O}}_{2} }}\) 49.9 (18.8) ml/min (p < 0.0001). Levene ‘s test verified normal distribution within the samples (p > 0.05) and Shapiro–Wilks test for the difference between methods \(({\text{cV}}_{{{\text{O}}_{2} }} - {\text{mV}}_{{{\text{O}}_{2} }} )\) also indicated normal distribution (p = 0.92). Bias between \({\text{mV}}_{{{\text{O}}_{2} }}\) and \({\text{cV}}_{{{\text{O}}_{2} }}\) was 6.5 (11.3) ml/min (LOA of − 15.7 and 28.7 ml/min) corresponding to an overestimation by \({\text{mV}}_{{{\text{O}}_{2} }}\) of 14.7% (Fig. 1). The percentage error was 47.7%.
Discussion
We found that the oxygen consumption measured by indirect calorimetry overestimated oxygen consumption calculated by the reverse Fick method by 14.7% in this cohort of young children after corrective heart surgery. The result from this study is consistent with findings in previous studies in adult patients where the overestimation has ranged between + 13 to + 30% [7, 8]. This difference has been explained in part by lung tissue oxygen consumption, but systematic errors and flaws in measurements (e.g. circuit leaks, calibration problems, possible oxygen consumption in blood gas samples at higher temperatures, lower oxygen uptake during anaesthesia and hyperdynamic response after cardiopulmonary bypass) may have influenced the results [23, 24]. A bias of this magnitude corresponds to a significant overestimation in CO of 0.19 l/min (from mean cardiac output of 1.28 l/min) in our cohort. As the direct Fick method is dependent on indirect calorimetry for estimation of cardiac output, this study indicates that this method is impaired by a considerable measurement error. Method comparison studies based on a reference technique already affected by this level of inaccuracy could unjustifily disqualify promising new alternatives for cardiac output measurements. This is especially true and extremly important in method comparison studies in children where the margin for error is less [25,26,27].
Statistical standards have evolved in the last few decades. In the beginning of cardiac output comparison studies, simple linear correlation coefficient analysis was used to compare two methods, but has since changed to Bland–Altman analysis [28]. Bland–Altman analysis is now required in all cardiac output comparison studies and a percentage error of less than 30% has been recommended between different methods in adult studies to be considered interchangeable [22]. Cecconi also highlighted another important fundamental demand in his review article regarding method comparison studies [29]. He insisted that the precision of the reference method be known and reported as it is a critical factor in differentiating between different methods. The precision of the direct Fick method is almost never available or reported as the sample of the respiratory gases has to be done over a certain time period and usually done together with only a single set of arterial and mixed venous blood gases for the cardiac output calculation. These limitations regarding precision were already known and addressed in the 1960s when direct Fick method was emerging as a new alternative for estimating of cardiac output [4]. In a recent systematic review and meta-analysis on accuracy and precision of minimally invasive cardiac output monitors in children, the direct Fick method was used as a reference method in 3 out of 20 studies [30]. In none of these studies was the precision of direct Fick method reported and, therefore, it was only possible to estimate if the tested methods were interchangeble. However, it was not known whether it was the new tested technique or the reference technique that contributed to the inaccuracy or the percentage error.
The direct Fick method is impractical in children as it is highly invasive and limited to general anesthesia in the catheterization lab. Although, the direct Fick method can be used to determine cardiac output in an individual child, it has to be kept in mind that systemic oxygen consumption is affected by metabolic needs and general anesthesia. Direct Fick method assumes equilibration between the measured pulmonary oxygen uptake and systemic oxygen consumption, which may be inaccurate during critical illness, high pulmonary oxygen consumption, and uncoupling between oxygen delivery and oxygen consumption. In addition, most metabolic evaluations relying on oxygen uptake measurements become difficult to interpret when higher fractions of inspiratory oxygen are needed.
Conclusions
Direct Fick method is sensitive to minor changes or errors in saturation and oxygen consumption that can result in false cardiac output calculations. Indirect calorimetry seems to overestimate oxygen consumption in young children. Caution is advised when using the direct Fick method as a reference method in cardiac output method comparison studies in young children.
References
Fick A (1870) Über die messung des blutquantums in den Herzventrikeln. Sitz der der Physiol-Med Ges zu Würzbg 2:16
Guyton AC, Hall JE (1996) Textbook of medical physiology, 9th edn. Saunders Company, Philadelphia, W.B, p 251
Laszlo G (2004) Respiratory measurements of cardiac output: from elegant idea to useful test. J Appl Physiol 96:428–437
Selzer A, Sudrann RB (1958) Reliability of the determination of cardiac output in man by means of the Fick principle. Circ Res 6:485–490
de Boode WP (2010) Cardiac output monitoring in newborns. Early Hum Dev 86(3):143–148
Lemson J, Nusmeier A, van der Hoeven JG (2011) Advanced hemodynamic monitoring in critically ill children. Pediatrics 128(3):560–567
Bizouarn P, Blanloeil Y, Pinaud M (1995) Comparison between oxygen consumption calculated by Fick’s principle using a continuous thermodilution technique and measured by indirect calorimetry. Br J Anaesth 75:719–723
Peyton PJ, Robinson GJB (2005) Measured pulmonary oxygen consumption: difference between systemic oxygen uptake measured by the reverse Fick method and indirect calorimetry in cardiac surgery. Anaesthesia 60:146–150
Bizouarn P, Soulard D, Blanloeil Y et al (1992) Oxygen consumption after cardiac surgery—a comparison between calculation by Fick’s principle and measurement by indirect calorimetry. Intensive Care Med 18:206–209
Cheong KF, Lee TL (1997) Oxygen consumption—a comparison between calculation by Fick’s principle and measurement by indirect calorimetry. Med J Malays 52(1):70–75
Sigurdsson TS, Aronsson A, Lindberg L (2019) Extracorporeal arteriovenous ultrasound measurement of cardiac output in small children. Anesthesiology 130:712–718
Healthcare GE (2006) Datex-Ohmeda S-5 Anaesthetic Monitor-technical reference manual. General Electric Company, Helsinki
Eisenkraft JB (2011) Respiratory gas monitoring. In: Reich DL (ed) Monitoring in anesthesia and perioperative care. Cambridge University Press, New York, pp 164–165
Meyer R, Briassouli E, Briassoulis G et al (2008) Evaluation of the M-COVX metabolic monitor in mechanically ventilated adult patients. Eur e-J Clin Nutr Metab 3:e232–e239
Krivitski NM, Kislukhin VV, Thuramalla NV (2008) Theory and in vitro validation of a new extracorporeal arteriovenous loop approach for hemodynamic assessment in pediatric and neonatal intensive care unit patients. Pediatr Crit Care Med 9(4):423–428
Stewart GN (1897) Researches on the circulation time and on the influences which affect it. J Physiol 22(3):159–183
Hamilton WF, Moore JW, Kinsman JM et al (1932) Studies on the circulation. Am J Physiol 99:534–551
Reuter DA, Huang C, Edrich T et al (2010) Cardiac output monitoring using indicator-dilution techniques: basics, limits, and perspectives. Anesth Analg 110(3):799–811
Crittendon I III, Dreyer WJ, Decker JA et al (2012) Ultrasound dilution: an accurate means of determining cardiac output in children. Pediatr Crit Care Med 13(1):42–46
Lindberg L, Johansson S, Perez-de-Sa V (2014) Validation of an ultrasound dilution technology for cardiac output measurement and shunt detection in infants and children. Pediatr Crit Care Med 15(2):139–147
Saxena R, Krivitski N, Peacock K et al (2015) Accuracy of the transpulmonary ultrasound dilution method for detection of small anatomic shunts. J Clin Monit Comput 29(3):407–414
Critchley LA, Critchley JA (1999) A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques. J Clin Monit Comput 15:85–91
Kelman GR, Nunn JF (1966) Nomograms for correction of blood PO2, PCO2, pH and base excess for time and temperature. J Appl Physiol 21:1484
Oudemans-van Straaten HM, Scheffer GJ, Eysman L et al (1993) Oxygen consumption after cardiopulmonary bypass—implications of different measuring methods. Intensive Care Med 19:105–110
de Boode WP, Vrancken SL, Lemson J et al (2013) Gold standard must be solid gold. Intensive Care Med 39:1330–1331
Floh AA, La Rotta G, Wermelt JZ et al (2013) Validation of a new method base on ultrasound velocity dilution to measure cardiac output in paediatric patients. Intensive Care Med 39(5):926–933
Lemson J, Tibby SM (2016) Reliability of the ultrasound cardiac output monitor for pediatric patients. Pediatr Cardiol 37(3):618
Le Manach Y, Collins GS (2016) Disagreement between cardiac output measurement devices: which device is the gold standard? Br J Anaesth 116(4):451–453
Cecconi M, Rhodes A, Poloniecki J et al (2009) Bench-to-bedside review: the importance of the precision of the reference technique in method comparison studies with specific reference to the measurement of cardiac output. Crit Care 13(1):201
Suehiro K, Joosten A, Murphy LSL et al (2016) Accuracy and precision of minimally-invasive cardiac output monitoring in children: a systematic review and meta-analysis. J Clin Monit Comput 30(5):603–620
Acknowledgements
Open access funding provided by Lund University. Thanks to our Pediatric Cardiac Surgical Team at Lund Childrens Hospital (Skane, Sweden) for their help, patience and understanding.
Funding
This work was supported by a project grant from The Swedish Children Heart Association (Stockholm, Sweden) and The Anna and Edwin Bergers Private Foundation (Lidingö, Sweden).
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors have no conflict of interest.
Ethical Approval
This study was approved by the Ethics Committee of Lund University, Sweden, and is in accordance with the Ethical Standards laid down in the 1964 Declaration of Helsinki and its later amendments.
Informed Consent
Written informed consent was obtained from parents of all patients.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Electronic supplementary material
Below is the link to the electronic supplementary material.
Rights and permissions
Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
About this article
Cite this article
Sigurdsson, T.S., Lindberg, L. Indirect Calorimetry Overestimates Oxygen Consumption in Young Children: Caution is Advised Using Direct Fick Method as a Reference Method in Cardiac Output Comparison Studies. Pediatr Cardiol 41, 149–154 (2020). https://doi.org/10.1007/s00246-019-02238-5
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00246-019-02238-5