Introduction

In the native knee, anteroposterior stability is granted by the anterior cruciate ligament (ACL), primarily in early flexion and in conjunction with the posterior cruciate ligament (PCL) throughout knee motion. The ACL contributes to the so-called screw-home mechanism that is associated with anterior location of the femur on the tibia near full extension, while the PCL drives posterior femoral rollback in high flexion [2, 7, 32, 33]. Hence, in the native knee, both ACL and PCL play a major role in joint stability and kinematics.

Current efforts to restore native knee function following total knee arthroplasty (TKA) aim to retain both ACL and PCL through the use of bi-cruciate-retaining (BCR) implants or bi-unicompartmental procedures. In vivo, in vitro (cadaver), and computational studies have shown that ACL preservation provides more normal kinematics than contemporary ACL-sacrificing TKA [10, 19, 23, 35, 43].

Several studies investigating TKA implants found the tibiofemoral contact to be shifted to the posterior portion of the tibia when the ACL was not present [9, 34, 42]. This is in contrast to studies investigating ACL-retaining implants [35, 36]. In functional outcome studies, TKA patients with contemporary ACL-sacrificing implants have reported abnormal feeling knees after joint replacement, which may be linked to these kinematic impairments [8, 25].

However, in spite of potential benefits of ACL retention, ACL-retaining total knee arthroplasty is currently not part of standard TKA practice. This is due to perceived technical difficulties in retaining and balancing both ACL and PCL, lack of availability of such implants, and clinical contraindications including the absence of a functional ACL at surgery. Patients undergoing TKA often present with an absent or nonfunctional ACL at the time of surgery due to the progression of arthritis or due to prior trauma. Incidence of a functional ACL at the time of TKA surgery is reported to range from 25 to 86% of the patients [13, 15]. Concerns have also been raised about the tibial baseplate design changes required to accommodate the ACL, specifically reduction in surface available for implant fixation and the strength of the tibial baseplate [26, 29]. Further, these designs create a tibial bone island around the ACL attachment, which could fracture [20], particularly in the presence of osteoporotic bone or due to improper ligament balancing resulting in increased ACL tension [28].

The hypothesis of this study was that an ACL-substituting tibial implant designed with an anterior post mechanism to replace the ACL function while allowing retention of the PCL (ASCR: ACL-substituting, cruciate (PCL)-retaining) could improve the kinematics of contemporary cruciate-retaining (CR) designs. The primary purpose of such an ACL-substituting implant would be to locate the femur anteriorly in extension like in native knees, while allowing the PCL to guide knee motion at higher flexion angles. An implant incorporating this new concept of an ACL-substituting post that also allows for PCL retention may be of significant clinical value. Such an implant may satisfy surgeons desire to improve kinematic abnormalities of CR implants, without the challenges posed by attempting to retain the native ACL.

Materials and methods

This hypothesis was tested by using dynamic computational simulations performed in KneeSIM software (LifeModeler, San Clemente, CA) to evaluate kinematics of implant designs that either retain (BCR), substitute (ASCR), or sacrifice the ACL (CR) in comparison with the native knee simulation.

KneeSIM is a previously validated software tool that mimics an oxford-type physical test set-up (Fig. 1) commonly used to test/analyse kinematics of knee implant designs in cadaver specimens [27]. This software tool uses rigid body dynamics coupled with elastic foundation contact modelling to simulate knee mechanics and has been used by several researchers to analyse kinematics of different knee implant designs, effect of variation in component positioning, etc. [6, 24, 37, 43]. In particular, Patil et al. validated their computational model within KneeSIM using experimentally measured kinematic and kinetic data, and found major trends plotted as function of knee flexion angle to be similar between the computational and experimental results [6]. Magnetic resonance imaging (MRI) data from a previous IRB-approved study (2003P000337) was used to create average bone and cartilage models of tibia, femur, and patella, and identify average insertion locations for medial/lateral collateral ligaments (MCL/LCL), cruciate ligaments (ACL/PCL), and the patellar tendon [43]. The quadriceps angle was determined to be 14.0° based on the average literature values [1, 18] and the proximal quadriceps insertion was chosen accordingly.

Fig. 1
figure 1

Graphical visualization of the KneeSIM set-up modelling an oxford-type physical test including a detailed image of the knee joint

The kinematics of an ACL-substituting implant (ASCR) was compared to that of the same tibial articular surface without a post but with an intact ACL (BCR). This allowed for a direct comparison of the kinematic function of the ACL-substituting post, and the modelled ACL, for a given femoral and tibial articular surface design (identical for all tested implants). The ACL-substituting design tested in this study included an anterior tibial post that substitutes for the native ACL by interacting with the anterior portion of the femoral intercondylar notch. This concept of replacing the native ACL function is analogous to the concept of replacing PCL function in posterior stabilized (PS) implants using a post–cam interaction. The ACL-substituting post is also designed to accommodate the intact PCL (Fig. 2).

Fig. 2
figure 2

Schematic showing the novel implant design with an ACL-substituting post and retention of the native PCL, the PCL is shown at different flexion angles

Kinematics of the ASCR implant were also compared to the ACL-sacrificing CR implant that consisted of the same articular surface but without an ACL or ACL-substituting post. Additionally, simulations were performed for the average native knee using the average articular cartilage geometry (femur tibia and patella) derived from MRI data (the “native knee”). The native knee model included ACL and PCL. Figure 3 shows a model and cross sections of all the implants and the native knee tested in this study. In the Appendix, there is a further comparison of the ASCR design to a commercially available BCR and CR implant.

Fig. 3
figure 3

Implants tested in KneeSIM (ASCR, BCR, CR, and native knee) together with a sagittal cross section of the femur on the tibia

Several activities were simulated to capture activities of daily living involving different ranges of knee motion (ROM): walking (60° flexion), stair-ascent (90° flexion), sitting on a chair (105° flexion), and deep knee bend (DKB, 135° flexion). These simulations were carried out with ideal (normative) component placements. The components were mounted perpendicular to the mechanical axis on the average bone models to restore the joint line on the lateral side according to standard surgical technique. The tibial posterior slope was 7.0° for all implants.

Ligaments were modelled as nonlinear, tension-only springs, with average stiffness values obtained from the literature (Table 1), [21, 30, 41]. An initial preload was applied to the collateral ligaments to simulate a balanced knee joint during surgery. The ACL was modelled with initial tension at full extension, while the PCL was modelled to be slack at the starting position (Table 1), [5, 16]. For a given knee flexion, the knee joint was free to move in all other degrees of freedom (internal/external and varus/valgus rotation as well as all translations), and no forces were imposed on the level of the knee joint for any activity. Tibiofemoral contact forces and knee kinematics are interdependent, and are in turn driven by implant geometry, implant placement, soft-tissue properties, and quadriceps muscle forces. Within KneeSIM, different combinations of built-in loading conditions are available to simulate a variety of activities, with the muscle loads being automatically modulated to balance forces applied to the hip/ankle joint centre. For deep knee bend, a constant load of 180.0 N was applied to the hip joint centre, while for chair-sit the same constant load at the hip joint was applied in conjunction with a varying anteroposterior (AP) load at the ankle joint. For stair-ascent, a variable vertical hip load was applied, with a peak force of 670.0 N. For walking simulation, a variable vertical hip load with a peak of 1100.0 N, together with a variable ankle torque around the tibia and an adduction/abduction moment, was applied.

Table 1 Initial slack (+) or tension (−) and preload of ligaments and capsule used for the knee model set-up in KneeSIM

For all simulations, tibiofemoral kinematics were reported as the average of medial and lateral femoral condyle motions relative to the tibia. The posterior femoral shift in extension, defined as shift of the midpoint between the medial and lateral condyle centres relative to KneeSIM’s built-in local tibial coordinate system, was of particular interest for the purpose of this study. Furthermore, the range of knee flexion angles where ACL and PCL were under tension, and range of knee flexion angles where there was contact between the femoral component and the ACL-substituting post were also reported.

Results

Walking Simulation

During walking, the results for the CR implant revealed notable posterior femoral shift in extension (Fig. 4) relative to the native knee (4.9 mm), and predominantly anterior femoral motion with increasing knee flexion (6.2 mm). Neither the ACL-retaining nor ACL-substituting implants showed this posterior femoral shift in extension relative to the native knee during the gait cycle (Table 2). Both ACL-retaining and ACL-substituting implants showed more similar motion trends compared to the native knee than the ACL-sacrificing implant. During walking simulations, the ASCR post was engaged with the intercondylar notch of the femoral component throughout the stance phase of gait (effective flexion range over which the “ACL substitute” was functional), which was similar to the portion over which the ACL was under tension in the BCR simulation and the native knee (Table 3).

Fig. 4
figure 4

Extended femoral position of all implants and activities relative to the tibia showing significant posterior femoral shift for the CR TKA

Table 2 Femoral posterior rollback (average of medial and lateral condyle) relative to full extension in KneeSIM [mm]
Table 3 Range of knee flexion [°] over which the ACL post is in contact with the femoral component or ACL/PCL are under tension

Stair-ascent simulation

For the stair-ascent simulation, the CR implant data again showed posterior femoral shift in extension relative to the native knee (5.6 mm) followed by paradoxical anterior sliding. In contrast to ASCR and both BCR implant and native knee that showed posterior femoral rollback, the CR implant had a more anterior location (1.6 mm) of the femur on the tibia at 90° flexion compared to full extension (Table 2).

Chair-sit simulation

The ASCR implant again showed motion similar to the BCR with net posterior femoral rollback of 5.0 mm for ASCR and 6.6 mm for BCR, respectively. The CR data for chair-sit again showed substantial posterior femoral shift in extension relative to the native knee (6.2 mm, Fig. 4), followed by anterior femoral sliding of 4.5 mm until 60° before rollback occurred with higher knee flexion (Table 2).

Deep knee bend simulation

During deep knee bend, the whole range of knee flexion was covered, and the comparison of all implants and the native knee is shown in detail in Fig. 5. Like for all the other simulated activities in low flexion (<30°), the ASCR tibial post-femoral notch interaction provided a similar kinematic effect to that of the retained ACL in the BCR simulation unlike the CR simulation. At higher flexion angles when there was no contribution from either the tibial post in ASCR or the ACL in the BCR implant, the kinematics for all designs were virtually identical. The simulations for the native knee showed very similar trends in low flexion until around 30° knee flexion. With further knee flexion, the native knee showed more femoral rollback than any of the implants (Fig. 5, Table 2).

Fig. 5
figure 5

Average motion of the femur relative to tibia during DKB showing initial posterior rollback of the ASCR and BCR implants unlike for the CR implant, compared to the native knee showing continuous femoral rollback

Table 3 shows the range of knee flexion angles where the tibial post was in contact with the femoral component in the ASCR implant, and where the ACL/PCL were under tension in the BCR/CR implants and the native knee.

Discussion

The most important finding of the present study was that within the simulation environment, the kinematic function of the ACL in low knee flexion was successfully replicated by ACL substitution involving engagement of a tibial post with the femoral component. The anterior substituting cruciate-retaining (ASCR) design showed kinematics close to that of the BCR design, which had the same articular surface geometry as the ASCR design. Thus, like the BCR design, the ASCR implant was able to improve the kinematic abnormalities of the CR implants across the simulated activities in this study (Figs. 4, 5; Table 2). In all activities, the CR implant showed substantial posterior femoral shift in extension (Fig. 4) followed by paradoxical anterior sliding. These findings for the CR implant are consistent with various in vivo, in vitro (cadaver), and simulation studies [23, 34, 39, 42, 43]. Further, we have also compared kinematics of the ASCR design to a commercial BCR and CR device in the appendix. This confirmed the original hypothesis of this study.

The abnormal posterior femoral location in CR implants is largely due to the missing ACL, which is under tension in extension and holds the femur anteriorly on the tibia (Fig. 6). Following this posterior shift, the force imbalance within the joint causes paradoxical anterior sliding of the femur in early flexion. As explained by Blaha, the line of action of the body weight in early flexion lies behind the knee joint. This is balanced by quadriceps activation, and the absence of the ACL and slack state of the PCL in early flexion causes anterior femoral sliding [4]. To reduce such paradoxical anterior sliding, contemporary TKA implants often utilize increased anterior tibial lips, which also cause the femur to sit posteriorly on the tibia in extension. With increased flexion, a reduced tibiofemoral constraint allows additional laxity, and paradoxical anterior sliding occurs until the PCL is adequately tensioned in mid-flexion to guide femoral posterior rollback.

Fig. 6
figure 6

Schematic showing the poster femoral shift at full extension for contemporary CR implants relative to the posterior aspect of the tibia (vertical lines) in contrast to a centrally located femur for the ASCR and BCR designs

In contemporary CR TKA, the ACL is resected and its function is lost, which alters native knee kinematics following TKA surgery [23, 35, 43]. ACL retention is one way to overcome these kinematic abnormalities but presents several challenges as previously discussed. Therefore, the goal of this study was to determine whether the kinematic abnormalities of CR implants could be improved by substituting for the resected ACL. This is achieved by the interaction of a tibial post with the intercondylar notch of the femoral component at low flexion angles to provide anteroposterior stability similar to that provided by the ACL in the native knee (Fig. 6), [33]. Further, the ACL-substituting tibial post is intended to work in conjunction with the native PCL. At higher flexion angles, the ACL post is designed to disengage from the femoral component and allow motion/stability to be governed by the native PCL.

Liu et al. [17] proposed ACL reconstruction following TKA as a different way to improve kinematics in CR implants based on the results of their computational simulations. Other attempts to provide improved knee kinematics through implant design include a bi-cruciate substituting (BCS) device with an anterior and posterior cam–post interaction (Smith and Nephew, London, UK). Publications related to this design have shown kinematic improvements over contemporary PS implants where both ACL and PCL are resected. While the kinematics of this implant has been compared to other contemporary devices and native knees, the effectiveness of the ACL- and PCL-substituting mechanism relative to the actual ligaments has not been directly evaluated [24, 38]. Another commercially available design uses a ball-and-socket-like joint on the medial side to provide AP stability while allowing greater AP motion on the less constrained lateral side also intends to improve joint stability and kinematics (Wright Medical Group, Arlington, TN, USA). Generally, these medially constrained designs are indicated for PCL-sacrificing applications, and use articular surface constraint instead of a post–cam mechanism to substitute for ligament function. Some publications have even debated whether PCL function either through retention (CR) or through substitution (PS) is necessary [31]. Nonetheless, CR and PS both are established and successful treatment options. Our research aimed to analyse the effect of direct substitution for the ACL with an ACL post while leaving the PCL intact providing an improved alternative to contemporary CR implants.

There are several limitations to the present study: The first limitation is that the kinematics evaluated in this study are based on the simulations of an average knee model in KneeSIM. The extent to which kinematics of an individual knee can be replicated by an ACL-substituting post designed for the average population is unclear. Therefore, subject-specific simulation studies should be conducted in future to evaluate inter-subject variations in knee kinematics for the different designs. The second limitation is that while the native knee simulations included geometry of the average articular cartilage, the simulation package did not allow for modelling of the menisci. However, the results for the native knee across different activities showed kinematic trends similar to published in vivo data of normal knees, particularly the activity depended range of AP translation showed very similar trends [12, 14, 22]. Another limitation of this study is although KneeSIM and other such computational tools have been used for design and evaluation of TKA implants, it is still uncertain whether such tools can fully predict kinematic behaviour of knees in vivo. Therefore, continued analysis of this concept via cadaver testing, more advanced full-body musculoskeletal simulations, and eventually in vivo kinematic evaluation is required. Another limitation of this study is also that the effect of articular geometry on implant kinematics was not evaluated. This was because we wanted to achieve a direct comparison of ACL substitution versus ACL retention and ACL sacrifice, without the confounding effect of articular surface variation. Prior studies have shown important effect of articular geometry on kinematics of BCR and CR implants [40, 43]. Therefore, future studies relating to the ASCR should evaluate the effect of changes in articular geometry coupled with the presence and absence of the ACL-substituting post.

The results of the present study suggest that the ASCR post can substitute for kinematic function of the ACL and provide a more natural location of the femur on the tibia. However, proprioception provided by the ACL would still be lost, which may have important implications for joint function [3, 11]. This may be a possible limitation of the ACL substitution concept.

While TKA procedures provide excellent pain relief, a significant portion of patients remain dissatisfied due to functional limitations, residual symptoms, and perception of joint instability [8, 25]. ACL has long been recognized as a missing puzzle in the quest for addressing these clinical challenges. However, retention of native ACL in TKA poses many challenges, and may also necessitate the use of advanced tools to obtain reliable outcomes. The proposed concept of ACL substitution may be an alternative to ACL retention, particularly for patients with an absent or nonfunctional ACL. If the kinematic improvements seen here are replicated in vivo, improved patient outcomes could be achieved with such ACL-substituting designs compared to contemporary CR implants. This would allow surgeons to provide better outcomes for their patients, without encountering challenges of ACL retention.

Conclusion

In conclusion, an ACL-substituting design that retains the native PCL showed important kinematic improvements over a CR TKA during dynamic simulations. Particularly, the abnormal posterior femoral shift and paradoxical anterior sliding in low knee flexion seen with the CR implants were addressed with the ASCR design through replacement of the native ACL by an ACL-substituting post. The kinematic results of the ASCR design were similar to an ACL-retaining implant and the native knee.