Abstract
Purpose
Polyethylene wear particle generation is one of the most important factors that affects the mid- to long-term results of total knee arthroplasties (TKA). Mobile-bearing total knee prostheses were developed to reduce polyethylene wear generation. However, whether mobile-bearing prostheses actually generate fewer polyethylene wear particles than fixed-bearing prostheses remains controversial. The aim of this study was to compare, within individual patients, the in vivo polyethylene wear particles created by a newly introduced mobile-bearing prosthesis in one knee and a conventional fixed-bearing prosthesis in other knee.
Methods
Eighteen patients receiving bilateral TKAs to treat osteoarthritis were included. The synovial fluid was obtained from 36 knees at an average of 3.5 years after the operation. The in vivo polyethylene wear particles were isolated from the synovial fluid using a previously validated method and examined using a scanning electron microscope and an image analyser.
Results
The size and shape of the polyethylene wear particles from the mobile-bearing prostheses were similar to those from the conventional fixed-bearing prostheses. Although the number of wear particles from the mobile-bearing prosthesis (1.63 × 107 counts/knee) appeared smaller than that from the fixed-bearing prosthesis (2.16 × 107 counts/knee), the difference was not statistically significant.
Conclusions
The current in vivo study shows that no statistically significant differences were found between the polyethylene wear particles generated by a newly introduced mobile-bearing PS prosthesis and a conventional fixed-bearing PS prosthesis during the early clinical stage after implantation.
Level of evidence
Therapeutic study, Level III.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
Polyethylene wear particles induce macrophages to release cytokines, which can lead to osteolysis and aseptic loosening of total joint prostheses [4, 13, 14, 17]. Generation of polyethylene wear particles is one of the most important factors that affects the mid- and long-term clinical results after total knee arthroplasty (TKA) [2, 6]. Therefore, to achieve better long-term results for patients with high levels of activity, the materials and design of total knee prostheses have been modified to reduce polyethylene wear particles after TKA. The mobile-bearing prosthesis was expected to produce fewer polyethylene particles than the conventional fixed-bearing prostheses. However, only a few reports on the in vivo polyethylene wear from mobile-bearing prostheses have been published [19, 20]. Therefore, whether mobile-bearing prostheses lead to a reduction in wear remains controversial.
Recently, a highly conforming mobile-bearing posterior stabilized (PS) prosthesis (Vanguard RP; Biomet, Warsaw, IL, USA) was introduced. Because it takes decades to establish the long-term performance of such a newly introduced total knee prosthesis, it is particularly important to report early feedback on the in vivo polyethylene wear generation before the prosthesis is used widely. However, measuring the in vivo polyethylene wear using post-operative TKA radiographs is difficult. To provide such early feedback, we have established a method for measuring the in vivo polyethylene wear by isolating and analysing the polyethylene wear particles in the synovial fluid from well-functioning knees after TKA. The characteristics of the in vivo polyethylene wear particles generated between different types of total knee prostheses have been compared previously [10, 11, 18–22]. One previous report [19] compared an old mobile-bearing cruciate-retaining prosthesis (MBK, Zimmer, Warsaw, IN, USA) and an old PS prosthesis (IB-II, Zimmer, Warsaw, IN, USA), but those prostheses are no longer used. There have been no reports on contemporary mobile-bearing PS prostheses.
Thus, the aim of this study was to compare the in vivo polyethylene wear particles generated by a newly introduced mobile-bearing PS TKA prosthesis in one knee and a conventional fixed-bearing PS TKA prosthesis in the other knee within individual patients. It was hypothesized that the number, size, and shape of the polyethylene wear particles generated by a newly introduced mobile-bearing PS TKA prosthesis and a conventional fixed-bearing PS TKA prosthesis would be significantly different.
Materials and methods
Patients who had bilateral TKAs at our university hospital between 2008 and 2012 were eligible to participate in this study, and patients with primary osteoarthritis and varus deformity of the knee were included. The exclusion criteria included inflammatory arthritis, post-traumatic osteoarthritis, osteonecrosis of the knee, and valgus deformity of the knee. A total of 18 patients who had bilateral TKAs agreed to participate in the present study. Twelve of the patients had simultaneous bilateral TKAs, and six patients had staged bilateral TKAs at an interval of 0.2 ± 0.2 years (mean ± standard deviation). All of the patients received a mobile-bearing PS prosthesis (Vanguard RP, Biomet) in one knee and a conventional fixed-bearing PS prosthesis (Vanguard PS; Biomet) in the other knee (Figs. 1, 2). The Vanguard RP prosthesis is characterized by its unique saddle-shaped post-cam design and large contact area. A previous report showed that this prosthesis has a much larger contact area on the upper surface of the polyethylene insert and a much lower contact stress throughout the full range of motion than other fixed-bearing PS prostheses [26]. The saddle-shaped post-cam mechanism is in contact during the early phase of flexion and prevents the paradoxical anterior movement of the femoral component [16]. The lower surface of the polyethylene insert contains a polyethylene stem and tibial baseplate hole-type mobile-bearing mechanism and induces medial pivot motion after the operation [16]. The surface of the tibial tray was polished in the mobile-bearing PS prosthesis, but was not polished in the fixed-bearing PS prosthesis.
All of the patients were female, and all of the operations were performed by a single surgical team using the medial parapatellar approach. Besides the design of the articulating surface, the prostheses in these two groups were identical. The femoral and tibial components were made from Co–Cr alloy, and the tibial insert was conventional polyethylene (ArCom, Biomet). All of the components were secured with bone cement. Synovial fluid was obtained from all 36 of the knees at an average of 3.5 years after the operation.
The in vivo polyethylene wear particles were isolated from the synovial fluid samples using previously validated methods and examined using a scanning electron microscope and an image analyser [12, 19–23]. Each synovial fluid sample was digested with the same amount of 10 M sodium hydroxide at 65 °C for 12 h, applied to a sucrose density gradient (5, 10, 20 %) in a 14-mL tube (14PA tube, Hitachi Koki Co., Ltd., Tokyo, Japan), and then ultracentrifuged at 28,000 rpm (103,700×g) at 4 °C for 3 h (CP100α, P28S1014 rotor, Hitachi Koki Co., Ltd.). Next, the top layer was collected and applied to an isopropanol-water density gradient (0.90, 0.96 g/mL) in a 40-mL tube (40PA tube, Hitachi Koki Co., Ltd.) and ultracentrifuged again at 28,000 rpm (103,200×g) for 1 h (CP100α, P28S1004 rotor, Hitachi Koki Co., Ltd.). The polyethylene particles were collected from the interface between the two layers and filtered through 0.1-µm polycarbonate filters (VCTP 013-00, Millipore Corporation, Bedford, MA, USA). The filters were dried, attached to an aluminium specimen mount (M4, Nisshin EM Co., Ltd., Tokyo, Japan), and sputter-coated with platinum (E-1030 ion sputter, Hitachi Science Systems Ltd., Tokyo, Japan) for scanning electron microscopic examination (S-4700SI, Hitachi Ltd., Tokyo, Japan).
The total number of polyethylene wear particles in each knee joint was calculated from (a) the number of particles on the filter, (b) the retrieval ratio (65 %) [19], (c) the synovial fluid collection ratio (92 %) [19], (d) the amount of synovial fluid used for the extraction process, and (e) the total volume of synovial fluid. Particle size was expressed as an equivalent circle diameter, which is the diameter of a circle having the same area as the particle. Particle shape was determined from the aspect ratio (length to width) and roundness as perimeter2/(4π × Area) [11, 19–23]. Particle size and shape were determined with a digital image analyser (Mac Scope, Minami Co., Tokyo, Japan).
The pre-operative and post-operative activity levels of each patient were evaluated using the University of California Los Angeles (UCLA) activity-level rating [1]. The Knee Society score [11], extension and flexion angles of the knee joint [15], post-operative implant alignment according to the Knee Society grading [7], and mechanical axis using full-length leg radiography were also evaluated. The extension and flexion angles of the knee joint were determined with a standard clinical goniometer according to the method of Kim et al. [15]. A single expert observer (Y.M.) evaluated these parameters for all patients.
The protocol for this study was approved by the Institutional Committee on Human Research at Osaka City University Graduate School of Medicine (number of approval: 1281). Informed consent was obtained from all of the patients before their enrolment.
Statistical analysis
The significance of differences between the two prostheses was assessed using the paired t tests with a commercially available software package (StatView 4.5; Abacus Concepts Inc., Berkeley, CA, USA). P values <0.05 were considered statistically significant. A sample size calculation showed that 16 knees in each group would allow detection of a difference of 1.5 × 107 polyethylene wear particle counts/knee with a power of 0.8 at an α of 0.05 for a standard deviation of 2 × 107 counts/knee.
To determine the interobserver reliability of the measurements, intraclass correlation coefficients (ICCs) were calculated using the method of absolute agreement and with a single measurement as the unit of analysis using the SPSS ver. 13.0 (SPSS Inc., Chicago, IL, USA) for 6 patients. ICCs below 0.4 represented poor reliability, those from 0.4 to 0.6 represented moderate reliability, those from 0.6 to 0.8 represented good reliability, and those from 0.8 to 1.0 represented very good reliability, according to Bland and Altman [3]. Total number (ICC = 0.85; 95 % CI [confidence interval], 0.12–0.98), ECD (ICC = 0.98; 95 % CI, 0.87–0.99), aspect ratio (ICC = 0.75; 95 % CI, 0.52–0.96), and roundness (ICC = 0.87; 95 % CI, 0.24–0.98) had very good or good interobserver reliability.
Results
All of the patients analysed were female. The average age at the time of the operation was 75 ± 7 years, and the average height and body weight were 148 ± 4 cm and 58 ± 8 kg. Additional salient patient data for both groups are shown in Table 1. The polyethylene insert was thicker in the mobile-bearing than in the fixed-bearing PS prosthesis (p = 0.011), but the differences in other pre- and post-operative parameters were not significant. There was no significant difference between the synovial fluid volumes collected from the mobile-bearing (6 ± 3 mL) and fixed-bearing (7 ± 3 mL) PS prostheses.
The quantity, size, and shape of the polyethylene wear particles from each group are shown in Table 2. The total number of polyethylene wear particles obtained from the knees with the mobile-bearing PS prosthesis tended to be less than that obtained from those with the fixed-bearing PS prosthesis, but the difference was not statistically significant. The size distribution of the polyethylene wear particles is shown in Fig. 3. Particles of size 0.8–1.0 µm occurred most frequently in both groups. The differences in the size and shape (aspect ratio and roundness) between the groups were not statistically significant.
Discussion
The most important findings of the present study were that the number, size, and shape of the in vivo polyethylene wear particles generated from a newly introduced mobile-bearing PS prosthesis were not statistically different than those generated from a conventional fixed-bearing PS prosthesis. Although the number of wear particles of the mobile-bearing PS prosthesis tended to be less than that of the fixed-bearing PS prosthesis, the difference was not statistically significant.
One previous report analysing the in vivo polyethylene wear particles also found that the number, size, and shape of the in vivo polyethylene wear particles generated by a mobile-bearing prosthesis were not statistically different from those generated by a fixed-bearing prosthesis [19]. However, the prosthesis investigated in that study was an old mobile-bearing cruciate-retaining prosthesis, which is no longer used. In fact, different mobile-bearing prostheses were assessed, and thus the clinical results were varied with the various mobile-bearing prostheses. Therefore, the results of the previous study could not be applied to the recently introduced mobile-bearing prosthesis examined in the present study.
Although several studies have reported simulations of the in vitro wear of mobile-bearing total knee prostheses, the results remain controversial. In mobile-bearing prostheses, the reduction in contact stress on the polyethylene caused by the large contact area may be offset by the increased number of articulating surfaces. Several studies reported advantages of mobile-bearing prostheses over fixed-bearing prostheses [5, 8, 18], but others did not [9, 24, 27]. However, there have been no studies on the newly introduced mobile-bearing PS prosthesis, which was investigated in the present study. The results reported here show that the mobile-bearing PS prosthesis did not generate more in vivo polyethylene wear particles than the conventional fixed-bearing PS prosthesis.
There are several strengths to the present study. First, we analysed polyethylene wear particles from patients who received a mobile-bearing PS prosthesis in one knee and a conventional fixed-bearing PS prosthesis in the other knee. The level of physical activity and body weight of patients strongly influence the amount of polyethylene wear after TKA. Here, the activity level and body weight were identical for the two different prostheses, which almost removed the influence of such factors on the polyethylene wear particle generation. Second, we used the same brand of prostheses for the two groups; only the design of the articulating surface was different between the mobile-bearing and fixed-bearing prostheses. Previous reports have shown that the materials used for the femoral component [20] and tibial insert [12] also influence in vivo polyethylene wear particle generation. Here again, the materials used for the femoral component, tibial component, and tibial insert were identical between the two different prostheses. Thus, we were able to eliminate the influence of the materials and properly evaluate the isolated effects of the difference in articulating surface on the in vivo polyethylene wear particle generation.
This study has several limitations. First, there may be a selection bias. The patients who agreed to participate in the present study showed good clinical results, with high Knee Society and UCLA activity-level scores and improved knee joint flexion angles post-operatively. The number of in vivo polyethylene wear particles may be smaller in other patients with lower activity scores and smaller flexion angles in the knee joint. Therefore, the results of this study may apply to active patients with high activity levels, but not to patients with low active levels. Second, the polyethylene wear particles analysed were not from the soft tissue, but from the synovial fluid. However, it is unethical to harvest soft tissue from well-functioning patients. Polyethylene wear particles are generated from the prostheses and dispersed into the synovial fluid. Some particles are captured in the capsule, while others migrate to the tissue interface and cause osteolysis and aseptic loosening. Previous report showed that the concentration of polyethylene wear particles in the synovial fluid was significantly correlated with the number of particles and histiocytes observed by histological analysis and the degree of osteolysis found during revision TKA [25]. Therefore, the number of polyethylene wear particles in the synovial fluid is related, but not necessarily equal, to the number of particles generated from the prosthesis and deposited at the tissue interface. Comparing the polyethylene wear particles in the synovial fluid of knees with the newly introduced total knee prosthesis and those with the established total knee prosthesis design for which long-term clinical results are already available may also be predictive of the incidence of osteolysis. Besides total hip arthroplasty, it is difficult to measure polyethylene wear using radiographs after joint replacement because the articulation of the joints is not ball and socket shape and the posture of the knee joint (extension–flexion and rotation) changes the contact area between the femoral component and the polyethylene insert. Therefore, in vivo polyethylene wear particle analyses on synovial fluid samples after TKA were performed. Third, the polyethylene inserts could not be observed in the present study. Therefore, the different types of polyethylene wear, such as abrasive, adhesive, cracks, and delamination, could not be distinguished. Further studies would be needed to determine the relationship between the characteristics of the polyethylene wear particles and the types of polyethylene wear. Fourth, it was difficult to make an accurate sample size calculation for this study. The characteristics of in vivo polyethylene wear particles are influenced by both the type of prosthesis and the patient activity level, and no reports on the relationship between the number of particles in the synovial fluid and clinical outcomes were available. Thus, the data from a recent study on contemporary fixed-bearing PS prosthesis [23] were used to make the sample size calculation. This may not be the best approach, but no better method was available. Fifth, the polyethylene insert was thicker in the mobile-bearing group than in the fixed-bearing group. This difference in thickness between the groups may have affected the characteristics of the in vivo polyethylene wear particles.
Although the newly introduced mobile-bearing PS prosthesis did not reduce the amount of in vivo polyethylene wear particles generated compared to a conventional fixed-bearing PS prosthesis in the early clinical stage after implantation, there were no adverse effects of the newly introduced prosthesis either. Therefore, there is currently no reason to abandon the newly introduced mobile-bearing PS prosthesis. However, further follow-up studies on the generation of polyethylene wear particles should be performed to determine whether there are advantages to the mobile-bearing PS prosthesis.
Conclusions
The current in vivo study shows that no statistically significant differences were found between the polyethylene wear particles generated by a newly introduced mobile-bearing PS prosthesis and a conventional fixed-bearing PS prosthesis during the early clinical stage after implantation.
References
Amstutz HC, Thomas BJ, Jinnah R, Kim W, Grogan T, Yale C (1984) Treatment of primary osteoarthritis of the hip. A comparison of total joint and surface replacement arthroplasty. J Bone Joint Surg 66A:228–241
Benevenia J, Lee FY, Buechel F, Parsons JR (1998) Pathologic supracondylar fracture due to osteolytic pseudotumor of knee following cementless total knee replacement. J Biomed Mater Res 43:473–477
Bland JM, Altman DG (1990) A note on the use of the intraclass correlation coefficient in the evaluation of agreement between two methods of measurement. Comput Biol Med 20:337–340
DeHeer DH, Engels JA, DeVries AS, Knapp RH, Beebe JD (2001) In situ complement activation by polyethylene wear debris. J Biomed Mater Res 54:12–19
Delport HP, Sloten JV, Bellemans J (2010) Comparative gravimetric wear analysis in mobile versus fixed-bearing posterior stabilized total knee prostheses. Acta Orthop Belg 76:367–373
Engh GA, Dwyer KA, Hanes CK (1992) Polyethylene wear of metal-backed tibial components in total and unicompartmental knee prosthesis. J Bone Joint Surg 74B:9–17
Ewald F (1989) The knee society total knee arthroplasty roentgenographic evaluation and scoring system. Clin Orthop Relat Res 248:9–12
Fisher J, Jennings LM, Galvin AL, Jin ZM, Stone MH, Ingham E (2010) 2009 Knee Society Presidential Guest Lecture: polyethylene wear in total knees. Clin Orthop Relat Res 468:12–18
Grupp TM, Kaddick C, Schwiesau J, Maas A, Stulberg SD (2009) Fixed and mobile bearing total knee arthroplasty–influence on wear generation, corresponding wear areas, knee kinematics and particle composition. Clin Biomech (Bristol, Avon) 24:210–217
Hinarejos P, Piñol I, Torres A, Prats E, Gil-Gómez G, Puig-Verdie L (2013) Highly crosslinked polyethylene does not reduce the wear in total knee arthroplasty: in vivo study of particles in synovial fluid. J Arthroplasty 28:1333–1337
Insall J, Dorr L, Scott R, Scott W (1989) Rationale of the knee society clinical rating system. Clin Orthop Relat Res 248:13–14
Iwakiri K, Minoda Y, Kobayashi A, Sugama R, Iwaki H, Inori F, Hashimoto Y, Ohashi H, Ohta Y, Fukunaga K, Takaoka K (2009) In vivo comparison of wear particles between highly crosslinked polyethylene and conventional polyethylene in the same design of total knee arthroplasties. J Biomed Mater Res B Appl Biomate 91:799–804
Kadoya Y, Revell PA, al-Saffar N, Kobayashi A, Scott G, Freeman MA (1996) Bone formation and bone resorption in failed total joint arthroplasties: histomorphometric analysis with histochemical and immunohistochemical technique. J Orthop Res 14:473–482
Kadoya Y, Revell PA, Kobayashi A, al-Saffar N, Scott G, Freeman MA (1997) Wear particulate species and bone loss in failed total joint arthroplasties. Clin Orthop Relat Res 340:118–129
Kim YH, Sohn KS, Kim JS (2005) Range of motion of standard and high-flexion posterior stabilized total knee prostheses. A prospective, randomized study. J Bone Joint Surg Am 87:1470–1475
Kurita M, Tomita T, Yamazaki T, Fujii M, Futai K, Shimizu N, Yoshikawa H, Sugamoto K (2012) In vivo kinematics of high-flex mobile-bearing total knee arthroplasty, with a new post-cam design, in deep knee bending motion. Int Orthop 36:2465–2471
Laskin RS, Davis J (2005) Total knee replacement using the Genesis II prosthesis: a 5-year follow up study of the first 100 consecutive cases. Knee 12:163–167
McEwen HM, Fisher J, Goldsmith AA, Auger DD, Hardaker C, Stone MH (2001) Wear of fixed bearing and rotating platform mobile bearing knees subjected to high levels of internal and external tibial rotation. J Mater Sci Mater Med 12:1049–1052
Minoda Y, Kobayashi A, Iwaki H, Miyaguchi M, Kadoya Y, Ohashi H, Takaoka K (2004) Characteristics of polyethylene wear particles isolated from synovial fluid after mobile-bearing and posterior-stabilized total knee arthroplasties. J Biomed Mater Res 71B:1–6
Minoda Y, Kobayashi A, Iwaki H, Iwakiri K, Inori F, Sugama R, Ikebuchi M, Kadoya Y, Takaoka K (2009) In vivo analysis of polyethylene wear particles after total knee arthroplasty: the influence of improved materials and designs. J Bone Joint Surg 91A(Suppl 6):67–73
Minoda Y, Kobayashi A, Iwaki H, Miyaguchi M, Kadoya Y, Ohashi H, Yamano Y, Takaoka K (2003) Polyethylene wear particles in synovial fluid after total knee arthroplasty. Clin Orthop Relat Res 410:165–172
Minoda Y, Kobayashi A, Iwaki H, Miyaguchi M, Kadoya Y, Ohashi H, Takaoka K (2005) Polyethylene wear particle generation in vivo in an alumina medial pivot total knee prosthesis. Biomaterials 26:6034–6040
Minoda Y, Hata K, Iwaki H, Ikebuchi M, Hashimoto Y, Inori F, Nakamura H (2014) No difference in in vivo polyethylene wear particles between oxidized zirconium and cobalt–chromium femoral component in total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc 22:680–686
Mueller-Rath R, Kleffner B, Andereya S, Mumme T, Wirtz DC (2007) Measures for reducing ultra-high-molecular-weight polyethylene wear in total knee replacement: a simulator study. Biomed Tech (Berl) 52:295–300
Piñol I, Torres A, Gil G, Prats E, Puig-Verdier L, Hinarejos P (2014) Polyethylene particles in joint fluid and osteolysis in revision total knee arthroplasty. Knee 21:402–405
Shiramizu K, Vizesi F, Bruce W, Herrmann S, Walsh WR (2009) Tibiofemoral contact areas and pressures in six high flexion knees. Int Orthop 33:403–406
Utzschneider S, Harrasser N, Schroeder C, Mazoochian F, Jansson V (2009) Wear of contemporary total knee replacements–a knee simulator study of six current designs. Clin Biomech (Bristol, Avon) 24:583–588
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Minoda, Y., Hata, K., Ikebuchi, M. et al. Comparison of in vivo polyethylene wear particles between mobile- and fixed-bearing TKA in the same patients. Knee Surg Sports Traumatol Arthrosc 25, 2887–2893 (2017). https://doi.org/10.1007/s00167-016-4027-z
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00167-016-4027-z