The enduring success of the low-friction arthroplasty, advanced by Sir John Charnley as a solution for hip arthrosis, may be appreciated by the fact that in 2016 almost 1.4 million primary and revision hip and knee arthroplasty procedures were performed in the United States, a number more than doubling on a global basis. Improvements in surgical technique and implant design over the last four decades have resulted in total knee arthroplasty (TKA) being deemed one of the most successful, contemporary orthopaedic procedures to effectively relieve pain and allow patients to resume the activities of their daily lives. The prevalence of aseptic loosening attributed to ultra-high molecular weight polyethylene (UHMWPE) wear debris-induced osteolysis is in the single digits in most knee series, with some reports describing prosthesis survival beyond 20 years. Despite this obvious success, UHMWPE wear is an inescapable consequence of total joint articulation and is of contemporary concern particularly as our population grays and lifestyle demands increase. While UHMWPE has remained the tibial insert and patellar component bearing material of choice over the last four decades, researchers are continually striving to increase its in-vivo longevity through alterations to processing and/or sterilization techniques. This chapter presents an overview of TKA bearing materials and how they integrate into an orthopaedic triad where patient outcomes are not only dictated by the implant but are highly dependent on patient factors and technical proficiency to assist the goal of avoiding revision arthroplasty.
Access provided by CONRICYT-eBooks. Download chapter PDF
Similar content being viewed by others
Keywords
The Enduring Goal
The enduring success of the low-friction arthroplasty, advanced by Sir John Charnley as a solution for hip arthrosis, may be appreciated by the fact that in 2016 almost 1.4 million primary and revision hip and knee arthroplasty procedures were performed in the United States, a number more than doubling on a global basis [1] (Table 2.1). Improvements in surgical technique and implant design over the last four decades have resulted in total knee arthroplasty (TKA) being deemed one of the most successful, contemporary orthopaedic procedures to effectively relieve pain and allow patients to resume the activities of their daily lives. The prevalence of aseptic loosening attributed to ultra-high molecular weight polyethylene (UHMWPE) wear debris-induced osteolysis is in the single digits in most knee series, with some reports describing prosthesis survival beyond 20 years [2,3,4,5,6,7,8,9,10,11,12,13,14,15,16,17,18,19,20,21,22,23,24,25]. Despite this obvious success, UHMWPE wear is an inescapable consequence of total joint articulation and is of contemporary concern particularly as our population grays and lifestyle demands increase [26,27,28,29,30,31,32,33,34,35,36,37,38,39,40,41,42,43,44]. Appreciating an orthopaedic triad where patient outcomes are not only dictated by the implant but are highly dependent on patient factors and technical proficiency assists the goal of avoiding total knee arthroplasty revision.
The Triad: The Implant
The Evolution of UHMWPE
The UHMWPE used in joint arthroplasty components results from polymerization of ethylene gas into a fine resin powder of submicron and micron size distribution. A number of resin mixtures exist, but GUR 1020 and GUR 1050 are the prevalent polymers utilized in contemporary devices. They are consolidated with the use of ram extrusion or compression-molding techniques. Structurally, UHMWPE is made up of repeating carbon-hydrogen chains that are arranged in ordered (crystalline) and disordered (amorphous) regions [45]. While UHMWPE has remained the tibial insert and patellar component bearing material of choice over the last four decades, researchers are continually striving to increase its in-vivo longevity through alterations to processing and/or sterilization techniques.
Short- to mid-term clinical reports of UHMWPE damage in the 1990s led to a review of manufacturing processes and determined that inadequate quality control resulted in fusion defects arising from incomplete polymerization, voids, and foreign body inclusions [46,47,48]. Recognizing the direct impact of these variables on the in-vivo degradation of the final parts, orthopaedic device manufacturers addressed the allowable tolerances for these components, and these issues have not reappeared in the peer-reviewed literature.
Previous attempts to improve UHMWPE performance have included carbon fiber reinforcement (Poly-II) [49] and polymer reprocessing by hot isostatic pressing (Hylamer) [50]. The former was withdrawn from the market because of an unexpectedly high wear rate [51] (Fig. 2.1), while the latter has been linked to debris-induced osteolytic response, especially when sterilized by gamma irradiation in air [52] (Fig. 2.2). Heat pressing was yet another attempt to improve the finish of the articular surface, but was associated with UHMWPE fatigue and early delamination [53] (Fig. 2.3). These material innovations had checkered pasts as they moved from the laboratory to clinical application.
Gamma irradiation in air was the predominant method of UHMWPE component sterilization, and, to this day, represents the long-term standard against which contemporary material improvements are measured. In the early 1990s, an increasing prevalence of tibial component failures associated with debris-induced osteolysis raised concerns over the long-term durability of contemporary devices [54, 55]. A clinical follow-up study reported by Bohl et al. suggested that this may be accounted for by the prolonged shelf storage prior to implantation of UHMWPE components gamma irradiated in air [56]. A 12–20% reduction in in-vivo survival was noted for shelf storage ranging from 4 to 11 years with a mean in-vivo time to revision of 2.5 years (Figs. 2.4 and 2.5).
Further, laboratory studies indicated that as shelf storage increased, the amount of UHMWPE exposed to high surface stresses during articulation increased dramatically and was a contributing factor to early in-vivo polymer failure [57,58,59] (Fig. 2.6).
The explanation for these observations lies in the mechanics of the sterilization process, which facilitates breakage of polymer chains by the incoming gamma radiation, creating free radicals, which preferentially combine with available oxygen [60, 61] (Fig. 2.7). The onset of mass UHMWPE component production and device modularity resulted in extended component shelf storage before use. This was not a previous consideration, but ongoing shelf life oxidation offered an explanation for mechanical compromise of the polymer in-situ [58, 60, 62, 63] (Fig. 2.8). It was also noted that in-vivo component oxidation occurred, but to a lesser degree [64].
At this point, attempts to remove oxygen from the sterilization process included the use of inert gas and vacuum environments or by avoiding gamma irradiation altogether through the use of ethylene oxide (EtO) or gas plasmas [65,66,67]. Acetabular components sterilized by these techniques demonstrated a reduction in UHMWPE wear in hip simulation studies (Fig. 2.9). Today, orthopaedic device manufacturers avoid the use of an air environment when packaging UHMWPE components, and sterilization dates are standard on device package labeling.
It is now also quantitatively appreciated that increasing the gamma radiation dose above the 2.5 Mrad level used in conventional UHMWPE component sterilization encourages free radicals to combine, creating cross-links between the molecules of adjacent chains, which is further enhanced in an oxygen-free environment [68,69,70]. Figure 2.10 from McKellop and coworkers [69] is descriptive of this phenomenon in a simulator comparison of acetabular cup components with the volumetric wear per million cycles dramatically reduced with increasing gamma radiation exposure.
There are clinical reports attributed to Oonishi and Grobbelaar, which describe in-vivo UHMWPE wear reduction in acetabular components realized through increased cross-linking [71,72,73,74,75,76]. However, these studies employed large doses of gamma radiation (>50 Mrad), which are known to cause polymer embrittlement and yellowing. Wroblewski, employing a chemically enhanced cross-linked polymer, achieved similar findings both in-vivo and in-vitro, when coupled with an alumina articulation [77].
These isolated studies pointed the way to a new class of UHMWPEs, whose common denominator was an appreciation of the importance of increased cross-linking while minimizing oxidative degradation to reduce wear. Initial methods used to manufacture these moderately to highly cross-linked UHMWPEs included (1) heating above or below the melt temperature of the polyethylene, (2) the type of radiation employed, (3) the radiation dose level, (4) the sequence of stepwise application, and (5) the endpoint sterilization. The one common factor is that radiation was integrated into the manufacturing process. All received U.S. Food and Drug Administration 510[k] clearance, allowing commercial distribution for both hip and knee components (Table 2.2).
However, changes in the mechanical properties of these materials, particularly in their reduced resistance to fatigue crack propagation (fracture toughness), raised concerns about their long-term suitability in hip and knee components where locking mechanisms offered foci for stress risers [78,79,80,81] (Figs. 2.11, 2.12, and 2.13). Short-term clinical reports for total hip arthroplasty demonstrated a significant reduction in wear volume and rate for these polymers [82,83,84,85,86,87], which supported the impressive preclinical hip simulation laboratory data [88,89,90,91,92]. However, the negative impact of extreme component positioning on outcome was also demonstrated through case and retrieval reports at this time [80, 93, 94].
An appreciation of the differing modes of hip (abrasion and adhesion) and knee (pitting and delamination) failure, confirmed through conventional UHMWPE component retrieval [95,96,97], also suggested that a universal, moderately to highly cross-linked polymer may not be appropriate. To counter the reported degradation in material properties, “enhanced” UHMWPEs now have antioxidants, predominantly vitamin E, infused or blended into the resin powder during the manufacturing process (Table 2.3). Laboratory studies have confirmed the maintenance of UHMWPE mechanical properties and wear resistance in addition to prevention of oxidative degradation for these polymers [98,99,100,101,102,103,104,105].
Contemporary peer-reviewed literature for the moderately to highly cross-linked UHMWPEs in total hip arthroplasty is reporting dramatic reduction of wear rate when compared to conventional UHMWPE in mid- to long-term follow-up studies with metal femoral heads [106,192,107,108,109,110,111,112,113,114,115,116,117,118,119,121]. Total knee arthroplasty clinical reporting focuses on aseptic loosening and mechanical failures rather than wear rate, but again, in short- to mid-term studies, these UHMWPEs are demonstrating efficacy [121,122,123,124,193,194,195,196,197,126]. While short- to mid-term clinical studies supporting the further advantages of the antioxidant-infused UHMWPEs in total hip arthroplasty are increasing [127–131], reporting for total knee arthroplasty has just begun [132]. While the overall clinical gains of these enhancements have been questioned [133], the passage of in-vivo time will be, as has always been, the defining factor in the continued use of these moderately to highly cross-linked UHMWPEs with or without antioxidants.
The Femoral Side
While the predominant focus for increasing the in-vivo longevity of total knee arthroplasty is alteration of the UHMWPE tibial insert, there are alternative bearing options for the femoral component as well. As example, oxidized zirconium, marketed under the trade name Oxinium (Smith + Nephew, Memphis, TN) in the United States, has the strength of metallic cobalt-chromium femoral components with the wear characteristics of ceramics [134], as has been shown in laboratory simulators for pairings with both conventional and highly cross-linked UHMWPE [135, 136]. The uniqueness of this material is that it offers patients with metal hypersensitivity, particularly to nickel, an implant option that has been shown to be clinically equivalent to cobalt-chromium femoral components [131,132,133,134,135,142].
The Tibial-Femoral Geometries
As knee designs have evolved, a growing appreciation of the avoidance of round-on-flat geometries through the ranges of knee flexion in favor of round-on-curved surfaces emerged [54]. The ability of a given design to minimize contact stresses during walking gait contributes to UHMWPE tibial component longevity [143]. With this, the trend toward more conforming design geometries also has associated with it the expectation that femoral component tolerances be maintained during the manufacturing process. Failure to achieve this can dramatically decrease contact surfaces, elevate peak stresses, and, concurrent with articulation, is the harbinger of material damage [144] (Fig. 2.14).
The attainment of femoral component tolerances has markedly improved with the use of computer-aided precision grinding as a standard finishing technique for metallic femoral knee components. This is particularly beneficial where small variations in surface contours have large effects on contact areas and surface stresses (Fig. 2.15). The implications of this technique have potentially far-reaching consequences. As design specifications are produced with tighter tolerances, the need for precision manufacturing is imperative (Fig. 2.16).
The Wear Particles Produced
Conventional wisdom and our experience particular to hip arthroplasty suggest that osteolytic response is associated with both particle size and debris volume. Laboratory hip simulator experiments have shown that UHMWPE particle volumes in various size ranges are dependent on radiation dose [145] (Figs. 2.17 and 2.18). The greatest potential for cytokine release, the first step in the sequelae leading to osteolysis, following macrophage debris encapsulation is at the <1 μm level. Ingram et al. suggested that highly cross-linked UHMWPE debris obtained from scratched surface articulation was bioreactive when placed in culture medium and appeared to be volume dependent [146].
The influence of surface roughness was further investigated by Scott et al. in a hip simulator comparison between conventional, EtO, and 10 Mrad UHMWPE components [147]. As one appreciates from Fig. 2.19, roughened surfaces have a negative influence on particle production where highly cross-linked UHMWPEs are employed. This was challenged by Muratoglu et al. in a study in which retrieved femoral components were articulated in knee simulation against a highly cross-linked UHMWPE [148]. Further, retrieved oxidized zirconium femoral components have demonstrated decreased surface roughness with time in-vivo, suggesting another benefit of this cobalt-chromium alternative for improving the long-term viability of knee articulations [143,144,145,152].
The Triad: The Patient
Overenthusiastic patient use following total knee arthroplasty has been cited as a factor influencing failure [147,148,155]. Its occurrence, however, has generally been described in singular case reports in much the same way as failure attributed to obesity. Series reports do not support a relationship between increased body mass index and device failure following arthroplasty [150,151,152,153,154,161]. Surgical preference, however, weighs in favor of the lightweight patient as the ideal arthroplasty candidate [162]. It is also known from both physical laboratory testing and finite element analysis that load magnitude in combination with displacement are factors influencing UHMWPE damage [157,158,159,160,161,162,163,170]. While a recommendation for patient weight loss before surgery may be justified from these laboratory investigations, the clinical reality of achieving this does not lie in the patient’s or surgeon’s favor [171].
With the patient population pursuing total knee arthroplasty getting younger and living longer, it is imperative that contemporary implant bearing materials address these increasing demands [172]. Clinical studies are now focusing more on patient-reported outcomes and relating them to comorbidities in an effort to align expectations for both the patient and the surgeon [167,168,175].
The Triad: The Surgery
The forces and torques that occur during walking gait, particularly during toe-off, promote articulation in the posteromedial quadrant of tibial inserts [170,171,172,173,180]. Retrieved components of failed knee arthroplasties demonstrate UHMWPE damage patterns in this area [175,176,177,178,185] (Fig. 2.20). Notwithstanding poor component design, causal factors include overloading the medial compartment, improper surgical correction or alignment of the bony structures, insufficient soft tissue balance and release, polyethylene cold flow near the edge of the tibial plateau, and surgical malrotation of the components [175,176,177,178,185]. In addition, the dynamic effects of lift-off and subsequent impact loading and unusual patient kinematics further increase the potential for posteromedial failures [186]. The influence of surgical malrotation may be appreciated in Fig. 2.21a, b, which demonstrate dramatic changes in location, contact area, and peak stresses for a PCL preserving knee in a laboratory investigation [187].
The continual emphasis on templating and the technological advances in computer-assisted and robotic navigation systems , intraoperative sensors as well as patient-specific instrumentation offer the promise that component malalignment may ultimately be minimized and patient satisfaction increased [182,183,184,185,186,187,194]. Eliminating the outliers in component placement will contribute to diminishing UHMWPE material damage in knee arthroplasty, however, the best technology to utilize in the achievement of this goal, is yet to be defined [189,190,197].
The Enduring Promise
The previous remarks have attempted to define problems, solutions, and unknown performance factors of bearing materials currently utilized in total knee arthroplasty as they relate to the implant, the patient, and the surgery. What is important for the reader to appreciate is that this is a continually evolving experience, which will find advocacy or limitations, with the passage of in-vivo time.
References
Orthopedic Network News. 28:3, Ann Arbor, MI: Mendenhall Associates, Inc.; 2017.
Aglietti P, Buzzi R, De Felice R, et al. The Insall-Burstein total knee replacement in osteoarthritis: a 10-year minimum follow-up. J Arthroplast. 1999;14(5):560–5.
Berger RA, Rosenberg AG, Barden RM, et al. Long-term followup of the Miller-Galante total knee replacement. Clin Orthop Relat Res. 2001;388:58–67.
Blumenfeld TJ, Scott RD. The role of the cemented all-polyethylene tibial component in total knee replacement: a 30-year patient follow-up and review of the literature. Knee. 2010;17(6):412–6.
Buechel FF Sr, Buechel FF Jr, Pappas MJ, D’Alessio J. Twenty-year evaluation of meniscal bearing and rotating platform knee replacements. Clin Orthop Relat Res. 2001;388:41–50.
Buechel FF. Long-term followup after mobile-bearing total knee replacement. Clin Orthop Relat Res. 2002;404:40–50.
Cross MJ, Parish EN. A hydroxyapatite-coated total knee replacement: prospective analysis of 1000 patients. J Bone Joint Surg Br. 2005;87(8):1073–6.
Epinette J-A. Long lasting outcome of hydroxyapatite-coated implants in primary knee arthroplasty: a continuous series of two hundred and seventy total knee arthroplasties at fifteen to twenty two years of clinical follow-up. Int Orthop. 2014;38(2):305–11.
Ewald FC, Wright RJ, Poss R, et al. Kinematic total knee arthroplasty: a 10- to 14-year prospective follow-up review. J Arthroplast. 1999;14(4):473–80.
Gill GS, Joshi AB. Long-term results of cemented, posterior cruciate ligament-retaining total knee arthroplasty in osteoarthritis. Am J Knee Surg. 2001;14(4):209–14.
Gill GS, Joshi AB, Mills DM. Total condylar knee arthroplasty. 16- to 21-year results. Clin Orthop Relat Res. 1999;367:210–5.
Goldberg VM, Kraay M. The outcome of the cementless tibial components: a minimum 14-year clinical evaluation. Clin Orthop Relat Res. 2004;428:214–20.
Gustke KA. The natural-knee system: 25 years of successful results. Am J Orthop. 2010;39(6 Suppl):5–8.
Jordan LR, Olivo JL, Voorhorst PE. Survivorship analysis of cementless meniscal bearing total knee arthroplasty. Clin Orthop Relat Res. 1997;338:119–23.
Kelly MA, Clarke HD. Long-term results of posterior cruciate-substituting total knee arthroplasty. Clin Orthop Relat Res. 2002;404:51–7.
Khaw FM, Kirk LMG, Morris RW, Gregg PJ. A randomised, controlled trial of cemented versus cementless press-fit condylar total knee replacement. J Bone Joint Surg Br. 2002;84(5):658–66.
Melton JTK, Mayahi R, Baxter SE, Glezos C. Long-term outcome in an uncemented, hydroxyapatite-coated total knee replacement: a 15- to 18-year survivorship analysis. J Bone Joint Surg Br. 2012;94(8):1067–70.
Pavone V, Boettner F, Fickert S, et al. Total condylar knee arthroplasty: a long-term followup. Clin Orthop Relat Res. 2001;388:18–25.
Park JW, Kim YH. Simultaneous cemented and cementless total knee replacement in the same patients: a prospective comparison of long-term outcomes using an identical design of NexGen prosthesis. J Bone Joint Surg Br. 2011;93(11):1479–86.
Rasquinha VJ, Ranawat CS, Cervieri CL, Rodriguez JA. The press-fit condylar modular total knee system with a posterior cruciate-substituting design. A concise follow-up of a previous report. J Bone Join Surg Am. 2006;88(5):1006–10.
Ritter MA, Keating EM, Sueyoshi T, Davis KE, Barrington JW, Emerson RH. Twenty-five-years and greater: results after nonmodular cemented total knee arthroplasty. J Arthroplast. 2016;31(10):199–202.
Rodriguez JA, Bhende H, Ranawat CS. Total condylar knee replacement: a 20-year followup study. Clin Orthop Relat Res. 2001;388:10–7.
Tarkin IS, Bridgeman JT, Jardon OM, Garvin KL. Successful biologic fixation with mobile-bearing total knee arthroplasty. J Arthroplast. 2005;20(4):481–6.
Watanabe H, Akizuki S, Takizawa T. Survival analysis of a cementless, cruciate-retaining total knee arthroplasty. Clinical and radiographic assessment 10 to 13 years after surgery. J Bone Joint Surg Br. 2004;86(6):824–9.
Whiteside LA. Long-term followup of the bone-ingrowth Ortholoc knee system without a metal-backed patella. Clin Orthop Relat Res. 2001;388:77–84.
Brander VA, Malhotra S, Jet J, et al. Outcome of hip and knee arthroplasty in persons aged 80 years and older. Clin Orthop Relat Res. 1997;345:67–78.
Diduch DR, Insall JN, Scott WN, et al. Total knee replacement in young, active patients. J Bone Joint Surg Am. 1997;79(4):575–82.
Duffy GP, Trousdale RT, Stuart MJ. Total knee arthroplasty in patients 55 years old or younger. 10- to 17-year results. Clin Orthop Relat Res. 1998;356:22–7.
Duffy GP, Crowder AR, Trousdale RR, Berry DJ. Cemented total knee arthroplasty using a modern prosthesis in young patients with osteoarthritis. J Arthroplast. 2007;22(6):67–70.
Gill GS, Joshi AB. Total knee arthroplasty in the young. J Arthroplast. 2004;19(2):255.
Hilton AI, Back DL, Espag MP, et al. The octogenarian total knee arthroplasty. Orthopedics. 2004;27(1):37–9.
Hofmann AA, Heithoff SM, Camargo M. Cementless total knee arthroplasty in patients 50 years or younger. Clin Orthop Relat Res. 2002;404:102–7.
Joshi AB, Markovic L, Gill GS. Knee arthroplasty in octogenarians: results at 10 years. J Arthroplast. 2003;18(3):295–8.
Laskin RS. Total knee replacement in patients older than 85 years. Clin Orthop Relat Res. 1999;367:43–9.
Long WJ, Bryce CD, Hollenbeak CS, Benner RW, Scott WN. Total knee replacement in young, active patients: long-term follow-up and functional outcome: a concise follow-up of a previous report. J Bone Joint Surg Am. 2014;96(18):e159.
Meftah M, Ranawat AS, Sood AB, Rodriguez JA, Ranawat CS. All-polyethylene tibial implant in young, active patients: a concise follow-up, 10 to 18 years. J Arthroplast. 2012;27(1):10–4.
Meftah M, White PB, Ranawat AS, Ranawat CS. Long-term results of total knee arthroplasty in young and active patients with posterior stabilized design. Knee. 2016;23(2):318–21.
Morgan M, Brooks S, Nelson RA. Total knee arthroplasty in young active patients using a highly congruent fully mobile prosthesis. J Arthroplast. 2007;22(4):525–30.
Nilsson KG, Henricson A, Norgren B, Dalen T. Uncemented HA-coated implant is the optimum fixation for TKA in the young patient. Clin Orthop Relat Res. 2006;448:129–39.
Pagnano MW, Levy BA, Berry DJ. Cemented all polyethylene tibial components in patients age 75 years and older. Clin Orthop Relat Res. 1999;367:73–80.
Rubin LE, Blood TD, Defillo-Draiby JC. Total hip and knee arthroplasty in patients older than age 80 years. J Am Acad Orthop Surg. 2016;24(10):683–90.
Tai CC, Cross MJ. Five- to 12-year follow-up of a hydroxyapatite-coated, cementless total knee replacement in young, active patients. J Bone Joint Surg Br. 2006;88(9):1158–63.
Tankersley WS, Hungerford DS. Total knee arthroplasty in the very aged. Clin Orthop Relat Res. 1995;316:45–9.
van der Ven A, Scott RD, Barnes CL. All-polyethylene tibial components in octogenarians: survivorship, performance, and cost. Am J Orthop. 2014;43(1):21–4.
Li S, Burstein AH. Ultra-high molecular weight polyethylene. The material and its use in total joint implants. J Bone Joint Surg Am. 1994;76(7):1080–90.
Tanner MG, Whiteside LA, White SE. Effect of polyethylene quality on wear in total knee arthroplasty. Clin Orthop Relat Res. 1995;317:83–8.
Won CH, Rohatgi S, Kraay MJ, et al. Effect of resin type and manufacturing method on wear of polyethylene tibial components. Clin Orthop Relat Res. 2000;376:161–71.
Wrona M, Mayor MB, Collier JP, et al. The correlation between fusion defects and damage in tibial polyethylene bearings. Clin Orthop Relat Res. 1994;299:92–103.
Sclippa E, Piekarski K. Carbon fiber reinforced polyethylene for possible orthopedic uses. J Biomed Mater Res. 1973;7:59–70.
Champion AR, Li S, Saum K, et al. The effect of crystallinity on the physical properties of UHMWPE. Trans Orthop Res Soc. 1994;19:585.
Busanelli L, Squarzoni S, Brizio L, et al. Wear in carbon fiber-reinforced polyethylene (poly-two) knee prostheses. Chir Organi Mov. 1996;81(3):263–7.
Collier JP, Bargmann LS, Currier BH, et al. An analysis of hylamer and polyethylene bearings from retrieved acetabular components. Orthopedics. 1998;21(8):865–71.
Wright TM, Rimnac CM, Stulberg SD, et al. Wear of polyethylene in total joint replacements. Observations from retrieved PCA knee implants. Clin Orthop Relat Res. 1992;276:126–34.
Collier JP, Mayor MB, McNamara JL, et al. Analysis of the failure of 122 polyethylene inserts from uncemented tibial knee components. Clin Orthop Relat Res. 1991;273:232–42.
Peters PC, Engh GA, Dwyer KA, et al. Osteolysis after total knee arthroplasty without cement. J Bone Joint Surg Am. 1992;74:864–76.
Bohl JR, Bohl WR, Postak PD, et al. The effects of shelf life on clinical outcome for gamma sterilized polyethylene tibial components. Clin Orthop Relat Res. 1999;267:28–38.
Collier JP, Sperling DK, Currier JH, et al. Impact of gamma sterilization on clinical performance of polyethylene in the knee. J Arthroplast. 1996;11:377–89.
Currier BH, Currier JH, Collier JP, et al. Shelf life and in vivo duration: impacts on performance of tibial bearings. Clin Orthop Relat Res. 1997;342:111–22.
Heim CS, Postak PD, Greenwald AS. The influence of shelf storage duration on gamma irradiated UHMWPE tibial components. Orthop Trans. 1998;22:149–50.
Collier JP, Sutula LC, Currier BH, et al. Overview of polyethylene as a bearing material: comparison of sterilization methods. Clin Orthop Relat Res. 1996;333:76–86.
Nusbaum HJ, Rose RM. The effects of radiation sterilization on the properties of ultrahigh molecular weight polyethylene. J Biomed Mater Res. 1979;13:557–76.
Ries MD, Weaver K, Rose RM, et al. Fatigue strength of polyethylene after sterilization by gamma irradiation or ethylene oxide. Clin Orthop Relat Res. 1996;333:87–95.
Rimnac CM, Klein RW, Betts F, et al. Post-irradiation ageing of ultra-high molecular weight polyethylene. J Bone Joint Surg Am. 1994;76:1052–6.
Leibovitz BE, Siegel BV. Aspects of free radical reactions in biological systems: aging. J Gerontol. 1980;35:45–56.
Fisher J, Reeves EA, Isaac GH, et al. Comparison of aged and non-aged ultrahigh molecular weight polyethylene sterilized by gamma irradiation and by gas plasma. J Mater Sci Mater Med. 1997;8:375–8.
Greer KW, Schmidt MB, Hamilton JV. The hip simulator wear of gamma-vacuum, gamma-air, and ethylene oxide sterilized UHMWPE following a severe oxidative challenge. Trans Orthop Res Soc. 1998;23:52.
McGloughlin TM, Kavanagh AG. Wear of ultra-high molecular weight polyethylene (UHMWPE) in total knee prostheses: a review of key influences. Proc Inst Mech Eng. 2000;214:349–59.
Kurtz SM, Muratoglu OK, Evans M, et al. Advances in the processing, sterilization, and crosslinking of ultra-high molecular weight polyethylene for total joint arthroplasty. Biomaterials. 1999;20(18):1659–88.
McKellop H, Shen FW, Lu B, et al. Development of an extremely wear-resistant ultra high molecular weight polyethylene for total hip replacements. J Orthop Res. 1999;17(2):157–67.
Muratoglu OK, Bragdon CR, O’Connor DO, et al. A novel method of cross-linking ultra-high-molecular weight polyethylene to improve wear, reduce oxidation, and retain mechanical properties. J Arthroplast. 2001;16(2):149–60.
Oonishi H, Kadoya Y, Masuda S. Gamma-irradiated cross-linked polyethylene in total hip replacements—analysis of retrieved sockets after long-term implantation. J Biomed Mater Res. 2001;58(2):167–71.
Oonishi H, Kadoya Y. Wear of high-dose gamma-irradiated polyethylene in total hip replacements. J Orthop Sci. 2000;5(3):223–8.
Oonishi H, Clarke IC, Masuda S, et al. Study of retrieved acetabular sockets made from high-dose, cross-linked polyethylene. J Arthroplast. 2001;16(8):129–33.
Oonishi H, Clarke IC, Yamamoto K, et al. Assessment of wear in extensively irradiated UHMWPE cups in simulator studies. J Biomed Mater Res. 2004;68(1):52–60.
Grobbelaar CJ, de Plessis TA, Marais F. The radiation improvement of polyethylene prostheses. A preliminary study. J Bone Joint Surg Br. 1978;60-B(3):370–4.
Yamamoto K, Masaoka T, Manaka M, et al. Micro-wear features on unique 100-Mrad cups: two retrieved cups compared to hip simulator wear study. Acta Orthop Scand. 2004;75(2):134–41.
Wroblewski BM, Siney PD, Dowson D, et al. Prospective clinical and joint simulator studies of a new total hip arthroplasty using alumina ceramic heads and cross-linked polyethylene cups. J Bone Joint Surg Br. 1996;78(2):280–5.
Baker DA, Hastings RS, Pruitt L. Study of fatigue resistance of chemical and radiation crosslinked medical grade ultrahigh molecular weight polyethylene. J Biomed Mater Res. 1999;46(4):573–81.
Tradonsky S, Postak PD, Froimson AI, et al. A comparison of disassociation strength of modular acetabular components. Clin Orthop Relat Res. 1993;296:154–60.
Halley D, Glassman A, Crowninshield RD. Recurrent dislocation after revision total hip replacement with a large prosthetic femoral head. J Bone Joint Surg Am. 2004;86(4):827–30.
Ries MD. Dissociation of an ultra-high molecular weight polyethylene insert from the tibial baseplate after total knee arthroplasty. J Bone Joint Surg Am. 2004;86(7):1522–4.
Digas G, Karrholm J, Thanner J, et al. The Otto Aufranc award. Highly cross-lined polyethylene in total hip arthroplasty: randomized evaluation of penetration rate in cemented and uncemented sockets using radiosterometric analysis. Clin Orthop Relat Res. 2004;429:6–16.
Heisel C, Silva M, dela Rosa MA, et al. Short-term in vivo wear of cross-linked polyethylene. J Bone Joint Surg Am. 2004;86(4):748–51.
Hopper RH Jr, Young AM, Orishimo KF, et al. Correlation between early and late wear rates in total hip arthroplasty with application to the performance of Marathon cross-linked polyethylene liners. J Arthroplast. 2003;18(7 Suppl 1):60–7.
Martell JM, Verner JJ, Incavo SJ. Clinical performance of a highly cross-linked polyethylene at two year in total hip arthroplasty: a randomized prospective trial. J Arthroplast. 2003;18(7 Suppl 1):55–9.
Muratoglu OK, Ruberti J, Melotti S, et al. Optical analysis of surface changes on early retrieval of highly cross-linked and conventional polyethylene tibial inserts. J Arthroplast. 2003;18(7 Suppl 1):42–7.
Sychterz CJ, Orishimo KF, Engh CA. Sterilization and polyethylene wear: clinical studies to support laboratory data. J Bone Joint Surg Am. 2004;86(5):1017–22.
Bragdon CR, Jasty M, Muratoglu OK, et al. Third-body wear of highly cross-linked polyethylene in a hip simulator. J Arthroplast. 2003;18(5):553–61.
McEwen HMJ, Farrar R, Auger DD, et al. Reduction of wear in fixed bearing total knee replacement using crosslinked UHMWPE. Trans Orthop Res Soc. 2003;28(2):1428.
Kurtz SM, Pruitt LA, Jewett CW, et al. Radiation and chemical crosslinking promote strain hardening behavior and molecular alignment in ultra high molecular weight polyethylene during multi-axial loading conditions. Biomaterials. 1999;20(16):1449–62.
Muratoglu OK, Bragdon CR, O’Connor DO, et al. Aggressive wear testing of a cross-linked polyethylene in total knee arthroplasty. Clin Orthop Relat Res. 2002;404:89–95.
Muratoglu OK, Merrill EW, Bragdon CR, et al. Effects of radiation, heat and aging on in vitro wear resistance of polyethylene. Clin Orthop Relat Res. 2003;417:253–62.
Bradford L, Baker DA, Graham J, et al. Wear and surface cracking in early retrieved highly cross-linked polyethylene acetabular liners. J Bone Joint Surg Am. 2004;86(6):1271–82.
Bradford L, Kurland R, Sankaran M, et al. Early fallure due to osteolysis associated with contemporary highly cross-linked ultra-high molecular weight polyethylene. A case report. J Bone Joint Surg Am. 2004;86(5):1051–6.
Blunn GW, Joshi AB, Minns RJ, et al. Wear in retrieved condylar knee arthroplasties. J Arthroplast. 1997;12(3):281–90.
Jasty M, Goetz DD, Bragdon CR, et al. Wear of polyethylene acetabular components in total hip arthroplasty. An analysis of one hundred and twenty-eight components retrieved at autopsy or revision operations. J Bone Joint Surg Am. 1997;79(3):349–58.
Landy MM, Walker PS. Wear of ultra-high-molecular-weight polyethylene components of 90 retrieved knee prostheses. J Arthroplast. 1988;3(Suppl):S73–85.
Shibata N, Tomita N. The anti-oxidative properties of alpha-tocopherol in gamma-irradiated UHMWPE with respect to fatigue and oxidation resistance. Biomaterials. 2005;26(29):5755–62.
Ries MD, Pruitt L. Effect of cross-linking on the microstructure and mechanical properties of ultra-high molecular weight polyethylene. Clin Orthop Relat Res. 2005;440:149–56.
Teramure S, Sakoda H, Terao T, Endo MM, Fujiwara K, Tomita N. Reduction of wear volume from ultrahigh molecular weight polyethylene knee components by the addition of vitamin E. J Orthop Res. 2008;26(4):460–4.
Oral E, Malhi AS, Wannomae KK, Muratoglu OK. Highly cross-linked ultrahigh molecular weight polyethylene with improved fatigue resistance for total joint arthroplasty: recipient of the 2006 Hap Paul award. J Arthroplast. 2008;23(7):1037–44.
Wannomae KK, Christensen SD, Micheli BR, Rowell SL, Schroeder DW, Muratoglu OK. Delamination and adhesive wear behavior of alpha-tocopherol-stabilized irradiated ultrahigh-molecular weight polyethylene. J Arthroplast. 2010;25(4):635–43.
Vaidya C, Alvarez E, Vinciguerra J, Bruce DA, DesJardin JD. Reduction of total knee replacement wear with vitamin E blended highly cross-linked ultra-high molecular weight polyethylene. Proc Inst Mech Eng H. 2011;225(10):1–7.
Micheli BR, Wannomae KK, Lozynsky AJ, Christensen SD, Muratoglu OK. Knee simulator wear of vitamin E stabilized irradiated ultrahigh molecular weight polyethylene. J Arthroplast. 2012;27(1):95–104.
Haider H, Weisenburger JN, Kurtz SM, Rimnac CM, Freedman J, Schroeder DW, Garvin KL. Does vitamin E-stabilized ultrahigh-molecular-weight polyethylene address concerns of cross-linked polyethylene in total knee arthroplasty? J Arthroplast. 2012;27(3):461–9.
Bitsch RG, Loudolt T, Heisel C, Ball S, Schmalzried TP. Reduction of osteolysis with use of Marathon cross-linked polyethylene: a concise follow-up, at a minimum of five years, of a previous report. J Bone Joint Surg Am. 2008;90(7):1487–91.
Dorr LD, Wan Z, Shahrdar C, Sirianni L, Boutary M, Yun A. Clinical performance of a Durasul highly cross-linked polyethylene acetabular liner for total hip arthroplasty at five years. J Bone Joint Surg Am. 2005;87(8):1816–21.
Engh CA Jr, Hopper RH Jr, Huynh C, Ho H, Sritulanondha S, Engh CA Sr. A prospective, randomized study of cross-linked and non-cross-linked polyethylene for total hip arthroplasty at 10-year follow-up. J Arthroplast. 2012;27(8):2–7.e1.
Flatoy B, Rydinge J, Dahl J, Rohrl SM, Nordsletten L. Low wear, high stability—promises of success in a moderately cross-linked cup? Hip Int. 2015;25(3):199–203.
Garcai-Rey E, Garcia-Cimbrela E, Cruz-Pardos A. New polyethylenes in total hip replacement: a ten- to 12-year follow-up study. Bone Joint J. 2013;95(3):326–32.
Greiner JJ, Callaghan JJ, Bedard NA, Liu SS, Gao Y, Goetz DD. Fixation and wear with contemporary acetabular components and cross-linked polyethylene at 10-years in patients aged 50 and under. J Arthroplast. 2015;30(9):1577–85.
Hanna SA, Somerville L, McCalden RW, Naudie DD, MacDonald SJ. Highly cross-linked polyethylene decreases the rate of revision of total hip arthroplasty compared with conventional polyethylene at 13 years follow-up. Bone Joint J. 2016;98-B(1):28–32.
Kurtz SM, Medel FJ, MacDonald DW, Parvizi J, Kraay MJ, Rimnac CM. Reasons for revision of first-generation highly cross-linked polyethylenes. J Arthroplast. 2010;25(6):67–74.
Lachiewicz PF, Soileau ES, Martell JM. Wear and osteolysis of highly cross-linked polyethylene at 10 to 14 years. The effect of femoral head size. Clin Orthop Relat Res. 2016;474(2):365–71.
Mall NA, Nunley RM, Zhu JJ, Maloney WJ, Barrack RL, Clohisy JC. The incidence of acetabular osteolysis in young patients with conventional versus highly crosslinked polyethylene. Clin Orthop Relat Res. 2011;469(2):372–81.
Nebergall AK, Greene ME, Rubash H, Malchau H, Troelsen A, Rolfson O. Thirteen-year evaluation of highly cross-linked polyethylene articulating with either 28-mm or 36-mm femoral heads using radiostereometric analysis and computerized tomography. J Arthroplast. 2016;31(9):269–76.
Ranawat CS, Ranawat AS, Ramteke AA, Nawabi D, Meftah M. Long-term results of a first-generation annealed highly cross-linked polyethylene in young active patients. Orthopedics. 2016;39(2):e225–9.
Reynolds SE, Malkani AL, Ramakrishnan R, Yakkanti MR. Wear analysis of first-generation highly cross-linked polyethylene in primary total hip arthroplasty: an average 9-year follow-up. J Arthroplast. 2012;27(6):1064–8.
Samujh C, Bhimani S, Smith L, Malkani AL. Wear analysis of second-generation highly cross-linked polyethylene in primary total hip arthroplasty. Orthopedics. 2016;9:1–5.
Surace MF, Monestier L, Vulcano E, Harwin SF, Cherubino P. Conventional versus cross-linked polyethylene for total hip arthroplasty. Orthopedics. 2015;38(9):556–61.
Thomas GE, Simpson DJ, Mehmood S, Taylor A, McLardy-Smith P, Gill HS, Murray DW, Glyn-Jones S. The seven-year wear of highly cross-linked polyethylene in total hip arthroplasty: a double-blind randomized controlled trial using radiostereometric analsysis. J Bone Joint Surg Am. 2011;93(8):716–22.
Hodrick JT, Severson EP, McAlister DS, Dahl B, Hofmann AA. Highly crosslinked polyethylene is safe for use in total knee arthroplasty. Clin Orthop Relat Res. 2008;466(11):2806–12.
Meneghini RM, Ireland PH, Bhowmik-Stoker M. Multicenter study of highly cross-linked vs. conventional polyethylene in total knee arthroplasty. J Arthroplast. 2016;31(4):809–14.
Meneghini RM, Lovro LR, Smits SA, Ireland PH. Highly cross-linked versus conventional polyethylene in posterior-stabilized total knee arthroplasty at a mean 5-year follow-up. J Arthroplast. 2015;30(10):1736–9.
Minoda Y, Aihara M, Sakawa A, Fukuoka S, Hayakawa K, Tomita M, Umeda N, Ohzono K. Comparison between highly cross-linked and conventional polyethylene in total knee arthroplasty. Knee. 2009;16(5):348–51.
Paxton EW, Inacio MC, Kurtz S, Love R, Cafri G, Namba RS. Is there a difference in total knee arthroplasty risk of revision in highly crosslinked versus conventional polyethylene? Clin Orthop Relat Res. 2015;473(3):999–1008.
Callary SA, Field JR, Campbell DG. Low wear of a second-generation highly crosslinked polyethylene liner: a 5-year radiostereometric analysis study. Clin Orthop Relat Res. 2013;471(11):3596–600.
Nebergall AK, Greene ME, Laursen MB, Nielsen PT, Malchau H, Troelsen A. Vitamin E diffused highly cross-linked polyethylene in total hip arthroplasty at five years: a randomised controlled trial using radiostereometric analysis. Bone Joint J. 2017;99-B(5):577–84.
Salemyr M, Muren O, Ahl T, Bodén H, Chammout G, Stark A, Sköldenberg O. Vitamin-E diffused highly cross-linked polyethylene liner compared to standard liners in total hip arthroplasty. A randomized, controlled trial. Int Orthop. 2015;39(8):1499–505.
Scemama C, Anract P, Dumaine V, Babinet A, Courpied JP, Hamadouche M. Does vitamin E-blended polyethylene reduce wear in primary total hip arthroplasty: a blinded randomised clinical trial. Int Orthop. 2017;41(6):1113–8.
Shareghi B, Johanson PE, Kärrholm J. Wear of vitamin E-infused highly cross-linked polyethylene at five years. J Bone Joint Surg Am. 2017;99(17):1447–52.
Flament EM, Berend KR, Hurst JM, Morris MJ, Adams JB, Lombardi AV Jr. Early experience with vitamin E antioxidant-infused highly cross-linked polyethylene inserts in primary total knee arthroplasty. Surg Technol Int. 2016;29:334–40.
Sakellarious VI, Sculco P, Poultsides L, Wright T, Sculco TP. Highly cross-linked polyethylene may not have an advantage in total knee arthroplasty. HSS J. 2013;9(3):264–9.
Sheth NP, Lementowski P, Hunter G, Garino JP. Clinical applications of oxidized zirconium. J Surg Orthop Adv. 2008;17(1):17–26.
Ezzet KA, Hermida JC, Colwell CW Jr, D’Lima DD. Oxidized zirconium femoral components reduce polyethylene wear in a knee simulator. Clin Orthop Relat Res. 2004;428:120–4.
Tsukamoto R, Chen S, Asano T, Ogino M, Shoji H, Nakamura T, Clarke IC. Improved wear performance with crosslinked UHMWPE and zirconia implants in knee simulation. Acta Orthop. 2006;77(3):505–11.
Gascoyne TC, Teeter MG, Guenther LE, Burnell CD, Bohm ER, Naudie DR. Vivo wear performance of cobalt-chromium versus oxidized zirconium femoral total knee replacements. J Arthroplast. 2016;31(1):137–41.
Hofer JK, Ezzet KAA. Minimum 5-year follow-up of an oxidized zirconium femoral prosthesis used for total knee arthroplasty. Knee. 2014;21(1):168–71.
Hui C, Salmon L, Maeno S, Roe J, Walsh W, Pinczewski L. Five-year comparison of oxidized zirconium and cobalt-chromium femoral components in total knee arthroplasty: a randomized controlled trial. J Bone Joint Surg Am. 2011;93(7):624–30.
Inacio MC, Cafri G, Paxton EW, Kurtz SM, Namba RS. Alternative bearing in total knee arthroplasty: risk of early revision compared to traditional bearings: an analysis of 62,177 primary cases. Acta Orthop. 2013;84(2):145–52.
Innocenti M, Matassi F, Carulli C, Nistri L, Civinini R. Oxidized zirconium femoral component for TKA: a follow-up note of a previous report at a minimum of 10 years. Knee. 2014;21(4):858–61.
Innocenti M, Carulli C, Matassi F, Carossino AM, Brandi ML, Civinini R. Total knee arthroplasty in patients with hypersensitivity to metals. Int Orthop. 2014;38(2):329–33.
Heim CS, Postak PD, Greenwald AS. Factors influencing the longevity of UHMWPE tibial components. In: Pritchard D, editor. Instructional Course Lectures, vol. 45. Chicago, IL: American Academy of Orthopaedic Surgeons; 1996.
Morra EA, Postak PD, Greenwald AS. Surface stress: a factor influencing ultra-high molecular weight polyethylene durability in mobile-bearing knee design. Orthopedics. 2007;30(5 Suppl):35–8.
Ries MD, Scott ML, Jani S. Relationship between gravimetric wear and particle generation in hip simulator: conventional compared with cross-linked polyethylene. J Bone Joint Surg Am. 2001;83(Suppl 2, Pt 2):116–22.
Ingram JH, Fisher J, Stone M, et al. Effect of crosslinking on biological activity of UHMWPE wear debris. Trans Orthop Res Soc. 2003;28(2):1439.
Scott ML, Ries MD, Jani S. Abrasive wear in total hip replacement: is crosslinked UHMWPE coupled to ceramic heads the answer? Proc Am Acad Orthop Surg. 2002;3:732.
Muratoglu OK, Burroughs BR, Christensen SD. In vitro simulator wear of highly crosslinked tibias articulating against explanted rough femoral components. Trans Orthop Res Soc. 2004;29(1):297.
Good V, Ries M, Barrack RL, et al. Reduced wear with oxidized zirconium femoral heads. J Bone Joint Surg Am. 2003;85(Suppl 4):105–10.
Heyse TJ, Elpers ME, Nawabi DH, Wright TM, Haas SB. Oxidized zirconium versus cobalt-chromium in TKA: profilometry of retrieved femoral components. Clin Orthop Relat Res. 2014;472(1):277–83.
Laskin RS. An oxidized Zr ceramic surfaced femoral component for total knee arthroplasty. Clin Orthop Relat Res. 2003;416:191–6.
Minoda Y, Hata K, Iwaki H, Ikebuchi M, Hashimoto Y, Inori F, Nakamura H. No difference in in vivo polyethylene wear particles between oxidized zirconium and cobalt-chromium femoral component in total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc. 2014;22(3):680–6.
Lavernia CJ, Sierra RJ, Hungerford DS, et al. Activity level and wear in total knee arthroplasty. J Arthroplast. 2001;16(4):446–53.
Mont MA, Rajadhyaksha AS, Marxen JL, et al. Tennis after total knee arthroplasty. Am J Sports Med. 2002;30(2):163–6.
Seedhom BB, Wallbridge NC. Walking activities and wear of prostheses. Ann Rheum Dis. 1985;44:838.
Benjamin J, Tucker T, Ballesteros PI. Obesity a contraindication to bilateral total knee arthroplasties under one anesthetic? Clin Orthop Relat Res. 2001;392:190–5.
Deshmukh RG, Hayes JH, Pinder IM. Does body weight influence outcome after total knee arthroplasty? A 1-year analysis. J Arthroplast. 2002;17(3):315–9.
Griffin FM, Scuderi GR, Insall JN, et al. Total knee arthroplasty in patient who were obese with 1 years followup. Clin Orthop Relat Res. 1998;356:28–3.
Mont MA, Mathur SK, Krackow KA, et al. Cementless total knee arthroplasty in obese patients. A comparison with a matched control group. J Arthroplast. 1996;11(2):153–6.
Spicer DD, Pomeroy DL, Badenhausen WE, et al. Body mass index as a predictor of outcome in total knee replacement. Int Orthop. 2001;25(4):246–9.
Wendelboe AM, Hegmann KT, Biggs JJ, et al. Relationships between body mass indices and surgical replacement of knee and hip joints. Am J Prev Med. 2003;25(4):290–5.
Vazquez-Vela Johnson G, Worland RL, Keenan J, et al. Patient demographics as a predictor of the ten-year survival rate in primary total knee replacement. J Bone Joint Surg Br. 2003;85(1):52–6.
Blunn GW, Walker PS, Joshi A, et al. The dominance of cyclic sliding in producing wear in total knee replacements. Clin Orthop Relat Res. 1991;273:253–60.
Kawanabe K, Clarke IC, Tamura J, et al. Effects of A-P translation and rotation on the wear of UHMWPE in a total knee joint simulator. J Biomed Mater Res. 2001;54(3):400–6.
Rose RM, Goldfarb HV. On the pressure dependence of the wear of ultrahigh molecular weight polyethylene. Wear. 1983;92:99–111.
Rostoker W, Galante JO. Contact pressure dependence of wear rates of ultra high molecular weight polyethylene. J Biomed Mater Res. 1979;12:957–64.
Sathasivam S, Walker PSA. Computer model with surface friction for the prediction of total knee kinematics. J Biomech. 1996;30(2):177–84.
Szivek JA, Anderson PL, Benjamin JB. Average and peak contact stress distribution evaluation of total knee arthroplasties. J Arthroplast. 1996;11(8):952–63.
Wimmer MA, Andriacchi TP. Tractive forces during rolling motion of the knee: implications for wear in total knee replacement. J Biomech. 1996;30(2):131–7.
Wimmer MA, Andriacchi TP, Natarajan RN, et al. A striated pattern of wear in ultrahigh-molecular-weight polyethylene components of Miller-Galante total knee arthroplasty. J Arthroplast. 1998;13(1):8–16.
Booth RE Jr. Total knee arthroplasty in the obese patient: tips and quips. J Arthroplasty. 2002;17(4 Suppl 1):69–70.
Robertson NB, Battenberg AK, Kertzner M, Schmalzried TP. Defining high activity in arthroplasty patients. Bone Joint J. 2016;98(1 Suppl A):95–7.
Issa K, Jauregui JJ, Given K, Harwin SF, Mont MA. A prospective, longitudinal study of patient activity levels following total knee arthroplasty stratified by demographic and comorbid factors. J Knee Surg. 2015;28(4):343–7.
Ledford CK, Millikan PD, Nickel BT, Green CL, Attarian DE, Wellman SS, Bolognesi MP, Queen RM. Percent body fat is more predictive of function after total joint arthroplasty than body mass index. J Bone Joint Surg Am. 2016;98(10):849–57.
Paxton EW, Torres A, Love RM, Barber TC, Sheth DS, Inacio MC. Total joint replacement: a multiple risk factor analysis of physical activity level 1-2 years postoperatively. Acta Orthop. 2016;87:44–9.
Apkarian J, Naumann S, Cairns BA. Three-dimensional kinematic and dynamic model of the lower limb. J Biomech. 1989;22:143–55.
LaFortune MA, Cavanaugh PR, Sommer HJ, et al. Three-dimensional kinematics of the human knee during walking. J Biomech. 1992;25:347–57.
Morrison JB. Function of the knee joint in various activities. Biomed Eng. 1969;4:573–80.
Murray MP, Drought AB, Kory RC. Walking patterns in normal men. J Bone Joint Surg Am. 1964;46:335–60.
Paul JP. Forces transmitted by joints in the human body. Proc Inst Mech Eng. 1967;181:358.
Cameron HU. Tibial component wear in total knee replacement. Clin Orthop Relat Res. 1994;309:29–32.
Eckhoff DG, Metzger RG, Vedewalle MV. Malrotation associated with implant alignment technique in total knee arthroplasty. Clin Orthop Relat Res. 1995;321:28–31.
Fehring TK. Rotational malalignment of the femoral component in total knee arthroplasty. Clin Orthop Relat Res. 2000;380:72–9.
Lewis P, Rorabeck CH, Bourne RB, et al. Posteromedial tibial polyethylene failure in total knee replacements. Clin Orthop Relat Res. 1994;299:11–7.
Swany MR, Scott RD. Posterior polyethylene wear in posterior cruciate ligament-retaining total knee arthroplasty: a case study. J Arthroplast. 1993;8:439–46.
Dennis DA, Komistek RD, Cheal EJ, et al. Vivo femoral condylar lift-off in total knee arthroplasty. Orthop Trans. 1997;21:1112.
Morra EA, Postak PD, Plaxton NA, et al. The effects of external torque on polyethylene tibial insert damage patterns. Clin Orthop Relat Res. 2003;410:90–100.
Baumbach JA, Willburger R, Haaker R, Dittrich M, Kohler S. 10-year survival of navigated versus conventional TKAs: a retrospective study. Orthopedics. 2016;39(3):S72–6.
de Steiger RN, Liu YL, Graces SE. Computer navigation for total knee arthroplasty reduces revision rate for patients less than sixty-five years of age. J Bone Joint Surg Am. 2015;97(8):635–42.
Gustke KA, Golladay GJ, Roche MW, Jerry GJ, Elson LC, Anderson CR. Increased satisfaction after total knee replacement using sensor-guided technology. Bone Joint J. 2014;96(1):1333–8.
Miyasaka T, Kurosaka D, Omori T, Ikeda R, Marumo K. Accuracy of computed tomography-based navigation-assisted total knee arthroplasty: outlier analysis. J Arthroplast. 2016. [Epub ahead of print].
Nodzo SR, Carroll KM, Mayman DJ. Disposable navigation for total knee arthroplasty. Am J Orthop. 2016;45(4):240–5.
Picard F, Deep K, Jenny JY. Current state of the art in total knee arthroplasty computer navigation. Knee Surg Sports Traumatol Arthrosc. 2016;24(11):3565–74.
Todesca A, Garro L, Penna M, Bejui-Hugues J. Conventional versus computer-navigated TKA: a prospective randomized study. Knee Surg Sports Traumatol Arthrosc. 2016. [Epud ahead of print].
Dyrhovden GS, Fenstad AM, Furnes O, Gothesen O. Survivorship and relative risk of revision in computer-navigated versus conventional total knee replacement at 8-year follow-up. Acta Orthop. 2016;24:1–8.
Gharaibeh MA, Solaya GN, Harris IA, Chen DB, MacDessi SJ. Accelerometer-based, portable navigation (KneeAlign) vs. conventional instrumentation for total knee arthroplasty: a prospective randomized comparative trial. J Arthroplast. 2017;32(3):777–82.
Steinhaus ME, McLawhorn AS, Richardson SS, Maher P, Mayman DJ. Handheld navigation device and patient-specific cutting guides result in similar coronal alignment for primary total knee arthroplasty: a retrospective matched cohort study. HSS J. 2016;12(3):224–34.
Author information
Authors and Affiliations
Corresponding author
Editor information
Editors and Affiliations
Rights and permissions
Copyright information
© 2018 Springer International Publishing AG
About this chapter
Cite this chapter
Heim, C.S., Greenwald, A.S. (2018). Implant Bearings in Total Knee Arthroplasty. In: Bono, J., Scott, R. (eds) Revision Total Knee Arthroplasty. Springer, Cham. https://doi.org/10.1007/978-3-319-67344-8_2
Download citation
DOI: https://doi.org/10.1007/978-3-319-67344-8_2
Published:
Publisher Name: Springer, Cham
Print ISBN: 978-3-319-67342-4
Online ISBN: 978-3-319-67344-8
eBook Packages: MedicineMedicine (R0)