Abstract
Purpose
N-butyl cyanoacrylate (NBCA) glue is a valuable liquid embolic agent for the endovascular treatment of brain arteriovenous malformations (BAVM). The use of NBCA carries a risk of embolic agent reflux and distal migration with ensuing possibility of venous drainage compromise. The aim of this technical note is to describe a single center case series of high-flow intranidal fistulas embolized though a dual lumen balloon under local flow-arrest conditions.
Methods
This article presents a retrospective description of a case series including five NBCA injections through a dual lumen balloon performed by a single operator in three patients with BAVM. Demographic, clinical, imaging and procedure-related data are reported.
Results
The three patients presented with ruptured BAVMs. Of the patients, one underwent proximal flow-related ruptured aneurysm coiling before planned BAVM embolization. In the three patients, staged BAVM embolization as a first line treatment was decided. Preliminary embolization of high-flow BAVM fistulas by NBCA using a dual lumen balloon under local flow-arrest conditions was performed and five separate injections were carried out without any complications. No balloon entrapment or rupture was observed.
Conclusion
High-flow BAVM fistulas can be treated with NBCA embolization using a dual lumen balloon under local flow-arrest conditions.
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Avoid common mistakes on your manuscript.
Introduction
Brain arteriovenous malformations (BAVM) are some of the most challenging cerebrovascular disorders to manage. Rupture is the major source of morbidity and mortality [1]. Only complete and definitive nidus obliteration protects against BAVM-related hemorrhage [2]. Several strategies are available to achieve this goal, including surgical excision, stereotactic radiosurgery, embolization or a combination of these different treatment options. The choice of the treatment modality depends on a variety of factors including BAVM angioarchitecture, presentation and surgeon experience. Since the first description of BAVM embolization with methylmethacrylate particles through a carotid arteriotomy in the 1960s [3], great progress and refinement have been achieved in the field of interventional neuroradiology. Embolization can be performed as a sole treatment with a curative goal, as an adjunct before surgery or radiosurgery for nidal reduction or to target specific angioarchitectural features, such as flow-related or nidal aneurysms [4]. Embolization of high-flow intranidal fistulas can be particularly challenging. Several embolic agents have been used to embolize high-flow fistulas, including detachable balloons, glue and coils [5,6,7,8]. For several decades now, acrylic glue, originally isobutyl 2‑cyanoacrylate (IBCA) and subsequently N‑butyl cyanoacrylate (NBCA), has been widely used for BAVM embolization [9,10,11]. Embolization of NBCA carries a number of technical difficulties including risk of reflux, proximal embolization, glue cast fragmentation, microcatheter rupture, microcatheter adhesion and glue adhesion during microcatheter retrieval [12,13,14]. Distal migration of the embolic agent into the venous compartment of the BAVM prior to nidal occlusion can lead to significant hemorrhagic complications [5]. The risk of distal migration is dramatically increased in patients with high-flow intranidal fistulas. Dual lumen balloons have been used for direct coiling and injection of liquid embolic agents such as Onyx (Medtronic, Dublin, Ireland) [15,16,17,18,19]. Onyx and related liquid embolic agents have gained more popularity in recent times but glue injection is still relevant in a number of situations. The aim of this study was to describe a single center case series of high-flow intranidal fistulas embolized though a dual lumen balloon under local flow-arrest conditions.
Methods
Ethical Statement
This work adheres to the World Medical Association Declaration of Helsinki.
Patients and Embolization Procedures
Single center retrospective records review of patients with BAVM treated by embolization between 1 January 2015 and 30 April 2018, was performed. During this time frame, a single senior operator (NS) performed 6 NBCA injections in 4 patients using the technique described. Of the patients, two gave consent for publication and a third patient was deceased at the time of manuscript preparation. Data regarding the 4th patient who refused permission for publication are not included in this report. The three patients included presented with ruptured BAVM. The decision to embolize was taken by a multidisciplinary team including interventional neuroradiologists, neurosurgeons, and anesthesiologists based on clinical and radiological data, such as computed tomography (CT) scan, magnetic resonance imaging (MRI), digital subtraction angiography (DSA) and 3D rotational angiography. Staged embolization was planned. Treatment was delayed after rehabilitation according to standard practice at this institution. All procedures were performed in a biplane angio suite with the patient under general anesthesia with intravenous heparinization. A 6 French Envoy (Codman, Reynham MA, USA) guiding catheter was introduced through a 6 French femoral access. The guiding catheter was placed distally in the cervical vertebral artery or internal carotid artery. A long nose Eclipse 2L (Balt, Montmorency, France) 6/9 mm hypercompliant dual-lumen balloon microcatheter was navigated over a 0.014-inch microguide wire as closely as possible to the shunting point of the high-flow fistula. After balloon inflation with a 1:1 mixture of saline and contrast medium solution, a contrast injection through the balloon working lumen was performed to analyze the downstream vascular bed. The working lumen dead space was rinsed with 5% dextrose. A 1:2 NBCA Glubran 2 (GEM, Viareggio, Italy) and lipiodol (Guerbet, Villepinte, France) solution was subsequently injected under flow-arrest via balloon inflation. In cases of NBCA reflux along the distal tip of the dual lumen balloon, which was observed even with balloon inflation, the balloon was temporarily deflated, then reinflated to continue with the glue injection. Deflation was permitted by the fact that glue injection under flow-arrest is slower that under free-flow, similar to injection under wedge conditions.
Results
A total of three patients (2 female and 1 male) were included in the study. The clinical and demographic features of the patients are summarized in Table 1. In all cases, the balloon could be navigated over a 0.014-inch Traxcess microguide wire (Microvention, Aliso Viejo, CA, USA) through an arterial feeder close to the shunt point. A total of five balloon-aided NBCA injections into high-flow arteriovenous fistulas were performed. In all cases, balloon inflation allowed local flow-arrest. All injections through the dual lumen balloon led to definitive occlusion of the intranidal high-flow arteriovenous fistula. No distal migration threatening the venous outflow occurred during the five injections under flow-arrest conditions. No microcatheter entrapment or balloon rupture occurred. All five injections and related procedures were uneventful. Figs. 1 and 2 demonstrate NBCA injections through a 6/9 mm Eclipse dual lumen balloon. A long nose balloon microcatheter (7 mm distal tip beyond the balloon) was chosen to decrease the risk of NBCA reflux and contact with the balloon surface.
Patient 1 was a 70-year-old male, admitted for spontaneous intraventricular hemorrhage related to a left frontal Spetzler-Martin III BAVM draining in a deep venous varix (Fig. 1). It was decided to treat the ruptured BAVM by embolization after rehabilitation but early intraventricular rebleeding on day 24 prompted rapid intervention. The intervention allowed successful high-flow fistula occlusion and near complete BAVM obliteration and disappearance of the large varicose draining vein. The patient died a few months later of an unrelated cause.
Patient 2 was a 46-year-old female with a right parietooccipital Spetzler-Martin IV BAVM, admitted for spontaneous subarachnoid hemorrhage related to a proximal basilar flow-related aneurysm rupture. The aneurysm was treated by coiling in the acute phase. After rehabilitation, the staged BAVM embolization was decided. High-flow fistulas were successfully targeted by embolization as demonstrated in Fig. 2 with final control showing residual shunting. A third embolization session by standard flow-guided NBCA injection was complicated by distal migration of the embolic material, venous drainage compromise and subsequent postprocedural hematoma. Further treatment will be discussed after rehabilitation.
Patient 3 was a 37-year-old female with a left temporal Spetzler-Martin III BAVM, admitted for isolated intraventricular hemorrhage. After rehabilitation, the patient underwent a first embolization session allowing successful occlusion of high-flow fistulas. A later embolization session, using standard flow-guided embolization was complicated by NBCA reflux in the left posterior cerebral artery, leading to an occipital infarction with symptomatic hemianopia. The residual nidus was treated by stereotactic radiosurgery and the patient is awaiting MRI control.
Discussion
This case series supports the feasibility of embolization of high-flow intranidal fistulas with NBCA using a dual lumen balloon under local flow-arrest conditions. The use of balloons in endovascular procedures has a longstanding history [20]. New generation dual lumen balloons have one lumen dedicated to balloon inflation and a second dimethyl sulfoxide (DMSO) compatible, termed the working lumen, dedicated to the microguide wire. A side benefit is that the working lumen can be used for stent or coil deployment [15, 19] and liquid embolic agent injection [15,16,17,18]. Over the last few years Onyx embolization of BAVMs, spinal AVMs, cervical AVMs, dural arteriovenous fistulas and aneurysms using dual lumen balloons has been extensively described in the literature [15, 17, 18, 21, 22]. The injection of NBCA is a valuable technique for BAVM embolization [9, 10]; however, flow-dependent injection is limited by poor control over NBCA distribution and progression. Several parameters including microcatheter design, blood flow, blood vessel caliber, temperature and operator experience determine the success of the procedure [12]. The NBCA is usually mixed with iodized oil which imparts radio-opacity to the embolic mixture and thus greatly enhances control over injection [23]. Varying the ratio between the two components modifies polymerization dynamics with an increased iodized oil concentration generating a buffer that delays contact with blood and increases polymerization time [24]. The NBCA can also be injected without iodized oil, further accelerating polymerization. In such cases, tantalum powder is added for radio-opacity [25]. Using highly concentrated NBCA in high-flow fistulas has the drawback of increasing the risk of microcatheter entrapment [26]. This seems less likely when using hydrophilically coated microcatheters [27]. Alternatively, distal progression and penetration of NBCA can be enhanced by continuous flushing of the guiding catheter with dextrose 5% solution during NBCA injection, limiting the risk of proximal occlusion [28]. Also, because an increasing iodized oil concentration also increases viscosity, adjunction of titrated quantities of acetic acid to the glue mixture has been proposed to prolong polymerization time without increasing glue viscosity [23]. Increased control over NBCA progression into high-flow arteriovenous shunts can sometimes be obtained by inducing hypotension to limit distal embolization [26]. An interesting alternative to flow-dependent embolization is the wedging technique where the microcatheter is advanced to such a distal position that it occludes the arterial lumen [29, 30]. The arterial pedicle beyond the microcatheter becomes a virtual extension of the lumen and facilitates both delivery of the liquid embolic agent across the fistula site in flow-arrest conditions and permeation of the perifistulous collateral network [12]; however, it is not always possible to navigate the microcatheter to reach such a distal position. Additionally, arterial pedicles of high-flow fistulas may be too large to wedge a microcatheter.
Utilizing an inflatable balloon for flow-arrest aided NBCA injection is a relatively old idea. In 1979 Kerber et al. reported using a single lumen balloon with a custom-made distal valve, allowing balloon inflation for flow-guided navigation and subsequent flow control [9]. Additional pressure by the operator then delivered the embolic agent into the BAVM through the single lumen balloon. More recently, several authors have reported the use of single lumen balloons for proximal high-flow control, subsequently facilitating NBCA injection through a second microcatheter navigated more distally [5, 31]. The concept was further refined by Hamaguchi et al. who demonstrated in a swine model the feasibility of NBCA injection through the working lumen of a dual lumen balloon microcatheter under flow-arrest conditions [12]. This was termed the B‑glue technique. In their study, a 50% NBCA (Histoacryl®, Aesculap, Tuttlingen, Germany) mixture with iodized oil was used [12]. The B‑glue technique demonstrated a significantly better control of the embolus range with less fragmentation and reflux in a swine model of intercostal artery embolization [12]. The B‑glue technique was also successfully used to treat a traumatic pseudoaneurysm arising from a small branch of the deep femoral artery without reflux [32]. More recently, Takao et al. reported successful embolization of a severe intrahepatic arterioportal shunt prior to chemoembolization of a hepatocellular carcinoma [33]. To the best of our knowledge, the present case series is the first description of NBCA embolization through a dual lumen balloon under local flow-arrest conditions in the setting of BAVM. This technique widens the armamentarium of the interventional neuroradiologist for the management of BAVM and is particularly useful for embolizing high-flow arteriovenous fistulas which have been reported as presenting a particular challenge for stereotactic radiosurgery, surgery and embolization [34].
Significant limitations may hinder the generalizability of the technique described here. First of all, dual lumen balloons are more difficult to navigate distally in cerebral arteries with a higher risk of navigation-related complications when compared to low-profile flow-dependent microcatheters usually used for NBCA embolization. Specific potential complications related to distal balloon inflation include arterial dissection and arterial rupture. Further development is required to make dual lumen balloons specifically tailored for this indication. Additional concern comes from a recent report by Park et al. regarding possible incompatibility between lipiodol and balloon microcatheters [35]. Park et al. reported a case of Hyperglide (Covidien, Irvine, CA, USA) single lumen balloon rupture upon contact with a 21% NBCA-lipiodol liquid embolic agent solution during a preoperative hemangiopericytoma embolization procedure [35]. Subsequent in vitro testing with several commercially available single and dual lumen balloon microcatheters demonstrated tolerance to NBCA [35]. Conversely, adding lipiodol to inflated balloons resulted in rupture. Moreover, scanning electron microscopy of balloon surfaces exposed to lipiodol solution for 10 s demonstrated surface cracks [35]. This particular concern must be addressed if the technique is to be widely applied by using dedicated balloons compatible with lipiodol; however, it is noteworthy that no balloon rupture was recorded in this short case series due to the contact with the NBCA-lipiodol mixture. When the glue refluxed too close to the distal aspect of the balloon, it was gently deflated to help the progression of the glue cast then the balloon was re-inflated for further NBCA injection.
Conclusion
High-flow BAVM fistulas can be treated with NBCA embolization using a dual-lumen balloon under local flow-arrest conditions.
References
Shotar E, Debarre M, Sourour NA, Di Maria F, Gabrieli J, Nouet A, Chiras J, Degos V, Clarençon F. Retrospective study of long-term outcome after brain arteriovenous malformation rupture: the RAP score. J Neurosurg. 2018;128:78-85.
Guo WY, Karlsson B, Ericson K, Lindqvist M. Even the smallest remnant of an AVM constitutes a risk of further bleeding. Case report. Acta Neurochir (Wien). 1993;121:212–5.
Luessenhop AJ, Spence WT. Artificial embolization of cerebral arteries. Report of use in a case of arteriovenous malformation. J Am Med Assoc. 1960;172:1153–5.
Vinuela F, Duckwiler G, Guglielmi G. Contribution of interventional neuroradiology in the therapeutic management of brain arteriovenous malformations. J Stroke Cerebrovasc Dis. 1997;6:268–71.
Andreou A, Ioannidis I, Nasis N. Transarterial balloon-assisted glue embolization of high-flow arteriovenous fistulas. Neuroradiology. 2008;50:267–72.
Guglielmi G, Viñuela F, Duckwiler G, Dion J, Stocker A. High-flow, small-hole arteriovenous fistulas: treatment with electrodetachable coils. AJNR Am J Neuroradiol. 1995;16:325–8.
Nesbit GM, Barnwell SL. The use of electrolytically detachable coils in treating high-flow arteriovenous fistulas. AJNR Am J Neuroradiol. 1998;19:1565–9.
Viñuela F, Drake CG, Fox AJ, Pelz DM. Giant intracranial varices secondary to high-flow arteriovenous fistulae. J Neurosurg. 1987;66:198–203.
Kerber CW, Bank WO, Cromwell LD. Calibrated leak balloon microcatheter: a device for arterial exploration and occlusive therapy. AJR Am J Roentgenol. 1979;132:207–12.
Yu SC, Chan MS, Lam JM, Tam PH, Poon WS. Complete obliteration of intracranial arteriovenous malformation with endovascular cyanoacrylate embolization: initial success and rate of permanent cure. AJNR Am J Neuroradiol. 2004;25:1139–43.
Letter KC. Intracranial cyanoacrylate: a new catheter therapy for arteriovenous malformation. Invest Radiol. 1975;10:536–8.
Hamaguchi S, Lohman BD, Ogawa Y, Arai Y, Hashimoto K, Matsumoto J, Nakajima Y. Preliminary findings of arterial embolization with balloon-occluded and flow-dependent histoacryl glue embolization in a swine model. Jpn J Radiol. 2015;33:344–51.
Mavili E, Donmez H, Ozcan N, Akcali Y. Endovascular treatment of lower limb penetrating arterial traumas. Cardiovasc Intervent Radiol. 2007;30:1124–9.
n-BCA Trail Investigators. N‑butyl cyanoacrylate embolization of cerebral arteriovenous malformations: results of a prospective, randomized, multi-center trial. AJNR Am J Neuroradiol. 2002;23:748–55.
Clarençon F, Nouet A, Redondo A, Di Maria F, Iosif C, Le Jean L, Chiras J, Sourour N. Occlusion of M1 segment after superficial temporal artery-middle cerebral artery bypass in a giant M1 aneurysm with Onyx-34 injected via a double-lumen balloon under balloon inflation. J Neurointerv Surg. 2014;6:e27.
Guenego A, Zerlauth JB, Puccinelli F, Hajdu S, Rotzinger DC, Zibold F, Piechowiak EI, Mordasini P, Gralla J, Dobrocky T, Daniel RT, Chapot R, Mosimann PJ. Balloon-assisted coil embolization and large stent delivery for cerebral aneurysms with a new generation of dual lumen balloons (Copernic 2L). J Neurointerv Surg. 2018;10:395–400.
Martínez-Galdámez M, Rodriguez-Arias CA, Utiel E, Arreba E, Gonzalo M, Arenillas JF. Paraspinal arteriovenous malformation Onyx embolization via an Ascent balloon. J Neurointerv Surg. 2014;6:e19.
Fifi J, Niimi Y, Berenstein A. Onyx embolization of an extensive mandibular arteriovenous malformation via a dual lumen balloon catheter: a technical case report. J Neurointerv Surg. 2013;5:e5.
Clarençon F, Pérot G, Biondi A, Di Maria F, Szatmary Z, Chiras J, Sourour N. Use of the Ascent balloon for a 2-in-1 remodeling technique: feasibility and initial experience: case report. Neurosurgery. 2012;70:170–3. discussion 173.
Alaraj A, Wallace A, Dashti R, Patel P, Aletich V. Balloons in endovascular neurosurgery: history and current applications. Neurosurgery. 2014;74(Suppl 1):S163–S90.
Chiu AH, Aw G, Wenderoth JD. Double-lumen arterial balloon catheter technique for Onyx embolization of dural arteriovenous fistulas: initial experience. J Neurointerv Surg. 2014;6:400–3.
Jagadeesan BD, Grigoryan M, Hassan AE, Grande AW, Tummala RP. Endovascular balloon-assisted embolization of intracranial and cervical arteriovenous malformations using dual-lumen coaxial balloon microcatheters and Onyx: initial experience. Neurosurgery. 2013;73(2 Suppl Operative):ons238-43; discussion ons243.
Gounis MJ, Lieber BB, Wakhloo AK, Siekmann R, Hopkins LN. Effect of glacial acetic acid and ethiodized oil concentration on embolization with N‑butyl 2‑cyanoacrylate: an in vivo investigation. AJNR Am J Neuroradiol. 2002;23:938–44.
Stoesslein F, Ditscherlein G, Romaniuk PA. Experimental studies on new liquid embolization mixtures (histoacryl-lipiodol, histoacryl-panthopaque). Cardiovasc Intervent Radiol. 1982;5:264–7.
Iizuka Y, Murata N, Kohda E, Tsutsumi Y, Nosaka S, Morota N, Konishi Y. High-flow neonatal macrocerebral arteriovenous fistulas in hereditary hemorrhagic telangiectasia. Neuroradiol J. 2011;24:772–8.
Hermier M, Turjman F, Bozio A, Duquesnel J, Lapras C. Endovascular treatment of an infantile nongalenic cerebral arteriovenous fistula with cyanoacrylate. Childs Nerv Syst. 1995;11:494–8.
Mathis JM, Evans AJ, DeNardo AJ, Kennett K, Crandall JR, Jensen ME, Dion JE. Hydrophilic coatings diminish adhesion of glue to catheter: an in vitro simulation of NBCA embolization. AJNR Am J Neuroradiol. 1997;18:1087–91.
Moore C, Murphy K, Gailloud P. Improved distal distribution of n‑butyl cyanoacrylate glue by simultaneous injection of dextrose 5% through the guiding catheter: technical note. Neuroradiology. 2006;48:327–32.
Nelson PK, Russell SM, Woo HH, Alastra AJG, Vidovich DV. Use of a wedged microcatheter for curative transarterial embolization of complex intracranial dural arteriovenous fistulas: indications, endovascular technique, and outcome in 21 patients. J Neurosurg. 2003;98:498–506.
Debrun GM, Aletich V, Ausman JI, Charbel F, Dujovny M. Embolization of the nidus of brain arteriovenous malformations with n‑butyl cyanoacrylate. Neurosurgery. 1997;40:112-20; discussion 120-1.
Kerim AA, Bonneville F, Jean B, Cornu P, LeJean L, Chiras J. Balloon-assisted embolization of skull base meningioma with liquid embolic agent. J Neurosurg. 2010;112:70–2.
Hamaguchi S, Ogawa Y, Arai Y, Hashimoto K, Nakajima Y. A case of pseudoaneurysm of the deep femoral artery successfully treated by NBCA embolization under occlusion. Jpn J Radiol. 2013;31:538–41.
Takao H, Shibata E, Ohtomo K. Double-balloon-assisted n‑butyl-2-cyanoacrylate embolization of intrahepatic arterioportal shunt prior to chemoembolization of hepatocellular carcinoma. Cardiovasc Intervent Radiol. 2016;39:1479–83.
Yuki I, Kim RH, Duckwiler G, Jahan R, Tateshima S, Gonzalez N, Gorgulho A, Diaz JL, De Salles AA, Viñuela F. Treatment of brain arteriovenous malformations with high-flow arteriovenous fistulas: risk and complications associated with endovascular embolization in multimodality treatment. J Neurosurg. 2010;113:715–22.
Park S, Hwang SM, Lim OK, Hwang C, Lee DH. Compliant neurovascular balloon catheters may not be compatible with liquid embolic materials: intraprocedural rupture of the protecting balloon during tumor embolization using n‑butyl cyanoacrylate and lipiodol mixture. J Neurointerv Surg. 2015;7:740–3.
Funding
This research received no specific grant from any funding agency in the public, commercial or not-for-profit sectors.
Author information
Authors and Affiliations
Contributions
Conception or design of the work: E. Shotar, F. Clarençon, N.‑A. Sourour; data acquisition: E. Shotar, A. Al Raaisi, N.‑A. Sourour; data analysis: E. Shotar, A. Al Raaisi, F. Clarençon, N.‑A. Sourour; drafting the work: E. Shotar, A. Al Raaisi, F. Clarençon, N.‑A. Sourour; critical revision of the work for important intellectual content: all co-authors; final approval of the version to be published: all co-authors; agreement to be accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved: all co-authors.
Corresponding author
Ethics declarations
Conflict of interest
F. Clarençon: consultant for Truffle Capital, Penumbra, Guerbet and Stryker. N.‑A. Sourour: consultant for Balt. E. Shotar, A. Al Raaisi, S. Lenck, K. Premat, V. Degos and B. Mathon declare that they have no competing interests.
Additional information
This work has been previously presented as an oral presentation at the 2018 ABC WIN Meeting in Val d’Isère, France (14–19 January 2018).
Rights and permissions
About this article
Cite this article
Shotar, E., Al Raaisi, A., Lenck, S. et al. Injection of N‑butyl Cyanoacrylate Through a Dual-Lumen Balloon for Embolization of High-flow Intranidal Fistulas in Brain Arteriovenous Malformations: Technical Note. Clin Neuroradiol 30, 313–319 (2020). https://doi.org/10.1007/s00062-019-00780-6
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00062-019-00780-6