Abstract
Heart rate variability (HRV) provides a valuable tool for early detection of cardiovascular abnormalities. Ectopic beats have been proven to have an influence on HRV results, but the effect of different amount of ectopic beats on analysis of congestive heart failure (CHF) patient rhythms has not been quantified. In this study, we tested the commonly used HRV indices for significant differences between 5-min RR segments with and without ectopic beats. Eight long-term CHF RR interval recordings from http://www.physionet.org were studied. Each recording was divided into non-overlapping segments of 5-min RR segments without or with different numbers of ectopic beats. Two time-domain HRV indices of SDNN and RMSSD and two frequency-domain indices of normalized low frequency (LFn) and high frequency (HFn) powers were employed. Results showed that ectopic segments had significantly larger values for SDNN (39 ± 18 vs. ectopic free segments 28 ± 16 ms, P < 0.05), RMSSD (47 ± 29 vs. 24 ± 23 ms, P < 0.05) and HFn (0.66 ± 0.13 vs. 0.52 ± 0.14, P < 0.01), and significantly lower values for LFn (0.34 ± 0.13 vs. ectopic free segments 0.48 ± 0.14, P < 0.01). Compared with the indices of RMSSD and frequency-domain indices, SDNN was least affected by a relatively small amount of ectopic beats (one to six beats). Compared with the time-domain indices, the frequency-domain indices responded more quickly to the appearance of ectopic beats.
Access provided by Autonomous University of Puebla. Download conference paper PDF
Similar content being viewed by others
Keywords
1 Introduction
Heart rate variability (HRV) studies the slight fluctuations in RR interval time series and has become an important tool for the assessment of cardiovascular autonomic regulation during various physiological and clinical conditions. However, ectopic beats consists in RR interval time series may cause bias in the reliable measurement of HRV [1, 2]. Researchers reported that some HRV time-domain parameters are more sensitive to the ectopic beats than others [1]. Researchers also stated that the presence of only one ectopic beat can introduce an increase in the high frequency (HF) power in HRV of around 10% [3]. The reason is that ectopic beats can induce the sudden changes in RR intervals and increase the high-frequency fluctuation in the tachogram of HRV, distorting the true measure of an HRV metric [4].
Although many detecting and editing (correcting) methods for ectopic beats have been proposed [2, 5,6,7], there is no agreed conclusion on which method is more efficient to handle the ectopic beats to obtain accurate HRV estimation [2, 4, 7]. Meanwhile, the effect of different amount of ectopic beats on the common time-domain and frequency-domain HRV indices for special cardiovascular disease has not been widely documented. This paper, therefore, aims to quantify the differences between ectopic-free and ectopic RR interval time series for congestive heart failure (CHF) patients and to quantify the effect of different amount of ectopic beats on the commonly used time-domain and frequency-domain HRV indices.
2 Methods
2.1 Database
All data used were from the PhysioNet CHF RR Interval Databases [8], a free-access, online archive of physiological signals. This database included 29 long-term RR interval records of subjects aged 34–79, with CHF. The original ECGs were digitized at 128 Hz, and the beat annotations were obtained by automated analysis with manual review and correction. A 5-min time window was used to segment the long-term RR interval records. In each 5-min RR segment, RR intervals greater than 2 s were first removed to exclude the influence from artifacts [9]. Then 5-min RR segments were classified as an ectopic segment if at least one ectopic beat was included or ectopic-free segment if no ectopic beat existed. Ectopic beats were from supra-ventricular or ventricular beats, depending on the localization of ectopic focus. Figure 1 shows the examples of an ectopic-free and ectopic 5-min RR segments. To exclude the record effect, only eight records with approximately equal amounts of ectopic-free and ectopic RR segments were selected for the following analysis (see Table 1).
2.2 HRV Indices
For a 5-min RR segment \( x\left( i \right) \), \( \left( {i = 1,2, \ldots ,N} \right) \), four common HRV indices were calculated in this study.
-
(1)
Time-domain indices:
The standard deviation of 5-min RR segment (\( {\text{SDNN}} \)) and the square root of the mean of the sum of the squares of differences between adjacent intervals of 5-min RR segment (\( {\text{RMSSD}} \)) were used and were defined as:
$$ {\text{SDNN}} = \sqrt {\frac{1}{N - 1}\sum\nolimits_{i = 1}^{N} {\left( {x\left( i \right) - \bar{x}} \right)^{2} } } $$(1)$$ {\text{RMSSD}} = \sqrt {\frac{1}{N - 1}\sum\nolimits_{i = 1}^{N - 1} {\left( {x\left( {i + 1} \right) - x\left( i \right)} \right)^{2} } } $$(2)where \( \bar{x} = \frac{1}{N}\sum\nolimits_{i = 1}^{N} {x\left( i \right)} \) is the mean value of RR intervals.
-
(2)
Frequency-domain indices:
Frequency-domain analysis provides a quantitative evaluation of sympathetic and parasympathetic activation. The low frequency (LF) component (between 0.04 and 0.15 Hz) reflects both sympathetic and parasympathetic nervous systems while the HF component (between 0.15 and 0.40 Hz) mainly reflects vagal activity [4, 10]. Burg modern spectrum estimation was used to acquire frequency parameters. The spectral powers of LF and HF were normalized using \( {\text{LFn}} = \frac{\text{LF}}{{{\text{LF}} + {\text{HF}}}} \) and \( {\text{HFn}} = \frac{\text{HF}}{{{\text{LF}} + {\text{HF}}}} \).
2.3 Statistical Analysis
Group t-student test was used to determine whether the results obtained from the ectopic-free and ectopic RR segments had significant differences. Comparisons were also performed between the ectopic-free group and the groups with different amount of ectopic beats. All statistical analyses were performed using the SPSS software (Ver. 20, IBM, USA). A statistical significance was accepted at P < 0.05.
3 Results
Table 2 gives the overall means and SDs of HRV indices (i.e., SDNN, RMSSD, LFn, and HFn) from both ectopic-free and ectopic 5-min RR segments. As shown in Table 2, there were significant differences in all four indices between the two groups (all P < 0.05). For time-domain indices, SDNN in ectopic group was significantly larger by 11 ms (39 ± 18 vs. 28 ± 16 ms, P < 0.05) and RMSSD was significantly larger by 23 ms (47 ± 29 vs. 24 ± 23 ms, P < 0.05). For frequency-domain indices, LFn in ectopic group was significantly lower by 0.14 (0.34 ± 0.13 vs. 0.48 ± 0.14, P < 0.01) while HFn was significantly larger by 0.14 (0.66 ± 0.13 vs. 0.52 ± 0.14, P < 0.01).
Then we divided the ectopic RR segments into sub-groups with different amount of ectopic beats. Figure 2 shows the distribution of RR segments classified by the number of ectopic beats. There are total 795 ectopic-free RR segments (also as shown in Table 2). For ectopic ones, there are 464, 243, 138, 81, 50 and 33 segments when ectopic beast are from one to six respectively, and there are 217 ectopic segments with ectopic beats more than six. Figure 3 shows the corresponding mean and SDs values of the four HRV indices for each RR segment group, to allow the ectopic beat number-related HRV changes to be observed and compared.
For time-domain indices, SDNN increased with increasing ectopic beat number. When compared with the ectopic-free group, statistically significant increases were observed at the group with one ectopic beat (P < 0.01), the group with two ectopic beats (P < 0.01) and the group with more than six ectopic beats (P < 0.01). However, the groups with three to six ectopic beats did not show the statistical significance. The reason maybe come from the small amount of RR segments. RMSSD also increased with the increase of ectopic beat number for each group. When compared with the ectopic-free group, statistically significant increases were observed for all seven ectopic groups (all P < 0.01).
For frequency-domain indices, LFn gradually decreased with the increase of ectopic beat number for each group. When compared with the ectopic-free group, statistically significant decreases were observed for all seven ectopic groups (all P < 0.01). HFn showed a reverse trend to LFn since mathematically, the sum of LFn and HFn was a constant.
For comparison of the results from time-domain and frequency-domain HRV indices, the latter responded more quickly to the appearance of ectopic beats. With the increase of ectopic beats from one to six, LFn (HFn) quickly decreased (increased) at these situations with relatively small amount of ectopic beats. When the number of ectopic beats was large and increased up to more than six, the decrease trend in LFn (increase trend in HFn) was not obvious. However, time-domain indices showed distinct changes when applied on the RR segments with more than six ectopic beats.
4 Discussion
In this study, we quantified the effect of different amount of ectopic beats on the common time-domain and frequency-domain HRV indices for CHF patients. The results showed that the ectopic RR segments had larger SDNN, RMSSD, and HFn but lower LFn values than the ectopic-free RR segments. Compared with the indices of RMSSD and frequency-domain indices, SDNN was least affected by a relatively small amount of ectopic beats (one to six beats). Compared with the time-domain indices, the frequency-domain indices responded more quickly to the appearance of ectopic beats. The results confirmed that ectopic beats in RR segments have a significant influence on the HRV results for CHF patients.
Ectopic beats are routinely removed/edited from the RR tachogram prior to HRV analysis. Our results agreed with the previously reported works. Salo et al. found that SDNN was least affected by the editing of RR intervals, and compared with the indices of SDNN and RMSSD, frequency-domain indices were more sensitive to the editing of RR intervals and reported larger errors after the ectopic beat editing [1]. This finding was consistent with our current study, where we showed that RMSSD and frequency-domain indices were more sensitive to the appearance of a small amount of ectopic beast (one to six beats). The reason is that the ectopic beats are usually step-like shapes and can result in sudden changes in RR interval time series. This effect is significant on the transient change of HRV reflected by RMSSD and the spectral components of HRV reflected by the frequency-domain LFn and HFn indices [4].
References
M. A. Salo, H. V. Huikuri, and T. Seppänen, “Ectopic beats in heart rate variability analysis: effects of editing on time and frequency domain measures,” Ann Noninvasive Electrocardiol, vol. 6, no. 1, pp. 5–17, 2001.
D. Nabil and F. B. Reguig, “Ectopic beats detection and correction methods: A review,” Biomedical Signal Processing and Control, vol. 18, no. 4, pp. 228–244, 2015.
G. G. Berntson and J. R. Stowell, “ECG artifacts and heart period variability: don’t miss a beat,” Psychophysiology, vol. 35, no. 1, pp. 127–132, 1998.
G. D. Clifford and L. Tarassenko, “Quantifying errors in spectral estimates of HRV due to beat replacement and resampling,” IEEE Trans Biomed Eng, vol. 52, no. 4, pp. 630–638, 2005.
T. H. Tarkiainen, T. A. Kuusela, K. U. Tahvanainen, J. E. Hartikainen, P. Tiittanen, K. L. Timonen, and E. J. Vanninen, “Comparison of methods for editing of ectopic beats in measurements of short-term non-linear heart rate dynamics,” Clin Physiol Funct Imaging, vol. 27, no. 2, pp. 126–133, 2007.
C. Y. Liu, L. P. Li, L. N. Zhao, D. C. Zheng, P. Li, and C. C. Liu, “A combination method of improved impulse rejection filter and template matching for identification of anomalous intervals in RR sequences,” Journal of Medical and Biological Engineering, vol. 32, no. 8, pp. 245–250, 2012.
M. A. Peltola, “Role of editing of R-R intervals in the analysis of heart rate variability,” Front Physiol, vol. 3, p. 148, 2012.
A. L. Goldberger, L. A. Amaral, L. Glass, J. M. Hausdorff, P. C. Ivanov, R. G. Mark, J. E. Mietus, G. B. Moody, C. K. Peng, and H. E. Stanley, “PhysioBank, PhysioToolkit, and PhysioNet: components of a new research resource for complex physiologic signals,” Circulation, vol. 101, no. 23, pp. 215–220, 2000.
L. N. Zhao, S. S. Wei, C. Q. Zhang, Y. T. Zhang, X. E. Jiang, F. Liu, and C. Y. Liu, “Determination of sample entropy and fuzzy measure entropy parameters for distinguishing congestive heart failure from normal sinus rhythm subjects,” Entropy, vol. 17, no. 9, pp. 6270–6288, 2015.
K. Zhu, D. Chemla, G. Roisman, W. Mao, S. Bazizi, A. Lefevre, and P. Escourrou, “Overnight heart rate variability in patients with obstructive sleep apnoea: a time and frequency domain study,” Clin Exp Pharmacol Physiol, vol. 39, no. 11, pp. 901–908, 2012.
Acknowledgements
The study was partly supported by the National Natural Science Foundation of China (Grant Number: 61571113 and Grant Number: 61671275), the Key Research and Development Programs of Jiangsu Province (Grant Number: BE2017735).
Author information
Authors and Affiliations
Corresponding author
Editor information
Editors and Affiliations
Rights and permissions
Copyright information
© 2019 Springer Nature Singapore Pte Ltd.
About this paper
Cite this paper
Liu, C. et al. (2019). Effect of Ectopic Beats on Heart Rate Variability Indices in Heart Failure Patients. In: Lhotska, L., Sukupova, L., Lacković, I., Ibbott, G. (eds) World Congress on Medical Physics and Biomedical Engineering 2018. IFMBE Proceedings, vol 68/2. Springer, Singapore. https://doi.org/10.1007/978-981-10-9038-7_68
Download citation
DOI: https://doi.org/10.1007/978-981-10-9038-7_68
Published:
Publisher Name: Springer, Singapore
Print ISBN: 978-981-10-9037-0
Online ISBN: 978-981-10-9038-7
eBook Packages: EngineeringEngineering (R0)