Abstract
In this chapter, an optoelectronic Lab-on-Chip system for DNA amplification integrating a polydimethylsiloxane microfluidic structure with indium tin oxide heaters on a 5 × 5 cm2 microscope glass slide is presented. The microfluidic structure comprises a channel for the fluid handling and a reaction chamber for the polymerase chain reaction (PCR). Two lateral heaters, located at the inlet and outlet positions of the chamber, actuate the valves and allow the chamber isolation, while the last one, positioned below the chamber, is dedicated to the PCR thermal cycle. In order to optimize heaters and valves geometry, the temperature distribution over the heaters and the membrane valve deformation have been studied using the commercial software COMSOL Multiphysics. The microfluidic structure has been fabricated by using soft lithography techniques on one side of the glass substrate, while the thin film heaters were deposited and patterned on the opposite glass side. Experiments show that valve activation begins around 60 °C, and complete closure is observed around to 100 °C, without any loss of liquid from the chamber.
Access provided by Autonomous University of Puebla. Download conference paper PDF
Similar content being viewed by others
Keywords
- Microfluidic Channel
- Polymerase Chain Reaction Technique
- PDMS Membrane
- Polymerase Chain Reaction Cycle
- Microfluidic Network
These keywords were added by machine and not by the authors. This process is experimental and the keywords may be updated as the learning algorithm improves.
1 Introduction
Lab-on-Chip (LOC) systems, in which functionalities as handling, treatment, and detection of biological solutions are downsized and integrated, represent one of the most attractive device for biomolecular and chemical analysis [1–4]. In particular, implementation of the polymerase chain reaction (PCR) in LOC devices has received a lot of attention [5–7] for applications such as DNA finger printing, genomic cloning, and genotyping for disease diagnosis. Indeed, its miniaturization results in reduced consumption of samples/reagents, shorter analysis times, and higher sensitivity and portability. Different solutions in terms of materials for the substrate and for the microfluidics have been proposed.
In this paper the authors present the design, fabrication, and characterization of a LOC system for DNA amplification based on a PolyDiMethylSiloxane (PDMS) microfluidic structure with integrated thermo-actuated valves and indium tin oxide (ITO) heaters. The use of these materials coupled with their deposition on a microscope glass presents features of easy fabrication, low cost, transparency, and biocompatibility.
2 LOC Design
The structure of the proposed LOC system (see Fig. 1) comprises:
-
1.
A microfluidic network, made in PDMS, that includes a rectangular channel for inlet and outlet, a chamber for the PCR, and two thermo-actuated valves.
-
2.
Three thin film heaters. Two of them, located at the inlet and outlet positions of the chamber, actuate the valves and allow the chamber isolation, while the last one, positioned below the chamber, is dedicated to the PCR thermal cycle.
-
3.
A glass substrate hosting the microfluidic network on one side and the heaters on the opposite side.
The solution to be amplified is injected in the microfluidic channel through the inlet. As the channel and the PCR chamber are filled, the two heaters below the valves are actuated to avoid sample losses, due to the pressure developed in the reaction chamber during the thermal cycles. Indeed, applying power to the heaters, the air contained in the valve reservoir (rectangular white spaces over the heaters in Fig. 1) heats up and increases its volume creating a pressure that pushes up the PDMS membrane into the channel. As the valves are closed, the PCR cycle may begin by turning on the heater below the PCR chamber.
4 Heater Design
In order to achieve a uniform temperature distribution inside the PCR chamber, the geometry of the PCR chamber heater and the distance between the three heaters have been optimized using the software COMSOL Multiphysics. We found that selecting a chirp geometry for the PCR chamber heater [8], a serpentine geometry for the valve heaters, and 6 mm distance between the valve heaters and the PCR chamber, the temperature uniformity in the PCR chamber is better than 3 % for the three temperature steps, satisfying the PCR technique requirements. A simulated temperature distribution during a PCR cycle along the microfluidic network is reported in Fig. 2.
5 Valve Design
The valves have a cylindrical shape with diameter and height equal to 500 μm and 50 μm, respectively. These sizes have been chosen, as a result of finite element simulations, to ensure that the deformation of the valve membranes closes the microfluidic channel when the heaters are turned on.
6 LOC Fabrication and Testing
Device fabrication begins with the deposition of the ITO thin film heaters by magnetron sputtering on one side of a 5 × 5cm2 ultrasonically cleaned glass substrate. The film thickness is 200 nm, the same utilized in the simulations.
The PDMS structure was obtained by using soft lithography techniques. Two 50-μm-thick SU-8 molds were fabricated and patterned by photolithography: the first for the thermo-actuated valves and the second one for the microfluidic channel and the PCR chamber. PDMS was spun on the first mold in order to obtain a 100-μm-thick layer and a 50-μm-thick PDMS membrane over the SU-8 cylinders. PDMS was also poured onto the second mold to achieve a 3-mm-thick structure. The two PDMS structures were merged using the partial curing method [9], and subsequently the overall microfluidic device was bonded to the heaters on the glass substrate by oxygen plasma (Fig. 3a).
The PDMS microfluidic channel was filled with a mix of water and a blue dye in order to easily monitor the fluid position during the experiment. The valve-heater temperature has been raised slowly, and the behavior of the valve has been monitored under the microscope. Valves activation began around 60 °C, and complete closure was observed around to 100 °C as shown in Fig. 3b, without any loss of liquid from the chamber.
7 Conclusions
This paper has presented the design and fabrication of a microfluidic network integrated with thin film heaters for implementation of the PCR technique in Lab-on-Chip systems. The heaters have been designed to achieve temperature uniformity inside the PCR chamber better than 3 % at each temperature of the DNA amplification cycle. The microfluidic network, made in PDMS, includes two thermally actuated valves, whose successful operation in avoiding loss of samples from the PDMS chamber has been demonstrated.
References
D. Erickson, L. Dongqing. Integrated microfluidic devices. Analytica Chimica Acta, 507 (1), 11–26 (2004)
A.G. Crevillén, M. Hervás, M.A. López, M.C. González, A. Escarpa. Real sample analysis on microfluidic devices. Talanta, 74 (3), 342–357 (2007)
D. Janasek, J. Franzke, A. Manz. Scaling and the design of miniaturized chemical-analysis systems. Nature 442 (7101) 374–380 (2006)
D. Caputo, M. Ceccarelli, G. de Cesare, A. Nascetti, R. Scipinotti. Lab-on-glass system for DNA analysis using thin and thick film technologies. in Materials Research Society Symposia Proceedings, 1191, 53–58 (2009)
M.A. Northrup, R.F. Hills, P. Landre, S. Lehew, D. Hadley, R. Watson. A MEMS-based DNA analysis system. in Tranducer ’95, Eighth International Conference on Solid State Sens Actuators, Stockholm, Sweden. ISBN:9 1-630-3473-5, 764–767 (1995)
N.C. Cady, S. Stelick, M.V. Kunnavakkam, C.A. Batt. Real-time PCR detection of Listeria monocytogenes using an integrated microfluidics platform. Sens. Actuators B Chem. 107, 332 – 341 (2005)
Z.Q Niu, W.Y. Chen, S.Y. Shao, X.Y. Jia, W.P. Zhang. DNA amplification on a PDMS–glass hybrid microchip. J. Micromech. and Microeng., 16, (2), 425–433 (2006).
D. Caputo, G. de Cesare, A. Nascetti, and R. Scipinotti. a-Si:H temperature sensor integrated in a thin film heater. Phys. Status Solidi A, 207 (3), 708–711 (2010).
M.A. Eddings, M.A. Johnson, B.K. Gale. Determining the optimal PDMS-PDMS bonding technique for microfluidic devices. J Micromech Microengineering, 18, (6), 06700 (2008).
Author information
Authors and Affiliations
Corresponding author
Editor information
Editors and Affiliations
Rights and permissions
Copyright information
© 2014 Springer International Publishing Switzerland
About this paper
Cite this paper
Caputo, D. et al. (2014). Thermally Actuated Microfluidic System for Polymerase Chain Reaction Applications. In: Di Natale, C., Ferrari, V., Ponzoni, A., Sberveglieri, G., Ferrari, M. (eds) Sensors and Microsystems. Lecture Notes in Electrical Engineering, vol 268. Springer, Cham. https://doi.org/10.1007/978-3-319-00684-0_5
Download citation
DOI: https://doi.org/10.1007/978-3-319-00684-0_5
Published:
Publisher Name: Springer, Cham
Print ISBN: 978-3-319-00683-3
Online ISBN: 978-3-319-00684-0
eBook Packages: EngineeringEngineering (R0)