Keywords

1 Introduction

Transtibial amputation constitutes a significant impairment to the neural, muscular, and skeletal systems, involving the loss of the ankle, distal tibia, and fibula articulation while retaining the knee joint. Individuals with this type of amputation experience challenges in walking and increased energy consumption [1]. Patients with transtibial amputation require a prosthesis that enhances efficiency and gait pattern during both static and dynamic phases. This involves cushioning impacts from body weight and achieving proper prosthetic suspension in the swinging phase as well as stable support during the stance phase [2]. Achieving this involves assessing biomechanical variables such as baropodometry, accelerometry, and electromyography to understand how prosthetic adaptation occurs in amputee patients. Consequently, there has been a growing focus on prosthetic alignment studies, particularly regarding the impact of socket alignment with the foot and the assessment of associated energy costs [2].

Among baropodometric variables, weight distribution assesses how weight is borne by the prosthetic device and the contralateral limb in an orthostatic position [3]. Additionally, the center of pressure trajectory is associated with postural stability and balance evaluation. On the other hand, ground reaction force (GRF) identifies the effects of misalignment in prosthetic gait, indicating that misalignments increase the load on the contralateral limb and likewise extend the support phase time, as well as causing differences in vertical GRF peaks [4, 5].

Acelerometry evaluation identifies propulsion indices, describing a subject’s ability to push the center of mass forward during the single support phase [6]. The symmetry index represents the difference in percentage terms between support and swing phases of the healthy and amputated limbs [7]. The quality index characterizes a single limb, assessing the subject’s ability to correctly time and balance their gait cycle between the healthy and amputated limbs [8]. Lastly, pelvic kinematics are assessed in different planes, providing information about inclination, obliquity, and rotation during the gait cycle. Ultimately, muscle coactivation informs about the overall compensatory strategy adopted by amputees in the intact limb, where high values indicate the need for amputees to increase simultaneous activation of multiple muscles for an extended duration. This may lead to compensatory increases in stiffness, force production, and support time [9].

Biomechanical assessments are crucial for prosthesis users as they provide quantitative information about prosthetic adaptation, aiming to improve patient function and prosthetic component effectiveness in terms of user safety and stability [2]. Therefore, the main objective of this study is to evaluate kinetic and kinematic signals of lower limbs using baropodometry, accelerometry, and electromyography during the walking motion in a patient with transtibial amputation.

2 Materials and Methods

A 66-year-old patient with left transtibial amputation was evaluated. The patient reported fatigue when attempting short walks with the prosthetic provided by their healthcare institution. The study took place at the Biomechanics and Rehabilitation Laboratory of the Metropolitan Technological Institute in Medellín, Colombia. The patient participated voluntarily after signing the appropriate informed consent, which was approved by the Ethics Committee of the Metropolitan Technological Institute and adhered to the principles outlined in the Helsinki Declaration (DoH), as recommended by the World Medical Association [10].

2.1 Baropodometry Analysis

The EcoWalk plantar pressure platform (Ecosanit, Arezzo, Italy) was utilized for the study. Data processing was conducted using the EcoFoot 4.0 software developed by the same company, which includes predefined protocols for defining specific characteristics in both static and dynamic positions. This encompasses weight distribution, calculation of the center of gravity trajectory, and the ground reaction force vector.

For static position data acquisition, the patient stood on the platform, facing a fixed point, with relaxed arms and feet shoulder-width apart for 60 s (see Fig. 1). Subsequently, the patient was instructed to perform a short walk, alternating steps, to ensure recording of at least 4 steps for each foot during the dynamic assessment.

Fig. 1.
figure 1

Static evaluation on the plantar pressure platform.

2.2 Accelerometry and Surface Electromyography (EMG) Analysis

The G-walk inertial sensor (BTS Bioengineering, Milan, Italy) was used for Accelerometry Analysis. This sensor includes an accelerometer (frequency: 4–1.000 Hz), a gyroscope (frequency: 4–8.000 Hz), and a magnetometer. For signal processing, the G-Studio software version 3.2.25.0 developed by the same company was implemented.

This device facilitates the spatial and temporal evaluation of gait parameters, including quality index, propulsion, temporal parameters, and pelvic kinematics.

The inertial sensor was placed at the S1 and S2 vertebrae level, specifically below the line connecting the two dimples of Venus, to align it accurately with the patient’s spine. Once the sensor was securely positioned, the patient received instructions to walk along a 6-m path, maintaining a natural pace. You were also asked to perform a wide turn and return to the starting point of the test. This sequence was repeated multiple times, with each repetition lasting no more than 1 min.

To analyze muscle activity using surface electromyography, the wireless electromyograph FREEEMG 1000 (BTS Bioengineering, Milan, Italy) was utilized. This device is equipped with 8 sensors and operates at a sampling frequency of 1000 Hz. For signal processing, the EMG-Analyzer software version 2.10.44.0, developed by the same company, was employed. The calculation of the coactivation index followed the methodologies outlined in previous studies [11, 12]. Electrodes were placed on the right tibialis anterior, right medial gastrocnemius, right rectus femoris, right semitendinosus, left rectus femoris and left semitendinosus as shown in Fig. 2. After the sensor was stabilized, the patient then walked a 6-m path at a natural pace until reporting fatigue, marking the end of the test.

Fig. 2.
figure 2

Positioning of the inertial sensor and electromyography probes. a. Front view. b. Rear view.

3 Results and Discussion

3.1 Baropodometry Analysis

In the static baropodometry results, it’s observed that the right foot bears 50.1% of the load, while the left foot carries 49.9%. Furthermore, the left foot has 38.1% of the load on the forefoot, whereas the right foot records 31.3%, totaling 69.4% on the forefoot. These values suggest that the metatarsals bear the majority of the patient’s weight, leading to a forward displacement of the center of pressure, causing instability, fatigue, and wear in this area [13].

In the posterior zone, a total load of 30.5% is presented, with an uneven distribution between the feet.

The left foot bears 11.8% of the load, whereas the right foot reaches 18.8%. This discrepancy indicates that the patient doesn’t fully support the heel of the prosthetic foot, and there is a force compensation in the right foot’s heel (see Fig. 3). It’s crucial that the body weight is evenly distributed, with 50% of the load on both lower limbs, as well as anteroposteriorly, to avoid potential discomfort or long-term injuries [6].

Fig. 3.
figure 3

Distribution of Body Loads

In the dynamic analysis, for the left foot, the ground reaction force (GRF) exerted is low (68.9 kPa), representing less than 20% of the total weight. In the rearfoot, there’s a percentage of 5.49%, midfoot 2.94%, and forefoot 6.47%. These percentages add up to a total of 14.9% of the total weight, indicating that the patient exerts very little force with their left foot, possibly due to the lack of confidence they feel with the prosthesis.

On the other hand, the right foot records a higher force percentage (253.6 kPa). Specifically, there’s 15.29% in the rearfoot, 9.80% in the midfoot, and 24.31% in the forefoot, adding up to a total of 48.88% of the total weight. Although all these percentages are below normal values, the right foot shows a higher force percentage compared to the left foot and a longer time in the support phase, which is consistent with findings reported in other studies [5] (see Fig. 4).

Fig. 4.
figure 4

Ground Reaction Force Vector Graph

3.2 Analysis of Accelerometry and Surface Electromyography

The global symmetry index is 1.7%, indicating that the gait cycle in both feet is not symmetric, as the value is below the normal range [8]. Furthermore, the gait cycle quality index is 88.5% in the right limb and 85.5% in the left limb, indicating that the subject lacks symmetry in accelerating the center of mass similarly during the walking cycle in the right and left feet. The closer the index is to 100, the more symmetrical the gait pattern is [7] (see Table 1).

Table 1. Symmetry gait index.

The propulsion index presents a normal value when it’s > 5.5. In the case of this study, the values are slightly higher than the normal range, indicating that the patient has good propulsion, but might not possess all the necessary strength to execute it in the most correct and comfortable manner (see Fig. 5).

Fig. 5.
figure 5

Propulsion

Figure 6 illustrates pelvic kinematics in all three anatomical planes. The graphs feature two continuous lines representing the right side (green) and the left side (red). The gray lines correspond to the normal range. On the right side of the figure, the symmetry index is generated for each evaluated angle, along with maximum and minimum values for each foot, and the range between these values [14].

The inclination displays a symmetry index of 73.7%, indicating alignment issues with the pelvis and a tendency towards positive values, signifying anterior pelvic movement. This relates to maximum metatarsal support and confirms the patient’s forward-leaning body posture. In pelvis obliquity, there’s a symmetry of 33.5%, with an opposing angular pattern in each limb. The right pelvis remains higher than the left, potentially indicating poor prosthetic adaptation due to gait pattern asymmetry. Finally, in pelvic rotation, a symmetry index of 62.3% is observed, indicating neutral movement (see Fig. 6). Changes in pelvic kinematics might arise from relative compensations due to different prosthetic components. Inclination can be influenced by stability in re-establishing the center of gravity, and obliquity might be influenced by prosthetic adjustments like prosthesis length and suspension system [15].

Fig. 6.
figure 6

Pelvic Kinematics

The results of muscular activity can be observed in Fig. 7. The gray band indicates the phase during which each muscle should be activated, and the absence of a band signifies the resting phase of muscular contraction. For the right foot, muscles display activation throughout most of the gait cycle, without clear relaxation during the appropriate phase. This finding is consistent with studies indicating that the intact limb increases simultaneous activation for a longer duration to compensate for stiffness, force production, and support time in the healthy limb [16]. The rectus femoris and semitendinosus muscles show peaks of electrical activity closer to normal values; however, the voltage values obtained are low, suggesting that these muscles might not be generating sufficient force to adequately contribute to the gait cycle. On the other hand, in the left lower limb, the semitendinosus muscle’s activation is not consistent but shows relatively low voltage values, indicating insufficient force generation for a proper gait cycle with the amputated limb. However, the rectus femoris displays better activation and less silence, indicating that this muscle tends to experience faster fatigue, given that it bears the weight of the prosthesis. Consequently, the foot support might be lighter.

Fig. 7.
figure 7

Muscle Activation Timing

The coactivation index reflects the postural stability of the patient [17]. According to the data reported in Table 2, the coactivation values are higher than normal, indicating that the patient has balance issues during gait.

Table 2. Muscle Coactivation Index.

4 Conclusions

The results of this study revealed altered biomechanical behaviors in both joint and muscular aspects in a patient with transtibial amputation, underscoring the importance of such assessments for proper prosthetic adaptation. The absence of these considerations can lead to muscular fatigue, instability, and gait difficulties, significantly impacting the patient’s quality of life. It’s important to note that this study was conducted with a single patient, thus further evaluation with a larger amputee population is crucial to obtain comparative results. Moreover, the devices used for measurements necessitate trained personnel to ensure their proper utilization. Despite limitations, this study provides vital insights in biomedicine and rehabilitation, highlighting overlooked healthcare research that ensures proper patient prosthetic adaptation and quality of life. These findings advance understanding and stress the need for broader future approaches.