1 Introduction

Noncovalent interactions between proteins and carbohydrates on the surface of cells, present as either part of membrane glycoproteins or glycolipids, are involved in many normal and pathologic cellular processes, including catalysis, signaling, and molecular recognition [1]. Studies of protein–carbohydrate interactions in vitro can provide fundamental insights into these important processes and guide the development of diagnostics and therapeutics for a variety of infections and diseases. There exist a number of analytical methods for the detection and characterization of protein–carbohydrate interactions. For example, glycan microarrays are now commonly used to screen libraries of carbohydrates for specific interactions with proteins [2], while isothermal titration calorimetry (ITC) [3], surface plasmon resonance (SPR) [4], nuclear magnetic resonance (NMR) spectroscopy [5], and enzyme-linked immunosorbent assays (ELISA) [6] are extensively used to quantify the thermodynamics (and in some instances the kinetics) of protein–carbohydrate binding. In recent years, electrospray ionization mass spectrometry (ESI-MS) has emerged as a powerful method for detecting protein–carbohydrate complexes in solution and measuring the affinities of the interactions [718].

In the direct ESI-MS binding assay, the protein–ligand binding equilibrium constant is determined by quantifying the relative abundances of the free and ligand-bound protein ions in the gas phase [1113]. The measurements are fast and can often be completed within a few min, the amount of sample consumed is low, typically pmol of protein and nano- to pmol of ligand, and there is no requirement for labeling or additional reagents, which makes the assay very versatile. Moreover, the direct ESI-MS assay is the only technique that directly measures binding stoichiometry. This feature is particularly beneficial to the study of protein–carbohydrate interactions, as many carbohydrate-binding proteins are composed of multiple subunits and possess multiple ligand binding sites. The ESI-MS assay also affords the opportunity to measure, simultaneously, the binding of multiple, distinct ligands, and is, therefore, well suited to carbohydrate library screening [16].

A drawback of the ESI-MS assay, which is typically carried out using aqueous ammonium acetate solutions [19], is that it suffers from general incompatibility with nonvolatile “physiological” buffers, such as phosphate buffered saline (PBS), citrate, HEPES, and Tris-HCl. Such buffers are often needed to keep the protein stable in solution and to minimize protein aggregation [20]. Several strategies have been proposed to allow ESI-MS analysis of solutions containing physiological buffers at relevant concentrations, including the use of high concentrations of ammonium acetate [21] or carrying out ESI in the presence of a high velocity gas [22]. A possible alternative approach involves separating the sample from the ESI process through the use of desorption electrospray ionization (DESI) [2325] or liquid sample DESI [2629]. In liquid sample DESI-MS, the liquid sample is ionized through collisions with charged droplets produced by ESI [26, 27, 30]. The ESI solution is typically a mixture of water and an organic solvent, such as acetonitrile or methanol [28]. Despite this, liquid sample DESI has been shown to produce multiply charged gaseous ions of proteins and noncovalent protein complexes without inducing significant unfolding of the protein [28]. A variation of liquid sample DESI, known as reactive liquid sample DESI, was recently described and used to screen a library of compounds for specific binding to a target protein and to quantify the interactions [27]. In this approach, the ligands are introduced (consecutively) into the ESI spray solvent, rather than to the sample solution, which contained the target protein. The advantage of reactive liquid sample DESI is that the premixing of protein with ligands can be avoided [27].

The goal of the present study was to assess the reliability of liquid sample DESI for the quantification of protein–carbohydrate binding in aqueous ammonium acetate solutions and the tolerance of assay to the presence nonvolatile buffers. The affinities of tri- and tetrasaccharide ligands for lysozyme (Lyz), a glycosyl hydrolase, and a single chain variable fragment (scFv) of a monoclonal antibody were measured by liquid sample DESI-MS and the results compared with those measured using ITC and the direct ESI-MS assay [3133]. The suitability of liquid sample DESI-MS for quantitative binding measurements carried out using solutions containing high concentrations of PBS was also explored. Binding measurements were performed on solutions of Lyz and a trisaccharide ligand in varying concentrations of PBS and the results compared with those obtained using ITC and ESI-MS.

2 Experimental

2.1 Materials

Ubiquitin (Ubq, MW 8565 Da), lysozyme (from chicken egg white, Lyz, MW 14310 Da), and maltotriose (L1, MW 504.44 Da) were purchased from Sigma-Aldrich Canada (Oakville, Canada), and β-D-GlcNAc-(1 → 4)-β-D-GlcNAc-(1 → 4)-D-GlcNAc (L2, MW 627.59 Da), and β-D-GlcNAc-(1 → 4)-β-D-GlcNAc-(1 → 4)-β-D-GlcNAc-(1 → 4)-D-GlcNAc (L3, MW 830.27 Da) were purchased from Dextra Science and Technology Centre (Reading, UK). The single chain variable fragment of Se155-4 (scFv, MW 26539 Da) was produced and purified as described previously [34, 35], and α-D-Galp-(1 → 2)-[α-D-Abep-(1 → 3)]-α-D-Manp-OCH3 (L4, MW 486.50 Da) and β-D-Glcp-(1 → 2)-[α-D-Abep-(1 → 3)]-α-D-Manp-OCH3 (L5, MW 486.50 Da) were gifts from Professor D. Bundle (University of Alberta). Stock solutions of each protein (in 50 mM ammonium acetate) and oligosaccharide (in deionized water) were prepared and stored at −20°C until needed. A 10× PBS stock solution (NaCl (1.37 M), KCl (27 mM), Na2HPO4 (0.1 M), and KH2PO4 (18 mM), pH 7.4) was prepared and stored at 4°C until needed. Sample solutions for ESI- and liquid sample DESI-MS analysis were prepared from the stock solutions of protein and oligosaccharide. Unless otherwise indicated, the sample solutions contained 20 mM ammonium acetate. In a limited number of experiments, PBS was added, at the concentrations indicated.

3 Apparatus

3.1 Mass Spectrometry

All of the ESI- and liquid sample DESI-MS measurements were carried out in positive ion mode using a Synapt G2 quadrupole-ion mobility separation-time-of-flight (Q-IMS-TOF) mass spectrometer (Waters UK Ltd., Manchester, UK) equipped with a 8 k quadrupole mass filter. All data were processed using MassLynx software (ver. 4.1). For the ESI-MS measurements, nanoflow ESI (nanoESI) tips, produced from borosilicate capillaries (1.0 mm o.d., 0.68 mm i.d.) and pulled to ~5 μm using a P-1000 micropipette puller (Sutter Instruments, Novato, CA, USA), were used. A platinum wire was inserted into the nanoESI tip and a capillary voltage of 1.0–1.3 kV was applied to initiate the spray. A cone voltage of 30 V was used and the source block temperature was maintained at 60°C. The trap and transfer ion guides were maintained at 5 and 2 V, respectively, and the argon pressure in these regions was 2.22 × 10-2 mbar and 3.36 × 10–2 mbar, respectively. For the liquid sample DESI-MS measurements, a modified OMNI SPRAY Ion Sources 2-D OS-6205 (Prosolia Inc., Indianapolis, IN, USA) was used. The liquid sample solution was delivered through a silica capillary (360 nm o.d., 100 nm i.d.) at a flow rate of 5–10 μL h–1 using a syringe pump (Chemyx Syringe Pumps Fusion 100; Chemyx Inc., Stafford, TX, USA). The end of the silica capillary was positioned between the ESI tip and inlet of the mass spectrometer. The ESI solution flow rate was between 2 and 4 μL min–1. Capillary and cone voltages of 3.0–3.5 kV and 30 V, respectively, were used and the pressure of the N2 nebulizing gas was 60–70 psi. The source block temperature was the same as for the ESI-MS measurements.

Prior to carrying out the liquid sample DESI-MS protein–carbohydrate binding measurements, several different spray solvent compositions were tested (deionized water, 20 mM ammonium acetate, 50/50 v/v water/methanol, 20/80 v/v water/acetonitrile, 50/50 v/v water/acetonitrile, and 80/20 v/v water/acetonitrile) for the analysis of aqueous ammonium acetate solutions of Lyz or scFv, the two model carbohydrate-binding proteins used in this study. Ultimately, it was found that a 50/50 water/acetonitrile solution gave mass spectra with the highest signal-to-noise ratios. Shown in Supplementary Figure S1 (Supplementary Information) are representative liquid sample DESI mass spectra acquired in positive ion mode for aqueous ammonium acetate (20 mM) solutions containing Lyz (10 μM) or scFv (10 μM). It can be seen that liquid sample DESI-MS produced abundant signal corresponding to the protonated ions of Lyz (Supplementary Figure S1a) and scFv (Supplementary Figure S1b). A 50/50 water/acetonitrile solution was used as the spray solvent for all of the liquid sample DESI-MS binding measurements reported in this study.

3.2 Isothermal Titration Calorimetry

The ITC measurements were carried out using a VP-ITC (MicroCal Inc., Northampton, MA, USA). For each ITC experiment, the Lyz solution (0.1–0.2 mM) in the sample cell was titrated with a solution of L2 or L3 (2 mM); both the protein and ligand solutions were aqueous ammonium acetate (50 mM, pH 6.8) or PBS (1×, pH 7.4) at 25°C.

3.3 Data Analysis

The general procedure for determining association constants (Ka) for protein–ligand interactions from ESI mass spectra has been described in detail elsewhere and only a brief description is given for the case where the protein has single ligand binding site [18, 36, 37]. The assay relies on the detection and quantification of the gas-phase ions of free and ligand-bound protein. The concentration ratio (R) of the ligand-bound protein (PL) to free protein (P) in solution is taken to be equal to the total abundance (Ab) of P and PL ions as measured by ESI-MS, Equation 1. It follows that Ka can be calculated from Equation 2:

$$ R=\frac{{\displaystyle \sum Ab\left(\mathrm{P}\mathrm{L}\right)}}{{\displaystyle \sum Ab\left(\mathrm{P}\right)}}=\frac{{\left[\mathrm{P}\mathrm{L}\right]}_{eq}}{{\left[\mathrm{P}\right]}_{eq}} $$
(1)
$$ {\mathrm{K}}_a=\frac{R}{{\left[L\right]}_{\mathrm{o}}-\frac{R{\left[\mathrm{P}\right]}_{\mathrm{o}}}{1+R}} $$
(2)

where [P]o and [L]o are the initial protein and ligand concentrations, respectively.

It was shown previously that during the ESI process, free L can form so-called nonspecific complexes with P (and specific PL complexes), as the ESI droplets evaporate to dryness [36, 37]. The extent of nonspecific ligand binding is sensitive to the concentration of free L and, consequently, is more prevalent when measuring low affinity interactions because high L concentrations are needed to produce detectable concentrations of the PL complexes [37]. The formation of nonspecific PL interactions changes to the measured abundances of the P and PL ions and, thereby, introduces error into the R and Ka values. The reference protein method was developed to quantitatively correct ESI mass spectra for nonspecific binding [37]. The method involves the addition of reference protein (Pref), which does not interact with P or L, to the solution. The presence of nonspecific binding is established from the appearance of ions corresponding to Pref bound to one or more molecules of L (i.e., PrefLx complexes). As described in detail elsewhere, the contribution of nonspecific binding to the apparent (measured) abundances of P [Ab app (P)] and PL [Ab app (PL)] can be accounted for using Equations 3a and 3b:

$$ Ab\left(\mathrm{P}\right)=A{b}_{app}\left(\mathrm{P}\right)/{f}_O $$
(3a)
$$ Ab\left({\mathrm{PL}}^{\mathrm{n}+}\right)=\left\{A{b}_{app}\left({\mathrm{PL}}^{\mathrm{n}+}\right)-{F}_1 Ab\left({\mathrm{PL}}^{\mathrm{n}+}\right)\right\}/{f}_O $$
(3b)

where f 0 is the fraction of free P and f 1 the fraction of P bound nonspecifically to one molecule of L. These fractions can be determined from the measured abundances of free and ligand-bound forms of Pref, Equations 4a and 4b:

$$ {f}_0= Ab\left({\mathrm{P}}_{\mathrm{ref}}\right)/\left\{ Ab\left({\mathrm{P}}_{\mathrm{ref}}\right)+ Ab\left({\mathrm{P}}_{\mathrm{ref}}\mathrm{L}\right)\right\} $$
(4a)
$$ {f}_1= Ab\left({\mathrm{P}}_{\mathrm{ref}}\right)/\left\{ Ab\left({\mathrm{P}}_{\mathrm{ref}}\right)+ Ab\left({\mathrm{P}}_{\mathrm{ref}}\mathrm{L}\right)\right\} $$
(4b)

To test the reliability of the reference protein method for correcting liquid sample DESI mass spectra for the occurrence of nonspecific carbohydrate-protein binding, control experiments were carried out on solutions containing a pair of proteins (Lyz and Ubq) and L1, which does not bind to either protein in solution [31, 32]. Shown in Figure 1 are liquid sample DESI mass spectra acquired in positive ion mode for aqueous 20 mM ammonium acetate solutions of Lyz (10 μM) and Ubq (10 μM) and L1 at concentration of 15 μM (Figure 1a) or 40 μM (Figure 1b). It can be seen that in addition to the protonated ions of Lyz (at charge states +6 to +9) and Ubq (at charge states +4 to +6), ions corresponding to nonspecific complexes with L1 are evident at both concentrations. Shown in the insets of Figure 1a and b are the normalized abundances of Lyz and Ubq in their free and bound (to L1) forms. Notably, the distributions of both proteins are identical, within experimental error, at both concentrations of L1. These results confirm that the extent of nonspecific carbohydrate binding during the liquid sample DESI process is the same for the two proteins and, further, supports the use of the reference protein method for correcting liquid sample DESI mass spectra for nonspecific carbohydrate-protein binding.

Figure 1
figure 1

Representative liquid sample DESI mass spectra acquired in positive ion mode for aqueous ammonium acetate (20 mM, pH 6.8 and 25°C) solutions containing Ubq (10 μM), Lyz (10 μM), and L1 at (a) 15 μM or (b) 40 μM concentrations. The ESI spray solution was 50/50 water/acetonitrile

4 Results and Discussion

To test the reliability of liquid sample DESI-MS for quantifying protein–carbohydrate interactions, the affinities of the tri- and tetrasaccharide ligands, L2 and L3 for Lyz, and the trisaccharide ligands L4 and L5 for scFv, were measured. The affinities of L4 and L5 for scFv were previously measured in this laboratory using the direct ESI-MS assay and found to be 1.2 × 105 M–1 [38] and (5.0 ± 1.0) × 103 M–1 [39], respectively. The affinities of L2 and L3 for Lyz have been measured using several different biophysical techniques. Values of 1.1 × 105 M–1 (L2) and 1.8 × 105 M–1 (L3) were obtained using fluorescence-based assay [40]. Quantitative ESI-MS studies have also been carried out. Oldham and coworkers measured affinities of 1.0 × 105 M–1 (L2) and 1.2 × 105 M–1 (L3) [31], whereas Zenobi and coworkers found somewhat lower values (ranging from 2 × 104 M–1 to 5 × 104 M–1) for L2 [33]. Given the range of the reported values for L2, the affinities of L2 and L3 for Lyz in aqueous ammonium acetate (50 mM, pH 6.8 and 25°C) were measured using ITC, which is widely regarded as the gold standard method for quantifying the thermodynamics of protein–carbohydrate interactions. Shown in Supplementary Figures S2 and S3 are the raw and integrated ITC data measured for binding of Lyz to L2 and L3, respectively. According to the best fit of a 1:1 binding model to the ITC data, the affinities of L2 and L3 for Lyz are (9.0 ± 0.3) × 104 M–1 and (1.1 ± 0.1) × 105 M–1, respectively. These results are in good agreement with the values obtained using the fluorescence-based assay [40] and those reported by Oldham and coworkers [31].

4.1 Binding of Lyz to L2 and L3

The affinities of L2 and L3 for Lyz in aqueous ammonium acetate (20 mM, pH 6.8 and 25°C) were measured at three different ligand concentrations. Shown in Figure 2a and c are representative liquid sample DESI mass spectra acquired for solutions of Lyz (10 μM) with L2 (15 μM) or L3 (15 μM), respectively. Ubq (5 μM), which served as Pref, was added to both solutions. For comparison purposes, ESI mass spectra were also acquired for these solutions (Figure 2b and d). From Figure 2a and c, it can be seen that liquid sample DESI produces ions corresponding to free Lyz and ligand-bound Lyz (i.e., the (Lyz + L2) or (Lyz + L3) complexes), at charge states +6 to +8, as well as free Ubq at charge states +4 and +5. Ion signal corresponding to the nonspecific (Ubq + L2) or (Ubq + L3) complexes was negligible. Similar results were obtained for solutions containing L2 or L3 at two other concentrations, 5 μM and 10 μM (data not shown). The Ka values, representing the average of the values obtained by liquid sample DESI-MS at the three ligand concentrations, are (1.0 ± 0.1) × 105 M–1 (L2) and (9.9 ± 0.6) × 104 M–1 (L3) (Table 1). The ESI mass spectra obtained for aqueous ammonium acetate (20 mM) solutions of Lyz (10 μM), Ubq (5 μM), and L2 (15 μM) or L3 (15 μM) (Figure 2b and d, respectively) are qualitatively similar to the liquid sample DESI mass spectra, although the average charge states (ACS) of Lyz are slightly higher than those observed with liquid sample DESI [ACS 6.98 (Figure 2a), 7.43 (Figure 2b), 7.02 (Figure 2c), 7.98 (Figure 2d)]. The lower ACS values measured with liquid sample DESI-MS may be due to a subtle enhancement in the extent of proton transfer from the protein ions to acetonitrile in the gas phase. Acetonitrile has a relatively low gas-phase basicity (178.8 kcal mol–1) compared with ammonia (195.7 kcal mol–1) [41], but is present at a high concentration in the spray solvent (~9.6 M) and is expected to be present at relatively high concentrations in the spray droplets. The resulting acetonitrile vapor could affect proton transfer from the gaseous Lyz ions. Support for this explanation can be found in an observed decrease in ACS measured for Lyz when carrying out ESI in the presence of acetonitrile vapor (data not shown), a phenomenon also observed by Oldham and coworkers [42, 43]. The average Ka values obtained by ESI-MS at the three ligand concentrations are (8.0 ± 0.5) × 104 M–1 (L2) and (6.3 ± 0.5) × 104 M–1 (L3) (Table 1). Notably, the absolute affinities measured by liquid sample DESI-MS for L2 and L3 agree within a factor of two with the values determined from the ESI-MS measurements. More importantly, the liquid sample DESI-MS values are in excellent agreement with the affinities determined by ITC.

Figure 2
figure 2

Representative (a), (c) liquid sample DESI, and (b), (d) ESI mass spectra acquired in positive ion mode for aqueous ammonium acetate (20 mM, pH 6.8, and 25°C) solutions containing Lyz (10 μM), L2 (15 μM), and Ubq (5 μM) [(a) and (b)] or Lyz (10 μM), L3 (15 μM), and Ubq (5 μM) [(c) and (d)]. For the liquid sample DESI-MS measurements, the ESI spray solution was 50/50 water/acetonitrile

Table 1 Comparison of Association Constants (Ka) Measured by Liquid Sample DESI-MS, ESI-MS, and ITC for the Interactions of L2 and L3 with Lyz, and L4 and L5 with scFv in Aqueous Ammonium Acetate Solutions at pH 6.8 and 25°C a

4.2 Binding of scFv to L4 and L5

The affinities of L4 and L5 for scFv in aqueous ammonium acetate (20 mM, pH 6.8, and 25°C) were also measured at three different ligand concentrations. Shown in Figure 3a and c are representative liquid sample DESI mass spectra acquired for solutions of scFv (10 μM) with L4 (15 μM) or L5 (40 μM), respectively. Lyz (5 μM), which served as Pref, was added to both solutions. For comparison purposes, ESI mass spectra were also acquired for these solutions (Figure 3b and d). In Figure 3a and c, ion signal corresponding to protonated free scFv and the (scFv + L4) or (scFv + L5) complexes, at charge states +8 to +10, as well as free Lyz and the (Lyz + L4) or (Lyz + L5) complexes, at charge states +6 to +9, is evident. The appearance of ion signal for the (Lyz + L4) and (Lyz + L5) complexes indicates the occurrence of nonspecific carbohydrate–protein binding during ion formation. Similar results were obtained for solutions at two other concentrations of L4 (5 and 10 μM) and L5 (20 and 30 μM). Following correction of the mass spectra for nonspecific binding, average Ka values of (7.6 ± 0.1) × 104 M–1 (L4) and (5.7 ± 0.2) × 104 M–1 (L5) were determined (Table 1). The ESI mass spectra measured for solutions of scFv (10 μM), Lyz (5 μM) with L4 (15 μM), or L5 (40 μM) (Figure 3b and d, respectively) are similar to the corresponding liquid sample DESI mass spectra (Figure a and c). However, the extent of nonspecific binding is less in the case of ESI—there was no significant signal corresponding to the nonspecific (Lyz + L4) complex and significantly less (Lyz + L5) detected. The reduced occurrence of nonspecific binding may be due to the small droplets produced with the nanoESI tips compared with those formed in liquid sample DESI [44]. The smaller nanoESI droplets will contain fewer ligand molecules and, therefore, produce less nonspecific binding compared with the larger ESI droplets used for liquid sample DESI-MS. Following correction for nonspecific ligand binding, the affinities of L4 and L5 are found to be (6.6 ± 0.3) × 104 M–1 and (5.0 ± 0.1) × 103 M–1, respectively. Importantly, the affinities measured by liquid sample DESI-MS for L4 and L5 are in good agreement with the values determined using the direct ESI-MS assay.

Figure 3
figure 3

Representative (a), (c) liquid sample DESI, and (b), (d) ESI mass spectra acquired in positive ion mode for aqueous ammonium acetate (20 mM, pH 6.8 and 25°C) solutions containing scFv (10 μM), L4 (15 μM), and Lyz (5 μM) [(a) and (b)] or scFv (10 μM), L5 (40 μM), and Lyz (5 μM) [(c) and (d)]. For the liquid sample DESI-MS measurements, the ESI spray solution was 50/50 water/acetonitrile

Taken together, the results obtained for these model carbohydrate binding proteins demonstrate that absolute affinities for protein–carbohydrate interactions can be accurately quantified using liquid sample DESI-MS. These findings further indicate that the lifetime of the ESI droplets that produce gaseous protein ions in liquid sample DESI-MS are sufficiently short that neither the presence of a high concentration of organic solvent in the ESI spray solution nor the inevitable dilution of the sample (protein and ligand) solution by the solvent spray results in a measurable shift in the binding equilibrium.

4.3 Comparison of Liquid Sample DESI-MS and Reactive Liquid Sample DESI-MS

It is also interesting to compare the affinity of L2 for Lyz measured by liquid sample DESI-MS with the value determined by Loo and coworkers using reactive liquid sample DESI-MS [27]. Notably, the value measured using reactive liquid sample DESI-MS, 5.9 × 103 M–1, is 18 times smaller than the value determined by liquid sample DESI-MS (and ITC). It has been suggested that the short time available for protein and ligand mixing in reactive liquid sample DESI (estimated to be <2 ms) might be insufficient for equilibration of the binding reaction [45]. To help rule out other alternative explanations, in particular the possibility of in-source dissociation, reactive liquid sample DESI-MS was carried out in the present study to measure the affinity of L2 for Lyz. The experimental and instrumental conditions were identical to those used for the liquid sample DESI measurements described above, with the exception that L2 was absent in the sample solution but present in the spray solvent. Shown in Supplementary Figure S5 is a representative reactive liquid sample DESI mass spectrum acquired for an aqueous ammonium acetate (20 mM) solution of Lyz (10 μM) and Ubq (5 μM); the spray solvent was a 50/50 water/acetonitrile solution containing L2 (50 μM). Ion signal corresponding to protonated and sodiated L2 monomer, dimer, and trimer was detected, along with protonated ions of Lyz and (Lyz + L2), at charge states +6 to +8, and Ubq and (Ubq + L2), at charge state +4. Following correction for nonspecific carbohydrate–protein binding, the Ka value was determined to be (7.9 ± 0.4) × 103 M–1, which is similar to the value reported by Loo and coworkers [27]. Given that the instrumental conditions were identical to those used for the liquid sample DESI-MS measurements, it can be concluded that the lower affinity is not due to artifacts associated with instrumental conditions, such as in-source dissociation of the protein–carbohydrate complexes. This finding further supports the suggestion that the lower affinity is, in fact, a kinetic artifact owing to the insufficient time in the droplets for the protein–ligand binding equilibrium to be established [20].

4.4 Tolerance of Liquid Sample DESI-MS to Nonvolatile Salts

The influence of nonvolatile salts on the performance of liquid sample DESI-MS for protein–carbohydrate binding measurements was assessed by through binding measurements performed on Lyz and L2 in aqueous solutions containing varying concentrations of PBS. Shown in Figure 4a, b and c are representative liquid sample DESI mass spectra acquired for aqueous solutions of ammonium acetate (20 mM), Lyz (40 μM), L2 (30 μM) and 0.1× PBS, 0.5× PBS, or 1× PBS. For comparison purposes, ESI mass spectra were also acquired for these solutions (Figure 4d, e and f). At PBS concentrations up to 1×, abundant signal corresponding to protonated ions of Lyz and (Lyz + L2) at charge states +7 to +9 was detected. However, adducts corresponding to the attachment of Na+, NaCl, K+, and KCl were also evident and their abundances increased with increasing PBS concentration. The Ka values determined at each PBS concentration are (1.0 ± 0.1) × 105 M–1 (0.1×), (5.0 ± 0.2) × 104 M–1 (0.5×), and (4.1 ± 0.2) × 104 M × 1 (1×). The value at 0.1× PBS is similar to the one measured in aqueous ammonium acetate (20 mM); however, the values at the higher PBS concentrations are about a factor of two smaller. A similar phenomenon was reported by Oldham and coworkers, who proposed that alkali metal ions may destabilize Lyz–ligand complexes in the gas phase [43]. To establish whether the lower values measured at higher PBS concentrations were reflective of an actual decrease in affinity, ITC was used to measure Ka in a solution of 1 × PBS (Supplementary Figure S4). Notably, the ITC-derived value of (8.4 ± 0.2) × 104 M–1 confirmed that the interaction between L2 and Lyz is slightly weakened at high PBS concentrations, although the magnitude of the effect is not as pronounced as suggested by the liquid sample DESI-MS measurements. Nevertheless, the affinity measured by liquid sample DESI-MS for the 1× solution agrees within a factor of two of the ITC value, indicating that liquid sample DESI-MS can be used to quantify protein–carbohydrate interactions in solutions containing relatively high concentrations of nonvolatile salts. The situation is very different in the case of the ESI mass spectra, which reveals the presence of significant nonspecific adducts (Figure 4d, e and f). In fact, even for the 0.1× PBS solution, it was not possible to positively identify ions corresponding to the (Lyz + L2) complex, making it impossible to quantify the binding interaction. Further investigation revealed that the direct ESI-MS binding measurements were restricted to PBS concentrations of less than 0.02 X .

Figure 4
figure 4

Representative (a), (b), and (c) liquid sample DESI, and (e), (f), and (g) ESI mass spectra acquired for aqueous solutions (pH 7.4 and 25°C) containing Lyz (40 μM) and L2 (30 μM) in 20 mM ammonium acetate and 0.1× PBS [(a) and (d)], 0.5× PBS [(b) and (e)], and 1× PBS [(c) and (f)]. For the liquid sample DESI-MS measurements, the ESI spray solution was 50/50 water/acetonitrile

The differences in the appearances of the liquid sample DESI and ESI mass spectra can be rationalized by considering the differences in the initial composition of the droplets in each case. In the ESI-MS experiments, the initial droplets will contain concentrations of buffer that are similar to that found in bulk solution, with some enrichment in cations (Na+ and K+) expected as a result of the applied electric field [19]. As a result of solvent evaporation, the concentration of buffer components in the droplets will further increase, with the highest concentrations found in the offspring droplets produced late in the ESI process [19]. In contrast, in liquid sample DESI-MS, the initial ESI droplets are devoid of buffer and contain only water and acetonitrile. It is only through collisions with the sample solution that buffer components are transferred to the ESI droplets. Consequently, the overall concentration of PBS in the droplets of liquid sample DESI that lead to the formation of gaseous protein ions is expected to be significantly lower than in the case of the direct ESI-MS measurements.

5 Conclusions

The application of liquid sample DESI-MS for quantifying protein–carbohydrate interactions in aqueous solutions is described. Notably, the affinities of tri- and tetrasaccharide ligands for Lyz and scFv measured using liquid sample DESI-MS are found to be in good agreement with values measured by ITC and the direct ESI-MS assay. It is also found that the reference protein method, which was originally developed to correct ESI mass spectra for the occurrence of nonspecific ligand–protein binding, can be used to correct liquid sample DESI mass spectra for nonspecific carbohydrate binding. The tolerance of liquid sample DESI-MS for quantitative binding measurements carried out using solutions containing high concentrations of PBS was also explored. The binding between Lyz and a trisaccharide ligand was successfully measured with liquid sample DESI-MS at concentrations up to 1× PBS. In contrast, direct ESI-MS binding measurements were limited to PBS concentrations less than 0.02X PBS.