1 Introduction

Spinal fusion surgery is a common treatment for spinal instability. Before this, many scholars have conducted a series of experimental studies and biomechanical analysis of fusion surgery. However, spinal fusion surgery with bilateral pedicle screw fixator (BPS) has several problems, including adjacent segment disease and implant breakage [1,2,3]. Some clinical reports indicated that spinal fusion might accelerate the degeneration of adjacent segments [4, 5]. These rigid fixators withstood most of the load, which reduced the load shared by the fused segment and produced the stress shielding effect [6]. Strong fixator also changed the load-bearing of intervertebral discs and facets at the adjacent segments and increased range of motion (ROM) and disc pressure at adjacent segments. It resulted in a series of problems, such as the decrease of bone fusion rate, Osteoporosis, and atrophy of fixed segment, degeneration of adjacent intervertebral discs and facet joints, screw breakage and rod breakage caused by excessive stress concentration [7]. To increase load sharing, many scholars have applied a variety of low stiffness fixators in biomechanical and animal experiments and have achieved encouraging results [8, 9]. However, low stiffness fixators cannot provide spinal stability. Besides, some flexible posterior spinal fixator systems have been developed and put into clinical practice [10]. These implants were designed to restore spinal stability and decreased the load on adjacent segments to reduce the damage to the lumbar spine caused by the fixator system. Some scholars believed that a semi-rigid pedicle screw fixator and a rigid pedicle screw fixator may provide spinal stability, but the former can reduce the stress shielding effect [11]. However, some scholars questioned the spinal stability. Dynesys pedicle screw fixator system was used to treat 50 cases of degenerative lumbar instability. Two years of follow-up showed that the reoperation rate of Dynesys system was also high [12].

Meanwhile, Chen et al. used topology optimization to design a new intervertebral cage to allow more bone graft volume to be placed [13]. The disadvantage was that the stress of the new cage was greater than that of the RF cage. The optimization of the implant directly affected the biomechanical properties of the spine. Furthermore, Lin et al. demonstrated the ability of topological optimization to change the biomechanics of the lumbar fusion cages [14]. Topology optimization algorithms generated an optimized material distribution for the set loads and constraints within the given design space [15, 16]. Topology optimization can distribute materials preferably, reduce stress concentration, change stiffness, and achieve different biomechanical behavior by changing the structure. Clinically, some reports have demonstrated that spinal fusion surgery accelerated the degeneration of adjacent segments because of the higher stiffness of rigid rods [4, 5]. The stress concentration caused by the rigid rod caused some problems such as broken screws and broken rods [17]. In this study, the stiffness of the connecting rod was reduced appropriately by reducing the volume of the connecting rod without affecting the lumbar motion, and the overall stiffness of the connecting rod that had been reduced in volume was maximized by adjusting the structure using topology optimization. This study used ABAQUS 6.14 (Dassault Simulia, Providence, USA) to design a new rod and designed to effectively reduce the stiffness of the rod to alleviate the above-mentioned problems. Therefore, the objective of this study is to design a new lumbar spinal fixator by using topology optimization (TOD fixator) to reduce overall stiffness and volume and solve the various problems [6, 18, 19] existing in traditional spinal fixators.

2 Materials and Methods

2.1 Finite Element Modeling and Materials

In this study, a three-dimensional nonlinear L1–L5 finite element model (FEM) of the intact lumbar spine was used. The geometry of the spine was obtained by computer scanning tomographic specimens [20]. The model mainly includes vertebrae, intervertebral discs, endplates, and various ligaments, as shown in Fig. 1. Each vertebral body in L1–L5 is composed of the outer cortical bone and the inner cancellous bone. The cortical bone was C3D8 elements and its thickness is 0.5–1.0 mm. Cancellous bones are tetrahedron elements. Intervertebral disc consists of the annular matrix, annulus fibrosus, and nucleus pulposus. Fibers are crisscrossed in the annular matrix. The Young's modulus of the fiber decreases proportionally from the outer layer to the inner layer. The nucleus pulposus is an incompressible substance. There is a 0.5 mm thick endplate between the vertebral body and the intervertebral disc (Fig. 1). The facet joints were modeled by surf-to-surf. The material properties of lumbar ligament is low elastic. The material properties were assumed to be homogeneous and isotropic [21,22,23], and the data were adopted from the literature and are given in Table 1.

Fig. 1
figure 1

Finite element model of the intact human L1–L5 lumbar spine

Table 1 Material properties of spinal components

2.2 The Validation of FEM

In this paper, based on the existing in vitro cadaveric experiments and numerical analysis, the L1–L5 spine finite element model was verified. The present model results are in good agreement with published results, including experimental data and numerical results [24, 25]. The investigations of Renner et al. [24] and Dreischarf et al. [25] presented the compressive deformation, intradiscal pressure, and range of motion of each segment of L1–L5 under follower loads. In summary, the present developed model can be used for further analysis. The detailed model verification results are shown in Table 2.

Table 2 Comparison of the results for the present study with the experimental data of Renner et al. [24] and numerical results of Dreischarf et al. [25] under static loading conditions

2.3 New Lumbar Spinal Fixator by Using Topology Optimization

Topology optimization theory is to satisfy the objective function as far as possible under constraint conditions. Structural energy is used to measure the stiffness of the rod, and minimizing the structural energy is equivalent to maximizing overall structural stiffness [13]. In this paper, topology optimization is to minimize the structural energy of the rod, and its structural volume (V) is the constraint.

$${\text{C}}\left( {\text{a}} \right) = {\text{EA}}/l,\;\left( {{\text{N}}/{\text{mm}}} \right)$$
(1)
$${\text{C}}\left( {\text{b}} \right) = {\text{3EI}}/l^{2} ,\left( {{\text{N}}/{\text{mm}}} \right)$$
(2)

where l the length, A the area, E Yong’s modulus, I the second moment of area, C(a) the axial stiffness of rod, C(b) the bending stiffness. Based on the above formula, the volume or parameters of the connecting rod directly affect its stiffness [26, 27]. Therefore, the reduction in the stiffness of the connecting rod is determined by the decrease in its volume (the constraint), and topology optimization is to maximize the structural rigidity of the rod that has been reduced in volume by adjusting the structure. Appropriate reduction of the rod volume (the constraint) determines that the stiffness of the rod can be decreased and ensures that the lumbar motion cannot be affected. Some numerical studies indicated that the stiffness of the connecting rod greater than 1000 N/mm had only a minor effect on intersegmental rotation [28, 29]. An in-vitro Biomechanical Study by Kwonsoo et al. investigated the effect of the connecting rod (Φ5.5 × L42.5 mm) stiffness on the motion of the lumbar spine instrumented with pedicle screw fixator at L4–L5 level. They found that the range of motion was only marginally increased when the implant stiffness beyond 70 GPa (aluminum) [30]. Since the bending stiffness of rod directly affects the stability of the diseased lumbar spine and its axial stiffness is much larger than the bending stiffness, many studies and present research have focused on the bending stiffness of the rod. In the present study, based on the parameters of rod, the change of the connecting rod stiffness between 5220 and 8947 N/mm may marginally affect the range of motion of the lumbar spine. To ensure the stability between the screws and the connecting rod, only the area about 16.5 mm in the middle of the rod was reduced, as Fig. 3 shown. Based on the formula of calculating the bending stiffness of the rod (2), when the volume of the middle area of the rod was reduced by 50%, the maximum stiffness of rod was approximately 7582 N/mm which may not affect the range of motion of the lumbar spine. Therefore, the process was expected to reduce the volume of the area approximately 16.5 mm in the middle of the rod by 50% and iterate 30 times. Convergence tolerance is defined as 0.0001.

Topology optimization is as follows: objective function: minimize Uc, Uc: the energy of structural compliance; limitation: 0 < ηi < 1 (i = 1,2,3,…,n), ηi: the internal pseudo-densities that are assigned to each element (i) in the topology

$$V = \sum\limits_{{\text{i}}}^{{\text{n}}} {\eta_{i} V_{i} }$$
(3)
$$V \le V_{0} V^{*}$$
(4)
$${\text{E}}_{{\text{i}}} = {\text{E}}\left( {\eta_{i} } \right)$$
(5)
$$\left\{ {\sigma_{i} } \right\} = \left[ {{\text{ E}}_{i} } \right]\left\{ {\varepsilon_{i} } \right\}$$
(6)

V: the computed volume, V0: the original volume, V*: the amount of material to be removed, Vi: the volume of element (i), Ei: the elasticity tensor for each element, E: the elasticity tensor, σi: the stress vector of element (i), εi: the strain vector of element (i).

The density variable, ηi varied between 0 and 1, where ηi near 1 represents the material to be retained, while ηi close to 0 represents the material that should be removed. (Fig. 2a) Topology optimization was carried out in flexion, extension, lateral bending, and torsion. Then, the new rod structure deriving from topology optimization was called the topological rod (Fig. 2b). Compared to the rigid rod (Φ5.5 × L42.5), the volume of the topology rod was reduced by approximately 19.8%. From the results of topology optimization, the effective materials were located on anterior–posterior and outside direction of the rod. Therefore, the invalid material located inside was removed, that is, the topology optimization of the connecting rod. Based on our calculation, the safety factor of the new connecting rod is sufficient to meet the actual engineering needs.

Fig. 2
figure 2

a The rigid rod in the L4–L5 motion segment, the arrow indicates the location of redundant material in inside aspect of spinal model, b the new design (topology rod) and the size of the topology rod

2.4 FEM of the Traditional BPS Fixator and New TOD Fixator

The BPS and TOD fixators were bilaterally implanted into the L4–L5 vertebral body finite element model [31, 32]. The reason for the L4–L5 level is that the prevalence of L4–L5 in individuals suffering from lumbar diseases is greater than other intervertebral discs [33, 34]. The traditional bilateral pedicle screw fixator system includes pedicle screws (Φ6 × L55 mm) and longitudinal rods (Φ5.5 × L42.5 mm) (Fig. 3a) [35]. Titanium (E = 120,000 MPa, v = 0.3) material properties were defined for pedicle screws and longitudinal rods. The TOD fixator system includes pedicle screws and rods designed by topology optimization (Φ5.5 × Φ4 × L42.5 mm) (Fig. 3b). Cage (L16 × W10 × H8 mm) provided support stability for intervertebral fusion. PEEK (E = 3600 MPa, v = 0.3) material properties were defined for the cage. This study adopted the single cage method [36], one of the commonly used fusion methods, to improve the stability of the lumbar spine.

Fig. 3
figure 3

(a) The finite element model of L1–L5 fusion spine with BPS fixator (b) The finite element model of L1–L5 fusion spine with TOD fixator

In this paper, the model of pedicle screw was simplified and structures such as threads and chamfers that have little effect on mechanical properties were removed. To implant the cage and pedicle screw, one of the facet joints between L4 and L5 vertebral bodies was removed. Moreover, the intact lumbar spine model (Intact) was created to compare with the BPS model and the TOD model.

2.5 Loading and Boundary Conditions

The constraint type ‘Tie’ was defined between pedicle screws and vertebral bodies, as well as pedicle screws and longitudinal rods. To simulate the condition after fusion, the interfaces between cage and endplates were bonded via node sharing. The lower surface of the L5vertebral body was fixed throughout the simulation process. The nodes on the upper surface of L1 were combined with reference points for load application. A preload of 150 N was applied to the superior surface of the L1 level [28]. The nodes on the uppermost surface of the L1 vertebra were coupled to a reference node for load application. A 10Nm moment was applied to this reference node. And the flexion, extension, torsion, and lateral bending of the model were achieved by changing the direction of torque.

3 Results

The numerical results about adjacent segment disease and stress concentration such as the disc stress, intradiscal pressure, and Von-Mises stress distributions of rod and pedicle screw were compared and listed. Essential indexes such as range of motion and stiffness of the spine were used to assess the stability of the structure and stable environment for the fused vertebrae. The Von-Mises stress distributions of the vertebrae with TOD fixator and BPS fixator under the different loads were compared and analyzed. The stress in L4/5 endplates was used to assess the load sharing of the TOD and BPS model.

3.1 Range of Motions

The ROMs of the fusion lumbar segment (L4–L5) in the TOD and BPS models were all under 1.06º and were reduced in all motions when compared to Intact. For the fused segment (L4–L5) level, the ROMs of TOD were similar to those of BPS. For example, in all motions, the ROMs of L4/5 were 0.65, 1.05, 0.68, and 0.87 degrees for TOD, and 0.57, 1.06, 0.52, and 0.88 degrees for BPS. For the adjacent segments (L1–L2, L2–L3, and L3–L4), the TOD model generated smaller ROMs in all motions compared with the BPS model. The stiffness of the BPS and TOD were, at most, 1.83 and 2 times the stiffness of the intact model in all motions (Table 3).

Table 3 ROMs and stiffness comparison for each motion segment in the three FE models

3.2 Stress in Intervertebral Discs and Facet Contact Forces

It was found in Fig. 4a, b that the stress of the L2/3 and L3/4 discs in the TOD model were smaller than those of the BPS model in all motions. Especially in torsion, the stresses of the L2/3 and L3/4 disc were 0.212, 0.236 MPa for the TOD model, and 0.458, 0.469 MPa for the BPS model. The TOD model decreased the intradiscal pressure of L2/3 and L3/4 discs in all motions compared with the BPS model.

Fig. 4
figure 4

Stress in intervertebral discs and intradiscal pressure of the L2/3 and L3/4 levels for the three FE models

As illustrated in Fig. 5a, b that the facet contact force between L2 and L3 in the TOD model were smaller than the BPS model in extension, torsion, and lateral bending. For example, the L2/3 facet contact forces were 32.9, 156, and 44.4 N in extension, torsion, and lateral bending for TOD, and 52.8, 204, and 52.6 N for BPS. The force between L3 and L4 of TOD in extension and torsion (196 and 223 N) decreased the facet contact force compared to the BPS model (205 and 203 N). The TOD model increased the Von-Mises stress of L4/5 endplates compared to the BPS model (Fig. 5c, d). Especially in flexion, the stresses of the L4 inferior and L5 superior discs were 0.37 (119%), 0.361 (118%) MPa for the TOD model, and 0.312 (100%), 0.305 (100%) MPa for the BPS model. This implied that the contact stress of TOD model at the fusion level increased by 18% compared with the BPS model, and this might also improve the stress shielding effect.

Fig. 5
figure 5

Facet contact forces of the L2/3 and L3/4 levels and stress in L4/5 endplates for the three FE models

3.3 Stress in Pedicle Screws and Rods

It was found in Fig. 6 that the stress was concentrated at the neck of the pedicle screw for TOD and BPS models in all motions. In flexion, extension, torsion, and lateral bending, the pedicle screw of TOD exhibited smaller stress than it did in BPS (Fig. 6). The max stresses of screws in TOD were 52%, 64.8%, and 79.3% of those of BPS in flexion, torsion, and lateral bending, respectively. What’s more, in torsion, the TOD model decreased the high-stress region in screws relative to the same region in the BPS model (Fig. 6). It was observed in Fig. 7 that the stress concentration region was located in the middle of the rod in all motions, and TOD decreased the stress concentration region of rods compared to BPS. Moreover, TOD reduced the maximum stresses of the rod in flexion, extension, torsion, and lateral bending compared to those of BPS, and the difference, at most, can be up to 50.3% in torsion.

Fig. 6
figure 6

Screw stress distribution in pedicle screws for all motions. The arrowhead indicates the position of maximum stress (left: BPS system, right: TOD system)

Fig. 7
figure 7

Screw stress distribution in rods for all motions (left: BPS system, right: TOD system)

3.4 Stress in L4 and L5 Vertebra Body

In order to better illustrate the impact of the pedicle screw on the L4 and L5 vertebral bodies, the positions of four screws were marked as position1 (upper left), position2 (lower left), position3 (upper right), and position4 (lower right), in Fig. 8, and the stresses of the position 1,2,3,4 in BPS were compared with the stresses of same positions in TOD, respectively. The stress concentrations of the L4–L5 vertebral body in BPS and TOD were mainly located at the junction between screws and the vertebral bodies, as shown in Figs. 8 and 9 (Positions 1, 2, 3, and 4). The stresses of the position1, 2, 3, and 4 in TOD were smaller than the stresses of the same position in BPS in all motions. And in torsion, the stresses of the position1, 3 in BPS were 2.03 and 1.39 times of those in TOD (Fig. 9).

Fig. 8
figure 8

The von-Mises stress distribution of the L4–L5 of TOD and BPS vertebrae in flexion and extension

Fig. 9
figure 9

The von-Mises stress distribution of the L4–L5 vertebrae of TOD and BPS in torsion and lateral bending

4 Discussion

At present, the rigid traditional bilateral pedicle screw lumbar fixator bears most of the load, which is prone to lead to degenerative disc and facet joint disease at the adjacent segments. At the same time, excessive concentration of stress may lead to problems such as broken rods and broken screws. To solve these problems, all kinds of new fixators came into being, but there were many problems. In this paper, a new lumbar fixator was designed by topology optimization to alleviate the above problems. Compared to the rigid rod, the volume of the topology rod was reduced by approximately 19.8%. The new lumbar fixator saved material and avoided damaging the tissue around the spine to a greater extent.

ROMs and facet contact forces were the main indexes to evaluate the performance of the fusion spine. Freudiger et al. found that the BPS model reduced ROMs compared to the intact spine in flexion, extension, and lateral bending [37]. Our results agree with previous studies including experiments and biomechanical analysis. The ROMs of the TOD model were similar to the BPS model. As far as ROMs are concerned, the ability of the TOD model is obviously not losing to the BPS model. In terms of the facet contact force, the performance of the BPS model was consistent with the result of Rohlmann et al.’s study [38]. They found that the bilateral pedicle screw fixator system increased facet joint forces at other levels for axial rotation and extension. In this paper, the results about the facet contact force showed that the new topology-optimized lumbar fixator significantly reduced the facet contact forces at adjacent segments, thereby reducing the possibility of degenerative lesions. Judging from previous tests and finite element analysis results, the stresses of the discs at adjacent segment increased when BPS was used, and our result was consistent with the earlier experimental and numerical results that the stress on disc and facet joints of the adjacent segments increased after implanting the pedicle screw fixator [39, 40]. The stresses and intradiscal pressures in the L2/3 and L3/4 disc (the adjacent segments) of the TOD model were less than those of the BPS model in all motions, especially in torsion. Based on our results, we believe that the new fixator may reduce the risk of adjacent segment disease and degenerative lesions caused by traditional BPS fixator. The TOD model increased the stress in the fused endplates (L4 inferior and L5 superior endplates) compared with the BPS model, especially in flexion. This implies that the new fixator increase the load sharing of the vertebrae and reduce the stress shielding effect compared with the traditional bilateral pedicle screw fixator. The traditional BPS fixator may provide stability for the fixed segment, but the rigid fixator led to stress concentration and led to broken screws and rods [7, 41, 42]. In this study, the stress concentration regions in rods and pedicle screws of TOD and BPS were in the middle of rods and the neck of the pedicle screws. And some researches predicted the same trend about the stress concentration region in rods and pedicle screws [43, 44]. The pedicle screws and rods in TOD exhibited smaller stress than it did in BPS in all motions. Especially in flexion and torsion, the maximum stresses in screws and rods of TOD reduced to half of screws and rods in BPS. The properties of the pedicle screw and vertebral cortical bone were different, and the pedicle screw may directly damage the vertebral cortical bone [45]. The stresses in the junction between pedicle screws and vertebral bodies are usually larger, so the vertebral bodies tend to be damaged easily. In this paper, the stress concentration regions in vertebral bodies of BPS and TOD were all located at the junction between screws and vertebral bodies. This result was consistent with the study of Chen et al. that the peak stress of vertebrae was located at the junction of the screw and bone, which was consistent with the location of bone fracture observed in a clinical setting [46]. The maximum stresses in vertebral bodies of the TOD model were similar to those of BPS in flexion, extension, and lateral bending. But in torsion, the maximum stress in the vertebral bodies of TOD decreased obviously compared with the BPS model. Therefore, we infer that the new instrumentation may decrease the likelihood of rod breakage, screw breakage, and vertebral injury compared to the traditional BPS fixator.

The potential limitations in this study are as follows. The linear simplification or piecewise linear simplification used in this study for the material property of spinal ligaments and intervertebral discs with non-linear behaviors might influence the accuracy of calculation results. The spinal ligaments and intervertebral discs show obvious nonlinearity, such as viscoelastic, poroelastic, degenerative, etc. In fact, in many finite element analyses [47], the linear simplification of material property of human spine tissues was still adopted. This simplification did not make a high influence on the accuracy of calculation results if choosing appropriate linear material parameters. And the fact that our results were consistent with the in vitro experiments confirms the reality of our spine FE model. The threads and the chamfer on the screws and cage were simplified. The small impact of the threads and chamfer was not enough to affect the overall results.

5 Conclusions

In this study, the topology optimization method was used to optimize the connecting rod, and a new lumbar fixator was designed to solve the problems of traditional bilateral lumbar fixator, such as the degenerative disease of the adjacent segments and stress concentration. The results of biomechanical finite element analysis showed that compared with the traditional fixator, the new fixator may reduce intradiscal pressure, the stresses of adjacent intervertebral discs, and the facet contact forces, thus reducing adjacent segment disease, degenerative lesions, and the stress shielding effect. What’s more, the new lumbar fixator may decrease stress concentration and reduce the likelihood of rod breakage, screw breakage, and vertebral injury. The topological rods in this study are less volume, more economical, and avoid damage to vertebral tissue. The present study may provide useful information for alleviating adjacent segment disease and stress concentration of the fixator in lumbar interbody fusion.