Abstract
Background
Estimation of left ventricular ejection fraction (LVEF) using equilibrium radionuclide angiography is an established method for assessment of left ventricular function. The purpose of this study was to establish normative data on left and right ventricular volumes and ejection fraction, using cadmium-zinc-telluride SPECT camera.
Methods and results
From routine assessments of left ventricular function in 1172 patients, we included 463 subjects (194 men and 269 women) without diabetes, previous potentially cardiotoxic chemotherapy, known cardiovascular or pulmonary disease. The lower limits defined as mean value minus two standard deviations for ventricular ejection fraction and end diastolic volume were LVEF (men: 50%, women: 50%), LEDV (men: 45 mL, women: 40 mL), RVEF (men: 29%, women: 28%), and REDV (men: 73 mL, women: 57 mL).There was a significant negative correlation between age and both left and right ventricular volumes in women (r = −0.4, P < .001) but only for right end systolic ventricular volume in men (r = −0.3, P = .001).
Conclusion
A set of reference values for cardiac evaluation prior to chemotherapy in cancer patients without other known cardiopulmonary disease is presented. There are age-related changes in cardiac dimensions with age depending on gender, although with only limited influence on LVEF or RVEF.
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Avoid common mistakes on your manuscript.
Introduction
The continuous development of more efficient cancer therapies and effective screening has led to an impressive rise in long-term cancer survivors.1 Thus, focus is increasingly being directed on the management of potentially long-term side effects, i.e., cardiotoxicity and chronic heart failure.1,2 There is an increasing need for more precise evaluation of cardiac function in order to navigate between the risk of cardiotoxicity and otherwise potentially lifesaving therapy.3
Cardiotoxicity especially with anthracyclines and trastuzumab is often monitored via measuring left ventricular volumes and function,2,3 primarily focusing on left ventricular ejection fraction (LVEF). This is traditionally performed non-invasively with echocardiography, magnetic resonance imaging (MRI) or radionuclide angiography with either planar acquisition or single photon emission computer tomography (SPECT).
The use of serial MUGA to prevent heart failure in patients receiving cardiotoxic drugs is well established and was originally proposed in 1979 by Alexander et al in 55 patients receiving doxorubicin for treatment of cancer.4 Later Schwartz el al5 suggested guidelines for serial MUGA’s and when to stop therapy. They showed adhering to these guidelines could reduce incidence of heart failure. A finding that still holds in the modern era.6 Evidence also suggest that long-term (>3 months post dosing) monitoring is also warranted at least in patients receiving epirubicin.7 The importance of the radionuclide angiography in evaluating LVEF is also highlighted by the use of this technique in large clinical oncological trials—i.e., in the evaluation of trastuzumab (Herceptin Adjuvant Trial).8
There are no current evidence-based guidelines stipulating how and when to monitor cardiac function. The most resent multicenter studies have used multigated acquisition scans (MUGA) or echocardiography.1,3 MUGA is often regarded as a non-invasive gold standard for the evaluation of cardiotoxicity, due to higher reproducibility and lower interobserver variability than echocardiography.3,9
Advances in radionuclide angiography include SPECT, which allows for calculation of additional variables on end systolic (ESV) and end diastolic volumes (EDV) for both ventricles.3,9,10 Since 2005, new dedicated cardiac gamma camera types equipped with cadmium-zinc-telluride (CZT) detectors have been commercially available, with improved spatial resolution, faster acquisition times and potentially lower radiation doses as well as improved reproducibility.11-13 This calls for establishing normative data for this new camera type given the fact that its increased spatial resolution might give rise to adjustments of previously established limits for normalcy as well as establishing normative data for right ventricular variables.
The primary aim of this study is to establish normative data on patients with cancer referred for measurement of baseline data of LVEF prior to potentially cardiotoxic chemotherapy. The secondary objective was to test for gender and age-specific changes or differences related to blood pressure and heart rate at rest.
Methods
From October 1st 2012 to September 30th 2014, we performed 2691 consecutive routine assessments of left ventricular function in 1172 patients after informed consent was given. Patients were excluded from this study if they had diabetes, pleural effusions, prior lung resection, previous cancer-related chemotherapy and known cardiovascular disease including cardiac insufficiency, valvular disease, myocardial infarction, coronary revascularization, and cardiac arrhythmias treated medically or with pacemaker. In all 463 patients (194 men and 269 women) were included (Figure 1). The study was approved by the Danish data protection Agency.
Information on hypertension was also recorded. The data were retrieved from patient files. Patients with large thoracic circumferences and patients unable to rest with their left arm above the head were scanned on other cameras and could subsequently not be included (Figure 1).
Image Processing
All acquisitions were performed at the Department of Clinical Physiology and Nuclear Medicine at Herlev Hospital. After a 9-month test period from January through September 2012 the CZT-cardiac SPECT gamma camera, GE Discovery 530c (GE Healthcare, Milwaukee, WI, USA) (DNM) was adapted for routine measurements of left ventricular function as previously described.11,12 In brief, a dose of 550-600 MBq (15-16 mCi) 99mTc-labeled human serum albumin (HSA) was administered intravenously to each patient. An acquisition protocol for multigated acquisition, using 16 bins and requesting 600 accepted beats was adapted and a 20% energy window centered on 140 keV was performed in all cases. A Xeleris 3 Imaging workstation (GE Wauwatosa, WI, US) equipped with software version no. 3.0562 was used for reorientation of the heart after applying a generic reconstruction algorithm and filtering. The matrix was resized from 74 × 74 to 64 × 64. Assessment of the left and right ventricular ejection fraction as well as estimations of end diastolic and end systolic volumes was carried out with the Cedars-Sinai QBS processing software (Cedars-Sinai, Ca, USA—revision 2009.0).
Each acquisition was analyzed twice by two experienced technologists. If LVEF differed more than 2% points between the two analyses, repeated processing was performed until sufficient agreement was obtained. All results were revised to ensure that the region of interest (ROI) automatically drawn by the software around the ventricles, were in accordance with visually estimated ventricular borders
Statistical Analysis
A paired two-tailed t test was applied to both datasets and used to compare differences between groups with normally distributed variables. One-way ANOVA was used to compare variations between more than two groups and changes in cardiac variables with age.
Pearson correlation was performed to evaluate relationship between continuous variables. A probability of less than .05 was regarded as significant.
Results
Demographic data, selected cardiovascular variables, smoking status, and prior medical history regarding hypertension and hypercholesterolemia are presented in Table 1 divided by gender. Men were older, taller, and had higher blood pressure than women. The majority of patients with hypertension on the day of investigation had only mild hypertension (81/114 patients) and 9 patients had severe grade 3 hypertension. There were no differences in ventricular volumes or ejection fraction (EDV, ESV, and LVEF) between age, gender, height, and weight matched hypertensive and normotensive patients (P = .42, .70, and .23).
Left and right ventricular volumes and ejection fraction on the included patients are presented in Tables 2 (women) and 3 (men). Overall differences between age groups as calculated by one-way ANOVA are presented by an overall P value in the far right column of each Table.
Figures 2A and 3A depict normal ranges for LVEF, LESV, and LEDV with 10 years intervals for men (men aged <45 were pooled into one age group because of low numbers) and women, respectively. Similar figures for right ventricular variables are given in Figures 2B and 3B.
Both right and left ventricular volumes declined significantly with increasing age for women along with a significant increase in LVEF and a trend toward an increase in RVEF. There was no significant development in left ventricular volumes or ejection fraction with increasing age for men, whereas there was a decline in right ventricular end systolic volume and an increase in RVEF with increasing age (r = −0.25, P = .001, and r = 0.29, P < .001, respectively). Correlation matrices are presented for each gender in Tables 4 and 5.
A high heart frequency was associated with low right and left ventricular volumes and higher ejection fractions. A high systolic blood pressure was associated with lower right ventricular volumes and higher ejection fractions in both genders. For women this relationship was also found for left ventricular variables. The distribution of patients in overall cancer groups is given in Table 6.
Discussion
With the data presented above it is our intention to provide reference values for chemotherapy naïve cancer patients over a wide range of age and for both genders using a new and very reproducible technique.
We have not directly compared our values with MRI or echocardiography. Instead, we must relate our findings to previous studies using either CZT-camera technology, radionuclide angiography SPECT, or estimates of left ventricular function based on gated myocardial perfusion scintigraphy (MPI).
Cadmium-Zinc-Telluride-Cameras
To our knowledge, this is the first attempt to produce reference values for gated tomographic radionuclide angiography using a CZT-detector gamma camera. Previously, only re-projecting gated blood-pool SPECT from CZT-camera to standard 2D-planar radionuclide angiography have been described.14 Absolute values were not given, but the authors found excellent correlation with the reprojected data compared to traditional imaging with sodium-iodine-cameras (NaI-cameras). It has also been shown that the CZT-camera is superior to both planar and SPECT radionuclide angiography obtained with traditional NaI-cameras regarding intra- and interobserver variation.12 Both SPECT systems tended to produce higher LVEF than the planar acquisition, and NaI-SPECT marginally higher values than CZT-SPECT. It has been also demonstrated that radionuclide angiography using CZT-cameras has excellent reproducibility when estimating left ventricular volumes and ejection fraction from repeated scans comparable to the gold standard MRI.11
CZT-camera technology has foremost been developed and used for myocardial perfusion imaging. The higher sensitivity, better energy resolution and improved spatial resolution allows for faster acquisition times,15 lower radiation doses and fewer equivocal results.16 Left ventricular volumes and ejection fraction estimated from perfusion imaging using CZT-cameras has been shown be highly correlated with standard NaI-SPECT15 and cardiac MRI.17-19 However, CZT tends to underestimate the EDV and ESV compared to MRI (i.e., 10-30 mL for EDV). This could be due to incomplete segmentation of base and outflow tract on SPECT as these structures contain fewer myocytes.19 Still, LVEF did not differ between CZT-SPECT and MRI.17-19
Comparing absolute values for the left ventricle, our findings are in agreement with those found on perfusion CZT-SPECT and MRI by Cochet et al19 but lower than Giorgetti et al18 and Baillez et al.17 However, in Baillez et al more than half of the patients had known LV hypertrophy, dilated, or ischemic cardiomyopathy and the majority of Giorgetti et al’s patients had known ischemic heart disease and were older, thus not comparable to our population.
Gated Radionuclide Angiography SPECT Using NaI-Cameras
First and foremost the presented data are in good accordance with results from De Bondt et al20 who used normal volunteers for assessment of gender-specific data on right and left ventricular volumes also using gated tomographic radionuclide angiography on NaI-cameras. RVEF, RVEDV, RVESV, and LVEF are all similar to those measured by De Bondt et al, whereas we found slightly smaller left ventricular volumes.
In 55 patients referred for routine LVEF assessment Harel et al21 conducted both gated radionuclide angiography SPECT, planar radionuclide angiography and cardiac MRI. Overall there was excellent correlation between LVEF, LESV, and LEDV between the different modalities. However, the radionuclide angiography tended to underestimate volumes and overestimate LVEF compared to cardiac MRI. The absolute values for ventricular volumes were higher and LVEF lower than our findings. However, their population had a high proportion of patients with ischemic cardiomyopathy (49%) and previous CABG or PCI (59%).
Gated Perfusion SPECT (MPI)
Ababneh et al investigated ejection fraction and left ventricular volumes in 1513 patients undergoing MPI SPECT with either technetium-99m or thalium-201. They found no difference between tracers or acquisition cameras, but did find a significant difference in LVEF and volumes amongst genders with women having smaller ventricles and larger LVEF. They found lower limits for LVEF to be 50% for women and 43% for men, and higher ventricular volume in men.22 Our lower limit for LVEF is a bit higher for men (50%) and similar for women (50%). Both mean values and upper limit of normal of left ventricular volumes are higher in our material for both genders. The discrepancy can be due to different imaging techniques (perfusion SPECT vs blood-pool SPECT) and difference in population. Ababneh used normal MPI in patients referred for angina or dyspnoea as inclusion.
When comparing gated myocardial perfusion tomography with gated blood-pool tomography there is overall a very good correlation between the measured left ventricular volumes and ejection fractions. Whereas the systematic error for LVEF measurement looks to be linear for the whole spectrum of LVEF, the difference in volume estimation is not with a tendency toward larger differences with larger volumes.23,24 Right ventricular estimation is even more troublesome with relatively large biases between the two methods.24
Different processing algorithms have shown, despite being well correlated, to produce varying ventricular volumes and ejection fractions.20,21,25 Often the bias between imaging techniques or software is linear with over- or underestimating “true” volumes, but this is not always the case.21,23,25 Therefore, in optimal circumstances individual patients should only be evaluated using same equipment regarding both hardware and software.
Right Ventricle
In a series of 58 patients26 referred for evaluation of ventricular function both gated radionuclide angiography SPECT, echocardiography and MRI was done, with the latter as the reference standard. Radionuclide angiography showed high correlation with MRI for right ventricular volumes (r values 0.67-0.84), whereas echocardiographic derived indices of volumes and function performed less well. Our absolute values for right ventricular volumes and RVEF was comparable to the SPECT findings in the above series, and especially similar to the MRI findings, although not stratified by gender, confirming the ability of gated radionuclide angiography SPECT to asses right ventricular size and function.
In a selection of healthy volunteers (29% female), however, Kjær et al27 found higher RVEF both on MRI and SPECT and somewhat lower volumes (when presuming BSA of 1.73 m2 on average) than our results. In their experience there was agreement between MRI and radionuclide angiography SPECT for absolute volumes, however, with wide limits of agreement.
Age-Related Changes in Cardiac Variables
In concordance with the findings of De Bondt et al,28 we found a significant negative correlation between age and left ventricular volumes, and positive correlation with LVEF. Both in our study and in that of De Bondt the correlation was only seen in women.
Cardiac MRI have demonstrated similar results with decreases in ESV and EDV for both ventricles together with a significant increase in ejection fraction with advancing age.29-31
Important cardiovascular changes occur with advancing age. Increased vascular intimal thickness and vascular stiffness lead to increased afterload and hypertension. In the heart, the number of myocytes decreases, while the remaining myocyte hypertrophies, thus myocardial mass, remain relatively stable. Left ventricular wall is thickened leading to changes in diastolic filling. Other changes included reduced heart rate variability and reduced SA-node cells and beta-adrenergic modulation.32,33
The changes in left ventricular wall size, compliance and ventricular filling leads to decreased left ventricular volumes but increased ejection fraction, the later perhaps due to changes in diastolic filling behavior and to more twisting and torsion of the left ventricle in the elderly. Another explanation could be reduced physical activity.30-32
In our study, we did not see any age-related change in left ventricular volumes or ejection fraction in men. Perhaps the smaller sample size and more skewed age distribution in our male group could account for this discrepancy.
Gender-Specific Differences
To our knowledge, our study is the first to address any gender difference in ventricular volumes and ejection fraction using CZT-cameras in a large sample of patients. We found that although there was a difference in absolute volumes the lower limit of ejection fraction did not differ between genders.
Giorgetti et al18 reported no gender difference in ventricular volumes in a smaller sample of 55 patients (50% women), however, their population might be too small to identify minor differences. Ababneh et al22 and Sharir et al34 found significant differences in both left ventricular volumes and ejection fraction between men and women, with men having higher volumes and a lower ejection fraction. We did find a similar difference regarding volumes, but identified equal lower limits for LVEF between genders. This could be attributed to differences in population (patients referred for MPI) and method (perfusion vs blood pool).
In a smaller samples of 36 healthy volunteers cardiac MRI have shown that men had larger volumes and ventricular mass than women.35 However, the LVEF was only slightly different and in the authors view relatively independent of gender. Again the small sample size might influence the finding, but overall the finding is in agreement with the presented data above.
In summary, our population is neither normal (cancer diagnosed) nor are they suspected of ischemic heart disease. Some of the large SPECT “normal” materials are patients referred for MPI with an intermediate risk of IHD—with a normal MPI. Thus, they do not have diagnosed IHD, but still are not normal as they where referred for MPI. In theory they could have other cardiac or pulmonary disease or false negative MPI.
The much larger Dallas Heart Study36 conducted cardiac MRI on 1435 women and 1183 men without self-reported valvular abnormalities or congenital defects. They found lower limits for LVEF to be 61% in women and 55% in men, thus higher than our results and with a gender difference. Their populations were at mean younger than ours and much more ethnical diverse. A recent MR-study evaluating a population more similar to ours (Framingham Heart Study, n = 852)—found only small difference in lower limit LVEF between genders (men 59% and women 61%).29
Thus, our results regarding modest gender difference in lower limits of LVEF are conflicting with SPECT studies and to a lesser extend MRI-data. However, the gender difference in lower limits in LVEF seems to be smaller in MR-studies (3-6%) than SPECT studies. As previously discussed CZT offers increased resolution and reproducibility compared to SPECT and accordingly our results with little/no difference in LVEF is closer to the findings from large MRI studies. Other reasons for some of this discrepancy could, as discussed, be difference in techniques and difference in populations. As a consequence, we are currently in the process of setting up a comparison of CZT and MRI.
Limitations
The included patients are without diabetes, arrhythmia, or known cardiovascular or valvular disease. Still, they do not reflect a random sample and the underlying angiographic data cannot be considered complete in order to represent normal values in the traditional sense. However, the presented data still provide reference values for patients as they present themselves to the clinical oncologist before potentially cardiotoxic chemotherapy.
Conclusion
Reference values in a large sample of patients with no known cardiovascular or pulmonary disease is now provided. Despite gender-related differences in ventricular volumes, the lower limit of LVEF and RVEF is similar in the two groups. Finally, there is a decrease in ventricular volumes and a slight increase in LVEF in women with advancing age.
New knowledge gained
Cardiac-dedicated CZT-detector cameras offer improved resolution and improved reproducibility. Although primarily intended for MPI, we here report normal values for gated radionuclide angiography using CZT-detector technique in chemotherapy naïve patients. Overall these values are in accordance with previous reporting’s of gated blood-pool SPECT.
Abbreviations
- CZT:
-
Cadmium-zinc-telluride
- MRI:
-
Magnetic resonance imaging
- MPI:
-
Myocardial perfusion imaging
- MUGA:
-
Multigated acquisition scans
- R/LVEF:
-
Right/left ventricular ejection fraction
- R/LEDV:
-
Right/left ventricular end diastolic volume
- R/LESV:
-
Right/left ventricular end systolic volume
- SPECT:
-
Single photon emission computer tomography
References
Carver JR, Shapiro CL, Ng A, Jacobs L, Schwartz C, Virgo KS, et al. American Society of Clinical Oncology clinical evidence review on the ongoing care of adult cancer survivors: cardiac and pulmonary late effects. J Clin Oncol 2007;25:3991-4008.
Curigliano G, Cardinale D, Suter T, Plataniotis G, de Azambuja E, Sandri MT, et al. Cardiovascular toxicity induced by chemotherapy, targeted agents and radiotherapy: ESMO Clinical Practice Guidelines. Ann Oncol 2012;23:vii155-66.
De Geus-Oei L-F, Mavinkurve-Groothuis AM, Bellersen L, Gotthardt M, Oyen WJ, Kapusta L, et al. Scintigraphic techniques for early detection of cancer treatment-induced cardiotoxicity. J Nucl Med Technol 2013;41:170-81.
Alexander J, Dainiak N, Berger HJ, Goldman L, Johnstone D, Reduto L, et al. Serial assessment of doxorubicin cardiotoxicity with quantitative radionuclide angiocardiography. N Engl J Med 1979;300:278-83.
Schwartz RG, McKenzie WB, Alexander J, Sager P, D’Souza A, Manatunga A, et al. Congestive heart failure and left ventricular dysfunction complicating doxorubicin therapy. Am J Med 1987;82:1109-18.
Mitani I, Jain D, Joska T, Burtness B, Zaret B. Doxorubicin cardiotoxicity: Prevention of congestive heart failure with serial cardiac function monitoring with equilibrium radionuclide angiocardiography in the current era. J Nucl Cardiol 2003;10:132-9.
Jensen BV, Skovsgaard T, Nielsen SL. Functional monitoring of anthracycline cardiotoxicity: A prospective, blinded, long-term observational study of outcome in 120 patients. Ann Oncol 2002;13:699-709.
Piccart-Gebhart MJ, Procter M, Leyland-Jones B, Goldhirsch A, Untch M, Smith I, et al. Trastuzumab after adjuvant chemotherapy in HER2-positive breast cancer. N Engl J Med 2005;353:1659-72.
Mitra D, Basu S. Equilibrium radionuclide angiocardiography: Its usefulness in current practice and potential future applications. World J Radiol 2012;4:421-30.
Clements IP, Akerem Khan SK, Chen HH, Mullan BP, Clements IP. Measurement of left and right ventricular volumes with tomographic equilibrium radionuclide angiocardiography and cardiac MRI. Nucl Med Commun 2012;33:481-5.
Jensen MM, Haase C, Zerahn B. Interstudy repeatability of left and right ventricular volume estimations by serial-gated tomographic radionuclide angiographies using a cadmium-zinc-telluride detector gamma camera. Clin Physiol Funct Imaging 2014. doi:10.1111/cpf.12178.
Jensen MM, Schmidt U, Huang C, Zerahn B. Gated tomographic radionuclide angiography using cadmium-zinc-telluride detector gamma camera; comparison to traditional gamma cameras. J Nucl Cardiol 2014;21:384-96.
Garcia EV, Faber TL, Esteves FP. Cardiac dedicated ultrafast SPECT cameras: new designs and clinical implications. J Nucl Med 2011;52:210-7.
Wells RG, Marvin B, Kovalski G, Ruddy TD. Planar radionuclide angiography with a dedicated cardiac SPECT camera. J Nucl Cardiol 2013;20:358-66.
Esteves FP, Raggi P, Folks RD, Keidar Z, Askew JW, Rispler S, et al. Novel solid-state-detector dedicated cardiac camera for fast myocardial perfusion imaging: Multicenter comparison with standard dual detector cameras. J Nucl Cardiol 2009;16:927-34.
Mouden M, Timmer JR, Ottervanger JP, Reiffers S, Oostdijk AHJ, Knollema S, et al. Impact of a new ultrafast CZT SPECT camera for myocardial perfusion imaging: Fewer equivocal results and lower radiation dose. Eur J Nucl Med Mol Imaging 2012;39:1048-55.
Bailliez A, Blaire T, Mouquet F, Legghe R, Etienne B, Legallois D, et al. Segmental and global left ventricular function assessment using gated SPECT with a semiconductor cadmium zinc telluride (CZT) camera: Phantom study and clinical validation vs cardiac magnetic resonance. J Nucl Cardiol 2014;21:712-22.
Giorgetti A, Masci PG, Marras G, Rustamova YK, Gimelli A, Genovesi D, et al. Gated SPECT evaluation of left ventricular function using a CZT camera and a fast low-dose clinical protocol: Comparison to cardiac magnetic resonance imaging. Eur J Nucl Med Mol Imaging 2013;40:1869-75.
Cochet H, Bullier E, Gerbaud E, Durieux M, Godbert Y, Lederlin M, et al. Absolute quantification of left ventricular global and regional function at nuclear MPI using ultrafast CZT SPECT: Initial validation versus cardiac MR. J Nucl Med 2013;54:556-63.
De Bondt P, Nichols KJ, De Winter O, De Sutter J, Vanderheyden M, Akinboboye OO, et al. Comparison among tomographic radionuclide ventriculography algorithms for computing left and right ventricular normal limits. J Nucl Cardiol 2006;13:675-84.
Harel F, Finnerty V, Grégoire J, Thibault B, Marcotte F, Ugolini P, et al. Gated blood-pool SPECT versus cardiac magnetic resonance imaging for the assessment of left ventricular volumes and ejection fraction. J Nucl Cardiol 2010;17:427-34.
Ababneh A, Sciacca R, Kim B, Bergmann S. Normal limits for left ventricular ejection fraction and volumes estimated with gated myocardial perfusion imaging in patients with normal exercise test results: Influence of tracer, gender, and acquisition camera. J Nucl Cardiol 2000;7:661-8.
Vanhove C, Walgraeve N, De Geeter F, Franken PR. Gated myocardial perfusion tomography versus gated blood pool tomography for the calculation of left ventricular volumes and ejection fraction. Eur J Nucl Med Mol Imaging 2002;29:735-41.
Odagiri K, Wakabayashi Y, Tawarahara K, Kurata C, Urushida T, Katoh H, et al. Evaluation of right and left ventricular function by quantitative blood-pool SPECT (QBS): Comparison with conventional methods and quantitative gated SPECT (QGS). Ann Nucl Med 2006;20:519-26.
De Bondt P, Claessens T, Rys B, De Winter O, Vandenberghe S, Segers P, et al. Accuracy of 4 different algorithms for the analysis of tomographic radionuclide ventriculography using a physical, dynamic 4-chamber cardiac phantom. J Nucl Med 2005;46:165-71.
Anderson K, Prylutska H, Ducharme A, Finnerty V, Grégoire J, Marcotte F, et al. Evaluation of the right ventricle: Comparison of gated blood-pool single photon electron computed tomography and echocardiography with cardiac magnetic resonance. Int J Cardiol 2014;171:1-8.
Kjær A, Lebech A, Hesse B, Petersen CL. Right-sided cardiac function in healthy volunteers measured by first-pass radionuclide ventriculography and gated blood-pool SPECT: Comparison with cine MRI. Clin Physiol Funct Imaging 2005;25:344-9.
De Bondt P, Van de Wiele C, De Sutter J, De Winter F, De Backer G, Dierckx RA. Age- and gender-specific differences in left ventricular cardiac function and volumes determined by gated SPET. Eur J Nucl Med 2001;28:620-4.
Yeon SB, Salton CJ, Gona P, Chuang ML, Blease SJ, Han Y, et al. Impact of age, sex, and indexation method on MR left ventricular reference values in the framingham heart study offspring cohort. J Magn Reson Imaging 2015;41:1038-45.
Fiechter M, Fuchs TA, Gebhard C, Stehli J, Klaeser B, Stähli BE, et al. Age-related normal structural and functional ventricular values in cardiac function assessed by magnetic resonance. BMC Med Imaging. 2013;13:6.
Chuang ML, Gona P, Hautvast GLTF, Salton CJ, Breeuwer M, O’Donnell CJ, et al. CMR reference values for left ventricular volumes, mass, and ejection fraction using computer-aided analysis: The Framingham Heart Study. J Magn Reson Imaging 2014;39:895-900.
Lakatta EG, Levy D. Arterial and cardiac aging: Major shareholders in cardiovascular disease enterprises: Part II: The aging heart in health: Links to heart disease. Circulation. 2003;107:346-354. http://circ.ahajournals.org/content/107/2/346.short.
Lakatta EG, Levy D. Arterial and cardiac aging: major shareholders in cardiovascular disease enterprises: Part I: Aging arteries: A “Set Up” for vascular disease. Circulation 2003;107:139-46.
Sharir T, Kang X, Germano G, Bax JJ, Shaw LJ, Gransar H, et al. Prognostic value of poststress left ventricular volume and ejection fraction by gated myocardial perfusion SPECT in women and men: gender-related differences in normal limits and outcomes. J Nucl Cardiol 2006;13:495-506.
Sandstede J, Lipke C, Beer M, Hofmann S. Age-and gender-specific differences in left and right ventricular cardiac function and mass determined by cine magnetic resonance imaging. Eur Radiol 2000;442:438-42.
Chung AK, Das SR, Leonard D, Peshock RM, Kazi F, Abdullah SM, et al. Women have higher left ventricular ejection fractions than men independent of differences in left ventricular volume: The Dallas Heart Study. Circulation 2006;113:1597-604.
Disclosure
The authors declare that they have no conflict of interest.
Author information
Authors and Affiliations
Corresponding author
Additional information
See related editorial, doi:10.1007/s12350-015-0274-5.
Rights and permissions
About this article
Cite this article
Haarmark, C., Haase, C., Jensen, M.M. et al. Pre-chemotherapy values for left and right ventricular volumes and ejection fraction by gated tomographic radionuclide angiography using a cadmium-zinc-telluride detector gamma camera. J. Nucl. Cardiol. 23, 87–97 (2016). https://doi.org/10.1007/s12350-015-0177-5
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s12350-015-0177-5