1 Introduction

Cerebral blood flow (CBF), cerebral blood volume (CBV), oxygen extraction fraction (OEF), and cerebral metabolic rate of oxygen (CMRO2) have been quantified by positron emission tomography (PET) using oxygen-15-labeled gases [1,2,3]. These quantitative parameters have been applied to pathophysiological evaluations and used to predict outcomes among patients with cerebral vascular disease [4,5,6,7].

Newer PET scanners comprised of lutetium-based scintillator acquire data only in three-dimensional (3D) mode, which is optimized for whole-body scanning. The high sensitivity of 3D PET has decreased the acquisition time and the radiation dose associated with whole-body imaging [8, 9]. However, the increase in scattered photons from radioactivity outside the field of view (FOV) not only degrades image quality but also decreases the quantitative accuracy of 3D PET [10,11,12,13,14,15,16]. Direct and scatter photons from large amounts of 15O in the lungs and in blood pools, such as the heart and major blood vessels, affect quantitative PET measurements. Neck shields have been developed to reduce the contribution of annihilation photons from outside the FOV [11, 15, 17,18,19]. Ibaraki et al. found that an additional neck shield made of lead plate reduced random events and improved the signal-to-noise ratio (SNR) derived from reconstructed image-based SNRs generated by 3D PET using 15O-labeled compounds [15]. Lutetium-based scintillators such as lutetium–yttrium–orthosilicate (LYSO) and Lu2SiO5 (LSO) have faster scintillation decay than Bi4Ge3O12 (BGO) and Gd2SiO5 (GSO) [20]. The high count-rate capability of LYSO scintillator reduces dead time and random coincidence events. However, Kloet et al. proposed that an additional neck shield would reduce the effects of radioactivity outside the FOV even when images are acquired using a PET scanner with LSO [16].

Previous neck shields have been made of lead plate [11, 15,16,17, 19]. We previously used the lead-plate neck shield used by Ibaraki et al. [15] during a 15O-labeled gas study using a conventional PET scanner that has a smaller trans-axial FOV than do most PET/CT scanners. The neck shield made of lead could be attached to PET scanners manufactured by SHIMADZU (SHIMADZU Corp., Kyoto, Japan) only [15]. In the present study, we used a novel neck shield comprised of tungsten-impregnated rubber and a stainless metal-alloy frame to reduce scatter radiation from radioactivity outside the FOV. Tungsten-impregnated rubber is softer and more flexible than lead; it can be shaped to fit curved surfaces. The neck shield was designed to avoid feelings of oppression and constriction during 15O-labeled gas studies regardless of the physique of the patient.

The present study aimed to validate the effects of the tungsten-impregnated rubber neck shield on the quality of images acquired from a phantom (phantom study) and from 15O-labeled gas inhalation (clinical study) using 3D PET and a high-end LYSO scanner.

2 Materials and methods

2.1 PET/CT equipment

All data were acquired using a Discovery PET/CT 710 (GE Healthcare, Milwaukee, WI, USA) [21]. The PET scanner comprised 13,824 LYSO crystals in a 4.2 × 6.3 × 25-mm3 block. The PET detector comprised a 150.42-mm axial FOV and a 700-mm trans-axial FOV with a 64-slice CT scanner. The spatial resolution according to the National Electrical Manufacturers Association (NEMA) NU 2-2007 was 4.52 mm (full width at half maximum, FWHM, at 1.0 cm off center) [22].

2.2 Novel neck shield

The novel neck shield (Universal Giken, Odawara, Japan) comprised a stainless alloy frame with an axial length of 163 mm and tungsten-impregnated rubber (6 mm thick) that was equivalent to >6 mm lead (Fig. 1). The 564 × 318 mm (width and height) rubber component consists of a base sheet, right and left covers, and top sheet that are easily connected and can cover the shoulders of patients. The weights of the neck shield components were 6.6, 5.6, and 3.6 kg for the base sheet, right and left covers, and top sheet, respectively.

Fig. 1
figure 1

Novel tungsten-impregnated rubber neck shield with stainless alloy frame. Whole (a), top (b), and front (c) views. Arrows tungsten-impregnated rubber

2.3 Phantom study

An acrylic cylindrical phantom (diameter, 20 cm; length, 16 cm) simulated a human head and another (diameter, 20 cm; length, 25 cm) simulated the body. The center of the PET axial FOV head-simulated phantom containing [15O]H2O was placed 16 cm from the body phantom containing an 18F solution. The reported brain to body radioactivity ratio of inhaled [15O]O2 is 1:8 [23]. The radioactivity ratio of brain to body was set to 1:7.6 and 1:7.3 with and without neck shield, respectively, at the start of each image acquisition. The amount of [15O]H2O radioactivity in the head-simulated phantom was determined by the mean radioactive concentration of the whole brain in the [15O]O2 images of three participants. The amount of radioactivity was 102.9 and 93.7 MBq with and without the neck shield, respectively, at the start of each image acquisition. Images were acquired from the head-simulated phantom for 3.5 min with and without the neck shield.

2.4 Clinical study

All retrospective experiments were approved by the institutional review board (IRB) and were performed in accordance with the IRB rules and policies (282559, and 28-371). We retrospectively analyzed 15O-labeled gas images acquired from three participants (average age, 22.5 years; average height, 172.0 cm; average weight, 64.0 kg). After a low-dose CT scan for attenuation correction, three sequential emission scans of inhaled [15O]CO2 (2000 MBq/min), [15O]O2 (3000 MBq/min), and [15O]CO (2000 MBq/min) were performed with and without the neck shield (Fig. 2). The acquisition of images for 3.5 min in [15O]CO2 and [15O]O2 scans were initiated simultaneously with the 1.5-min inhalation of these gases [24]. Then, participants inhaled [15O]CO gas for 4 min and [15O]CO images were acquired for 4 min after 3 min of [15O]CO gas inhalation. The interval between each gas inhalation was 10 min.

Fig. 2
figure 2

Schema of 15O-labeled gas study using a positron emission tomography (PET) auto-radiographic method

The amount of radioactivity in arterial whole blood was continuously measured via arterial puncture by a physician using a BeCON Model 01 Gd2SiO5:Ce (GSO) coincidence detector (Molecular Imaging Labo, Suita, Japan) while the images of inhaled [15O]CO2 and [15O]O2 were acquired [25]. The amount of radioactivity was measured in samples of arterial whole blood (1.0 mL) that were manually collected at 0, 2, and 4 min during the acquisition of PET images of [15O]CO inhalation using a BeWell Model-QS03 F/B well counter (Molecular Imaging Labo, Suita, Japan).

2.5 Data processing

PET images were reconstructed under the following conditions: three-dimensional-ordered-subsets expectation–maximization (3D-OSEM) algorithm, 128 × 128 matrix, 47 slices, 2.0 mm/pixel, 3.27 mm/slice, 4 iterations, 16 subsets, and Gaussian filter of 3.0 mm (FWHM). The clinical [15O]CO2 and [15O]O2 images consisted of 12 frames of 5 s and 10 frames of 15 s. Attenuation was corrected using the corresponding CT image. We applied scatter limitation correction (SLC) to correct the SC error derived from high levels of radioactive 15O gas inside a facemask [23, 26]. The SLC limits the amount of scatter outside the object when the scatter fraction of the emission profile exceeds a specific threshold and allows SC scaling to levels typical of clinical PET images. That is, significantly scaled scatter profiles are scaled down by adjusting the amount of scatter outside the object in the process of single scatter simulation [26]. We calculated the CBF, CBV, OEF, and CMRO2 from functional images and amounts of radioactivity in arterial blood using the PET autoradiographic method with Xeleris software (GE Healthcare, Milwaukee, WI, USA) [3].

2.6 Data analysis

2.6.1 PET count rate characteristics

Noise equivalent count rates (NECRs) [27] were calculated as follows:

$${\text{NECR}} = \left( {1 - {\text{SF}}} \right)^{2} \frac{{\left( {T + S} \right)^{2} }}{{\left( {T + S} \right) + R}} \left[ {\text{cps}} \right],$$

where SF is the scatter fraction, T is the true coincidence count rate, S is the scatter coincidence count rate, and R is the random coincidence count rate. We calculated (T + S) by subtracting the random from the prompt coincidence count rate (P). The NECR was calculated for the image sequences of each gas in the clinical study. We calculated NECR gain as NECR with/without the neck shield. Values for P, R, (T + S), and NECR in each frame of [15O]CO2 and [15O]O2 images with and without the neck shield in the clinical study were compared using the Wilcoxon matched-pairs signed rank test. Values of p < 0.05 were considered significant.

2.6.2 Image uniformity

Circular regions of interest (ROIs) of 18 cm in diameter were placed on all slices of the center of the head-simulated phantom images to determine mean radioactivity (kBq/mL) and standard deviation (SD). The coefficient of variance (CV) to evaluate image uniformity for each slice was calculated as

$${\text{CV}}_{i} = {\raise0.7ex\hbox{${{\text{SD}}_{i} }$} \!\mathord{\left/ {\vphantom {{{\text{SD}}_{i} } {{\text{mean radioactivity}}_{i} }}}\right.\kern-0pt} \!\lower0.7ex\hbox{${{\text{mean radioactivity}}_{i} }$}} \times 100\left( \% \right),$$

where i is the slice number. The radioactivity concentration and CVs between with and without the neck shield were compared using the Wilcoxon matched-pairs signed rank test. Values of p < 0.05 were considered significant.

2.6.3 Variation of voxel value

Circular ROIs with a 10-mm diameter were placed manually on three adjacent slices of CBF, CBV, OEF, and CMRO2 images (24 ROIs per region) of the cerebellar cortex and frontal, temporal, and occipital lobes in the clinical study (Fig. 3). The SDs inside ROIs from images of the three participants were measured. The average SD of functional images was calculated from all SDs from the three participants. We calculated SD change from the ratio of average SD with and without the neck shield. We compared SDs with and without the neck shield using paired t tests. Values of p < 0.05 were considered significant. All ROIs were analyzed using Dr. View/Linux (AJS, Tokyo, Japan).

Fig. 3
figure 3

Location of regions of interest (ROI) on cerebral blood flow images. Circular 10-mm ROI are set on three adjacent slices

2.6.4 Inter-acquisition variability

We also compared the mean value, SD, and CV (SD/mean value) of quantitative metrics of CBF (mL/100 g/min), CBV (mL/100 g), OEF (%), and CMRO2 (mL/100 g/min) in each region with and without the neck shield.

3 Results

3.1 PET count rate characteristics

Table 1 shows the PET count characteristics in the phantom and clinical studies. The neck shield decreased the prompt and random count rates by 37.1 and 58.8%, respectively, and increased the true + scatter count rate and NECR by 15.0 and 61.3%, respectively, in the phantom study. Prompt and random count rates in the clinical study were decreased by 9.5 and 24.5% in [15O]CO2 images, by 22.1 and 36.3% in [15O]O2 images, and by 20.8 and 44.9% in [15O]CO images, respectively. The neck shield significantly decreased the prompt and random count rates of [15O]CO2 and [15O]O2. The true + scatter count rates and NECRs were increased by 14.2 and 43.9%, 15.5 and 65.9%, and 9.3 and 55.0% in the [15O]CO2, [15O]O2, and [15O]CO images, respectively. True + scatter count rates and NECRs were significantly increased with the neck shield except for the true + scatter count rate in the [15O]CO2 images. The count characteristics of the clinical [15O]CO images tended to be the same as those of the other images, i.e. the difference was not statistically significant. The values of the scatter fraction were not different between with the neck shield and without the neck shield in the phantom and clinical studies.

Table 1 PET count characteristics in phantom and clinical studies

3.2 Image uniformity

Figure 4 shows the mean radioactivity of all ROIs in the phantom study with and without the neck shield. The mean radioactivity was lower in caudal slices than in cranial slices. The average of mean radioactivity and the CV of all ROIs with and without the neck shield were 12.2 ± 0.1 kBq/mL and 15.4 ± 8.2%, and 11.1 ± 0.1 kBq/mL and 16.3 ± 8.8%, respectively. The neck shield significantly decreased radioactivity concentration and CV (both p < 0.05). The ratio of injected radioactivity in the phantom with and without the neck shield was equivalent to the ratio of mean radioactivity concentration in the phantom image with to that without the neck shield.

Fig. 4
figure 4

Mean radioactivity of slices in a head phantom with and without the neck shield. Black and white circles with and without the neck shield, respectively

3.3 Variation of voxel value

Table 2 shows the average SDs with and without the neck shield and SD change using the neck shield in functional images. The neck shield significantly decreased SD in the CMRO2 images (p < 0.05). Further, the SD of the cerebellum region in the CBF images, occipital region in the CBV images, and all regions in the CMRO2 images decreased significantly with the neck shield (p < 0.05). There was no change in SD after using the neck shield in the entire CBF, CBV, and OEF images, nor were there region-specific changes. Figure 5 shows the CBF, CBV, OEF, and CMRO2 images from a participant with and without the neck shield. The image noise on functional images with the neck shield was the same as that without it.

Table 2 SDs (mean ± SD) and SD changes of CBF, CBV, OEF, and CMRO2 with and without neck shield
Fig. 5
figure 5

Parametric images of cerebral blood flow (CBF), cerebral blood volume (CBV), oxygen extraction fraction (OEF), and cerebral metabolic rate of oxygen (CMRO 2 ) with and without the neck shield. Color scales in the CBF and CMRO2 images represent 0–80% of the maximum value. The ranges of the color scales in the CBV and OEF images are 0–20 (mL/100 g) and 0–80 (%), respectively

3.4 Inter-acquisition variability

The mean ± SD of the quantitative parameters of CBF, CBV, OEF, and CMRO2 in total region with and without the neck shield were 54.5 ± 11.8 and 55.7 ± 14.8 mL/100 g/min; 3.3 ± 1.1 and 3.4 ± 1.1 mL/100 g; 36.2 ± 3.4 and 38.3 ± 5.0%; and 6.3 ± 2.3 and 8.1 ± 3.5 mL/100 g/min, respectively. Table 3 shows the SD and CV of quantitative parameters of CBF, CBV, OEF, and CMRO2 in each region with and without the neck shield.

Table 3 Inter-acquisition variability with and without the neck shield in each region

4 Discussion

We investigated the ability of a novel neck shield made of tungsten-impregnated rubber to improve the quality of 15O-labeled gas images acquired using 3D PET/CT. The neck shield improved NECRs from 43.9 to 65.9% by reducing random coincidence in the phantom and clinical studies. However, an improvement of physical and visual image quality with the neck shield was not observed in the phantom and clinical studies.

Radioactivity outside the FOV degrades 3D PET whole-body and brain image quality and quantitation [10,11,12,13,14,15,16]. A neck shield reduces direct and scattered photons from outside of the FOV and improves count rate performance [11, 15, 17,18,19]. Thompson and Moreno-Cantu found that an additional shield during brain PET imaging with [11C]raclopride, [15O]H2O, and [18F]FDG improved NEC in three studies. They concluded that the additional shield conferred the greatest benefit upon [15O]H2O or 15O-oxygen images [19]. Spinks et al. also found that an additional shield improved PET count rates, more so in head than in body images [11]. Our phantom and clinical results showed that the random count rate of [15O]O2 decreased to 36.3% and that the NECR increased to 65.9% using the neck shield. Ibaraki et al. [15] showed that random counts were reduced by about 40% and NECR increased by about 40% when a 7-mm lead-plate neck shield was applied during [15O]O2 image acquisition [19]. Our PET count characteristics using a tungsten-impregnated rubber neck shield with a lead equivalent of 6 mm were similar to those reported by Ibaraki et al. [15]. The values of scatter fraction were not much different with and without the neck shield. The radioactivity outside the FOV was found to have no effect on the scatter fraction in a previous study [14]. Our findings suggest that the benefit of the neck shield is in the reduction of random counts. Our PET/CT scanner exhibited higher count rates and decreased dead time because the scanner uses LYSO, in which fluorescence decays faster than the GSO used by Ibaraki et al. [15]. The NECR results suggested that the quality of PET images acquired using an LYSO scanner and 15O-labeled gas is not degraded by scatter photons from outside the FOV.

The variation of voxel value of CMRO2 decreased significantly upon using the neck shield, although the CMRO2 image did not improve visually. The neck shield not only improved the statistical noise in functional images in the clinical study but also the parameters of image quality such as CV in the phantom study. Ibaraki et al. showed that a neck shield improved SNR calculated using the bootstrap method by about 20% [15]. The neck shield did not improve measures of image quality such as CV and SD in the present study because our iterative 3D-OSEM method of reconstruction might produce less image noise without the neck shield, compared to other methods.

Thus, the quality of functional images was equivalent between the procedures done with and without the neck shield when the PET system included an LYSO scanner and the reconstruction method was 3D-OSEM. The relationship between the count characteristics and image quality is nonlinear when using 3D-OSEM in a PET/CT scanner that does not correct for scatter and random events by subtraction of the estimated scatter and random events, respectively [28]. Furthermore, the improvements of image quality based on SD change were slight in comparison with NECR gain in our study.

In the phantom study, the difference of mean radioactive concentration with and without the neck shield was equivalent to the difference in injected dose in the phantom with and without the neck shield. Thus, we demonstrated that the neck shield did not affect the quality and quantity of 15O-labeled gas images from the phantom study. However, the range of CVs, which represent the inter-acquisition variability between acquisitions with and without the neck shield, was 9.5–12.3%, 12.0–26.2%, 7.3–9.4%, and 15.5–19.4% in the CBF, CBV, OEF, and CMRO2, respectively. The inter-acquisition variability of CBV and CMRO2 was worse than those found in a previous study [29]. Although we defined the ROIs to exclude large blood vessels in the CBV image, spillover from small blood vessels contained in ROIs might have degraded the inter-acquisition variability. Further, the quantitative parameter of CMRO2 was calculated from the CBV images, meaning that the inter-acquisition variability of CMRO2 might also be degraded.

The present study has two limitations. One is that the sample size of three participants was insufficient to establish definite evidence. The other is that some technologies such as point-spread-function [30], time-of-flight [31], and a novel reconstruction algorithm [32] might also improve the image quality of 15O-labeled gas imaging. We did not make comparisons with these various other correction methods.

5 Conclusion

A tungsten-impregnated rubber neck shield reduced the random count rate and improved count-based SNR such as NECR in a 3D PET study using 15O-labeled gas. The image quality with the neck shield was as same as that without the neck shield. The neck shield is useful in terms of improvement of count rate performance, and our PET system and acquisition condition without the neck shield is applicable to qualitative and quantitative 15O-labeled gas PET imaging.