Introduction

18F-fluorodeoxyglucose (FDG) positron emission tomography/computed tomography (PET/CT) has become one of the most useful tools in diagnostic imaging for cancer. Many studies have demonstrated the efficacy of whole-body FDG-PET/CT in staging or re-staging, in monitoring the response to therapy, and for prediction of the prognosis in patients with breast cancer [1,2,3]. It is important to detect breast cancer at an early stage when it is small, given that mortality increases with tumors exceeding 1 cm in size [4, 5]. However, detection of small breast cancers by whole-body PET/CT is challenging, because of its limited spatial resolution [6].

High-resolution dedicated breast PET (dbPET) scanners have been developed to detect small breast lesions. There are two types of high-resolution dbPET, i.e., positron emission mammography (PEM) and a tomographic technique using a ring-shaped scanner (ring-shaped dbPET) [7]. PEM systems depict breast tissue via soft compression of the breast with two opposing plate-like detectors and have higher sensitivity than whole-body PET/CT [8,9,10], while ring-shaped dbPET scanners can visualize breast cancer more clearly than whole-body PET/CT [11, 12]. These high-resolution breast PET systems have greater photon sensitivity and can improve spatial resolution by setting the detector close to the breast, reducing respiratory movement, and using smaller detection units with reconstruction methods that are different to those used for whole-body PET/CT. In Japan, combined use of dbPET and whole-body PET/CT is necessary in some indications for health insurance and the Japanese Society of Nuclear Medicine also recommends their combined use [13]. dbPET can evaluate local lesions of breast cancer, which is difficult to do by whole-body PET due to its spatial resolution limitations. Conversely, whole-body PET can scan for metastasis, including the lymph nodes, which cannot be evaluated by dbPET. Therefore, a more accurate diagnosis can be achieved using both modalities, rather than one examination, which would lead to the selection of an appropriate treatment method and improvement of the prognosis of the breast cancer patient.

The standard performance of dbPET has been evaluated previously using a rat-sized phantom [NU4-2008, National Electrical Manufacturers Association (NEMA), Rosslyn, VA, USA] [14, 15]. Furthermore, a phantom experiment procedure manual for the NEMA NU-2 2007, a body phantom, has been created for standardization of whole-body PET images in Japan [16]. The spatial resolution of most whole-body PET scanners is approximately 5 mm [17, 18], whereas that of a dbPET system is approximately 2 mm [14]. Although combined use of dbPET and whole-body PET/CT is recommended; the performance of these devices when using a common phantom has not been reported. Therefore, the purpose of this phantom study based on the Japanese FDG-PET/CT data acquisition protocol was to demonstrate and compare the performance of high-resolution dedicated breast positron emission tomography (dbPET) with whole-body PET/CT.

Materials and methods

Ring-shaped dbPET scanner

The ring-shaped dbPET scanner (Elmammo, Shimadzu Corp., Kyoto, Japan) consists of 36 detector modules arranged in three contiguous rings, has a diameter of 195 mm and an axial length of 156.5 mm, and has depth-of-interaction measurement capability [19]. The transaxial effective field-of-view (FOV) is 185 × 156.5 mm2. Each detector block consists of a four-layered 32 × 32 array of lutetium oxyorthosilicate crystals coupled to a 64-channel positron-sensitive photomultiplier tube via a light guide. Attenuation correction was calculated using a uniform attenuation map with object boundaries obtained from emission data [20] and scatter corrections were applied for all images. Scatter correction was performed using the convolution-subtraction method [21] with kernels obtained by background tail fitting. Performance metrics included 1.5-mm FWHM resolution in standard mode in the axial, sagittal, and coronal views, detector sensitivity of 0.09–0.13 cps/Bq at the center of the detector, and detector sensitivity of 0.05–0.08 cps/Bq at a quarter depth of the detector. The peak noise equivalent count was 600–800 kcps. The characteristics of this scanner have been reported in detail previously [14]. The standard performances are shown in Table 1.

Table 1 Standard performances of dbPET and PET/CT

Whole-body PET/CT scanner

PET/CT scans were obtained using a Biograph Horizon TrueV FDG-PET/CT system (Siemens Medical Solutions, Knoxville, TN, USA). This system has 52 detector rings consisting of 160 blocks, with each block containing an array of 13 × 13 lutetium oxyorthosilicate crystals (4 mm × 4 mm × 20 mm) covering an axial FOV of 221 mm and having a transaxial FOV of 690 mm. The coincidence timing window and time-of-flight system timing resolution were 4.1 ns and 540 ps, respectively. A CT scan was performed for attenuation correction (130 kV; 15–70 mA; tube rotation time, 0.6 s per rotation; pitch, 1; a transaxial FOV, 700 mm; and section thickness, 5 mm). The standard performances are shown in Table 1.

Development and preparation of the breast phantom

A cylindrical breast phantom made of acrylic resin was used to compare the detectability of lesions on dbPET with that on PET/CT (Fig. 1). The phantom was constructed to contain four plastic spheres of different diameters. The inner and outer diameters of the cylinder were 100 mm and 140 mm, respectively, and the height was 170 mm. The diameters of the spheres arranged inside were 5, 7.5, 10, and 16 mm. The cylinder and four spheres were filled with 18F-FDG solution at sphere-to-background radioactivity ratios (SBRs) of 8:1, 4:1, and 2:1 to account for the fact that FDG uptake in the mammary gland is physiologically variable according to age, the amount of fibroglandular tissue, and the menstrual cycle, ranging from SUVmax 0.65 to 2.05 [22, 23]. The background radioactivity at the start of data acquisition by PET/CT and dbPET was set to 3.01 and 2.46 kBq/mL, respectively. The phantom was prepared in these three conditions separately. One scan was performed under each condition. The phantom was placed in the detector, so that the spheres were located at the center plane of the detector. Sphere placement at the center of the detector was confirmed visually on the image.

Fig. 1
figure 1

Cylindrical breast phantom. A cylindrical breast phantom consisting of an outer cylinder (a) and four spheres of different diameter (5, 7.5, 10, and 16 mm) (b)

Data acquisition and image reconstruction

The phantom was placed at the center of the column in the dbPET or PET/CT detector, so that the four spheres were aligned horizontally and then scanned by both devices for 20 min in list mode. The dbPET images were reconstructed using a three-dimensional list mode dynamic row-action maximum-likelihood algorithm with one iteration and 128 subsets, a relaxation control parameter of β = 20, a matrix size in the axial view of 236 × 200 × 236 with a post-reconstruction smoothing Gaussian filter (1.17 mm FWHM), and scatter correction. The PET/CT images were reconstructed using the ordered subset expectation maximization method and the time-of-flight algorithm with four iterations and ten subsets. The CT data were resized from a 512 × 512 matrix to a 180 × 180 matrix to match the PET data and construct CT-based transmission maps for attenuation correction of the PET data with a post-reconstruction smoothing Gaussian filter (5 mm FWHM). These conditions were determined based on those used in a previous study using dbPET [24].

Analyses of phantom image quality

Visual analyses of the phantom images were performed using syngo. via VB10 (Siemens Healthcare GmbH, Erlangen, Germany). An experienced nuclear medicine physician and two experienced PET technologists evaluated the hot spheres. Evaluations were performed using the slices displayed in the transverse image slice containing the centers of the spheres. The images were displayed in an inverse gray scale with a standardized uptake range of 0–6. The hot sphere of 5 mm in diameter was visually graded as follows: 2, identifiable; 1, visualized, but similar hot spots observed elsewhere; and 0, not visualized. Spheres with visual scores ≥ 1.5 were deemed to be detectable. The final score was the mean of the scores from three readers. The visual assessment was performed based on the Japanese guideline [13]. It was decided by consensus that the SBR was suitable for physical analysis.

The physical analysis was also performed using syngo. via VB10. According to the Japanese guideline for the FDG-PET/CT oncology data acquisition protocol, dbPET image quality was also analyzed using the parameters described below. The three formulas were modified from those described in the standard PET imaging protocols and phantom test procedures and criteria [25].

The coefficient of variation and uniformity of the background (CVbackground and SDΔSUVmean, respectively) were calculated by evaluation of various regions of interest (ROIs) in the transverse image slice that contained the centers of the spheres. Ten ROIs with a diameter of 16 mm were placed in the background region in that slice and adjacent ± 5 mm and 10 mm slices (50 ROIs in total), as illustrated in Fig. 2a.

Fig. 2
figure 2

Positioning of the regions of interest for the analyses of image quality. Regions of interest measuring 16 mm diameter were placed over the phantom background (yellow) of the reconstructed PET images at a sphere-to-background radioactivity ratio of 8:1 and were used for analysis of CVbackground and SDΔSUVmean (a). Regions of interest measuring 5 mm diameter were placed over the smallest active sphere (red) and the phantom background (yellow) of the reconstructed PET images at a sphere-to-background radioactivity ratio of 8:1 and were used for analysis of %QH,5mm and %N5mm (b). PET positron emission tomography

CVbackground and SDΔSUVmean were calculated using the following:

$${\text{CV}}_{{{\text{background}}}} = {\text{ mean of}}\left[ {\frac{{{\text{SD}}_{{16\,{\text{mm}}}} }}{{C_{{{\text{B}},16\,{\text{mm}}}} }}} \times 100\right]{ } { }(\%)$$
$${\text{ SD}}_{{\Delta {\text{SUV}}_{{{\text{mean}}}} }} { } = \sqrt {\frac{1}{n}\mathop \sum \limits_{i = 1}^{n} \left( {\Delta {\text{SUV}}_{{{\text{mean}},i}} } \right)^{2} }$$

SD16mm is the standard deviation of the background ROI values that was 16 mm in diameter, CB,16 mm is the average of the background ROI values that was 16 mm in diameter, ΔSUVmean,i is the relative error of each background ROI value that was 16 mm in diameter. n is 50 which is the sum of 10 ROIs set for each slice.

The percentages of contrast and background variability (%QH,5mm and %N5mm) and their ratio (%QH,5mm/N5mm) were also calculated by evaluation of various ROIs. The 12 ROIs that were 5 mm in diameter were placed on the background region in that slice and adjacent ± 5 mm and 10 mm slices (60 ROIs in total) as illustrated in Fig. 2b. %QH,5mm and %N5mm were used as a measure for the image contrast and noise for sphere and their ideal values were 100 and 0%, respectively.

%QH,5mm was calculated using the following:

$$\% Q_{{{\text{H}},{5}\,{\text{mm}}}} = \frac{{C_{{{\text{H}},5\,{\text{mm}}}} /C_{{{\text{B}},5\,{\text{mm}}}} - 1}}{{a_{{\text{H}}} /a_{{\text{B}}} - 1}}{ } \times 100(\%){}$$

CH,5mm is the average pixel value in the ROI for the 5-mm sphere, CB,5mm is the average pixel value in the background regions of all ROIs, aH is the measured activity concentration in the hot spheres, aB is the measured activity concentration in the background.

In addition, %N5mm was calculated using the following:

$$\% N_{{{5}\,{\text{mm}}}} = \frac{{{\text{SD}}_{{5\,{\text{mm}}}} }}{{C_{{{\text{B}},5\,{\text{mm}}}} }}{ } \times 100\,(\%)$$
$${\text{SD}}_{{{5}\,{\text{mm}}}} = \sqrt {\frac{{\mathop \sum \nolimits_{k = 1}^{60} \left( {C_{{{\text{B}},5\,{\text{mm}},{ }k}} - C_{{{\text{B}},5\,{\text{mm}}}} } \right)^{2} }}{{\left( {60 - 1} \right)}}}$$

SD5mm is the standard deviation of the background ROI values for a 5-mm-diameter sphere, CB,5mm is the average pixel value in the background regions of all ROIs, The ratio of %QH,5mm and %N5mm (QNR) was also calculated:

$${\text{QNR }} = \, \% Q_{{{\text{H}},{5}\,{\text{mm}}}} /\% N_{{{5}\,{\text{mm}}}}$$

The recovery coefficient (RC) was defined as the ratio of the measured sphere activity to the true sphere activity. The SUVmax of the four hot spheres was determined from a circular ROI placed on the center slice of the images of each sphere. The diameter of each ROI was equal to that of each sphere (5, 7.5, 10, and 16 mm).

The relative RC was calculated using the following:

$${\text{RC}}_{{\text{j}}} = C_{{\text{j}}} /C_{{16\,{\text{mm}}}}$$

Cj is the SUVmax of each ROI in each sphere, C16mm is the SUVmax of the ROI with a 16-mm diameter for a sphere with a 16-mm diameter.

Human imaging

Patients fasted at least 6 h prior to administration of 18F-FDG (3 MBq/kg) and were scanned by whole-body PET/CT for 90 s per bed and dbPET for 7 min per breast. Scans were performed at 60-min and 90-min post injection, both in the prone position.

The PET/CT and dbPET images were reconstructed using the same conditions as for the phantom images. The SUVmax of the abnormal FDG uptake and SUVmean of the normal background were measured. To measure the SUVmean of the background, a VOI with a diameter of approximately 2–2.5 cm was placed on the contralateral normal mammary gland, avoiding structures such as the nipple and the chest wall. The size of the VOI on the background varied slightly depending on the volume of the mammary gland in each patient. Where patients had previously undergone unilateral mastectomy, a VOI was placed on the ipsilateral mammary gland as far as possible from the abnormal uptake. Two representative cases were shown in this study.

This component of the study was approved by our institutional review board and written informed consent was obtained from the study participants.

Results

The visual scores on the dbPET and PET/CT images were, respectively, 0 and 0 for an SBR of 2:1, 1 and 0 for an SBR of 4:1, and 2 and 0.7 for an SBR of 8:1 (Table 2). Representative phantom images acquired by the two scanners for various SBRs and their three-dimensional pixel maps at an SBR of 8:1 are shown in Fig. 3.

Table 2 Results of the phantom image quality analyses
Fig. 3
figure 3

Images of the breast phantom scanned by dbPET and PET/CT. The background on dbPET was noisier than that on PET/CT. However, the hot spots were more clearly visualized. The smallest spheres were visible, so the physical assessment was made with an SBR of 8:1 (a). Two- and three-dimensional pixel maps for four spheres (5, 7.5, 10, and 16 mm in diameters) of the breast phantom images scanned at a SBR of 8:1 also showed that the background on dbPET is rougher, but the hotspots are more sharply delineated on dbPET than on PET/CT (b). dbPET dedicated breast positron emission tomography, PET/CT positron emission tomography, SBR sphere-to-background ratio

The CVbackground, SDΔSUVmean, and %N5mm values on the images scanned by dbPET and PET/CT were 6.14 and 2.21, 0.024 and 0.021, and 4.55 and 3.11, respectively. The CVbackground and %N5mm values on dbPET were higher than those on PET/CT; in contrast, the %QH,5mm values were 16.67 and 5.34, respectively, and the %QH,5mm/%N5mm values were 3.66 and 1.72 (Table 2). The RCs for the 5-mm sphere on dbPET and PET/CT were 0.33 and 0.18, respectively (Fig. 4).

Fig. 4
figure 4

Recovery coefficients for four spheres on dbPET and PET/CT. Recovery coefficients measured for different sphere sizes. Images with 20 min of acquisition were used. dbPET dedicated breast positron emission tomography, PET/CT positron emission tomography, RC recovery coefficient, SUV standardized uptake value

Figure 5 shows representative clinical images acquired by dbPET and PET/CT in the prone position. The focal FDG uptakes were visualized on both dbPET and PET/CT images. However, they had clearer and more obvious on dbPET, despite the degree of background FDG uptake by the mammary gland.

Fig. 5
figure 5

Clinical images of dbPET and PET/CT. Representative clinical images of dbPET (upper) and PET/CT (lower) with higher (case A) and lower (case B) focal FDG uptake in the background mammary gland tissue. The focal uptakes were visualized on both the dbPET and PET/CT images. However, they were clearer and more obvious on dbPET and showed more contrast despite the degree of FDG uptake in the background mammary gland tissue (asterisk in a, double asterisk in b). a Image from a 52-year-old woman. The focal FDG uptake corresponded to a 4-mm mass (BI-RADS category 3) on ultrasonography. b Image from a 41-year-old woman. The focal FDG uptake corresponded to a 4-mm enhanced mass (BI-RADS category 3) on magnetic resonance imaging. BI-RADS Breast Imaging Reporting and Data System, FDG18F-fluorodeoxyglucose

Discussion

In this study, we investigated the difference in quality of a common breast phantom and clinical images between dbPET and whole-body PET/CT to define the parameters needed to standardize dbPET images according to the Japanese protocol for whole-body PET/CT.

In the phantom study, although there was no significant difference in uniformity (SDΔSUVmean) between dbPET and whole-body PET/CT, the CVbackground and %N5mm values were higher on dbPET than on whole-body PET/CT (Table 2), indicating that the background on dbPET was noisier than that on whole-body PET, as shown in the phantom images (Fig. 3) and clinical images (Fig. 5). Despite the noisier background, the %QH,5mm and %QH,5mm/%N5mm values on dbPET were much higher than those on whole-body PET/CT (Table 2). This finding means that the lesion-to-background contrast of dbPET was better than that of whole-body PET/CT, which is consistent with previous reports [10, 26]. Furthermore, the RCs for the smaller spheres were higher on dbPET than on PET/CT, indicating that the partial volume effect of dbPET was less than that of PET/CT. These results demonstrate that dbPET delineates lesions about 5 mm in size more clearly than whole-body PET/CT.

In this study, we referenced the whole-body FDG-PET/CT data acquisition protocol for oncology that is designed to standardize PET/CT images in Japan. However, there is no dbPET imaging protocol at present. The protocol for PET/CT recommends the use of the NEMA NU-2 2007, a widely used body phantom, with an SBR of 4:1 for acquisition. The NU-2 2007 contains six spheres with diameters of 32, 28, 22, 17, 13, and 10 mm. There were some differences in the phantom used in the whole-body PET/CT protocol and that used in our study in terms of the size of the sphere, a suitable SBR, and acquisition time. Moreover, given that the acquisition time that is considered clinically acceptable differs between dbPET and whole-body PET/CT, a different PET/CT protocol for the phantom test will be needed for standardization of the dbPET images.

In the PET/CT protocol devised for oncology, the standard values recommended for the physical indicators are as follows: SDΔSUVmean, < 10%; SDΔSUVmean, < 0.025; %N10mm, < 5.6%; QNR, > 2.8; and RC10mm, > 0.38. It is noteworthy that a standard value for %QH is not defined in the guideline for whole-body PET/CT. It is important for detection of a lesion by whole-body PET/CT that %N is sufficiently small [13]. In contrast, despite the %N for dbPET being higher than that for PET/CT, the %QH was much higher for dbPET, resulting in a higher QNR and a high-contrast image. It seems that a high %QH is more important for detection of lesions in dbPET images. Given these results, standard index values that are different from those for PET/CT will be needed for dbPET.

We evaluated both dbPET and whole-body PET/CT using a common breast phantom. Currently, not only ring-shaped dbPET, but also different types of PEM devices are commercially available and used. However, there is no acquisition protocol for a test using a common phantom. In the case of PET/CT, although the specifications of the detector and the image reconstruction method are different, the shape of the detectors and the images obtained are almost the same for each device. Therefore, standardization of the imaging methods for whole-body PET/CT can be performed by a test using the same phantom with physical and visual assessments. However, there are two types of high-resolution breast PET, i.e., dbPET, which has a ring-shaped detector, and PEM, which has two plate-like detectors with markedly different shapes, so it is difficult to standardize images obtained from these devices by a test using the same phantom. A previous study evaluated dbPET (MAMMI breast PET, Oncovision, Valencia, Spain) and whole-body PET/CT using a specially made phantom containing gelatin and hydrogel [27] that would be difficult to use repeatedly in a multicenter setting. A common phantom with the durability and reproducibility like the NEMA body phantom would be preferable. Furthermore, standardization of quantitativeness is also a problem, because the quantification method varies depending on the device.

The harmonization of quantitative PET parameters is necessary to use them for diagnostic, prognostic, or therapeutic monitoring, regardless of the scanners [28]. In Japan, some standard physical indicators of whole-body PET/CT images have been defined and phantom test procedures also have been established [16, 25]. Currently, there are two types of commercially available high-resolution breast PET: dbPET and PEM, which have detectors with significantly different shapes. The next challenge is how to harmonize these scanners; however, harmonization will be very difficult to achieve for breast PET due to the large number of differences in the quantification methods of these two types of breast PET scanners. It would be appropriate to define some standards such as a recovery coefficient, to provide relative values, and to standardize the image using a common phantom, which can be scanned with both scanners. As with whole-body PET/CT, the breast PET imaging guidelines and the phantom test procedure should also be published.

This study has some limitations. First, to evaluate the standard performance of dbPET and to compare it with PET/CT, the phantoms were scanned at the center of the ring-shaped detector, as in a previous report [14]. However, many Japanese women have small breasts and their mammary glands are often located near the chest wall above the center of the detector even if they are in the prone position. This tendency is particularly common in young women, who are less likely to have breast ptosis than older women. Therefore, the position of the phantom in the detector needs to be changed more often. Second, we did not consider the optimal reconstruction conditions for dbPET; unlike with phantom images, it may be necessary to consider other factors when examining clinical images such as physical uptake in the mammary gland, heart, and pectoral muscles. Third, the phantom was scanned only once for each condition. It would have been better in terms of reproducibility to calculate the average results of several scans under each condition.

Conclusion

This phantom study showed that the background was noisier and that the contrast was much higher in dbPET images than in PET/CT images. The acquisition protocol and standards for dbPET will need to be different from those used for whole-body PET/CT. Standardization of the quantitative evaluation of dbPET will enable more accurate assessment of treatment effects and prognosis prediction.