Abstract
Measuring and monitoring of intracranial pressure is considered standard of care in patients with suspected intracranial hypertension. Sonographic assessment of the optic nerve sheath diameter (ONSD) has been promising and potentially useful for noninvasive intracranial hypertension screening. ONSD measurements are easy to perform, repeatable at bedside, fast, low cost, and radiation-free. However, they are still burdened by inter-rater variability, lack of ultrasound (US) setting standardization (e.g., US frequency, focus depth, etc.), and possible artifacts. To overcome this problem, we propose the CLOSED protocol associated with equipment specifications, as a guide to minimize the occurrence of such artifacts enabling a more reliable and accurate measurement. We suggest that color Doppler could be used as a new standard evaluation for the ONSD.
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Avoid common mistakes on your manuscript.
Introduction
Intracranial pressure (ICP) measuring and monitoring is considered care standard for suspected intracranial hypertension patients. Invasive measurement techniques still have significant side effects and complications such as bleeding and infections [1,2,3]. ICP’s basic principles are based on anatomical continuity of the optic nerve (ON) sheath with the intracranial dura mater and cerebrospinal fluid (CSF) movement according to pressure gradients. ICP variations are detectable by measuring the sheath diameter of the ON, due to the direct connection between ON sheath diameter (ONSD) compartment and cerebral ventricular system. Sonographic assessment of the ONSD can be considered promising and potentially useful for noninvasive intracranial hypertension screening [4, 5]. ONSD is easy to perform, repeatable at bedside, fast, cheap and radiation free. Previous studies reported a high intra- and interobserver reliability [6, 7]. Despite standardization of ultrasound (US) settings (e.g., US frequency, focus depth, etc.), a clear cutoff for intracranial hypertension, and guidelines/recommendations are yet not available. Furthermore, the measurement of artifacts could alter the real value [8]. We propose and describe a new simple and systematic protocol—the CLOSED protocol (Color Doppler–Low power examination–Optic disk clarity–Safety [short examination duration]–Elevate frequency–Dual measurements)—to perform the sonographic assessment of ONSD to obtain safe, reliable and comparable measurements. The CLOSED protocol is the result of a narrative review on this topic and results of several expert meetings.
Methods and Technique
After a standard learning curve of 25 measurements [9], two different operators of a single-center refined US settings supported by an experienced sonographer. All patients with suspected high ICP or normal pressure hydrocephalus were submitted to ONSD measurements according to the research protocol approved by the local Ethical Committee. US examinations and measurements of the ONSD were performed in B-mode and color Doppler mode on a MyLab™Twice US system (Esaote, Italy), using an 11–3 MHz linear array probe.
Correct identification of the margins, ON course, and its sheath is of paramount importance for the sonographic ONSD assessment. ONSD should be measured 3 mm behind the optic disc, because at this distance the sheath is subject to maximum diameter fluctuations due to ICP [10,11,12]. Color Doppler visualizes the central retinal artery (CRA) and vein (CRV) (Fig. 1), which run in the middle of the distal segment of the ON [13, 14]. CRA and CRV visualization helps to correctly identify the ON course, which can be difficult due to artifacts, wrong angulation of the probe, and nerve tortuosity [15,16,17]. Furthermore, identifying the ophthalmic artery (OA) with color Doppler, which runs parallel to the ON, increases accurate identification of ONSD margins (Fig. 2).
ONSD Assessment with the CLOSED Protocol
CLOSED is our proposed acronym to recall the key elements for an accurate ONSD measurement. A brief description of each element is provided:
-
Color Doppler used to identify CRA, CRV, and OA in order to better detect the nerve course;
-
Low power output a sensitive organ, such as the retina, requires the lowest possible output power for a safe use of diagnostic US;
-
Optic disk clarity the clearance of the optic disk is a minimum requirement for a correct landmark 3 mm behind the papilla.
-
Safety (short examination duration) minimize image information loss recording an US video clip to reduce the examination duration for the safety of the patient (ALARA principle [as low as reasonable achievable]);
-
Elevate frequency for a better quality of US images.
-
Dual measurements two measurements should be taken for each eye: horizontal and vertical.
Device Setting
-
Optimal frequency for US ONSD measurements should be balanced to avoid any potential damage to the ocular structure [18]. Although the high US frequencies produce a better spatial resolution [19, 20], the ALARA principle must always be considered [21]. Despite the availability of very high frequency probes, we suggest using a conventional 10 MHz probe to safeguard the patient’s eye from possible damage. Potential biological damage caused by thermal and mechanical effects was noticed in experimental US conditions [21]. The mechanical index (MI) and thermal index (TI) should be below a cutoff value where harmful effects are possible [29]. According to the Food and Drug Administration recommendations for ophthalmic use, the following upper limit parameters should be observed: TI = Max (TIS_as, TIC) should be ≤ 1 °C; spatial-peak temporal-average intensity (ISPTA.3) ≤ 50 mW/cm2; and MI ≤ 0.23 [22].
-
According to our protocol, the procedure must be stopped immediately if the patient reports any discomfort. Actually, no cases have been reported. No safety concerns for the duration of the insonation are reported. The standard procedure usually requires 1 min for each eye [18].
-
It is important to reduce the power output of the US, usually expressed as a percentage. Authors obtained good images with 20% power and even 25% can be used in compliance with MI and TI limits.
-
The depth should be set at 40–45 mm to obtain a sufficiently large image of the eyeball (which should be kept in the center of the screen) to visualize the anatomical structures in proximity of the nerve.
-
The focus should be kept at a depth of about 25–30 mm—the depth of focus must be positioned where the ON is usually displayed and where the measurement is to be performed. However, adjustments should be taken into consideration according to the anatomy of each patient.
-
The frequency of the probe should be kept at around 10 MHz to obtain the best balance between wave penetration and resolution.
-
The gain compensates signal attenuation (a reduction in sound amplitude), as the waves propagate through tissues. In our experience, gain at 55–60% produces adequate brightness. An excessive gain could lead to excessive contrast and less clear images.
-
The Doppler value of the pulse repetition frequency gives sufficiently clear images when set to 1.0 kHz.
-
Power Doppler can be used to assess the OA waveform to distinguish it from veins based on the classical incisural notch.
-
Authors consider a pediatric linear probe on some patients for anatomical conformation useful. As reported by other authors, the probe is crucial in the ONSD US and only measurements obtained with the same probe-type should be compared [22].
Patient Positioning
-
The technique and safety measures (even though side effects are very low) should be explained to the patient whenever possible [23, 24].
-
The patient should be supine, in neutral head position, with an chest elevation of about 15°–25° to facilitate venous drainage, even though position-related side effects are still debated and controversial [25]. The performer should sit behind the patient’s head. The probe-cable should not be in tension or suspension, in order to minimize tension, which limits fine movements.
-
The lens of the eye is, by definition, a refractive lens and might deflect the angle of US wave incidence. Patients should keep a stable gaze, looking straight ahead, because even with closed eye, pupils could be pointed upward and outward (Bell’s phenomenon)—a natural orientation of gaze—leading to a tortuous ON image similar to a situation of increased ICP [15]. Therefore, keeping the gaze centered, in longitudinal and transverse measurements, reduces the likelihood of error. The Doppler image can help confirming the correct trajectory of the nerve to obtain accurate measurements of the ONSD.
Probe Positioning
-
US gel should be used to avoid air between probe and skin. Eyelashes can present a challenge. Some authors proposed the use of Tegaderm on the closed eyelid to prevent possible direct contact between the eye and the gel [26]. It is important to minimize the pressure with the probe on the eyeball, because it is a sensitive organ and a deformed US image is unreliable.
-
The probe marker should always be oriented in the same way. Conventionally, the authors recommend keeping the probe laterally in the horizontal measurement and superiorly in the vertical measurement.
-
During horizontal measurement, the probe should be placed at an approximate angle of 15°–20° on the patient’s closed upper eyelid. For the vertical measurement, the probe should be placed on the center of the patient’s closed eye, slightly toward the corner of the nose (latero-to-medial direction), trying to intercept the nerve direction (Fig. 3).
Imaging and Measurements
-
An accurate image of the eye and the posterior orbital structures, including the ON and its sheath, should appear on the US screen after checking correct device settings, patient, and probe positioning. The sheath diameter must be measured 3 mm behind the merging point of the nerve from the papilla, which is considered the point of greatest distensibility [10,11,12]. The nerve appears in B-mode, the CSF surrounding the nerve appears black and, externally, the nerve sheath appears as two hyperechoic binaries; furthermore, the papilla of the retina must appear free from false images, clean, and usually hyperechoic. We recommend to take measurements on images in which nerve and sheath are completely visible from the eye to the 3 mm landmark into consideration.
-
The quality of B-mode is often inadequate and the diameter measurement may be unsatisfactory because of unclear margins. To minimize errors induced by acoustic shadow cones and false images, the use of color Doppler is mandatory. The authors suggest to visualize the CRA and/or CRV and to use these vessels as landmarks to identify the correct anatomical plane for reliable ONSD measurements (Fig. 4).
-
The OA is another useful landmark provided by the color Doppler, which is not always visible on the same plane as that of CRA and CRV. The OA runs parallel to the ON [13] and is easy to distinguish from other small arteries for its typical power Doppler curve (Fig. 1). This further anatomical marker may allow the sonographer to be more confident about the margins of the ONSD. However, possible color Doppler artifacts (like the blooming effect) should be kept in mind [27].
-
To reduce the examination time, we recommend capturing a short video clip, from which the performer can then extract the best frame for measurement.
-
Two measurements (dual measurements) should be taken for each eye: horizontal and vertical. The best-captured frame should be used for measurements to minimize any operator-dependent error. In case of doubt or clear discrepancies between the horizontal and vertical measurements, the horizontal measure should be considered as most reliable [22]. At least during the learning stages, we recommend that two operators should select the frame to agree on the four measurement points.
-
Once measurements have been completed, the frame is named and saved, and additional notes are added if necessary (e.g., idiopathic normal pressure hydrocephalus, subarachnoid hemorrhage, idiopathic intracranial hypertension, etc.). The authors suggest using the conventional sequence for measurements (e.g., left eye horizontal; left eye vertical; right eye horizontal; right eye vertical).
Novelty in US Assessment
The novelty of this paper consists of two main points, in terms of US technique: (1) the definition of the recommended setting for the US; and (2) the use of arteriovenous color Doppler to better define the anatomy of the nerve, sheath, and course to obtain a reliable and objective measurement.
The CLOSED protocol includes two important elements allowing a safe and accurate ON sheath evaluation. First, appropriate US settings are essential to keep the procedure safe. In fact, ocular structures, containing a large amount of collagen, like the cornea and eye lens, are US absorbers, which heat up during US exposure. US exposure, especially if prolonged, may cause transient chemosis, conjunctival injection, corneal clouding, lens opacities, reduction in intraocular tension, or permanent destruction of the ciliary body [28, 29]. The CLOSED protocol allows the operator to perform a safe and fast examination adjusting settings to obtain proper and fast image acquisition. In this regard, as recently described by Rasulo and Bertuetti, the use of color Doppler to visualize retinal vessels may play an important role. In fact, visualization of retinal artery allows the operator to reduce the likelihood of artifacts and to better identify borders of the ON sheath [30]. This will increase safety, effectiveness, and accuracy of the examination, reducing US exposure of the ocular structures.
Limitations
Our protocol is based on the results of a narrative review and several expert meetings without strict methodological approach. Furthermore, multiple US systems and machines are currently available and a standard setting may not be achieved. Our purpose was to provide a broad vision to suggest some basic settings rather than a rigid and dogmatic standard. Finally, US measurements are dependent on the operator’s experience and skills. In case of clinical uncertainty, a more objective diagnostic investigation (e.g., computed tomography if rapid/urgent assessment is needed or magnetic resonance imaging in more stable cases) should be performed.
Conclusion
The noninvasive measurement of ICP remains an important element of scientific debate and the role of ONSD in this field is more and more promising. However, ONSD measurements are often not accurate, as a number of artifacts can affect them. We propose the CLOSED protocol as a guide to minimize the occurrence of such artifacts enabling safer, more reliable, and accurate measurements. Color Doppler use could be proposed as a new standard of care for ONSD evaluation. Of course, clinical studies are needed to prove feasibility, applicability, and safety of the CLOSED protocol to test sensitivity, specificity, and overall accuracy compared with a standard technique not including color Doppler visualization of the CRA.
Abbreviations
- ALARA:
-
As low as reasonably achievable
- CRA:
-
Central retinal artery
- CRV:
-
Central retinal vein
- ICP:
-
Intracranial pressure
- ICU:
-
Intensive care unit
- IIH:
-
Idiopathic intracranial hypertension
- iNPH:
-
Idiopathic normal pressure hydrocephalus
- I SPTA.3 :
-
Spatial-peak temporal-average intensity
- MI:
-
Mechanical index
- ms:
-
Millisecond
- ON:
-
Optic nerve
- ONSD:
-
Optic nerve sheath diameter
- PRF:
-
Pulse repetition frequency
- SAH:
-
Subarachnoid hemorrhage
- TI:
-
Thermal index
- TIC:
-
Thermal index cranium
- TIS_as:
-
Soft tissue thermal index at surface
- US:
-
Ultrasound
References
Rossi S, Buzzi F, Paparella A, Mainini P, Stocchetti N. Complications and safety associated with ICP monitoring: a study of 542 patients. In: Marmarou A, et al., editors. Intracranial pressure and neuromonitoring in brain injury., Acta neurochirurgica supplementsVienna: Springer; 1998.
Rebuck JA, Murry KR, Rhoney DH, Michael DB, Coplin WM. Infection related to intracranial pressure monitors in adults: analysis of risk factors and antibiotic prophylaxis. J Neurol Neurosurg Psychiatry. 2000;69:381–4.
Ma R, Rowland D, Judge A, Calisto A, Jayamohan J, Johnson D, et al. Complications following intracranial pressure monitoring in children: a 6-year single-center experience. J Neurosurg Pediatr. 2018;21:278–83.
Robba C, Cardim D, Tajsic T, Pietersen J, Bulman M, Donnelly J, et al. Ultrasound non-invasive measurement of intracranial pressure in neurointensive care: a prospective observational study. PLoS Med. 2017;14:e1002356.
Robba C, Santori G, Czosnyka M, Corradi F, Bragazzi N, Padayachy L, et al. Optic nerve sheath diameter measured sonographically as non-invasive estimator of intracranial pressure: a systematic review and meta-analysis. Intensive Care Med. 2018;44:1284–94.
Moretti R, Pizzi B, Cassini F, Vivaldi N. Reliability of optic nerve ultrasound for the evaluation of patients with spontaneous intracranial hemorrhage. Neurocrit Care. 2009;11:406.
Lochner P, Coppo L, Cantello R, Nardone R, Naldi A, Leone MA, et al. Intra- and interobserver reliability of transorbital sonographic assessment of the optic nerve sheath diameter and optic nerve diameter in healthy adults. J Ultrasound. 2016;11:406.
Popovic D, Khoo M, Lee S. Noninvasive monitoring of intracranial pressure. Recent Patents Biomed Eng. 2009;2:165–79.
Zeiler FA, Ziesmann MT, Goeres P, Unger B, Park J, Karakitsos D, et al. A unique method for estimating the reliability learning curve of optic nerve sheath diameter ultrasound measurement. Crit Ultrasound J. 2016;8:9.
Helmke K, Hansen HC. Fundamentals of transorbital sonographic: evaluation of optic nerve sheath expansion under intracranial hypertension. II. Patient study. Pediatr Radiol. 1996;26:701–10.
Helmke K, Hansen HC. Fundamentals of transorbital sonographic evaluation of optic nerve sheath expansion under intracranial hypertension. I. Experimental study. Pediatr Radiol. 1996;26:701–5.
Hansen H-C, Helmke K. Validation of the optic nerve sheath response to changing cerebrospinal fluid pressure: ultrasound findings during intrathecal infusion tests. J Neurosurg. 1997;87:34–40.
Tsutsumi S, Rhoton AL. Microsurgical anatomy of the central retinal artery. Neurosurgery. 2006;59:870–9.
Ertl M, Barinka F, Torka E, Altmann M, Pfister K, Helbig H, et al. Ocular color-coded sonography—a promising tool for neurologists and intensive care physicians. Ultraschall Med. 2014;35:422–31.
Pribojszki M, Komáromi K, Csomor A, Vörös ES. Tortuous optic nerve-ECR 2017/C-2914. 2017.
Feldman MK, Katyal S, Blackwood MS. US artifacts. RadioGraphics. 2009;24:1179–89.
Baad M, Lu ZF, Reiser I, Paushter D. Clinical significance of US artifacts. RadioGraphics. 2017;37:1408–23.
Food and Drug Administration. Marketing clearance of diagnostic ultrasound systems and transducers—draft guidance for industry and Food and Drug Administration Staff. 2019. https://www.fda.gov/media/71100/download. Accessed 30 Aug 2019.
Hewick SA, Fairhead AC, Culy JC, Atta HR. A comparison of 10 MHz and 20 MHz ultrasound probes in imaging the eye and orbit. Br J Ophthalmol. 2004;88:551–5.
Coleman DJ, Silverman RH, Chabi A, Rondeau MJ, Shung KK, Cannata J, et al. High-resolution ultrasonic imaging of the posterior segment. Ophthalmology. 2004;111:1344–51.
Toms DA. The mechanical index, ultrasound practices, and the ALARA principle. J Ultrasound Med. 2006;25:560–1.
Shah S, Kimberly H, Marill K, Noble V. Ultrasound techniques to measure the optic nerve sheath: is a specialized probe necessary? Med Sci Monit. 2009;15:63–8.
Nelson TR, Fowlkes JB, Abramowicz JS, Church CC. Ultrasound biosafety considerations for the practicing sonographer and sonologist. J Ultrasound Med. 2009;28:139–50.
Safety Group of the British Medical Ultrasound Society. Guidelines for the safe use of diagnostic ultrasound equipment. Ultrasound. 2010;18:52–9.
Romagnuolo L, Tayal V, Tomaszewski C, Saunders T, Norton HJ. Optic nerve sheath diameter does not change with patient position. Am J Emerg Med. 2005;23:686–8.
Roque PJ, Hatch N, Barr L, Wu TS. Bedside ocular ultrasound. Crit Care Clin. 2014;30:227–41.
Rubens DJ, Bhatt S, Nedelka S, Cullinan J. Doppler artifacts and pitfalls. Radiol Clin N Am. 2006;44:805–35.
Shankar H, Pagel PS. Potential adverse ultrasound-related biological effects: a critical review. Anesthesiology. 2011;115:1109–24.
Ertl M, Gamulescu M-A, Schlachetzki F. Application of orbital sonography in neurology. In: Thoirs K, editor. sonography. Rijeka: InTech; 2012.
Rasulo FBR. Transcranial Doppler and optic nerve sonography. J Cardiothorac Vasc Anesth. 2019;33:S38–52.
Acknowledgements
We thank Dr. Monica Ferri for color Doppler learning skills, Dr. Veronica Salvatore for assistance with US settings support, and Dr. Franziska Lohmeyer for the revision of our manuscript.
Funding
The author(s) received no financial support for the research, authorship, and/or publication of this article.
Author information
Authors and Affiliations
Contributions
Aspide, Bertolini was involved in conception and design. Aspide, Bertolini, Albini was involved in acquisition of data. Aspide, Bertolini, Biasucci contributed in drafting the article. All authors critically revised the article and reviewed the submitted version of manuscript. Albini was also involved in Technical/material support. Mazzatenta, Palandri, and Biasucci contributed in study supervision.
Corresponding author
Ethics declarations
Conflicts of Interest
The authors declare that the article content was composed in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
Ethical approval
The local Institutional Review Board approved the present study. CLOSED protocol adhered to ethical standards and Helsinki Declaration.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Springer Nature or its licensor (e.g. a society or other partner) holds exclusive rights to this article under a publishing agreement with the author(s) or other rightsholder(s); author self-archiving of the accepted manuscript version of this article is solely governed by the terms of such publishing agreement and applicable law.
About this article
Cite this article
Aspide, R., Bertolini, G., Albini Riccioli, L. et al. A Proposal for a New Protocol for Sonographic Assessment of the Optic Nerve Sheath Diameter: The CLOSED Protocol. Neurocrit Care 32, 327–332 (2020). https://doi.org/10.1007/s12028-019-00853-x
Published:
Issue Date:
DOI: https://doi.org/10.1007/s12028-019-00853-x