Introduction

Sagittal rotation of a femoral component can influence the kinematics of a replaced knee in various ways [11, 22, 25, 35, 38]. If a femoral component is placed in an overly flexed position, it can cause limited extension or polyethylene post wear resulting from impingement between the anterior part of the polyethylene insert and the anterior margin of the intercondylar box in TKA using a post-cam mechanism [23, 25, 28, 38]. When a surgeon places a femoral component in an overly extended position relative to the femur, he or she may create a notch in the anterior femoral cortex, and some fear these notches may increase the prospective risk of a supracondylar fracture [19, 25, 29, 31]. Few studies have explored the clinical implications of femoral component rotation in the sagittal plane. A couple studies have suggested femoral component sagittal rotation could be influenced by the presence of femoral bowing [35, 38].

Numerous studies report use of a computer-assisted navigation system limits the prevalence of outlier values of limb and/or implant alignment measurements in the coronal plane [2, 5, 8, 9, 12, 14, 21, 30, 32, 37]. Whether the same benefits are realized in the sagittal plane when a surgical navigation system is used is unclear [2, 68, 12, 18, 32]. One reason is simply that fewer studies have sought to quantify alignment in the sagittal plane. A second is that several methods have been used to define reference axes in the sagittal plane [2, 68, 12, 18, 32]. When using a navigation system, a femoral component is placed by referring to the mechanical axis identified by the navigation system [2, 5, 12, 14]. Two different sagittal mechanical axes typically are used by current systems, ie, the line connecting the identified hip center and the registered center of the distal femur, which typically is where an entry point is made for the intramedullary guiding system [2, 5], and the line connecting the femoral head center and the distal femur center identified from kinematically and manually registered points by the navigation system [12, 14]. In contrast, two distal femoral axes frequently are used as references for intraoperative or postoperative assessments of sagittal rotations of femoral components. The distal anterior cortical axis typically is used as an intraoperative reference for correct positioning of a femoral component in the sagittal plane, whereas the distal medullary axis is used as an anatomic reference during postoperative radiographic assessments of sagittal alignment of femoral components [10].

We considered the reported disparities in sagittal rotation of femoral components could be the result of differences between the sagittal mechanical axes used by the navigation system and distal femoral axes used during postoperative radiographic assessments [2, 68, 12, 18, 32]. We therefore sought to (1) quantify the sagittal plane angular deviation between the mechanical axis of the femur and its distal femoral axis in the preoperative TKA population, (2) compare the magnitudes of differences between the mechanical and distal femoral axes measures in the right and left femurs of the same patients, and (3) identify variables that predict the magnitude of deviation between the mechanical and distal axes of a femur. We hypothesized (1) meaningful differences do exist, (2) there would be no difference between right and left femurs of the same patients, and (3) the degree of femoral bowing would be the strongest predictor.

Materials and Methods

We performed radiographic measurements in 200 knees of 100 selected patients undergoing TKA for advanced osteoarthritis between July 23, 2007, and May 15, 2008. We excluded patients who had a prior fracture of the femur, congenital anomaly of the femur, or prior knee or hip arthroplasty. Ninety-five patients (190 knees) were women and five (10 knees) were men. The average age of the patients was 68.1 years (range, 57–80 years), and the average body height, weight, and body mass index were 151.8 cm (range, 134.2–174.4 cm), 64.1 kg (range, 41.9–90.8 kg), and 27.8 kg/m2 (range, 20.0–40.2 kg/m2), respectively. This study was approved by the Institutional Review Board of Seoul National University Bundang Hospital, and all patients provided informed consent for use of their radiographs and medical records.

All radiographic measurements were obtained from true lateral radiographs covering the whole femur using PACS working software (Impax™; Agfa, Antwerp, Belgium). We established a new technique for obtaining a true lateral view of the whole femur in which the femoral head is seen clearly and the contours of both femoral condyles are viewed as one line. Patients placed a thigh on a 17-inch × 17-inch digital flat detector (General Diagnostic Radiographic X-Ray System; Philips, Hamburg, Germany) in a diagonal position, and the xray beam tube then was tilted to a 15° cephalad position (Fig. 1). The radiographs were made preoperatively in all patients.

Fig. 1
figure 1

To obtain a true lateral view of the whole femur, patients placed their thigh on a 17-inch × 17-inch digital flat detector in a diagonal position. The xray beam tube was tilted 15° to directly aim the center of the guide beam toward the center of the digital detector (solid arrow) and the midpoint of the patient’s thigh (asterisk). A sponge block (open arrow) was placed to hold the patient’s leg in a right position.

To ensure the sagittal femoral mechanical axes were representative of the axes used by contemporary navigation systems, we defined two sagittal mechanical axes. Mechanical axis 1 was defined as the line connecting the femoral head center to a point 1 cm anterior to the end of Blumensaat’s line (a line extending through the intercondylar notch on a lateral view of the knee) [4, 24], which is used as the registration point for the distal femur center in many navigation systems, including VectorVision® (BrainLAB, Redwood, WA) [7] and the Stryker Navigation System (Stryker Howmedica Osteonics, Allendale, NJ) [32] (Fig. 2) [2, 5]. To validate our definition of mechanical axis 1, we performed a separate study using five knees, in which we checked the position of a pin placed 1 cm anterior to the top of intercondylar notch on a lateral radiograph taken intraoperatively. We found the position of the pin was within 1 mm of the point 1 cm anterior to the end of Blumensaat’s line in all knees. Mechanical axis 2 was defined as the line connecting the femoral head center and a point identified 65% posteriorly on the line between the anterior cortex and the most prominent point of the posterior medial femoral condyle (Fig. 3). We assumed mechanical axis 2 was equivalent to the sagittal mechanical axis used in the OrthoPilot® navigation system (B. Braun-Aesculap, Tuttlingen, Germany) [12, 14, 26]. The angle between mechanical axis 1 and mechanical axis 2 was measured to evaluate the relationship between the two axes.

Fig. 2A–C
figure 2

(A) A radiograph shows how mechanical axis 1 was defined. (B) An intraoperative photograph and (C) a radiograph display the registration point (asterisk) for the distal femur center and the top point of the intercondylar notch (solid arrow).

Fig. 3A–B
figure 3

(A) A schematic drawing shows how the distal femur center is identified by the OrthoPilot® system. (B) A lateral radiograph shows mechanical axis 2 was defined to represent the point 65% posteriorly in the connecting line between the anterior cortical line and the most prominent point of the posterior femoral condyle.

We identified two distal femoral axes. The distal anterior cortex axis was defined as the line connecting two points on the anterior cortex at 5 cm and 10 cm proximal to the knee line and the distal medullary axis connecting midpoints of the outer cortical diameter at 5 cm and 10 cm proximal to the knee line (Fig. 4). With these definitions, we defined the angles between the two distal femoral axes and each of the two sagittal femoral mechanical axes as proxies of the differences. Negative values were assigned to knees in which the anatomic reference was oriented in an extended position (recurvatum) with reference to the sagittal mechanical axes.

Fig. 4A–D
figure 4

The radiographic measurements of the differences are shown between (A) mechanical axis 1 versus the distal anterior cortex axis, (B) mechanical axis 1 versus the distal medullary axis, (C) mechanical axis 2 versus the distal anterior cortex axis, and (D) mechanical axis 2 versus the distal medullary axis.

To assess the reliability of measurements using the references mentioned, two orthopaedic surgeons (CBJ, KTK) and one clinical investigator (KYG) performed measurements twice with a 1-week interval in 20 patients (40 knees), which were randomly selected from among the 100 patients. The degree of measurement reliabilities was assessed using intraclass correlation coefficients. The intraclass correlation coefficient for intrarater and interrater agreement was greater than 0.86 for all measurements. These findings allowed us to rely on the validities of measurements using the references and measurement data produced by one investigator (CBJ).

To obtain anthropometric femoral dimensions, femur length and degree of femoral bowing were measured (Fig. 5). Femur lengths were measured using long cassette radiographs with grids covering the entire lower limb, which we considered less sensitive to magnification variation than the lateral radiographs. Degrees of femoral bowing were assessed using the angle between the anterior cortical lines of the proximal and distal femur. Proximal anterior cortical lines were drawn as the line connecting the two points 10 cm and 15 cm distal to the proximal end of femoral head and distal anterior cortical lines as the line connecting the two points 5 cm and 10 cm proximal to the distal end of the medial femoral condyle.

Fig. 5A–B
figure 5

(A) The femur length was measured on long cassette AP radiographs using grids, and (B) the degrees of bowing were determined by measuring the angle between the anterior cortical line and the proximal cortical line.

All data were summarized as means, standard differences, and ranges. We calculated the proportion of knees with a deviation greater than 2°, which we assumed to be clinically important considering the navigation system has an accuracy of ± 1° [27] and our radiographic measurements would have 1° variation. The Kolmogorov-Smirnov test was used to confirm measured data followed a normal (Gaussian-shaped) distribution. To determine whether both knees of a patient had similar patterns, measured angles for right and left knees of the same patient were compared, and the differences between right and left knees were determined using the paired-sample t test. To identify predictors of differences between sagittal femoral mechanical axes and distal femoral axes, stepwise multiple linear regression analyses were performed for demographic factors (gender, age, laterality [right or left], body height, weight, body mass index) and anthropometric femoral dimensions (femur length, femoral bowing angle). A priori power analyses were performed to confirm the study sample size had a power greater than 99% to detect a difference of 1° from laterality (right knees versus left knees) with an alpha level of 0.05. The analyses were performed using the SPSS® for Windows® statistical package (Version 12.0; SPSS Inc, Chicago, IL).

Results

Notable differences sometimes were observed between distal femoral axes and the sagittal femoral mechanical axes (Table 1). On average, the distal anterior cortex axis and distal medullary axis were in a flexed position with respect to mechanical axis 1 by 3.0° and 3.8°, respectively, whereas they were in a position almost parallel to mechanical axis 2 by 0° and 0.9°, respectively. The mean angle between mechanical axis 1 and mechanical axis 2 was 3° (standard deviation, 0.3; range, 1°–5°), indicating mechanical axis 2 was oriented in the flexion position to mechanical axis 1 in all knees. Mechanical axis 1 differed by as much as 8.5° from either of the two distal femoral axes, whereas mechanical axis 2 differed by as much as 5.6° from either of the two distal axes. Greater than 20% of the 200 knees had a deviation greater than 2° from the mean value.

Table 1 Summary of the measured parameters

No differences were detected when the differences between the mechanical and distal femoral axes of the right and left knees of the same patients were compared (Table 2). However, in greater than 5% (range, 6%–10%) of the patients, the absolute difference of the deviation between the mechanical axis and distal femoral axis of the two knees was greater than 2°.

Table 2 Comparisons of right and left knees in the same patient

The degree of femoral bowing and femur length predicted differences between the sagittal femoral mechanical axes and the distal femoral axes. The degree of femoral bowing was the strongest predictor of differences followed by femoral length (Table 3). The major portion (ΔR2, 51.9%–73.8%) of the differences was explained by the degree of femoral bowing, and the femur length explained the minor portion (ΔR2, 5.3%–7.1%). For each 1° of femoral bowing or 1 cm of femur length, the deviation between the distal femoral axis and mechanical axis increased by 0.38° to 0.47° or by 0.17° to 0.32°, respectively. The deviation between the mechanical axis and the distal femoral axis was not associated with the other studied factors (age, gender, height, weight, body mass index, knee laterality).

Table 3 Results of the multiple regression analyses*

Discussion

In performing TKA with a computer-assisted navigation system, sagittal mechanical axes identified by the navigation system play a key role as the reference to determine positioning of the femoral component in the sagittal plane. Information on the relationship between the sagittal mechanical axis and distal femoral axes is important for intraoperative evaluations or postoperative radiographic assessments of femoral component sagittal rotation, which can influence the kinematics of a replaced knee [22, 25, 35, 38]. Current navigation systems typically provide a default target value for positioning the femoral component in the sagittal plane using the reference sagittal mechanical axis identified by the system. Target sagittal alignments intended in previous studies that have used a navigation system vary widely from 0° to 5°, but fixed values were used and individual variations were not considered (Table 4) [2, 3, 7, 12, 18, 21, 33]. We therefore sought to (1) quantify the sagittal plane angular deviation between the mechanical axis of the femur and its distal femoral axis in a preoperative TKA population, (2) compare the magnitudes of differences between the mechanical and distal femoral axes measures in the right and left femurs of the same patients, and (3) identify variables that predict the magnitude of deviation between the mechanical and distal axes of a femur.

Table 4 Summary of the studies

Several limitations should be noted when considering our findings. First, anatomic features of the femur may depend on ethnicity. All subjects enrolled in the current study were ethnically Korean, and thus, it may not be possible to directly extrapolate our results to other ethnicities. Second, the majority of our patients were women (only 10 men were included), which limits the statistical power of our study to detect subtle differences by gender. However, this study was conducted in a consecutive series, and the gender proportion of our study population is similar to other TKA series reported from Korea [1517]. Finally, despite our efforts to obtain radiographs of suitable quality, our study is inherently limited by the use of plain radiographs, eg, by issues such as magnification, limb positional variances, and inability to assess three-dimensional features [20, 34]. In addition, we assumed the point 1 cm anterior to the end of Blumensaat’s line represents a typical registration point for the distal femur center, but the position of this registration point depends on technique.

Our findings show substantial differences do exist between the sagittal femoral mechanical axes and distal femoral axes. Mechanical axis 1 differed by as much as 8.5° from either of the two distal femoral axes, whereas mechanical axis 2 differed by as much as 5.6° from either of the two distal axes (Table 1). Greater than 20% of the 200 knees had a deviation greater than 2° from the mean value. Our findings have several clinical implications in applying the navigation technology to TKA. First, observed average deviation values can be used to modify a default target value if the target value intends to position the femoral component with reference to the distal femoral axes. Second, if surgeons want to set the sagittal femoral component rotation with reference to the distal femoral axes, different target values for sagittal femoral component rotation should be applied when using different navigation systems that identify a different sagittal mechanical axis. We found the mean angle between mechanical axis 1 and mechanical axis 2 was 3°, and mechanical axis 2 had smaller differences from the distal femoral axes than mechanical axis 1. However, our findings do not advocate a specific system because there are no documented ideal target values, and surgeons can adjust the target values with the relevant information of the navigation system being used. Third, our findings may explain why the sagittal alignments of femoral components in TKA performed using navigation technology, which relies on the sagittal femoral mechanical axis, vary as determined by postoperative radiographic assessments in a manner that depends on the distal femoral axes.

Intuitively, it might be expected the two femurs of an individual have identical anthropometric features, and thus, we assumed both femurs would yield identical results, and as was expected, we found no major differences between right and left knees in terms of the mean values of differences between the mechanical and the distal femoral axes (Table 2). In contrast, it also would be expected there are tendencies for the right femur to have larger or smaller differences than the left femur, but we found the right and left femurs of the same patient tended to have similar differences between the mechanical and distal femoral axes in the majority (90% or more). However, knees with a deviation greater than 2° sometimes were seen (6%–10%), which may suggest optimization on a knee-by-knee basis is a better option if surgeons prefer to determine sagittal femoral component rotation with reference to distal femoral axes.

We found the degree of femoral bowing to be the strongest predictor of differences between the sagittal mechanical axes and the distal femoral axes. For each 1° of anterior femoral bowing, the angular deviation between the mechanical and distal femoral axes increased nearly 0.5°. Previous studies have reported the existence of anterior femoral bowing [11, 13, 35]. Because of differences in the methods used to evaluate femoral bowing, direct comparisons are difficult, but our study confirms the existence of anterior femoral bowing in Korean patients. Femoral length was the only other studied factor associated with the angular deviation between the mechanical axis and distal femoral axis; a 1-cm increase in femoral length would be associated with an approximately 0.25° increase in the deviation between the mechanical axis and distal femoral axis. Furthermore, based on our results, it seems most important to consider the consequences of using a mechanical axis as an intraoperative reference (as is commonly the case in computer-assisted surgery), a distal femoral axis as an intraoperative reference (as is commonly the case in non-computer-assisted surgery), or a distal femoral axis as a postoperative radiographic reference when the femur is more anteriorly bowed or is longer.

Our data suggest the mechanical and distal femoral axes may differ by as much as 6° to 9° in the sagittal plane, depending on the measurement method used, and identify anterior bowing as the best predictor of such differences. This study suggests, when using navigation technology in TKA, surgeons should consider these differences between sagittal femoral mechanical axes and distal femoral axes, especially if the femur is bowed anteriorly. This is especially important for surgeons who prefer to customize the sagittal flexion angle of the femoral component such that its anterior flange is nearly parallel to the anterior cortex of the femur (Figs. 6, 7). Our findings also may be relevant when measuring rotation of the femoral component in the sagittal plane from postoperative radiographs or when interpreting femoral component sagittal rotation results reported in other studies.

Fig. 6A–D
figure 6

(A) A preoperative radiographic assessment of a patient shows the anterior cortical line is oriented in a position of 6° flexion relative to the mechanical axis 2. (B) Projected into the anterior cortex are the line (dotted line) oriented in the position of 3° flexion reflecting the design feature of 3° flexion in the anterior flange of a femoral component. (C) A postoperative radiograph shows distal femur resection was performed at an angle of 3° flexion relative to the mechanical axis 2. (D) The anterior flange of the femoral component (asterisk) sits perfectly on the anterior cortical surface without anterior notching or anterior flange prominence.

Fig. 7A–D
figure 7

(A) A preoperative radiographic assessment of a patient shows the anterior cortical line is oriented in a position of 1° extension with respect to the mechanical axis 2. (B) Projected off the anterior cortex is the line (dotted line) oriented in the position of 3° reflecting the design feature of 3° flexion in the anterior flange of a femoral component. (C) A postoperative radiograph shows distal femur resection was performed at an angle of 3° extension relative to the mechanical axis 2. (D) The anterior flange of the femoral component (asterisk) sits perfectly on the anterior cortical surface without anterior notching or anterior flange prominence.