Introduction

The long-term results of TKA have been favorable in terms of implant survival; some studies [17, 43, 44] have reported greater than 90% survival at 10 to 15 years. However, this gives no indication of the patient’s subjective impression of function, and several studies have reported dissatisfaction rates of 20% to 40% [1, 15, 29]. Dissatisfaction may be explained partly by abnormal kinematics affecting muscle moment arms and sensations of instability. The advent of Roentgenstereophotogrammetric analysis and videofluoroscopy has enhanced our ability to describe these kinematics.

Uvehammer et al. used Roentgenstereophotogrammetric analysis in vivo to show cruciate-retaining and -sacrificing designs exhibited abnormal kinematics when compared with controls [4042]. They reported all designs had increased femoral anterior translation with flexion when compared with the normal knee and that there was abnormal tibial anteroposterior (AP) motion with knee extension.

Single-plane fluoroscopy has further enhanced understanding and various studies have shown normal kinematics are lost after TKA [2, 4, 6, 25, 31, 38]. Banks and Hodge [3] found a closer reproduction of normal knee kinematics with posterior cruciate-sacrificing designs with an average medial center of axial rotation during knee extension from flexion. After cruciate-retaining TKA, the average center of rotation was displaced lateral to the center of the tibia; this was associated with paradoxic medial femoral condylar roll-forward with flexion for some designs. Wimmer and Andriacchi [46] noted abnormal tractive-rolling forces are generated at the articular interface after TKA because the coefficient of friction (μ) of artificial joints (μ = 0.03–0.10 [16]) is higher than in the natural joint (μ = 0.002–0.03 [14]). This phenomenon also may induce abnormal kinematics by altering the normal rolling/sliding behavior.

The properties of the implants and the nonanatomic shape of their articulations suggest normal knee kinematics are unlikely to be replicated. Erratic stick-slip behavior (the femoral component suddenly sliding forward over the tibia when the expected physiologic motion should be progressive roll-back) arising from the raised friction [46] may partly explain patient perceptions after TKA. However, the unpredictability of knee kinematics after TKA [39] may, at least partly, result from variability in component positioning and the resulting effects on soft tissue behavior [5, 7, 8]. Prior studies have been unable to account for that variability, because they did not document implant positions and measure the kinematics after surgery. The introduction of computer assistance to TKA allows more precise positioning relative to predefined axes that are based on current knowledge and guidelines, reducing variability from the desired alignment [11] and so reducing the shortcomings of earlier methods.

One would anticipate that even with carefully navigated component alignment, TKAs would still cause changes in the path of motion and joint laxity, particularly tibial internal-external and varus-valgus rotations, plus anterior-posterior translations, across the range of knee flexion-extension. For example, take a femoral component with equal medial and lateral fixed radii of curvature in the sagittal plane and a tibial bearing surface that is only partially conforming but with the same sagittal plane geometry medially and laterally. Given the putative accuracy of a navigation system balancing the ligaments, we presume the kinematic changes primarily would be attributable to the articular geometry and secondarily to the loss of the ACL.

We formulated three hypotheses: (1) The design features noted above would inhibit tibial internal-external rotation during knee flexion-extension motion, giving motion more compatible with a fixed axis; (2) Internal-external rotation and varus-valgus laxity would not be changed substantially across the arc of knee flexion-extension; and (3) The normal AP tibiofemoral relationship would not be maintained after TKA, specifically excision of the ACL would result in anterior subluxation under drawer testing and larger than normal AP laxity during flexion-extension motion.

Materials and Methods

We measured natural knee kinematics and laxity across the range of knee flexion-extension for several loading conditions: active knee extension, AP drawer, internal-external rotation, and varus-valgus moments. We then repeated the measurements after TKA so paired comparisons could be made to find changes.

We obtained 10 adult fresh-frozen disarticulated limbs in conformity with local legal requirements. We used two limbs in developing this method, leaving eight knees for analysis unless specified otherwise. A post hoc power analysis indicated that, with the standard deviations calculated, we could detect changes of 6 mm or 11° with 80% power at the 95% level. All were left-sided, had normal alignment, and no evidence of gross arthritic changes, ligamentous instability, or previous surgery. Navigation trackers (Stryker Knee Navigation System; Stryker Leibinger, Freiburg, Germany) were fixed securely to the outer cortices of the femora and tibiae 150 mm from the joint line. For each specimen, we recorded the position of standard navigation reference points at the center of the femoral head and ankle. The legs were divided 200 mm above and below the joint line and mounted in a dedicated loading rig with the transepicondylar axis aligned approximate to the flexion-extension axis of the rig [12] (Fig. 1). The rig allowed unconstrained tibial motion relative to the femur, apart from control of flexion-extension, so the knee did not need its axis of flexion aligned to that of the test rig; the axis alignment had no effect on the relative tibiofemoral motions that were calculated only between the two bone-mounted motion trackers. The tibia hung free, allowing flexion-extension of the knee (0°–120°) by moving the femur in an approximate vertical arc and applying displacing loads to the quasistatic tibia [18].

Fig. 1
figure 1

The experimental setups to measure the kinematics and envelopes of knee laxity limits are shown.

To obtain active knee extension, we applied a quadriceps tension of 200 N to the patella through a cable attached to a pneumatic cylinder acting parallel to the femoral axis. This tension was limited by the fragility of the specimens from elderly cadavers; it was sufficient to ensure tibiofemoral joint compression through the range of motion (ROM). Tibial internal-external rotation moments of 5 Nm could be applied by hanging weights on two cords that pulled in opposite directions and were attached at the opposite sides of a large pulley connected to an intramedullary rod in the distal tibia. This induced tibial rotation without causing a resultant translation force and has been used in similar work [18]. Similarly, we applied tibial varus-valgus moments of 3.5 Nm using a cord-pulley system pulling medially or laterally on the distal end of the intramedullary rod. Anteroposterior forces of 70 N were applied to the proximal tibia through a low-friction bearing on a hoop so the resultant force was always through the center of the tibial plateau and secondary rotations were not inhibited. The displacing forces and moments were chosen to reflect those imposed during clinical examination of joint laxity, again with limits imposed by the fragility of the specimens; clinical anterior drawer tests often use 89 N force [13]. We then extended and flexed the femur by hand, with and against the quadriceps action, with each motion cycle taking approximately 5 seconds. This rate of motion allowed collection of sufficient kinematic data points.

The navigation system recorded the movements of the femur and tibia of the intact knee during active knee extension motion from 90° to 0° for the following loading conditions: internal rotation moment (5 Nm), external rotation moment (5 Nm), anterior drawer (70 N), posterior drawer (70 N), varus moment (3.5 Nm), valgus moment (3.5 Nm), and neutral (no additional loading other than the simulated extensor load).

We then inserted a Scorpio CR (Stryker Orthopaedics, Mahwah, NJ) posterior cruciate-retaining prosthesis in a classic standardized alignment [21]. This implant was widely used and linked to a navigation system that would allow accurate implant positioning and measurement of kinematics. The implant had design features that were expected to affect the kinematics and laxity of the knee. The femoral component had equal-sized medial and lateral condyles that had a fixed radius of curvature in the sagittal plane; the tibial bearing surface was partially conforming and had the same sagittal plane geometry medially and laterally. The distal femur was cut perpendicular to the computer estimation of the mechanical axis in coronal and sagittal planes; the resection was 0 to 1 mm deeper than the thickness of the distal part of the femoral component. We determined the rotation of the femoral component according to Berger et al. [9]. The final cuts were made after ensuring accurate AP positioning with the navigation. We cut the tibia perpendicular to its mechanical axis. The resection level matched the thickness of the insert, and the rotational alignment was referenced using the PCL and medial one-third of the tibial tuberosity. We cemented the tibial component and the femoral component was press fit. The arthrotomy was closed in the same manner as in clinical practise with Number 1 Vicryl (Ethicon, Somerville, NJ) to fascial layers and 2–0 Vicryl to fat and skin. We repeated the kinematic measures for the same conditions as the intact knee.

Kinematics data were described with six degrees of freedom according to the Grood and Suntay joint coordinate system [19] without adjustment for clinical abduction-adduction. Flexion-extension was a tibial rotation about the digitized femoral epicondylar axis [9]; internal-external rotation was about the long axis of the tibia, the zero point of which was defined by the relative position of the malleolar mediolateral axis and the femoral epicondylar axis; and varus-valgus motion (change of alignment in the coronal plane is, mechanically, a rotation) was about a floating axis mutually perpendicular to the other two axes in the AP direction [19]. We measured translations along these axes; only AP translation is presented. Care was taken to define the centers of the hip, knee, and ankle to minimize crosstalk, which is an inaccuracy in the calculated kinematics arising from misalignment of coordinates [33]; the center of the femoral head was known; the center of the knee was defined as the highest point of the anterior-distal outlet of the intercondylar notch; and the center of the ankle was the midpoint of a line joining the malleoli. The average of three extension cycles over the range 90° to 0° was used for analysis.

We used paired two-way t-tests to compare the neutral paths of motion and the joint laxity with the knee in two states: without and with the TKA, and across the range of knee flexion-extension. The dependent variables were the primary motions of the knee, ie, tibial AP translation, tibial internal-external rotation, or tibial varus-valgus angulation. We did not analyze coupled motions. When describing the difference between intact and TKA implanted knees, the 95% confidence interval bands of this difference were calculated and plotted in the graphs of results. Data relating to each of the hypotheses (tibial internal-external rotation, varus-valgus rotation, and AP translation) were calculated for the neutral path of motion and the limits of laxity, across the range of knee flexion-extension, and at the forces or moments applied (sometimes known as the envelope of laxity [10]). The neutral path of motion was that followed by the tibia when the knee extended with only the specimen weight and quadriceps tension acting.

Results

When the intact knee extended from 25° flexion to full extension, the tibia rotated externally by a mean of 8°; this was the screw-home mechanism. After TKA, there was almost no tibial rotation as the knee extended (Fig. 2). The tibia was in the same mean rotation after TKA as it had been when the intact knee was externally rotated after the screw-home in extension. After TKA, the tibia was rotated more externally (0.003 ≤ p < 0.05) than the intact knee had been over the extension range of 63° to 11°.

Fig. 2A–B
figure 2

Comparison of the internal-external rotation pathways of the natural and prosthetic knees shows that the screw-home mechanism was abolished by the prosthesis. (A) The neutral pathway of tibial internal-external rotation for the intact knee (n = 8, ±standard deviation), and (B) the neutral pathway of tibial internal-external rotation after TKA implantation (n = 8, ±95% confidence interval of the difference between intact and TKA) are shown. Considerable differences are indicated with the bracket.

The tibial internal-external rotation laxity for the intact and post-TKA knees reduced progressively as the knee extended (Fig. 3). For the intact knee, both limits of the rotational laxity envelope moved into tibial external rotation in terminal knee extension, similar to the screw-home motion. Because the screw-home was lost after TKA, the limits of internal and external rotation laxity after TKA were more externally rotated than for the intact knee: for external rotation, 0.026 ≤ p < 0.05 from 32° to 14° knee extension, and for internal rotation, 0.007 ≤ p < 0.05 from 41° to 20° extension (Fig. 3).

Fig. 3A–B
figure 3

Comparison of the ranges of internal-external rotation of the natural and prosthetic knees shows that, although the range of laxity was maintained, it was shifted by the loss of the screw-home mechanism after TKA, reflecting the loss of the screw-home. (A) Tibial internal-external rotation envelope of laxity for the intact knee (n = 8, ±standard deviation), and (B) tibial internal-external rotation envelope of laxity after TKA implantation (n = 8, ±95% confidence interval of the difference between intact and TKA) are shown. Differences are indicated with the brackets.

Neither the neutral pathway nor the limits of laxity of tibial varus-valgus motion was changed (p > 0.05) by TKA (Figs. 4, 5). On average, the neutral paths of the tibiae were aligned within 1° to the mechanical axis over the full ROM examined. However, the large confidence intervals of the difference between the normal and implanted knees show some had a large difference from average behavior. Data for one knee were lost in the varus-valgus loading condition, therefore, data for seven are presented here.

Fig. 4A–B
figure 4

Implantation of the prosthesis did not affect the mean varus-valgus alignment of the knee; the large CI indicates that some knees had motion differing from the mean. (A) The neutral pathway of tibial varus-valgus angle for the intact knee (n = 7, ±standard deviation), and (B) the neutral pathway of tibial varus-valgus angle after TKA implantation (n = 7, ±95% confidence interval of the difference between intact and TKA) are shown.

Fig. 5A–B
figure 5

The TKA did not affect the range of varus-valgus laxity as the knee flexed, but there was a tendency for the range of laxity to shift into varus. (A) Tibial varus-valgus angle envelope of laxity for the intact knee (n = 7, ±standard deviation), and (B) tibial varus-valgus angle envelope of laxity after TKA implantation (n = 7, ±95% confidence interval of the difference between intact and TKA) are shown.

The TKA did not have a large effect on either the neutral path of AP motion or the AP laxity when the knee was near extension (Figs. 6, 7). In the flexed knee, the tibial positions were shifted considerably posteriorly after TKA for the neutral path (0.024 ≤ p < 0.05 from 87° to 75° extension) and with a posterior drawer force (0.0002 ≤ p < 0.05 from 87° to 44° extension). The anterior laxity did not change very much (p > 0.05) at any angle of knee flexion examined.

Fig. 6A–B
figure 6

After TKA, the tibial position moved progressively posteriorly with knee flexion compared with the natural position. (A) The neutral pathway of tibial AP translation for the intact knee (n = 8, ±standard deviation), and (B) the neutral pathway of tibial AP translation after TKA implantation (n = 8, ±95% confidence interval of the difference between intact and TKA). Considerable differences are indicated with the bracket.

Fig. 7A–B
figure 7

Although the range of AP laxity was maintained after TKA, it shifted progressively posteriorly as the knee flexed compared with the natural range of tibial positions. (A) Tibial AP translation envelope of laxity for the intact knee (n = 8, ±standard deviation), and (B) tibial AP translation envelope of laxity after TKA implantation (n = 8, ±95% confidence interval of the difference between intact and TKA). Considerable differences are indicated with the bracket.

Discussion

It has been difficult to relate knee kinematics to specific details of a surgical procedure or implant design, partly because of the inherent lack of precision in surgery. That has been overcome to a large extent by the introduction of optical navigation systems that typically can measure to ±0.5 mm or ±0.5°, thus allowing consistent TKA positioning in relation to anatomic features. Improving kinematics should improve the function of patients after TKA, particularly with increasing use in younger patients. We report data arising from placing one type of prosthesis at a position to which the surgeon was guided by navigation software; the resulting construct led to some deviations from the natural path of motion and laxity limits. Analysis of the design of the chosen prosthesis, along with a review of the literature, led to several hypotheses relating to alterations from the natural motion and laxity of the knee caused by the TKA. The chosen implant design had equal-sized femoral condyles, of constant radius, that articulated with partly conforming fixed tibial bearing surfaces that also were the same size. We hypothesized these features would lead to motion compatible with that of a fixed axis, losing rotational components of motion about other axes. Because of the constant-radius femoral condyles and accuracy of the navigation system when setting up the collateral ligaments, it was hypothesized that internal-external rotation and varus-valgus laxity would remain within normal limits across the arc of knee flexion-extension. Published data regarding loss of roll-back led to the hypothesis that the normal AP tibiofemoral roll-back relationship would not be maintained after TKA. Excision of the ACL led to the hypothesis that the tibia would subluxate anteriorly under drawer testing and have larger than normal AP laxity.

These findings must be interpreted in light of the specific limitations of these experiments in vitro: the use of TKA with a fixed bearing and equal-sized medial and lateral femoral condyles with one radius of curvature in the sagittal plane; loading imposed only by a quadriceps tension; and analysis only of knee extension motion. Although the specimens did reflect the ages of patients undergoing TKA, these specimens were more normal than those to be replaced; however, we believe it is important to know how well a TKA restores normal function. We also imposed only a quadriceps load to the knees to provide an extension moment; simulation with other muscles could alter the kinematics and numerous combinations are possible. We chose not to add muscle cocontractions because that would have added to the complexity of an already complex experiment. The important point is the geometry and loading were well-defined; we believe differences between kinematics studies may result from lack of control of the forces across the knee, especially in vivo. The quadriceps tension was limited to 200 N because we found in an earlier pilot experiment that we sometimes damaged frail cadaveric knees when applying 400 N. However, this force was sufficient to ensure that the articular surfaces were compressed together during the tests, so the knee motion was controlled by the implant geometry. We postulate that a larger compressive force would have reduced the limits of laxity for the natural and artificial joints. The validity of the data resulting from this work in vitro may be judged by reference to data obtained in vivo. Our experiments differed from those of Stiehl et al. [39] in that we did not observe the erratic AP motion they reported. However, loss of screw-home rotation was reported in a fluoroscopic study of the same prosthesis in vivo [26].

We confirmed the hypothesis that symmetric and conforming condyles of the TKA components would inhibit the natural screw-home rotation kinematics. The prosthesis maintained a constant rotation that matched the natural knee in full extension. That stable position reflected the close-packing of the articular surfaces with simultaneous tightening of the soft tissues around the knee. The natural screw-home motion in the intact knees was in accordance with previous studies [10, 20, 24, 27], although variations have been reported and Lafortune et al. [28] did not find screw-home motion. These variations might reflect differences in loading conditions and definition of axes of motion [33]. Screw-home sometimes is lost after TKA [30, 31, 42], although that is not inevitable [23, 37]. Differences in behavior may reflect the use of symmetric versus asymmetric condyles, inherent constraint between the components, or differing alteration of the soft tissues during implantation. In particular, tightening of the ACL as the natural knee reaches extension imposes an external rotation torque on the tibia [20], which was lost by ACL excision during TKA.

We also confirmed the hypothesis that the limits of tibial rotation laxity after TKA would remain similar to those of intact knees. The pattern of reducing laxity as the natural knee extended, from 27° to 6° internal-external rotation, was matched closely by the prosthetic knee. This suggests progressive tightening of the soft tissues as the natural knee extended was preserved after TKA and constraint in the articulation was not excessive. However, the limits of rotational laxity after TKA also reflected the loss of the screw-home and so were considerably externally rotated compared with normal, from approximately 40° to 10° knee extension. Whiteside et al. [45] suggested rotational constraint is not necessary in a TKA to maintain normal rotational laxity as long as the soft tissues are tensed correctly.

The data confirmed the hypothesis that after TKA, varus-valgus laxity would not be altered substantially across the arc of knee flexion-extension. Varus-valgus alignment was aided by accuracy of the navigation system when cutting the femoral condyles and setting up the collateral ligaments to have constant tension when moving around the constant-radius prosthetic condyles. It was expected that knee varus-valgus alignment would be maintained because of the accuracy of the navigation system. The tendency to move into varus in flexion after TKA reflected the femoral component rotation, that followed from the surgeon’s judgement when digitizing the transepicondylar axis, which deviates from the posterior condylar line [36]. However, the extra bone resected from the posterior medial condyle was matched at the tibia, maintaining a parallel space for the implant.

Finally, the data do not support the hypothesis that excision of the ACL and preservation of the PCL during TKA would lead to the tibia moving in a relatively anteriorly translated position. After TKA, the AP position of the tibia was normal when the knee was in extension, but it shifted considerably posteriorly in flexion, the opposite tendency to what was expected. This indicates a loss of femoral roll-back in flexion and loss of normal restraint by the PCL. We confirmed, by reopening the knees, that the PCL always remained intact. This implies the size, position, and shape of the femoral component, which acts like a cam against the tibial component concavity, had effectively allowed the PCL to slacken as the knee flexed. The abnormal posterior translation of the tibia did not imply stretching of the PCL if the implant geometry had allowed the tibia to move proximally to maintain contact, as has been reported [22]. The kinematic data showed that there had been a mean proximal translation of 1.8 mm at 70° flexion. Conversely, the tibia tended not to be drawn forward to its physiologic anterior laxity limit in the flexed knee, despite the ACL having been excised. These observations suggest the prosthetic articulation forced the tibia posteriorly in the flexed knee. In addition, the quadriceps tension would have acted to prevent tibiofemoral distraction, which would result from the femoral component sliding up the slope of the posterior part of the tibial articular surface during anterior drawer testing. Thus, the mechanics of the loaded articulation prevented femoral roll-back (the posterior movement of the tibiofemoral contact points [35]) seen in the natural knee [34]. This has been reported with other implants [32] and causes posterior impingement and loss of knee flexion [6].

The clinical relevance of this in vitro study relates to the desire to improve function of patients after TKA, particularly with younger patients undergoing TKAs. In the past, it has been difficult to relate knee kinematics to specific details of surgical procedure or implant design, partly because of the inherent lack of precision in surgery. That has been overcome to a large extent by the introduction of optical navigation systems that can measure typically to ±0.5 mm or ±0.5°, thus allowing consistent prosthesis positioning in relation to anatomic features. We report data arising from placing one type of prosthesis at a position to which the surgeon has been guided by navigation software; the resulting construct led to some deviations from the natural path of motion and laxity limits. Much of that deviation is now explainable by reference to the prosthetic articular geometry, as discussed previously. It is not known whether the datum position in the software will give the best restoration of physiologic kinematics, but the technology now can be applied to study this question. Such a study carries the inherent assumption that a search for restoration of normal kinematics is desirable for optimizing patient function, and that also is not yet known but does appear a reasonable assumption.

Our data suggest the kinematics of the natural knee were not maintained in all respects after arthroplasty with an implant that had equal-sized circular femoral condyles, when loaded and moved in vitro. Use of a navigation system enabled accurate maintenance of tibiofemoral coronal alignment and varus-valgus limits of laxity, across the range of knee flexion-extension. However, the conforming cylindrical articular geometry led to loss of the natural pattern of tibial rotation, the screw-home mechanism, whereas the range of internal-external rotation laxity was maintained. Contrary to the initial hypothesis, excision of the ACL during arthroplasty did not lead to abnormal tibial anterior translation. On the contrary, there was lack of physiologic femoral roll-back on the tibial plateau when the knee was flexed after arthroplasty, in line with published studies using video fluoroscopy in vivo. Although these findings are implant-specific, the differences between the kinematics of natural and artificial knees suggest that an adjustment of the guidance given by the navigation system might lead to an implant position that provides better maintenance of physiologic motion.