Introduction

Thermal Spraying: Plasma-Spraying Process

Thermal spraying is a coating deposition technique. Powder or wire materials are heated to a molten or partially molten state by certain heat sources (such as electric arc, flame, plasma, and laser, etc.). The molten droplets are accelerated to impinge on the substrate at high velocities, solidify, and form the coating. The thermal-spraying method and spray materials primarily determine the coating quality (Ref 13). Among current thermal-spraying methods, the plasma-spray process using conventional DC discharge plasma torches is widely used in industry. Different construction principles exist with single or multiple electrodes (cathode, anode) as well as with one-piece or cascaded nozzles. A conventional DC plasma torch with a rod-shaped tungsten cathode surrounded concentrically by the plasma gas flow and a one-piece nozzle operating as the anode is shown in Fig. 1 (Ref 4, 5).

Fig. 1
figure 1

Plasma-spray process with conventional DC plasma torch (F4-type)

Particle Diagnostics During the Thermal-Spray Process

High-quality plasma-sprayed coatings require a uniform treatment of the particles injected by the plasma jet. Diagnostic systems (Ref 6, 7) help us to understand the fundamental characteristics and physics of the spray process. These analytic systems focus on one or more significant particle properties (temperature, velocity, number density, etc.). The measurement principles (Ref 610) are based on different physical effects, e.g., scattering, radiation, and absorption. Such diagnostic systems provide results helping us improve the process, and they are also used as validation or start data in numerical computer simulations. Thermal-spray diagnostics can be divided into four areas of application (system diagnostics, flame or plasma diagnostics, particle diagnostics, coating diagnostics, and tests). Table 1 provides an overview of typical particle diagnostics (Ref 1115).

Table 1 Overview of different particle diagnostics

Our main goal was to identify spray parameters (distance, temperature, pressure, etc.) for optimal coating properties, as well as to monitor the process conditions. The kinetic and thermal characteristics of the powder particles determine the quality of the coating. The most important particle properties are (Ref 6, 7, 10) size \( d_{\text{P}} \), shape S P, number density n P, velocity \( \vec{v}_{\text{P}}, \) and temperature T P. Many diagnostic systems have been developed to obtain information about these properties and most of them are well suited for online operation.

Principles of the Particle Diagnostic Systems

Laser Doppler Anemometry

Laser Doppler anemometry (LDA) is a flexible and well-established nonintrusive method of investigating the local velocity components of gases, liquids, and multicomponent flow mixtures with the help of tracer particles. In this case, LDA was used to investigate thermal spraying, whereby the particles of the injected spray powder served as tracer particles. This allows the velocity of injected powder particles to be measured. In addition, the velocity of the powder-carrying plasma flow can be determined. However, a lag between the particles and the particle-carrying medium has to be taken into account. Commercial LDA configurations are available from DantecDynamics and TSI (Ref 1214), and interesting setups and developments (e.g., mobile robust system miniaturization) have been accomplished. LDA is based on the principle of light scattering, whereby particles are illuminated by a laser beam (Fig. 2). It can be described by Mie theory. The light-scattering process is classified according to the Mie parameter q, which describes the ratio between size and the incident wavelength and is defined as

$$ q = \frac{{\uppi d_{\text{p}} }}{{\uplambda_{\text{laser}} }}, $$
(1)
Fig. 2
figure 2

Experimental setup used for LDA

where d p is the particle diameter, and \( \uplambda_{\text{laser}} \) is the wavelength of the laser beam. Scattering is thus divided into q ≪ 1 Rayleigh scattering, q ≈ 1, q > 1 Mie scattering (the typical range for LDA), and q ≫ 1 applied for geometrical optics. Figure 3 represents the dependence between maximum, minimum, and the velocity component perpendicular to the interference fringes. The LDA measurement volume in our case is built by the following parameters: MV-measurement volume (l x  = 0.32 mm × l y  = 21.34 mm × l z  = 032 mm ellipsoid, fringe value N = 37), laser beam radius 0.16 mm, focus lens 1000 mm, laser wavelength 532 nm, and LDA 2Θ-angle 1.7188°. The LDA process can be explained in two alternative but equivalent ways (Ref 9, 1620): as an interference pattern and by means of the Doppler effect. To produce an interference pattern, two laser beams are superposed in the measurement volume to produce a periodic pattern of intensity maxima and minima parallel to the bisecting line of the beams (Fig. 3). The distance \( d \) between two consecutive maxima and minima is \( d = \frac{{\uplambda_{\text{laser}} }}{2\sin (\Uptheta /2)} \), where \( \Uptheta \) is the angle between the two incident laser beams. These particles flying through the measurement volume produce a periodic scattering signal (frequency f), when crossing the interference fringes. A photomultiplier records these signal. The velocity component \( v_{x} \) perpendicular to the interference fringes is calculated according to

$$ v_{x} = df = \frac{{\uplambda_{\text{laser}} f}}{2\sin (\Uptheta /2)}. $$
(2)
Fig. 3
figure 3

LDA measurement volume (interference fringe pattern)

The intensity of the signal depends on the detection direction and the object size (particle diameter). Assuming spherical particles and using Mie theory (Ref 10, 21), this angular function can be determined. Alternatively, the Doppler shift in the light scattered by the flying particles can be used. The Doppler effect occurs twice, once when the incident laser light (beam 1 with wave vector \( \overrightarrow {{k_{1} }} \) and beam 2 with wave vector \( \overrightarrow {{k_{2} }} \)) of the transmitter system, characterized by the wavelength \( \uplambda_{\text{laser}} = \frac{2\uppi }{{k_{1} }} = \frac{2\uppi }{{k_{2} }} \) and frequency \( f_{\text{laser}} = f_{1} = f_{2} \) impinges on the moving target (velocity \( \overrightarrow {v} \)), and once when the light with frequencies \( f_{{{\text{p}}1}} \) and \( f_{{{\text{p}}2}} \) (subscript p for particle; subscripts 1 and 2 for laser beams 1 and 2) is scattered by the moving particle and received by the detector with frequencies \( f_{{{\text{r}}1}} \) and \( f_{{{\text{r}}2}} \) (subscript r for receiver). The frequency shifts are calculated by

$$ \Updelta f_{{{\text{p}}1}} = f_{{{\text{p}}1}} - f_{\text{laser}} = - \frac{{\overrightarrow {{k_{1} }} \;\overrightarrow {v} }}{2\uppi }\;\;\;\;{\text{and}}\;\;\;\;\Updelta f_{{{\text{p}}2}} = f_{{{\text{p}}2}} - f_{\text{laser}} = - \frac{{\overrightarrow {{k_{2} }} \;\overrightarrow {v} }}{2\uppi }; $$
(3)
$$ \Updelta f_{{{\text{r}}1}} = f_{{{\text{r}}1}} - f_{\text{laser}} = - \frac{{\overrightarrow {{k_{1} }} \;\overrightarrow {v} }}{2\;\uppi } + \frac{{\overrightarrow {{k_{S} }} \;\overrightarrow {v} }}{2\;\uppi }\;\;\;\;{\text{and}}\;\;\;\;\Updelta f_{{{\text{r}}2}} = f_{{{\text{r}}2}} - f_{\text{laser}} = - \frac{{\overrightarrow {{k_{2} }} \;\overrightarrow {v} }}{2\uppi } + \frac{{\overrightarrow {{k_{\text{S}} }} \;\overrightarrow {v} }}{2\uppi }. $$
(4)

The difference \( \Updelta f_{\text{D}} \) between the two frequencies of the light components scattered by the moving particle and recorded by the photo multiplier is

$$ \Updelta f_{\text{D}} = f_{{{\text{r}}2}} - f_{{{\text{r}}1}} = \Updelta f_{{{\text{r}}2}} - \Updelta f_{{{\text{r}}1}} = \frac{{\overrightarrow {{k_{1} }} \;\overrightarrow {v} }}{2\uppi } + \frac{{\overrightarrow {{k_{2} }} \;\overrightarrow {v} }}{2\uppi } = \frac{{\left( {\overrightarrow {{k_{1} }} - \overrightarrow {{k_{2} }} } \right)\;\overrightarrow {v} }}{2\uppi }. $$
(5)

Using \( \left| {\overrightarrow {{k_{1} }} - \overrightarrow {{k_{2} }} } \right| = 2\left| {\overrightarrow {k} } \right|\sin \left( {\frac{\Uptheta }{2}} \right) = \frac{4\uppi }{{\uplambda_{\text{laser}} }}\sin \left( {\frac{\Uptheta }{2}} \right), \) the particle velocity component \( v_{x} \) can be determined:

$$ v_{x} = \frac{{\Updelta f_{\text{D}} \uplambda_{\text{laser}} }}{2\sin (\Uptheta /2)}. $$
(6)

DPV-2000 Particle Diagnostic System

The DPV-2000 (Ref 11) enables measuring particle velocities, temperatures, and diameters. Figure 4 shows the basic operating principles of the system. The velocity is obtained by determining the time between two signals triggered by a radiating particle passing the two-slit mask of the optoelectronic sensor head. In conjunction with the distance of the slits \( d_{\text{slit}} \) and the magnification factor of the lens, \( f_{\text{magn}} \) the particle velocity \( v_{\text{p}} \) can be calculated by

Fig. 4
figure 4

Basic principles of the particle diagnostic system DPV-2000

$$ v_{\text{p}} = \frac{{d_{\text{slit}} }}{\Updelta t}f_{\text{magn}} . $$
(7)

The particle temperature T P is acquired by two-color pyrometry, i.e., by calculating the ratio of the energy E(λ i ) radiated at two different wavelengths λ i assuming that the particles are gray-body emitters with the same emissivity at both color bands by

$$ T_{\text{p}} = \frac{{c2\left( {\frac{1}{\uplambda 1} - \frac{1}{\uplambda 2}} \right)}}{{\ln \left( {\frac{E(\uplambda 1)}{E(\uplambda 2)}\left( {\frac{\uplambda 1}{\uplambda 2}} \right)^{5} } \right)}}, $$
(8)

where c 2 = 1.438775 10−2 mK is a constant. The particle diameter d P is obtained from the radiation energy emitted at one wavelength, assuming that the melted particles are spherical or almost spherical by

$$ d_{\text{p}} = \sqrt {\frac{E(\uplambda i)}{{f_{\text{diam}} }}} . $$
(9)

Since it is necessary to know the real emissivity of the particle, usually a powder with known diameter distribution has to be measured in advance to determine the calibration factor f diam. However, since possible changes in emissivity with temperature or molten state could affect the accuracy of particle size measurement for a given calibration, the size distributions of the powders applied in this study were independently measured by laser diffraction, and the DPV was calibrated correspondingly as described later. As the measurement volume is relatively small (<1 mm3), the data are collected for individual particles and can subsequently be analyzed statistically. A certain measurement time is necessary to support the mean and standard deviations by sufficient individual particle number of datasets (Ref 2224).

Experimental Setups

Experimental Environment

Measurements of the in-flight particle velocity of the plasma-sprayed particles were performed at IEK-1, Forschungszentrum Jülich GmbH. Two different diagnostic methods were compared. One of them was LDA, which was set up and operated by IPM at the Universität der Bundeswehr. The other diagnostic system was the DPV-2000 [TECNAR Automation Ltd., St-Bruno, QC, Canada (Ref 11, 23)], which was operated by Forschungszentrum Jülich, IEK-1. The comparative measurements were carried out on a Multicoat™ facility (Sulzer Metco, Wohlen, Switzerland) during atmospheric plasma spraying with a single-cathode F4 gun. The spray parameters are listed in Table 2.

Table 2 F4 plasma-spray parameters

Two 8 wt.% yttria partially stabilized, zirconia-spray-dried powders were used. The finer powder (in the following, designated F) was a Sulzer Metco 204NS-G (−106/+11 μm). The particle diameters determined by laser diffraction were d 10 = 8 μm, d 50 = 26 μm, and d 90 = 51 μm. The coarser powder (in the following, designated C) was a Sulzer Metco 204NS (−125/+11 μm). The particle diameters determined by laser diffraction were d 10 = 28 μm, d 50 = 57 μm, and d 90 = 92 μm. Figure 5 shows the measured particle size distributions. The morphology was spheroidal due to the manufacturer’s proprietary HOSP™ process (Sulzer Metco Europe GmbH, Kelsterbach, Germany).

Fig. 5
figure 5

Particle size distributions of the applied YSZ powder feedstock determined by laser diffraction method

Laser Doppler Anemometry

The system was developed by the Universität der Bundeswehr München. The principal setup was built comprising a CW laser (coherent, type Compass 315M, Nd-YAG Laser 100 mW), a photodetector (Dantec Type 57X08PM (500 MHz)), and a computer (ASYST software). The measuring positions within the plasma jet are arranged in a quadratic planar grid perpendicular to the gun axis. The center point of the grid is positioned on the gun axis. The gun was moved by an x-y-z positioning unit so that the grid points were subsequently located in the measurement volume (Fig. 6) (Ref 8, 25). The measurement time per grid point was 1 s or less if the maximum number of 1500 detected particles had been reached earlier. In the periphery of the measurement grid, only very few particles were detected. As their characteristics are not representative for the total particle fraction, the results at these points were not considered. At every point of the grid, a complete evaluation was carried out, the results of which can be presented in different ways (histogram of velocity distribution, contour plots of number density or velocity) (Ref 8, 25).

Fig. 6
figure 6

Measurement setup for laser Doppler anemometry

DPV-2000 Particle Diagnostic System

For the DPV-2000 measurements (Ref 11, 2224), the gun was positioned at a fixed location, and the measurement head was moved in the y-z plane perpendicular to the gun axis at a spray distance of x = 100 mm (Fig. 7). The 13 × 13 point measurement grid was 30 × 30 mm2. The measurement time per grid point was 10 s or less if the maximum number of 1000 detected particles had been reached earlier. In our measurements, we did not find any further statistical improvement of the results if more than 1000 particles were evaluated. The measurement head was adjusted so that the measurement grid was centered at the location of maximum particle flow. Again, the results at grid points in the periphery of the grid were not considered because very few particles were detected there.

Fig. 7
figure 7

Measurement setup for the particle diagnostic system DPV-2000

Measurement Results and Discussion

Initially, two LDA measurements (denoted LDA 1 and LDA 2) were performed with powder F at a spray distance of x = 100 mm in the planes perpendicular to the gun axis: LDA 1 on a 11 × 11 point grid measuring 80 × 80 mm2, and LDA 2 on a 13 × 13 point grid measuring 60 × 60 mm2. The DPV results were obtained on a 13 × 13 point grid measuring 30 × 30 mm2 at the same spray distance. Shorter or longer spray distances than this operational stand-off were not investigated.

In order to compare the LDA and the DPV-2000 results, the measurement grids of both methods must be congruent. As they were not located at physically identical positions, the grids were aligned with each other by horizontal and vertical displacement of the DPV grid. Only the qualitative shape of the curves is suitable as a criterion for agreement because the measured results are quantitatively different. The speeds achieved by the DPV-2000 averaged approximately 25% higher than those obtained by LDA. Figure 8 shows the vertical and horizontal developments of the measured particle velocities through the center points of the measurement grids when they were aligned with each other. The two LDA results show good reproducibility. Qualitatively, the LDA and DPV curves are fairly in good agreement. The quantitative differences between the results of the two methods will be discussed in the following.

Fig. 8
figure 8

Vertical and horizontal courses of the LDA and DPV-2000 particle speeds through the measurement grid center points for powder F

The two diagnostic methods LDA and DPV-2000 generally have limited capabilities for detecting small particles. According to Mie theory, the intensity of the light scattered by the particles depends among other parameters on the particle size. Thus, the sensitivity of the LDA detector is determined by the smallest measurable particle size. The DPV-2000 measurements are based on the particle thermal radiation. The radiated power is determined by the particle temperature, by its surface area, and by the emissivity of the material. This means that for a given material, there is also a minimum detectable particle size, which is additionally dependent on its temperature. Assuming a particle temperature of 2000°C, the manufacturer of the DPV-2000 indicates that the detectivity based on the minimum measurable energy ranges down to particle diameters of 14 μm for an emissivity of 0.2, and 8 μm if the emissivity is 0.8. At temperatures higher than 2000°C, even smaller particles are captured. Unfortunately, no precise data are available on the emissivity of molten YSZ. The acceleration of the particles in the plasma plume depends on their masses and thus on the particle diameters. Larger particles tend to have a smaller velocity. The quantitative differences between the velocities measured by LDA and DPV-2000 may be caused by the different detectivities of the two methods. Depending on the particle size, unequal fractions may be captured with different mean velocities. This is investigated in the following.

First, the DPV-2000 results were examined in more detail. As explained above, the system must be calibrated for diameter measurements. This was done using the particle size distribution determined by laser diffraction (see Fig. 5). The calibration factor f diam was set in such a way that the mean diameter d50 measured with DPV-2000 agreed with the result of the laser diffraction measurement of the powder feedstock. It should be noted that the feedstock was a spray-dried powder, i.e., many particles were hollow. When they are molten in flight, the diameters of the liquid droplets are certainly smaller. However, since we do not know this effect quantitatively, we could not consider it while calibrating the DPV. Furthermore, the laser diffraction measurement delivers volumetrically weighted frequencies, which must also be accounted for. Thus, the DPV-2000 was operated in the volumetric mode during the calibration procedure. The smallest diameters detected by the DPV-2000 were approximately 6 μm. These particles showed an average temperature of 3055°C, which is considerably higher than the overall mean particle temperature of 2905°C. As smaller particles were obviously not detected by the DPV-2000 in the investigated case, particles smaller than 6 μm were excluded from the evaluation of the laser diffraction data. This fraction is equivalent to 5.3% of the overall particle volume. Figure 9 shows a comparison of the two cumulated frequency curves. As a result of the DPV-2000 calibration, they intersect at the mean diameter d 50. As the curves agree well, it can be concluded that the diameters measured by DPV-2000 are reliable down to a specific detection limit, which was 6 μm in this case.

Fig. 9
figure 9

Volumetrically weighted cumulated frequencies of particle diameters measured by laser diffraction and by DPV-2000 for powder F

Unfortunately, no quantitative information is available on the detectivity of the applied LDA system. To verify the assumption that a lower detection limit also exists for LDA and that it affects the representativeness of this method, cumulative frequencies and density distributions for the particle velocities of the LDA 1 and LDA 2 measurements as well as for those obtained with DPV-2000 were compared at the center point of the measurement grid. There is a large difference in the mean values between LDA and DPV results, as mentioned above. By evaluating the DPV-2000 results, mean particle velocities were calculated not only for the totality of the detected particles but also for fractions above specific particle diameters. Considering particles with diameters greater or equal to 32 μm only, a mean velocity of 214 m s−1 was obtained. As smaller and faster particles were excluded, this average velocity was smaller than the mean velocity for the totality of particles. This value corresponds to the LDA results (215 m s−1 for LDA 1 measurement and 213 m s−1 for LDA 2 measurement). The associated DPV cumulative frequencies and density distribution for this particle fraction are also included in Figure 10. They show very good agreement with the measurement results of LDA 1 and LDA 2. Thus, presuming that no systematic errors occurred, it can be concluded that in the investigated case, the lower detectivity limit of the LDA was reached at a particle diameter of 32 μm.

Fig. 10
figure 10

Comparison of cumulative frequencies (top) and density distributions (bottom) for particle velocities measured using LDA and those obtained by DPV-2000 for powder F

To confirm the assumption of a lower detection limit in LDA, a second set of two LDA measurements and one DPV measurement was performed with the coarser powder C. As before, DPV diameter measurements were calibrated based on the particle size distribution determined by laser diffraction. Figure 11 compares LDA and DPV results. As expected for this coarser powder, the curves are closer to each other because a smaller fraction of the powder particles was subjected to the detection limit of LDA compared with the finer powder F. The best agreement was found when only particles with a diameter ≥28 μm were considered in the DPV evaluation. This value is quite close to the threshold of 32 μm, which was found for the finer powder F.

Fig. 11
figure 11

Comparison of cumulative frequencies (top) and density distributions (bottom) for particle velocities measured using LDA and those obtained by DPV-2000 for powder C

A lower detectability threshold for particle diameters of 28-32 μm in LDA is therefore consistent with the intensity scattered by the particles (and detected by LDA). The intensity of the signal depends on the detection angular direction and the object size (particle diameter). Assuming spherical particles and using Mie theory (Ref 21), this angular function can be determined. Figure 12 shows the calculation results for the Mie scattering of nonpolarized light with a wavelength of 532 nm at a single spherical particle of diameter d, carried out using the code developed in Ref 26 and 27. It is obvious that the scattered intensities at YSZ particles with diameters less than 35 μm are at least two orders of magnitude lower than for larger particles. Thus, the conclusion that LDA has a lower detection limit is justified.

Fig. 12
figure 12

Scattered intensity of a single spherical particle with diameter d for an incident wavelength of 532 nm and a refractive index of 2.15 (YSZ, no absorption)

Conclusion

A lower detectability threshold for particle diameters of 28-32 μm in LDA was found in comparative measurements versus the DPV-2000 diagnostic system. The calculation of the Mie-scattering confirms that this lower bound for LDA particle size measurement is consistent with the scattered intensities (and detected by LDA). At values greater than this threshold, the results obtained by LDA and DPV particle diagnostic systems are in very good agreement. This is of particular importance, since it could be suggested that the LDA technique for plasma spraying is generally limited by the experimental environment (e.g., plasma fluctuations, huge temperature, and flow velocity gradients).