Introduction

State of the art of soil nutrient analysis

Monitoring soil nutrients is an essential task for sustainable agriculture (Bindraban et al. 2000; Roberts et al. 2013). In this respect, it is important to know if the soil gets back what has been taken from it during crop growth and if the procedures for soil fertilisation do not exceed the plant’s demand. With this information, environmental hazards may be reduced and expenses optimised.

The availability of soil macronutrients N, P and K in the ionic form of NO3, NH4, PO4 and K is of special interest, as these ions are directly available for plant roots. Amongst them, NO3 is of central importance for plant physiology (see Ho and Tsay 2010). In standard laboratory tests, these ions are extracted by shaking soil samples with an extraction solution and the ion concentrations in the extracts are typically measured with Inductively Coupled Plasma Atomic Emission Spectroscopy (ICP-AES) or Atomic Absorption Spectroscopy (AAS) after a filtration procedure. The cost of this analysis is high, the sample treatment is complicated (e.g. soil samples have to be stored at low temperatures (4 °C) for N concentration analysis) and results are obtained only after a delay, especially in regions with low density of laboratory infrastructure. Devices for fast on-site analysis of soil nutrients are therefore under intensive research.

Many researchers have tried simplified extraction procedures, which allow on-site measurements in soil sample extracts and subsequent measurement with portable devices. Ion selective sensors [ion selective electrodes (ISE) or ion selective field effect transistors (ISFET)] were used by Artigas et al. (2001), Kim et al. (2007) and Shaw et al. (2013). Colorimetric test strips with optical read out devices have been frequently investigated (Scholefield and Titchen 1995; Wetselaar et al. 1998). These techniques have to be tuned to the analyte ions by the selection of suitable ion selective membranes or test strips. Hence the simultaneous measurement of various nutrients is complicated or demands the use of several instruments.

Fertility related parameters of soil samples have been investigated with reflectance measurements on soil samples in the visible, near and mid-infrared ranges of the electromagnetic spectrum [soil reflectance spectroscopy (SRS), for literature overviews see e.g. Schirrmann et al. (2013), An et al. (2014), Vohland et al. (2014)]. Together with site-specific multivariate calibrations (see e.g. Shi et al. (2012) and Viscarra Rossel and Behrens (2010) and cited papers therein) remarkable results were achieved e.g. in measurements of soil organic matter (SOM) and total nitrogen content (Ntot). The short term availability of N, P and K was studied with SRS as well but especially for NO3 concentrations, a key parameter for short term nutrient availability, no clear correlation with standard tests was found [R 2 = 0.5 by Shibusawa (2003) and R 2 = 0.0 by Viscarra Rossel et al.(2006)].

Several mobile units for testing of soil fertility parameters from a field vehicle have been investigated (on-the-go soil sensor systems). Soil samplers on a tractor allowing on-the-go NO3 measurements with ISE or ISFET sensors were presented by Birrell and Hummel (1997, 2001), Sibley et al. (2010). Vehicles using SRS can directly scan soil parameters without the need for a sampling mechanism. A system presented by Christy et al. (2006) was marketed by Veris Technologies, USA. The technique showed promising performance in measurements of fertility related parameters like SOM and Ntot (Kodaira and Shibusawa 2013; Schirrmann et al. 2013). Kodaira and Shibusawa (2013) evaluated the correlation of NO3 concentrations predicted by SRS with laboratory measurements but found a low correlation with R 2 = 0.5.

In summary, a technology for sensing ionic forms of soil macronutrients N, P and K with one portable device is still missing. The present study expands the techniques of mobile ion concentration measurements in soil sample extracts with recent research in the fields of Microfluidics. This research area aims at the integration of laboratory procedures with miniaturised systems (lab-on-a-chip (LOC) systems or micro total analysis systems (µTAS)) with functionalities like fluid handling (mixing, separation, pre-concentration), activation of micro-reactions, chemical analysis and many more. The central interest lies in miniaturised portable and user-friendly devices for environmental analysis, chemistry, medical technology and many more applications.

Objectives

A LOC device capable of measuring the most important plant nutrient ions in soil sample extracts is presented. The paper focuses on the following objectives:

  1. 1.

    Demonstration of the sensitivity of the LOC device to macronutrient ions NO3, NH4, K and PO4.

  2. 2.

    Characterization in field conditions with water-extracted soil samples including comparisons to standard laboratory measurements.

  3. 3.

    Discussion about the potential of the sensor for measurements of additional ions (Cl, SO4, Ca and Na).

  4. 4.

    Discussion of needed advancements of the sensor performance as well as of the soil nutrient extraction procedures.

Materials and methods

Basics of capillary electrophoresis

The sensor presented uses the principle of capillary electrophoresis (CE) (Ewing et al. 1989; Kuhr 1990; Roth et al. 1991) to separate different ion species of a liquid sample by an electric field and to move them past a detector one after another. CE is performed in a buffer-filled capillary to which the sample is introduced from one end, so that a few per cent of the capillary length are filled. With an electric voltage between the ends of the capillary, the sample ions migrate with a speed dependant on their size and charge. This causes the separation of the ion species into plugs which independently pass an ion detector at the opposite capillary end. Each ion plug causes a peak in the recording of the detector signal (electropherogram) and the peak area scales linearly with the ion concentration (at low concentrations where the ion activity equals the ion concentration).

CE was originally setup with tubular capillaries with typical length of several hundred mm. The long capillaries allow good separation quality in numerous analytical applications. In cases where also short capillaries (length of a few tens of mm) yield separation of all analytes (e.g. analysis of small inorganic ions), the integration of the setup to a microfluidic chip is an attractive alternative [microchip capillary electrophoresis MCE, see e.g. Fercher et al. (2010a), Laugere et al. (2003), Tanyanyiwa and Hauser (2002)]. This allows strongly miniaturized setups especially for mobile devices [see e.g. Becker et al. (2008), Gärtner et al. (2009), Kubán et al. (2007), Puchberger-Enengl et al. (2013), Smolka et al. (2013)].

Setups of CE for the analysis of soil nutrients have been presented by Howald et al. (1995) and O’Flaherty et al. (2000). Due to the set-up with classic capillaries, the devices are bulky and not suitable for untrained users. Studies of soil nutrient measurements based on MCE devices have not been presented so far.

CE and MCE ion detectors are often based on a conductivity measurement [for general reviews see e.g. Guijt-van Duijn et al. (2004), Swinney and Bornhop (2000), Vandaveer et al. (2004)]. As the conductivity is influenced by all charged sample components, this principle is practically universal for ion detection (Coltro et al. 2012). A frequently used conductivity measurement is built with two electrodes which are separated from the liquid in the analysis capillary by an insulating material (the capillary wall in the case of CE or an insulation layer in MCE). One electrode is excited with a high frequency (HF) voltage signal. The second electrode is connected over a sensitive current measurement device to the ground potential of the HF source. This setup generates and measures a local capacitively coupled current in the liquid of the capillary. The measured signal varies when passing ions change the conductivity near the electrodes. This technique has been studied by many researchers on microchips and is typically called capacitively coupled contactless conductivity detection or measurement (C4D or C4M) (Fercher et al. 2010b; Fracassi da Silva and do Lago 1998; Kubáň and Hauser 2009; Zemann et al. 1998).

The research on the device presented was focused on the set-up of a stable operating mobile device. In practice, such a mobile device will not be kept aligned horizontal and thus pressure differences will occur between the liquid containers used. This effect complicates the liquid analysis, which demands well-defined injection of nanoliters of the sample (Karlinsey 2012). Stable conditions of the liquids in the measurement system have been achieved by the integration of a nano-porous filter at the sample inlet of the chip, which drastically dampens pressure driven liquid flow but allows penetration of ions driven by electric fields. Furthermore, the presented device is simpler than other designs as only three microfluidic connects are used and no mechanical pumping of the sample is required.

Device fabrication

The central part of the nutrient sensor is the MCE chip, which mainly consists of a micro-channel with the C4D detection electrodes at one end. The channel has one inlet to inject a buffer solution and one inlet for the sample injection (sample injection point). Designs of the microchip and prototype chips are shown in Fig. 1. Each chip is 16 mm × 26 mm. In all measurements in this paper, the channel was 44 mm long (distance between sample injection point and detection electrodes), 100 µm wide and 55 µm high.

Fig. 1
figure 1

Two chip layouts have been developed, one with a 44 mm long channel (a), one with a 64 mm long channel (b). The design (a) was used for all measurements in this paper. Unit of three fully processed chips with mounted sample reservoirs (c)

The chips were fabricated on a rectangular glass wafer (100 mm × 100 mm). In total, 18 chips were realized per wafer. The fabrication steps are shown in Fig. 2. Briefly, the measurement electrodes are structured on the glass substrate by a standard lift-off process. After application of a 17 µm dry film photoresist (Ordyl, Elga Europe, Italy) as an insulation layer, the channel is patterned in another 55 µm thick resist layer by photolithography. The channel is sealed by a cover sheet containing a nano-porous membrane. Further details are given in Smolka et al. (2013).

Fig. 2
figure 2

The chip production process in cross-section along the channel: a Forming of the Cr–Au electrodes by evaporation and lift off process. b Hot roll lamination of 17 µm Ordyl dry film resist, flood exposure with ultraviolet light (UV) and deposition of 55 µm Ordyl dry film resist. c Generation of the channel structure by selective UV exposure and development of the photosensitive dry film resist Ordyl. d Processing of the polymethylmethacrylate (PMMA) cover sheet by computerized numerical control (CNC) drilling. e Bonding of a nano-porous polycarbonate track etched (PCTE) membrane. f Bonding of the cover sheet to the chip by hot roll lamination. g Processing of the protection layer by CNC drilling. h Bonding of the protection layer by hot roll lamination

Measurement set-up and injection procedure

Figure 3 depicts the schematic of the chip in the measurement setup. First, 0.16 ml of the sample is inserted into the sample reservoir with a manual pipette. The buffer inlets of the chip are connected to tubes, each leading into a bottle. One of these bottles (pressure bottle) can be switched to 100 kPa over-pressure to pump the buffer through the chip into the second bottle (collection bottle). The over-pressure is generated with an air-filled syringe of a syringe pump (neMESYS, cetoni GmbH, Germany). This procedure is used to fill the chip with buffer and to flush it after each measurement.

Fig. 3
figure 3

Schematic of the sample injection: a high potential difference U inj is applied between pressure bottle and sample reservoir. This causes sample ions to pass the membrane and to enter the separation channel. In the case depicted, the voltage pattern is suited for anion injection

A Pt electrode dips in each bottle and into the sample reservoir. By applying electric voltages between the electrodes, the electrophoretic effect transports the sample ions into the chip and through the micro-channel.

Without electric voltage, the nano-porous membrane blocks the sample flow into the micro-channel. This prevents self-driven injection by pressure differences due to unequal levels of the liquid in the reservoirs or due to diffusion. In contrast, a high injection voltage (U inj) between sample reservoir and the pressure bottle causes an electrophoretic motion of sample ions through the pores of the membrane into the micro-channel. After switching off of the injection voltage, a sample plug has entered the chip. The duration t inj of U inj defines the length of the sample plug in the microfluidic channel. This process is visualised in Fig. 3 with a voltage U inj (typically −2000 V) between sample reservoir and pressure bottle.

Varying the plug length allows adjusting the sensor properties: a prolonged injection time increases the amount of injected sample and therefore increases the sensitivity of the sensor. However, excessive sample loading leads to overlapping peaks. Thus the injection time has to be suited to the investigated chemical compounds.

After sample injection, the high voltage U sep between the two bottles is applied. To suppress further flow of sample ions into the micro-channel, the voltage in the sample reservoir is adjusted to get a slight current flowing back to the sample electrode. This means that all remaining sample ions are drawn back towards the sample electrode, as is the case for a voltage of −1600 V in the sample reservoir and −2000 V in the collection bottle.

The applied voltage pattern causes the injected ion plug to migrate towards the detection electrodes. During the motion, the individual ion species split according to their electrophoretic mobilities and ion packages pass the detector independently. The detector signal is recorded (electropherogram) and the peak areas are evaluated with an algorithm for baseline and peak search and peak area integration (implemented in OriginLab, OriginLab Corporation, USA).

The voltage patterns described are suited for the injection and separation of anions. For cation measurements, all voltages have to be reversed. A summary of the applied voltages and switching times is given in Table 1.

Table 1 Steps of the measurement procedure for anion and cation analysis (U inj and U sep are defined in Fig. 3)

Reagents and samples

Chemicals were purchased from Sigma-Aldrich GmbH, Austria. All chemical solutions were prepared with de-ionised water from a Millipore system (Hydrolab, Poland, ion exchange resin H+/OH, electric conductivity ≤0.2 µS/cm). The buffer consisted of 30 mmol/l dl-histidine (His), 30 mmol/l 2-(N-morpholino)ethanesulfonic acid (Mes), 4 mmol/l crown-6-ether and 0.1 % methyl cellulose at pH of 6.5.

All calibration samples were prepared in 30 mmol/l His/Mes solution from 200 mmol/l stock solutions of KNO3 (NH4)H2PO4, CaCl2, and Na2SO4. All filtered soil extracts were mixed with 200 mmol/l His/Mes stock solution in 5.55–1 volume ratio, so that the same concentration of 30 mmol/l His/Mes was reached. This high background concentration of 30 mmol/l His/Mes increases the sample conductivities so that the effect of varying sample composition on the injected sample amount is reduced [effect of field amplified sample stacking, see e.g. Bharadwaj and Santiago (2005)]. Further on, the addition of His/Mes buffer stabilises the sample pH around 6.5 where PO4 is mainly present in the single charged form H2PO4 . In the following discussions of electrophoresis experiments, PO4 and H2PO4 are therefore used synonymously.

On-site soil nutrient extraction and reference measurements

The sensor was tested with a soil nutrient extraction with de-ionised water (DIW), the concentrations of the four macro-nutrient ions in the DIW extracts are named \( C_{{{\text{NO}}_{3} }} \), \( C_{{{\text{NH}}_{4} }} \), C K, \( C_{{{\text{PO}}_{4} }} \). DIW extraction was chosen for this first sensor test, as no additional ions are added to the extracted sample. Cross-sensitivities to high background ion concentrations are thus avoided. This procedure is expected to be adequate for on-site extraction of NO3 as good correlations to KCl extraction were shown in a measurement series by Shaw et al. (2013). In contrast, the extracted amount of the macronutrients NH4 +, K+ and PO4 does typically not correlate to extracted amount with standard solutions like CaCl2 for NH4 +, Ca acetate lactate (CAL) for K+ and PO4. DIW extraction is therefore yielding the water-extractable amount of these three ions.

14 soil samples were taken from experimental fields of Warmia and Mazury University in Olsztyn, Poland (53°46′N, 20°28′E). The samples represented four different soil types: Haplic Cambisol, Stagnic Luvisol, Gleyic Phaeozem and Hortic Anthrosol. At each sampling point, three partial soil samples were taken using an Edelman Auger from the topsoil down to a depth of 300 mm. These partial samples were mixed to obtain 200 g of fresh mass. The mixed samples were ground with pestle and mortar and sieved through a 1 mm mesh to discard organic debris and large particles. For the on-site extraction, 15 g of the sieved soil were dosed with a portable scale.

The remaining sieved soil was frozen and sent to a soil laboratory for extraction and analysis. The values of \( C_{{{\text{NH}}_{4} }} \) and \( C_{{{\text{PO}}_{4} }} \) in the samples obtained were too low to be quantified by the sensor (see results in Sect. 3). Thus the laboratory procedure is only described for NO3 and K+. Other ion concentrations like SO4 2−, Ca2+ and Na+ were not quantitatively considered in this first sensor test.

The following laboratory extraction and analysis methods were used:

  • NO3 was extracted in un-buffered 1 % solution of K2SO4 and the ion concentrations were measured by colorimetry.

  • K+ was extracted with Egner–Riehm’s method [see e.g. Oreshkin (1980)] in 0.1 mol/l calcium lactate (buffered at pH 3.6) and the ion concentrations were measured by atomic emission spectroscopy (AES).

For the on-site extraction, each 15 g soil sample was mixed with 30 ml of de-ionized water in a 50 ml plastic container (1:2 ratio of soil to water) and shaken for 15 min with an orbital mini shaker (rotation frequency 1 Hz). The suspension was filtered through 0.22 µm syringe filters (combined glass and polymer filter from Carl Roth GmbH, Austria). 11 ml of filtered extract was produced for each sample. 1 ml was taken for the sensor measurements. 0.18 ml of concentrated His/Mes buffer (200 mmol/l) was added to adjust the concentration of 30 mmol/l His/Mes buffer in the sample. The remaining 10 ml of the filtered soil sample extract were frozen and sent to a chemistry laboratory for measurements of NO3 , NH4 +, K+ and PO4 ion concentrations by inductively coupled plasma atomic emission spectroscopy.

Results and discussion

Measurement parameters

Solutions containing Cl, NO3 , SO4 2−, H2PO4 , NH4 +, Na+, Ca2+, and K+ were used for the sensor calibration. The concentrations of NO3 , NH4 +, H2PO4 and K+ were varied in the range of 5–500 µmol/l. The other ions were added for testing the complete separation of the typical soil sample matrix. The maximum injection time that allowed separation of the investigated ions was 0.6 s. The sensitivity of the H2PO4 measurement was only half the values for NO3 , K+ and NH4 +. This can be explained by the comparably low electrophoretic mobility of H2PO4 ions, which causes lower signal response of the detector.

The sensitivity of the H2PO4 measurement can be increased with a longer injection time. With t inj of 1.2 s, it reaches a similar value to NO3 , K+ and NH4 + measurements with 0.6 s injection time. However, the NO3 peak is covered by the Cl and SO4 2− peaks and can no longer be resolved (see Fig. 4). Thus this sensitive H2PO4 measurement has to be done in a separate run.

Fig. 4
figure 4

Electropherograms of 50 µmol/l Cl, 100 µmol/l NO3 , 100 µmol/l SO4 2− and 50 µmol/l H2PO4 with two different injection times

Table 1 shows all steps of a nutrient measurement including the optional measurement of H2PO4 with increased sensitivity. Prior to each measurement, the chip is flushed with fresh buffer by applying 100 kPa over-pressure in the pressure bottle for 15 s. Pressure differences are equilibrated during a delay time of 15 s before the measurement begins with the injection of sample anions. The flushing procedure is automatically started again after each ion separation. Each measurement is repeated three times for statistical evaluations.

Sensor calibration

Each measurement day started with sensor calibration using the standard solutions. The standard concentrations of NO3 , NH4 +, K+ and H2PO4 were 5, 10, 50, 100, 200, 500 µmol/l ± 5 %. Several calibration solutions were measured again throughout the day to characterize the reproducibility of the calibration measurements. All these calibration measurements were considered in the subsequent evaluations (see Fig. 5). A new chip was used each day to avoid contamination by bio fouling.

Fig. 5
figure 5

Calibration curves for the macro-nutrients NO3 (a), NH4 + (b), K+ (c), H2PO4 (d) measured on two different days

The peak areas in each electropherogram were calculated and basic sensor parameters were evaluated for each analyte ion (results in Table 2):

Table 2 Parameters from the calibration measurements of Fig. 5: calibration constant, coefficient of determination (R2), limit of detection (LOD), limit of quantification (LOQ)
  • Calibration constant slope of the trend line in the graph of ion concentration versus peak area.

  • Average noise level standard deviation of the detector signal from the baseline in the time before the peak arrival (10–25 s).

  • Limit of detection (LOD) ion concentration which generates a peak three times higher than the average noise level; lowest concentration which can be detected.

  • Limit of quantification (LOQ) ion concentration which generates a peak ten times higher than the average noise level; lowest concentration which can be quantitatively evaluated.

The calibration curves in Fig. 5 show the linear response of the sensor to all four macro-nutrient concentrations. The calibration curves of H2PO4 show stronger variation than the curves of the other three ions. This might be caused by the lower and broader peaks of H2PO4 which increase the influence of noise on the peak area evaluation, but more tests are necessary for statistical statements.

Soil nutrient analysis

Typical measurements of DIW soil sample extracts are presented in Fig. 6. The anion signal contains peaks of NO3 and SO4 2− (Cl concentration was below LOD in the chosen example). H2PO4 values were below LOQ in all 14 soil samples, thus no peak signals are shown here. The corresponding cation curve contains peaks of NH4 +, K+, Ca2+ and Na+ (further peaks were not identified). The ion concentrations of NO3 , NH4 + and K+ can be calculated by multiplying the peak area values with the calibration constants of Sect. 3.2. SO4 2−, Ca2+ and Na+ concentrations were not evaluated.

Fig. 6
figure 6

Typical electropherograms for anion (a) and cation (b) measurement (before baseline correction)

The macro-nutrient measurements yielded the following result.

NO3 measurement

All sensor values of \( C_{{{\text{NO}}_{3} }} \) were above LOQ so that a comparison to the laboratory references can be done.

Sensor versus laboratory measurements of on-site soil sample extracts: Sensor and laboratory values of \( C_{{{\text{NO}}_{3} }} \) measured in the on-site soil sample extracts are compared in Fig. 7a. The strong correlation with R 2 = 0.985 demonstrates the potential of the sensor for on-site NO3 measurements. This result resembles the findings obtained with an ISE [R 2 = 0.968–0.989 by Shaw et al. (2013), R 2 = 0.98 by Kim et al. (2007)], an ISFET [R 2 > 0.9 by Birrell and Hummel (1997)] and colour reaction strip sensors [R 2 = 0.96–0.97 by Wetselaar et al. (1998)].

Fig. 7
figure 7

a Fit of \( C_{{{\text{NO}}_{3} }} \) data from sensor and laboratory measured in the same on-site soil sample extract. b Fit of \( C_{{{\text{NO}}_{3} }} \) data from sensor after on-site extraction and laboratory analysis with standard extraction procedure

On-site versus laboratory extraction: The sensor data of the on-site soil sample extracts was compared to laboratory data after the standard soil extraction procedure in Fig. 7b. The strong correlation with R 2 = 0.950 confirms that the on-site soil nutrient extraction is an alternative to the standard laboratory extraction procedure in agreement with Shaw et al. (2013).

NH4 + measurement

The \( C_{{{\text{NH}}_{4} }} \) values were below LOQ except in two cases so that no comparison between laboratory and sensor measurements can be done. Anyhow, the average percentage of \( C_{{{\text{NH}}_{4} }} \) of the mineral nitrogen concentration (sum of \( C_{{{\text{NO}}_{3} }} \) and \( C_{{{\text{NH}}_{4} }} \)) was only 6.4 %. NH4 + was therefore a minor source of plant available N. For analysis of the sensor performance in \( C_{{{\text{NH}}_{4} }} \) measurements, NH4 + rich soils will have to be analysed in future studies.

K+ measurement

All sensor values of C K were above LOQ.

Sensor versus laboratory measurements of on-site soil sample extracts: Sensor and laboratory values of C K measured in the on-site soil sample extracts are compared in Fig. 8a. The strong correlation with R 2 = 0.985 demonstrates the potential of the sensor for on-site measurements of water-extractable K.

Fig. 8
figure 8

a Sensor data verification for K K sensor data versus laboratory data, measured with the same soil sample extract obtained by on-site extraction. b On-site extraction verification K concentrations measured with the sensor after on-site extraction versus K concentration from standard laboratory extraction and laboratory analysis

On-site vs laboratory extraction: The water-extractable K+ concentrations measured with the sensor did not correlate with K+ concentrations obtained with the standard laboratory extraction (see Fig. 8b). The reason is the binding of K+ ions to soil particles, which demands far stronger extraction solutions like CAL in future studies.

PO4 measurements

The extracted amount of PO4 was not sufficient for quantification with the sensor as all \( C_{{{\text{PO}}_{4} }} \) values were below LOQ. Thus, the use of stronger PO4 extractants (e.g. CAL) and a sensitivity increase has to be achieved in future studies.

Additional ions SO4 2−, Na+ and Ca2+

The electropherograms of Fig. 6 contain peak signals of SO4 2−, Na+, Ca2+ ions which would allow quantification of these soil components as well. Although no quantitative analysis was done in this study, this highlights the potential of the sensor for multi-ion soil analysis.

Conclusions

The application of a micro-chip capillary electrophoresis sensor device for soil nutrient analysis was demonstrated for the first time. A central advantage of this sensor is the general sensitivity to ions in liquids. The device is thus especially sensitive to the four plant macro-nutrient ions NO3, NH4, K and PO4. A strong linearity of the sensor calibration for these four ions was found. Although the measurement time was high compared to on-the-go sensors from the literature, the sensor can detect the most important plant nutrients in their short term available form. Nutrient measurements near the field with simplified extraction would strongly simplify the sample logistics (no cooled transport) and would reduce analysis costs in comparison to standard laboratory analysis. With these characteristics, the presented instrument could become a candidate for a mobile sensor for plant available macro-nutrients.

The sensor was tested in on-site measurements on samples of different soil type with water extraction. This extraction was chosen for the first sensor test as no ions that can interfere with the sensor reading are added to the samples.

From the measurements, it is concluded that:

  1. (1)

    The sensor resolved the entire NO3 concentration range of the analysed DIW extracts; the values were strongly correlated to laboratory data measured in the same extracts. Also a plot of the sensor data versus laboratory data obtained with a standard extraction procedure yielded a strong correlation. This demonstrates the potential of the sensor for on-site NO3 measurements and of the on-site extraction procedure for NO3 extraction. These findings have to be tested with larger statistics in future.

  2. (2)

    NH4 concentrations were mainly too low to be quantified by the sensor. They amounted on average to 6.4 % of the mineral N concentration (sum of NO3 and NH4 concentrations). Thus NH4 was only a minor source of plant available nitrogen and consequently of minor importance for soil fertility estimations. Measurements in NH4 rich soils have to be performed in the future for a characterisation of the sensor in the analysis of this ion.

  3. (3)

    Water soluble K amounts measured with the sensor correlated strongly with laboratory measurements of the same extracts. The sensor is therefore useful for on-site measurement of water-extractable K. In contrast, it was confirmed, that the sensor data after DIW on-site extraction did not correspond to data after standard laboratory extraction.

  4. (4)

    In PO4 measurements, the water-extracted amount of all samples was too low to be quantified by the sensor. Future investigations must therefore focus on the use of stronger extraction solutions and sensitivity enhancements of the sensor.

  5. (5)

    The electropherograms of DIW soil sample extracts resolved peak signals corresponding to SO4, Na, Ca ions. Future studies, focusing on the evaluation of these parameters, could make the device a versatile multi-ion analysis tool.

In summary, future studies have to focus on tests of the sensor with strong extraction solutions for NH4, K and PO4. The major challenge will be the separation of nutrients from the high concentrated background of extraction ions. Further measurements with a wider nutrient concentration range (e.g. due to variations between spring and autumn) have to prove the sensor functionality with higher statistical significance.