Introduction

Efficiency calibration of a gamma ray detector is carried out using a set of standards emitting gamma rays over the energy range of 50–1500 keV. Full energy peak (FEP) efficiency calibration is usually done using the mono-energetic gamma ray sources e.g. 241Am, 109Cd, 57Co, 203Hg, 51Cr, 137Cs and 65Zn. The availability of mono-energetic primary standards is a constraint. Moreover, some of these sources are not very long lived e.g. 203Hg (t1/2-46.6 d), 51Cr (t1/2-27.7 d), so they need to be replaced periodically. Multi-gamma ray sources like 152Eu and 133Ba are often preferred for efficiency calibration as they cover a wide energy range and are longer lived. These sources emit gamma rays in cascade and for close sample to detector distances, there may be substantial coincidence summing leading to inaccuracy in efficiency calibration. Therefore, standard multi-gamma sources cannot be used for the gamma spectrometric analysis of low level samples where there is a need to count the samples as close as possible to the detector.

These difficulties can be circumvented by using Monte Carlo simulation for FEP efficiency. It is a powerful tool to simulate the detector response and is applicable to a variety of matrices and source geometries [14]. These codes are particularly useful to get efficiency for cases where experimental determination is impossible or difficult such as samples of non-standard geometry and for large sized samples e.g. environmental samples [5] and waste drums where standards are not easily available. Thus the simulation codes have been increasingly used for efficiency calibration. Also, these simulation methods are decay scheme independent and are consequently free of any coincidence summing and can be used at closest sample to detector distance where multi-gamma ray sources cannot be used. Additionally, these simulation codes can be used to get total efficiency required for coincidence summing correction. However, as reported in the literature, there is often a mismatch between the simulated and experimental efficiencies. This is due to the sensitivity of the Monte Carlo calculations to the detector geometry specifications. The dimensions provided by the manufacturer correspond to the time of assembly of the detection system at room temperature. But there can be changes in the mechanical support of the crystal due to contractions at low temperature leading to changes in the detector configuration [6]. Also, there can be uncertainty in the parameters e.g. dead layer thickness, detector end cap to Ge crystal distance etc. There are reports in literature in which the physical characteristics of the detector have been measured experimentally to minimize these uncertainties [710]. The length of the sensitive region of the detector has been determined by scanning the detector along its axis and accuracy less than 1% has been achieved between experimental and calculated efficiency [7, 8]. The shape of the detector crystal and its position in the housing has been measured by gamma or X-ray radiography [9, 10]. The average dead layer thickness is generally measured by measuring the attenuation of a collimated photon beam. The average dead layer thickness on front face and on the cylindrical side of the crystal has been measured by using the 241Am source [10]. It has been observed that there can be increase in dead layer thickness from 0.35 mm to 1.16 mm after 9 years of operating time [11].

The effect of different detector parameters on the FEP efficiency has been studied in the literature [12, 13]. Detector efficiency has been observed to vary significantly with crystal diameter and its position [12]. It has been observed that it is the detector overall volume and not the detector diameter or length that has a major effect on the FEP efficiency [13]. The influence of dead layer thickness on the detector efficiency response has been studied [4]. It has been observed that increase of dead layer leads to the decrease in the efficiency, not only due to attenuation of gamma rays in the dead layer but also due to the reduction of the active volume of the detector.

Due to the uncertainty in the detector parameters, which may result in uncertainty in efficiency calculation by Monte Carlo method, the general approach has evolved whereby the different detector parameters are optimized to reproduce the experimental point source efficiencies as a function of energy. Subsequently, efficiency for other geometries can be obtained accurately using the optimized parameters. There are other studies in literature where the detector geometry has been optimized by adjusting the detector parameters. Crystal dimensions such as the crystal diameter, length and crystal front dead layer thickness has been adjusted and attained results with a relative deviation within 5–10% [1416]. Tzika et al. [17] optimized the detector geometry by changing the dead layer thickness and could get an overall agreement of 10%. Karamanis et al. [18] also increased Ge–Al distance by 8 mm and the entrance Li dead layer by 600 μm to get results within 5%. In EUROMET exercise [19], it was concluded that most of the simulation codes can be used for routine measurements where uncertainties of 5–10% in efficiencies are acceptable. There is literature where much better accuracy (<2%) has been obtained by optimizing the detector geometry [7, 8, 20]. The objective of this work was to optimize the detector parameters using experimental efficiencies for standard geometries such as point sources. The optimized detector parameters have been used for efficiency transfer to other geometries. Results are discussed in the light of existing observations in the literature.

Experimental

The detector used in the present work is a closed end coaxial p-type DSG HPGe detector. The detector has a 20% relative efficiency and a resolution of 2.1 keV for the 60Co gamma ray at 1332 keV and is used for routine gamma spectrometric analysis in our lab. The schematic diagram of the detector geometry as specified by the manufacturer is shown in Fig. 1. The detector was first calibrated for its efficiency by counting point sources of standard 133Ba and 152Eu at 21.7 cm from the detector end cap. Such a large distance was chosen for efficiency calibration with these multi gamma sources so as to avoid any uncertainty in the efficiency curve due to coincidence summing. The counting of the samples was done for a sufficiently long time so as to keep statistical uncertainty less than a percent. All the spectra were analyzed by PHAST software [21]. The Eu–Ba efficiencies were fitted into a fourth order log–log polynomial curve by non linear least square fitting. These efficiencies were then used to calibrate point sources of 109Cd, 57Co, 203Hg, 51Cr, 137Cs and 65Zn procured from Board of Radiation and Isotope Technology, Mumbai. For calibration, these sources were also counted at d = 21.7 cm. The mono-energetic sources were then used to determine FEP efficiencies (ε γ ) at closer distances (d = 1.7 and 12.6 cm) by using the experimentally determined disintegration rate (dps) given by:

$$ \varepsilon_{\gamma } = {\frac{{cps}}{{{{I}}_{\gamma } {{x\,dps}}}}} $$
(1)

where cps is the count rate at the energy of interest and I γ is the gamma ray emission probabilities taken from Table of Isotopes [22]. The count rates were corrected for the background peak counts in all the cases.

Fig. 1
figure 1

Schematic diagram of the detector geometry supplied by manufacturer

Similarly, 5 ml sources of 109Cd, 57Co, 203Hg, 137Cs and 65Zn were calibrated using standard 5 ml sources of 152Eu and 133Ba in same HPGe detector at d = 10.3 cm. These sources were then used to obtain efficiencies at d = 2.0 cm. Also, 100 ml standard source of 152Eu taken in a glass bottle was counted in the same HPGe detector at d = 19.8 cm and FEP efficiencies were determined.

Monte Carlo simulation

In this work, the version MCNP4c [23] was used to simulate the HPGe detector response. This code tracks the particle from its origin point to the point of its complete absorption. All primary as well as secondary interactions are taken into account. The efficiency was obtained with F8 tally which is a pulse height tally. Mode P was used. The description of the detector geometry was given in detail in the cell and surface cards of the MCNP input file. The detector housing including the absorbing materials (aluminum end cap, germanium dead layer), were all included in the geometry. In each run ~108 particles were sampled to reduce statistical uncertainties.

Results and discussion

Initial MCNP simulations were performed for point source geometry with the dimensions of the detector obtained from manufacturer as given in the second column of Table 1. The MCNP simulated and experimental values of the FEP efficiencies are shown in Fig. 2. The error on experimental efficiencies is also shown in the figure and is obtained by propagating error on peak areas, abundances and efficiency fitting error. The error bars are of the size of the symbol.

Table 1 Detector parameters provided by manufacturer and optimized by MCNP simulation
Fig. 2
figure 2

Experimental and MCNP simulated FEP efficiency for point source geometry using the manufacturer supplied detector geometry at three sample to detector distances, d = 1.7, 12.6 and 21.7 cm

At all sample to detector distances (d = 1.7, 12.6 and 21.7 cm), the MCNP efficiencies were found to be higher than the experimental values indicating the inadequacy of the manufacturer supplied detector dimensions. Similar observations have also been made in literature [7, 10, 12, 24]. A more detailed comparison of the MCNP and experimental efficiencies at d = 1.7, 12.6 and 21.7 cm has been given in Table 2. At a particular distance, the ratios of MCNP to experimental efficiencies were observed to be within 2–4% over the energy range of 88–1115 keV. But, these ratios were found to vary strongly from one distance to other with the MCNP to experimental ratio being higher for the closer distance. It can be seen from the table that the average MCNP to experimental ratio at d = 21.7 cm is 1.14 while it increases to 1.20 at d = 12.6 cm and to 1.46 at d = 1.7 cm. This indicates that the detector parameters have to be adjusted such that the constancy of efficiency over the energy range is not disturbed but the strong dependency on the sample to detector distance is removed. There are a number of geometrical parameters which can be uncertain like detector crystal radius, its length, inner hole radius and length, dead layer thickness, Al end cap thickness, the Al end cap to detector crystal distance. The effect of systematic change in these parameters on the FEP efficiency is discussed in the following.

Table 2 The comparison of experimental and MCNP efficiencies at three sample to detector distances, d = 1.7, 12.6 and 21.7 cm

Detector crystal geometry

Since crystal dimensions have been reported to have major effect on the detector efficiency, the radius of the detector was reduced in steps of 1 mm and detector efficiency was computed by MCNP at d = 1.7 and 21.7 cm. Table 3 gives the comparison of FEP experimental and MCNP efficiency results. The average efficiency ratio has reduced from 1.46 to 1.36 at 1.7 cm and from 1.14 to 1.03 at d = 21.7 cm. The detector radius was further reduced by 1 mm, to see if the ratio further reduces and approaches unity. As given in Table 3, although at d = 1.7 cm, the ratio has reduced to 1.16 but it became less than 1 at d = 21.7 cm. Thus, it can be concluded that the reduction in crystal radius by 1 mm can reproduce the experimental efficiency at long distance, but at shorter distance (= 1.7 cm), the discrepancy still exists. Since the MCNP to experimental ratio should be close to unity at all sample to detector distance, variation of crystal radius alone is not sufficient to match MCNP and experimental efficiency.

Table 3 The ratio of MCNP and experimental efficiencies at two sample to detector distances, d = 1.7 and 21.7 cm

Since the crystal length can also have appreciable effect on efficiency, as a next step, it was reduced by 4.9 mm keeping the inner hole length constant, and as given in Table 3, the average MCNP to experimental ratio at both the distances decreased by about 2–3% only even after a reduction of detector length by 10%. Since the value of length quoted by the manufacturer cannot be erroneous by more than 10%, this parameter was not further changed and its original value was restored. The increment in inner hole radius to 5.75 mm and thereby reduction in the overall detector volume, as given in Table 3, introduced energy dependence which was earlier not present with the manufacturer supplied detector geometry. This was expected as the lower energy gamma rays will be absorbed in a small detector length and so will not be affected by increasing the inner hole radius but some part of the high energy gammas will escape due to increase in the inner hole radius. Therefore, in our case, this is not the parameter to be changed. The detector efficiency also depends upon the dead layer thickness, as low energy photons may be highly attenuated in it. The influence of dead layer thickness on the detector efficiency was seen by increasing the dead layer thickness by 0.4 mm. There was a reduction of only 4–5% in the MCNP to experimental efficiency ratios as given in Table 3 after an increase in dead layer thickness by 67%. From this, it was concluded that, in our case the detector crystal parameters are not the one responsible for such a high efficiency ratio of 1.46 at d = 1.7 cm.

Detector end cap thickness and end cap to crystal distance

Since the effect of increasing the thickness of Al end cap will be similar and less in magnitude to increasing the Ge dead layer thickness, the Al thickness was not changed. The distance between the Al end cap and detector crystal (d alc) was increased by 3 mm and MCNP efficiencies were computed at d = 1.7 and 21.7 cm. Table 4 gives the ratios of MCNP and experimental efficiency values. The MCNP efficiencies at 1.7 cm decreased by about 13% while the efficiencies at d = 21.7 cm reduced by only 2%. Since in our case, there is distance dependence in the MCNP to experimental ratio with the ratio being higher at closer distance, the Al end cap to crystal distance may be the cause of discrepancy between MCNP and experimental efficiency values. This is in accordance with Hardy et al. [8] who observed an uncertainty of ~36% in the Al end cap to crystal distance. The dalc was then systematically changed and the results are presented in Table 4. When the Al end cap to crystal distance is kept as 8 mm, the ratio of efficiencies was further decreased, but still a difference between the efficiency ratios at d = 1.7 and 21.7 cm can be seen. The principle of efficiency transfer says that this difference should not exist, so the Al end cap to crystal distance was further incremented in steps to 1 mm. The results are given in Table 4. It can be seen that as the Al end cap to crystal distance is increased, the ratios of MCNP to experimental efficiencies at d = 1.7 and 21.7 cm come closer and almost coincides at Al end cap to crystal distance of 9 mm and after that again diverges. This indicates that the Al end cap to crystal distance should be taken as 9 mm instead of 3 mm (quoted by the manufacturer) to get a MCNP to experimental efficiency ratio independent of sample to detector distance.

Table 4 The comparison of experimental and MCNP efficiencies at two sample to detector distances, d = 1.7 and 21.7 cm

But, the MCNP efficiencies were still found to be 10% higher than the experimental values at both the sample to detector distances. Further optimization was carried out to bring this factor close to unity. Previously, from Table 3, it was observed that when the detector radius is reduced by 1 mm, the efficiency ratio also reduces by about 7–9% at the two sample to detector distances. This will bring the efficiency ratio close to 1 if the optimized Al end cap to detector crystal distance is used for MCNP calculation. Thus, the radius of the detector crystal was systematically reduced, keeping the Al end cap to crystal distance as 9 mm instead of 3 mm (given by manufacturer) and efficiency ratios at two distances were obtained. Initially, when the radius was reduced by 0.8 mm, the efficiency ratio at the two distances, d = 1.7 and 21.7 cm, reduced to 1.03 and 1.01, respectively (Table 5). This indicates that the ratio is approaching 1 as desired. Therefore, the radius was further reduced systematically in steps of 0.1 mm and the results are given in Table 5. When the radius is reduced by 0.9 mm, the average ratio at d = 1.7 and 21.7 cm was found to be close to unity, indicating that the manufacturer supplied crystal radius should be reduced by 0.9 mm to match the MCNP efficiencies with experimental efficiencies. But still at d = 1.7 cm, a slight systematic bias showing the dependence on energy was observed. The ratio of MCNP and experimental efficiencies were found to be higher at lower energies (1.10 at 88 keV) compared to higher energies (1.00 at 1115 keV). To remove this slight energy dependence, the dead layer thickness was increased by 0.1 mm. These calculations were done using the optimized Al end cap to crystal distance and the optimized detector crystal radius. With this variation, the efficiency ratios were found to be within 5% at all energies and over all sample to detector distances. This experimental and the MCNP efficiencies calculated using this optimized geometry has been shown in Fig. 3. The detector parameters for the optimized geometry have been given in the last column of Table 1.

Table 5 The comparison of experimental and MCNP efficiencies at two sample to detector distances, d = 1.7 and 21.7 cm
Fig. 3
figure 3

Experimental and MCNP simulated FEP efficiency for point source geometry using the optimized detector geometry at three sample to detector distances, d = 1.7, 12.6 and 21.7 cm

Vargas et al. [12] reported the effect of systematic variation of detector parameters on the simulated efficiencies of point and volumetric sources. The study showed significant deviations in the efficiency, depending on the source geometry and the photon energy, could be produced by varying only slightly some of the detector parameters. Of particular importance in the context of the present work, is the variation of efficiencies with the Al end cap to crystal distance for different source geometries. It has been observed [12] that this distance is critical for volumetric sources, but less important for point sources. However, our studies show that there is a strong distance dependence of simulated efficiencies with respect to experimental efficiencies for point sources also when manufacturer supplied detector parameters are used. It was also observed that this can be eliminated by adjusting the Al end cap to crystal distance.

In accordance with the principle of efficiency transfer, this optimized geometry should be valid for all the extended sample geometries. As a check of the optimized geometry, the efficiencies of 5 ml aqueous solutions of 109Cd, 57Co, 203Hg, 137Cs and 65Zn in cylindrical vials at d = 2.0 and 10.3 cm were compared with the MCNP calculated efficiencies. The MCNP efficiencies for these samples geometries were calculated by using the optimized detector geometry and the detailed sample geometry. Table 6 gives the experimental and MCNP efficiencies along with their ratios for 5 ml sources. The two efficiencies differ by about 7% at d = 2.0 cm and by 3% at d = 10.3 cm for this extended geometry. The validity of the optimized geometry was also checked for another extended source of standard 100 ml 152Eu. The experimental efficiencies at d = 19.8 cm for this geometry were compared with the efficiencies calculated by using MCNP. The ratios of the two efficiencies are given in the last column of the Table 6. In this case also, the two efficiencies are within 3–4%. The slight difference in the efficiencies for volumetric samples at closer distance can be due to some inaccuracy in sample geometry such as in its radius, glass thickness, its density and height. This has been also observed by Johnston et al. [6] and Vargas et al. [12] where it was observed that the sample dimensions could have a significant effect on the detector efficiency. However, our results of simulated efficiencies using adjusted detector parameters based on point source efficiencies can be used for volumetric sources also and one can get results within 5%.

Table 6 Ratios of MCNP to experimental efficiencies for 5 ml sources of 109Cd, 57Co, 203Hg, 137Cs and 65Zn and for 100 ml 152Eu source

Conclusion

This methodology is useful in getting an efficiency calibration not only in the cases where standards are not available but also where multi-gamma standards are available but low level counting is required.