Abstract
Samples of high alloy Stainless Steel SS 1 were analyzed by instrumental neutron activation analysis (INAA) at 5 research institutes in 4 European countries within the frame of an Inter-Laboratory Comparison (ILC) exercise aiming at checking if the results were independent of the standardization methods, and if INAA was accurate enough to contribute to certification. The mean element concentration values yielded by INAA were also compared with mean values obtained by atomic emission spectroscopy techniques at other laboratories that took part in the International Proficiency Testing organized by KIMAB Institute of Sweden, producer of the SS 1. The performances of the nuclear and atomic techniques were compared in terms of the z-score values calculated for 11 element concentrations evaluated by the two analytical ways. Finally, consideration is given to the ways in which highest accuracy and precision can be ensured for certifying stainless steels as CRMs.
Article PDF
Similar content being viewed by others
Avoid common mistakes on your manuscript.
References
W. Tian, B. Ni, P. Wang, L. Cao, Y. Zhang, Accred. Qual. Assur., 6 (2001) 488.
W. Tian, B. Ni, P. Wang, L. Cao, Y. Zhang, Accred. Qual. Assur., 7 (2002) 7.
P. Bode, E. A. De Nadai Fernandes, R. R. Greenberg, J. Radioanal. Nucl. Chem., 245 (2000) 109.
I. Gustavsson, B. Larsson, F. Reinholdsson, Proficiency Testing for High Alloy Steel in 2004 — Spectrometric discs, Research Report, SIMR (KIMAB), Stockholm, June 2005.
Standard ISO/IEC Guide 43-1:1997, Proficiency Testing by Interlaboratory Comparisons — Part 1: Development and Operation of Proficiency Testing Schemes.
Standard ISO 13 528:2005, Statistical Methods for Use in Proficiency Testing by Interlaboratory Comparisons, Geneva, Switzerland, 2005.
M. Thompson, S. L. R. Ellison, R. Wood, IUPAC Technical Report, Pure Appl. Chem., 78 (2006) 145.
F. de Corte, The k 0 Standardization Method. A Move to the Optimization of Neutron Activation Analysis, Rijksuniversiteit Gent, Faculteit van de Wetenschappen. Profeschrift voorgeleged tot het verkrijgen van de graad van Geaggreerde voor het Hoger Onderwijs, 1987.
M. Balla, Zs. Molnár, Á. Körös, J. Radioanal. Nucl. Chem., 259 (2004).
N. Etxebarria, P. Robouch, J. Pauwels, S. Pomme, F. Hardeman, Proc. 2nd Intern. k 0-Users Workshop, Ljubljana, Slovenia, 1996, p. 137.
R. Dybczyński, B. Danko, K. Kulisa, H. Polkowska-Motrenko, Z. Samczyński, M. Stepniewski, Z. Szopa, Chem. Anal., 44 (1999) 471.
Em. Cincu, D. Barbos, I. Manea, V. Manu, The advantageous way in getting the lowest uncertainty values of elemental concentration by INAA, Proc. MTAA-11 Intern. Conf., June 15–19, 2004, Guildford, UK, Workshop, J. Radioanal. Nucl. Chem., 271 (2007) 655.
O. Sima, D. Arnold, C. Dovlete, J. Radioanal. Nucl. Chem., 248 (2001) 359.
C. Păunoiu, D. Barbos, Reactor neutron activation analysis using Zr foils, 4th Intern. k 0-Users Workshop, 11–14 September 2005, Funchal, Portugal.
ISO/IEC/ EN 17 025: 2005 Standard, General Requirements for the Competence of Calibration and Testing Laboratories, ISO, Geneva, Switzerland.
NIPNE, Report, presented in the International Workshop ‘IWIRAD 2005’, June 20–21, 2005, Bucharest, Romania, on the website: http://iwirad-2005.nipne.ro
S. J. Parry, Handbook of Neutron Activation Analysis, Viridian Publishing, Surrey, UK, 2003.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Cincu, E., Manea, I., Manu, V. et al. Comparative performance of INAA and other spectroscopy techniques in the elemental analysis of stainless steel materials. J Radioanal Nucl Chem 274, 199–205 (2007). https://doi.org/10.1007/s10967-006-6912-2
Received:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10967-006-6912-2