1 Introduction

Bone graft materials are necessary during orthopedic surgery, such as for critically sized bone defects or in the presence of pseudarthrosis. Autogenous bone grafts are still the gold standard for bone replacement because they have displayed osteoconduction, osteoinduction, and osteogenesis without causing an immunologic response [13]. Additional incisions in healthy skin, however, are needed for that technique [4]. Furthermore, it is difficult to reconstruct bone tissue in large defects using only autogenous bone grafts because of the limited amount of available bone. In such cases, artificial bone substitutes are utilized.

Clinically, hydroxyapatite (HAp) [Ca10(PO4)6(OH)2] has been commonly used as an artificial bone substitute because of its excellent tissue response and osteoconductivity [13, 57]. After implantation in bone defect, however, HAp remains at the implant site for a long time [1, 8]. Such remaining artificial bone substitute, although without immune function, is associated with a risk of infection [9]. Therefore, a resorbable apatite bone substitute has been sought. Carbonate apatite [CO3Ap: Ca10-a(CO3)b(PO4)6-c(OH)2-d] artificial bone substitute [915] is closer to natural bone than HAp because bone apatite also contains carbonate in its apatitic structure [16]. This synthetic CO3Ap bone substitute has the potential to be an ideal bone substitute that is gradually replaced by new bone during a bone-remodeling cycle [9].

A CO3Ap bone substitute can be fabricated by compositional transformation via a dissolution−precipitation reaction using a precursor. To date, a calcite (CaCO3) block [9, 10], a calcium sulfate (CaSO4) block [11, 12], and an α-tricalcium phosphate (α-TCP) block [1315] have been used as precursors for fabricating the CO3Ap block.

Recently, dicalcium phosphate dihydrate [DCPD: CaHPO4·2H2O] was found to be a satisfactory precursor for fabricating the CO3Ap block [17]. The DCPD block can be fabricated based on the setting reaction of DCPD-forming cement. In other words, it is easy to create any shape of DCPD block. DCPD composition includes Ca and PO4 and is suitably soluble for the dissolution–precipitation reaction.

To date, however, there have been no studies regarding the in vivo behavior of CO3Ap block fabricated from DCPD. Also, no comparison study has been performed to evaluate the rate at which the CO3Ap block has been replaced by bone at various bone defect sites. In this study, therefore, CO3Ap block fabricated from DCPD blocks and sintered HAp were implanted in femoral and tibial bone defects of rabbits. The sintered HAp block was employed as a typical, commercially available and non-resorbable bone substitute. The implants were examined using micro-computed tomography (CT) up to 24 weeks after implantation in 19 male rabbits.

2 Materials and methods

2.1 Preparation of cylindrical CO3Ap blocks

Cylindrical CO3Ap blocks (6 mm diameter, 4 mm high) were fabricated via compositional transformation of cylindrical DCPD blocks, as described elsewhere [17]. First, the DCPD block was fabricated based on the setting reaction of DCPD forming cement [1820]. β-TCP [β-Ca3(PO4)2] (Taihei Chemicals, Osaka, Japan) and monocalcium phosphate monohydrate [Ca(H2PO4)2·H2O] (Sigma Aldrich, St. Louis, MO, USA) powders were mixed with methanol so the Ca/P molar ratio would be 1.0 (the molar ratio of DCPD). After the methanol was evaporated, the powder mixture was placed in a mold followed by a sufficient amount of water to set the block.

Second, composition of the DCPD precursor block was changed to CO3Ap using compositional transformation via a dissolution–precipitation reaction. The DCPD block was immersed in Na2CO3 solution (2 mol/L) at 80 °C for 14 days.

As a typical, commercially available and non-resorbable bone substitute, a cylindrical HAp block (6 mm diameter, 4 mm high) was prepared using the following procedure. HAp powder (HAP-200; Taihei Chemicals) was placed in a stainless steel mold and uniaxially pressed by loading 20 MPa using an oil pressure press machine (MT-50HD; NPa System, Saitama, Japan). The obtained, compacted material was then heated in an electronic furnace (SBV-1515D; Motoyama, Osaka, Japan) up to 1300 °C at 5 °C/min and then maintained at this temperature for 6 h, followed by cooling to room temperature inside the furnace.

2.2 Characterization of specimens

The obtained blocks were subjected to powder X-ray diffractometry (XRD) (D8 Advance; Bruker AXS GmbH, Karlsruhe, Germany) [CuKα (λ = 0.1542 nm), 40 kV, 40 mA] during continuous scanning from 10° to 40° at 2θ at a scanning rate of 2°/min. Fourier transform infrared (FT-IR) spectra were measured with an FT-IR spectrometer (FT/IR-6200, JASCO, Tokyo, Japan) using KBr method. The spectral resolution of 4 cm−1 was employed to examine structural changes.

The surface morphology of the obtained specimens was evaluated using a scanning electron microscopy (SEM) (S-3400N; Hitachi High-Technologies, Tokyo, Japan) at 15 kV accelerating voltage after applying a gold–palladium coating with a magnetron sputtering machine (MSP-1S; Vacuum Device Co., Ibaraki, Japan). The specific surface area (SSA) was measured using the Brunauer-Emmett-Teller (BET) N2 adsorption method (Gemini 2370; Micromeritics, Norcross, GA, USA). The porosity was calculated using the bulk density of the specimen (dspecimen) and the theoretical density of HAp (dHAp 3.16 g/cm3) [21], as shown in [Eq. 1].

$${\rm{Porosity}}\left( \% \right){\rm{ = }}{{\rm{d}}_{{\rm{HAp}}}} - \frac{{{{\rm{d}}_{{\rm{specimen}}}}}}{{{{\rm{d}}_{{\rm{HAp}}}}}}{\rm{ \times }}100$$
(1)

Carbonate contents were estimated from the mass% of carbon in the CO3Ap block. A CHN coder (MT-6; Yanako Analytical Instruments, Kyoto, Japan) was used to analyze the mass% of carbon.

2.3 Procedure for the animal experiment

A total of 19 Japanese white male rabbits (Japan SLC Inc., Hamamatsu, Japan) (age range 19−20 weeks; mean body weight 3335 ± 322 g) were studied. The animals were housed in the animal center of our institution and maintained on a standard diet and water. All experiments followed protocols that were approved by the Animal Care and Use Committee (Approval Number A-27-328-0, issued on March 16, 2016). The experiments were conducted in accordance with the Guidelines for Animal Experiments of our institution, Japanese law (No. 105), and notification No. 6 of the government of Japan.

Rabbits were anesthetized by intramuscular injection of ketamine-xylazine (35–10 mg/kg). The rear limb was shaved and then disinfected using isodine. To understand the effects of the location of bony tissue on the resorption of CO3Ap and HAp block specimens, we selected two sites for implantation. The epiphysis of the distal femur and metaphysis of the proximal tibia were employed because of the availability of sufficient bone volume and easy surgical exposure. A medial, longitudinal skin incision was made at each distal femur, and the medial femoral condyle was exposed. A 6.1-mm hole was drilled into the epiphysis of the distal femur, into which the block specimen was inserted. The fascia was then sutured. The next skin incision was in the proximal tibia. A similar 6.1-mm hole was drilled into the metaphysis of the proximal tibia and the block specimen inserted. In all rabbits used, both CO3Ap and HAp blocks were implanted bilaterally. All rabbits were allowed unrestrained movement in their cages after recovery from anesthesia. One rabbit was used for the collecting data for day 0. Day 0 means that samples with surrounding bone were collected just after the implantation (n = 1 in each group). For other implantation periods, the rabbits were euthanized, in batches, at 4, 12, and 24 weeks after implantation, and the epiphyses of the distal femurs and metaphyses of the proximal tibias were harvested (n = 6 in each group).

2.4 Micro-CT

The excised distal femur and proximal tibia with block specimens were scanned using micro-CT (Skyscan 1075 KHS; Skyscan, Kontich, Belgium) (source voltage 60 kV, source current 170 μA, 0.5-mm aluminum filter). The samples were scanned using the high-resolution mode (9 μm voxel resolution). Isotropic slice data were obtained from micro-CT measurements and reconstruction to two-dimensional images. These slice images were compiled and analyzed to render three-dimensional (3D) images using analyzing software. The volume of the residual block specimen was calculated using a quantitative 3D evaluation program that had been installed in an analytical personal computer [22]. Before implantation, the volume of each block specimen was calculated using its diameter and height. The resorption rate (%) was calculated using [Eq. 2].

$$ {\rm{Resorption}}\,{\rm{rate}}\ \left( \% \right) {\rm{ = }}100 \\ - \left( {\frac{{{\rm{volume}}\,{\rm{of}}\,{\rm{residual}}\,{\rm{block}}\,{\rm{specimen}}}}{{{\rm{volume}}\,{\rm{of}}\,{\rm{block}}\,{\rm{specimen}}\,{\rm{before}}\,{\rm{implantation}}}}} \right){\rm{ \times }}100$$
(2)

2.5 Statistical analysis

Resorption rates are expressed as means ± SD. The Mann−Whitney U test was used to assess the difference in the resorption rates between the CO3Ap and HAp blocks. The differences in the resorption rates between the femur group and the tibia group were assessed in the same way. Statistical analyses were performed using JMP Software (version 11.0; SAS Institute, Cary, NC, USA). A probability value of <0.05 was considered to indicate statistical significance.

3 Results

Figure 1 shows typical photographs of (a) a DCPD block used as a precursor for fabricating a CO3Ap block, (b) a CO3Ap block, and (c) an HAp block (used as the control). As shown, the size of the DCPD and CO3Ap cylindrical blocks (6 mm diameter, 4 mm high) were the same, indicating that the CO3Ap block fabricated by compositional transformation via a dissolution–precipitation reaction using DCPD as a precursor maintained its macroscopic structure. The DCPD and CO3Ap blocks had opaque bodies, whereas the HAp block had a relatively translucent body, which is typical for well-sintered ceramic creations.

Fig. 1
figure 1

Typical photographs of a a dicalcium phosphate dehydrate (DCPD) block used as a precursor for fabrication of a carbonate apatite (CO3Ap) block, b CO3Ap block and c sintered hydroxyapatite (HAp) block (used as a control)

Figure 2 shows typical SEM images of (a) a DCPD block, (b) a CO3Ap block, and (c) an HAp block. The DCPD block consists of plate-like crystals, whereas the CO3Ap block consists of small polygon-like crystals. In both cases, the crystals were interlocked, similar to those in set gypsum. This structure is typical for products made via a dissolution−precipitation reaction using a precursor. The different crystal morphologies indicated that compositional transformation via the dissolution−precipitation reaction maintained their macroscopic structure but not their microstructure. In contrast, sintered HAp showed a typical surface for sintered ceramic objects. In other words, grains and grain boundaries were seen on its surface.

Fig. 2
figure 2

Typical scanning electron microscopy (SEM) micrographs of a DCPD block, b CO3Ap block, and c HAp block

Figure 3 shows the powder XRD patterns of (a) DCPD block used as a precursor for CO3Ap block fabrication, (b) CO3Ap block, and (c) HAp blocks. XRD patterns of (d) DCPD and (e) HAp are shown for comparison. XRD analysis demonstrated clearly that DCPD block completely transformed to a pure CO3Ap block without by-products. Peaks of CO3Ap were broad, indicating that CO3Ap with low crystallinity was formed. In contrast, sintered HAp showed sharp peaks, indicating high-crystallinity apatite. In this case as well, all peaks were attributed to HAp. The difference in crystallinity was based on the temperature needed for its fabrication. CO3Ap block was fabricated at 80 °C in aqueous solution, whereas sintered HAp was fabricated at 1300 °C.

Fig. 3
figure 3

Powder X-ray diffractometry (XRD) patterns of a DCPD block, b CO3Ap block, c HAp block. XRD patterns of d DCPD powder and e HAp powder were used for comparison

Figure 4 shows the FT-IR spectra for the (a) DCPD block used as a precursor for CO3Ap block fabrication, (b) CO3Ap blocks, and (c) HAp blocks. The FT-IR spectra of (d) DCPD and (e) HAp are shown for comparison. FT-IR spectra also confirmed a DCPD block transformed to a CO3Ap block. The peaks at 567, 606, 1042, and 1092 cm−1 are assigned to PO4 3− groups [23]; those at 875, 1418, and 1474 cm−1 are assigned to CO3 2− groups [24]; and peaks at 640 cm−1 are assigned to OH groups [23]. PO4 3− groups were observed for all DCPD, CO3Ap, and HAp specimens, whereas CO3 2− groups were observed only for the CO3Ap specimens. OH groups were observed for HAp but not for CO3Ap specimens. The peak positions of CO3 2− groups and absence of an OH group imply that the obtained CO3Ap is similar to an AB-type carbonated apatite that has been reported in our previous paper [17]. In short, CO3 2− ions were substituted at both OH and PO4 3− sites in the apatite crystal lattice and some of CO3 2− ions were adsorped on the surface of CO3Ap crystal [16, 24, 25]. The broad peaks at 1647 cm−1 were relevant to H2O. No peak near 1647 cm−1 was observed from the HAp specimen because it was sintered at 1300 °C.

Fig. 4
figure 4

Fourier transform infrared (FT-IR) spectra of a DCPD block, b CO3Ap block, and c HAp block. FT-IR spectra of d DCPD powder and e HAp powder were used for comparison

Table 1 summarizes the SSA, total porosity, and carbonate contents of the CO3Ap and HAp blocks. The porosity and SSA of the CO3Ap block were significantly higher than those of the HAp block. The CO3Ap block contained approximately 16 wt% CO3 2− groups, whereas the HAp block was free of these groups. CO3Ap block has an SSA of approximately 22 cm2/g, whereas that of the HAp block is very small. In other words, the value could not be measured based on the BET analysis. The porosity of the CO3Ap block was approximately 49%, which is approximately 10 times larger than that of the HAp block. This porosity is consistent with the SEM observation shown in Fig. 2.

Table 1 Specific surface area (SSA), porosity, and carbonate contents of CO3Ap and HAp blocks used in this study

Figure 5 shows the micro-CT images (side view of the cylindrical block) of the CO3Ap blocks (a−d) and HAp blocks (e−h) immediately after implantation (day 0) (a, e) and at 4 (b, f), 12 (c, g), and 24 (d, h) weeks after implantation in the rabbits’ distal femurs. HAp maintained its structure even at 24 weeks after the operation. In other words, there was no sign of resorption in the case of sintered HAp.

Fig. 5
figure 5

Micro-computed tomography (CT) images (side view of the cylindrical block) of CO3Ap blocks (ad) and HAp blocks (eh) immediately after implantation (day 0) (a, e) and at 4 (b, f), 12 (c, g), and 24 (d, h) weeks after implantation in the rabbits’ distal femurs

CO3Ap block showed less opacity than the HAp block throughout the experimental period. In contrast to the HAp block, the size of the CO3Ap block decreased with increasing time after implantation. At 4 weeks after implantation, the CO3Ap block basically maintained its structure, although the edge of the CO3Ap block inside the bone became rounded, indicating that the area was being resorbed and replaced with bone. At 12 weeks after implantation, there was much more resorption and replacement than was seen at 4 weeks. All of the CO3Ap edges had become rounded. In addition, other surfaces were being resorbed and replaced with bone. As a result, the CO3Ap block, which had had a rectangular shape had become more rounded. At 24 weeks after implantation, there was even more resorption of the CO3Ap block, and more of it had been replaced with bone. Resorption of the CO3Ap block and its replacement with bone seemed to be accelerated when compared with that at the initial stage.

Figure 6 summarizes the micro-CT sagittal images of the rabbits’ distal femurs from a top view of the cylindrical block. Gaps were observed between bone and both the HAp and CO3Ap blocks at this stage. At 4 weeks, bone had grown to contact the surface of the CO3Ap, whereas the gap between the HAp block and bone remained, although it had decreased. At 12 and 24 weeks, the CO3Ap block had decreased in size, whereas the size of the HAp remained the same even at 24 weeks.

Fig. 6
figure 6

Micro-CT sagittal images of the rabbits’ distal femurs from a top view of the CO3Ap blocks (ad) and HAp blocks (eh) immediately after implantation (day 0) (a, e) and at 4 (b, f), 12 (c, g), and 24 (d, h) weeks after implantation in rabbits’ distal femurs

Resorption of the CO3Ap block is seen more clearly when the sample’s structure is picked up from the surrounding bone, as shown in Fig. 7. The CO3Ap block became smaller along with the increasing duration of the implantation period whereas the size of the HAp block was the same during the whole experimental period.

Fig. 7
figure 7

Three-dimensional images of the residual volume of the CO3Ap blocks (ad) and HAp blocks (eh) immediately after implantation (day 0) (a, e) and at 4 (b, f), 12 (c, g), and 24 (d, h) weeks after implantation in rabbits’ distal femurs

Figure 8 shows micro-CT images (side view of the cylindrical block) of the CO3Ap (a−d) and HAp (e−h) blocks just after implantation (day 0) (a, e) and at weeks 4 (b, f), 12 (c, g), and 24 (d, h) after implantation in the rabbits’ proximal tibias.

Fig. 8
figure 8

Micro-CT images (side view of the cylindrical block) of the CO3Ap blocks (ad) and HAp blocks (eh) just after implantation (day 0) (a, e) and at 4 (b, f), 12 (c, g), and 24 (d, h) weeks after implantation in the rabbits’ proximal tibias

Similar to the result for implantation of the HAp block at the femoral bone defect, HAp maintained its structure at 24 weeks postoperatively, even when the HAp block was implanted in the tibial bone defect. In other words, there were no signs of resorption in the case of sintered HAp.

In contrast to the HAp block, when the CO3Ap block was implanted at the tibial defect, its size decreased with the increasing implantation period. At 4 weeks after implantation, the CO3Ap block basically maintained its structure. However, the edge of the CO3Ap block inside the bone had become rounded, indicating that the area was being resorbed and replaced with bone. At 12 weeks after implantation, resorption and replacement with bone had proceeded in much larger quantities than at 4 weeks. All of the edge of the CO3Ap had become rounded. In addition to the edge of the CO3Ap block, other surfaces were resorbed and replaced with bone at this stage. As a result, the CO3Ap block, which had been rectangular, was becoming increasingly rounded. At 24 weeks after implantation, there was more resorption of the CO3Ap block and replacement with bone than had been seen at 12 weeks. That is, resorption of the CO3Ap block and its replacement with bone seemed to have accelerated when compared with the initial stage.

Figure 9 summarizes micro-CT sagittal images (from top view of a cylindrical block) when the sample was implanted in rabbits’ tibial bone defects. A gap was observed between the block and bone at this stage for both HAp and CO3Ap blocks. At 4 weeks, however, the bone was in contact with the surface of the CO3Ap block, whereas the gap between the HAp block and bone remained although it had decreased somewhat. At 12 and 24 weeks, the CO3Ap block had become smaller linearly with the increasing duration of the implantation period. In contrast, the HAp was the same size even at 24 weeks.

Fig. 9
figure 9

Micro-CT sagittal images of the rabbits’ proximal tibias from a top view of the CO3Ap (ad) blocks and HAp blocks (eh) immediately after implantation (day 0) (a, e) and at 4 (b, f), 12 (c, g), and 24 (d, h) weeks after implantation in rabbits’ proximal tibias

Figure 10 shows the images of the remaining tibial CO3Ap and HAp blocks at 24 weeks after implantation. Similar to the results when implanted in the femur, CO3Ap block became smaller with the increasing implantation period whereas the size of the HAp block was the same during the whole experimental period. Resorption of the CO3Ap block was faster at the tibial bone defects than at the femoral bone defects.

Fig. 10
figure 10

Three-dimensional images of the residual volume of the CO3Ap blocks (ad) and HAp blocks (eh) immediately after implantation (day 0) (a, e) and at 4 (b, f), 12 (c, g), and 24 (d, h) weeks after implantation in rabbits’ proximal tibias

Figure 11 shows the resorption rates for the CO3Ap and HAp blocks implanted at different sites. Resorption of the CO3Ap showed an approximately linear change with passing time regardless of the implantation site. Interestingly, the resorption rate of the CO3Ap block in the proximal tibia was significantly faster than that in the distal femur throughout the post-implantation period. In contrast, the HAp block showed almost no resorption during the post-implantation period regardless of the implantation site.

Fig. 11
figure 11

Resorption rate (%) for the CO3Ap and HAp blocks implanted in different sites of bony tissue (femur and tibia)

Resorption of the CO3Ap block prepared from a DCPD precursor block showed an approximately linear change throughout the post-implantation period, as shown in Fig. 11.

4 Discussion

Purpose of this study is to evaluate the osteoconductivity and resorbability of CO3Ap block fabricated from DCPD block in two different implantation sites (rabbit femoral and tibial defects) by the evaluation using micro-CT. Since we would like to compare the ability of the obtained CO3Ap block to clinically used HAp bone substitutes, a sintered HAp was chosen as a control in this experiment.

Both CO3Ap and HAp showed good osteoconductivity indicated by micro-CT analysis after 4 weeks implantation (Figs. 5, 6, 8 and 9) regardless of the implantation site. An initial osteoconductivity of CO3Ap in femur seemed to be higher than that of HAp because the surrounding bone had grown to contact the edge of CO3Ap block whereas the gap between HAp and surrounding bone was remained (Figs. 5 and 6). However, in tibia, the surrounding bone had grown to contact both CO3Ap and HAp. For making discussion on osteoconductivity more in detail, a histological evaluation in each implantation period must be necessary. Therefore, we will perform the histological evaluation in the near future.

The results obtained by the micro-CT analysis clearly demonstrated that the CO3Ap block prepared from a DCPD precursor block was gradually resorbed in the rabbit’s bony defect, whereas the sintered HAp block showed no sign of resorption throughout of post-implantation period. It is thought that the difference in resorption between the CO3Ap and HAp blocks must be due to the different degrees of SSA, porosity, and carbonate content of each specimen as shown in Table 1.

Resorption of the CO3Ap block prepared from a DCPD precursor block showed an approximately linear change throughout the post-implantation period, as shown in Fig. 11. At 24 weeks after implantation, resorption rate of the CO3Ap block at tibia and femur were about 50% and 34%, respectively. By extrapolation of the results, we estimated that it would take approximately 48 weeks (1 year) in tibia and 72 weeks (1.5 year) in femur to resorb the CO3Ap block completely. This resorption is probably due to the osteoclasts, not physical dissolution. It is because, among the calcium phosphates, apatite (including carbonate apatite) is the most thermodynamically stable phase [9]. In fact, we previously reported that CO3Ap granules (prepared from the CaCO3 precursor [9]) and CO3Ap granules (derived from the CaSO4 precursor [12]) were resorbed by osteoclasts and gradually replaced to create new bone. Therefore, we found that CO3Ap fabricated through a dissolution−precipitation reaction can be resorbed in a bony defect regardless of differences in the precursors. Although various precursors are available for fabricating CO3Ap bone substitute, the resorption rate for CO3Ap when using the dissolution−precipitation reaction is thought to be affected by the difference in the precursor’s solubility. The HAp block fabricated using an α-TCP precursor block showed greater porosity than that fabricated using a β-TCP precursor block, depending on their precursor’s solubility [26]. Because osteoclastic resorption may be affected by the porosity of the CO3Ap block, the effect of the precursor’s difference in resorption of the CO3Ap block must be examined as the next research target using the same size specimen but prepared from different precursors, such as CaCO3, CaSO4, and α-TCP.

The results obtained in this in vivo evaluation by micro-CT clearly demonstrated that resorption of the CO3Ap block was affected by the implantation site. As shown in Figs. 511, the metaphyseal site on the proximal tibia showed greater resorption of the CO3Ap blocks than the epiphyseal site on the distal femur, suggesting higher osteoclastic activity at the metaphysis. Blood vessel development is greater at the metaphysis than at the epiphysis [27], suggesting that the larger blood supply may contribute to the existence of more osteoclasts and thus greater resorption of the CO3Ap block. It might be one of the reasons for the different resorption levels of the CO3Ap block between the epiphyseal site of the distal femur and the metaphyseal site of the proximal tibia. Further histologic data for confirming new bone formation and the activities of various cells are necessary to understand the detailed mechanism. Either way, CO3Ap blocks showed far better resorption than HAp blocks.

5 Conclusions

A CO3Ap block prepared from a DCPD precursor block showed good resorption in bone tissue, whereas the sintered HAp block showed almost no resorption 6 months after implantation. The resorption rate of the CO3Ap block was faster at the metaphysis of the proximal tibia than at the epiphysis of the distal femur. Based on our results, complete resorption for a disk-shaped CO3Ap block (6 mm diameter, 4 mm high) was estimated to be approximately 1−1.5 years. Resorption of a CO3Ap block was probably achieved by osteoclasts although it should be confirmed by the histological evaluation in the future.