1 Introduction

Titanium and titanium alloys are commonly used in the fabrication of dental implants such as dental screws due to their light weight, high biocorrosion resistance, biocompatibility and mechanical properties [1]. Previous studies have shown that the long term clinical success of dental implants depends largely on the osteointegration of the implant [2]. Osteointegration is the formation of an interface between bone and the implant, without the presence of non-bone tissue between the bone and implant interface [3]. There are a number of factors that influence the rate of osteointegration such as surface roughness and surface chemistry [46].

Increasing the surface roughness of implants via grit-blasting, sand-blasting and acid etching has been shown to enhance the osteointegration of implants [79]. Surface roughness has been shown to influence cell attachment, proliferation and differentiation [1012]. Grit blasted screws have been shown to exhibit a higher bone-to-metal contact than as-machined screws and require more torque to remove from bone than smooth screws [1316]. The increase in surface roughness provides a greater surface area for mechanical interlock between the bone and the implant [17]. A range of surface roughness (R a ) between 1–10 µm maximises the interlocking between the implant and the bone [3].

The osteointegration of an implant can be enhanced by altering the surface chemistry via the application of an osteoconductive coating. An osteoconductive surface is one that permits bone to grow on its surface or down into its pores by permitting cells to attach, proliferate and deposit bone matrix [4, 18]. The formation of bone on the surface of the implant is a prerequisite for osteointegration [4]. Hydroxyapatite (HA, Ca10(PO4)6(OH)2) [19, 20] is the most widely used osteoconductive coating for hard-tissue applications as it closely resembles the natural mineral phase of bone and teeth [21]. HA is biocompatible, biodegradable and osteoconductive [18]. In spite of their widespread use, HA coatings have been reported to be susceptible to delamination and dissolution and this can result in a loss of coating integrity and ultimately aseptic loosening of the implant [2226]. To address this problem, other bioactive ceramics, such as fluorapatite (FA, Ca10(PO4)6(F)2), are being investigated as potential osteoconductive coatings [27].

Fluorapatite is analogous to HA. Like HA, FA is biocompatible, biodegradable and osteoconductive. However, FA has a number of advantages compared with HA: it is more chemically stable than HA due to its crystal structure. The F ions produce lower lattice strains than the OH ions in hydroxyapatite, this results in a more stable crystal structure [19]. The greater chemical stability of FA compared to HA results in a slower rate of degradation and this reduces the rate of coating resorption in vivo [28].

This paper investigates HA and FA coatings deposited on to dental screws using a novel coating technique, CoBlast™ [2935]. CoBlast is an ambient temperature process developed to address the problems associated with high temperature coating techniques, such as the formation of unwanted phases, amorphization of the coating and poor adherence of the coating [23, 29, 30, 32, 34, 3639]. The process utilises a co-incident stream of abrasive blast medium and a stream of coating medium particles to modify the substrate surface. The abrasive roughens the surface while simultaneously disrupting the passivating oxide layer of the substrate and exposing the reactive metal. The stream of dopant then reacts with the exposed reactive metal to form an intimate chemical bond [32]. The adhesion of the coating to the substrate is due to a combination of tribo-chemical bond formation and mechanical interlock between the bioceramic and the metal substrate [29].

Previous studies using the CoBlast process have deposited bioactive coatings onto flat metallic coupons. This paper investigates influence of this novel process on the deposition HA and FA onto more complex geometrical shapes (dental screws) and examines the influence of the process on the coating adhesion, substrate microstructure, coating crystallinity and the biological response of the deposited coating in vitro.

2 Materials and methods

2.1 Materials

Five titanium (Grade 2) dental screws 8.5 mm in length with a maximum diameter of 3.75 mm were coated per set. Five titanium (Grade 2) 15 × 15 mm coupons of 1 mm thickness were coated per set. The titanium coupons were used in the surface characterisation of the coating and substrate using X-ray diffraction and surface roughness measurements. Hydroxyapatite (HA) and fluorapatite (FA) were used as the coating media. The HA (SAI, France) particle size used was 25–60 μm. The FA (HIMED, U.S.A.) particle size was 5–30 μm. A sintered apatite (sHA) (HIMED, U.S.A.) was used as the blast medium. Two particles sizes of sHA (<106, <180 μm) were used to assess the influence of the blast medium particle size on the screw morphology. Figure 1 shows SEM images of the coating media and abrasive blast media used in this study.

Fig. 1
figure 1

SEM images of HA, FA, <106 and <180 μm sHA powders

2.2 Sample preparation

The titanium screws and coupons were ultrasonically cleaned in 1 M HCl and then in isopropanol to remove any contaminants prior to surface modification. The titanium screws were rotated at 20 mm/min. The coating media (HA, FA) and blast medium (sHA) were sprayed simultaneously in the same stream through a single de Laval nozzle at the rotating titanium screws with a jet pressure of 80 psi, at a height of 50 mm from the screws with an axial travel speed of 13 mm/min. The duration of the coating process for each 8.5 mm long screw was 40 s.

The flat titanium coupons were placed on a stationary tray. The coating media (HA, FA) and blast medium (sHA) were sprayed at the stationary titanium coupons with a jet pressure of 80 psi, at a height of 50 mm from the substrate with a nozzle travel speed of 13 mm/min. And a raster offset of 4.5 mm. Figure 2 shows a schematic of the coating process. To act as a control for the cell studies, a duplicate set of samples coated with HA/180 µm sHA were etched in 1 M HCl to remove the HA coating. This provided a rough surface without a bioactive coating for the in vitro studies.

Fig. 2
figure 2

Schematic of surface modification process

2.3 Surface characterisation

X-ray diffraction was carried out on the titanium coupons using a Siemens D500 diffractomer with rotating stage and beam monochromator (Cu Kα) (Munich, Germany). The spectra were measured using HA and FA feedstock powders and also the flat monolithic coated substrates to assess the influence of the CoBlast treatment on the HA and FA crystallinity. The analysis was carried out over the angular range of 20 to 60° (2θ). All XRD scans were carried out with a resolution of 0.04° (2θ) and a sampling time of 3 s per step. Silicon was used as an internal standard to account for non-linear peak shift. The lattice parameters of the HA and FA powders were calculated using the method of Holland and Redfern and refined using d-spacings [40].

Surface roughness of the coated titanium coupons was measured using a Taylor-Hobson Pneumo Form Talysurf Series 2 (Leicester, U. K.). Four measurements per each of the five samples were used to determine the average roughness (R a ). In order to determine the effect of the process on the titanium substrates, samples were acid etched with 1 M HCl for 90 s in an ultrasonic bath to remove the HA layers; the samples were subsequently ultrasonically cleaned in iso-propanol.

Cross sections of the titanium screws were made in order assess coating thickness and examined using a Lecia MEF4M (Wetzlar, Germany) optical microscope with reflected light. To prepare the cross sections for analysis the samples were mounted in a polyester resin, then ground and polished to a 0.04 µm finish, rinsed with water and iso-propanol respectively and dried with a stream of warm air.

2.4 Microstructural characterisation

The microstructure of the metallic substrates was examined using a Leica MEF4M (Wetzlar, Germany) optical microscope. The cross-sections were prepared as described in Sect. 2.3 and etched in Kroll’s solution for 15 s to reveal the microstructure [41] and subsequently rinsed thoroughly with water and then isopropanol and dried with a stream of warm air.

2.5 Coating adhesion

A bone model material (Renshape Grade 5460 (Mouldlife, U. K.)) was used to examine the adhesion of the coating to the screws. The model bone was composed of a cured polyurethane resin that mimics the hardness of cortical bone. This material has been used previously to examine the adhesion of coatings to screws [42]. Pilot-holes 3 mm in diameter were drilled into the board and the coated screws were screwed in and out of the holes. The screws were subsequently rinsed with distilled water and iso-propanol to remove debris. Cross-sections of the screws were prepared using the method described in Sect. 2.3. Cross-sections of the screws before and after ‘implantation’ were examined to ascertain whether there was delamination of the coating during implantation and/or removal. After removal of the screws, the bone model material was sectioned and energy-dispersive X-ray spectroscopy (EDX) analysis was carried out on the model bone material to examine coating loss during insertion, using an Oxford Swift-TM EDX system (Toronto, Canada). All scans were carried out for the same time duration (90 s), at the same magnification (×1000) at a voltage of 15 kV. EDX is limited for quantitative surface analysis due to the emission of X-rays from beneath the surface [43], thus here it is used for indicative proposes only.

2.6 Cell culture

MG-63 cells (European Collection of Cell Cultures), a human osteosarcoma cell line derived from a 14 year old male, were maintained in DMEM (Life Technologies) supplemented with 10 % heat inactivated foetal bovine serum (FBS) and 1 % l-glutamine in a humidified atmosphere at 37 °C and 5 % CO2. Cells were cultured up to ten passages.

2.7 Cell attachment and morphology

Prior to cell seeding, titanium samples were rinsed in 70 % ethanol followed by sequential washes in sterile water. Samples were then conditioned in complete media. MG-63 cells were seeded at 4 × 103cells per well in 12 well plates with titanium samples. Cells were fixed with 3 % paraformaldehyde (PFA; Sigma Aldrich) at room temperature for 20 min following a 1 or 5 day incubation. Fixed cells were stained with Hoechst 33342 (nuclei; Sigma) and phalloidin-AlexaFluor568 (F-Actin; Life Technologies). Images were acquired with Olympus Scan^R Automated Image Acquisition Software on a Scan^R IX81 microscope (Olympus) with an UPLSAPO 20× objective (Olympus) and 1344 × 1024 resolution cooled CCD camera. Standard filter sets and 64 subpositions were used. Images were analysed using Columbus image analysis software (Perkin Elmer) to determine nuclear size and roundness and cell number, size and roundness. One-way ANOVA was used for group comparison and a P < 0.05 was deemed to be significant.

3 Results

3.1 Surface characterisation

XRD analysis was carried out on the as-received powder and the deposited HA and FA coatings. Figure 3 shows the XRD patterns for HA and FA starting powders. The HA and FA powders exhibited similar diffraction patterns. Examination of (300) reflection shows a shift to a higher angle for the FA compared to the HA, Fig. 4. This is due to a reduction in the a-axis dimension of the FA compared to the HA. This reduction is due to the presence of the smaller F ions in the FA lattice compared to the larger OH ions in the HA lattice [44]. To confirm there was a reduction in the a-axis dimension of the FA, the lattice parameters of both powders were calculated as described in Sect. 2.3. The computed lattice parameters showed a reduction in the a-axis dimension for the FA compared to the HA. The computed lattice parameters for both the HA and FA powder are shown in Table 1.

Fig. 3
figure 3

XRD patterns of the as received HA powder and FA powder, the dashed lines correspond to silicon peaks

Fig. 4
figure 4

Shift of (300) reflection between HA and FA

Table 1 Lattice parameters of as received HA and FA powders

Figure 5 shows XRD plots of the as-supplied titanium and the coated substrates. The coated substrates did not exhibit any additional peaks associated with α/β-TCP or an amorphous halo in the 30–35° (2θ) region. This indicates that there was minimal change to the precursor HA and FA powders during processing. This is as expected due to the low temperature (<50 °C) nature of the process. The samples exhibited peaks associated with titanium which are present due to the underlying titanium substrate due to the very thin HA and FA layers deposited (<5 µm). The samples coated with HA exhibited a stronger trace of the coating compared to the sample coated with FA. The ratio of the most intense HA peak, at the (211) plane, to the most intense Ti peak, at the (101) plane, was 0.37 and 0.28 for the HA/180 and HA/106 µm sHA, respectively while the ratio of the most intense FA peak, at the (211) plane, to the most intense Ti peak, at the (101) plane, was 0.10 and 0.09 for the HA/180 and HA/106 µm sHA, respectively. This is due to a thicker coating of HA compared to FA.

Fig. 5
figure 5

XRD patterns of the as supplied titanium coupons, HA/180, FA/180, HA/106 and FA/106 µm sHA prepared surfaces. The dashed lines correspond to silicon peaks. The titanium peaks are marked with an X while all other peaks correspond to HA and FA respectively

The surface topography was examined using contact profilometry and SEM analysis. The average surface roughness and the roughness profile along the surface of the samples after the surface modification is shown in Fig. 6. The average surface roughness (R a ) of each of the samples increased after surface modification compared to the unmodified titanium. One-way analysis of variance (ANOVA) was performed to determine whether the difference in roughness between the various substrates was statistically significant. A P value < 0.05 was deemed to be significant. The test yielded a P value = 1.5 × 10−28. This indicates that the roughness was significantly influenced by the type of blast medium used. A Tukey post hoc test was performed to determine which surfaces had statistically significant different roughness. The results are shown in Table 2. The surface roughness of the as-supplied titanium yielded a P value < 0.05 when compared to each of the coated surfaces indicating that the blasting process significantly alters the surface roughness.

Fig. 6
figure 6

The average surface roughness and the roughness profile (5 mm) along the surface of the samples after the surface modification. Error bars indicate standard deviation

Table 2 P-values from post hoc Tukey test of the substrate surface roughness

The particle size also affected the surface roughness as substrates blasted with the larger 180 µm sHA particle gave rise to a significantly rougher surface than substrates blasted with the 106 µm sHA particle. The 106 sHA increased the roughness by a factor of 3.8 while the 180 sHA increased the surface roughness by a factor of 6.3.

The test was also used to assess the influence of the blast media on the underlying substrate in isolation by measuring the surface roughness after removal of the coating via an acid etch. The roughness of the underlying titanium is important to allow for mechanical interlock between the implant and the host tissue after resorption of the HA [45]. Comparison of the coated and etched substrates blasted with the same blast medium and coating medium yielded P values > 0.05, indicating no statistical significance in the roughness of the coated and etched samples.

Figure 7 shows the results of SEM imaging of the surface of the as-supplied titanium and the surfaces of the treated substrates before and after removal of the coating via the acid etch. The substrates exhibit a variety of features after acid etching: there are clear tears, scratches, and ridges evident. These features are more evident on the coupons blasted with the 180 µm sHA than the 106 µm sHA due to the greater abrasion of the larger particle.

Fig. 7
figure 7

SEM images of the surface of as received titanium coupon and the coated samples before and after removal of the coating via an acid etch

Cross-sections of the as-received dental screws and the coated screws are shown in Fig. 8. The figure highlights the thin coating (<5 µm) of HA and FA produced by the CoBlast process. The threads of the CoBlasted screws exhibited rounding compared to the unmodified screw. This is due to the abrasive nature of the CoBlast process. The degree of rounding was more significant for the screws blasted with the 180 µm sHA than the 106 µm sHA. Figure 9 shows SEM images of the coatings on the surface of the screws. The coating thickness varied depending on the type of blast media and coating particle used. The average coating thickness of the HA/106, HA/180, FA/106, FA/180 µm sHA coating were 4.9, 6.3, 2.4 and 2.8 µm respectively. The difference coating thickness between the HA and the FA coatings may be due to the difference in the particle size of the powders used. The HA a particle size of 25–60 μm and the FA had a particle size of 5–30 μm.

Fig. 8
figure 8

Optical micrographs of the as received dental screws and the coated screws. The black layer represents the HA and FA coatings

Fig. 9
figure 9

Cross-section SEM images of the coatings on the surface of the screws

3.2 Coating adhesion

Figure 10 shows the coated screws that were not inserted into the model bone (left) and screws after insertion into and subsequent removal from the model bone (right). From the optical micrographs the coating did not exhibit signs of delamination. The screws after insertion retain similar feature to the un-inserted screws. It should be noted that there is an apparent difference in screw width between the un-inserted FA/180 µm screw and the FA/180 µm screw after insertion into the model bone. This is due to the depth of grinding of the screw during the preparation of a cross section for optical microscopy. EDX analysis of the model bone material after removal of the screw yielded no elements from the uncoated screw while trace amounts of calcium only were detected on all of the model bone material that had coated screws inserted.

Fig. 10
figure 10

Optical micrographs of dental screws coated with HA and FA. Un-inserted screws (left) and screws after insertion into and subsequent removal from the model bone (right). The inserted screws do not show signs of delamination of the coating

3.3 Microstructural characterisation

The microstructures of the dental screws were examined to assess what influence the coating process had on the screws’ microstructure. As shown in Fig. 11, the screws treated via the CoBlast process exhibit a change in their microstructure compared to the as supplied cp-titanium screws. Three different regions were observed in the in the substrates treated using this process, a severely deformed surface layer, a region characterised by gross deformation of the grains and the un-deformed substrate consisting of equiaxed α-grains. These three different regions have previously been identified by Jiang et al. for sandblasted cp-titanium [46]. The region characterised by gross deformation of the grains exhibited twinning.

Fig. 11
figure 11

Etched cross-sections of titanium screws (As supplied titanium screw, FA/106, HA/106, FA/180 and HA/180 sHA, respectively). The treated screws exhibit signs of plastic deformation and twin formation. The white line outlines this deformation zone

3.4 Cell attachment and morphology

In order to assess cell attachment and cell morphology MG-63 cells were cultured on the various surfaces followed by fixation, staining and analysis using a high content automated imaging and analysis platform. The cell attachment and cell morphology results are shown in Fig. 12. Cell attachment was measured as cell counts and cell morphology determined as nuclear size, nuclear roundness, cell size and cell roundness.

Fig. 12
figure 12

Cell numbers (left) and cell roundness (right) on the prepared surfaces after 1 and 5 days. Error bars indicate standard deviation. Images of the cells seeded onto the various surface types (×20 magnification), a Blank, b HA/sHA etched, c HA/sHA and d FA/sHA (after 5 days/s). Cell nuclei are stained with Hoechst 33342 (blue) and cell boundaries (F-actin) are visualised with phalloidin-AlexaFluor568 (red) (Color figure online)

Since the surfaces of the samples are uneven it is not possible to focus on all cells within a single field of view. This limitation affected cell segmentation during image analysis and meant accurate cell number comparisons between rough and smooth surfaces could not be determined, hence only the results for the rough surface (HA/180 µm sHA (etched), HA/180 µm sHA and HA/180 µm sHA) are presented. This was deemed valid as previous studies have shown that roughened surfaces enhance cellular attachment, proliferation and differentiation compared to smooth surfaces [1012]. Examining the cell numbers after 1 and 5 days there was no statistical difference (P 1 day < 0.54, P 5 days < 0.06) between roughened titanium sample (HA/sHA (etched) and the CoBlasted surfaces which contained a rough bioactive surface. Additionally there were no statistical significances for cell morphological measurements between the different surfaces (P 1 day < 0.31, P 5 days < 0.07). The cells seeded on each surface exhibited similar morphology at 1 and 5 days. Figure 12 shows the cells seeded onto the various surfaces after 5 days.

4 Discussion

CoBlast is a novel coating technique that has been used in previous studies to deposit hydroxyapatite onto various metallic substrate types [2935]. In these previous studies the HA was deposited onto flat coupons. This study investigates the disposition HA and FA using this novel coating technique onto more complex geometrical shapes (dental screws).

The optical micrographs of the titanium screws after the deposition process exhibit a coating <5 µm thick on the surface of the screws. Previous studies have shown that altering the abrasive to dopant ratio, powder chemistry, blast pressure and blast height can alter the coating thickness marginally, but coatings in excess of 10 microns are rare and are not likely with the materials described herein [2935]. The coating thickness is independent of spraying time, as the process gives a balance of abrasion and deposition. Additional spraying merely produces additional substrate erosion, but the coating thickness does not change significantly.

Comparing the coated screws with the unmodified screw, there is a change in screw morphology. The threads of the coated screws exhibit rounding; this is due to the abrasive nature of the coating process. The degree of abrasion and rounding of the screw threads was dependent on the size of the blast medium particle used. The screws blasted with the 106 µm sHA broadly retained the trapezoidal shape of the screw with minimal rounding of the threads. The screws blasted using the 180 µm sHA particles exhibited significantly greater abrasion and rounding. This is due to the larger particle size of the 180 µm sHA particle. Although, the change in screw morphology is undesirable, previous studies suggest this does not adversely affect bone formation around the screw. Previous studies have reported grit blasted screws to exhibit a higher bone-to-metal contact than as-machined screws and require more torque to remove them from bone [1315].

Examining the surface roughness (R a ) of the coupons, Fig. 6, the 180 µm sHA particles significantly increase the surface roughness compared to the coupons blasted with the 106 µm sHA particles. SEM images of the etched coupons show that the coupons blasted with the 180 µm sHA particles had deeper and wider tears and scratches than the coupons blasted with 106 µm sHA particles, Fig. 7. Although, there is a significant difference in the roughness of the coupons blasted with the 106 and 180 µm sHA, both have roughness within the reported optimum range for osteointegration (1–10 µm) [3].

The assessment of the coating adhesion via insertion the screws into a model bone material provides a good assessment of how the coating will respond to biomechanical forces in clinical applications [47]. EDX scans of the bone model material after removal of the unmodified screw did not detect the presence of any elements. This indicates that there was no wear or fretting of the titanium during insertion. For the model bone inserted with coated screws, trace amount of calcium only were detected. This calcium is due to HA and FA particles liberated during insertion. It should be noted that calcium signal was very weak and that no phosphorous was detected indicating that the coating loss was minimal. As discussed in Sect. 2.5, EDX is limited for quantitative analysis surface analysis and thus was used here for indicative purposes only. Further, analysis is required to quantify the degree of wear and fretting of the titanium and coating loss during insertion of the screws. Examining the cross-sections of the screws before and after removal from the model bone, Fig. 10, there appears to be no obvious signs of damage to the coating indicating that there was minimal coating loss during to insertion of the screws.

The CoBlast process resulted in a change in the screws microstructure compared to the unmodified screws, coated screws exhibit a severely deformed surface layer, a region characterised by gross deformation of the grains and an un-deformed substrate consisting of equiaxed α-grains, within the region characterised by gross deformation of the grains twinning was observed. Twinning has previously been observed in shot peened cp-titanium [48]. The formation of twins is promoted by a high strain rate, shock loading and relatively large grain size [49]. This deformation and twinning is most pronounced at sites normal to the blast stream, such as the tip of the screw threads. This is due to the orientation of the screw to the blast stream. The tip of the threads are normal to the blast stream, thus a higher percentage of the particle energy is imparted to the surface compared to the sides of the threads [50]. For the screws blasted with the 180 µm sHA, the change in microstructure was identified to a depth of 30 µm from the surface of the tip of the thread while at the sides of the thread the change in microstructure was to a depth of 10 µm. The screws blasted with 106 µm sHA exhibited a change in microstructure to a depth of 20 µm from the surface of the tip of the thread while at the sides of the thread the change in microstructure was to a depth of 5 µm. The formation of the severely deformed surface layer may be beneficial as the associated compressive strength in the surface layer improves fatigue performance, a key material property for medical device implants [46, 51].

The in vitro tests indicate that there is no statistical difference in cell numbers at 1 and 5 days between the roughened titanium sample (HA/sHA (etched) and the CoBlasted surfaces which contained a rough bioactive surface (HA/sHA and FA/sHA). Previous studies have shown that rough surfaces enhance cell attachment, proliferation and differentiation compared to smooth surfaces [1012]. This indicates that this coating process can incorporate a bioactive coating onto a roughened surface without adversely affecting the initial cell response. Similarly, there was no statistical difference in the nuclear or cell morphologies between the cells seeded onto the blanks and the cells seeded onto the other surfaces; this would indicate that the process does not adversely affect cell health.

It should be highlighted that in vitro models are simplistic models that do not account for the complex biological interactions within biological systems and that further in vivo testing is required to fully assess the biological response to bioactive coatings deposited using this process.

5 Conclusions

The CoBlast process can be used to deposit highly adherent coatings of HA and FA onto dental screws. The process alters the morphology of the screw threads the degree of alteration is significantly influenced by the size blast media used. Previous studies indicate that moderate alteration of the screws morphology does this does not adversely affect performance in vivo. Initial in vitro tests on substrates coated using this process indicate that it does not adversely affect the biological response of cells.