Abstract
Objective
This study aimed to characterize the interactions of pacemakers with magnetic resonance imaging (MRI) and to identify device characteristics that could predict adverse interactions.
Background
The safety of MRI in patients with indwelling pacemaker systems remains uncertain. Previous studies demonstrated safety in most patients, but unpredictable, potentially concerning changes in pacemaker behavior have occurred.
Methods
We prospectively studied patients with pacemaker devices in situ who were not pacemaker dependent and in whom MRI was essential for adequate diagnosis and treatment. All patients were monitored by electrocardiography and pulse oximetry during scanning; devices were interrogated and cardiac enzymes were measured before and after scanning.
Results
Of 32 patients studied (46 MRI examinations), 28 patients had a dual-chamber system and one had a biventricular device. Regions scanned were the head and spine. Devices were reprogrammed to asynchronous pacing or sense-only mode in all except six patients before MRI. During six scanning episodes (five patients), “power-on” resetting of the device was noted. Magnet-mode pacing was noted during four episodes (three patients). Occasional premature ventricular contractions were noted in one patient. No significant changes in battery voltage, sensed P wave and R wave, pacing thresholds, lead impedance, or cardiac enzymes were noted immediately after MRI or at 1-month follow-up.
Conclusions
Overall, no significant changes were seen in pacemaker device function, and no adverse clinical events were observed. A minority of patients with older devices had unpredictable changes in device behavior, which stresses the need for close monitoring during and careful device interrogation after scanning.
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Avoid common mistakes on your manuscript.
1 Introduction
Magnetic resonance imaging (MRI) is an important diagnostic imaging modality that has had increasing utility in the field of cardiology [1]. The safety of MRI in patients with permanent pacemaker systems remains uncertain and continues to provoke controversy [2–5]. Because of the increasing prevalence of cardiac implantable electronic devices (CIEDs) and the increased demand for MRI, it has been estimated that up to 75% of patients who currently have a CIED are likely to require an MRI scan during their lifetime [6].
The current guidelines issued by the American Heart Association regarding use of MRI in patients with CIEDs discourage performing MRI in patients with a pacemaker who are not pacemaker dependent (“non-pacemaker-dependent patients”) and recommend that it should only be considered if there is a strong clinical indication and if the benefits clearly outweigh the risks [7]. They further recommend that MRI examinations not be performed on pacemaker-dependent patients and those with implantable cardioverter defibrillators (ICDs) [7]. Indeed, all of the major CIED manufacturers have stated that they do not support or recommend the use of devices currently available in the USA with MRI [8–10]. A similar position has also been taken by the US Food and Drug Administration (FDA) [11].
The concerns in relation to CIEDs exposed to the strong magnetic fields during MRI scanning principally extend to changes in programmed parameters and inappropriate sensing and pacing and the potential effect of lead tip heating on myocardial tissue. Although recent studies have demonstrated the safety of MRI in most patients with pacemakers included in specific protocols [12, 13], abnormalities in pacemaker function during pacing continue to be reported [13–16], some of which are potentially concerning, even with the current generation of available devices [17, 18]. Nonetheless, the increasing demand for use of MRI in patients with CIEDs has motivated clinicians and engineers to continue to research pacemaker and MRI interactions to permit the development of MRI-safe devices [13, 19, 20].
In this prospective study, we aimed to characterize pacemaker–MRI interactions in the present era at a single center and to identify device characteristics (generator model and brand, lead model and brand, programming mode during scanning, and implant site) that could predict any adverse interactions. We hypothesized that MRI can be safely conducted in non-pacemaker-dependent patients in an environment with beat-to-beat heart rhythm monitoring and a medical team consisting of members with expertise in radiology and cardiac devices.
2 Methods
2.1 Patient selection
2.1.1 Inclusion criteria
Patients seen at Mayo Clinic, Rochester, MN, USA, between January 2008 and December 2009 and in whom MRI examination was indicated and who had permanent pacemakers in situ were eligible for inclusion. In consultation with the referring physician, the radiologist discussed whether alternative imaging choices could provide similar information at less risk to the patient. If it was determined that MRI was the preferred imaging modality for a given patient, the referring clinician requested a pacemaker clinic consultation, including a device interrogation/examination (by nursing staff) and a clinical consultation (by a staff cardiologist). The main aim of the pacemaker clinic consultation was to determine whether the patient was pacemaker dependent and to determine the relative need for pacing (percentage of pacing) with the baseline device setting. Only non-pacemaker-dependent patients were considered for MRI. Furthermore, the lead system was required to be mature (implanted for more than 90 days). During the initial phase of this study, efforts were made to limit cases to head MRI and to scanners for which MRI-compatible monitoring equipment was available. In consultation with MRI physicists, scanning protocols and parameters were planned.
The patient and referring physician were informed that the imaging examination may affect some pacemakers and possibly discharge the battery [21]. After assessing the patient’s pacer function and reviewing the goals and risks of the procedure, the cardiologist obtained written consent from each patient that was scanned into the patient’s electronic medical record. The study was approved by the Mayo Clinic Institutional Review Board.
2.1.2 Exclusion criteria
Patients who were pacemaker dependent, as evidenced by absence of a stable escape rhythm of more than 30 beats per minute (bpm) or the presence of a potentially life-threatening arrhythmia or hemodynamic instability while the pacemaker was turned down to noncapture or subthreshold pacing, were excluded. Other exclusion criteria for this study were age younger than 18 years, presence of more than one implanted pulse generator, abnormal values for troponin I (>0.03 ng/mL) or creatine kinase-MB (>6.2 ng/mL), presence of any type of ICD device, and patients requiring sedation, anesthesia, or continuous intravenous medication, especially for cardiovascular support. In addition, patients were excluded if they had evidence of inadequate pacemaker function, as evidenced by one of the following: (1) high pacing threshold—capture at 0.5-ms pulse width at more than 3.0 V, (2) pacing lead impedance greater than 2,000 Ω, (3) battery voltage less than 2.7 V, or (4) battery longevity prediction of less than 6 months or at the elective replacement indicator.
2.2 Device programming before MRI examination
On the day of examination, at the MRI site, device interrogation and thresholds were documented and the patient’s intrinsic heart rate determined. The pacemaker was programmed to asynchronous pacing at the intrinsic heart rate plus 20 bpm (not exceeding 110 bpm) in AOO, VOO, or DOO mode. If the intrinsic heart rate was greater than 90 bpm, the device was programmed to monitor only (OAO, OVO, or ODO). If the monitor-only feature was not available, the device was programmed to a subthreshold output (AAI, VVI, or DDD). All other diagnostic and therapeutic features were programmed off (rate response, capture management, and mode switch).
2.3 MRI protocol
Mayo Clinic recently developed a pilot clinical protocol for body and head MRI in patients with cardiac devices (Fig. 1). This protocol incorporated common elements from protocols in the published literature [13, 17, 22, 23]. Development of the protocol was a collaboration between the Department of Radiology and the Electrophysiology group in the Division of Cardiovascular Diseases. The Safety and PM Monitoring Committee included two radiologists, one physicist, three electrophysiologists, and two pacemaker nurses. During each MRI examination, particularly early in the study, at least one radiologist and one physicist member of the Safety and PM Monitoring Committee were present to assess any potential clinical or device adverse events. Later in the study, other radiologists and physicists, who could be in immediate contact with members of the Safety and PM Monitoring Committee, participated in some of the studies once they were properly trained in the protocol.
2.4 Monitoring
The patient was monitored by a cardiologist or pacemaker nurse throughout the MRI procedure by pulse oximetry, CO2 measurement, electrocardiography (3-lead), and blood pressure measurement at 2-min intervals. In order to assess for the effects of lead tip heating, cardiac enzyme levels (creatine kinase and troponin I) were measured before and after each study as normal levels of these parameters after scanning are a helpful indicator of the absence of significant myocardial injury [16, 24]. Heart rate was monitored continually during the examination, with pulse oximetry and appropriate electrocardiographic monitoring. The MRI acquisition would be halted immediately whenever the cardiologist supporting the examination deemed necessary. The cardiologist or, for later studies, pacemaker nurse reviewed all physiologic data before and after the patient was moved into the center of the magnet bore to be familiar with normal signal artifacts. The examination was stopped if adequate monitoring could not be achieved. Patients were not anesthetized and were instructed to inform the MRI technologist and monitoring cardiologist if pain or discomfort occurred during the examination.
2.5 MRI settings
The MRI pulse sequences used were tailored by the radiologist and physicist supporting the examination. Magnetization transfer pulses were not used, and the specific absorption rate was limited to the minimum needed to acquire adequate imaging. The recommended specific absorption rate limit was 1.5 W/kg for up to 30 min of acquisition time, as described by Sommer et al. [13].
When possible, before imaging, the automatic and manual prescan sequences were used to assess the patient’s sensitivity to radiofrequency: If the patient’s heart rate synchronized to the transmission–receive interval of the radiofrequency coil, the scan would be discontinued.
2.6 Post-MRI patient assessment
The cardiologist examined the patient after imaging was completed. A pacemaker interrogation was performed and data collected before returning the pacemaker to the patient’s original settings.
2.7 Statistical analysis
Fisher exact tests and Pearson χ 2 tests were performed for bivariate analysis of categorical variables. The effect of time on continuous variables was analyzed by one-way ANOVA analysis. P < 0.05 was accepted as statistically significant. All continuous numerical data are represented as mean (SD), except where otherwise indicated.
3 Results
3.1 Patients and examinations
A total of 32 patients with pacemaker devices in situ met all the inclusion criteria for the study; 17 (53%) were men. The mean (SD) patient age was 67 (14) years. The clinical indications for MRI examination are shown in Table 1. Most studies were MRI scans of the head (n = 35), and the remainder were spine studies (12 cervical, 7 thoracic, 5 lumbar) and one pelvic examination. All MRI examinations performed were diagnostic, and in none of the studies was artifact from the device sufficient to interfere with interpretation.
Forty-six MRI scanning episodes were studied in the 32 patients. Twenty scanning episodes were of more than one region. The devices and leads used and specific indications for MRI are shown for each patient in Table 2. One patient had a biventricular device, and the rest had a single right ventricular lead. Twenty-six patients also had a functioning right atrial lead. In the case of one patient, the generator was in an abdominal location with a ventricular epicardial lead, but this was a nonfunctioning device.
3.2 Device–MRI interaction during MRI
In the study group, 11 occurrences of abnormal pacemaker function were observed in nine patients (Table 3). Occasional premature ventricular contractions were noted during one scanning episode. During six scanning episodes in five patients, the device underwent “power-on” resetting (POR). The device reverted to magnet-mode pacing (Fig. 2) during four further scanning episodes (three patients), but normal function resumed after the scan. The patient with a nonfunctioning abdominally placed device had abdominal pain during the scan, but this was not believed to be related to change in position of the pacemaker generator or lead. Otherwise, no other device function abnormalities were noted. No significant clinical adverse events were observed during the MRI examinations, and all patients completed the study.
The occurrence of premature ventricular contractions during scanning was only noted with the single Biotronik device studied in this series (P = 0.02). POR appeared to occur more frequently with the Medtronic Kappa devices during MRI (Table 4). This observation was on the borderline of statistical significance (P = 0.05). This finding held regardless of whether a synchronous or asynchronous mode of pacing was selected before MRI (P = 0.69).
Lead model did not predict any of the observations of abnormal pacemaker function noted in this series, nor did region scanned. Because only one patient had a left ventricular lead, the effect of this was not analyzed statistically.
3.3 Device function after MRI
No effect on generator voltage was noted in the immediate period after MRI or at 1 month post-MRI. Similarly, no significant effect was noted on lead function as assessed by P wave and R wave detection, measurement pacing threshold, or lead impedance at the time intervals above (Table 5).
3.4 Cardiac muscle enzymes
No patient had an increase in cardiac muscle enzymes observed when these were checked before and 24 h after MRI scanning (mean [SD] creatine kinase-MB, 3.2 [1.7] ng/mL; mean [SD] cardiac troponin I, 0.01 [0.01] ng/mL) (n = 22).
4 Discussion
The current study demonstrates the safety of MRI scanning of the head and body in our population of non-pacemaker-dependent patients. Our data add to similar data from other centers. Pacemaker–MRI interactions in this cohort were not infrequent; POR (n = 6), magnet-mode pacing (n = 4), and frequent premature ventricular contractions (n = 1) were all presumed to occur because of this interaction. However, all interactions were transient and reversible and resulted in no adverse consequences for the patient or for device function. We observed some relationships between pacemaker–MRI interactions and specific pacemaker models, but the relatively small numbers of each device model studied advocates caution in the interpretation of these results.
A very important component of the practice approach described in this paper is the collaborative protocol between cardiology and radiology, which ensured optimal patient safety with real-time continuous monitoring, while maintaining the diagnostic quality of the studies performed. Other centers have described the presence of an electrophysiologist and radiologist during the MRI procedure, along with continuous monitoring [13, 22, 23]. This has been incorporated into our current clinical practice with the establishment of a Safety and PM Monitoring Committee, which included representation from the departments of radiology, medical physics, and cardiology and pacemaker nursing staff. We believe that this collaborative clinical model is relevant to most institutions in which radiology and cardiology/electrophysiology are distinct and separate clinical practices.
In the current study, several pacemaker–MRI interactions were observed that have been previously documented by other groups. The most frequent of these was POR in six patients. POR occurs when the pacemaker generator battery voltage drops below a critical preset level determined by the manufacturer because below this voltage, the operation of the device is believed to be unpredictable. Thus, after the onset of POR, all device functions are disabled until the battery voltage exceeds this preset level, known as the POR trip voltage. Typically, the device then “resets” all pacing and sensing parameters to nominal settings determined by the manufacturer. Usually, the pacing mode is VVI, with nominal manufacturer-determined pacing rates, amplitudes, pulse width settings, sensitivities, and refractory periods. Occasionally, these parameter changes may be permanent. In pacemaker-dependent patients, POR can have potentially disastrous consequences if MRI interference inhibits VVI pacing, as was reported recently in a pacemaker-dependent patient in whom asystole developed for this reason. A review of 115 patients with Medtronic implants by Sommer et al. [13] demonstrated POR in seven patients’ devices (all Sigma and Thera models). However, other reviews of pacemaker and MRI interactions demonstrated no incidence of POR at 1.5 T [22]. Our experience with POR seems similar to that reported recently in a study of pacemakers and ICDs in the magnetic field of a remote magnetic navigation system, in which a preponderance of POR was noted in Medtronic Kappa model pacemakers [25].
The next most common form of pacemaker–MRI interaction seen was asynchronous pacing at the magnet rate in four patients, resulting from reed-switch activation in the MRI magnetic field. This incidence is somewhat lower than the experience of Sommer et al. [13] (26 of 47 patients) and Nazarian (10 of 55 devices, including ICDs) [22]. Furthermore, some devices may revert to magnet-mode pacing even after switching them to an asynchronous mode, as was our experience. In many cases in which programming is set to asynchronous for the scan, reversion to magnet-mode pacing simply results in a change in rate. However, in patients with higher heart rates in which the device is set to sense-only mode, the change in setting to asynchronous pacing could be clinically significant. Some experts in the field have suggested switching the magnet mode off when possible and then programming the device as desired [26]. Furthermore, it has been suggested that it would be ideal for electrocardiographic gating during MRI to take place only during the ventricular refractory period, but that, in practical terms, this would most likely increase the duration of the scan unacceptably.
In our study, we saw no evidence of device malfunction immediately after MRI or at the time of repeat testing 1 month after scanning. Furthermore, in contrast to the findings of previous authors, we noted no abnormality in cardiac enzymes after MRI [13]. These findings agree with the collective experience to date that scanning at 1.5 T and higher has been generally safe, with only minimal threshold changes [12, 13] and, in one study, a minor change in cardiac troponin levels [13]. In other studies, no significant changes in pacing thresholds, cardiac enzyme levels, or device function were noted [22, 27]. This is in accordance with previous work which has demonstrated the absence of significant heating at the lead tips in animal and ex vivo studies [26, 28, 29]. These encouraging collective clinical and laboratory findings have led to a proposed protocol for MRI in patients with CIEDs [22] and safety recommendations [7, 30].
In Europe, MRI-conditional pacemakers already have a CE mark and are thus approved for implantation. In the USA, FDA approval is awaited for these devices but is likely to occur in the next year or so. However, even if MRI-conditional pacing systems achieve widespread availability, patients with existing pacing systems will continue to need MRI for some time. Furthermore, with the exception of new implants (i.e., MRI-conditional pulse generator and leads, after FDA approval), simply upgrading patients to MRI-safe generators and continued use of the existing lead(s) or abandoning and capping the existing lead(s) will not resolve concerns related to MRI. Such concerns would be alleviated only if all existing hardware were removed, which is associated with additional morbidity and mortality related to lead extraction.
5 Conclusions
The current study reports the findings of a collaborative approach to MRI in patients with implanted permanent pacemakers at a single institution. We demonstrate that in non-pacemaker-dependent patients, with a closely monitored MRI environment, patient safety is achieved. In devices that are available currently, some pacemaker–MRI interactions were observed but did not result in any clinically significant adverse events or device dysfunction. Despite pending approval of MRI-conditional devices, the current study provides a clinical model for patients with pacemakers in situ who require MRI examination.
Abbreviations
- CIED:
-
Cardiac implantable electronic device
- FDA:
-
US Food and Drug Administration
- ICD:
-
Implantable cardioverter defibrillator
- MRI:
-
Magnetic resonance imaging
- POR:
-
Power-on resetting
References
Duru, F., Luechinger, R., Scheidegger, M. B., Luscher, T. F., Boesiger, P., & Candinas, R. (2001). Pacing in magnetic resonance imaging environment: Clinical and technical considerations on compatibility. European Heart Journal, 22, 113–124.
Duru, F., Luechinger, R., & Candinas, R. (2001). MR imaging in patients with cardiac pacemakers. Radiology, 219, 856–858.
Helfer, J. L., Gray, R. W., MacDonald, S. G., & Bibens, T. W. (2006). Can pacemakers, neurostimulators, leads, or guide wires be MRI safe? Technological concerns and possible resolutions. Minimally Invasive Therapy & Allied Technologies, 15, 114–120.
Shinbane, J. S., Colletti, P. M., & Shellock, F. G. (2007). MR in patients with pacemakers and ICDs: Defining the issues. Journal of Cardiovascular Magnetic Resonance, 9, 5–13.
Stevenson, B., Dabney, W., & Frysz, C. (2007). Issues and design solutions associated with performing MRI scans on patients with active implantable medical devices. Conference Proceedings—IEEE Engineering in Medicine and Biology Society, 2007, 6167–6170.
Kalin, R., & Stanton, M. S. (2005). Current clinical issues for MRI scanning of pacemaker and defibrillator patients. Pacing and Clinical Electrophysiology, 28, 326–328.
Levine, G. N., Gomes, A. S., Arai, A. E., et al. (2007). Safety of magnetic resonance imaging in patients with cardiovascular devices: An American Heart Association scientific statement from the Committee on Diagnostic and Interventional Cardiac Catheterization, Council on Clinical Cardiology, and the Council on Cardiovascular Radiology and Intervention: Endorsed by the American College of Cardiology Foundation, the North American Society for Cardiac Imaging, and the Society for Cardiovascular Magnetic Resonance. Circulation, 116, 2878–2891.
Levine, P. A. (2005). Industry viewpoint: St. Jude Medical: Pacemakers, ICDs and MRI. Pacing and Clinical Electrophysiology, 28, 266–267.
Smith, J. M. (2005). Industry viewpoint: Guidant: Pacemakers, ICDs, and MRI. Pacing and Clinical Electrophysiology, 28, 264.
Stanton, M. S. (2005). Industry viewpoint: Medtronic: Pacemakers, ICDs, and MRI. Pacing and Clinical Electrophysiology, 28, 265.
Faris, O. P., & Shein, M. (2006). Food and Drug Administration perspective: Magnetic resonance imaging of pacemaker and implantable cardioverter-defibrillator patients. Circulation, 114, 1232–1233.
Naehle, C. P., Sommer, T., Meyer, C., et al. (2006). Strategy for safe performance of magnetic resonance imaging on a patient with implantable cardioverter defibrillator. Pacing and Clinical Electrophysiology, 29, 113–116.
Sommer, T., Naehle, C. P., Yang, A., et al. (2006). Strategy for safe performance of extrathoracic magnetic resonance imaging at 1.5 Tesla in the presence of cardiac pacemakers in non-pacemaker-dependent patients: A prospective study with 115 examinations. Circulation, 114, 1285–1292.
Luechinger, R., Duru, F., Scheidegger, M. B., Boesiger, P., & Candinas, R. (2001). Force and torque effects of a 1.5-Tesla MRI scanner on cardiac pacemakers and ICDs. Pacing and Clinical Electrophysiology, 24, 199–205.
Luechinger, R., Duru, F., Zeijlemaker, V. A., Scheidegger, M. B., Boesiger, P., & Candinas, R. (2002). Pacemaker reed switch behavior in 0.5, 1.5, and 3.0 Tesla magnetic resonance imaging units: Are reed switches always closed in strong magnetic fields? Pacing and Clinical Electrophysiology, 25, 1419–1423.
Luechinger, R., Zeijlemaker, V. A., Pedersen, E. M., et al. (2005). In vivo heating of pacemaker leads during magnetic resonance imaging. European Heart Journal, 26, 376–383. discussion 25–7.
Gimbel, J. R., Wilkoff, B. L., Kanal, E., & Rozner, M. A. (2005). Safe, sensible, sagacious: Responsible scanning of pacemaker patients. European Heart Journal, 26, 1683–1684.
Gimbel, J. R. (2009). Unexpected asystole during 3T magnetic resonance imaging of a pacemaker-dependent patient with a ‘modern’ pacemaker. Europace, 11, 1241–1242.
Mitka, M. (2009). Researchers seek MRI-safe pacemakers. JAMA, 301, 476.
Sutton, R., Kanal, E., Wilkoff, B. L., et al. (2008). Safety of magnetic resonance imaging of patients with a new Medtronic EnRhythm MRI SureScan pacing system: Clinical study design. Trials, 9, 68.
Rozner, M. A., Burton, A. W., & Kumar, A. (2005). Pacemaker complication during magnetic resonance imaging. Journal of the American College of Cardiology, 45, 161–162. author reply 2.
Nazarian, S., Roguin, A., Zviman, M. M., et al. (2006). Clinical utility and safety of a protocol for noncardiac and cardiac magnetic resonance imaging of patients with permanent pacemakers and implantable-cardioverter defibrillators at 1.5Tesla. Circulation, 114, 1277–1284.
Naehle, C. P., Strach, K., Thomas, D., et al. (2009). Magnetic resonance imaging at 1.5-T in patients with implantable cardioverter-defibrillators. Journal of the American College of Cardiology, 54, 549–555.
Mollerus, M., Albin, G., Lipinski, M., & Lucca, J. (2008). Cardiac biomarkers in patients with permanent pacemakers and implantable cardioverter-defibrillators undergoing an MRI scan. Pacing and Clinical Electrophysiology, 31, 1241–1245.
Jilek, C., Tzeis, S., Reents, T., et al. (2010). Safety of implantable pacemakers and cardioverter defibrillators in the magnetic field of a novel remote magnetic navigation system. Journal of Cardiovascular Electrophysiology, 21, 1136–1141.
Irnich, W., Irnich, B., Bartsch, C., Stertmann, W. A., Gufler, H., & Weiler, G. (2005). Do we need pacemakers resistant to magnetic resonance imaging? Europace, 7, 353–365.
Goldsher, D., Amikam, S., Boulos, M., et al. (2006). Magnetic resonance imaging for patients with permanent pacemakers: Initial clinical experience. The Israel Medical Association Journal, 8, 91–94.
Roguin, A., Zviman, M. M., Meininger, G. R., et al. (2004). Modern pacemaker and implantable cardioverter/defibrillator systems can be magnetic resonance imaging safe: In vitro and in vivo assessment of safety and function at 1.5 T. Circulation, 110, 475–482.
Shellock, F. G., Fischer, L., & Fieno, D. S. (2007). Cardiac pacemakers and implantable cardioverter defibrillators: In vitro magnetic resonance imaging evaluation at 1.5-Tesla. Journal of Cardiovascular Magnetic Resonance, 9, 21–31.
Roguin, A., Schwitter, J., Vahlhaus, C., et al. (2008). Magnetic resonance imaging in individuals with cardiovascular implantable electronic devices. Europace, 10, 336–346.
Acknowledgments
The authors gratefully acknowledge the assistance of Nicholas E. Payne RN and Jeff Burrows (Medtronic, Minneapolis, MN, USA) and Gene A. Bornzin, Ph.D. (St. Jude Medical, Sylmar, CA, USA) for their technical and engineering advice in the preparation of this manuscript.
Disclosures
Dr. Friedman is involved in sponsored research with Medtronic, Boston Scientific, Bard, St. Jude Medical, and Pfizer; has intellectual property rights with Bard EP, Hewlett Packard, Medical Positioning, Inc., Aegis Medical, and NeoChord; and has spoken for and/or consulted for Medtronic, Boston Scientific, and St. Jude Medical. David L. Hayes serves as a member of the advisory board for St. Jude Medical, Boston Scientific. He also serves as a speaker at educational venues for St. Jude Medical, Boston Scientific, Sorin Medical, Biotronik, and Medtronic. Dr. Shen receives funding from Medtronic for an investigator initiated research project.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Boilson, B.A., Wokhlu, A., Acker, N.G. et al. Safety of magnetic resonance imaging in patients with permanent pacemakers: a collaborative clinical approach. J Interv Card Electrophysiol 33, 59–67 (2012). https://doi.org/10.1007/s10840-011-9615-8
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10840-011-9615-8