Introduction

Multidetector row computed tomography (MDCT) is a rapidly evolving technology that has been positioned as the non-invasive diagnostic approach with the highest predictive accuracy for the detection of coronary stenosis in selected populations [1, 2]. Nevertheless, recent data suggesting that revascularization does not improve the prognosis of patients with intermediate coronary artery stenosis if the stenosis does not impair flow during stress, renders sole anatomical assessment of coronary stenosis without myocardial perfusion information a very useful albeit insufficient approach for clinical decision making [3]. In this regard, myocardial perfusion imaging has shown to be a useful and accurate tool in the diagnosis and prognosis of patients with coronary artery disease (CAD) [4].

Until recently, noninvasive coronary angiography by MDCT was restricted to the evaluation of coronary stenoses from an anatomical standpoint, whereas the assessment of the functional significance of coronary lesions remained outside of its scope [5]. A number of subsequent studies have challenged this limitation by demonstrating that myocardial perfusion during first-pass adenosine stress contrast-enhanced MDCT is feasible and related to microsphere-derived myocardial blood flow [611].

The advent of prospectively gated acquisition techniques for 64 slice CT coronary angiography has allowed a significant reduction in dose exposure. Consequently, a combined approach of angiography and myocardial perfusion imaging with MDCT might potentially become feasible at a total radiation dose of less than 10 mSv, particularly for the assessment of patients with established coronary artery disease, who are likely to have diffuse calcification [12, 13].

Since the assessment of myocardial perfusion by MDCT is based on the myocardial signal density (SD), it is pivotal to determine the normal values of myocardial SD and to identify potential mechanisms of misinterpretation of perfusion defects. In routine MDCT acquisitions, we commonly visualize a considerable myocardial SD drop at the posterobasal wall resembling perfusion defects (Fig. 1), being attributed to beam hardening artifacts. We therefore explored the myocardial SD levels in asymptomatic patients without history of CAD and evaluated the impact of beam hardening artifacts.

Fig. 1
figure 1

Basal short axis at the left ventricular outflow tract (panel a), conventional basal short axis (panel b), and axial view (panel c) showing decreased myocardial density at the posterobasal wall (arrows) and normal attenuation at the remaining segments

Methods

The present was a single-center, investigator-driven, observational study, that involved consecutive asymptomatic patients without history of CAD and low probability of CAD who were referred for MDCT evaluation at our institution due to inconclusive or discordant functional tests. All patients included were >18 years old, in sinus rhythm, able to maintain a breath-hold for ≥15 s, without a history of contrast related allergy, renal failure, or haemodynamic instability. Patients with a pre-scan heart rate >65 bpm received beta-blockers either as a single oral dose or intravenously. In addition, patients with intrascan mild heart rhythm abnormalities leading to motion artifacts such as premature beats and heart rate <40 bpm were excluded. Scans were performed using a 64-channel MDCT scanner (Brilliance 64, Philips Healthcare, Cleveland, Ohio, USA). A bolus of 80–120 ml of iodinated contrast material (Optiray®, Ioversol 350 mg/ml, Mallinckrodt, St. Louis, U.S.A.) was injected through an arm vein at 5–6 ml/s. A bolus tracking technique was used to synchronize the arrival of contrast at the level of the coronary arteries with the start of acquisition. Scan parameters of the MDCT acquisitions were a collimation of 64 × 0.625 mm, rotation time 0.42 s, tube voltage 120 kV, and effective tube current-time product of 600–1,000 mAs corresponding to an approximate mean radiation dose of 12 mSv. An ECG-triggered dose modulation protocol (DoseRight Cardiac, Philips Healthcare, Cleveland, Ohio, USA) was applied to reduce radiation dose during systole whenever deemed possible by the operator [14], with approximate dose saving of 42% at a heart rate of 60 bpm, yielding an approximate mean effective radiation dose of 7 mSv in these patients. An ECG was recorded simultaneous to the CT scan to enable retrospective gating of the image data. A dedicated cardiac gating algorithm was used that identified the same physiological phases of the cardiac cycle while taking into account the non-linear changes in the individual cardiac states with the heart rate variations during the CT acquisition [15]. A cardiac adaptive multi-cycle (or multi-segment) reconstruction technique was used that combined data from consecutive cardiac cycles, thus significantly improving temporal resolution between 53 and 210 ms [16].

MDCT analysis

All analyses were performed by consensus of 2 experienced (>1,500 MDCT performed) observers using dedicated software (Comprehensive Cardiac Analysis, version 3.5), on a CT workstation (Brilliance Workspace, Philips Healthcare, Cleveland, Ohio, USA).

MDCT images were reconstructed at 75% of the cardiac phase using axial planes, multiplanar reconstructions, and maximum intensity projections at 1 mm slice thickness. Short axis views were obtained initially using 5-mm slice reformatted images at the basal, mid ventricle and apical levels. The posterobasal wall (Fig. 1), a basal segment not included in the American Heart Association (AHA) 17-segment model [17], was also evaluated at the short axis plane including the mitral valve and the left ventricular outflow tract (Fig. 1). Using standarized regions of interest of 20 mm2, myocardial signal density (SD) was determined for every segment according to the AHA 17-segment model [17] (Fig. 2). AHA segment 17 corresponding to the apical wall evaluated from the long axis was excluded from the analysis since it encompasses a thin myocardial wall and is therefore prone to measurement error. Left ventricular and right ventricular chamber mean SD were evaluated at basal, mid and apical short axis.

Fig. 2
figure 2

Myocardial signal density (above) and signal density ratio (below) levels of left ventricular myocardium segments showing decreased signal density of posterobasal (PB) segments

Definitions

“Perfusion defects” were defined as myocardium having a SD two standard deviations below the mean myocardial SD of the 16 AHA segments. Myocardial SD was evaluated at the posterobasal wall using the same approach. SD ratio, which is highly related to myocardial blood flow measured by microspheres [6], was determined as previously described: myocardial SD/left ventricular blood pool SD [6].

Myocardial SD and myocardial SD ratios were evaluated for the posterobasal segment and for AHA segments. In addition, myocardial SD and SD ratios were compared between patients with and without “perfusion defects” at the posterobasal wall.

In order to assess image quality and relate it to the presence of beam-hardening artifacts, we determined image noise and contrast-to-noise ratios and evaluated posible associations with posterobasal wall perfusion defects. Image noise was derived from the standard deviation of the SD values (in Hounsfield units) within a large region of interest in the left ventricle. The contrast-to-noise ratio was defined as the difference between the mean density of the contrast-filled left ventricular chamber and the mean density of the left ventricular wall, which was divided by image noise.

The study was approved by our Institution′s Ethics Committee, and all the patients enrolled gave their written informed consent.

Statistical analysis

Discrete variables are presented as counts and percentages. Continuous variables are presented as mean ± SD or median (25th, 75th percentile) as indicated. Comparisons among groups were performed using paired samples t-test, independent samples t-test, analysis of variance (ANOVA), chi square tests, Fisher′s exact test, or Mann–Whitney U tests as indicated. We explored correlations between the posterobasal wall SD and variables thought to be related to the presence beam hardening artifacts using Spearman correlation coefficients. A two-sided p value of less than 0.05 indicated statistical significance. Statistical analyses were performed with use of SPSS software, version 13.0 (Chicago, Illinois, USA).

Results

Thirty six asymptomatic patients constituted the study population. The mean age was 55.9 ± 10.6 years, 29 (81%) were male and 1 (3%) was diabetic.

Myocardial SD was evaluated in 576 AHA segments and 36 posterobasal segments. The mean myocardial SD at the posterobasal segment was 53.5 ± 35.1 HU, whereas the mean myocardial SD at the basal, mid and apical myocardium was 97.4 ± 17.3, with significant differences (p < 0.001) between posterobasal and all AHA segments (Table 1). Similarly, myocardial SD ratio at the posterobasal segment was 0.14 ± 0.09, whereas the SD ratio at the basal, mid and apical myocardium was 0.25 ± 0.06, with significant differences (p < 0.001) between posterobasal and all AHA segments (Table 2). With regard to the spatial distribution of SD, the septal segments (AHA 2, 3, 8, 9 and 14) had significantly higher myocardial SD than the other segments (Table 3).

Table 1 Myocardial signal density (Hounsfield units, HU) of left ventricular segments and posterobasal wall, and frequency of perfusion defects defined as myocardium having a signal density two standard deviations below the mean signal density of all left ventricular segments
Table 2 Myocardial signal density ratio of left ventricular segments and posterobasal wall, defined as myocardial signal density/left ventricular chamber signal density
Table 3 Spatial distribution of myocardial signal density

Posterobasal “perfusion defects”

Posterobasal “perfusion defects” at the short axis plane including the mitral valve and the left ventricular outflow tract, were identified in 26 (72%) patients (Fig. 1; Table 1). In addition, though infrequently, perfusion defects were identified at several AHA segments, particularly at segments 4, 5 and 13, corresponding to the inferobasal (segments 4 and 5) and anteroapical walls, with a prevalence of 11, 17 and 17%, respectively (Table 1). The only demographical variable associated to the presence of posterobasal “perfusion defects” was heart rate (61.8 ± 6.2 bpm vs. 56.3 ± 8.1 bpm, p = 0.04), whereas body mass index (BMI) was not related to the presence of perfusion defects (28.0 ± 3.7 kg/m [2] vs. 27.1 ± 4.6 kg/m [2], p = 0.55). Blood densities of the left and right ventricular chambers were not associated to the presence of posterobasal “perfusion defects” (Table 4). In addition, contrast-to-noise ratio and signal noise were not related to the presence of “perfusion defects” (Table 4). Regarding the presence and extent of coronary atherosclerosis, patients with posterobasal “perfusion defects” had similar Agatston calcium scores (0.0 (0.0-29.5) vs. 0.0 (0.0-259.0), p = 0.81), number of lesions (1.5 ± 2.6 vs. 1.7 ± 2.5, p = 0.81) and number of significant lesions (0.2 ± 0.6 vs. 0.4 ± 1.3, p = 0.63) than patients without posterobasal “perfusion defects”.

Table 4 Relationship between patient′s demographics, acquisition parameters, image quality, atherosclerosis and the presence of myocardial signal density deficit mimicking perfusion defects at the posterobasal wall

Finally, we did not find a significant relationship between posterobasal wall SD and heart rate (r = −0.16, p = 0.37), BMI (r = −0.12, p = 0.49), basal left ventricular chamber SD (r = 0.16, p = 0.35), basal right ventricular chamber SD (r = 0.16, p = 0.36), basal signal noise (r = −0.05, p = 0.79), basal contrast-to-noise ratio (r = 0.14, p = 0.42), and calcium score (r = 0.10, p = 0.55). In turn, a weak positive correlation was identified between posterobasal wall SD and descending aorta SD (r = 0.35, p = 0.04).

Discussion

For decades, myocardial perfusion imaging has been undoubtedly established as the gold-standard for prognosis and clinical decision making of patients with CAD, and has predominantly been assessed by single-positron emission tomography (SPECT) and, more recently, by positron emission tomography and magnetic resonance imaging (MRI).

After standing as an accurate tool to evaluate coronary stenosis in a non-invasive fashion, several non-coronary applications of MDCT have emerged and were tested with success, including myocardial function, viability and perfusion [811, 1720]. Among them, myocardial perfusion imaging by MDCT has been explored by a number of preclinical and clinical studies showing that hypoenhanced regions on contrast-enhanced MDCT correlate well to hypoperfused myocardial regions, becoming an accurate tool to evaluate myocardial infarction, with a good agreement with gated SPECT and MRI [611, 21, 22]. The pathophysiological basis of this concept is based on the kinetics of the iodinated contrast agent used for MDCT that, parallel to gadolinium-DTPA in contrast enhanced magnetic resonance, has a decreased inflow to the myocardium in the setting of a coronary stenosis, resulting in early hypoenhancement during contrast inflow [6, 23].

The present study is the first to establish the normal values of myocardial SD of left ventricular segments in an asymptomatic population, as well as SD values normalized according to the left ventricular chamber SD (SD ratio). Acquaintance of these normal values and identification of artifacts mimicking perfusion defects is essential to avoid misinterpretation of myocardial perfusion by MDCT. Our main findings can be summarized as follows:

  • Beam hardening artifacts are a common finding in MDCT-CA of asymptomatic patients, and affect predominantly the posterobasal wall appearing as myocardial perfusion defects.

  • “Perfusion defects” at the short axis plane including the mitral valve and the left ventricular outflow tract were not related to technical issues such as BMI, contrast in the aorta, right ventricle, contrast-to noise ratio or signal noise, whereas heart rate was associated with this finding.

  • Although very rarely, myocardial “perfusion defects” can be identified at the inferior and anteroapical segments of the left ventricular wall.

We identified a high prevalence of SD deficit at the posterobasal wall resembling perfusion defects. It should be stressed however, that SD deficit was commonly identified opposite to the left ventricular outflow tract, a segment not included in the AHA classification.

In addition, we found a trend towards lower myocardial SD, and even a few perfusion defects at the anteroapical and inferior (mid and basal) segments of the left ventricular wall. We ascribe this finding to beam-hardening artifact most probably from the spine for posterobasal segments and from the sternum for anteroapical segments that have an appearance similar to a myocardial perfusion deficit in contrast-enhanced coronary CTA images. The selective filtration of low-energy photons by highly attenuating cardiac structures, such as the contrast-enhanced left ventricle, the descending aorta, and occasionally the right ventricle, as well as bony structures such as the spine, sternum, and ribs, may create focal areas of nonphysiologic hypoenhancement in the myocardium. Since all these highly attenuating structures are aligned along the same x-ray path as these myocardial segments, we typically see worse beam hardening selectively in posterobasal segments. That being said, contrary to what it was expected, technical issues such as contrast in the right and left ventricular chambers, contrast-to-noise ratio, signal noise and BMI, as well as the presence and extent of coronary atherosclerosis, were not related to the presence of perfusion defects in our study. Nevertheless, patients with lower heart rate were more likely to have normal SD, although we did not find a significant relationship between posterobasal wall SD and heart rate.

Occurrence of attenuation artifacts during SPECT imaging has been considered an important limitation of the technique [24]. Our results indicate that attenuation artifacts also occur in MDCT perfusion imaging and should be accounted for in order to avoid being misinterpreted as perfusion defects.

Limitations

Although we included asymptomatic patients with no history of CAD and small, if any, atherosclerotic burden, confirmation of the absence of ischemia with SPECT or MRI was not performed. In addition we did not use 2-phase or 3-phase contrast injection protocols with saline to minimize right-heart contrast. Nevertheless, we did not find an association between the presence of contrast in the right ventricular chamber and perfusion defects.

Finally, image acquisition requires a number of heart beats leading to potential non-uniform distribution of contrast in myocardial segments. Nevertheless, blood signal density at both left and right ventricles (at basal, mid and apical levels), as well as at the ascending and descending aorta, did not differ between patients with and without SD deficit. In addition, acquisition time was similar in patients with and without “perfusion defects” therefore the number of cardiac cycles probably was not associated to the presence of defects.

Conclusions

In an asymptomatic population with no history of coronary artey disease, myocardial signal density deficit mimicking perfusion defects is a common finding at the posterobasal wall and is not related to body mass index or scan quality.