Introduction

Biodegradable polymer stents are currently being revisited as a solution to the adverse potential for delayed endothelialization and late thrombosis recently documented for first generation drug-eluting stents.12,15 While still providing mechanical support to prevent recoil and restenosis, biodegradable alternatives may also provide additional benefits, such as compatibility with noninvasive imaging technologies, stented segments suitable for future surgical revascularization, as well as positive remodeling and a physiological response to pulsatile blood flow.6,29

The purpose of this study was to develop a biodegradable polymer stent with high expansion speed, since quick delivery of intravascular stent systems is an important safety factor largely contributing to the success of vascular interventions. In particular coronary interventions require rapid balloon expansion protocols to avoid the risk of myocardial infarction or other adverse cardiac events.1 Similar requirements must be met in peripheral interventions, such as in carotid or neurovascular procedures, where rapid stent expansion is also required to prevent stroke, minimize ischemia, and to ensure immediate vascular scaffolding.

While high stent expansion speeds are part of everyday clinical practice with currently used metallic stents, both self-expanding and balloon-expandable, they are not necessarily feasible with biodegradable polymeric stents due to their viscoelastic material behavior.

In general, major challenges in the development process of polymeric stent concepts are being encountered in: (i) the selection of appropriate structural polymer materials, which must be biodegradable and biocompatible at the same time, (ii) the development of a stent design adapted to the polymeric material properties, and (iii) the identification of a stent expansion protocol adapted to the intended application site.

Practically, it is an immanent problem of polymeric materials, that their structural properties are inferior to those of metals. For this reason, the design of a polymeric stents with similar mechanical properties, such as low recoil and high radial strength, will require greater strut dimensions. To keep strut thickness as low as possible, polymeric stent designs will need to be highly optimized towards the polymeric material properties to take most advantage of them. Furthermore, polymeric stent deliverability may be restricted, for instance due to higher profile and longer expansion time, which may require modified delivery devices and predilation of target lesions.

Due to these hindrances only a small number of different biodegradable polymeric stent concepts have so far been investigated in clinical trials as an alternative to permanent stents. In 2000, Tamai et al., described the first clinical application of a self-expanding PLLA zig-zag fiber stent, which can be expanded in 0.5 min with a balloon catheter heated to 50 °C and inflated to 6–14 atm.23 A peripheral size version of this stent is being investigated in the PERSEUS clinical trials.2

Currently, the everolimus-eluting BVS stent (Abbott Vascular, Santa Clara, USA), a coronary polymer stent based on PLLA, is being investigated in the ABSORB clinical trial.19 Recent data on the in vivo performance of this stent, which has a strut thickness of 150 μm at a nominal diameter of 3.0 mm, is very promising and demonstrates an acute recoil comparable to a metallic stent.25

A larger number of groups have already demonstrated the high potential of their technology in the experimental stage. Lauto et al., have reported on experimental work on a self-expanding chitosan helical coil stent, fully expanding within 3 min.16 In a similar design approach, Vogt et al., developed a balloon-expandable helical coil stent from poly(d,l-lactide) (PDLLA), capable of being expanded in 1 min with balloon pressures of 10–12 atm.28 Venkatraman et al., are also pursuing another similar design approach using a self-expanding elastic memory helical coil stent based on PLLA and poly(l-lactide-co-glycolide) (PLGA), which self-expands in a few minutes.27 Su et al., have described a balloon-expandable fiber stent based on PLLA, which can be expanded in 1.5 min with balloon inflation to 10 bar.22 Not disclosing their exact expansion protocol, Uurto et al., have reported on their animal experience with a self-expanding fiber stent based on PDLLA, which is dilated with a balloon to support self-expansion.26 In the non-vascular field, Isotalo et al., have clinically investigated self-expanding urethral PLLA stents.14 Properties of this stent have recently been improved by changing the original spiral design (0.7 mm wire) to a braided mesh (0.35 mm wire).13

In contrast to the above mentioned fiber or coil based self-expanding or balloon-expandable polymer stent concepts, our group is pursuing the development of a balloon-expandable slotted tube stent concept.

While drug incorporation may be equally challenging, and self-expanding fiber stents may be safer in applications vulnerable to external mechanical impact, balloon-expandable slotted-tube concepts may be particularly advantageous where precise sizing and high radial strength are desired.

We have reported earlier on a PLLA stent that requires expansion over 8 min, which was successfully applied in a surgical anastomosis model in the porcine A. iliaca with a follow-up of 6 weeks.4,8

The present study describes in detail the mechanical properties and in vitro degradation of peripheral size (6.0 × 25 mm) slotted tube stents made of a novel PLLA/P4HB blend material, developed and fabricated by our group, that allow for rapid balloon-expansion.

Materials and methods

Polymer Materials and Stent Prototype Manufacture

The stents were manufactured from a polymer blend material of high molecular weight PLLA (Resomer® L214, Boehringer Ingelheim Pharma, Ingelheim, Germany) and P4HB (TephaFLEX®, Tepha, Inc., Cambridge, MA, USA), with a mass ratio of 78/22%. The novel blend component P4HB is a thermoplastic biopolymer produced in a recombinant fermentation process using E. coli K12. Its degradation product 4HB is a natural human metabolite present in several organs. P4HB is currently evaluated for a number of medical applications.17

In a first step, a 1.6% (w/v) polymer solution in chloroform was prepared to fabricate polymer mini-tubes with an inner diameter of 2.8 mm by dip-coating of stainless steel mandrels. The dip-coating procedure was performed in a repetitive, automated dipping process until a nominal wall thickness of 300 μm of the polymer tubes was achieved. The polymer tubes were then removed from the mandrels and washed in methanol and water for 24 h each to remove all solvent remnants.

In a second step, the polymer tubes were laser machined with a CO2 laser plotter (LP 1007, Wild Emco Lasertec, Hallein, Austria) to produce peripheral balloon-expandable slotted tube stent prototypes with nominal dimensions of 6.0 × 25 mm. The design pattern used for laser cutting has been reported on earlier by our group.8 Figure 1 shows a sample stent prototype.

Figure 1
figure 1

Balloon-expandable slotted tube PLLA/P4HB stent prototype in the unexpanded (top) and expanded state (bottom). Comparison of the two states illustrates the small stent foreshortening of <3%

Mechanical and Thermal Analysis

Mechanical properties of the tubular blanks were assessed at 37 °C in uniaxial tensile tests with a ZWICKI ZN 2.5 (Zwick, Ulm, Germany). The gauge length and crosshead speed were 20 mm and 40 mm/min, respectively.

Thermo-mechanical analysis was carried out on a Dynamic Mechanical Analyzer DMA 2980 (TA-Instruments, New Castle, Delaware, USA). The specimens were cooled and equilibrated at a temperature of −80 °C using a nitrogen gas cooling unit. Then temperature was ramped at a heating rate of 2 °C/min up to 120 °C. During heating, the specimens were oscillated with an amplitude of 25 μm at a frequency of 1 Hz.

Melting behavior was measured by differential scanning calorimetry (DSC) on a Mettler-Toledo Thermosystem FP900 with a DTA/DSC-sensor FP84 (Mettler-Toledo, Schwerzenbach, Switzerland). The heating rate was 5 °C/min. Degree of crystallinity was determined separately for the P4HB and PLLA phase of the blend, based on the specific heat of fusion known for 100% cystalline PLLA (93.7 J/g)7 and P4HB (76 J/g).18

Stent Bench Testing

The stents were mounted on standard 6.0 × 30 mm balloon catheters (Pheron, Biotronik GmbH & Co. KG, Berlin, Germany). Stent expansion was achieved within 1 min by steady balloon inflation at a rate of 8 bar/min to a balloon pressure of 8 bar. Prior to expansion, the stents were preconditioned in a 37 °C water bath for 5 min. Expansion and recoil measurements were conducted in a dedicated test chamber, described in detail in.21 Briefly, the stent systems were installed in a 37 °C water bath. A computer-controlled pump was used for balloon inflation. Stent profile during balloon inflation and stent recoil at balloon deflation were measured optically with a dual-axis laser measuring head (ODAC 32 XY, ZUMBACH Electronic AG, Orpund, Switzerland), and recorded with a PC.

Stent collapse was assessed after inserting dilated stent specimens into thin walled polyurethane tubes open to the ambient atmosphere. The stented tubes were then immersed in a 37 °C water filled pressure chamber according to.21 The hydraulic pressure in the chamber was raised in steps of 0.05 bar at a rate of 0.5 bar/min by a computer controlled piston pump. Simultaneously, the stent diameter was monitored optically in two orthogonal directions with a dual-axis laser measuring head (ODAC 32 XY, ZUMBACH Electronic AG, Orpund, Switzerland), and recorded with a PC.

Scanning Electron Microscopy

Expanded stent specimens were prepared for detailed analysis of strut deformation and surface morphology, and viewed using a Philips XL 30 ESEM (Philips Electron Optics, The Netherlands) scanning electron microscope in SEM mode.

In vitro Stent Degradation

Expanded stent specimens were incubated in Sørensen buffer solution (0.1 mol/l, pH = 7.4) at 37 °C to evaluate hydrolytic degradation in vitro. After 0/2/4/8/12/24 w, stent specimens were removed from storage and tested for their remaining collapse stability, as described in the collapse testing section. Pieces of collapsed stents were analyzed by gel permeation chromatography (GPC) to determine molecular weight. The molecular weight of stent material was plotted over 48 w.

The GPC system (Thermo Separation Products, San Jose, USA) utilized a Shodex RI 71 refractive index detector (Showa Denko K.K., Tokyo, Japan) and a viscosity detector η-1000 (WGE Dr. Bures GmbH & Co KG, Dallgow, Germany). The system further contained three PSS (Polymer Standards Service GmbH, Mainz, Germany) SDV 10 μm columns (103, 105, 106 Å) in series at 35 °C. Chloroform was used as solvent and eluant at a flow rate of 1 ml/min. The sample concentration was 1.0 mg/ml, and the injection volume was 100 μl. Results were obtained using the method of universal calibration, based on 12 narrow dispersed polystyrene standards (Polymer Standards Service GmbH, Mainz, Germany).

Data Analysis

Expansion, recoil, and collapse data were obtained on the basis of n ≥ 3 specimens tested for each measurement. Degradation data are based on the testing of n = 3 samples at each time point. Arithmetic mean and standard deviation of the mean values were calculated. Statistical analysis was conducted using the two-tailed t-test for independent samples with unequal variances. A p-level < 0.05 was considered statistically significant.

Results

Mechanical and Thermal Analysis

Comparison of the tensile properties of the PLLA/P4HB blend with the cited PLLA tensile data illustrates, that the addition of 22% P4HB to PLLA lowers the elastic modulus and the tensile strength by approx. 52% and 20%, respectively, see Table 1. However, elongation to break increases approximately 16-fold.

Table 1 Tensile test material data of solution cast tubular PLLA/P4HB (22% P4HB) specimens at 37 °C

Dynamic thermo-mechanical analysis confirms the lowered modulus of the blended material (Fig. 2), and also shows that addition of P4HB produces a second relaxation slope, as well as a shift of the glass transition range towards lower temperatures. Glass transition of the P4HB phase in the blended material was observed at −62 °C (onset of E′). The main relaxation of the blend was measured at 34 °C (onset of E′). Storage modulus E′ was 2.0 GPa at 20 °C, and 1.5 GPa at 37 °C. Relaxation of the pure PLLA material was measured at 60 °C (onset of E′), with E′ being 3.2 GPa (20 °C), and 3.1 GPa (37 °C).

Figure 2
figure 2

Modulus–temperature relationship for PLLA/P4HB and pure PLLA

The DSC data given in Table 2 indicate the marked reduction of crystallinity of the PLLA phase by approx. 49% in the blended material compared to the referenced crystallinity data for pure PLLA. Similarly, the crystallinity of the P4HB phase in the blend was reduced by approx. 83% compared to earlier studies where we determined a crystallinity of 53% for similarly processed pure P4HB film samples (unpublished data). There was, however, no lowered melting temperature of PLLA in the blended material, see Table 2.

Table 2 Melting behavior and crystallinity of solution cast tubular PLLA/P4HB specimens

Stent Expansion and Recoil

The stent expansion protocol, preconditioning of the stents for 5 min at 37 °C and subsequent balloon-inflation to 8 bar in 1 min, allowed for successful expansion of all tested stent specimens. All stents were expanded evenly with uniformly circular cross sections, and completely without the occurrence of strut ruptures. Figure 3 shows the stent expansion behavior during balloon pressurization and demonstrates that the PLLA/P4HB stents expand at balloon pressures between 1 and 2 bar and are fully expanded at nominal balloon pressure (6 bar). Figure 3 also indicates, that stent diameter can be further increased by over-expansion to 8 bar.

Figure 3
figure 3

Stent pressure-diameter behavior during balloon expansion. The curve reflects the stent diameter measured in the stent midsection

The well tolerated overexpansion of the stent design is also demonstrated by the electron microscopical analysis. It can be seen in Fig. 4, that CO2 laser cutting causes melting of the strut edges, which may contribute to the prevention of crack initiation in the highly deformed outer strut regions.

Figure 4
figure 4

Electron micrographs of an expanded stent specimen at 20× (a) and 150× (b). The images show in detail the melted strut edges caused by the CO2 laser cutting and the integrity of the strut deformation zones after stent expansion

The mean elastic recoil of the stents was 4.2 ± 2.5%, measured after deflation of the balloon-catheter from 8 bar.

Stent Collapse

Stent collapse measurements showed that external pressurization of the stents initially results in a uniform diameter reduction until just over 0.6 bar (see Fig. 5). Further increase of the external pressure subsequently leads to instability and induces buckling of the stent, resulting in a progressive ovalization of the cross section and ultimately stent collapse. Therefore, stent behavior under external pressure can be expressed in terms of the radial displacement parameters Δr 1 and Δr 2, as depicted in Fig. 5.

Figure 5
figure 5

Representative radial displacement curves of a stent under increasing external pressure. The inset is composed of superimposed images from numerical analysis of stent collapse, showing axial views of the stent in the expanded and collapsed configuration. It illustrates the ovalization of the stent under external pressure, which can be expressed in terms of the radial displacement parameters Δr 1 and Δr 2. Between 0.2 and 0.4 bar the curves show a typical artifact of the measurement which is created when the increasing outer load initiates an evasive rotational movement of the stent in the laser scanning field

Figure 5 also shows the representative radial displacement curves of a PLLA/P4HB stent under external pressure. The curves illustrate the uniform diameter reduction of the stent up to a pressure of approximately 0.6 bar, which can be interpreted as radial compliance of the stent. Above this pressure, the two displacement curves Δr 1 and Δr 2 bifurcate, and the stent starts to ovalize. The abrupt splitting of the curves at a pressure of 1.1 bar is an indicator of stent collapse.

The mean collapse pressure measured for the stents was 1.1 ± 0.25 bar.

In vitro Degradation

In the first phase of incubation in Sørensen buffer solution (pH = 7.4, 37 °C) for in vitro degradation, the stents exhibited an increase in collapse pressure. The maximum collapse pressures were obtained after 4 weeks, when the mean collapse pressure was 175% of the original value at the beginning of in vitro incubation. The collapse pressures subsequently decreased, but at 24 weeks they were still above the initial value, see Fig. 6.

Figure 6
figure 6

Stent collapse pressure as a function of incubation time in Sørensen buffer (pH = 7.4) at 37 °C

The GPC measurements showed a gradual and steady molecular weight decrease of the PLLA/P4HB blend material during the in vitro degradation study. The molecular weight at the beginning of in vitro incubation was determined as a single peak for the polymer blend with M w = 357,000 g/mol. After 48 weeks, the molecular weight of the blend material had decreased by 82% to approximately 64,000 g/mol, see Fig. 7. However, there was no mass loss apparent after this time.

Figure 7
figure 7

Molecular weight M w as a function of incubation time in Sørensen buffer (pH = 7.4) at 37 °C

Discussion

Safe clinical application of a biodegradable polymeric stent requires rapid expansion, low recoil, and high collapse pressure of the stent. The results from the mechanical bench tests show that the presented stent fulfils these demands. The stent can be expanded at moderate balloon pressures, and even exhibits safety in application in overexpansion situations, which commonly occur in clinical practice to adjust stent sizing.5 The described preconditioning period of 5 min in 37 °C water, however, does not actually fit in current clinical routines, and may need to be shortened.

Rapid stent expansion within 1 min, compared to 8 min for our previous PLLA stent, was made possible by the blending of PLLA with P4HB. Although further mechanical testing, addressing the non-linear material behavior, may be necessary to fully understand the effects of P4HB blending, the tensile tests and the thermo-mechanical analyses indicate, that this effect results from the increased ductility and lowered glass transition of the material.

The in vitro degradation tests show that the stents maintain their load bearing capacities for a time period of at least 24 weeks. The detected increase in collapse pressure can be attributed to recrystallization of the polymer material during in vitro incubation. We have also observed this with pure PLLA,8 and it has also been reported by Tan and Venkatraman et al., for PLGA.24

The steady molecular weight decrease of the PLLA/P4HB blend material during the in vitro incubation period of 48 weeks is more pronounced than with pure high molecular weight PLLA, for which we have observed a reduction of molecular weight of 28% after 24 w,8 and 55% after 48 w.9

Accelerated degradation of the PLLA/P4HB blend material can be attributed in part to the lower molecular weight P4HB blend component, the lowered glass transition and the lower degree of crystallinity of the polymer blend, as compared to pure PLLA. The blended material may also allow better water diffusion throughout the material and increase the rate of polymer hydrolysis.

Interestingly, the increase in stent collapse pressure during in vitro storage is progressing conversely to the observed decrease in molecular weight starting immediately after in vitro incubation. However, after 4 weeks of in vitro storage the progressive trend of the collapse pressure curves is reversed and follows the decline of molecular weight.

Conclusions

The low recoil, high collapse pressure and rapid balloon-expandability of the slotted-tube PLLA/P4HB stent described herein provide adequate mechanical characteristics for a broad range of peripheral vascular and non-vascular applications. In particular recoil (4.2%) and collapse pressure (1.1 bar) are comparable to those of commercial peripheral metallic stents, for which recoil values of 2.5–4.8%, and collapse pressures of 0.8–1.2 bar are known.20 These characteristics were achieved with a strut thickness of 300 μm, which compares to a range of 160–220 μm for current peripheral 6.0 mm balloon-expandable metallic stents.11

However, the key feature of the stent concept presented here is the fact that the PLLA/P4HB polymeric blend material provides a stent with both, load bearing capacity and the capability for rapid expansion, at the same time.

This is a pivotal advantage over our previous technology based solely on PLLA, and may ultimately permit the development of a coronary stent concept. Other vascular applications in reach are the treatment of renal, iliac or femoral artery stenoses, and stenoses of prosthetic arteriovenous grafts. Potential non-vascular applications may be, for instance, the treatment of urethral or biliary tract stenoses.

Important issues that will have to be addressed during further stent development are long-term degradation behavior (>1 year), fatigue testing under dynamic conditions, and the identification of appropriate sterilization protocols and options for drug incorporation. The in vivo performance and degradation of the stent will also need to be evaluated. Recently the PLLA/P4HB stent has been successfully used by our group as an anastomotic support stent in the porcine iliac artery. Stent patency and integrity was maintained during the follow-up period of 6 w.3