Abstract
The human eye optical model with variable-focus liquid lens based on Chinese human eye model is presented for the first time. The Varioptic Arctic 316 liquid lens is used as a substitute for the crystalline lens in the model. The dioptric power and the MTF of the human eye model with variable-focus liquid lens are analyzed theoretically. The result shows that the human eye model with the liquid lens under applied voltage of about 53 V can be used as an emmetropia model. This research work can provide a novel idea for suitable model for clinical regulation and control with Chinese eye.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
1 Introduction
A schematic eye model is helpful in teaching aids in optics, optometry, ophthalmology, psychology (vision and visual perception) and visual ergonomics (Smith 1995). Moreover, this can be used to understand and research the optical performances of eye in refractive surgical procedures, such as photorefractive keratectomy (PRK), laser-assisted in situ keratomileusis (LASIK) (Gobbi et al. 1999; Curatu et al. 2002; Guo et al. 2005) and intraocular lens implants (Norrby et al. 2007; International Organization for Standardization 1999; Rosales and Marcos 2007). In current research of the human eye optical system, the information about the human crystalline lens is sparse because of the difficulties of measurement and the complexity itself. In order to build a practical eye model that is more suitable for clinical regulation and control, the human eye model with the replacement of human crystalline lens based on variable-focus liquid lens is proposed in the paper.
Variable-focus liquid lens is a novel optical element, whose focal length is varied by changing the surface curvature or internal refractive index of the lens without mechanically moving the lens (Kong et al. 2016). The variable-focus liquid lens presents the simplest and the highest-quality performances, so the lens has been studied extensively for eye glasses, cameras, camcorders, projectors, as well as other machine vision (Ren and Shin-Tson 2005).
The liquid lens based on the electrowetting effect is highly regarded because of its outstanding performance (Peng et al. 2013), which can provide considerable power savings and eliminate wear associated with moving parts. So it is widely used in a large variety of application areas, such as mobile phones (Wippermann et al. 2007), surgical instruments (Kuiper and Hendriks 2004) and miniature cameras (Tsai et al. 2008). Currently, the French Varioptic Company and the Netherlands Institute of Philips have the mature technology of liquid lens, they have produced focusable liquid lens for mobile phone camera (Xin 2010). So the Varioptic Arctic 316 liquid lens which is a commercial variable-focus liquid lens (Varioptic, France) with good stability and controllability is used here.
Based on our previous work result on Chinese human eye model (Kong et al. 2009; Kong and Li 2014), a double-liquid variable-focus lens (the Varioptic Arctic 316 liquid lens) is placed in the human eye model in this work.
2 Chinese generic eye model with Arctic 316 liquid lens
The Arctic 316 liquid lens consists of two kinds of liquid (insulating liquid and conducting liquid) and two glass windows (Berge 2005). Based on the electrowetting effect (Berge 2005), the focal length of the double-liquid lens is varied by means of an external voltage, which can change the curvature radius of the liquid interface. The photograph and structure schematic diagram of Arctic 316 liquid lens with the external voltage of 30 V are shown in Fig. 1 (Model Arctic 2010). Figure 1b shows this liquid lens structure: Between two glass windows, two non-miscible liquids are trapped in a closed cell. One of the liquid (conducting liquid) is based on a water solution and thus it is conducting electricity. The other liquid (insulating liquid) is apolar and should be non-conducting (the oil phase). The natural interface between the liquids thus forms a natural diopter, due to the refractive-index difference of the two liquids (Berge 2005). The interfacial shape between the two liquids is controlled by the applied external voltage.
Based on the corresponding data of the optical characteristic parameters of Arctic 316 liquid lens surface and the applied voltage value (Model Arctic 2010), the human eye models with Arctic 316 liquid lens under different applied voltages are built by Zemax optical design software.
As shown in Fig. 2, in Chinese generic eye model (Kong et al. 2009; Kong and Li 2014), the variable-focus performance of the crystalline lens is simulated by the replacement of Arctic 316 liquid lens. In Fig. 2a, the optical element in the dashed box is Arctic 316, incident to the human eye model under applied voltage of 53 V, the light rays converge good at the last surface which is represented as the retina. From Fig. 2b–g, the applied voltages are 25, 30, 35, 40, 45 and 55 V, respectively. As the voltage increases, the curvature radius of the liquid interface (red solid line) changes from negative to positive gradually, the convergence of light rays increased gradually, so the human eye model changes from the form of hyperopia to the form of myopia gradually. The data of curvature radius of the liquid interface at different voltages are shown in Table 1.
Figure 3 shows how the dioptric powers of human eye models vary with the applied voltages. The dioptric power of human eye models increases gradually with the increase in the applied voltage. The dioptric power of the normal aging human eye is about 60D (Smith et al. 1992). So from Fig. 3, it is found that the human eye model under applied voltage of about 53 V can be used as an emmetropia model. Moreover, according to the different diopters of myopia and hyperopia, the corresponding human eye models can be chosen to represent different myopia and hyperopia models.
Figure 4 displays the monochromatic MTF of Chinese generic eye model with Arctic 316 liquid lens under applied voltage of 53 V (a) compared with Chinese generic eye model (Kong et al. 2009) for the pupil diameter of 3 mm. In Fig. 4a, the agreement of the MTF curves is good. From Fig. 4b–g, the applied voltages are 25, 30, 35, 40, 45 and 55 V, respectively; Chinese generic eye model is an emmetropia model (Kong et al. 2009), so there is no MTF curve comparison under these voltages. From Fig. 4, it is found that the MTF of eye model with liquid lens under applied voltage of 53 V is clearly better than the ones under other voltages. This result indicates that the imaging quality of the human eye model under applied voltage of about 53 V, which can be used as an emmetropia model, is better than the ones under other voltages, which can be used as myopia and hyperopia models. This result of the MTF analysis is consistent with the result of the previous dioptric power analysis of the human eye models.
3 Discussion and conclusion
Therefore, it is feasible that the liquid lens is used as a substitute for the crystalline lens in the human eye model. And this method to build the human eye model with variable-focus liquid lens can provide a novel idea for the practical human eye model for clinical regulation and control in the future.
Limited by the aperture size of the liquid lens (shown in Fig. 1), only the MTF analysis of the human eye model for 3-mm pupil diameter is presented here; in the future work, the liquid lens with large aperture size, which can be used for the human eye model with large pupil diameter such as 6 mm, will be studied.
On the other hand, in addition to the voltage-driven liquid lens in this paper, there are some other liquid lenses based on different working mechanisms, such as magnetic-driven liquid lens (Malouin et al. 2010), acoustics-driven liquid lens (López and Hirsa 2008) and pressure-driven liquid lens (Zhang et al. 2004). Therefore, the comparative study of liquid lenses with different driving methods as a substitute for the crystalline lens in the human eye model would be a very good research direction, and we will do more work in this direction in the future.
References
Berge B (2005) Liquid lens technology: principle of electrowetting based lenses and applications to imaging. In: Micro electro mechanical systems, 2005. MEMS 2005. 18th IEEE international conference on IEEE, p 227
Curatu EO, Pettit GH, Campin JA (2002) Customized schematic eye model for refraction correction design based on ocular wavefront and corneal topography measurements. Proc SPIE 4611:165
Gobbi PG, Carones F, Brancato R (1999) Optical eye model for photo-refractive surgery evaluation. Proc SPIE 3591:10
Guo HQ, Wang ZQ, Wang Y, Zhao QL, Wang Y (2005) A new method to calculate corneal ablation depth based on optical individual eye model. Optik 116:433
International Organization for Standardization (1999) ISO 11979-2, Ophthalmic implants—intraocular lenses—part 2: optical properties and test methods
Kong M, Li Z (2014) Research of bright pupil effect based on Chinese human eye model. Chin Opt Lett 12(11):113301
Kong M, Gao Z, Li X, Ding S, Xiao-mei Q, Mei-qun Y (2009) A generic eye model by reverse building based on Chinese population. Opt Express 17(16):13283
Kong M, Zhu L, Chen D, Liang Z, Zhao R, Enming X (2016) Effect of the liquid density difference on interface shape of double-liquid lens. J Opt Soc Korea 20(3):427
Kuiper S, Hendriks BHW (2004) Variable-focus liquid lens for miniature cameras. Appl Phys Lett 85(7):1128–1130
López CA, Hirsa AH (2008) Fast focusing using a pinned-contact oscillating liquid lens. Nat Photon 2(2):610
Malouin B, Olles J, Cheng L, Hirsa A, Vogel M (2010) Electromagnetic liquid pistons for capillarity-based pumping. Lab Chip 11(3):393
Model Arctic (2010) Model arctic 316 liquid lens technical data sheet: products (Varioptic, SA, 2010)
Norrby S, Piers P, Campbell C, Mooren M (2007) Model eyes for evaluation of intraocular lenses. Appl Opt 46:6595
Peng R, Wang D, Hu Z, Chen J, Zhuang S (2013) Focal length hysteresis of a double-liquid lens based on electrowetting. J Opt 15(2):205–216
Ren H, Shin-Tson W (2005) Variable-focus liquid lens by changing aperture. Appl Phys Lett 86:211107
Rosales P, Marcos S (2007) Customized computer models of eyes with intraocular lenses. Opt Express 15:2204
Smith G (1995) Schematic eyes: history, description and applications. Clin Exp Optom 78:176
Smith G, Atchison DA, Pierscionek BK (1992) Modeling the power of the aging human eye. J Opt Soc Am A 9(12):2111
Tsai FS, Cho SH, Lo YH, Vasko B, Vasko J (2008) Miniaturized universal imaging device using fluidic lens. Opt Lett 33(3):291–293
Wippermann FC, Schreiber P, Bräuer A, Craen P (2007) Bifocal liquid lens zoom objective for mobile phone applications. Proc SPIE 6501:650109–650109-9
Xin Q (2010) Trends in liquid lens. Opt Instrum 32(6):70 (in Chinese)
Zhang DY, Justis N, Lo YH (2004) Fluidic adaptive lens of transformable lens type. Appl Phys Lett 84(21):4194
Acknowledgement
This research is supported in part by the National Natural Science Foundation (No. 61302026), the Natural Science Foundation of the Jiangsu Higher Education Institutions (No. 13KJB510025), Research Center of Optical Communications Engineering & Technology (No. ZSF0402), Jiangsu Province. We thank Manager Shao of Shenzhen Brilliant Optics Co., Ltd., for his help with the model Arctic 316 liquid lens technical data.
Author information
Authors and Affiliations
Corresponding author
Additional information
This article is part of the topical collection “2016 International Conference of Microfluidics, Nanofluidics and Lab-on-a-Chip, Dalian, China” guest edited by Chun Yang, Carolyn Ren and Xiangchun Xuan.
Rights and permissions
About this article
Cite this article
Kong, M., Chen, D., Chen, X. et al. Research of the human eye model with variable-focus liquid lens. Microfluid Nanofluid 21, 40 (2017). https://doi.org/10.1007/s10404-017-1857-z
Received:
Accepted:
Published:
DOI: https://doi.org/10.1007/s10404-017-1857-z