Abstract
In this study, an inexpensive and robust microfluidic droplet manipulation device based on gravity-actuated technique is described. The device is mainly composed of a turntable and some disposable infusion sets. The movement of oil-phase and water-phase in the microchannel is controllable. Droplet volume of different sizes is precisely generated by controlling droplet length using the turntable. A series of essential functional units have been demonstrated including on-demand sampling, droplets generation of controllable size and controllable number, droplets fusion of different sizes and contents, even droplets-stopping. These functions can well meet the requirements of routine droplet-based microfluidic biological/chemical assays, cell research, and high content drug screening. The presented technique shows unique advantages in inexpensiveness, flexibility, and portability when compared with general droplet manipulation systems containing precision syringe pump and complicated micromechanical/pneumatic valve. Furthermore, this device can be used as a portable droplet manipulation tool in outdoor and remote areas, because it does not need electrical source and is very facile to set up in any place.
Avoid common mistakes on your manuscript.
1 Introduction
Droplet-based microfluidic systems have attracted increasing attention in the past 10 years due to its remarkable advantages in miniaturized bio-analytical and chemical field (Song et al. 2006; Fair 2007, Teh et al. 2008; Huebner et al. 2008a, b; Zhang et al. 2008, Chiu et al. 2009). A series of droplet-based microfluidics successful applications have been reported, including chemical reaction (Cygan et al. 2005), synthesis of micro- and nano-particles (Tan and Takeuchi 2007; Rondeau and Cooper-White 2008; Um et al. 2008; Prasad et al. 2009), cell sorting (Baret et al. 2009) and assay (Clausell-Tormos et al. 2008), enzyme assay (Srinivasan et al. 2004; Huebner et al. 2008a, b; Wang et al. 2009a, b), and drug discovery (Zheng et al. 2003; Dittrich and Manz 2006). All the above applications are based on the mature droplet manipulation technique, therefore it is crucial whether droplets are precisely manipulated or not.
Essential droplet manipulation technique mainly includes droplet generation, transport, fusion, and mixing. Usually precise syringe pump and syringe needle are necessary for droplet generation by T-junction (Garstecki et al. 2006) or flow-focusing configuration (Anna et al. 2003). Although the method shows extremely good capability in obtaining monodisperse droplets (Thoren et al. 2001; Yobas et al. 2006), it lacks flexibility in the process of droplet manipulation. Therefore, a series of styles, including passive and active, for enhancing flexible performance have been developed. According to the principles of fluid dynamics, microchannel structures with specific function have been successfully designed and fabricated for the control of droplet fission, fusion, mixing, and sorting (Song et al. 2003; Tan et al. 2004; Tan et al. 2008; Maenaka et al. 2008). The main shortage of passive style chips is difficult to integrate multiple functional units into a chip although relatively easy to fabricate. For the active style chips, droplets are manipulated by external force mainly including valve actuation force (Lin and Su 2008) and electrical force (Velev et al. 2003; Xu and Attinger 2008). Microvalve is a highly efficient tool to precisely and flexibly control the generation, size, composition, and fusion of droplets in microfluidic devices (Sassa et al. 2008; Lee et al. 2009; Zeng et al. 2009). Although it does not require the precise control of fluid velocity, syringe pumps controlled by computers are still necessary. Another important approach for flexible droplet manipulation is to integrate microelectrodes into chips (Zeng and Korsmeyer 2004), including dielectrophoresis (DEP) (Jones et al. 2001; Gascoyne et al. 2004) and electrowetting on dielectric (EWOD) (Moon et al. 2002; Cho et al. 2003). The size and uniformity of the droplets depend on the magnitude and the frequency of the applied voltage. By the technique, some essential functional units have been demonstrated including creating, transporting, cutting, and merging. However, the complexity of chip design and fabrication are greatly increased by the integration of electronic components.
In addition to the methods mentioned above, other active droplet manipulation techniques have also been developed including magnetic force (Shikida et al. 2006; Pipper et al. 2007; Zhang et al. 2009), optical force (Chiou et al. 2003), surface acoustic wave force (Franke et al. 2009), and thermocapillary force (Darhuber et al. 2003; Glockner and Naterer 2005). Most of them are either lack in enough flexibility to meet diverse requirements of droplet manipulation, or make chips too complicated to fabricate for common laboratories due to integrating micromechanical/pneumatic valves and electronic components into devices. Furthermore, outdoor operation and application are greatly limited by bulky and expensive external control facilities.
Gravity is economical and easy to use (Yamada et al. 2008), which provides an alternative method for droplet manipulation. However, for most of published papers in relation to droplet manipulation, playing as a driving force, gravity is just like a pump (Huh et al. 2007) and fluid flow velocity is difficult to adjust (Liu et al. 2008). These passive droplet manipulation methods lack independence and flexibility. Herein, we first systematically present a facile and robust gravity-actuated technique for active flexible microfluidic droplet manipulation. The overall process, from sampling to on-demand droplets generation, transport, collision, fusion, mixing, and stopping, is accomplished by the technique. The necessary driving force of promoting fluids movement in the microchannel is provided by the gravity of oil-phase (silicon oil) stored in a reservoir. And all the reservoirs are hanged on both the sides of a turntable which is used to adjust the height difference between reservoirs and chips. The main advantages of this system are the capabilities for meeting the requirements of flexible droplet manipulation as well as completely eliminating the demand of electrical resource, complicated fabricate techniques, and bulky external control facilities.
2 Experimental section
Standard PDMS (Sylgard 184 polydimethysioxane silicone elastomer, Dow Corning, USA) fabrication techniques were used to fabricate microchips (Duffy et al. 1998) and all the depth of microchannels were 100 μm. Before use, the chips were left at 100°C for 2 h to ensure full recovery of the PDMS hydrophobicity. Dimethylsilicon oil (with dynamic viscosity at room temperature of 35 mPa s and a density of 950 kg m−3, purchased from Hangping, China) as oil-phase was filtered through a 0.22-μm Nylon membrane filter (Jinteng, China), and HP Inkjet inks of different colors (viscosity 2 mPa s) were used as water-phase. No surfactant was used in the experiment. The necessary gravity was provided through disposable infusion sets (each length 30 cm, OD 4 mm, and ID 3 mm, provided by hospital) filled with oil-phase. PTFE tubings (each length 5 cm, OD 1.4 mm, and ID 0.6 mm, purchased from Xingchen, China) used to store water-phase (samples/reagents) were inserted directly into the PDMS holes connected with the channels. The diameter of PTFE tube was slightly larger than that of PDMS hole (1.2 mm). Since PDMS has excellent elasticity, no leakage phenomenon is observed. Images were obtained and analyzed using a CCD (Daheng, China).
3 Results and discussion
Precision syringe pumps and syringes are commonly considered to be the essential tools in microfluidic-based droplet generation. Both sampling and droplets generation are accomplished only by gravity through the technique, which completely eliminates the use of any syringe pump/syringe and also further reduces the demand of samples/reagents volume. Oil-phase is first loaded into disposable infusion sets, and then water-phase (samples/reagents) is pumped and stored into PTFE tubing by utilizing the gravity of oil-phase (see Fig. 1a). The method of sampling is easy to operate and the consumption of samples/reagents volume is extremely decreased by reducing the diameter of the PTFE tubing (typical volume consumption between 1 and 30 μl). Usually a drop of sample/reagent (~20 μl) is enough in the experiment. After sampling, the two reservoirs are hanged on both the sides of the turntable (diameter 7 cm) with the same length of Nylon line (each 10 cm, see Fig. 1b). The turntable fixed on a plastic board can be smoothly turned to left/right manually with a screwdriver. Each inlet is manipulated by individual reservoir.
The main principle of this technique is that the size of fluid pressures (P, Pa), including oil-phase (P O) and water-phase (P W), at the T-junction can be varied quickly and precisely by turning the turntable to left/right. When water-phase height difference (Δh W, m) is higher than that of oil-phase (Δh O, m), the droplets are generated. We call this state switch-on. According to the Pascal’s expression:
where ρ is the fluid density (kg m−3), g is the gravity acceleration (m s−2), and Δh is the height difference between the reservoir and channel (m). P is proportional to Δh. Therefore, P W > P O (or \( {\frac{{P_{\text{W}} }}{{P_{\text{O}} }}} = {\frac{{\Updelta h_{\text{W}} }}{{\Updelta h_{\text{O}} }}} > 1 \)) represents the state of switch-on, namely droplet generation, and P W < P O (or \( {\frac{{P_{\text{W}} }}{{P_{\text{O}} }}} = {\frac{{\Updelta h_{\text{W}} }}{{\Updelta h_{\text{O}} }}} < 1 \)) represents the state of switch-off, namely no droplet generation (see Fig. 1d).
The sizes of droplets by this method are closely related to total flow velocity in the main channel (v T, m s−1), time period of the state of switch-on (t on, s) and the size of \( {\frac{{P_{\text{W}} }}{{P_{\text{O}} }}} \). However, the volume of each droplet (V droplet, m3) is difficult to be precisely controlled by managing t on and \( {\frac{{P_{\text{W}} }}{{P_{\text{O}} }}} \) simultaneously unless v T matches well with the turntable sensitivity. Flow velocity (v, m s−1) is described by Poiseuille equation (Morier et al. 2004; Liu et al. 2008):
where R is the radius of fluid channel (m), η is the fluid viscosity (Pa s), and l is the length of fluid channel (m). Combining Pascal and Poiseille equations:
We can see from the Eq. 3 that v is proportional to Δh and R 2 but inversely proportional to η. However, for a given liquid and a given microfluidic chip, the effects of variation of R and η on the experiments can be ignored because PDMS channels are perfectly wetted by oil-phase. Therefore, it should be possible to control v by manipulating Δh. When the turntable is fixed at a height, the total height difference (Δh T, Δh T = Δh O + Δh W) is always constant. So v T (v T = v O + v W) is also constant whether the state is switch-on or switch-off. The phenomenon has been verified by observing droplets movement. In the experiment, the typical v T ranges from 0 (called droplet-stopping v T) to 103 μm s−1 (called droplet transport v T), and corresponding range of Δh T and range of total fluid pressure (P T, P T = P W + P O) are 0–40 cm and 0–104 Pa, respectively. However, in order to precisely control droplet length in the channel (L droplet, m), a relatively low flow velocity range (called droplet generation, v T, 1–300 μm s−1) is selected for droplet generation. Within the range of droplet generation v T, the turntable has a good sensitivity and L droplet can be precisely controlled with the help of the standard scale. According to the theoretical analysis (Eq. 4), we find that V droplet can be quantitatively expressed by L droplet no matter what the state of droplet is “slug” or “plug” (see Fig. 2):
where H is the depth of microchannel (m). Therefore, V droplet of different sizes can be precisely generated by the control of L droplet using the turntable, and turning angle (θ) and t on do not require repeatable manipulation during droplet generation. Typical θ of droplet generation ranges from −45° (\( {\frac{{P_{\text{W}} }}{{P_{\text{O}} }}} \) = 0.76, no droplet generation) to +45° (\( {\frac{{P_{\text{W}} }}{{P_{\text{O}} }}} \) = 1.32, droplet generation). Considering the combined effect of t on and θ, the actual droplet generation frequency ranges from 0 to 30 droplets per minute and V droplet ranges from picoliter (10−12 l) to microliter (10−6 l). The approach for droplet manipulation is simple, flexible, and reproducible (RSD < 5%). Moreover, no unwanted satellite droplets (Xu and Attinger 2008) are generated due to low values of the Capillary number (Ca = vη/γ, 0–3.75 × 10−4, dimensionless) and low values of the Reynolds number (Re = vlρ/η, 0–8 × 10−4, dimensionless). Where v is the flow velocity in the channel (0–10−4 m s−1), η is the fluid viscosity (35 × 10−3 Pa s), γ is the surface tension at the water-phase/oil-phase interface (20 × 10−3 N m−1), l is the diameter of fluid channel (10−4 m), and ρ is the fluid density (0.95 × 103 kg m−3).
Monodisperse droplets of the same length are obtained (see Fig. 3a; movie 1). The droplet generation technique is not only very simple but also flexible. On-demand individual droplets can be continuously generated, including alternate droplets of different sizes (see Fig. 3b; movie 2) and droplets of any size (see Fig. 3c; movie 3). As the state of switch-on/off can be varied easily and precisely, the number and spacing of droplets can also be well controlled to meet the requirements of a wide range of experiments. Moreover, droplets-stopping (at droplet-stopping v T) at designated position in the microchannel is achieved by reducing the height of turntable (see movie 4), which allows in situ monitoring, real-time detection, and study of the biological/chemical assays process without need of a high speed camera (Wang et al. 2009a, b).
Droplets fusion is very important for droplet-based microfluidic system. Therefore, in addition to flexible way to generate individual droplets, as a practical droplet manipulation technique it should have the capability for highly efficient fusing two or more droplets of different reagents/samples. The gravity-actuated technique demonstrates a good ability in this aspect.
All four reservoirs are hanged on both the sides of the turntable and two of them connected with water-phase (1 and 2) are separated (see Fig. 4a). Microfluidic chip 2 of four inlets is fabricated and the connection between chip and tubings is showed in Fig. 4b. In order to clearly observe the fusion process of two droplets, red and yellow ink solutions are used to represent the different reagents/samples, respectively. No droplet is generated when the system is in the balance (namely the state of switch-off), i.e., P O1 > P W1 and P O2 > P W2. Red and yellow droplets are generated in sequence by turning the turntable to right firstly (P W1 > P O1, P O2 > P W2, only red droplet generation, see Fig. 4cII) and then left (P W2 > P O2, P O1 > P W1, only yellow droplet generation, see Fig. 4cIII). In order to ensure synchronous arrival of the two droplets to the microchannel expansion, P 2 (P 2 = P O2 + P W2) should be slightly larger than P 1 (P 1 = P O1 + P W1). P is inversely proportional to the length of Nylon line (l N) which is easily implemented by adjusting connected circle, therefore l N2 (l N2 = l NO2 + l NW2) is supposed to be shorter than l N1 (l N1 = l NO1 + l NW1). The experimental results demonstrate that when the total difference of l N is about 4 cm (namely l NO1 = l NO2 + 2 and l NW1 = l NW2 + 2), the two droplets almost simultaneously reach the expansion. The oval-shaped microchannel expansion (1000 μm in length and 400 μm in width, see Fig. 4c) is specifically designed to increase the possibility of two droplets fusion. Since the shape changes from “plug” to “slug” and the flow velocity reduces, the opportunities of droplets contact with droplets increase when the droplets flow into the expansion. As no surfactant is added into the oil-phase, the two droplets will highly efficient fuse when they collide with together (see Fig. 4c; movie 5). The variation of fluid pressures has been listed in detail (see Fig. 4d).
The various mixed droplets of different concentration gradients are generated because the size of each droplet is precisely controllable. Here, blue ink and yellow ink are used to represent different reagents/samples, respectively. Seven kinds of mixed droplets of different length ratios (L Blue/L Yellow) are continuously obtained, including 1:3, 1:2, 2:3, 1:1, 3:2, 2:1, and 3:1 (see Fig. 5). The described method is more simple and flexible when compared with other approaches generating droplets of concentration gradients (Lorenz et al. 2008; Damean et al. 2009). Moreover, the generation of array droplets of different compositions and sizes can also be simply implemented by our technique, which might promote the development of a cartridge approach so as to perform screening experiments (Li et al. 2007; Zeng et al. 2009).
As an alternative active droplet manipulation approach, the gravity-actuated technique offers obvious advantages in flexibility, simplicity, economy, and portability. However, limitations still exist in this technology. First, fast droplet generation is relatively difficult within the range of current turntable sensitivity. Secondly, owing to manual operation, it is hard to generate droplets for a long span of time. The two limitations may be easily resolved using a small motor (less than 10 dollars) which can precisely control turntable to-and-fro movement. Future work would include measures to increase the sensitivity of the system and also the use of real biological samples.
4 Conclusions
In summary, we have systematically presented a facile, flexible, and robust gravity-actuated technique for microfluidic droplet manipulation. The total cost of the device is less than 1 dollar and a series of essential functional units have been demonstrated by our technique, including gravity-based sampling, droplets generation of controllable size and controllable number, on-demand fusion of two droplets containing different contents, even droplets stopping for end-point detection. All these can meet the requirements of routine microfluidic biological/chemical assays, cell research and high content drug screening. Since the overall system is simple to set up and related materials are very easy to obtain without involving any syringe pump/syringe, bulky external control facility or complicated fabrication technique, the device can also be used as a portable tool which may accelerate the application of droplet-based microfluidic systems in outdoor and remote areas.
References
Anna SL, Bontoux N, Stone HA (2003) Formation of dispersions using “flow focusing” in microchannels. App Phys Lett 82:364–366
Baret JC, Miller OJ, Taly V, Ryckelynck M, EI-Harrak A, Frenz L, Rick C, Samuels ML, Hutchison JB, Agresti JJ, Link DR, Weitz DA, Griffiths AD (2009) Fluorescence-activated droplet sorting (FADS): efficient microfluidic cell sorting based on enzymatic activity. Lab Chip 9:1850–1858
Chiou PY, Moon HJ, Toshiyoshi H, Kim CJ, Wu MC (2003) Light actuation of liquid by optoelectrowetting. Sens Actuators A 14:222–228
Chiu DT, Lorenz RM, Jeffries GDM (2009) Droplets for ultrasmall-volume analysis. Anal Chem 81:5111–5118
Cho SK, Moon HJ, Kim CJ (2003) Creating, transporting, cutting, and merging liquid droplets by electrowetting-based actuation for digital microfluidic circuits. J Microelectromech Syst 12:70–80
Clausell-Tormos J, Lieber D, Baret JC, EI-Harrak A, Miller OJ, Frenz L, Blouwolff J, Humphry KJ, Koster S, Duan H, Holtze C, Weitz DA, Griffiths AD, Merten CA (2008) Droplet-based microfluidic platforms for the encapsulation and screening of mammalian cells and multicellular organisms. Chem Biol 15:427–437
Cygan ZT, Cabral JT, Beers KL, Amis EJ (2005) Microfluidic platform for the generation of organic-phase microreactors. Langmuir 21:3629–3634
Damean N, Olguin LF, Hollfelder F, Abell C, Huck WTS (2009) Simultaneous measurement of reaction in microdroplets filled by concentration gradients. Lab Chip 9:1707–1713
Darhuber AA, Valentino JP, Troian SM, Wagner S (2003) Thermocapillary actuation of droplets on chemically patterned surfaces by programmable microheater arrays. J Microelectromech Syst 12:873–879
Dittrich PS, Manz A (2006) Lab-on-a-chip: microfluidics in drug discovery. Nature 5:210–218
Duffy DC, McDonald JC, Schueller OJA, Whitesides GM (1998) Rapid prototyping of microfluidic systems in poly(dimethylsioxane). Anal Chem 70:4974–4984
Fair RB (2007) Digital microfluidics: is a true lab-on-a-chip possible? Microfluid Nanofluid 3:245–281
Franke T, Abate AR, Weitz DA, Wixfort A (2009) Surface acoustic wave (SAW) directed droplet flow in microfluidics for PDMS devices. Lab Chip 9:2625–2627
Garstecki P, Fuerstman MJ, Stone HA, Whitesides GM (2006) Formation of droplets and bubbles in a microfluidic T-junction-scaling and mechanism of break-up. Lab Chip 6:437–446
Gascoyne PRC, Vykoukal JV, Schwartz JA, Anderson TJ, Vykoukal DM, Current KW, McConaghy C, Becker FF, Andrews C (2004) Dielectrophoresis-based programmable fluidic processors. Lab Chip 4:299–309
Glockner PS, Naterer GF (2005) Thermocapillary control of microfluidic transport with a stationary cyclic heat source. J Micromech Microeng 15:2216–2229
Huebner A, Olguin LF, Bratton D, Whyte G, Huck WTS, deMello AJ, Edel JB, Abell C, Hollfelder F (2008a) Development of quantitative cell-based enzyme assays in microdroplets. Anal Chem 80:3890–3896
Huebner A, Sharma S, Art MS, Hollfelder F, Edel JB, deMello AJ (2008b) Microdroplets: a sea of applications? Lab Chip 8:1244–1254
Huh D, Bahng JH, Ling YB, Wei H-H, Kripfgans OD, Fowlkes JB, Grotberg JB, Takayama S (2007) Gravity-driven microfluidic particle sorting device with hydrodynamic separation amplification. Anal Chem 79:1369–1376
Jones TB, Gunji M, Washizu M, Feldman MJ (2001) Dielectrophoretic liquid actuation and nanodroplet formation. J Appl Phys 89:1441–1448
Lee WS, Jambovane S, Kim D, Hong JW (2009) Predictive model on micro droplet generation through mechanical cutting. Microfluid Nanofluid 7:431–438
Li L, Boedicker JQ, Ismagilov RF (2007) Using a multijunction microfluidic device to inject substrate into an array of preformed plugs without cross-contamination: comparing theory and experiments. Anal Chem 79:2756–2761
Lin BC, Su YC (2008) On-demand liquid-in-liquid droplet metering and fusion utilizing pneumatically actuated membrane valves. J Micromech Microeng 18:115005
Liu JJ, Lin JM, Knopp D (2008) Using a circular groove surrounded inlet to generate monodisperse droplets inside a microfluidic chip in a gravity-driven manner. J Micromech Microeng 18:095014
Lorenz RM, Fiorini GS, Jeffries GDM, Lim DSW, He MY, Chiu DT (2008) Simultaneous generation of multiple aqueous droplets in a microfluidic device. Anal Chim Acta 630:124–130
Maenaka H, Yamada M, Yasuda M, Seki M (2008) Continuous and size-dependent sorting of emulsion droplets using hydrodynamics in pinched microchannels. Langmuir 24:4405–4410
Moon HJ, Cho SK, Garrell RL, Kim CJ (2002) Low voltage electrowetting-on-dielectric. J Appl Phys 92:4080–4087
Morier P, Vollet C, Michel PE, Reymond F, Rossier JS (2004) Gravity-induced convective flow in microfluidic systems: electrochemical characterization and application to enzyme-linked immunosorbent assay tests. Electrophoresis 25:3761–3768
Pipper J, Inoue M, Ng LFP, Neuzil P, Zhang Y, Novak L (2007) Catching bird flu in a droplet. Nat Med 113:1259–1263
Prasad N, Perumal J, Choi CH, Lee CS, Kim DP (2009) Generation of monodisperse inorganic-organic Janus microspheres in a microfluidic device. Adv Funct Mater 19:1656–1662
Rondeau E, Cooper-White JJ (2008) Biopolymer microparticle and nanoparticle formation within a microfluidic device. Langmuir 24:6937–6945
Sassa F, Fukuda J, Suzuki H (2008) Microprocessing of liquid plugs for bio/chemical analyses. Anal Chem 80:6206–6213
Shikida M, Takayanagi K, Inouchi K, Honda H, Sato K (2006) Using wettability and interfacial tension to handle droplets of magnetic beads in a micro-chemical-analysis system. Sens Actuators B 113:563–569
Song H, Tice JD, Ismagilov RF (2003) A microfluidic system for controlling reaction networks in time. Angew Chem Int Ed 42:767–772
Song H, Chen DL, Ismagilov RF (2006) Reactions in droplets in microfluidic channels. Angew Chem Int Ed 45:7336–7356
Srinivasan V, Pamula VK, Fair RB (2004) Droplet-based microfluidic lab-on-a-chip for glucose detection. Anal Chim Acta 507:145–150
Tan WH, Takeuchi S (2007) Monodisperse alginate hydrogel microbeads for cell encapsulation. Adv Mater 19:2696–2701
Tan YC, Fisher JS, Lee AI, Cristini V, Lee AP (2004) Design of microfluidic channel geometries for the control of droplet volume, chemical concentration, and sorting. Lab Chip 4:292–298
Tan YC, Ho YL, Lee AP (2008) Microfluidic sorting of droplets by size. Microfluid Nanofluid 4:343–348
Teh SY, Lin R, Hung LH, Lee AP (2008) Droplet microfluidics. Lab Chip 8:198–220
Thoren T, Roberts RW, Arnold FH, Quake SR (2001) Dynamic pattern formation in a vesicle-generating microfluidic device. Phys Rev Lett 86:4163–4166
Um E, Lee DS, Pyo HB, Park JK (2008) Continuous generation of hydrogel beads and encapsulation of biological materials using a microfluidic droplet-merging channel. Microfluid Nanofluid 5:541–549
Velev OD, Prevo BG, Bhatt KH (2003) On-chip manipulation of free droplets. Nature 426:515–516
Wang JB, Zhou Y, Qiu HW, Huang H, Sun CH, Xi JZ, Huang YY (2009a) A chip-to-chip nanoliter microfluidic dispenser. Lab Chip 9:1831–1835
Wang W, Yang C, Li CM (2009b) On-demand microfluidic droplet trapping and fusion for on-chip static droplet assays. Lab Chip 9:1504–1506
Xu J, Attinger D (2008) Drop on demand in a microfluidic chip. J Micromech Microeng 18:065020
Yamada H, Yoshida Y, Terada N, Hagihara S, Komatsu T, Terasawa A (2008) Fabrication of gravity-driven microfluidic device. Rev Sci Instrum 79:1–124301
Yobas L, Martens S, Ong WL, Ranganathan N (2006) High-performance flow-focusing geometry for spontaneous generation of monodispersed droplet. Lab Chip 6:1073–1079
Zeng J, Korsmeyer T (2004) Principles of droplet electrohydrodynamics for lab-on-a-chip. Lab Chip 4:265–277
Zeng SJ, Li BW, Su XO, Qin JH, Lin BC (2009) Microvalve-actuated precise control of individual droplets in microfluidic devices. Lab Chip 9:1340–1343
Zhang K, Hu P, Liang QL, Luo GA (2008) Control and application of microdroplet in microfluidic chip. Chinese J Anal Chem 36:556–562
Zhang K, Liang QL, Ma S, Mu X, Hu P, Wang YM, Luo GA (2009) On-chip manipulation of continuous picoliter-volume superparamagnetic droplets using a magnetic force. Lab Chip 9:2992–2999
Zheng B, Roach LS, Ismagilov RF (2003) Screening of protein crystallization conditions on a microfluidic chip using nanoliter-size droplets. J Am Chem Soc 125:11170–11171
Acknowledgments
This research was supported by funding from National Basic Research Program (973 Program) of China (2007CB714505) and Ministry of Education of China (No. 20080031012).
Author information
Authors and Affiliations
Corresponding authors
Electronic supplementary material
Below is the link to the electronic supplementary material.
Rights and permissions
About this article
Cite this article
Zhang, K., Liang, Q., Ma, S. et al. A gravity-actuated technique for flexible and portable microfluidic droplet manipulation. Microfluid Nanofluid 9, 995–1001 (2010). https://doi.org/10.1007/s10404-010-0611-6
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10404-010-0611-6