Introduction

As of late 2019, the outbreak of a novel human virus, severe acute respiratory syndrome coronavirus-2 (SARS-CoV-2), which causes the coronavirus disease 2019 (COVID-19), has infected more than 562 million people all around the world, at the time of writing, as shown by the Coronavirus Resource Center at the Johns Hopkins University School of Medicine (https://coronavirus.jhu.edu/map.html). Death numbers have reached over six million as of July 2022 (COVID Live—Coronavirus Statistics—Worldometer 2022). The development of detection kits and tools has been extremely important for the early diagnosis of such disease. Besides in vitro molecular tests, current coronavirus diagnosis protocols are based on clinical symptom. Yet, the latter resembles the ones of influenza and common cold, pointing to the need of accurate and rapid molecular tests for a reliable diagnosis.

Among currently available in vitro diagnostics (IVD), polymerase chain reaction (PCR)–based kits, serological tests, and metagenomic next-generation sequencing (mNGS) were the validated methods to detect SARS-CoV-2. Serological tests detect SARS-CoV-2 antigen or antibodies (immunoglobulins G and M (IgG and IgM)) from patients’ sera to identify recent exposure or immunity (Behera et al. 2021). One of the most popular serological tests to detect SARS-CoV-2 is the enzyme-linked immunosorbent assay (ELISA) (Iruretagoyena et al. 2021; Vernet et al. 2021). It relies on binding antibodies from patient’s sera or blood to antigens coated on microtiter wells, resulting in complexes that changes the color of the solution after adding a substrate (Aileni et al. 2022). Also, lateral flow assays caught the attention of researchers, because they are small portable devices that can be used as a point-of-care (POC) testing technique. These assays use the immunochromatography principle. Thus, drops of a patient sample (blood, swabs, or saliva) flow through a test pad conjugated with IgG and/or IgM, which form antigen–antibody complexes if the person is infected (Grant et al. 2020; Li et al. 2020). In addition, neutralization assays measure the ability of cultured cells or patients’ samples of producing neutralizing antibodies, when exposed to SARS-CoV-2 antigens (Abe et al. 2020). Aside from serological tests, clustered regularly interspaced short-palindromic repeats (CRISPR)–based technology along with its associated enzymes (Cas 9, Cas 12, and Cas13) has been widely invested, showing high accuracy in COVID-19 diagnosis (Broughton et al. 2020; F. Zhang et al. 2020a, b). Biosensors have also caught attention during the COVID-19 pandemic. These devices include a transduction element that quantifies biochemical reactions from antigen–antibody complexes. They generate an output signal from bioreceptor element, including immobilized antibodies or nucleic acids, being analyzed by a detection system (Aileni et al. 2022; Behera et al. 2021). Biosensors effectively demonstrated high sensitivity and specificity against SARS-CoV-2, in addition to providing real-time detection (Chaibun et al. 2021; Seo et al. 2020).

PCR-based tests are useful for initial diagnosis. Yet, they require expert personnel and equipped laboratories that can substantially delay testing time even for days, depending on laboratory turnaround and sample-shipping time (Sheridan 2020). On the other hand, serological tests could rapidly diagnose within minutes, but they can only identify persons with detectable blood antibody levels, and they take a longer time to develop (Ravi et al. 2020). Others, like CRISPR-based techniques, are prone to false results when viral mutations are recurrent. Finally, biosensors require a considerably higher amount of analyte to result in a detectable response (Abid et al. 2021). Therefore, due to the limitations of the currently available methodologies, development of new cost- and time-efficient approaches with on-site testing capability is desirable. In this regard, loop-mediated isothermal amplification (LAMP)–based diagnostics are having considerable attention, thanks to their simple design, easy implementation, and POC testing capability, as well as reduced cost and testing time to generate results (Wong et al. 2018).

LAMP is a nucleic acid amplification technology using four to six specially designed primers—forward outer (F3), forward inner (FIP), loop forward (LF), backward outer (B3), backward inner (BIP), and loop backward (LB)—and a DNA polymerase with high strand-displacement activity, at 60–65 °C constant temperature (Kashir and Yaqinuddin, 2020). It can also be combined with reverse transcription to target RNA (RT-LAMP), where reverse transcription and amplification take place simultaneously (Yu et al. 2020). LAMP results can be detected by various approaches, such as change in turbidity caused by magnesium pyrophosphate precipitate (Mori et al. 2004), probe-based with nanoparticles (Bhat et al. 2019), strip pH indicators (Bhat and Rao 2020), fluorescence reading by using intercalating dyes (Thi et al. 2020), gel electrophoresis followed by UV detection (Aoki et al. 2021), or even with the naked eye, as a reaction-color change (Zhang et al. 2020a, b).

In site testing capability of RT-LAMP, even with equivalent performance to commercial PCR-based tests, broadened the convenience and scope of its applicability. Successful applications of LAMP, coupled with different methodologies, have been demonstrated in many studies, including COVID-19 (Yu et al. 2020; Zhang et al. 2020a, b; Thi et al. 2020; El-Tholoth et al. 2020; Aoki et al. 2021), MERS-CoV (Shirato et al. 2018), EcoV (Nemoto et al. 2015), SADS-CoV (Wang et al. 2018), FcoV (Stranieri et al. 2017), HIV-1 (Rudolph et al. 2015), and many others.

Among the current coronavirus diagnostic technologies such as RT-PCR, LAMP-based ones are highly prioritized due to their lower cost and time efficiency, DNA/RNA targeting ability, simple design, and easy implementation. Also, such latter assays do not require expensive qPCR instrumentation when used as colorimetric. The current study demonstrates both colorimetric and fluorometric LAMP assays by targeting N, RdRp, S, and E genes of SARS-CoV-2 genomes, belonging to the most recent variants of concern (VOC), including B.1.1.7 (Alpha), B.1.351 (Beta), P.1 (Gamma), B.1.617.2 (Delta), and B.1.1.529 (Omicron).

Materials and methods

RT-LAMP primer design

A total of 1000 SARS-CoV-2 genomes belonging to VOC, including the five described in the previous paragraph, were downloaded from Global Initiative on Sharing Avian Influenza Data (The GISAID 2022) (https://www.gisaid.org). Virus genome sequences were selected from different countries, belonging to five continents. In addition, genome sequences of SARS-CoV relatives such as AY278491.2, AY502924.1, AY502927.1, AY559094.1, AY613947.1, and NC_004718.3 were downloaded from the National Center for Biotechnology Information (NCBI) database (https://www.ncbi.nlm.nih.gov/sars-cov-2). Sequences were first aligned by using MAFFT software version 7 (https://mafft.cbrc.jp/alignment/server) (Katoh et al. 2019). Then, mutation sites (polymorphisms) on the genes were screened by using the JalView program version 2.11.1.3 (https://www.jalview.org) (Waterhouse et al. 2009). Then, 100% conserved (non-mutated; monomorphic) regions specific to SARS-CoV-2, but non-homolog for other viruses, were selected as potential primer-binding sites. PrimerExplorer program version 5 (https://primerexplorer.jp/e) was used to design six LAMP primers in each set, targeting the conserved region of specific genes (RdRp, N, E, and S, genes; Table S1). Primers designed for the S gene have also been used by others (Yan et al. 2020). Since the target region is narrow and the same design program was used, it was expected to have the same primer design. To avoid forming secondary structures (homodimers, hetero-dimers, and hairpins), OligoAnalyzer tool from Integrated DNA Technologies (IDT) (https://eu.idtdna.com/pages/tools/oligoanalyzer) was used. Many different primer sets were designed and tested. Primers showing the best performance without primer dimers in no-template control (NTC) were selected. Primers were synthesized by Alpha DNA (Montreal, Canada) (http://www.alphadna.com), desalted, and lyophilized. Concentrations of oligonucleotides in the reaction mixtures (20 µl) were adjusted as 0.2 (2 µM), 0.2 (2 µM), 1.6 (16 µM), 1.6 (16 µM), 0.8 (8 µM), and 0.8 (8 µM), for F3, B3, FIP, BIP, LB, and LF primers, respectively.

Samples and RNA extraction

This study was approved by the Institutional Review Board (IRB) at Imam Abdulrahman Bin Faisal University (IAU) with IRB no. IRB-2020–13-406. According to the institutional ethics committee regulations, no participants’ consent was required, since samples used in this study were post-diagnostic leftover tests, being de-identified. The nasopharyngeal/throat (NP/T) swabs of COVID-19-suspected patients were collected and transferred into a liquid viral transport medium (VTM) from Copan (Murrieta, CA, USA) (https://www.copanusa.com). Specimens were kept at 4 °C for less than 8 h before RNA extraction. RNA isolation was carried out according to the suggested protocol of the QIAamp Viral RNA mini kit from Qiagen (Hilden, Germany) (https://www.qiagen.com). The SARS-CoV-2 Omicron synthetic-RNA control (Twist synthetic RNA control 51; EPI_ISL_7718520) was obtained from Twist Bioscience (South San Francisco, CA, USA) (https://www.twistbioscience.com). The specificity of the primers was determined by testing other respiratory viruses’ RNA extracted from clinical samples, including parainfluenza virus 3, enterovirus, rhinovirus, human metapneumovirus A + B, parainfluenza virus 4, bocavirus, and coronavirus 229 E.

RT-LAMP assay

RT-LAMP reaction mixtures were composed of the following components: 10 × LavaLAMP RNA buffer and 1 µl LavaLAMP RNA enzyme from Lucigen (Middleton, WI, USA) (https://www.lucigen.com), 100 mM MgSO4, 10 mM dNTP, 2.5 µl primer mixture, 5 µl template, and ddH2O up to 20 µl. For colorimetric assays, 2.5 mM neutral red was added to the mixture, being replaced with green fluorescent dye from Lucigen in the fluorometric assays. Colorimetric RT-LAMP mixtures were incubated in a thermal block at 70 °C for 45 min. Reactions were terminated at 95 °C for 5 min and immediately cooled down at 4 °C, being crucial to allow post-reaction analyses and validations. Fluorometric RT-LAMP was performed in a 7500 Fast Real-Time PCR System from Thermo Fisher Scientific (Waltham, MA, USA) (https://www.thermofisher.com), using 80 cycles of 70 °C for 30 s. Fluorescence readings were performed at amplification steps of each cycle. FAM filter was selected as the reporter dye channel. In addition, assays were tested by using different light cyclers, such as StepOnePlus Real-Time PCR System from the same manufacturer, CFX96 Touch Real-Time PCR Detection System from Bio-Rad Laboratories (Hercules, CA, USA) (https://www.bio-rad.com), Q2000 Real-Time PCR System from LongGeneScientific Instruments (Hangzhou, China) (http://en.longgene.com), and LightCycler 480 System from Hoffmann-La Roche (Basel, Switzerland) (https://diagnostics.roche.com).

Validation of results by RT-PCR

The RT-PCR was applied according to our previous protocol (Tombuloglu et al. 2021, 2022). In multiplex RT-PCR, viral RdRp and N, and human RP genes were targeted. A positive control containing synthetic copies of each target gene was used (EURM-019, single-stranded SARS–CoV-2 RNA fragments) from the European Commission, Joint Research Centre Directorate F—Health, Consumers and Reference Materials (Geel, Belgium) (https://crm.jrc.ec.europa.eu/p/EURM-019). For NTC, RNase/DNase-free ddH2O was added instead of the RNA template. To validate LAMP results, reaction mixtures were segregated using 2% agarose gel electrophoresis (AGE), stained with VisualaNA (A) DNA Stain from Molequle-On (Auckland, New Zealand) (http://molequle-on.com). Gels were imaged under a UV transilluminator (ChemiDoc XRS + System with Image Lab Software) from Bio-Rad Laboratories.

Results

Colorimetric and fluorometric identification of positive samples

Figure 1 depicts tested individuals, according to the RT-LAMP assay. Different primer sets targeting N, S, RdRp, and E genes were evaluated, via both colorimetric and fluorometric identification. Using the pH-sensitive neutral red dye, a color change to red was evident in positive control (PC) samples after colorimetric RT-LAMP reactions, while NTC was yellow. Also, in the fluorometric identification, PC samples produced a fluorescence amplification after ~ 40 cycles (around 20 min), and no amplification was evident in NTC of all tested primer sets. Development of yellow to red color was more pronounced in positive reaction tubes prepared with N or S primer sets, compared to RdRp and E ones (Fig. 1a-d). However, there was no significant difference in Ct scores of positive amplification curves among tested target genes (Fig. 1al-dl).

Fig. 1
figure 1

RT-LAMP-based (a–d) colorimetric and (a1–d1) fluorometric identification of SARS-CoV-2-positive controls (PC, +) and negative template controls (NTC, –) samples, by using N (a, a1), S (b, b1), RdRp (c, c1), and E (d, d1) gene primer sets. SARS-CoV-2-positive samples undergo color change from yellow to red/pink in the colorimetric-based reactions, producing amplification curves in fluorometric-identification methods

RT-LAMP assay on clinical samples

Since colorimetric identification leads to more noticeable color development for N and S primer sets, performance of those primer sets was tested on clinical specimens (n = 51), using both colorimetric and fluorometric RT-LAMP assays. Results were then validated using AGE, after colorimetric RT-LAMP reactions. Figure 2 demonstrates colorimetrically and fluorometrically tested specimens (S1–S51), highlighting color differences between positives and negatives, comparing to PC and NTC in each sample set. Among the 51 tested samples, RT-LAMP assays agreed with RT-qPCR results on 48 samples (94.1%), as 34 are detected as SARS-CoV-2 positives and 14 negatives. Three samples were found inconclusive. Thus, S18 was positive in colorimetric RT-LAMP, but negative in RT-qPCR. On the other hand, S30 and S32 were negative in colorimetric RT-LAMP, but positive in RT-qPCR. Therefore, S18 must be regarded as positive, since amplicon formation was evident via color change, being undetected by RT-qPCR (Fig. S1). Color intensities varied among these samples, based on the amount of viral RNA present (Fig. 2a). Also, some of these samples were left at room temperature to check if any further color was developed, as some of them had an immediate color change (within 20 min), whereas others took over 60 min, thus causing some variety in color intensity. Nonetheless, results were in agreement with fluorometric identification (94.1%), where positive specimens produced fluorescence amplifications with high fluorescence intensities (Fig. 2b). Ct values in positive specimens varied between 35 and 75, and no Ct value was evident in NTC (Fig. 2c). It should be taken into consideration that differences in Ct values may also be due to different viral loads among specimens.

Fig. 2
figure 2

(a) Colorimetric and (b) fluorometric S-gene-specific RT-LAMP assays applied on SARS-CoV-2-positive and SARS-CoV-2-negative RNA specimens. PC, positive control; NC, negative control. (c) Ct spectrum of samples

In order to confirm both colorimetric and fluorometric results, colorimetric RT-LAMP products were loaded in a 2% agarose gel and visualized under a UV transilluminator. Amplification products with different sizes appeared for SARS-CoV-2-positive and PC specimens. No bands were evident in negative samples and NTC, confirming successful amplification of the SARS-CoV-2 S gene in the former. Figure 3 shows agreement between colorimetric, fluorometric, and gel electrophoresis techniques on the patients’ specimens, along with PC and NTC.

Fig. 3
figure 3

Triple confirmation of S-specific LAMP primer sets. (a) Colorimetric, (b) agarose gel electrophoresis, and (c) fluorometric, showing agreement among these techniques

Assay sensitivity and specificity of primer sets

SARS-CoV-2 RNA with a known concentration (EURM-019, single-stranded SARS–CoV-2 RNA fragments) were serially diluted (1, 101, 102, 103, 104, 105, and 106 fold) to determine the limit of detection (LOD) of RT-LAMP reactions. Figure 4 shows the LOD of colorimetric and fluorometric RT-LAMP assays, targeting the S gene. Results showed that color development from yellow to red is prominent in 1, 101, 102, and 103 fold dilutions. The limit of detection corresponds to 50 copies/µl in reactions, for the 103-fold diluted specimen (Fig. 4a). As expected, samples with higher viral load produced a more prominent or darker red color, as compared to lower ones. Likewise, Ct values in fluorometric identifications progressively increased with an increasing dilution factor, further confirming colorimetric results (Fig. 4b). Besides, the color intensity and Ct values were highly dependent on viral load. Once reactions were carried out, they were subjected to AGE in 2% agarose and visualized under a UV transilluminator, to further check if band pattern intensities decreased with reduced viral loads (Fig. 4c). Results indicate a ladder-type banding pattern in detected dilutions, while no bands appeared for others. In summary, there was agreement between agarose gel electrophoresis, fluorometric, and colorimetric identifications.

Fig. 4
figure 4

Assay sensitivity of (a) colorimetric and (b) fluorometric RT-LAMP tests, targeting the S gene. Dilution factors are represented on the graph. (c) Agarose gel electrophoresis of diluted samples (from 100 to 106 fold). M and NC denote marker and negative control, respectively. (d) Comparison of colorimetric (col.) and fluorometric (flu.) RT-LAMP with RT-qPCR assays. Among 51 tested samples, RT-LAMP assays, both colorimetric and fluorometric, agreed with RT-qPCR on 48 samples, as 34 positives and 14 negatives. Three samples were found inconclusive. (e) Comparison of time to positive (Tp) of RT-LAMP and cycle threshold (Ct) of RT-qPCR assays, based on N (black dots) and RdRp (red dots) genes

When compared with RT-PCR, RT-LAMP assays result in 94% agreement (48 over 51 tested samples) with both fluorometric and colorimetric identifications, where three samples were found inconclusive (Fig. 4d). For instance, sample 18 (S18) was not detected as positive in multiplex RT-qPCR simultaneously targeting both N and RdRp genes, while RP amplification was shown. This reflects an amplification pattern of a negative specimen (Fig. S1). However, both fluorometric and colorimetric RT-LAMP reactions generated positive results, which conflicts with the ones of RT-qPCR. Opposite patterns were found for S30 and S32 specimens (Fig. S1), being therefore considered as inconclusive. As a result, RT-LAMP colorimetric assays showed sensitivity of 94.6%, specificity of 92.9%, and accuracy of 94.1%. Among the tested 51 clinical specimens, the assay’s positive predictive value (PPV) was 97.2% and the negative predictive value was 86.7% (Table S2; S3). In addition, RT-LAMP time-to-positive (Tp) was compared with the Ct values of N and RdRp genes by RT-qPCR (Fig. 4e) (14 positive and 16 negative clinical samples). Fluorometric LAMP showed 100% agreement with RT-qPCR results.

Similarly, the N gene primer set was tested against SARS-CoV-2-positive (PC) and negative (NC) specimens, using the fluorescence RT-LAMP detection technique. This primer set successfully amplified PC samples, shown in the amplification plot (Fig. S2). Melting curve analyses of SARS-CoV-2-positive samples lead to a single peak, pointing out the generation of an abundant amplicon, rather than multiple unspecific- or spurious-priming products. Such abundant target specificity is also shown in Fig. 4b. Although LAMP reactions are known to produce multiple amplicons with randomized sizes, some amplicons could be more abundant, while rare ones could be masked, leading to a single peak at melting curve analyses. Indeed, it is known that PCR follow the law of all-or-nothing. In other words, if a molecule is amplified in the first cycles, it will probably be logarithmically amplified, taking over the reaction. Yet, if a molecule is not amplified in such first cycles, it will likely not undergo such exponential amplification, not being eventually detected.

In addition, the assay’s cross-reactivity was tested against eight different respiratory viruses from different clinical samples, which included coronavirus 229 E, parainfluenza virus 3, parainfluenza virus 4, human metapneumovirus A + B, bocavirus, enterovirus, and rhinovirus. The resulted colorimetric RT-LAMP tests showed color change only in SARS-CoV-2 samples, but not for the others. RT-LAMP products were then loaded in 2% agarose gel 40 min post-reaction. Only the SARS-CoV-2 sample showed ladder-type patterns (Fig. S3). Both results indicate high specificity of this primer set against COVID-19 infection.

Discussion

The colorimetric RT-LAMP technique’s simplicity allows for its use as a POC testing alternative for large-scale screening. Its diagnostic accuracy, however, must be comparable to the RT-qPCR gold standard to consider such methodology reliable, avoiding false results and their undesired consequences. In the colorimetric detection, 94.1% of results were found to be consistent with RT-qPCR outcomes. However, three samples (S18, S30, and S32), among the tested specimens with the S gene primers set, contradicted the RT-qPCR results, and hence were considered inconclusive. Two of the samples (S30 and S32) were RT-qPCR positive without color development in colorimetric LAMP. In opposite, the other inconclusive sample (S18) was RT-qPCR negative, while color development was obvious upon LAMP reaction. Also, fluorometric LAMP led to consistent findings with colorimetric LAMP (Fig. S1). It is well known that the viral load has a great impact on colorimetric RT-LAMP and RT-qPCR sensitivity. The positive identification of the S18 specimen by RT-LAMP, but not by RT-qPCR, points out the sensitivity of the current assay. Consistency with RT-qPCR gold standard results must be validated, to assure reliability of colorimetric RT-LAMP. The colorimetric RT-LAMP tests are reliable when testing samples with moderate to high viral loads, which usually corresponds to samples with Ct values lower than 40 in RT-qPCR. For example, over 700 clinical specimens were tested, via targeting both N and RdRp genes, showing color change in positive ones after 30-min reaction time at 65 °C with 97.5% sensitivity, for samples with high to moderate viral loads (Thi et al. 2020). In that regard, most studies show agreement on colorimetric RT-LAMP-reaction time of 30 min (Amaral et al. 2021; Chow et al. 2020; Huang et al. 2020; Lalli et al. 2021; Promlek et al. 2022), with temperatures not exceeding 65 °C (Amaral et al. 2021; Aoki et al. 2021; Chow et al. 2020; Thi et al. 2020; de Oliveira et al. 2021; Huang et al. 2020; Lalli et al. 2021; Luo et al. 2022; Nawattanapaiboon et al. 2021; Promlek et al. 2022), in moderate to high viral load samples, to avoid false results. Authors discussed that reaction time over 30 min resulted in nonspecific amplification, and temperature settings above 65 °C gave false-positive results in their assays, especially in NTC (Aoki et al. 2021; Promlek et al. 2022). However, time and temperature settings vary depending on the type of DNA-polymerase enzymes, primers, and samples in reaction mixtures. In this study, we used the LavaLAMP RNA enzyme (Lucigen), which has an activation temperature of 68–74 °C. Based on the melting temperature (Tm) of primers, and after optimizing reaction conditions, 40 min at 70 °C was the optimum setting in our colorimetric RT-LAMP assay. On average, it takes around 20 min (Ct 40) to 37 min (Ct 74) for SARS-CoV-2-positive specimens to develop color. When compared with RT-qPCR results, colorimetric RT-LAMP results are consistent, except for S18, which was positive in colorimetric RT-LAMP, but not detected in RT-qPCR, highlighting the sensitivity of this technique. To better distinguish between positives and negatives, it was found that placing samples on ice immediately after RT-LAMP reaction enhanced color formation (de Oliveira et al. 2021). Overall, the sensitivity of the colorimetric RT-LAMP reaction greatly depends on the assay’s LOD, sample quality, viral load, and time of sample collection from the disease onset (Promlek et al. 2022).

Furthermore, it is common to validate post-reaction amplifications by loading samples in agarose gel electrophoresis and observing distinct band patterns in positive ones (Amaral et al. 2021; Aoki et al. 2021; Thi et al. 2020; de Oliveira et al. 2021; Huang et al. 2020; Luo et al. 2022; Nawattanapaiboon et al. 2021). It was observed that intermediate results showed band patterns in AGE, indicating that amplification has occurred, regardless of color development. These samples may have low amplicon formation or a low pH reduction, being probably affected by the elution buffer used during the RNA extraction process (Aoki et al. 2021).

According to our results, the N and S gene primer sets showed the best analytical performance and thus were chosen for further analyses. These primer sets were designed to target conserved (non-mutated) regions of the most recent VOC. So, this assay is expected to detect any SARS-CoV-2 variant, including the heavily mutated Omicron. Since most Omicron mutations are on the S gene, primer-binding sites of such gene across VOC were screened (Fig. S4). Accordingly, 99.2% of sequences in SARS-CoV-2 Omicron (B.1.1.529) variant (among 136 nt in total primer length, one was found as variant/mutated) exhibited similarity of the ones for Delta (B.1.617.2) variant. Omicron’s specific point mutation (C/A) exists at the LB binding site, which is five nucleotides away from the 3′-site of the primer. Although it is far from the critical 3′-end for the polymerase enzyme, and seems to be negligible for amplification, its potential effect on the amplification of the Omicron variant was examined. Indeed, this primer set successfully detected Omicron, which confirms that it targets a non-mutated region in the S gene. In addition, we confirmed the specificity of the primers against other respiratory viruses, and no cross-reactivity was observed. Overall, the analytical performance of four primer sets targeting N, S, RdRp, and E genes was elucidated. Among the tested genes, N and S had the highest analytical sensitivity when detected by both colorimetric and fluorometric RT-LAMP.

Conclusions

Alternative COVID-19 diagnostic techniques to the RT-qPCR gold standard are being broadly investigated, to address shortages in some areas. Colorimetric RT-LAMP has the potential to be used as a POC testing technique, showing a 94.6% sensitivity and 92.9% specificity comparable to RT-qPCR, with shorter reaction time and simpler implementation. Besides, it is cheaper, not requiring specialized equipment and trained personnel, and results can be interpreted by simple color visualization. This LAMP test is expected to detect any SARS-CoV-2 variant, including the heavily mutated Omicron. That makes it an attractive candidate to be used in local emergency centers, and even for in-home testing.