Abstract
Real-time, contrast-specific ultrasonography (US) uses low-mechanical-index, harmonic software to stimulate echo emission from resounding second-generation contrast medium microbubbles. At our institution, contrast-enhanced US is increasingly being used in the evaluation of acute abnormalities of the spleen, mainly to overcome some limitations of conventional (basic) US. This pictorial essay illustrates the appearance of several acute splenic lesions, both traumatic and nontraumatic, on contrast-specific US.
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Avoid common mistakes on your manuscript.
Introduction
By using a low acoustic pressure (low mechanical index), real-time, contrast-specific ultrasonography (US) techniques produce images based on nondestructive, nonlinear acoustic effects of sound interaction with microbubble contrast media [1, 2]. The high harmonic emission capability of stabilized microbubbles containing gases other than air (the so-called second-generation US contrast media) allows gray-scale display of tissue enhancement and continuous exploration of the spleen during circulation of these contrast agents. This is at the cost of a slight loss in spatial resolution and overall image quality.
The blood-pool, second-generation contrast agent SonoVue can be prepared in a few seconds and is immediately administrable. The overall time for contrast-enhanced US is 6 min at most. All these features make SonoVue optimal for use in emergency patients [3, 4].
Our technique
At our institution radiologists perform US studies directly [3, 4]. All patients first undergo baseline, conventional US examination. Thereafter, a contrast-enhanced study is obtained, employing the continuous-mode, low-mechanical-index technology called contrast-tuned-imaging (CnTI, Esaote, Italy). CnTI software allows selective tuning of the scanner with the contrast medium signal. A specific transmitting and receiving resonance frequency is used to avoid interference from tissue signals (lowered to minimum). We use an ultra-low acoustic power setting (40–50 kPa derated pressure, expressing a mechanical index of 0.06–0.07). The ultrasound beam is focused on the deepest portion of the region of interest.
We rapidly inject intravenously 2.4–4.8 ml of sulfur-hexafluoride-filled, phospholipid-stabilized microbubbles (SonoVue, Bracco, Italy) [5, 6]. The contrast agent is administered through a peripheral vein and a 20G needle using a three-way stopcock, and is followed by 5–10 ml normal saline flush.
Real-time imaging starts immediately after injection and lasts 5–6 min. The spleen is scanned right through, exploring repeatedly the whole of the parenchymal portion. A timer is pushed at the moment of injection and the whole examination movie is archived on the scanner. Stored clips are subsequently sent to a personal computer and converted to avi files.
Spectrum of findings
Spleen injury
US has recently become very popular in the evaluation of abdominal trauma, especially as a rapid way to rule out hemoperitoneum in unstable subjects (FAST technique) [7, 8, 9, 10]. The role of FAST technique in imaging stable patients is a matter of controversy [11, 12, 13] and many institutions, including our hospital, still prefer to obtain a complete US survey of abdomen and pelvis to detect combined organ injuries. In any case, it is known that US is less sensitive than CT in the evaluation of parenchymal lesions. The use of contrast enhancement may partially overcome this limitation [3,14, 15, 16, 17].
Spleen injuries appear as parenchymal portions lacking contrast enhancement (Figs. 1, 2, 3). Injury-to-parenchyma contrast increases progressively as spleen echogenicity rises, reaching its maximum at 2–3 min from injection. Lacerations are recognizable as hypoechoic bands, linear or branched, perpendicular to the spleen capsule, while lacerocontusive areas and parenchymal hematomas appear as inhomogeneous hypoechoic regions without mass effect or vessel displacement.
Contrast-enhanced US may also display findings that do not appear on conventional US imaging, including perfusion defects and contrast extravasation, as shown in experimental studies [18, 19, 20] and clinical series [3]. Partial hypoperfused areas (post-traumatic infarctions) appear as wedge-shaped or polar hypoechoic areas, whereas total or subtotal lack of enhancement is detectable in cases of pedicle avulsion or severe hypovolemia (“shock spleen”). In the latter case overall parenchymal enhancement is poor, clearly less than expected, and less than the adjacent left kidney.
Intrasplenic contrast pooling is recognizable as a small and persistent parenchymal area of hyperechogenicity, isoechoic to opacified parenchymal arteries. Extrasplenic contrast extravasation appears as a subcontinuous jet reaching the spleen capsule (possibly from the tip of an opacified artery) and spreading around the organ (Figs. 4, 5). Contrast medium leakage indicates ongoing bleeding and has been repeatedly shown [21, 22, 23, 24, 25, 26, 27] to be a CT predictor of conservative management failure and an indicator of requirement for surgery. The new possibility of identifying this sign by using contrast-specific US opens new perspectives in blunt trauma sonographic evaluation and follow-up [3].
Contrast-enhanced US has been shown to be comparable to contrast-enhanced CT in detecting spleen trauma [3]. Small injuries may sometime be missed but, in our experience, this occurrence does not have practical consequences (Fig. 6).
Spleen infarction
Typical splenic infarctions appear as wedge-shaped or polar areas of persistent hypoechogenicity (opacification defect) (Fig. 7, 8). Absence of contrast enhancement and detection of a perihilar, truncated artery are the diagnostic landmarks. The texture can be homogeneous or inhomogeneous while infarction margins can be well- or ill-defined [4].
Atypical round-shaped infarcts, sometimes simulating focal lesions, are readily diagnosed because of total lack of enhancement.
Subtotal splenic infarction is seen as a lack of parenchymal opacification, with possible upper pole peripheral areas of preserved perfusion. Comparison with the clearly enhancing left renal parenchyma allows immediate diagnosis.
Other acute disorders
Less commonly seen disorders of the spleen include spontaneous hematomas and abscesses. Both are recognizable as hypoechoic areas surrounded by enhancing normal parenchyma (Fig. 9). Abscess collections are surrounded by opacified vessels and may show peripheral enhancing rim and enhancing septa [4].
Pitfalls and limitations
There are a few potential causes of misinterpretation to be aware of [3]. During the early phase of parenchymal opacification the spleen may appear inhomogeneous (a finding similar to the well-known zebra pattern of dynamic CT and MRI). However, the transient nature of pseudolesions distinguishes them from real disorders (Fig. 10). At about 2–3 min from injection the densely opacified superficial parenchyma may cause transient obscuration of deeply located portions. Large hemangiomas may also opacify densely and thus obscure the deeply located parenchyma. Splenic lobulation, incisures, and accessory tissue may be mistaken for traumatic lesions, an well-known occurrence on conventional US imaging (Fig. 11). Opacified parenchymal arteries may mimic intrasplenic contrast extravasation and vice versa. Finally, it should be noted that regions difficult to explore using conventional US, such as the deep pole of the spleen or subphrenic regions, are still hard to image using contrast-specific US. The same applies to cases where a meteoric left colonic angle partially overshadows the spleen: this limitation clearly persists on contrast-enhanced images.
Conclusion
The spleen is a superficial organ in the abdomen, with an almost homogeneous echotexture and dense and persistent contrast enhancement. These features make the spleen an optimal organ to be study with contrast-specific US [4].
US contrast media injection allows a significant increase in lesion-to-parenchyma conspicuity, with hypovascular abnormalities becoming clearly evident in comparison with the undamaged hyperechoic parenchyma. Parenchymal changes that were very subtle on baseline US became evident after contrast administration. Moreover, the true size of parenchymal abnormalities is better defined on contrast-enhanced US, showing greater correlation with CT findings than conventional US.
Traumatic injuries, infarctions, and other acute disorders of the spleen can be optimally evaluated using real-time, contrast-specific US. Radiologists should be aware of the related findings, since this technique is expected to be increasingly used in clinical practice.
References
Bauer A, Solbiati L, Wessman N (2002) Ultrasound imaging with SonoVue: low mechanical index real-time imaging. Acad Radiol 9(Suppl 1):282–284
Lencioni R, Cioni D, Bartolozzi C (2002) Tissue harmonic and contrast-specific imaging: back to gray scale in ultrasound. Eur Radiol 12:151–165
Catalano O, Lobianco R, Sandomenico F, Siani A (2003) Splenic trauma: evaluation with contrast-specific sonography and a second-generation contrast medium: preliminary experience. J Ultrasound Med 22:467–477
Catalano O, Lobianco R, Sandomenico F, D’Elia G, Siani A (2003) Real-time, contrast-enhanced sonography of the spleen: examination technique and preliminary clinical experience. Radiol Med 106:338–356
Schneider M (1999) SonoVue, a new ultrasound contrast agent. Eur Radiol 9(Suppl 3): 347–348
Morel DR, Schwieger I, Hohn L, Terrettaz J, Llull JB, Cornioley YA, Schneider M (2000) Human pharmacokinetics and safety evaluation of SonoVue, a new contrast agent for ultrasound imaging. Invest Radiol 35:80–85
Bode PJ, Edwards MJ, Kruit MC, van Vugt AB (1999) Sonography in a clinical algorithm for early evaluation of 1671 patients with blunt abdominal trauma. AJR Am J Roentgenol 172:905–911
Dolich MO, McKenney MG, Varela JE, Compton RP, McKenney KL, Cohn SM (2001) 2,576 ultrasounds for blunt abdominal trauma. J Trauma 50:108–112
Brown MA, Casola G, Sirlin CB, Patel NY, Hoyt DB (2001) Blunt abdominal trauma: screening US in 2,693 patients. Radiology 218:352–358
Richards JR, McGahan JP, Jones CD, Zhan S, Gerscovich EO (2001) Ultrasound detection of blunt splenic injury. Injury 32:95–103
Shanmuganathan K, Mirvis SE, Sherbourne CD, Chiu WC, Rodriguez A (1999) Hemoperitoneum as the sole indicator of abdominal visceral injuries: a potential limitation for screening abdominal US for trauma. Radiology 212:423–430
Chiu WC, Cushing BM, Rodriguez A, Ho SM, Mirvis SE, Shanmuganathan K et al (1997) Abdominal injuries without hemoperitoneum: a potential limitation of focused abdominal sonography for trauma (FAST). J Trauma 42:617–623
Poletti P-A, Kinkel K, Vermeulen B, Irmay F, Unger P-F, Terrier F (2003) Blunt abdominal trauma: should US be used to detect both free fluid and organ injuries? Radiology 227:95–103
Beckman M (2003) Ultrasound with contrast enhancement as a means to assess trauma patients—an initial experience. Digital Imagery (ASER 14th Annual Scientific Meeting book):84–85
Poletti P-A (2003) Contrast-enhanced ultrasound in blunt abdominal trauma: the Geneva experience. Digital Imagery (ASER 14th Annual Scientific Meeting book):85–86
Martegani A, Cosgrove DO, Del Favero C, Aiani L, Harvey CJ (2002) Contrast enhanced abdominal ultrasound in trauma using Sonovue. Radiology 225(p): 358 (Radiological Society of North America, oral presentation)
Poletti P-A, Platon A, Becker C, Terrier F (2003) The value of contrast-enhanced sonography to improve detection of liver and spleen traumatic injuries: a comparison with contrast enhanced CT. Oral presentation, Radiological Society of North America, Radiology (p):45
Brown JM, Quedens-Case C, Alderman JL, Greener Y, Taylor KJ (1997) Contrast enhanced sonography of visceral perfusion defects in dogs. J Ultrasound Med 16:493–499
Goldberg BB, Merton DA, Liu J-B, Forsberg F (1998) Evaluation of bleeding sites with a tissue-specific sonographic contrast agent: preliminary experiences in an animal model. J Ultrasound Med 17:609–616
Liu J-B, Merton DA, Goldberg BB, Rawool NM, Shi WT, Forsberg F (2000) Contrast-enhanced two- and three-dimensional sonography for evaluation of intra-abdominal hemorrhage. J Ultrasound Med 21:161–169
Schurr MJ, Fabian TC, Gavant M, Croce MA, Kudsk KA, Minard G, et al (1995) Management of blunt splenic trauma: computed tomography contrast blush predicts failure of nonoperative management. J Trauma 39:507–513
Gavant ML, Schurr M, Flick PA, Croce MA, Fabian TC, Gold RE (1997) Predicting clinical outcome of nonsurgical management of blunt splenic injury: using CT to reveal abnormalities of splenic vasculature. AJR Am J Roentgenol 168:207–212
Federle MP, Courcoulas AP, Powell M, Ferris J, Peitzman AB (1998) Blunt splenic injury in adults: clinical and CT criteria for management, with emphasis on active extravasation. Radiology 206:137–142
Shanmuganathan K, Mirvis SE, Boyd-Kranis R, Takada T, Scalea TM (2000) Nonsurgical management of blunt splenic injury: use of CT criteria to select patients for splenic arteriography and potential endovascular therapy. Radiology 217:75–82
Omert LA, Salyer D, Dunham CM, Porter J, Silva A, Protetch J (2001) Implications of the “contrast blush” findings on computed tomographic scan of the spleen in trauma. J Trauma 51:272–27
Yao DC, Jeffrey RB, Mirvis SE, Weejes A, Federle MP, Kim C, et al (2002) Using contrast-enhanced helical CT to visualize arterial extravasation after blunt abdominal trauma. AJR Am J Roentgenol 178:17–20
Willmann JK, Roos JE, Platz A, Pfammatter T, Hilfiker PR, Marincek B, et al (2002) Multidetector CT: detection of active hemorrhage in patients with blunt abdominal trauma. AJR Am J Roentgenol 179:437–444
Author information
Authors and Affiliations
Corresponding author
Additional information
Presented as an exhibit at the 2003 ASER 14th Annual Scientific Meeting, Las Vegas.
Rights and permissions
About this article
Cite this article
Catalano, O., Cusati, B., Nunziata, A. et al. Real-time, contrast-specific sonography imaging of acute splenic disorders: a pictorial review. Emerg Radiol 11, 15–21 (2004). https://doi.org/10.1007/s10140-004-0343-6
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10140-004-0343-6