Abstract
The purpose of this study was to evaluate the temperatures on the root surfaces during Nd:YAG laser irradiation in root canals using pulse durations of 180 and 320 μs. Thirty extracted human teeth were used in this study. The teeth were enlarged up to ISO 40 (multi-rooted) or up to ISO 60 (single-rooted) by conventional technique using K-files. Then the teeth were placed into a water bath with a constant temperature of 37 °C and then irradiated with an Nd:YAG laser having an output power of 1.5 W, a frequency of 15 Hz, using an optic fiber of 200 μm diameter. The temperature on the root surface was measured by means of attaching thermocouples in three areas (coronal, mesial, and apical regions) of the root canals. The thermographic study showed that the average temperature elevation for both pulse durations on the root surfaces was less than 9 °C. There was no significant difference in the observed temperatures in coronal and mesial areas. Though a higher increase of temperature was observed in the apical region when the pulse length of the Nd:YAG laser was 320 μs. The results of the study showed that the temperature rises during Nd:YAG laser irradiation with parameters used in this study minimal to cause damage on bone and periodontal tissues. Moreover, it was suggested that in order to have lower temperature in the apical region, an Nd:YAG laser with a pulse length of 180 μs is preferred than one with a pulse length of 320 μs.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
The importance of sterilizing a root canal
The primary target of an endodontic therapy is to remove the organic material from the root canal, then to shape the root canal in order to introduce the filling material in it and then to fill the root canal with this material in order to seal it from the surrounding oral tissues [1, 2]. The most critical point in the procedure is to obtain “sterile” root canals since microorganisms are one of the major causes of the development of endodontic disease. That means that the target of a “microbe-free” root canal is crucial for a successful therapy [3–5].
The smear layer
It has been shown that the smear layer is not beneficial to the overall success of the treatment because it is responsible for leakage between the canal walls and the filling materials and should be removed prior to the root filling. It has been also shown that this layer contains microorganisms, which if the canal cannot be completely sealed, can result in subsequent pathologic conditions [6]. Conventional procedures in endodontics are focusing in removing debris and microorganisms from the root canal with hand instrumentation, sometimes in combination with motor-driven files, or with ultrasonic instruments [7].
Chemical irrigants are also recommended to dissolve debris and smear layer. Solutions such as ethylenediaminetetraacetate are used. However, both irrigants are unable to remove smear layer effectively and they are not able to reach the bacterial colonizations harbored in the dentinal tubules. Kouchi et al. could demonstrate that bacteria are capable of invading the dentine up to a depth of 1.100 μm [8]. On the other hand, chemical irrigants penetrate no more than 130 μm into the dentine as indicated by Lindhe [9].
The introduction of lasers in endodontic therapy
In order to achieve a high bactericidal effect, lasers were introduced in 1971 in endodontic treatment, and different types of lasers since then have been investigated for their efficacy in root canal sterilization [10, 11]. Experiments like those of Gutknecht et al. demonstrate that Nd:YAG laser irradiation has bactericidal effects in depths of 1.000 μm and more [12].
Numerous studies have been performed using Nd:YAG lasers [7, 13, 14, 15, 16]. It is the most popular laser used in the root canal sterilization due to its fiber optic delivery system which enters the narrow root canals, its high absorption in pigmented tissue (bacteria), and its high transmission through dentine walls [17–21].
High-power pulsed Nd:YAG lasers are mostly used in the field of sterilizing a root canal due to its high bactericidal effect. In the investigation performed by Gutknecht, Conrads, and Sievert, an average of 99.91 % of the bacteria examined (Enterococcus faecalis) could be eliminated [13].
The heat produced by the lasers has become a great factor for clinical acceptance in endodontic treatment. The American Association of Endodontists has warned that the temperature increase inside the root canal by the laser could potentially damage the periodontal tissues. The magnitude of heat generated depends on a number of variables, such as the type of the laser, the power settings, the mode of energy delivery, the presence of water, and the type of the target tissue [22–24].
Since the introduction of different pulse durations from manufacturers, the heat generation during laser irradiation in the root canal differs. Hence the study measured the temperatures on the root surfaces during Nd:YAG laser irradiation in root canals using pulse durations of 180 and 320 μs. This has not been done so far and we needed to prove that clinicians working with these settings for laser-supported endodontic treatment are on the safe side regarding thermal damage on the periapical tissues.
Materials and methods
Sample preparation
Thirty human freshly extracted non-carious teeth, 15 single- and 15 multi-rooted with straight canals, were used in this study. Only teeth with totally formed roots were selected.
The access cavities were prepared with diamond burs (Shofu, CA, USA). The root canals were cleaned and enlarged with a step-back technique using hand instruments (K-files Dentsply–Maillefer, Ballaigues, Switzerland). The enlargement was up to an apical size of ISO #60 for single-rooted teeth and up to an apical size of ISO #40 for multi-rooted teeth. All the canals were completely cleaned and washed with normal saline. Then, each root canal was dried completely with paper cones.
Laser apparatus
A Fidelis Plus II (Fotona d.d., Ljubljana Slovenia) was used. For this study, we used only the Nd:YAG laser which emits photons at a wavelength of 1,064 nm. This laser is pulsed with three options of pulse width: VSP (100 μs), short pulse (SP) (180 μs), and long pulse (LP) (320 μs). Its maximum frequency is 100 Hz and its maximum output power is 15 W. The delivery system consists of a glass fiber of 200 μm diameter. The aiming laser is a semiconductor diode laser operating at a wavelength of 650 nm.
Experimental procedure
Each tooth was placed in a thermal bath, which was constantly fixed to 37 °C in order to simulate the human temperature in the mouth. On the external surface of the root, three copper–constantan thermocouples (type T 202 KC, Digitron, England) were positioned and attached in order to register thermal changes at the coronal, mesial, and apical third of the root.
Paraffin wax was used to isolate the thermocouple to prevent it being influenced by environmental temperature. The room temperature during the experiment was 25 °C.
To ensure that the 200-μm glass fiber reaches the apex, the working length of each root canal was transferred exactly to the fiber optic waveguide. Using the method described by Gutknecht, the fiber is first positioned in place without activating the laser, and then the fiber is removed from the canal in circular movements from apical to coronal with a rate of 2 mm per second (that means that in a root canal of 20 mm length, the laser irradiation procedure should reach 10 s).
The power settings that was selected were 1.5 and 15 Hz repetition rate which has been proved to be most efficient [7]. Each root canal was first irradiated with SPs (pulse duration 180 μs) and then with LPs (pulse duration 320 μs). Between the two irradiations, adequate time was given for the root canal to return to initial temperature of 37 °C. During the procedure, the highest temperature indicated from each thermocouple (apical, mesial, and coronal) was recorded.
Results
The temperature rise between the two pulse durations in all three points of the tooth can be better seen in the box plot graphs. The chart and the graphs show that mean temperatures for the apical area are the highest, followed by those of mesial and coronal. In accordance, similar results can be seen for the standard deviation, which is also higher for the apical areas.
Comparing the two pulse duration, it shows that the LP gives systematically higher mean values. The difference of the mean values between LP and SP is higher for apical points followed by those of mesial and coronal points.
The temperature distributions seem symmetrical. Using the Kolmogorov–Smirnov tests, it showed that they could be assumed as normal. The results of this study tests can be seen below. All p values were much higher than the formal significance level of 5 %, indicating that the null hypothesis of normality of temperatures is a reasonable assumption.
Statistics
The mean difference in temperature between SP and LP was nonsignificant for coronal areas (mean difference −0.20, t = −0.073, p = 0.942), nonsignificant for mesial areas (mean difference −0.37, t = −1.34, p = 0.191), while in apical points, the mean difference was significant (mean difference −1.29, t = −2.73, p = 0.011) (Table 1). The paired differences of the mean temperatures are high between the three points of the tooth for SP and LP.
For SP, the differences between coronal and mesial were significant (t = −4.16, p < 0.001). Also significant were the differences between coronal and apical (t = −5.78, p < 0.001), and mesial and apical (t = −2.64, p = 0.013). Similar results were drawn for LP and for the differences between coronal and mesial (t = −5.09, t < 0.001), coronal and apical (t = −8.32, p < 0.001), and mesial and apical (t = −3.90, t = 0.001).
For reasons of completeness, the corresponding results based on nonparametric tests are shown in the following table. The results confirm the conclusions that were found in the parametric tests (Figs. 1 and 2).
Discussion
From the analysis, it was shown that the temperatures do not differ between SP and LP for the coronal and mesial points. These mean differences remain statistically nonsignificant even if one-sided tests are used, mainly in case where information from the experiment exists as to the direction of the mean difference of temperatures. In apical points though, the mean differences were statistically significant.
Statistically significant were also the paired differences between the three points of the tooth either in SP or LP. This study was aimed to compare the temperature elevation on human tooth surface induced by a pulsed Nd:YAG laser with same power settings but different pulse duration.
It is clear that the most susceptible area for thermal damage is the apical part of the root where there is only a thin layer of dentine and cement between the root canal and periodontal ligaments and surrounding alveolar bone. Several reports have attempted to find a safe limit that would allow lasers to be used without causing thermal injury. Most studies resulted that the in vitro temperature rise after laser-assisted root canal therapy is between 47.3 and 51.4 °C and that higher temperatures than 53 °C will result in bone necrosis [25].
In this study, it was shown that the highest temperature rises were observed in the apical point of the root either using short pulses or long pulses. The mean temperature measured for SP was 45.5 °C and for LP was 46.8 °C. That is a mean temperature elevation of 8.54 °C for SP and 9.83 °C for LP taking under consideration that the experiment was done under stable temperature condition of 37 °C.
This temperature rise is only present for the minimal time of the pulse duration, so it is impossible to produce thermal damage on the apical region of the root canal. We must also consider that the blood vessels of the area will accelerate the heat conductivity in an in vivo situation, and thus helping the periodontal tissues to cool down faster and the expected temperature elevation will be even smaller [26].
Conclusion
Based on the conditions of this study, the Nd:YAG laser application in the root canal will not increase the temperature of surrounding tissue more than 10 °C which is the threshold for the occurrence of evident bone tissue damage according to the study of Eriksson and Albrektsson [27], since there is only a mean temperature elevation of 8.54 °C for SP and 9.83 °C for LP in apical points.
There was no significant difference in the observed temperatures in coronal and mesial areas. The results of the study demonstrated that the temperature rises during Nd:YAG laser irradiation at the parameters used in the study are minimal to cause damage on bone and periodontal tissues.
References
Hargreaves KM, Goodis HE (2002) Dental pulp. Quintessence, Chicago
Manogue M, Patel S, Walker R (2005) The principles of endodontics. Oxford University Press, Oxford
Pong-Yin Ng R (2004) Sterilization in root canal treatment: current advantages. HongKong Dental Journal 1:52–7
Berutti E, Marini R et al. Penetration ability of different irrigants into dentinal tubules. J Endod. 1997; 23: 725
Gutknecht N (1999) Lasertherapie in der zahnartzlichen praxis. Qintessenz Verlag, Berlin
Gutknecht N, Franzen R, Meister J, Vanweersch L, Mir M (2005) Temperature evolution on human teeth root surface after diode laser assisted endodontic treatment. Lasers Med Sci 20:99–103
Gutknecht N, Kaiser F, Hassan A, Lambert F (1996) Long term clinical evaluation of endodontically treated teeth by Nd:YAG lasers. J Clin Laser Med Surg 14(1):7–11
Kouchi Y, Ninimiya J, Yasuda H, Fukui K, Moriyama T, Okamoto H (1980) Location of Streptococcus mutans in the dentinal tubules of open infected root canals. J Dent Res 59(12):2038–2046
Lindhe J (1998) Clinical periodontology and implant dentistry, 3rd edn. Munksgaard, Copenhagen
Walsh LJ (2003) The current status of laser applications in dentistry. Aust Dent J 48(3):146–155
Kimura Y, Wilder-Smith P, Matsoumoto K (2000) Laser in endodontics: a review. Int Endod J 33:173–185
Klinke T, Klimm W, Gutknecht N (1997) Antibacterial effect of Nd:YAG laser irradiation within root canal dentin. J Clin Laser Med Surg 15(1):29–31
Gutknecht N, Moritz A, Conrads G, Sievert T, Lampert F (1996) Bactericidal effect of the Nd:YAG laser in in vitro root canals. J Clin Laser Med Surg 14:77–80
Rooney J, Midda M, Leeming J (1994) A laboratory investigation of the bactericidal effect of a Nd:YAG laser. Br Dent J 176(2):61–64
Ramsköld LO, Fong CD, Strömberg T (1997) Thermal effects and antibacterial properties of energy levels required to sterilize stained root canals with an Nd:YAG laser. J Endodont 23(2):69–100
Goodis HE, White JM, Marshall GW Jr, Yee K, Fuller N, Gee L, Marshall SJ (1997) Effects of Nd: and Ho:yttrium-aluminium-garnet lasers on human dentine fluid flow and dental pulp-chamber temperature in vitro. Archives of Oral Biology 42(12):845–854
Lan WH (1999) Temperature elevation on the root surface during Nd:YAG laser irradiation in the root canal. J Endod 25(3):155–156
Moritz A, Schoop U, Goharkhay K, Jakolits S, Kluger W, Wernisch J, Sperr W (1999) The bactericidal effect of Nd:YAG, Ho:YAG and Er:YAG laser irradiation in the root canal: an in vitro comparison. J Clin Laser Med Surg 17(4):161–164
Shoop U, Kluge W, Moritz A, Nedjelik N, Georgopoulos A, Sperr W (2004) Bactericidal effect of different laser systems in deep layers of dentin. Lasers Surg Med 35:111–116
De Moraes FG, Bramante CM, De Moraes IG, Carneiro E, Menezes R (2004) Influence of EDTA, Nd:YAG laser and association of both on the filling of artificial lateral root canals. J Appl Oral Sci 12(1):22–26
Arisu H, Bala O, Alimzhanova G, Turkoz E (2004) Assessment of morphological changes and permeability of apical dentin surfaces induced by Nd:YAG laser irradiation through retrograde cavity surfaces. J Contemp Dent Pract 5(2):102–113
Kimura Y, Yonaga K, Murakoshi M, Yokoyama K, Watanabe H, Matsoumoto K (2004) Effects on periradicular periodontal tissues of root canal irradiation with Er:YAG laser in rats. Photomed Laser Surg 22(4):335–341
Ishizaki NT, Matsoumoto K, Kimura Y, Wang X, Kinoshita W, Okano SM, Jayawardena JA (2004) Thermographical and morphological studies of Er, Cr:YSGG laser irradiation on the root canal walls. Photomed Laser Surg 22(4):291–297
Lea SC, Landini G, Walmsley AD (2004) Thermal imaging of ultrasonic scaler tips during tooth instrumentation. J Clin Periodontology 31:370–375
Moritz A, Gutknecht N, Goharkhay K, Schoop U, Wernisch J, Sperr W (1997) In vitro irradiation of infected root canals with diode laser: results of microbiologic, infrared spectrometric and stain penetration examinations. Quintessence Int 28(3):205–209
W O’Brien (1996) Thermal conductivity, biomaterials properties database, University of Michigan. Quintessence, Chicago
Eriksson A, Albrektsson T, Grane B, McQueen D (1982) Thermal injury to bone. A vital-microscopic description of heat effects. Int J Oral Surg 11(2):115–121
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Strakas, D., Franzen, R., Kallis, A. et al. A comparative study of temperature elevation on human teeth root surfaces during Nd:YAG laser irradiation in root canals. Lasers Med Sci 28, 1441–1444 (2013). https://doi.org/10.1007/s10103-012-1203-3
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10103-012-1203-3