Abstract
A 64-year-old woman underwent aortic valve replacement with a 21-mm Advancing The Standard (ATS) open-pivot mechanical heart valve for bicuspid aortic valve stenosis. In addition to the appearance of a new cardiac murmur, echocardiography performed 3 years after surgery showed a high pressure gradient across the ATS valve and a reduction in the valve orifice area. Cineradiography of the valve revealed restricted leaflet opening. Subsequent multidetector-row computed tomography clearly demonstrated pannus overgrowth on the inflow aspect of the ATS valve. During a repeat operation, subvalvular overgrown pannus was confirmed and the ATS valve was replaced with a bioprosthetic valve. This is the first reported case of prosthetic valve dysfunction resulting from pannus formation in a patient with an ATS valve in the aortic position.
Article PDF
Similar content being viewed by others
Avoid common mistakes on your manuscript.
References
N Vitale A Renzulli L Agozzino A Pollice N Tedesco LDLT Schinosa M Cotrufo (1997) ArticleTitleObstruction of mechanical mitral prostheses: analysis of pathological findings Ann Thorac Surg 63 1101–1106 Occurrence Handle9124913 Occurrence Handle1:STN:280:ByiB2cfjt1M%3D
H Teshima N Hayashida H Yano M Nishimi E Tayama S Fukunaga H Akashi T Kawara S Aoyagi (2003) ArticleTitleObstruction of St. Jude Medical valves in the aortic position: histology and immunohistochemistry of pannus J Thorac Cardiovasc Surg 126 401–407 Occurrence Handle10.1016/S0022-5223(03)00702-5 Occurrence Handle12928636
S Aoyagi Y Nishimi H Kawano E Tayama S Fukunaga N Hayashida H Akashi T Kawara (2000) ArticleTitleObstruction of St. Jude Medical valves in the aortic position: significance of a combination of cineradiography and echocardiography J Thorac Cardiovasc Surg 120 142–147 Occurrence Handle10.1067/mtc.2000.106524 Occurrence Handle10884667 Occurrence Handle1:STN:280:DC%2BD3czovVWisA%3D%3D
S Aoyagi T Kawara S Fukunaga T Mizoguchi Y Nishi H Kawano K Arinaga (1997) ArticleTitleCineradiographic evaluation of ATS open pivot bileaflet valve J Heart Valve Dis 6 258–263 Occurrence Handle9183724 Occurrence Handle1:STN:280:ByiA38nptFU%3D
DJ Radford PG Pohlner (2002) ArticleTitleFormation of pannus on prosthetic valves in a child with pseudoxanthoma elasticum Cardiol Young 12 183–185 Occurrence Handle12018727
H Teshima N Hayashida S Fukunaga E Tayama T Kawara S Aoyagi M Uchida (2004) ArticleTitleUsefulness of multidetector-row computed tomography scanner for detecting pannus formation Ann Thorac Surg 77 523–526 Occurrence Handle10.1016/S0003-4975(03)01531-5 Occurrence Handle14759431
RW Emery CC Krogh DJ Jones DM Nicoloff DP Blake KV Arom (2004) ArticleTitleFive-year follow up of the ATS mechanical heart valve J Heart Valve Dis 13 231–238 Occurrence Handle15086262
GJ Van Nooten Y Van Belleghem F Caes K Francois H Van Overbeke T Bove Y Taeymans (2003) ArticleTitleLower-intensity anticoagulation for mechanical heart valves: a new concept with the ATS bileaflet aortic valve J Heart Valve Dis 12 495–502 Occurrence Handle12918853
T Mori M Asano H Ohtake A Bitoh S Sekiguchi Y Matsuo M Aiba M Yamada T Kawada T Takaba (2002) ArticleTitleAnticoagulant therapy after prosthetic valve replacement – optimal PT-INR in Japanese patients Ann Thorac Cardiovasc Surg 8 IssueID2 83–87 Occurrence Handle12027793
H Koertke K Minami Th Breymann D Seifert O Wagner N Atmacha A Krian J Ennker U Taborski WP Klövekorn R Moosdorf W Saggau R Koerfer (2003) ArticleTitleINR self-management permits lower anticoagulation levels after mechanical heart valve replacement Circulation 108 IssueIDsuppl II II-75–II-78 Occurrence Handle1:STN:280:DC%2BD3svivV2itA%3D%3D
A Renzulli L De Luca A Caruso R Verde D Galzerano M Cotrufo (1992) ArticleTitleAcute thrombosis of prosthetic valves: a multivariate analysis of the risk factors for a lifethreatening event Eur J Cardiothorac Surg 6 412–420 Occurrence Handle10.1016/1010-7940(92)90065-6 Occurrence Handle1389247 Occurrence Handle1:STN:280:ByyD3c7isVE%3D
E Tayama Z Feng H Oda H Tomoeda N Hayashida S Fukunaga M Umezu S Aoyagi (2000) ArticleTitleATS prosthetic valve motion: an in vitro analysis J Heart Valve Dis 9 408–414 Occurrence Handle10888099 Occurrence Handle1:STN:280:DC%2BD3M%2FhsVOktA%3D%3D
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Teshima, H., Aoyagi, S., Hayashida, N. et al. Dysfunction of an ATS valve in the aortic position: the first reported case caused by pannus formation. J Artif Organs 8, 270–273 (2005). https://doi.org/10.1007/s10047-005-0317-9
Received:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/s10047-005-0317-9