Abstract
We report on newly synthesized Ag(I)-filled multiwall carbon nanotubes as a potential sensing element in ion-selective carbon paste electrodes for the determination of free cyanide in aqueous solution. The electrode was obtained by entrapping the silver-filled nanotubes into a carbon paste and displays a Nernstian response with a slope of 59.8 ± 0.3 mV decade−1, a very wide linear range (from 21.0 nM to 0.1 M of cyanide), a lower detection limit of 13.0 nM, and a response time of <2 min. The operational lifetime is up to 3 months without significant deviation in normal function.
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Avoid common mistakes on your manuscript.
Introduction
Cyanide is one of the most toxic chemical substances on Earth and is toxic to most aquatic life and humans, even at low concentrations. The importance of cyanides in technological and industrial fields is well known from using to extract gold and silver from their mines to key role as constituents of amino acids in animal and plant structures [1–3]. The high degree of toxicity that these compounds present to living organisms alone warrants their study [2–4]. Some procedures have been reported for measurement of cyanides, which generally suffer from high experimental steps, unfavorable precision, low accuracy, and demand for fast operations [4–12].
Carbon paste electrodes (CPEs) are widely used in various electrochemical applications due to their simplicity, wide potential range for analysis, non-toxic character, easily preparation and easy surface renewing [13–16]. Furthermore, all contacts in this type of electrode are solid-state which provides maintenance-free towards common ion sensors to eliminate the inner filling solution. Various electrochemical detection schemes have been reported by employing different inorganic and organic modifiers in carbon pasts [17–22]. Although considerable attention has been given to preparation of CPEs so far, the applications of these CPEs mainly have been focused on the field of voltammetric determination and only few of these types of the electrodes have been used in potentiometry.
Among modifiers, carbon nanotubes (CNTs) have become the most outstanding material in most recent works on the determination of chemical species employing CNTs as sensing materials because the unique properties of CNTs make them very attractive for operation of chemical sensors in general, and electrochemical detection in particular [23–26]. On the other hand, substances that are incorporated into the channels of CNTs can take part in chemical reactions. Filling of CNTs with proper species has recently become a very interesting field in nanotube science. This can be accomplished either during or after their synthesis. By thermal decomposition of oxides and their reduction, metal-containing nanotubes were obtained [27]. The inner spaces of tubes could be filled by chemical deposition from the gas phase by utilizing, e.g., volatile metal compounds [26, 28].
So far, to the best of our knowledge, there is no report on the study of filled CNTs used in CPEs. This is the first report on construction a carbon paste selective electrode based on a filled CNTs by potentiometric method. In addition, this is the first report on determination of highly toxic cyanide ion by the proposed system. Few reports for determination of cyanide by CPEs are available in literature [29–31]. In this study, we are describing the use of a newly synthesized Ag(I)-filled-multiwall carbon nanotube (Ag-MWCNT) as a sensing ion-selective element in the construction of a novel CPE for determination of free cyanide ion in aqueous solution.
Experimental
Reagents
Graphite powder and paraffin oil were purchased from Fluka (www.sigmaaldrich.com), Switzerland. Multiwall carbon nanotubes (MWCNT), filled with Ag(I), prepared as described elsewhere [32]. Sodium and potassium of all anions were of analytical reagent grade from Merck (www.merck-chemicals.com). Solutions were prepared with doubly distilled and deionized water.
Apparatus
All potentiometric measurements were recorded on a CRISON BASIC 20 (www.crison.es, Spain) pH/mV-meter. The reference electrode was a silver/silver chloride electrode. A Jenway pH-meter (UK 3020, www.jenway.com) was used for pH measurements. All potentiometric measurements were carried out by using the following assembly.
Preparation of Ag-MWCNT/CPE
The ion selective electrode was prepared by mixing 4 mg Ag-MWCNT and 80 mg pure graphite powder, which was completely mixed to obtain a homogenized powder. Then, 50 mg of Paraffin oil was added to the mixture and mixed fully until it was uniformly wetted. Subsequently, the carbon paste (CP) was packed into a polypropylene electrode holder (inner diameter of 3 mm) and the electrical contact was established with a copper rod through the back of the electrode (Fig. 1). Before using, the electrode surface was polished by a page of paper contaminated with alumina powder to produce a reproducible working surface.
Measuring procedure
The electrodes were immersed directly into the test solutions. A citrate buffer (0.015 M) was used to adjust the pH of the solutions at 10.0 ± 0.1. Then proper amounts of potassium cyanide solution were added to the test solution and temperature was controlled at 20.0 ± 0.1 °C with a water-circulating bath system. The solution was stirred immediately until the response of the system reached constant (about 2 min). The potentiometric data was gathered after each cyanide solution addition versus the reference electrode. Highly toxic cyanide ion can be adsorbed on the walls of glass vessels, so all experiments were carried out in a polyethylene beaker.
Results and discussion
As reported by [30], the Ag+ ion and its nitrate counter ion are in the MWCNT channel where the electrostatic ionic forces are operating. This fact could be due to the requirement for electro-neutrality of the system. It would be a good idea if we provided an optimal condition in which the nitrate ion be exchanged with another proper ion [33–35]. This is the basic idea to prepare a CPE for which nitrate anion could be replaced by cyanide ion in an exchange fashion at the electrode-solution interface. Visualization of such a mechanism is demonstrated in Fig. 2.
It is well known that several parameters must be considered during the development of a useful ion-selective electrochemical sensor.
Effect of composition
Influence of carbon paste and membrane compositions on the ISEs’ performances are crucial. Thus, different aspects of the composition of the proposed CPE based on Ag-MWCNT were optimized for determination of cyanide ion. For this purpose, some electrodes with different compositions of basic compounds were prepared. The used ratios of the reactants showed that the corresponding slopes and dynamic ranges of response calibration curves were closely depended on the proportion of the modifier in the electrode. According to the results that are shown in Table 1, there was no considerable response for the electrode in which no sensing element was used (No. 1). The effect of the amount of Ag-MWCNT on the electrode performances were investigated (Nos. 2–4). Electrodes 3, 5 and 6 are showing the influence of the paraffin ratio on the electrode response. Electrodes 3, 7 and 8, demonstrate the graphite effect. From this table, optimum amounts of the reactants are 4:50:80 (mg, No. 3) for Ag-MWCNT:paraffin:graphite, respectively.
Response time of proposed electrode
From the literature and our preliminary experiments, the response time value (t 95%) of the sensor, when reaches to 95% of its final value (steady state), depends on the sensing membrane composition [36]. The electrode response time was evaluated by measuring the average time while it was in a 10−4 M solution of cyanide ion. Reading the response potential was started from the time of addition of cyanide solution to the time after which the electrode response reached a steady state, for which the results are plotted in Fig. 3. This time (<2 min) was evaluated from the intersection of the two extrapolated segments of the response time curve (by creating proper regression equations from the data). Thus, the response time of 2 min was used for subsequent potential readings.
Effect of pH
As an important factor, pH value influences the determination of cyanide ion directly. Cyanide is a strong base (pK b = 4.79) and at pH values below 9.5, cyanide species is converted into a volatile and highly toxic weak acid, HCN. Thus, the effect of pH on the potential response of the electrode (No. 3) was evaluated from a fixed concentration of CN− (1.0 × 10−4 M) in the solution at different pH values. The desired pH was adjusted by adding small volumes of concentrated hydrochloric acid or sodium hydroxide solutions to the test solution and the variation of potential was followed after each addition. The potential variations were plotted as a function of pH, as seen in Fig. 4. From this figure, the response of the electrode remains unchanged in the pH range of 8–12 and the electrode works satisfactorily in this pH range. At pH values lower than 8 the response of the electrode influences from the pH of the solution tested that can be due to HCN formation and reduction of free cyanide mole fraction. So pH = 10.0, which was prepared from citric acid/sodium citrate buffer solution (0.015 M), was used as an optimum pH in subsequent potentiometric measurements.
Analytical performances
The critical response characteristics of the cyanide—selective electrode (No. 3) were investigated according to IUPAC recommendations [37, 38]. A series of standard solutions of CN− were examined by the electrode. The potential responses of the electrode at varying CN−concentrations follows a linear trend for concentrations ranging from 2.1 × 10−8 to 1.0 × 10−1 M with a near-Nernstian slope of 59.8 ± 0.3 mV decade−1, which are demonstrated in Fig. 5.
The detection limit (DL) was estimated to be 1.3 × 10−8 M of CN− which was evaluated from the intersection of the two extrapolated segments of the above calibration plot at the lowest portion.
In order to obtain a comparison between the behaviors of the electrode (No. 3) toward some common anions and cyanide ion, the electrode performances were separately examined in the presence of some other sodium salts of anions such as thiocyanate, chloride, iodide, hydroxide and acetate. As shown in Fig. 6, the electrode response to cyanide ion is reasonably quantitative.
The selectivity power of the system was investigated by estimating potentiometric selectivity coefficients for different anions by using fixed interference method [38–40]. The potential responses of a cell including the ISE and a reference electrode (Ag/AgCl) were measured for solutions of constant activity of the interfering ion (0.01 M), a B , and varying activity of the primary ion, a A . The obtained potential values were plotted versus the logarithm of the activity of the primary ion. The intersection of the extrapolated linear portions of this plot indicated the value of a A to calculate selectivity coefficients from the following equation (Eq. 1):
where z is the charge of ion. The results are summarized in Table 2. From this table, it is obviously seen that the cyanide—selective electrode is quite selective with respect to some chemical species were tested to check their levels of interference in cyanide determination. Table 2 also shows that the selectivity coefficients are small enough for all tested diverse ions indicating they would not significantly disturb the functioning of the modified electrode, so it possesses a high selectivity. This highly selective behavior of the electrode toward CN– could be attributed to the capability of Ag+ ion to form well-known complex(es) with CN– as AgCN –2 and Ag[AgCN2].
The practical reversibility and precision required for the sensor, to reach a potential within ±1 mV of the final equilibrium value, were investigated. This was evaluated by potential reading for the electrode dipped alternatively into two different solutions of cyanide ion 1.0 × 10−3 and 1.0 × 10−5 M, respectively. Six potentiometric experiments were repeatedly carried out with the same CPE. The resulted relative standard deviations (RSDs, (%)) of ± 0.5 and ± 0.7 were calculated for the lower and the upper CN− concentrations (for 15 replicate measurements), respectively. These RSDs show that the electrode had unchanged its sensing behavior while the potentials were being recorded either from the low to the high concentrations or vice versa. The results are demonstrated in Fig. 7. Furthermore, the system responses are reasonably repeatable over the potentiometric measurements. As this figure reveals, the electrode potential responses are quite reversible for reading the potential of each solution.
Reproducibility of the system was tested when several electrodes with the same compositions were immersed in the same standard solution of cyanide ion. The responses were nearly the same (within ±5 mV) which indicate there is a good reproducibility.
The lifetime of the modified electrode was also tested over a period of 3 months. During this period, the electrode was in daily use over extended period of time (1 h a day) and only a small decrease of potential sensitivity (less than 4%) for 1.0 × 10−4 M CN− was observed, which can be attributed to the excellent stability of the sensor.
As can be realized from the above-mentioned characteristics of the optimized electrode, a Nernstian response, low DL, a wide calibration range with good reversibility and reasonable reproducibility are the suitability of this electrode to determine the cyanide free ion, within the offered concentration range, in solution.
Applications
Samples of aqueous solutions including cyanide ion were collected from different sources, an electroplating manufactory waste, a photographic film development waste solution and a cyanide added tap water. After filtration to remove suspension solids and some organic substances the concentration values of CN− in the samples were determined by using the proposed method. The cyanide contents of the samples were also determined by a standard spectrophotometric method [11]. The concentration values obtained from both methods (for six repeated measurements, n = 6) were listed in Table 3. Obvious differences were not found between the results obtained by the electrode and the standard spectrophotometric method. Furthermore, the evaluated RSDs, average 0.80% for all measurements in Table 3, are small enough (comparable with the standard method, 1.0%) which are expected for such valuable potentiometric electrode.
Conclusions
In this work, we applied a modified carbon paste electrode (MCPE) for determination of free cyanide anion in aqueous solution using potentiometric method. According to our knowledge, this is the first report on cyanide selective electrode based on the chemically MCPE using potentiometric method. Not only the method is simple and inexpensive, the electrode has the advantages of being selective for cyanide in the presence of variety of anions such as SCN−, I−, Cl−, Br− and oxalate that usually are serious interfering species for most of cyanide ISEs. In addition, application of a new type of materials, a silver(I)-filled carbon nanotube develops the area of employment of modified CNTs as sensing elements in ion selective devices and promising applications in analytical chemistry for constructing CNT-based sensors.
References
Rosentreter JJ, Skogerboe RK (1991) Trace determination and speciation of cyanide ion by atomic absorption spectroscopy. Anal Chem 63:682–688
Zagury GJ, Oudjehani K, Deschenes L (2004) Characterization and availability of cyanide in solid mine tailings from gold extraction plants. Sci Total Environ 320:211–224
Zaknun JJ, Stieglbauer K, Trenkler J (2005) Cyanide-induced a kinetic rigid syndrome: clinical, MRI, FDG-PET, β-CIT and HMPAO SPECT findings. Parkinsonism Relat Disord 11:125–129
Karmarkar S (2002) Anion-exchange chromatography of metal cyanide complexes with gradient separation and direct UV detection. J Chromatogr A 956:229–235
Mirmohseni A, Alipour A (2002) Construction of a sensor for the determination of cyanide in industrial effluents: a method based on quartz crystal microbalance. Sens Actuators B 84:245–251
Anzenbacher P, Tyson DS, Jursíková K, Castellano FN (2002) Luminescence lifetime-based sensor for cyanide and related anions. J Am Chem Soc 124:6232–6233
Suzuki T, Hioki A, Kurahashi M (2003) Development of a method for estimating an accurate equivalence point in nickel titration of cyanide ions. Anal Chim Acta 476:159–165
Weinberg HS, Cook SJ (2002) Segmented flow injection, UV digestion, and amperometric detection for the determination of total cyanide in wastewater treatment plant effluents. Anal Chem 74:6055–6063
Chattaraj S, Das AK (1991) Indirect determination of free cyanide in industrial waste effluent by atomic absorption spectrometry. Analyst 116:739–741
Tracqui A, Raul JS, Geraut A (2002) Determination of blood cyanide by HPLC-MS. J Anal Toxicol 26:144–148
Yari A, Kargosha K (2006) Simple photometric determination of free cyanide ion in aqueous solution with 2,6-dichlorophenolindophenol. Centr Europ J Chem 4:329–337
Dadfarnia S, Shabani AMH, Tamadon F, Rezaei M (2007) Indirect determination of free cyanide in water and industrial waste water by flow injection-atomic absorption spectrometry. Microchim Acta 158:159–163
Ravichandran K, Baldwin RP (1981) Chemically modified carbon paste electrodes. J Electroanal Chem 126:293–300
Wang J, Kirgoz UA, Mo JW, Lu J, Kawde AN, Muck A (2001) Glassy carbon paste electrodes. Electrochem Commun 3:203–208
Ravichandran K, Baldwin RP (1983) Phenylenediamine-containing chemically modified carbon paste electrodes as catalytic voltammetric sensors. Anal Chem 55:1586–1591
Kalcher K, Grabec I (1990) Chemically modified carbon paste electrodes in voltammetric analysis. Electroanalysis 2:419–433
Ivan S, Carel V, Jiri B, Jiri Z (2001) Carbon paste electrodes in modern electroanalysis. Crit Rev Anal Chem 31:311–345
Walcarius A, Bessiere J (1997) Silica-modified carbon paste electrode for copper determination in ammoniacal medium. Electroanalysis 9:707–713
Arrigan DWM (1994) Voltammetric determination of trace metals and organics after accumulation at modified electrodes. Analyst 119:1953–1966
Fiorucci AR, Cavalheiro ETG (2002) The use of carbon paste electrode in the direct voltammetric determination of tryptophan in pharmaceutical formulations. J Pharmac Biomed Anal 28:909–915
Vytras K, Svancara I, Metelka R (2009) Carbon paste electrodes in electroanalytical chemistry. J Serb Chem Soc 74:1021–1033
Wildgoose GG, Banks CE, Leventis HC, Compton RG (2006) Chemically modified carbon nanotubes for use in electroanalysis. Microchim Acta 152:187–214
Rao CN, Satishkumar BC, Govindaraj A, Nath M (2001) Nanotubes Chem Phys Chem 2:78–105
Baughman RH, Zakhidov A, de Heer WA (2002) Carbon nanotubes: the route toward applications. Science 297:787–792
Zhao Q, Gan Z, Zhuang Q (2002) Electrochemical sensors based on carbon nanotubes. Electroanalysis 14:1609–1613
Wang J, Kawde AN, Musameh M (2003) Carbon-nanotube-modified glassy carbon electrodes for amplified label-free electrochemical detection of DNA hybridization. Analyst 128:912–916
Cook J, Sloan J, Green MLH (1997) Opening and filling carbon nanotubes. Fullerene Sci Technol 5:695–704
Sinha AK, Hwang DW, Hwang LP (2000) A novel approach to bulk synthesis of carbon nanotubes filled with metal by a catalytic chemical vapor deposition method. Chem Phys Lett 332:455–460
Ulakhovich NA, Zakieva DZ, Yu GG, Budnikov GK (1993) Voltammetric determination of cyanide ions using preconcentration at a carbon-paste electrode modified with liquid crystals. J Anal Chem (Transl of Zh Anal Khim) 48:731–733
Amine A, Alafandy M, Kauffmann JM, Pekli MN (1995) Cyanide determination using an amperometric biosensor based on cytochrome oxidase inhibition. Anal Chem 67:2822–2827
Abbaspour A, Asadi M, Ghaffarinejad A, Safaei E (2005) A selective modified carbon paste electrode for determination of cyanide using tetra-3,4-pyridinoporphyrazinatocobalt(II). Talanta 66:931–936
Sepahvand R, Adeli M, Astinchap B, Kabiri R (2008) New nanocomposites containing metal nanoparticles carbon nanotube and polymer. J Nanopart Res 10:1309–1318
Liu H, Murad S, Jameson CJ (2006) Ion permeation dynamics in carbon nanotubes. J Chem Phys 125:084713–084726
Beu TA (2010) Molecular dynamics simulations of ion transport through carbon nanotubes. I. Influence of geometry, ion specificity, and many-body interactions. J Chem Phys 132:164513–164528
Sumikama T, Saito S, Ohmine I (2006) Mechanism of ion permeation in a model channel: free energy surface and dynamics of K+ ion transport in an anion-doped carbon nanotube. J Phys Chem B 110:20671–20677
Lindner E, Toth K, Pungor E (1986) Definition and determination of response time of ion selective electrodes. Pure Appl Chem 58:469–479
Buck RP, Lindner E (1994) Recommendations for nomenclature of ion-selective electrodes. Pure Appl Chem 66:2527–2536
Macca C (2004) Response time of ion-selective electrodes: current usage versus IUPAC recommendations. Anal Chim Acta 512:183–190
Umezawa Y (1990) Handbook of ion selective electrodes: selectivity coefficients. CRC Press, Boca Raton
Pretsch E, Bakker E, Buhlmann P (2000) Selectivity of potentiometric ion sensors. Anal Chem 72:1127–1133
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Yari, A., Sepahvand, R. Highly sensitive carbon paste electrode with silver-filled carbon nanotubes as a sensing element for determination of free cyanide ion in aqueous solutions. Microchim Acta 174, 321–327 (2011). https://doi.org/10.1007/s00604-011-0629-9
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00604-011-0629-9