In 1958, Adams [1] first introduced the carbon paste electrode (CPE) which usually comprises a Teflon bore, into which is inserted a copper, steel or carbon contact. The bore is filled with a paste made by mixing powdered carbon with an adequate mineral oil. In 1964, Kuwana and French [2] added modifier materials to the carbon paste in order to improve the selectivity and sensitivity of CPE. Since then, modifier materials have been widely researched in electroanalysis and in the context of chemical sensors [35]. In recent years, nanomaterials with special physical and chemical properties have been an attractive field of research [69]. Gold nanoparticles, one of the new promising materials, were used to modify electrodes due to their stable physical and chemical properties, small dimensional size and useful catalytic activities [1012].

Acetaminophen (N-acetyl-p-aminophenol, APAP) is an extensively used antipyretic analgesic used in the symptomatic management of pain and fever [13]. Several methods have been proposed for the determination of acetaminophen, including titrimetry [14, 15], fluorimetry [1618], spectrophotometry [1922], chromatography [2326], Fourier transform infrared spectrometry [27, 28], voltammetry [29, 30], capillary electrophoresis [31, 32], colorimetry [33] and flow-injection systems with colorimetric detection, and spectrophotometric detection [34]. However, the above mentioned methods have some disadvantages, for example, relatively expensive instruments, time-consuming procedures, and inconvenient operation. Flow injection analysis (FIA) with amperometric detection is an alternative method for the acetaminophen determination primarily because of the generally simple, inexpensive instrumentation and time-saving operation. In this paper, a gold nanoparticles modified carbon paste electrode (GNMCPE) was applied to FIA with an in-house amperometric detector for the determination of acetaminophen. The GNMCPE has some advantages, such as easy preparation, relatively fast response and low cost. The method of FIA-GNMCPE was applied to the determination of acetaminophen in six commercial pharmaceuticals. A satisfactory result was obtained which is consistent with that of high performance liquid chromatography (HPLC).

Experimental

Chemicals and reagents

All chemicals were of analytical grade and used without further purification. Sodium dihydrogen phosphate, sodium hydroxide, hydrochloric acid, anhydrous ethanol, acetone, ascorbic, urea and glucose were purchased from Chengdu Kelong Chemical Reagent Company (Chengdu, China). Acetaminophen was obtained from the National Institute for the Control of Pharmaceutical and Biological Products (Beijing, China). Carbon powder (particle size of 60 μm) and heavy mineral oil were purchased from Shanghai Chemical Reagent Company (Shanghai, China). Hydrogen tetrachloroaurate (HAuCl4) was purchased from Sinopharm Chemical Reagent. Co., Ltd. (Shanghai, China).

All solutions were prepared using deionized water. Phosphate buffer solutions were prepared from sodium dihydrogen phosphate and adjusted to pH with 30% sodium hydroxide solution and 3 mol L−1 hydrochloric acid. The solution of standard acetaminophen was freshly prepared by dissolving 50 mg acetaminophen in 50 mL of anhydrous ethanol, and diluting to 100 mL with deionized water.

Apparatus

Electrochemical measurements were carried out with a microcomputer-based electrochemical analyzer (LK98BII, Tianjin Lanlike Chemical and Electron High Technology Co., Ltd, Tianjin, China, http://lanlike.instrument.com.cn). The amperometric measurements were performed in a flow-through cell with three electrodes (Fig. 1): a GNMCPE or a bare CPE as the working electrode, a saturated calomel electrode (SCE) as the reference electrode, and a platinum electrode as the auxiliary electrode. The amperometric detection flow-through cell was fabricated with a polyethylene tube, silicon rubber tube and polytetrafluoroethylene (PTFE) pipe with an inner diameter of 0.8 mm and an outer diameter of 1 mm. X-ray diffraction (XRD) analysis was performed on a Dandong Fangyuan 800 X-ray diffractometer (Dandong, China) with Cu Kα radiation. The transmission electron microscopy (TEM) images were obtained from a Hitachi H-800 transmission electron microscope (Tokyo, Japan) operating at 200 kV. The HPLC system was an Agilent Technologies 1100 series purchased from Agilent Technologies Co. Ltd. (Shanghai, China). The flow injection analysis system was supplied by Xintong Science Instrument Company (Beijing, China). The sonicator was purchased from Kunshan Ultrasonic Instrument Co. Ltd. (KQ218, 100W, 40 kHz, Jiangsu, China, http://www.ks-csyq.com). The microwave oven was purchased from Panasonic Electric Industrial Co., Ltd (Panasonic National, 950 W, Osaka, Japan).

Fig. 1
figure 1

Schematic diagram of the amperometric detection flow-through cell used in the flow injection system. 1 Platinum auxiliary electrode; 2 modified carbon paste electrode; 3 saturated calomel reference electrode (SCE); 4 polyethylene tube; 5 flow inlet; 6 PTFE pipe; 7 copper rod; 8 polyethylene tube; 9 flow outlet; 10 silicon rubber tube; 11 carbon paste

Procedures

Fabrication of gold nanoparticles modified carbon powder

A given weight of pure carbon powder was first carefully washed with acetone, ethanol and acetone in sequence, then three times with deionized water, transferred into an oven and dried at 80 °C. The dried powder was immediately separated into four parts and each part placed into a small beaker. The HAuCl4 solution of equal volume with different concentrations (1.3 × 10−3, 2.6 × 10−3, 5.2 × 10−3 and 9.6 × 10−3 mol L−1) was added into each beaker, respectively, and mixed simultaneously. Then the four beakers were kept in a microwave oven and heated for 5 min. The modified carbon powders were washed with deionized water until the yellow-green colors of the decanted solution had disappeared completely. Finally the products were kept in an oven and dried at 80 °C to obtain the gold nanoparticles modified carbon powder.

Preparation of the GNMCPE

GNMCPE was prepared as follows: firstly, 0.5 g of gold nanoparticles modified carbon powder and 0.18 mL of mineral oil were carefully mixed. Then, the mixture was homogenized by carefully mixing in an agate mortar and pressing with a pestle. Finally, the well-prepared paste was packed into a polyethylene cannula (inner diameter 2.5 mm, outer diameter 4.0 mm, length 70 mm) equipped with a copper rod (diameter 2.5 mm) as an external electric contact. This rod can move up and down to renew the surface of the electrode. The bare CPE was prepared in a similar way.

A shiny surface was obtained by smoothing the electrode onto a clean weighing paper, and then washing with deionized water several times. After several times of use, a new electrode surface can be obtained by piling out a small amount of the paste, rubbing away the excess against a clean weighing paper and polishing the electrode on the paper to obtain a shiny surface again.

FIA with amperometric detection for acetaminophen

As shown in Fig. 2, the flow injection analysis system consisted of a flow-through cell for amperometric detection, a main peristaltic pump, an auxiliary peristaltic pump, a sampling valve and an injection port with a 150 μL injection loop.

Fig. 2
figure 2

Schematic diagram of the flow injection system used for the determination of acetaminophen. P1 Main peristaltic pump; P2 auxiliary peristaltic pump; C1 phosphate buffer solution; C2 deionized water; IV injection valve; S sample; RC reaction coil; D amperometric detector; W waste

Deionized water was used as the carrier solution with a flow rate of 4.0 mL min−1, phosphate buffer solution (0.1 mol L−1, pH 4.7) was regulated by a reagent delivery module at a flow rate of 4.0 mL min−1.

Preparation of samples

All commercial samples were purchased from a local drugstore. These tablets were Riyebaifunin (Shanghai Ltd., Batch No. 0510565, Shanghai, China), Baijiahei (Jiangsu Ltd., Batch No. H10940251, Jiangsu, China), Gankang (Jilin Ltd., Batch No. H22026193, Jilin, China), Tainuo (Shanghai Ltd., Batch No. H20010115, Shanghai, China), Gandinuo (Xinjiang Ltd., Batch No. H19991427, Xinjiang, China) and Haiwangyindeifei (Shenzheng Ltd., Batch No. H44023557, Shenzheng, China). The tablet sample was crushed and powdered in a mortar, respectively. An 0.1 g amount of finely powdered sample was accurately weighed into a 100 mL flask, 50 mL of anhydrous ethanol was added before sonicating for 20 min, then the solution was diluted to volume with deionized water and filtered through an 0.45 μm membrane filter. Finally the filtrate was provided for assaying according to the method proposed in this paper and HPLC, respectively.

HPLC for acetaminophen

To evaluate the accuracy of the discussed method, the contents of acetaminophen in six commercial pharmaceuticals were determined by HPLC. The chromatographic parameters of the HPLC system are given in the electronic supported material S1.

Results and discussion

The properties of gold nanoparticles modified carbon paste and bare carbon paste

The XRD pattern of bare carbon paste (a) and gold nanoparticles modified carbon paste (b) is shown in Fig. 3. The characteristic peaks at 38.10°, 44.20° and 64.20° indicate that the modification of gold nanoparticles on carbon paste was successful. The morphologies of the bare carbon paste and gold nanoparticles modified carbon paste are also shown in Fig. 3, respectively. The average crystallite size of the gold nanoparticles was about 20 nm (Fig. 3B). When comparing Fig. 3A with Fig. 3B, the gold nanoparticles (two are marked with white arrows in Fig. 3B) throughout the carbon paste can be observed. Thus, we can safely draw the conclusion that gold nanoparticles had attached to the carbon paste electrode.

Fig. 3
figure 3

The XRD of the bare carbon paste (a) and gold nanoparticles modified carbon paste (b) electrode. Inset: the TEM micrograph of the bare carbon paste (A) and gold nanoparticles modified carbon paste (B) was verified

Electrocatalysis of acetaminophen at GNMCPE

The reaction of acetaminophen (APAP) is an electrochemical oxidation by a two-electron, two-proton process to generate N-acetyl-p-quinoneimine (NAPQI) [35, 36]. The oxidation mechanism is shown in Scheme 1.

Scheme 1
figure 1

The mechanism of electrochemical oxidation of acetaminophen

First the electrocatalytic activities of CPE and GNMCPE towards the oxidation of acetaminophen in 0.1 mol L−1 phosphate buffer solution of pH 4.7 were demonstrated. Figure 4 displays the cyclic voltammograms (CVs) of 50 mg L−1 acetaminophen in 0.1 mol L−1 phosphate buffer solution of pH 4.7 at CPE (dotted line) and GNMCPE (solid line) respectively. It was found that both exhibited well-defined peak currents. For the CPE, an anodic peak occurred at approximately +0.8 V (peak 1) versus Ag/AgCl. However, in the case of the GNMCPE, a very well-defined oxidation peak at approximately +0.75 V (peak 2) versus Ag/AgCl was observed. The anodic peak (peak 1 and peak 2) represented the oxidation of APAP to NAPQI, while the cathodic peak (peak 3 and peak 5) represented the reverse reaction (NAPQI to APAP). At lower pHs, NAPQI would easily be protonated as the electroactive intermediate (c), which is then hydrolyzed to form hydrated NAPQI (d) which is finally transformed into benzoquinone (e). The last reaction (step E) was an electrochemical reduction by a two-electron, two-proton process to produce hydroquinone (f). The peak appeared at nearly +0.05 V (peak 4) maybe because of the oxidation of hydroquinone to benzoquinone. In addition, the voltammograms obtained with the GNMCPE provided higher signal-to-background (S/B) ratios in the same electrode area. Furthermore, the cyclic voltammograms of acetaminophen solution at different scan rates are shown in Fig. S1. The peak current for anodic oxidation of acetaminophen was proportional to the square root of the scan rate. This illustrated that the electrochemical oxidation reaction was diffusion-controlled.

Fig. 4
figure 4

Cyclic voltammograms for 50 mg L−1 acetaminophen at pH 4.7 at CPE (dotted line) and GNMCPE (solid line). The sweep rate was 0.05 V s−1; area of electrode 0.049 cm2

We also studied the electrochemical properties of acetaminophen at GNMCPE with various concentrations of HAuCl4 (1.3 × 10−3, 2.6 × 10−3, 5.2 × 10−3 and 9.6 × 10−3 mol L−1) by cyclic voltammetry. The results showed that the oxidation potential of acetaminophen first increased and then decreased with the increase of HAuCl4 concentrations. In addition, the oxidation potential gradually shifted to a negative potential. Therefore, we chose the GNMCPE-modified concentration of HAuCl4 solutions to be 2.6 × 10−3 mol L−1 for the next experiments.

In order to improve its performance, various factors affecting the electrocatalytic activities of GNMCPE were investigated, including pH and sodium dihydrogen phosphate concentration. The effect of pH on the oxidation of acetaminophen was investigated over a pH range of 3.5 to 5.5. The results are presented in Fig. S2. The peak currents of acetaminophen first increased and then decreased with the increase in pH, and the maximum peak current was obtained when the pH was 4.7. The effect of phosphate buffer solution concentration was also investigated from 0.05 to 0.5 mol L−1 with a pH of 4.7. The results are presented in Fig. S3. It is clear that the maximum peak current was obtained when the concentration was 0.1 mol L−1. So 0.1 mol L−1 pH 4.7 phosphate buffer solution was selected and used for the following experiments.

Flow injection analysis with amperometric detection

In order to get the best response of the electrode in FIA, the following factors were optimized: flow rate, injection volume, injection time, and detection volume.

The effect of the flow rate on the response was verified using different flow rates (from 1.2 to 4.0 mL min−1) for the same sample volume. As can be seen from Fig. S4, as the flow rate increased, the peak current became higher and the shape of the peak became narrower. The maximum peak height was obtained with the flow rate of 4.0 mL min−1. So this flow rate (the maximum flow rate for the instrument) was selected. The sample injection volume varied from 150 to 400 μL (shown in Fig. S5). A volume of 150 μL was employed for further experiments. The effect of injection time on the magnitude of the amperometric response was investigated. The result shows that the peak current increased rapidly when increasing the injection time from 1 to 3 s (see Fig. S6). A further increase of the injection time has no effect on peak current. Thus, a 3 s injection time was chosen for further experiments. The amperometric response decreased with the increase of the volume of the flow-through cell from 30 to 80 μL (see Fig. S7). A cell volume of 30 μL was selected.

Once the conditions of the FIA procedure were optimized, amperometric detection of solutions containing different acetaminophen concentrations was carried out in phosphate buffer solution (pH 4.7) at the detection potential of +0.75 V versus Ag/AgCl. Figure 5 shows that the current values increased linearly with increasing acetaminophen concentrations. A linear response of the GNMCPE (Fig. 5a) to acetaminophen from 0.1 to 80 mg L−1 (ipGNMCPE (μA) = 0.1134 + 0.1123 C; r = 0.9994; n = 7) was obtained with a detection limit (3σ) of 0.05 mg L−1, where the unit of C was milligrams per liter. The relative standard deviation was less than 3%. Yet the response for CPE (Fig. 5b) to acetaminophen was only linear up to 40 mg L−1 (ipCPE (μA) = 0.1452 + 0.0754 C; r = 0.9928; n = 7) with a detection limit of 0.08 mg L−1. For higher concentrations (40 to 80 mg L−1), a non-linear relationship with decreasing slope was obtained. In conclusion, the GNMCPE compared with the CPE showed a relatively wide linear range for acetaminophen determination. Moreover, the GNMCPE improved the sensitivity and detection limit of acetaminophen determination.

Fig. 5
figure 5

Calibration curves for acetaminophen at a CPE (b) and a GNMCPE (a) in 0.1 mol L−1 phosphate buffer pH 4.7

Effect of concomitant substances, recovery and analysis of acetaminophen in pharmaceutical samples

In order to investigate the analytical application of this method, the specificity of GNMCPE to acetaminophen in the presence of some interfering substances such as ascorbic acid, urea and glucose was examined by carrying out the determination of 1 mg L−1 acetaminophen. Ascorbic acid and glucose represent electroactive materials, urea stands for non-electroactive matters. The results are compiled in Table 1. The interferences of ascorbic acid and glucose observed in the flow injection procedure for the determination of acetaminophen are obvious. However, the contents of these interferences in some relevant pharmaceuticals are commonly lower than the proportions of Table 1. This method can also be used for determination of acetaminophen.

Table 1 Results of interference experiments using the FIA amperometric procedure in 1 mg L−1 acetaminophen at GNMCPE

The recovery tests of acetaminophen in the range of 5 to 25 mg L−1 were performed using the FIA-GNMCPE method. The results are listed in Table S1. The recoveries lay in the range of 98.40–104.1%, and the relative standard deviation was lower than 3%. This fact showed that the precision of this method was good, which testified to the applicability of FIA-GNMCPE to real samples.

FIA-GNMCPE was applied to the amperometric determination of acetaminophen in six commercial tablets with different compositions. The results are presented in Table 2. The analysis of acetaminophen for each sample was performed in triplicate. As can be observed, the relative errors were <1%, and the relative standard deviations were <2% in the results provided by FIA-GNMCPE. The results showed good accuracy and precision obtained with FIA-GNMCPE.

Table 2 Determination of acetaminophen in pharmaceutical formulations using FIA-GNMCPE compared with HPLC (n = 3)

Conclusion

GNMCPE with an in-house amperometric detector was applied in flow injection analysis for the determination of acetaminophen. The GNMCPE shows a good electrochemical response to acetaminophen, and it exhibits a more sensitive and relatively fast response compared to CPE. The amperometric detection flow-through cell was made with inexpensive and easily obtained silicon rubber tubes of different diameters, its volume could be changed easily by adjusting the relative position of the silicon rubber tubes. The FIA-GNMCPE method developed represents excellent performance for the determination of acetaminophen in commercial tablets. The results are in agreement with HPLC. This method exhibited remarkable advantages including low cost, rapid response, simple detector configuration and high sensitivity. Apart from this, this method can be applied to the rapid determination of the filtration of analytes after ultrasonically assisted dissolution of the samples in water and anhydrous ethanol (without any previous reaction or derivatization process being necessary).