Abstract
Purpose
LigaSure is a bipolar diathermy system, which achieves vessel sealing with reduced thermal spread. The device has been used successfully in abdominal operations and because of its features, it has been applied recently in thyroid surgery to minimize the risk of complications such as laryngeal nerve palsy and hypocalcemia, and also to reduce the operating time.
Methods
Between June and December 2005, we performed total thyroidectomy for various disorders in 105 patients. We used the LigaSure diathermy system in 53 patients and traditional hemostatic procedures in the other 52. We evaluated the demographic features, histopathological diagnosis, operating times, and relevant postoperative complications.
Results
The two study groups had similar demographic and histopathological features. The mean operating time ± SD was not significantly shorter in the LigaSure group than in the traditional group (104 ± 12.7 vs 110 ± 15.6 min). None of the patients in either group suffered hemorrhage, definitive hypocalcemia, or definitive laryngeal nerve palsy. Transient hypocalcemia and transient laryngeal nerve palsy developed in 7.54% and in 1.88%, respectively, of the patients in the LigaSure group, and in 7.69% and 1.92%, respectively, of the patients in the traditional group; these differences were not significant.
Conclusions
LigaSure is a safe and effective method of hemostasis control, but it did not reduce the incidence of complications or operating times compared with traditional hemostatic procedures; moreover, the operative costs were higher.
Article PDF
Similar content being viewed by others
Avoid common mistakes on your manuscript.
References
Gluckman JL. Total thyroidectomy. In: Bailey BJ, Calhoun KH, editors. Atlas of head and neck surgery: otolaryngology, 2nd ed. Philadelphia: Lippincott Williams & Wilkins; 2001. p. 228–229.
Hambley R, Hebda PA, Abell E, Cohen BA, Jegasothy BV. Wound healing of skin incisions produced by ultrasonically vibrating knife, scalpel, electrosurgery, and carbon dioxide laser. J Dermatol Surg Oncol 1988;14:1213–1217.
Siperstein AE, Berber E, Morkoyun E. The use of the harmonic scalpel vs conventional knot tying for vessel ligation in thyroid surgery. Arch Surg 2002;137:137–142.
Voutilainen PE, Haapiainen RK, Haglund CH. Ultrasonically activated shears in thyroid surgery. Am J Surg 1998;175:491–493.
Voutilainen PE, Haglund CH. Ultrasonically activated shears in thyroidectomies: a randomized trial. Ann Surg 2000;231:322–328.
Meurisse M, Defechereux T, Maweja S, Degauque C, Vandelaer M, Hamoir E. Evaluation of the Ultracision ultrasonic dissector in thyroid surgery: prospective randomized study. Ann Chir 2000;125:468–472.
Dubuc-Lissoir J. Use of a new energy-based vessel ligation device during laparoscopic gynecologic oncologic surgery. Surg Endosc 2003;17(3):466–468.
Heniford BT, Matthews BD, Sing RF, Backus C, Pratt B, Greene FL. Initial results with an electrothermal bipolar vessel sealer. Surg Endosc 2001;15(8):799–801.
Kennedy JS, Stranahan PL, Taylor KD, Chandler JG. High-burststrength, feedback-controlled bipolar vessel sealing. Surg Endosc 1998;12:876–878.
Harold KL, Pollinger H, Matthews BD, Kercher KW, Sing RF, Heniford BT. Comparison of ultrasonic energy, bipolar thermal energy, and vascular clips for the hemostasis of small-, medium-, and large-sized arteries. Surg Endosc 2003;17(8):1228–1230.
Stranahan PL, Buysse SP, Ryan TP, Pearce JA, Thomsen S, Sieve VD, et al. Healing process and histologic evaluation following use of bipolar energy for vessel sealing. Boulder, CO: Valleylab, a Division of Tyco Healthcare Group LP; 1999.
Petrakis IE, Kogerakis NE, Lasithiotakis KG, Vrachassotakis N, Chalkiadakis GE. Ligasure versus clamp-and-tie thyroidectomy for benign nodular disease. Head Neck 2004;26:903–909.
Lachanas VA, Prokopakis EP, Mpenakis AA, Karatzanis AD, Velegrakis GA. The use of Ligasure Vessel Sealing System in thyroid surgery. Otolaryngol Head Neck Surg 2005;132:487–489.
Kirdak T, Korun N, Ozguc H. Use of ligasure in thyroidectomy procedures: results of a prospective comparative study. World J Surg 2005;29:771–774.
Shen WT, Baumbusch MA, Kebebew E, Duh QY. Use of the electrothermal vessel sealing system versus standard vessel ligation in thyroidectomy. Asian J Surg 2005;28(2):86–89.
Kiriakopoulos A, Dimitrios T, Dimitrios L. Use of diathermy system in thyroid surgery. Arch Surg 2004;139:997–1000.
Sandonato L, Cipolla C, Graceffa G, Fricano S, Li Petri S, Prinzi G, et al. Bipolar electrothermic coagulation (Ligasure bipolar vessel sealing system) in thyroid surgery. Chir Ital 2003;55:411–415.