Abstract
Purpose
Few complications have been reported for lumbar total disc replacement (TDR) and hybrid TDR fixations. This study evaluated retrieved implants and periprosthetic tissue reactions for two cases of osteolysis following disc arthroplasty with ProDisc-L prostheses.
Methods
Implants were examined for wear and surface damage, and tissues for inflammation, polyethylene wear debris (polarized light microscopy) and metal debris (energy-dispersive X-ray spectroscopy).
Results
Despite initial good surgical outcomes, osteolytic cysts were noted in both patients at vertebrae adjacent to the implants. For the hybrid TDR case, heterotopic ossification and tissue necrosis due to wear-induced inflammation were observed. In contrast, the non-hybrid implant showed signs of abrasion and impingement, and inflammation was observed in tissue regions with metal and polyethylene wear debris.
Conclusions
In both cases, wear debris and inflammation may have contributed to osteolysis. Surgeons using ProDisc prostheses should be aware of these rare complications.
Avoid common mistakes on your manuscript.
Introduction
TDR is an established alternative to lumbar fusion for the treatment of back and leg pain that is associated with degenerative disc disease (DDD). In cases of 2-level DDD, hybrid fixation is a new approach that involves combining the advantages of TDR with spinal fusion at the adjacent level. This approach preserves motion at one level and maintains stiffness in the lower segment to prevent adjacent segment degeneration. Although the majority of patients attain clinically significant pain reduction after 1-level TDR or hybrid fixation, foreign-body response to wear debris and rare osteolysis instances have been noted for other devices. Historical generations of polyethylene-core TDRs such as the CHARITÉ (originally Waldemar Link, Hamburg, Germany, later fabricated by DePuy Spine, Raynham, MA, USA and currently discontinued) have shown evidence of polyethylene wear debris in periprosthetic tissues, accompanied by histological changes, the presence of histiocytes and multinuclear giant cells [2, 3]. As a consequence of adhesive/abrasive wear mechanisms, polyethylene wear particles are released and then ingested by resident macrophages initiating a chronic immune response that can lead to osteolysis [4]. Evidence of lumbar periprosthetic osteolysis appeared in 1 of 21 implant revisions from 18 patients who received a CHARITÉ in a previously reported study, but little is known about osteolysis around ProDisc-L implants [3].
Interestingly, in the aforementioned study, osteolysis occurred in a patient that underwent hybrid fixation. This raises the question of whether the combination of TDR and fusion may create a loading and kinematic environment conducive to potential osteolysis. Another potential factor that may contribute to osteolysis in the spine is the use of bone morphogenic protein (BMP) during fusion. McKay et al. [5, 6] noted that resorption rates around the implant increase with the use of BMP-2, presumably due to BMP-induced enhancement of osteoclast activity, which results in vertebral osteolysis.
The purpose of this study was to report two unusual osteolysis cases after ProDisc-L lumbar disc replacement—one in which the patient underwent TDR at the level superior to BMP-induced interbody fusion at L5-L6 and another who had similar osteolytic lesions after 1-level TDR without any exposure to BMP-2. Both patients underwent TDR with ProDisc-L (Synthes, West Chester, PA, USA) prosthesis, which consists of an ultra-high molecular weight polyethylene (UHMWPE) core and two metallic endplates made of a cobalt-chromium (CoCr) alloy, similar to the CHARITÉ [7]. However, unlike the CHARITÉ, the UHMWPE core of ProDisc-L is locked into the inferior endplate, thus allowing relative motion only between the core and the superior endplate. To our knowledge, there have been no previous reports of osteolysis with the use of ProDisc-L.
Materials and methods
Two patients who suffered from lumbar disc herniation and radiculopathy underwent surgery. One patient required multi-level treatment and opted for hybrid fixation with ProDisc-L TDR and fusion, while the other received 1-level ProDisc-L TDR. Both TDRs were extracted during revision surgery and periprosthetic tissue specimens selected from regions adjacent to the implant were obtained. Retrievals, operative notes and radiographs were de-identified and collected in accordance with an IRB-approved protocol.
Implant retrieval analysis
The two sets of retrieved components were cleaned in 10 % bleach and examined under a stereomicroscope equipped with a digital camera (Leica DFC490) to assess for surface damage and gross fracture. All components were inspected to identify surface damage mechanisms (plastic deformation, scratching, burnishing, pitting, and embedded debris). Damaged regions of the implants were analyzed using scanning electron microscopy (SEM; Supra 50 VP, Zeiss Peabody, MA, USA), energy-dispersive X-ray spectroscopy (EDS) and x-ray fluorescence (XRF).
Tissue Preparation and Histological Analysis
Tissues collected from revision surgeries were fixed in Universal Tissue Fixative (Sakura Finetek USA, Inc., Torrance, CA, USA), and decalcified based on the presence of heterotopic ossification determined by microCT (µCT 80, Scanco Medical, Brüttisellen, Switzerland). One to two 4-mm punches from each tissue, considering variations in color, texture, and size of specimen, were embedded in paraffin blocks for 6-µm serial sectioning and staining with Alcian blue (Electron Microscopy Sciences, Hatfield, PA, USA), hematoxylin, and eosin (H&E) (ThermoFisher Scientific, Waltham, MA, USA). Entire tissue sections were imaged under transmitted light microscopy using a Motic BA300POL microscope (Motic, Richmond, British Columbia, Canada), equipped with an elliptically polarized light imaging system and ProgRes SpeedXT core 5 (Jenoptik, Jena, Germany) microscope camera. Inflammatory cells were confirmed using the Wright-Giemsa stain (Electron Microscopy Sciences, Hatfield, PA, USA). Tissues with notable chronic inflammation were examined using environmental scanning electron microscopy (ESEM; XL-30 ESEM-FEG, FEI Company, Hillsboro, OR, USA) with backscatter and were analyzed with EDS.
Case reports
Case 1
A 40-year-old male suffering from discogenic collapse with lower back pain and radiculopathy at L5-S1 underwent anterior discectomy at L5-S1 and interbody arthrodesis with 17 × 24 mm titanium-threaded fusion cages filled with BMP-2. A few months later, a posteriorly displaced cage, along with osteophyte formation and foraminal stenosis, required the patient to undergo revision of the cage. Segmental pedicle screws were used to stabilize the cage on the right side; 6.5 mm Xia (Stryker Spine, Allendale, NJ, USA) pedicle screws were inserted at L5 and S1. Once the cages were locked, posterior interbody arthrodesis was then implemented at L5-S1 with the use of BMP-2 in two small sponges placed between each cage. One year later, with continued back pain and disc herniation at L4–L5, the patient opted for hybrid fixation and underwent anterior interbody placement of 12 mm-large and 6-degree ProDisc-L at L4–L5. The following year, the pedicle screw instrumentation at L5-S1 was removed and there was solid fusion at the level. The patient also required mass resection of osteophytes, other bony spurs and scar tissue.
Three years after fusion, persistent patient pain and severe arthropathy and degeneration was noted at right L4–L5. Posterior fusion was undertaken; 6.5 mm screws were inserted and a PEEK rod was implemented at L4–L5, followed by posterolateral arthrodesis using local bone autograft, BMP-2, and DBX Demineralized Bone Matrix Allograft (Musculoskeletal Transplant Foundation, Edison, NJ, USA). However, computed tomographic (CT) scans in the following year revealed osteolytic cysts at L4–L5 (Fig. 1). The PEEK rod stabilization system and artificial disc were removed and sent for retrieval analysis. Preoperative work up and intraoperative cultures ruled out infection. Tissue samples adjacent to the disc were also removed for histological analysis. An overview of clinical information is provided in Table 1.
Device retrieval analysis
The ProDisc-L prosthesis was retrieved 5 years after implantation. Due to iatrogenic damage, we could not determine if the core’s rims experienced impingement with the superior endplate (see Online Resource Fig. 1). While the backside surface also experienced iatrogenic damage, there was evidence of burnishing and scratching that occurred in vivo. There were no obvious signs of impingement on the metallic endplates, and the abrasive scratches were not patterned in any physiological manner, suggesting they occurred during device removal. Analysis using SEM and EDS revealed no abnormal surface deposits on the metallic endplates. As expected, XRF scans consistently detected cobalt-chromium ratios matching ASTM F-75 cobalt alloy weight-standards in the interior of the endplates, and the exterior plasma-coated elements consisted of alloy compositions seen in commercially pure titanium.
Tissue analysis
Periprosthetic tissues from this patient were obtained from two unclassified regions around the implant. One region was identified as predominantly fibrocartilage and the other region was mature trabecular bone. One isolated region in the fibrocartilage tissue contained hemosiderin deposits and macrophages, consistent with an innate response to hemorrhage prior to explantation. Large areas of the surrounding tissue were also necrotic due to inflammation. There was no evidence of metal wear debris in any tissues and only minor polyethylene debris was detected in isolated regions of the trabecular bone at a mean density of 1.13 particles/mm2. In bony tissue samples, the fatty marrow in the intertrabecular spaces contained only a small number of viable cells; isolated regions of these samples consisted of necrotic bone marrow along with necrotic bone with empty osteocyte lacunae (see Online Resource Fig. 2). An overview of tissue morphology is provided in Table 2.
Case 2
A 56-year-old male with a herniated disc and radiculopathy at L4–L5 underwent anterior TDR with a 10 mm-large and 6-degree ProDisc-L. Three years after, subsidence of disc was noted at L4 on the right side, along with the formation of osteolytic cysts in CT scans at L5 that appear similar to the lesions seen in case 1 (Fig. 2). Progressive back pain led to removal of ProDisc-L, followed by placement of 22 mm PEEK interbody graft filled with BMP-2 for interbody fusion at L4–L5. Preoperative workup and intraoperative cultures were negative for infection.
Device retrieval analysis
The ProDisc-L prosthesis was retrieved 3 years after implantation. There was clear evidence of chronic impingement between the endplates and burnishing at the core’s edge; microscopic scratches of fan-shaped pattern were found on the interior of the metallic plates and a glossy appearance on the polyethylene core, respectively (Fig. 3). SEM images of the impinged regions showed a polished appearance in comparison to the as-manufactured texture seen in non-impinged regions of the metallic plate. The unidirectional and circumferential wear patterns seen on the endplates suggest the wear may have occurred during axial rotation and/or lateral bending of the articulating surfaces. The impingement was most likely due to implant subsidence which was observed by the surgeon during surgery. The dome of the core also had evidence of multi-directional scratches and burnishing. There were no indications of fatigue wear or fracture of the polyethylene core. No abnormal surface deposits were observed by SEM/EDS analysis. XRF scans showed the metallic surface-constituents on the interior of the endplates consistently matched CoCr ratios seen in ASTM F-75 cobalt alloy standards, and the exterior of plates consisted of weight compositions seen in commercially pure titanium.
Tissue analysis
The periprosthetic tissues in this patient showed several abnormalities such as progressive degeneration, varied inflammation levels, and metal and polyethylene wear debris (Table 2). While degeneration was observed in tissues from various regions, inflammation was predominantly in the intradiscal and cyst tissue; there were no signs of inflammation in tissues from the lateral annulus, left lateral, lateral spur, superior end plate and anterior intradiscal regions. Cyst tissue from L5 regions showed signs of both innate and adaptive immune response; macrophage ingested metal-wear-debris was present throughout the tissue and isolated areas of lymphocytes were also present (Fig. 4). Cyst and intradiscal tissues also contained hemosiderin deposits (not shown), suggesting a prior hemorrhage that may have contributed to or exacerbated the chronic inflammation. To confirm metallic wear debris, tissues with notable inflammation were examined by ESEM using backscatter and were analyzed with EDS. The particles from cyst and intradiscal tissues were predominantly cobalt and chromium, however, titanium was also detected (Fig. 5). Polyethylene wear debris was present in relatively low numbers in all cyst, posterior-intradiscal, anterior-intradiscal and posterior-lateral tissues. The mean polyethylene particles in these tissues was 2.08 particles/mm2. These particles varied from oval to amorphous in shape and were localized to regions of chronic inflammation (Fig. 6).
Discussion
This study reported two unusual cases of osteolysis in TDR patients with a ProDisc-L. The first patient suffered from multi-level DDD and opted for hybrid fixation, while the second had a 1-level TDR. Both patients developed osteolytic lesions in vertebrae adjacent to the prostheses. Infection was ruled out in both cases. Retrieval and histological analysis of the hybrid case showed minor amounts of wear, however, tissue responses included fibrocartilage generation, heterotopic ossification, and necrosis due to inflammation. The second case showed signs of endplate impingement and adverse local tissue reactions (ALTRs) in intradiscal and cystic tissues. In this second, impingement case, the amount of polyethylene wear debris was relatively low, but there was CoCr wear debris and associated inflammation. In both cases, inflammatory tissue responses may have contributed to the osteolytic lesions.
One difference between the hybrid and non-hybrid case was the use of BMP-2. The hybrid patient was exposed to BMP-2 on five occasions, in comparison to the patient that had a 1-level TDR who received none. Although osteolysis was a late observation after BMP-2 application, the hybrid patient only sought medical attention due to pain. A number of tissue responses to BMP-2 have been noted after spine surgery including heterotopic ossification observed in the hybrid patient’s retrieved tissue [8–10]. Although previously believed to be asymptomatic, heterotopic ossification can lead to delayed neural compression and pain [10]. Furthermore, recent studies reported increased resorption rates with the use of BMP-2 around implants, raising the question whether BMP-2 contributes to osteolysis in regions adjacent to the fused segment [5, 6]. Authors have reported asymptomatic osteolysis after interbody fusion and attributed bone loss to endplate violation during disc space preparation and/or to overdosing of BMP-2 [11–13]. Whether BMP-2 has a dose-responsive effect on the activation of osteolytic pathways remains unclear, since an optimal BMP-2 dose for fusions is still not agreed upon.
Osteolysis, along with tissue reactions involving mixed immune responses have been previously reported in a CHARITÉ by Kurtz et al. [3], but these devices consisted of polyethylene cores that were gamma-air-sterilized. ProDisc devices utilize conventional gamma-inert-sterilized polyethylene that has been previously reported in total joint replacements to significantly lower oxidation, wear debris generation and inflammation [16, 17]. ProDisc TDRs have been approved by the Food and Drug Administration, and only a limited number of complications have been documented with the use of either ProDisc-L or ProDisc-C. While hybrid fixation with ProDisc-L and fusion remains under clinical evaluation, there have been no previous reports of osteolysis with the use of ProDisc-L. However, with use of ProDisc-C for cervical TDR, one exceptional case of progressive osteolysis was reported and attributed to a possible immune-mediated metal sensitivity reaction [18].
In contrast to the minimal immune reaction in the hybrid case, there were substantial ALTRs in the intradiscal and cyst tissue of the non-hybrid case. Wear debris-induced inflammation is known to mediate osteolysis; thus, impingement and subsequent pro-inflammatory processes may explain the clinical symptoms and radiographic progression seen in the non-hybrid case [19]. All samples of cyst tissue showed signs of chronic inflammation and lymphocytic infiltration which were similar to ALTRs to metal ions from implant corrosion of metal-on-metal total hip replacements (THRs) [20]. These findings suggest that ALTRs from THRs share some characteristics with the cyst tissue from the non-hybrid case.
The present study reported two rare osteolysis cases following implantation of the Prodisc-L. In one case, wear-debris induced inflammation; in the second case, inflammation induced heterotopic ossification. As surgeons incorporate ProDisc technology into their clinical practice, the rare complication of osteolysis and its occurrence should be taken into account when defining contraindications for spinal arthroplasty.
References
Aunoble S, Meyrat R, Al Sawad Y, Tournier C, Leijssen P, Le Huec JC (2010) Hybrid construct for two levels disc disease in lumbar spine. Eur Spine J 19:290–296. doi:10.1007/s00586-009-1182-7
van Ooij A, Kurtz SM, Stessels F, Noten H, van Rhijn L (2007) Polyethylene wear debris and long-term clinical failure of the charite disc prosthesis: a study of 4 patients. Spine(Phila Pa 1976) 32:223–229. doi:10.1097/01.brs.0000251370.56327.c6
Kurtz SM, van Ooij A, Ross R, de Malefijt Waal J, Peloza J, Ciccarelli L, Villarraga ML (2007) Polyethylene wear and rim fracture in total disc arthroplasty. Spine J 7:12–21. doi:10.1016/j.spinee.2006.05.012
Campbell P, Ma S, Yeom B, McKellop H, Schmalzried TP, Amstutz HC (1995) Isolation of predominantly submicron-sized UHMWPE wear particles from periprosthetic tissues. J Biomed Mater Res 29:127–131. doi:10.1002/jbm.820290118
McKay B, Sandhu HS (2002) Use of recombinant human bone morphogenetic protein-2 in spinal fusion applications. Spine 27:S66–S85. doi:10.1097/01.Brs.0000020738.16636.B3
Choudhry OJ, Christiano LD, Singh R, Golden BM, Liu JK (2012) Bone morphogenetic protein-induced inflammatory cyst formation after lumbar fusion causing nerve root compression case report. J Neurosurg Spine 16:296–301. doi:10.3171/2011.11.Spine11629
Kurtz SM, Villarraga ML, Ianuzzi A (2009) The clinical performance of UHMWPE in the spine. In: UHMWPE Biomaterials Handbook. Academic Press, Burlington, pp 180
Anderson CL, Whitaker MC (2012) Heterotopic ossification associated with recombinant human bone morphogenetic protein-2 (infuse) in posterolateral lumbar spine fusion. Spine 37:E502–E506
Mannion RJ, Nowitzke AM, Wood MJ (2011) Promoting fusion in minimally invasive lumbar interbody stabilization with low-dose bone morphogenic protein-2–but what is the cost? Spine J off J North Am Spine Soc 11:527–533
Chrastil J, Patel AA (2012) Complications associated with posterior and transforaminal lumbar interbody fusion. J Am Acad Orthop Surg 20:283–291
Lewandrowski KU, Nanson C, Calderon R (2007) Vertebral osteolysis after posterior interbody lumbar fusion with recombinant human bone morphogenetic protein 2: a report of five cases. Spine J off J North Am Spine Soc 7:609–614. doi:10.1016/j.spinee.2007.01.011
McClellan JW, Mulconrey DS, Forbes RJ, Fullmer N (2006) Vertebral bone resorption after transforaminal lumbar interbody fusion with bone morphogenetic protein (rhBMP-2). J Spinal Disord Tech 19:483–486. doi:10.1097/01.bsd.0000211231.83716.4b
Vaidya R, Sethi A, Bartol S, Jacobson M, Coe C, Craig JG (2008) Complications in the use of rhBMP-2 in PEEK cages for interbody spinal fusions. J Spinal Disord Tech 21:557–562. doi:10.1097/Bsd.0b013e31815ea897
Glassman SD, Howard JM, Sweet A, Carreon LY (2010) Complications and concerns with osteobiologics for spine fusion in clinical practice. Spine 35:1621–1628. doi:10.1097/Brs.0b013e3181ce11cc
Benglis D, Wang MY, Levi AD (2008) A comprehensive review of the safety profile of bone morphogenetic protein in spine surgery. Neurosurgery 62:423–431. doi:10.1227/01.neu.0000326030.24220.d8
Kurtz SM, MacDonald D, Ianuzzi A, van Ooij A, Isaza J, Ross ER, Regan J (2009) The natural history of polyethylene oxidation in total disc replacement. Spine 34:2369–2377. doi:10.1097/Brs.0b013e3181b20230
Choma TJ, Miranda J, Siskey R, Baxter R, Steinbeck MJ, Kurtz SM (2009) Retrieval Analysis of a ProDisc-L Total Disc Replacement. J Spinal Disord Tech 22:290–296. doi:10.1097/Bsd.0b013e31816dd2b6
Tumialan LM, Gluf WM (2011) Progressive vertebral body osteolysis after cervical disc arthroplasty. Spine 36:E973–E978. doi:10.1097/Brs.0b013e3181fd863b
Purdue PE, Koulouvaris P, Nestor BJ, Sculco TP (2006) The central role of wear debris in periprosthetic osteolysis. HSS J Musculoskelet J Hosp Spec Surg 2:102–113. doi:10.1007/s11420-006-9003-6
Campbell P, Ebramzadeh E, Nelson S, Takamura K, De Smet K, Amstutz HC (2010) Histological features of pseudotumor-like tissues from metal-on-metal hips. Clin Orthop Relat Res 468:2321–2327. doi:10.1007/s11999-010-1372-y
Acknowledgments
We thank Eual Phillips for his help in performing the image analysis for this study.
Conflict of interest
This study was supported by a grant from the NIAMS (NIH R01 AR56264). Institutional support was received from Medtronic and DePuy Synthes for research unrelated to this work (SMK). For THL, he is a consultant and receives royalties from Medtronic.
Author information
Authors and Affiliations
Corresponding author
Electronic supplementary material
Below is the link to the electronic supplementary material.
Rights and permissions
About this article
Cite this article
Veruva, S.Y., Lanman, T.H., Hanzlik, J.A. et al. Rare complications of osteolysis and periprosthetic tissue reactions after hybrid and non-hybrid total disc replacement. Eur Spine J 24 (Suppl 4), 494–501 (2015). https://doi.org/10.1007/s00586-014-3535-0
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00586-014-3535-0