Abstract
Background. Successful radiofrequency-induced ablation is dependent on correct placement of the device. The ultraguide system is a three-dimensional positioning system creating an overlay of the real-time ultrasound image and a virtual image of the device.
Methods. Tumor mimics were produced by injecting an agarose/cellulose/glycerol gel into pig livers on table. The precision of device placement was evaluated either using a freehand ultrasound procedure or using the aid of the ultraguide system either by an experienced or inexperienced surgeon. Tumor mimics were ablated by a protocol yielding an ablation only discretely larger than the mimics to enhance the importance of precise positioning.
Results. The sizes of the 40 tumor mimics were: largest diameter 14.1±2.2 mm, volume 0.89±0.40 cm3. The largest diameter of ablation was 25.6±3.7 mm, the smallest diameter 21.9±2.9 mm, and the volume 7.20±2.38 cm3. The experienced surgeon was successful in 7 of 10 cases with and without the ultraguide, the inexperienced surgeon in 4 of 10 without and 7 of 10 with the ultraguide.
Conclusions. The ultraguide system may facilitate precise device placement for the less experienced surgeon. It seems worthwhile to evaluate a possible benefit of the system during placement of devices under operating room conditions.
Article PDF
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Author information
Authors and Affiliations
Additional information
Electronic Publication
Rights and permissions
About this article
Cite this article
Stippel, D.L., Böhm, S., Beckurts, T.K. et al. Experimental evaluation of accuracy of radiofrequency ablation using conventional ultrasound or a third-dimension navigation tool. Langenbecks Arch Surg 387, 303–308 (2002). https://doi.org/10.1007/s00423-002-0315-9
Received:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/s00423-002-0315-9