Abstract
An 84-year-old woman was unsuccessfully resuscitated for 3 min using standard cardiopulmonary resuscitation (CPR), followed by 15 min of active compression-decompression (ACD). The autopsy revealed that death was due to myocardial infarction complicated by rupture of the infarcted area and pericardial tamponade was diagnosed. Furthermore, a series of rib fractures, a transverse fracture of the sternum, rupture of the pericardial sac, the right ventricle, both atria and lacerations of the ascending aorta, were found with no signs of a vital reaction. To our knowledge, such extensive cardiac injury after CPR has not been previously reported. It is suggested that the pre-existing pericardial tamponade, the age of the patient and the application of the ACD-device to incorrect areas of the chest contributed to the extent of the cardiac injury. This case further adds to the suspicion of an increased risk of cardiac injuries when using an ACD device for cardiac massage.
Article PDF
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Author information
Authors and Affiliations
Additional information
Received: 16 April 1997 / Received in revised form: 10 July 1997
Rights and permissions
About this article
Cite this article
Klintschar, M., Darok, M. & Radner, H. Massive injury to the heart after attempted active compression-decompression cardiopulmonary resuscitation. Int J Leg Med 111, 93–96 (1998). https://doi.org/10.1007/s004140050123
Issue Date:
DOI: https://doi.org/10.1007/s004140050123