Abstract
The objectives of the study were to report a vibrant soundbridge (VSB) implant revision surgical method involving adhesiolysis at the short incus process under local anesthesia and demonstrate successful hearing performance after surgery. Three cases of VSB surgery, performed in 2016, were enrolled. All cases had diagnoses of device failure. This ‘seven-incision line’ exposed the floating mass transducer directly, after which the three steps (adhesiolysis, curettage, and hydrocortisone injection) were performed. Upon fitting the VSB, sound fields were evaluated immediately and at 3 months after the revision. During the revisions of surgery, all patients achieved immediate hearing gains and noticed differences in the outer devices with different amplifications. Satisfactory improvements in hearing thresholds and speech recognition abilities were confirmed by improvements of 20–30 dB in hearing loss 3 months after revision surgery. The VSB implant revision surgical method involving adhesiolysis is safe and efficient for patients who experience a VSB device failure. This method will reduce the requirement for surgery under general anesthesia, reduce the overall period of clinical therapy and, therefore, minimize patients’ medical costs.
Level of Evidence 4
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
The vibrant soundbridge (VSB; Vibrant MED-EL, Innsbruck, Austria), a semi-implantable middle ear device created in 1997, provides a treatment option for patients with mild-to-severe sensorineural hearing loss who are unable to wear or are dissatisfied with conventional hearing aids [1]. Multiple published reports describe the applications of VSB with a focus on both the surgical and audiological outcomes [2, 3]. In addition to the benefits, however, these reports presented long-term data on device complications related to medical or technical problems, including the need for diagnostic magnetic resonance imaging and device failure. Sterkers et al. [4] reported that during a short-term follow-up of 17 months, 5 device failures (4%) occurred. In addition, Mosnier et al. [2] reported results collected from 77 patients with a VSB during an average follow-up of 6 years. Five patients (6%) required revision surgery, mainly because of problems with coupling of the floating mass transducer (FMT) onto the incus, magnet issues, and fibrous tissue in the middle ear. Rameh et al. [5] reported a revision rate of 9% because of implant dysfunction. As the frequency of VSB surgery continues to increase, the incidence of VSB revision for several reasons is also expected to increase.
Recently, attachment of the FMT to the short incus process, a procedure comparable to stimulation when the FMT is attached to the long incus process [6], was introduced. A comparison of FMT attachment to the short and long incus processes revealed similar stapes footplate and round window membrane velocity responses with both procedures. However, no reports have described complications, especially VSB revision, associated with the new techniques used to fix the FMT to the short incus process.
The current study aimed to introduce adhesiolysis around the FMT during revision surgery under local anesthesia for a non-functioning VSB at the short incus process and to demonstrate the successful immediate hearing gains achieved immediately after the revision. This is the first report and guideline for this VSB revision method at the short incus process.
Methods
Subjects
Three patients who underwent initial VSB surgery at the short incus process in 2016 were subjected to revision surgery following complaints of no improvements in hearing with the VSB. The patients ranged in age from 50 to 65 years old. All had ‘no hearing gain with the device’ after several rounds of VSB fitting over a 3-month period [7]. Pure tone audiometry (PTA), sound field of the device, and transocular X-ray view were evaluated before the revision operation. The PTA and sound field results with VSB are shown in Fig. 1. The patients achieved no subjective hearing gains and no gains on the sound field hearing test. Transocular X-ray revealed that the FMTs were well positioned in the middle ear cavity and that their locations had not changed compared to the postoperative images after initial VSB surgery. Upon receiving written consent from the patients, we decided to perform VSB revision surgeries under local anesthesia to address hearing performance immediately after the adhesiolysis around the VSB.
Surgical technique
During VSB revision, the surgical field was sufficient to expose a 2 cm-wide postauricular groove from the skin to the hairline, without any further haircut. Xylocaine (1% with 1:100,000 epinephrine) was injected subcutaneously into the postauricular sulcus between the mastoid fascia and ear cartilage to avoid injury to the VSB electrode. After identifying the previous periosteal flap (Palva flap), flap incision was initiated with a blade at the mastoid tip and continued along the external auditory canal (Fig. 2a; Suppl movie 1). Coagulation was performed only with a bipolar electrosurgical instrument. The incision finally terminated at the temporal line to expose the antrum. This ‘seven-incision line’ was the best way to expose the FMT while avoiding electrode injury. ‘Adhesiolysis’ is the term for the surgical lysis of (intra-abdominal) adhesions, usually by laparoscopy. We used the adhesiolysis in removing the adhesion around the FMT. Adhesions to the electrode were removed via dissection with cold, sharp scissors along with bipolar backup to ensure the least risk of reformation [7]. Three steps were performed for adhesions around the FMT in the antrum. First, microsurgical instruments (picks and microforceps) were used for adhesiolysis from the lateral sides and inferior side, and finally to the upper side (Fig. 2b). Adhesiolysis on the upper side of FMT requires caution because the area where the adhesion interfered with the vibration of the FMT was located between the fossa incudis and upper side of the FMT. Second, minimal curettages (Fig. 2c) were needed around the FMT, especially on the fossa incudis (upper side of FMT), to ensure that the FMT did not touch bone. During this step, the achievement of subjective hearing gains should be monitored by attaching different external audio processors with individually adjusted and full gains and comparing the status relative to performance without the external device. Under the surgical drape, the audiologist changed the audio processors and checked hearing gains using several simple tests (ling 6, simple word lists, and simple sentences). Finally, after confirming that patients achieved hearing gains, hydrocortisone was spread around the FMT to prevent adhesion reformation. The skin flap was then closed sequentially.
Audiologic evaluation
PTA was performed according to the standard procedures with standard equipment. Air conduction thresholds were obtained at 0.25, 0.5, 1, 2, 4, and 8 kHz, and bone conduction thresholds were obtained at 0.5, 1, 2, and 4 kHz. Functional gains, or the differences between unaided and aided sound field thresholds, were determined using warble tones from 250 to 8 kHz at a modulation frequency of 5%. The sound field setup was calibrated according to Morgan et al. [8]. Functional gains were calculated at the implant ear, with the contralateral ear occluded.
When fitting the Amadé audio processor of the VSB, each patient’s threshold and dynamic range (uncomfortable level, UCL) were defined using the Vibrogram. DSL I/O was applied as the fitting formula in the fitting method. The acclimatization level setting reflected the audiologist’s experience with amplification from conventional hearing systems. The input/output (gain) curve display could be drawn using Connexx software (Piscataway, NJ, USA).
Results
All three subjects underwent immediate check-ups performed by an audiologist after the adhesiolysis, but before skin closure over the surgical field. A Vibrogram test was performed for all subjects, and the results indicated good coupling between the FMT and stapes. External audio processors with individually adjusted and full gains were attached in regular sequence. All patients achieved unexpected, immediate hearing gains and sensed differences when the outer devices were used with different amplifications. Two patients were more satisfied with the full gains than the adjusted gains used before the revision surgery, whereas one patient complained the hauling of self-voice despite a greater hearing gain. No complications, including hearing loss, peripheral facial paralysis, tinnitus, and vertigo, were observed immediately after the revision. During a 3-month follow-up, no events such as FMT dislocation, coil extrusion, facial palsy, headache, or skin dehiscence were observed.
Activation was performed 1 week post-operation because of the untouched FMT-incus joint. During programming, the Vibrogram test result was inputted as a fundamental parameter to create a fitting map. The patients’ fitting maps are presented in Fig. 3 as the level of output (Fig. 3a–c) and gains in specific frequency (Fig. 3d–f). After activating the processor, the hearing levels (0.5, 1, 2, 3, 4 and 8 kHz) were tested in the sound field; the speech discrimination test (MCL most comfortable loudness level and PB max maximum score of phonetically balanced with monosyllabic words HL in MCL) was tested 3 months after adhesiolysis. Test results confirmed the obvious improvements in both the hearing threshold (Fig. 4) and speech recognition ability (Table 1). The auditory results from the long-term follow-up will be reported later.
Discussion
Since its introduction in 1996, the standard surgical approach for VSB positioning involves a subtotal mastoidectomy combined with a posterior tympanotomy at the long incus process.
This surgical approach resulted in good audiological outcomes and stability during long-term use [1, 2, 9,10,11,12,13]. However, subtotal mastoidectomy and posterior tympanotomy could potentially cause facial nerve or chorda tympani injury [6]. Although no reports have discussed facial nerve injury during VSB implantation, the reported incidence of temporary or permanent facial nerve palsy during cochlear implantation with this surgical approach ranges from 0.3 to 2.2% [14]. The dimensions of the FMT necessitate a large opening in the facial recess and exposure of the chorda tympani [15]. Therefore, fixation of the FMT at the short incus process is another surgical issue. Compared with attachment to the long incus process, coupling of the FMT to the short incus process with wide antrostomy yielded similar stapes footplate and RW membrane velocity responses [6]. Polanski et al. [16] also reported that surgical coupling of the FMT to the short process of the incus yielded good clinical and audiological outcomes, especially when the long process and the oval or round window were inaccessible. Since 2015, we have performed 18 cases of VSB at the short incus process. Only four of these cases did not show hearing gain with the device. Because one case involved a mixed hearing loss with conductive postoperative loss, we repositioned the FMT on the round window. In the other three cases, we noticed adhesions around the FMTs, particularly between the bones and the FMTs. In the surgical field, we were able to check patients’ immediate responses by attaching an external audio processor after adhesiolysis of the revision. We were convinced that the adhesion around the FMT might be a major cause of poor VSB performance.
The adhesions in our cases were mostly observed between the surrounding bones and FMT. Although the FMT is composed of titanium, the biomaterial could trigger a foreign body reaction, leading to adverse reactions such as inflammation and fibrosis. Shortly after implantation, biomaterials are covered with a layer of plasma proteins, especially albumin, fibrinogen, IgG, fibronectin, and von Willebrand factor [17, 18]. These proteins, which are possibly adsorbed via hydrophobic interactions, tend to assume an altered conformation and expose hydrophobic domains that become tightly adherent to hydrophobic biomaterial surfaces [19]. Conformational changes in these adsorbed proteins are thought to be responsible for initiating the above-mentioned adverse reactions [20]. The two points between the fossa incuidis and the FMT and between the dula plate and the FMT were problematic. During the initial surgeries, the FMTs were positioned to avoid contact with bones; however, we were not able to achieve a sufficient space between the FMT and bone. When adhesions occurred between the bone and FMT, the presence of a narrow gap could increase the severity of the adhesion and thus limit movement of the FMT. All patients achieved hearing gains immediately after adhesiolysis, as demonstrated during surgery, and these improvements of 20–30 dB HL were also observed 3 months later. In the speech discrimination test, the patients achieved high PB max scores (>50%), even under the condition of a lower MCL. Polanski et al. [16] reported a pure tone average of 31 dB during an audiological evaluation with a VSB implant. Another study showed that speech discrimination with VSB improved significantly as compared with unaided conditions, and increased from 63 to 80% in a quiet environment [21].
In our cases of VSB revision with adhesiolysis, the ‘seven-incision line’ was the most important procedure with regard to avoiding electrode injury. The injections in the local anesthesia would be enough similar in the tympanoplasty via retroauricular approach. The incision along the EAC wall to the temporalis line was extended as far as possible to visualize the adhesions directly around the FMT. In an earlier case of VSB revision, we had attempted to initiate dissection from the body of the VSB and follow the electrode to the FMT. However, this dissection procedure was too lengthy. The patient could not tolerate the long revision time under local anesthesia. After modifying the procedure to include the ‘seven-incision line’, the surgical time was reduced to within 30 min. None of the patients complained during VSB revision under local anesthesia. This three-step surgical procedure is necessary for adhesiolysis during a VSB revision. Adhesiolysis, minimal bone curettage, and hydrocortisone application must be mastered to ensure a consistent recovery of FMT movement. After these three steps, we got good results with more than 20 dB gains of hearing. Bone curettage was performed to prevent fibrous layer formation between the bone and FMT. The ability of cortisol to prevent adhesion reformation following microsurgical adhesiolysis is well known. Hydrocortisone may also be used intraperitoneally at the conclusion of pelvic microsurgery [22].
In this report, we included three patients from among a small group of 18 cases. We state that the VSB failure rate of 16% is acceptable, and that these cases could be addressed with adhesiolysis. However, the overall VSB hardware failure rate was high (28%), which could be explained by technical failures of early generation implants [23]. Despite the currently low incidence of VSB revision surgery, we must consider the reasons for device failure and perform revision surgeries without hesitation according to this adhesiolysis guideline. If the impedance test of the VSB was good which meant the device was ok and the audiometry after the surgery did not show the air–bone gap which meant ossicles were intact, we should question the adhesion of VSB and decide the revision.
In summary, hearing restoration was achieved following revision surgery involving adhesiolysis around the FMT under local anesthesia in patients who experienced device failure. The seven-incision line and three adhesiolysis steps (adhesiolysis, curettage, and hydrocortisone injection) will facilitate the performance of adhesiolysis revision surgery during a short surgical period under local anesthesia. Moreover, these modifications will reduce the incidence of operation under general anesthesia, reduce the overall period of clinical therapy, and thus minimize patients’ medical costs.
References
Fisch U, Cremers WRJ, Lenarz T et al (2001) Clinical experience with the vibrant soundbridge implant device. Otol Neurotol 22:962–972
Mosnier I, Sterkers O, Bouccara D et al (2008) Benefit of the vibrant soundbridge device in patients implanted for 5 to 8 years. Ear Hear 29:281–284
Maier H, Hinze AL, Gerdes T et al (2015) Long-term results of incus vibroplasty in patients with moderate-to-severe sensorineural hearing loss. Audiol Neurootol 20:136–146
Sterkers O, Boucarra D, Labassi S et al (2003) A middle ear implant, the symphonix vibrant soundbridge: retrospective study of the first 125 patients implanted in France. Otol Neurotol 24:427–436
Rameh C, Meller R, Lavieille JP et al (2010) Long-term patient satisfaction with different middle ear hearing implants in sensorineural hearing loss. Otol Neurotol 31:883–892
Schraven SP, Dalhoff E, Wildenstein D et al (2014) Alternative fixation of an active middle ear implant at the short incus process. Audiol Neurootol 19:1–11
Reich H (1995) Laparoscopic surgery for adhesiolysis. In: Arregui ME, Fitzgibbons RJ Jr, Katkhouda N et al (eds) Principles of laparoscopic surgery. Springer, New York, pp 283–298
Morgan DE, Dirks DD, Bower DR (1979) Suggested threshold sound pressure levels for frequency modulated (warble) tones in the sound field. J Speech Hear Disord 44:37–54
Lenarz T, Weber BP, Issing PR et al (2001) Vibrant soundbridge system: ein neuartiges hörimplantat für innenohrschwerhörige—teil 2: audiologische ergebnisse. Laryngo Rhino Otol 80:370–380
Todt I, Seidl RO, Gross M et al (2002) Comparison of different vibrant soundbridge audioprocessors with conventional hearing aids. Otol Neurotol 23:669–673
Beltrame AM, Martini A, Prosser S et al (2009) Coupling the vibrant soundbridge to cochlea round window. Otol Neurol 30:194–201
Böheim K, Mlynski R, Lenarz T et al (2012) Round window vibroplasty: long-term results. Acta Otolaryngol 132:1042–1048
Huber AM, Mlynski R, Müller J et al (2012) A new vibroplasty coupling technique as a treatment for conductive and mixed hearing losses. Otol Neurotol 33:613–617
Brito R, Monteiro TA, Leal AF et al (2012) Complicações em 550 cirurgias consecutivas de implante coclear. Braz J Otorhinolaryngol 78:80–85
Truy E, Eshraghi AA, Balkany TJ et al (2006) Vibrant soundbridge surgery. Otol Neurotol 27:887–895
Polanski JF, Soares AD, Dos Santos ZM et al (2016) Active middle-ear implant fixation in an unusual place: clinical and audiological outcomes. J Laryngol Otol 130:404–407
Pankowsky DA, Ziats NP, Topham NS et al (1990) Morphologic characteristics of adsorbed human plasma proteins on vascular grafts and biomaterials. J Vasc Surg 11:599–606
Bonfield TL, Colton E, Anderson JM (1992) Protein adsorption of biomedical polymers influences activated monocytes to produce fibroblast stimulating factors. J Biomed Mater Res 26:457–465
Tang L, Thevenot P, Hu W (2008) Surface chemistry influences implant biocompatibility. Curr Top Med Chem 8:270–280
Collier TO, Anderson JM (2002) Protein and surface effects on monocyte and macrophage adhesion, maturation, and survival. J Biomed Mater Res 60:487–496
Simpson A (2009) Frequency-lowering devices for managing high-frequency hearing loss: a review. Trends Amplif 13:87–106
Li TC, Cooke ID (1995) The value of an absorbable adhesion barrier, Interceed®, in the prevention of adhesion reformation following microsurgical adhesiolysis. Obstet Gynecol Surv 50:190–192
Zwartenkot JW, Mulder JJ, Snik AF et al (2016) Active middle ear implantation: long-term medical and technical follow-up, implant survival, and complications. Otol Neurotol 37:513–519
Acknowledgements
This research was supported by the Ministry of Trade, Industry & Energy (MOTIE, Korea) under Industrial Strategic Technology Development Program (10051518) and Program (10070232).
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
All authors declare that they have no conflict of interest.
Research involving human participants and/or animals
Yes (participants).
Informed consent
Yes.
Electronic supplementary material
Below is the link to the electronic supplementary material.
Supplementary material 1 (MP4 91420 kb)
Rights and permissions
About this article
Cite this article
Park, Y.A., Kong, T.H., Chang, J.S. et al. Importance of adhesiolysis in revision surgery for vibrant soundbridge device failures at the short incus process. Eur Arch Otorhinolaryngol 274, 3867–3873 (2017). https://doi.org/10.1007/s00405-017-4715-4
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00405-017-4715-4