Abstract
We report a male with a coronary bifurcation lesion in the mid circumflex artery (CX). After predilatation, the lesion was treated with two drug-coated balloons (DCB). Primary success in the posterolateral branch was good; however the CX lesion had a residual stenosis including a non-flow-limiting type A dissection. After 6 months, angiography showed slight lumen enlargement in both branches of the bifurcation. Intravascular ultrasound identified about 35 % atherosclerotic plaque load within the inner area of the bifurcation but more than 50 % concentric atherosclerotic plaque burden in the vessel areas proximal and distal to the DCB-treated area.
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Avoid common mistakes on your manuscript.
Clinical case
Coronary stents were developed for emergency treatment of acute vessel closure after balloon angioplasty, especially in the case of flow-limiting dissections [1]. However, this technology created two new diseases: in-stent restenosis (ISR) and stent thrombosis. Furthermore, a caged vessel excludes late lumen enlargement and advantageous vascular remodelling. Local intravascular drug delivery by drug-eluting stents (DES) reduced the problem of restenosis [2]. However, DES therapy is limited by delayed healing, late acquired malapposition [3], and neo-atherosclerosis [4]. Drug-coated balloons (DCB) are a new clinical treatment modality for coronary and peripheral artery disease [5, 6]. Proposed advantages of this approach are a homogeneous drug delivery to the vessel wall, an immediate drug release without the use of a polymer, the potential of reducing anti-platelet therapy, and finally the option of leaving no foreign object behind in the body. By far, the largest clinical evidence in coronary artery disease has been reported for DCB coated with paclitaxel-iopromide [5, 7–14].
We report a 54-year-old male patient presenting with unstable angina pectoris undergoing invasive coronary angiography. He was a slim man with a body mass index of 19 kg/m² (height 184 cm, 64 kg); the only coronary risk factor was continued smoking. Left ventricular angiography showed a normal systolic function without regional wall motion abnormalities. Coronary angiography revealed a single-vessel disease with a bifurcation lesion including a 95 % stenosis of the mid circumflex artery (CX) immediately distal to a large posterolateral branch (PL-CX) originating in a pointed angle. The CX had a slight post-stenotic vessel dilatation (Fig. 1).
We predilated the lesion with a 2.5/20-mm balloon with an acceptable angioplasty result in the CX. However, relevant carina-shift occurred resulting in a significant stenosis of the origin of the PL-CX. The PL-CX was treated with a 3.0/15-mm DCB followed by a 3.0/20-mm DCB in the CX (each SeQuent™ Please, B.Braun, Germany). Primary success in the PL-CX was good; however, the CX lesion had a residual stenosis of about 30 % including a non-flow-limiting type A dissection. This primary result fulfilled the criteria of a successful angioplasty according to the drug-eluting balloon consensus group recommendations [15]; therefore, the lesion was left without stent implantation (Fig. 1). The patient was discharged with dual antiplatelet therapy (aspirin and clopidogrel) for 4 weeks followed by aspirin alone.
After 6 months, the patient was scheduled for control angiography. The CX showed a lumen enlargement of 0.25 mm with a healed dissection. The vessel diameter of the PL-CX increased by 0.18 mm (Table 1; all measurements done by quantitative coronary angiography with the CAAS II Research software, Pie Medical, The Netherlands). Quantitative analysis of intravascular ultrasound (IVUS) showed about 35 % atherosclerotic plaque load within the inner area of the bifurcation (the area with the angiographic high-grade stenosis at baseline). Interestingly, the vessel areas proximal and distal to the DCB-treated area had more than 50 % concentric atherosclerotic plaque area which was not reflected in a lumen reduction by angiography (Fig. 2).
Conventional angioplasty and stent-based approaches are frequently associated with late lumen loss due to elastic recoil, negative vessel remodeling, or neointimal proliferation. In contrast, DCB therapy of native coronary arteries appears to be the first transluminal intravascular therapy allowing for a late lumen increase [16]. A possible explanation of this finding may be compensatory expansive remodeling similar to the compensatory mechanisms of early atherosclerosis (Kleber 2013, submitted for publication). The present case also opens up the possibility of a modification of the atherosclerotic plaque itself by DCB-only treatment.
The pioneering work of Andreas Grüntzig [17] was the beginning of the era of percutaneous transluminal coronary angioplasty (PTCA). Later on, the introduction of stents allowed the control of elastic recoil and flow-limiting dissections [1]. However, a caged vessel excludes late lumen enlargement and advantageous vascular remodelling. Fully bioresorbable scaffolds (BVS) imitate a DES within the first 6 months after implantation [18]. Afterwards, a late lumen enlargement between 6 and 24 months has been reported for some of the first-in-man patients treated with the everolimus eluting lactid-acid BVS, especially in small coronary vessels [19].
The present case report supports the observation of lumen enlargement within the first few months after DCB-only treatment. Furthermore, this treatment modality may allow for a contemporary modification of the underlying atherosclerotic plaque. Of course, this observation is only hypothesis generating. Therefore, further trials using intravascular imaging modalities like OCT or IVUS should clarify the significance of this observation as a consequence of DCB-only treatment.
References
Sigwart U, Puel J, Mirkovitch V, Joffre F, Kappenberger L (1987) Intravascular stents to prevent occlusion and restenosis after transluminal angioplasty. New Engl J Med 316:701–706
Morice MC, Serruys PW, Sousa JE et al (2002) A randomized comparison of a sirolimus-eluting stent with a standard stent for coronary revascularization. New Engl J Med 346:1773–1780
Joner M, Finn AV, Farb A et al (2006) Pathology of drug-eluting stents in humans: delayed healing and late thrombotic risk. J Am Coll Cardiol 48:193–202
Nakazawa G, Vorpahl M, Finn AV, Narula J, Virmani R (2009) One step forward and two steps back with drug-eluting-stents: from preventing restenosis to causing late thrombosis and nouveau atherosclerosis. JACC Cardiovasc Imaging 2:625–628
Scheller B, Hehrlein C, Bocksch W et al (2006) Treatment of coronary in-stent restenosis with a paclitaxel-coated balloon catheter. New Engl J Med 355:2113–2124
Tepe G, Zeller T, Albrecht T et al (2008) Local delivery of paclitaxel to inhibit restenosis during angioplasty of the leg. New Engl J Med 358:689–699
Scheller B, Clever YP, Kelsch B et al (2012) Long-term follow-up after treatment of coronary in-stent restenosis with a paclitaxel-coated balloon catheter. JACC Cardiovasc Interv 5:323–330
Unverdorben M, Vallbracht C, Cremers B et al (2009) Paclitaxel-coated balloon catheter versus paclitaxel-coated stent for the treatment of coronary in-stent restenosis. Circulation 119:2986–2994
Unverdorben M, Kleber FX, Heuer H et al (2010) Treatment of small coronary arteries with a paclitaxel-coated balloon catheter. Clin Res Cardiol 99:165–174
Wohrle J, Birkemeyer R, Markovic S et al (2011) Prospective randomised trial evaluating a paclitaxel-coated balloon in patients treated with endothelial progenitor cell capturing stents for de novo coronary artery disease. Heart 97:1338–1342
Rosli R, Zambahari R et al (2011) Paclitaxel-eluting balloon angioplasty and cobalt-chromium stents versus conventional angioplasty and paclitaxel-eluting stents in the treatment of native coronary artery stenoses in patients with diabetes mellitus. EuroInterv J EuroPCR Collab Work Group Interv Cardiol Eur Soc Cardiol 7:K83–K91
Habara S, Mitsudo K, Kadota K et al (2011) Effectiveness of paclitaxel-eluting balloon catheter in patients with sirolimus-eluting stent restenosis. JACC Cardiovasc Interv 4:149–154
Rittger H, Brachmann J, Sinha AM, et al (2012) A randomized, multicenter, single-blinded trial comparing paclitaxel-coated balloon angioplasty with plain balloon angioplasty in drug-eluting stent restenosis: the PEPCAD-DES study. J Am Coll Cardiol 59(15):1377–1382
Wohrle J, Zadura M, Mobius-Winkler S et al (2012) SeQuent Please World Wide Registry: clinical results of SeQuent please paclitaxel-coated balloon angioplasty in a large-scale, prospective registry study. J Am Coll Cardiol 60:1733–1738
Kleber FX, Mathey DG, Rittger H, Scheller B (2011) How to use the drug-eluting balloon: recommendations by the German consensus group. EuroIntervention 7(Suppl K):K125–K128
Werk M, Albrecht T, Meyer DR et al (2012) Paclitaxel-coated balloons reduce restenosis after femoro-popliteal angioplasty: evidence from the randomized PACIFIER trial. Circ Cardiovasc Interv 5:831–840
Gruntzig A (1978) Transluminal dilatation of coronary-artery stenosis. Lancet 1:263
Serruys PW, Onuma Y, Dudek D et al (2011) Evaluation of the second generation of a bioresorbable everolimus-eluting vascular scaffold for the treatment of de novo coronary artery stenosis: 12-month clinical and imaging outcomes. J Am Coll Cardiol 58:1578–1588
Diletti R, Farooq V, Girasis C et al (2013) Clinical and intravascular imaging outcomes at 1 and 2 years after implantation of absorb everolimus eluting bioresorbable vascular scaffolds in small vessels. Late lumen enlargement: does bioresorption matter with small vessel size? Insight from the ABSORB cohort B trial. Heart 99:98–105
Conflict of interest
Bruno Scheller has received lecture fees from B.Braun and Medtronic, travel support from Medtronic and B.Braun; he was named on a patent application of Charite university hospital, Berlin and is a shareholder of InnoRa GmbH, Berlin. Dieter Fischer has received lecture fees from B.Braun. Franz X. Kleber has received lecture fees and study grants from B.Braun. Ulrich Speck was named on a patent application of Charite university hospital, Berlin and is a shareholder of InnoRa GmbH, Berlin. Bodo Cremers has received lecture fees from B.Braun.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Scheller, B., Fischer, D., Clever, Y.P. et al. Treatment of a coronary bifurcation lesion with drug-coated balloons: lumen enlargement and plaque modification after 6 months. Clin Res Cardiol 102, 469–472 (2013). https://doi.org/10.1007/s00392-013-0556-3
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00392-013-0556-3