Abstract
Aim
To investigate the shape and the volume of ablation zones obtained with microwave ablation (MWA) performed with multiple antennas in liver tumours.
Materials and methods
Tumour volume, number of antennas, size (long diameter (Dl), along the antenna axis; short diameter (Ds), perpendicular to the antenna axis; vertical diameter (Dv), vertical to both Dl and Ds) and shape (roundness index (RI); 1 corresponds to a sphere) of the ablation zone, ablation volume, and complications were evaluated.
Results
Mean Dl, Ds, and Dv were 4.7 ± 1.4 cm, 3.9 ± 1.4 cm, and 3.8 ± 1.0 cm, respectively. Mean RIs (Ds/Dl, Dv/Dl, and Dv/Ds) were 0.83 ± 0.13, 0.83 ± 0.17, and 1.02 ± 0.23, respectively, without any difference between the mean RI obtained with the double (0.84 ± 0.01) and that with the triple-antenna (0.93 ± 0.13) approach (p = 0.25). Mean ablation volume was 41 ± 32 cm3 (vs. mean tumour volume 13 ± 10 cm3; range 1–40; p < 0.001). No complications were noted.
Conclusions
Simultaneous multi-antenna MWA of liver tumours results in large nearly spherical ablation zones.
Key Points
• Simultaneous multi-antenna microwave ablation of liver tumours results in nearly spherical ablation zones.
• The multi-antenna approach generates oversized ablation volumes compared with the target tumour volume.
• The multi-antenna approach is safe.
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Avoid common mistakes on your manuscript.
Introduction
Microwave ablation (MWA) generates relatively large necrosis [1]. To further increase the ablation zone, sequential MWA or simultaneous applications of multiple antennas have been proposed [2,3,4]. Recent laboratory studies performed in ex vivo and in vivo animal liver models have highlighted that the latter approach results in larger ablation zones compared with the former [2, 4]. Nevertheless, there is a substantial lack of data about this new treatment modality in patients presenting with liver tumours.
The aim of our study was to retrospectively investigate the shape and the volume of ablation zones obtained with MWA performed with multiple antennas in liver tumours.
Materials and methods
This study was approved by the institutional review board with a waiver of written informed consent. Liver tumours treated with the multi-antenna approach from December 2017 to February 2019 were included.
Percutaneous MWA
All procedures were performed under general anaesthesia using US/CT-guidance. Different 15-17G antennas were used with a commercially available MWA system (NEUWAVE™; Johnson & Johnson). According to authors’ experience, in the perspective of obtaining large safety margins, when the largest axis of the tumour measured ≥ 30 mm, a triple-antenna approach was used; for tumours with the largest axis between 2 and 3 cm, a double-antenna approach was used.
MWA was performed by simultaneous activation of all the antennas (heating cycle, 65 W; 10 min). Antennas were spaced 2 cm apart (Fig. 1). Hydro-dissection was utilised to protect nearby non-target structures if needed.
Data collection
The following data were collected: patient characteristics (age, sex); tumour characteristics (histology, location in liver segments, proximity to large (> 3 mm) vessels, maximum diameter on multiplanar pre-operative imaging, volume); procedural details (number of antennas, number of heating cycles, number of times antennas were repositioned, additional hydro-dissection, size, shape, and volume of the ablation zone, complications); and local tumour control (LTC).
Tumour volume was calculated on contrast-enhanced pre-operative imaging by using dedicated segmentation software (Centricity™ Universal Viewer; GE Healthcare).
The size of the ablation zone was obtained by measuring the long diameter (Dl, along the antenna axis), the short diameter (Ds, perpendicular to the antenna axis), and the vertical diameter (Dv, vertical to Dl and Ds) on immediate post-ablation contrast-enhanced CT. The shape of the ablation zone was evaluated through roundness indexes (RIs) by calculating the following ratios: Ds/Dl, Dv/Dl, and Dv/Ds (i.e. values close to 1 were considered spherical). The volume of the ablation zone was calculated on immediate post-ablation contrast-enhanced CT through the same aforementioned segmentation method. LTC was evaluated on follow-up contrast-enhanced CT or MRI.
Statistical analysis
The Wilcoxon and Fisher tests were used. Significance levels were set at 0.05.
Results
Baseline characteristics
Patient and tumour characteristics are summarised in Table 1. Fifteen tumours were treated (mean maximal diameter 3.0 ± 0.8 cm; range 1.6–4.4; mean volume 12 ± 9 cm3; range 1–40) in 11 sessions. Six tumours (40%) were close to large vessels. Nine tumours (60%) were treated with 3 antennas and 6 (40%) with 2. Hydro-dissection was used in 8 sessions (72.7%). No complications were noted.
Ablation area and local tumour control analysis
The results of MWA are summarised in Tables 2 and 3. After excluding tumours treated with multiple heating cycles or antenna repositioning (tumour n. 1, 3, and 12) and tumours without immediate post-ablation contrast-enhanced CT (tumour n. 6), the mean calculated diameters of the ablation zone were Dl 4.7 ± 1.4 (range 2.1–6.3); Ds 3.9 ± 1.4 (range 1.6–6.7); and Dv 3.8 ± 1.0 (range 1.8–5.5). Mean RIs were Ds/Dl 0.83 ± 0.13 (range 0.59–1.08); Dv/Dl 0.83 ± 0.17 (range 0.56–1.12); and Dv/Ds 1.02 ± 0.23 (range 0.70–1.48). Overall, the mean RI did not differ between the double- and the triple-antenna approach (0.84 ± 0.01; range 0.70–0.91; vs 0.93 ± 0.13; range 0.70–1.1; p = 0.25).
The mean ablation volume was 41 ± 32 cm3 (range 4–95), significantly larger than the mean tumour volume (13 ± 10 cm3; range 1–40; p < 0.001).
At imaging follow-up (mean 4 ± 3 months; range 1–9) available in 13/15 tumours (87%), LTC was 92% (i.e. a 7-mm residual tumour was noted on the 4-month MRI in one metastasis close to the inferior cava vein).
Discussion
The mean size of the diameters of the ablation zone was almost 4 cm, which is approximately 1 cm larger than the mean maximal diameter of treated tumours, thus indicating a satisfactory oversizing of the ablation area. This impression was further confirmed by the volumetric analysis, which proved that the mean ablation volume was threefold larger than the mean tumour volume. Moreover, although not statistically significant probably due to the small sample size, when 3 antennas were used, the mean ablation volume was nearly threefold larger than that achieved with 2 antennas, which is in line with data obtained from pre-clinical experiences [2, 5].
In our series, the RI was equal to 1 in only one axis, thus implying a more oval-shaped ablation zone. Nevertheless, the other two RIs were 0.83, thus suggesting a tendency towards roundness. Nevertheless, a more pronounced roundness was noted with the triple-antenna approach (0.93) rather than with the double one (0.84), which is also reflected by more the prominent increase in size of Dv and Ds with the former approach compared with the latter.
In Zhang et al [4], 4 activated antennas were used simultaneously and 4 cm was the best-suggested spacing interval between antennas. Nevertheless, their study was conducted with an ex vivo model, thus neglecting the “heat-sink” effect; which is not the case in vivo, where a more fitted approach is probably more adapted in accordance with Harari et al [2] and the present experience of ours.
Vogl et al [6] have recently proved that spherical ablations result in larger ablative margins. For this reason, new MWA technologies are designed to achieve such ablation geometry. In this perspective, results in vivo and ex vivo are limited to single-antenna approaches that seem effective for tumours sized less than 3 cm [1, 6,7,8] (which was not much reflected in our experience, and consequently a multi-antenna approach was started in our institution also for 2–3-cm-sized tumours). On the contrary, for tumours larger than 3 cm, the synchronous multiple-antenna approach providing larger ablation volumes seems more adapted. However, in vivo experience with such approach remains limited. Ziemlewicz et al [9] treated 107 hepatocellular carcinomas with a single-, double-, or triple-antenna approach. They did not stratify their results according to the number of antennas and reported a 1-month LTC 91.8% for tumours ≤ 3 cm; 100% for tumours sized 3.1–4.0 cm; and 50% for tumours > 4 cm.
The large ablation zones achieved in our experience were not obtained at the cost of an increased morbidity. Nevertheless, extensive hydro-dissection was used as suggested in the literature [10].
Limitations of our study deal with the retrospective nature and the small sample size.
In conclusion, simultaneous multi-antenna MWA applied to treat liver tumours results in large, nearly spherical ablation volumes.
Abbreviations
- Dl:
-
Long diameter
- Ds:
-
Short diameter
- Dv:
-
Vertical diameter
- MWA:
-
Microwave ablation
- LTC:
-
Local tumour control
- RI:
-
Roundness index
References
Saccomandi P, Schena E, Massaroni C et al (2015) Temperature monitoring during microwave ablation in ex vivo porcine livers. Eur J Surg Oncol 41(12):1699–1705
Harari CM, Magagna M, Bedoya M et al (2016) Microwave ablation: comparison of simultaneous and sequential activation of multiple antennas in liver model systems. Radiology 278(1):95–103
Yang PL, He XJ, Li HP, Zang QJ, Wang GY (2017) Image-guided minimally invasive percutaneous treatment of spinal metastasis. Exp Ther Med 13(2):705–709
Zhang TQ, Huang SM, Gu YK et al (2019) Sequential and simultaneous 4-antenna microwave ablation in an ex vivo bovine liver model. Cardiovasc Intervent Radiol. https://doi.org/10.1007/s00270-019-02241-6
Wright AS, Lee FT Jr, Mahvi DM (2003) Hepatic microwave ablation with multiple antennae results in synergistically larger zones of coagulation necrosis. Ann Surg Oncol 10(3):275–283
Vogl TJ, Basten LM, Nour-Eldin NE et al (2018) Evaluation of microwave ablation of liver malignancy with enabled constant spatial energy control to achieve a predictable spherical ablation zone. Int J Hyperthermia 34(4):492–500
Imajo K, Tomeno W, Kanezaki M et al (2018) New microwave ablation system for unresectable liver tumors that forms large, spherical ablation zones. J Gastroenterol Hepatol 33(12):2007–2014
Groeschl RT, Pilgrim CH, Hanna EM et al (2014) Microwave ablation for hepatic malignancies: a multiinstitutional analysis. Ann Surg 259(6):1195–1200
Ziemlewicz TJ, Hinshaw JL, Lubner MG et al (2015) Percutaneous microwave ablation of hepatocellular carcinoma with a gas-cooled system: initial clinical results with 107 tumors. J Vasc Interv Radiol 26(1):62–68
Garnon J, Cazzato RL, Caudrelier J et al (2019) Adjunctive thermoprotection during percutaneous thermal ablation procedures: review of current techniques. Cardiovasc Intervent Radiol 42(3):344–357
Funding
The authors state that this work has not received any funding.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Guarantor
The scientific guarantor of this publication is Afshin Gangi.
Conflict of interest
The authors of this manuscript declare no relationships with any companies, whose products or services may be related to the subject matter of the article.
Statistics and biometry
No complex statistical methods were necessary for this paper.
Informed consent
Written informed consent was waived by the Institutional Review Board.
Ethical approval
Institutional Review Board approval was obtained.
Methodology
• retrospective
• observational
• performed at one institution
Additional information
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
About this article
Cite this article
Cazzato, R.L., De Marini, P., Leclerc, L. et al. Large nearly spherical ablation zones are achieved with simultaneous multi-antenna microwave ablation applied to treat liver tumours. Eur Radiol 30, 971–975 (2020). https://doi.org/10.1007/s00330-019-06431-1
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00330-019-06431-1