Introduction

Contrast-enhanced computed tomography is the imaging method of choice for initial and follow-up imaging in oncologic patients. Iodine-based contrast agents are used to enhance soft tissue contrast, while the assessment of bone lesions remains challenging [1, 2]. Therefore, patients with high suspicions for bone metastasis frequently undergo additional imaging, e.g. magnetic resonance imaging (MRI), scintigraphy, or positron emission tomography [2].

New techniques such as dual-energy CT (DECT) may improve the assessment of bone lesions in CT by allowing for a multivariable material mapping, e.g., enabling substraction of bone structures by use of calcium suppression, quantitation of iodine, and dedicated depiction of bone marrow or gadolinium [3,4,5,6,7,8,9]. Early results suggested that virtual non-calcium (VNCa) images might improve bone lesion assessment: Several studies investigated VNCa from dual-source computed tomography (DSCT) for visualization of bone marrow edema in degenerative and/or traumatic altered bone [10,11,12,13,14]. In an oncologic context, the benefit of VNCa images for the detection of bone marrow infiltration in multiple myeloma has been investigated on a DSCT system, representing an enhanced detection of pronounced diffuse bone marrow edema [4]. However, to the author’s knowledge, no study conducted structured investigations of multi-level VNCa for investigating contrast-enhanced bone lesions such as metastases.

The goal of our study was to investigate the feasibility of VNCa from SDCT for the assessment of metastatic bone lesions in intra-individual comparison to non-metastatic bone using MRI as standard of reference.

Materials and methods

Patient identification

The institutional review board approved this mono-centric study; informed consent was waived due to its retrospective character. Identification of patients was performed based on a structured search within the picture archiving and communication system (PACS).

Inclusion criteria comprised (a) age ≥ 18 years, (b) SDCT and MRI of the corresponding region to serve as reference, and (c) examination between April 1, 2017, and August 30, 2017. The inclusion process and details on exclusion criteria are reported in Fig. 1. All examinations were carried out according to clinical indications; no scan was performed for the purpose of this study.

Fig. 1
figure 1

Flowchart of inclusion and exclusion criteria for this study. After selection of patients (left), a more granular selection and classification were performed on a per vertebra basis (right)

In total, 21 patients with a mean age of 56.0 ± 11.3 years, of which 15 were female, were included in this study. Cancer types comprised breast (n = 8), melanoma (n = 5), lung (n = 4), and colon (n = 4). Time difference between SDCT and MRI was 12 (5–56) days.

MRI protocol

All images were acquired using a 3.0-T MRI scanner using a body coil (Ingenia, Philips Healthcare). The imaging protocol included a sagittal short tau inversion recovery (STIR; time to repetition/inversion/echo, 3411/200/70 ms; slice thickness, 3 mm; gap, 0.3 mm) sequence, and an unenhanced sagittal T1-weighted turbo spin echo sequence (T1w; time of repetition/echo, 556/12 ms; slice thickness, 3 mm; gap, 0.3 mm). After review of the unenhanced acquisitions, a body weight–adapted volume of gadolinium-based contrast agent (Dotarem, 0.2 mg/kg body weight, Guerbet) was administered through a peripheral vein and sagittal T1w as well as axial T1w mDixon imaging was performed on suspicious areas only (ceT1w, analogously to unenhanced T1w; axial T1 mDixon, time of repetition/echo, 687/10 ms; slice thickness, 3 mm; gap, 0.3 mm).

SDCT imaging protocol

CT imaging was performed using a spectral detector CT (IQon, Philips). Patients were scanned in a supine, head-first position. For contrast-enhanced studies, a body weight–adapted volume of non-ionic, iodine-based contrast media was applied through a cubital vein with a constant flow of 3.5 ml/s (Accupaque 350 mg/ml, GE Healthcare) followed by a 30-ml saline flush. Bolus tracking was applied using a region of interest within the descending aorta and a threshold value of 150 HU; scans were started with a delay of 50 s. Tube current modulation was applied in all examinations (DoseRight 3D-DOM; Philips Healthcare). Relevant scan parameters comprised collimation, 64 × 0.625 mm; rotation time, 0.5 s; pitch, 0.671; matrix, 512 × 512; tube voltage, 120 kVp; tube current, 116 mAs.

Establishment of a standard of reference

Conventional CT and MRI images (STIR, T1w, ceT1w) were viewed side by side in consensus by two radiologists with 5 and 8 years of experience in oncologic imaging. If bone exhibited unremarkable signal intensity in STIR, T1w, and ceT1w, it was considered normal. Lesions showing hyperintensity in STIR, hypointensity in unenhanced T1w, and enhancement after gadolinium administration were considered metastatic.

With the given 21 patients, there were a total of 357 vertebrae (5 lumbar, 12 thoracic) eligible for inclusions. Here, exclusion criteria comprised fractures or implants for vertebral fusions, resulting in a total of 336 vertebral bodies included in the final analysis, of which 203 were metastatic (Fig. 1).

SDCT image reconstruction and post processing

SDCT provides conventional image reconstructions by combining projection data from both detector layers and using a hybrid-iterative reconstruction algorithm with a constant kernel (referred to as CI; iDose4, level 3, Philips Healthcare). For dual-energy image post processing, a PACS-compatible file is generated, referred to as spectral base images (SBI). Technically, the SBI contains photoelectric-like and Compton-like images that allow for real-time post processing in the vendor’s viewer (IntelliSpace Portal, Spectral Diagnostics Suite, Philips Healthcare). All SDCT images were reconstructed with a slice thickness and section increment of 2 mm.

We used a prototype image re-composition algorithm that allows for reconstruction of VNCa images based on these SBI files. It employs the acquired spectral data to identify and suppress calcium by replacing the HU of voxels containing calcium with a virtual HU value as similar as possible to the expected HU without calcium contribution to the attenuation. Further, it allows to freely choose a calcium suppression index. While a low calcium suppression index targets tissues with a low calcium composition weight, a high index value targets tissues with a high calcium composition weight.

The computation of VNCa images is described in detail elsewhere [4, 14]. A pre-defined soft tissue material and a user-defined target level of calcium content (indicated by the calcium suppression index) are converted to corresponding attenuation values using the National Insitute of Standards and Technology database. These are then projected to the SDCT photoelectric- and scatter-like decomposition functions (CaVector). As in all DECT systems, each material is defined by a vector representing the ratio photoelectric- and scatter-like component while its magnitude represents the quantities of the materials. The VNCa image is then derived using the calculated attenuation at 70 keV for all voxels leaving out any attenuation corresponding to the the CaVector.

For this study, images were reconstructed with a low (level 25), medium (level 50), and high (level 90) calcium suppression index (referred to as VNCalow, VNCamed, VNCahigh, respectively). All images were reconstructed in the sagittal plane (slice thickness, 3 mm).

Objective image assessment

Objective image analysis was carried out based on a circular region of interest (ROI) with a size of 100 mm2 (size was possibly reduced to avoid unrepresentative tissue). A maximum of three ROI per patient was drawn in each normal and metastatic trabecular bone on CI (NORM and MET, respectively). The ROI were then copied and pasted to identical positions in VNCa. From each ROI, averaged attenuation (HU) and standard deviation (SD) were recorded. To evaluate the separation between normal and metastatic bone, a modified contrast-to-noise ratio (mCNR) was calculated to represent differences between normal and metastatic bone as \( \mathrm{mCNR}=\frac{\mid {\mathrm{HU}}_{\mathrm{MET}}\hbox{--} {\mathrm{HU}}_{\mathrm{NORM}}\mid }{\sqrt[2]{{\left({\mathrm{SD}}_{\mathrm{MET}}\right)}^2+{\left({\mathrm{SD}}_{\mathrm{NORM}}\right)}^2}} \), adapted from earlier studies [15, 16].

Subjective image assessment

After objective image analysis was completed, two readers with 3 and 8 years of experience in oncologic imaging independently reviewed CI, VNCalow, and VNCamed images in sagittal plane and were asked to determine the presence of a suspicious bone lesion on a per vertebra basis using a binary scale. VNCahigh images were excluded from final analysis due to their poor performance in objective analysis. Due to their characteristic appearance, readers were not blinded regarding the calcium suppression indices. Different reconstructions were interpreted in three different sessions with a minimum of 2 weeks apart from each other to avoid a recall bias. The readers were explicitly free to adjust window center and width for all image reconstructions.

Statistic assessment

Interval scaled data is reported as mean ± standard deviation. Statistical analysis was performed using JMP (SAS Institute) and MedCalc (MedCalc Software). Student’s T test or ANOVA with Tukey HSD adjustment for multiple comparisons was used to compare interval scaled data. For objective analysis, sensitivity and specificity were calculated and receiver operating characteristic (ROC) curves were analyzed; the area under the curve was compared using DeLong’s method. For subjective analysis, pooled sensitivity and specificity from both readers were calculated. Youden’s index was used to determine optimal cut-off values for the CT numbers in VNCa reconstructions [17]. Statistic comparison of subjective analysis was conducted using McNemar’s test. A p ≤ 0.05 was considered to be statistically significant. Cohen’s kappa was calculated to determine inter-rater reliability and interpreted as earlier suggested: excellent (ĸ ≥ 0.8), good (ĸ ≥ 0.6), moderate (ĸ ≥ 0.4), and poor (ĸ ≤ 0.4) [18,16,20].

Results

Objective image analysis

On conventional images, attenuation within normal and metastatic bone ranged from − 46.3 to 238.8 HU (109.8 ± 60.1 HU), exhibiting a large overlap while CT numbers for VNCalow showed a wider range (− 343.3 to 60.2 HU (− 163.5 ± 103.2 HU); p ≤ 0.05; Fig. 2). Ranges for VNCamed and VNCahigh were rather small (− 89.0 to 42.6 HU and − 25.9 to 64.6 HU, respectively; further details in electronic supplement 1).

Fig. 2
figure 2

a Imaging characteristics of normal and metastatic bone CT numbers for normal and metastatic bone (NORM and MET) in conventional images (CI) and virtual non-calcium (VNCa) images with different calcium suppression indices (low, medium, and high). b Further, the standard deviation in the different measurements is reported

Noise for normal and metastatic bone was comparable with CI and VNCalow (24.6 ± 11.4 HU and 29.6 ± 16.1 HU, p = 0.451), while VNCamed and VNCahigh exhibited lower image noise as compared with CI (17.5 ± 6.1 HU and 17.5 ± 9.2 HU, p ≤ 0.05; further details in electronic supplement 1).

Overlap between normal and metastatic bone was smaller in VNCalow and VNCamed as compared with CI and VNCahigh. The latter showed the largest overlap, indicating an impaired separation of normal and metastatic bone (Fig. 2; further details in electronic supplement 1). mCNR was able to further substantiate a significantly improved differentiability in VNCalow (Fig. 3a; further details in electronic supplement 1).

Fig. 3
figure 3

Separation between normal and metastatic bone. a Contrast between normal and metastatic bone (as indicated by the modified contrast-to-noise ratio, mCNR) was improved in virtual non-calcium images with a low calcium suppression index (VNCa low) as compared with conventional images (CI) and different calcium suppression indices. Significant differences as indicated by ANOVA are shown and indicated by an asterisk (p ≤ 0.05). b Receiver operating characteristics substantiate this finding further by comparing the area under the curve (AUC)

Further, the discrimination of normal versus metastatic bone as illustrated by the area under the curve in ROC analysis was significantly higher in VNCalow and VNCamed as compared with CI and VNCahigh (Fig. 3b; CI vs VNCalow, 0.74 vs 0.95, p ≤ 0.05; VNCalow vs VNCamed, 0.95 vs 0.98, p ≥ 0.05; VNCamed vs VNCahigh, 0.98 vs 0.73, p ≤ 0.05; CI vs VNCahigh, 0.74 vs 0.73, p ≥ 0.05). Ideal cut-off values were identified by determination of Youden’s index for all reconstructions. Youden’s index was found to be highest in VNCalow and VNCamed (CI 89.6, 115.3; VNCalow − 174.9, − 143.2; VNCamed − 39.2, − 30.8; VNCahigh 24.8; [two values are reported if Youden’s index had two maxima]).

Subjective image analysis

In CI, readers correctly identified 78% of metastatic vertebrae with a specificity of 82%. VNCalow yielded a clear improvement in sensitivity to 85% while specificity increased slightly as well (84%, p ≤ 0.05 and p > 0.05, respectively) (Figs. 4 and 5). VNCamed showed a decrease in sensitivity while specificity increased (71% and 88%, both p ≤ 0.05, respectively). Detailed reports on outcome of subjective analysis are summarized in Table 1. Inter-rater reliability was good (ĸ = 0.630–0.653; Table 1).

Fig. 4
figure 4

Conventional and virtual non-calcium images. Sagittal image reconstructions of a 58-year-old patient as conventional images (CI) and virtual non-calcium images with different calcium suppression indices (VNCa low, medium, and high). Notably, iodine is suppressed in low indices illustrating the limits of spectral separation (arrow in VNCa low)

Fig. 5
figure 5

Multimodality imaging of a metastasis in thoracic vertebra 6. Image of a 63-year-old female patient with a metastasis in thoracic vertebra 6. The lesion is not visible in the conventional image (CI), while the virtual non-calcium image with low calcium suppression index clearly illustrates the small metastasis in the right posterior portion of the vertebral body (VNCalow); however, applying moderate or high calcium suppression indices, the metastasis is not visible either (VNCamed). Sagittal MRI corresponds well with non-contrast-enhanced T1 hypointensity and hyperintensity in STIR imaging (T1 NCE and STIR, respectively). Follow-up CT after radiation therapy clearly demonstrates osteoblastic changes (CT follow-up)

Table 1 Sensitivity, specificity, and inter-rater reliability (Cohen’s kappa) for conventional images (CI) and virtual non-calcium (VNCa) images with different strengths (low and medium). In addition, 95% confidence intervals for sensitivity and specificity and the area under the curve are reported. An asterisk indicates significant differences as compared with CI

Discussion

This study investigated multi-level calcium suppression in SDCT images and its benefit in assessing vertebral metastases. We found that calcium can be suppressed in SDCT images using the investigated prototype post processing algorithm. The obtained images allow for a better discrimination of normal and metastatic bone as indicated by ROC analysis. Further, subjective assessment demonstrated improved sensitivity and specificity as compared with CI, however, only if using low-level calcium suppression indices (level 25). This is of particular interest, since Neuhaus et al using the same scanner and a comparable imaging protocol reported that high-level calcium suppression (level 70–100) is recommended for the detection of fracture-associated edema [4].

The reason for this is likely the closeness of the attenuation profiles of calcium and iodine; therefore, higher calcium suppression levels may suppress iodine content along with calcium content. Material decomposition is limited by the spectral capability, which is inherent to the used DECT technique [3]. Tube-modifying setups allow for a modification and therefore improvement of spectral separation in a wide range (70–140 kV) [5, 21, 22]. In SDCT, the spectral separation is an inherent characteristic of the dual-layer detector [23, 24]. According to the vendor, the separation between low- and high-energy photons is optimized in a way that both detector layers exhibit comparable noise characteristics when examining an adult. Naturally, this results in an overlap of low- and high-energy photon spectra; however, besides spectral overlap, noise plays a crucial role in the spectral capability of any DECT system as noise arises from various factors which add quadratically [14, 25]. Due to the temporospatial coherence of low- and high-energy datasets inherent to the concept of the dual-layer detector, anti-correlation of noise can be exploited for image reconstruction resulting in a considerable decrease of image noise which in turn increases spectral capability [16, 26,24,25,26,30]. Outside bone, the capability and accuracy of SDCT for material decomposition have been demonstrated earlier in phantoms for iodine and gadolinium and for urinary concrements [8, 30, 31]. In contrast to the detection of bone metastases, the use of calcium suppression for the detection of diffuse bone edema needs targeted calcium suppression to the bone mineral content [4], and usually targets at the highest calcium suppression indices. Although different DECT systems vary with respect to spectral separation, noise, and temporal and/or spatial misalignment, a multi-level calcium suppression in order to allow for a better delineation of metastases may be beneficial to these DECT approaches as well.

Subjective image analysis substantiated the objective findings showing an improvement of both sensitivity and specificity using VNCalow. Nevertheless, it needs to be emphasized that conventional images need to be reviewed in the first place and independently since they represent the standard of care. In a meta-analysis, Yang et al reported sensitivity and specificity regarding the detection of metastasis on conventional CT images of 77.1% and 83.2% [32]. Their reports correspond well to obtained values in our analysis using CI; however, by using VNCalow, we found values clearly higher which further substantiates our objective analysis and a benefit from calcium suppression in CT imaging [5, 32]. Besides VNCa, there are other tools and software such as temporal subtraction toolkits available which may further improve diagnostic performance [33, 34]. To use VNCa images in daily routine, overlay with conventional images may be especially helpful since they can help to raise suspicion in patients with known metastatic disease. While we demonstrated that VNCalow images from SDCT improve visualization and detection of vertebral metastases, future large-scale studies are necessary to understand possible pitfalls of the algorithm and the impact of the technology on follow-up investigations in the clinical routine. Another limitation of this study is that only lesions within the vertebral body were included. In line with earlier studies, we used MRI as a reference standard, but could not obtain histologic proof of every lesion [5, 6, 10, 35]. Due to the small sample size, we were unable to perform a subgroup analysis to differentiate between malignant diseases. It remains elusive if VNCa images provide a particular benefit for a specific cancer, and larger, more specific studies are encouraged to investigate this.

This is one of the first studies to investigate calcium subtraction in SDCT examinations. We demonstrated feasibility of calcium suppression in SDCT and that such images are capable to improve differentiation between metastatic and normal bone. Multi-level VNCa assessment reconstructed from SDCT improved separation of normal and metastatic bone and diagnostic performance when using low to intermediate calcium suppression indices.