Introduction

Cancer is diagnosed during pregnancy in approximately 1:1,000–2,000 pregnancies [1]. The incidence rate is expected to rise in the coming years due to a rising trend of delaying pregnancy to a later age [2]. The most common types of cancer diagnosed during pregnancy include breast cancer, haematological malignancies, melanoma and cervical uterine cancer [3]. A standardised approach is often lacking and poses significant conflicts between maternal benefit and foetal risk. Fear of foetal radiation exposure often leads to suboptimal staging. Computed tomography (CT) or radionuclide examinations should not be withheld from pregnant patients if a scan is medically indicated for the benefit of the mother or the foetus [4]. However, ionizing radiation such as CT and fluorodeoxyglucose positron emission tomography/CT (FDG-PET/CT) should be avoided during pregnancy whenever possible [5]. The most obvious methods to use to avoid radiation are ultrasound and magnetic resonance imaging (MRI). Both modalities are mainly used for detailed local-regional disease assessment or a specific organ (e.g. liver). While providing excellent information on the organ system examined, this approach often requires multimodality and multistep diagnostics.

MRI has a number of advantages for oncological staging as the technique allows more reproducible evaluation of entire organ systems and – more recently – whole body (WB) evaluation. Moreover, MRI allows evaluation of functional tissue properties through the use of diffusion-weighted imaging (DWI) without the need for a contrast-agent. DWI visualises tumoral lesions by combining heavy diffusion-weighting and background signal suppression of organs, blood vessels and body fluids [6]. DWI generates image contrast by probing differences in water molecule movement (Brownian molecular motion) between tissues with different cellularity, extracellular microstructure and microcirculation. The degree of impediment or restriction of water diffusion in biological tissue increases with increasing tissue cellularity. The more water molecule movement is restricted, the brighter lesions appear at DWI with heavy diffusion weighting up to b1000. This results in high signal contrast with tumoral lesions being depicted as bright foci in contrast to the suppressed background tissue [7].

Technological innovations have made whole-body diffusion-weighted magnetic resonance imaging (WB-DWI/MRI) a time-efficient method, with thin slice acquisition, millimetre-sized spatial resolution and robust performance. These advantages explain its promising development for tumour screening and staging [6, 8, 9]. There is already a decent body of evidence showing that DWI has satisfying sensitivity and specificity for the detection of nodal and visceral metastases, including peritoneal, liver, bone and pleural involvement [10,11,12].

We hypothesise that WB-DWI/MRI is a radiation-free single-step modality for diagnosis and staging of cancer during pregnancy, while reducing the need for multimodality, invasive staging. The objective of our study is to evaluate the feasibility of whole-body diffusion-weighted magnetic resonance imaging (WB-DWI/MRI) for detecting primary tumour, nodal and distant metastases in women with cancer during pregnancy.

Materials and methods

Patients

Approval for this prospective single-centre academic pilot study was obtained from the local institutional ethics review board. Written informed consent was provided by all included patients. The inclusion criterion was clinical diagnosis of cancer during pregnancy. Exclusion criteria were previous history of malignancy prior to conception and contraindications to MRI (e.g. pacemaker, claustrophobia).

Between September 2012 and January 2015, 22 consecutive pregnant patients (mean age 35.8 years, range 29–40) with suspected malignancy were invited to participate in this study; two patients declined due to claustrophobia. Twenty patients underwent WB-DWI/MRI in addition to routine staging procedures including diagnostic clinical/laboratory, surgical and imaging work-up. Extent and types of routine clinical staging modalities – hereafter called conventional staging – were chosen by the treating physician. The conventional staging methods compared to WB-DWI/MRI are depicted in Table 1. TNM-classification [13] and Ann-Arbor classification in case of suspected lymphoma was used for staging.

Table 1 Applied modalities for conventional staging versus whole-body diffusion-weighted MRI (WB-DWI/MRI)

The study was designed so that – after conclusion of all diagnostic work-up – metastases relevant for therapeutic decisions detected by WB-DWI/MRI were disclosed to the treating physician in order to allow for biopsy or correlative imaging to conclude the diagnostic process and in order to adapt treatment when necessary.

Imaging technique: whole body diffusion MRI

All patients underwent WB-DWI/MRI at 3 Tesla field-strength with parallel radiofrequency transmission and phased-array head–neck and surface coils (Ingenia Philips Medical Systems, Best, The Netherlands). The MRI-system has a bore diameter of 70 cm, which is helpful to comfortably scan pregnant patients. Free breathing short-tau inversion recovery (STIR) WB-DWI/MRI was acquired in the transverse plane at b0 and b1000 s/mm2, from the head to below the pelvis. Whole body images were generated automatically by the scanner’s software by reconstructing multiplanar reformatted (MPR) coronal and sagittal WB-DWI/MRI images from the transverse b1000 images.

For anatomical reference for visceral organs and lymph nodes, whole body coronal non-fat-suppressed T2-weighted (w) single-shot turbo spin-echo (SS-TSE) and a thoracic 3D T1-weighted sequence were used. When indicated for skeletal evaluation, a T1w TSE sequence over the spine and pelvis was added. Detailed pulse sequence parameters are provided in Table 2. No oral or intravenous contrast agent was given.

Table 2 Overview of sequence parameters

Image interpretation: WB-DWI/MRI

WB-DWI/MRI was evaluated by two abdominal radiologists (10 and 2 years of experience, respectively). Observers were blinded to all information regarding the other imaging tests, clinical, laboratory and pathological findings but were aware of the clinical diagnosis of cancer.

As previously described, at WB-DWI/MRI, primary lesions and visceral metastases were recorded when nodular or mass-like b1000 hyperintensity was present, not attributable to T2 shine-through – defined as fluid-like hyperintensity at T2-weighted imaging – or physiological impeded diffusion in anatomical structures. Skeletal metastases were recorded when b1000 hyperintense lesions were present, correlated to T1 hypo-intense lesions and not attributable to T2 shine-through at T2-weighted imaging. Lymph nodes were qualitatively assessed on the basis of b1000 SI; lymph nodes showing (heterogeneous) b1000 SI equal to or higher than the solid component of the primary tumour as compared with surrounding lymph nodes were considered malignant, irrespective of nodal size. In the absence of a primary tumour or suspected lymphoma, lymph nodes showing (heterogeneous) b1000 SI higher than the surrounding lymph nodes were considered malignant [6]. Co-registered anatomical MR-images were for anatomical correlation of DWI findings, to classify physiological b1000 hyperintensity and lesions smaller than 4 mm with intermediate b1000 SI due to possible partial volume effects [6].

WB-DWI/MRI assessment was divided in the following anatomical subsites: the primary tumour site, including assessment of location and size; nodal regions, including Waldeyer’ ring, cervical region (left and right), supraclavicular region (left and right), mediastinum, pulmonary hilum, axillar region (left and right), retrocrural region, retroperitoneal region, iliac region (left and right) and inguinal region (left and right), and distant sites including the skeleton (axial and non-axial), visceral organs (liver, lungs, kidneys, pancreas, spleen) and surface lining organs (pleura and peritoneum).

Reference standard

Histopathology, either at staging laparotomy, diagnostic laparoscopy, core biopsy or fine needle aspiration cytology, was primarily used to confirm detected lesions at primary, nodal and distant potentially metastatic sites whenever possible.

For disease sites marked negative at WB-DWI/MRI or conventional staging and for lesions without histopathological correlation, post-treatment patient follow-up and remission status was used as reference standard to exclude development of metastatic lesions. For this purpose, the following imaging criteria were defined: lesions appearing significantly larger (at least 20 % increase) during follow-up or showing a significant decrease after chemotherapy (at least 30 % decrease) were considered to be true positive. Lesions initially detected that had resolved without therapy were considered false positive. Sites that were initially classified as negative but unequivocally showed tumour at follow-up examinations were considered false negative. Sites initially designated as negative and not showing any tumour during follow-up were considered true negative. Clinical and follow-up information was collected from the INCIP-registration study (www.cancerinpregnancy.org).

Statistical analysis

Analyses were performed using SAS software (version 9.2, SAS System for Windows, Cary, NC, USA). Conventional staging and WB-DWI/MRI based staging were compared for sensitivity, specificity, accuracy and positive (PPVs) and negative (NPVs) predictive values for detection of primary tumour, nodal and distant metastases. Exact 95 % confidence intervals (CIs) based on the binomial distribution are reported for all diagnostic indices (sensitivity, specificity, PPVs, NPVs, accuracy). Kappa statistics were used to quantify inter-rater reliability. Since the sample size is small when primary lesions and distant metastases are considered, a confidence interval based on the exact bootstrap distribution for the kappa value is reported instead of the asymptotic one [14]. Note that we have adapted their SAS program to handle the presence of zero values.

Since for many patients multiple sites were assessed, the CIs for the diagnostic indices and the kappa value are too liberal in the evaluation of the nodal metastases. To adjust the CI for the diagnostic indices, an approach based on the ratio estimator for the variance of clustered binary data was used [15]. However, this approach is only asymptotically valid and no CI is obtained when the diagnostic index equals 100 %. Therefore, to be conservative, we have decided to report in each setting the confidence limits from the approach yielding the widest one. The CI for the kappa on the nodal metastases is adjusted by multiplying the asymptotic variance by the design effect. The latter equals 1+rho*(m-1), where rho equals the intraclass coefficient calculated on the agreement and m the average number of sites (=10) [16]. With rho=0.036, the design effect equalled 1.324 in the current study.

Results

Patients

Ten of the 20 included patients had breast cancer (50 %), three Hodgkin’s lymphoma (15 %), two cervical uterine cancers (10 %), one ovarian borderline tumour (5 %), two colon cancers (10 %), one lung cancer (5 %) and one a conjunctival malignant tumour (5 %). Disease extent and staging according to the reference standard and according to WB-DWI/MRI for both readers is displayed in Table 3.

Table 3 Disease extent and staging according to conventional staging, whole-body diffusion-weighted MRI (WB-DWI/MRI) reader 1 and 2 and the reference standard

Three patients were referred for staging after primary tumour resection, with the question for residual tumour: one patient with cervical uterine cancer after conisation, one patient with colon cancer after appendectomy and caecal resection, and one patient with conjunctival tumour.

Comparison of conventional staging with WB-DWI/MRI

In total, 8/20 (40 %) of patients were upstaged after WB-DWI/MRI. Comparative sensitivities, specificities, accuracies, PPVs and NPVs of conventional staging and WB-DWI/MRI for both readers, are shown in Table 4.

Table 4 Comparative sensitivities, specificities, accuracies, PPVs and NPVs of conventional staging and whole-body diffusion-weighted MRI (WB-DWI/MRI)

Identical to conventional staging, WB-DWI/MRI allowed correct identification of the primary tumour in all but one patient.

For nodal staging, conventional staging underestimated clinically relevant metastatic extent in five patients, all with breast cancer (Fig. 1). Three of these five patients were presumed node negative based on conventional staging. In one patient WB-DWI/MRI additionally detected a supraclavicular adenopathy. WB-DWI/MRI allowed better detection of involved nodal disease sites in the three lymphoma patients that did not lead to significant change of stage. Reader 1 falsely assigned an ipsilateral hilar lymph node as metastatic in the patient with lung cancer.

Fig. 1
figure 1

Patient with breast cancer. (A, B) Visualisation of breast cancer by ultrasound and contrast-enhanced MRI. (C) Axillary ultrasound shows a lymph node with central hilum and regular cortex. Fine-needle aspiration cytology did not show the presence of malignant cells in the lymph node. Coronal WB-DWI/MRI consisting of a (D) b1000 WB-DWI sequence and (E) WB T2-weighted sequence shows a small axillary lymph node with b1000 increased signal (arrow). (F) Transverse b1000 DWI image shows (arrow) the right breast cancer characterised by increased signal intensity to the surrounding breast tissue. (G) As on the coronal image, the right axillar lymph node (arrow) shows increased signal intensity, similar to the breast cancer. (F, G, H) Arrowheads indicate axillar, inguinal and pelvic lymph nodes with lower signal intensity, considered as normal. Histopathology during a sentinel procedure confirmed the presence of a right axillar lymph node metastasis. The lymph nodes at other regions did not show any sign of progression during follow-up

Of the three patients with distant metastases, conventional staging correctly detected skeletal metastases in one patient; WB-DWI/MRI reader 1 correctly identified distant metastases in two patients and WB-DWI/MRI reader 2 detected distant metastases in all three patients (Fig. 2, patient with colon cancer; and Fig. 3, patient with cervical cancer). In a breast cancer patient, WB-DWI/MRI reader 1 identified a b1000 hyperintense lesion at DWI in the rib requiring additional CT for definitive diagnosis of skeletal haemangioma. To allow comparison in this study, this was considered a false-positive finding at WB-DWI/MRI for reader 1.

Fig. 2
figure 2

Patient with cancer of the transverse colon (A, B, C: *). Conventional MRI of the abdomen performed at time of diagnosis for which a T2-weighted sequence without (A) and with (C) fat suppression and (B) T1-weighted sequence were performed reveals no liver lesions. (D) WB-WI shows a b1000 hyperintense lesion corresponding to the lateral subcapsular area of segment 6/7 of the liver on (E) the WB T2-weighted image compatible with a liver metastasis and was confirmed during surgery

Fig. 3
figure 3

Patient with cervical uterine cancer. T2 weighted image in the (A) transverse and (B) sagittal plane shows an exophytic tumoral mass from the exocervix. (C) WB-DWI shows bilateral iliac lymphadenopathies (dashed arrows). (D) WB-DWI shows a hyperintense lesion corresponding to the inferior angle of the scapula at the corresponding (E) T2-weighted image compatible with a skeletal metastasis (arrows)

Both WB-DWI/MRI readers showed good to very good agreement for lesion detection [primary lesions: Kappa (95 % CI) = 1.00 (0.44–1.00); lymph nodes: Kappa (95 % CI) = 0.90 (0.78–1.00); distant metastases: Kappa (95 % CI) = 0.61 (-0.07–1.00)].

Gestational outcome

Four patients were staged during the first trimester of pregnancy (between 7 weeks (w) 2/7 days (d) and 8w 6/7d), of which one patient with a Hodgkin’s lymphoma decided to terminate the pregnancy before starting treatment, and another patient had a missed abortion at 10 w 2/7 d (histopathological examination of the curettage material revealed a partial mola hydatidiformis). Sixteen patients were staged during the second and third trimester of pregnancy, of which one patient with a monochorionic diamniotic twin pregnancy developed a twin-to-twin transfusion syndrome. Chorioamnionitis and intrauterine death of both foetuses at 18 w occurred after foetal surgery. Fifteen neonates were born healthy and without congenital anomalies (median gestational age at birth 37 w 5/7 d (range 33 w 2/7 d–40 w 0/7 d)).

Discussion

The findings of this pilot study demonstrate the feasibility of WB-DWI/MRI for staging of women with cancer during pregnancy. With good to very good observer agreement, WB-DWI/MRI allowed accurate identification of the primary tumour site and more accurate staging of nodal and distant metastases compared to conventional staging.

The improved detection of nodal metastases by WB-DWI/MRI was most beneficial in the patients with breast cancer. In one patient, an unknown supraclavicular lymphadenopathy was detected; changing treatment from upfront surgery to neoadjuvant chemotherapy and in three other patients the better detection of axillary nodal involvement could have enabled upfront axillary lymphadenectomy, potentially sparing sentinel lymph node biopsy. As patients with lymph node metastases have a poorer prognosis compared to node-negative patients (10-year recurrence risk of 70 % and 15–30 %, respectively), the ability to detect small nodal metastases by DWI may enable earlier stratification of patients at risk of recurrence or requiring more aggressive treatment [17, 18].

Separate studies have investigated DWI for regional nodal staging in breast cancer using both qualitative visual and quantitative assessment with reported sensitivities between 72.4 % and 97 % with specificities between 54.4 % and 91.7 % [19,20,21,22,23].

In comparison, reported sensitivities and specificities for ultrasound are between 26.4 % and 92 % with specificities of 55.6–98.1 % [24].

Importantly, the increase of sensitivity by DWI may come with a decrease of specificity. This was only to a minor degree in our patient series. WB-DWI/MRI poses particular potential challenges towards nodal staging, due to the potentially high workload of manual region of interest (ROI) delineation and the possible lack of reproducibility and stable threshold of apparent diffusion coefficient (ADC) measurements if quantitative evaluation would be performed for nodal characterisation [6]. Therefore, we opted to assess lymphadenopathies in a qualitative way. The high accuracy obtained in this patient series using predetermined qualitative interpretative criteria for DWI relating the b1000 SI of nodal disease to that of the primary tumour and surrounding lymph nodes is in line with previous studies in abdominopelvic and pulmonary cancers [6, 25, 26]. The high accuracy found in our and these previous studies as compared to studies in breast cancer might be due to a number of reasons, including coil position, field homogeneity, type of applied sequence (STIR-based) and lower acquisition resolution, leading to a higher signal-to-noise ratio and fewer artefacts in the axillar regions on WB-DWI/MRI compared to DWI in dedicated breast MRI examinations.

WB-DWI/MRI enabled better description of total nodal disease involvement in the three lymphoma patients, but this did not lead to substantial change in staging. In the literature, arguments both favouring and questioning MRI for staging lymphoma during pregnancy are found. As the vast majority of patients are initially treated with chemotherapy, it has been proposed to limit imaging procedures to chest X-ray and ultrasound [27]. However, correct staging and clinical management aids in optimising long-term survival [28]. Importantly, ultrasound results are not always conclusive in lymphoma due to the low sensitivity for detecting abdominal lymphadenopathies while the superimposition of air and bone inhibits ultrasonographic assessment of the mediastinum [29]. WB-DWI/MRI has shown excellent agreement up to 99.4 % with FDG-PET/CT for staging lymphoma and is indicated as a non-irradiating alternative to FDG-PET/CT for staging lymphoma and other malignancies including lung cancer [9, 25]. Contrary to ultrasound/chest X-ray, WB-DWI/MRI allows for a maximal staging effort similar to that for non-pregnant patients and also better exclusion of primary tumours other than lymphoproliferative disease in the diagnosis process.

With slightly lower inter-rater agreement compared to nodal staging, WB-DWI/MRI enabled better detection of distant metastases at primary diagnosis compared to conventional staging.

Recent studies have demonstrated a good diagnostic performance of WB-DWI/MRI for detecting distant metastatic spread, equivalent to FDG-PET/CT [12]. WB-DWI/MRI has shown particularly high diagnostic value for assessing hepatic and peritoneal metastases in digestive and ovarian cancer compared to contrast-enhanced MRI, contrast-enhanced CT or FDG-PET/CT [6, 30, 31]. WB-DWI/MRI also has shown higher accuracy than bone scintigraphy for detecting skeletal metastases [32].

The safety profile of MRI towards the foetus is subject to debate and contributes to a reluctance to use MRI for first-line clinical assessment during pregnancy. Assumed concerns include potential heating effects from radiofrequency pulses and acoustic noise possibly resulting in foetal growth restriction, premature birth and hearing impairment. A retrospective case-control study in 751 neonates did not show impaired hearing or low birth weight secondary to MRI exposure, confirming findings of previous case series [33,34,35,36].

To date, no studies have indicated that any pulse sequences at 1.5 Tesla field-strength cause significant increases in temperature [37]. A 3 Tesla MRI has the benefit of better signal-to-noise ratio, thereby improving diagnostic quality or decreasing imaging time while maintaining high image quality. A recent study evaluating intrauterine heating effects by 3 Tesla MRI in pregnant miniature pigs showed only minimal temperature increase when limiting scan time to 30 min and using low SAR sequences but cautioned for heating effects when using prolonged scan time with multiple high SAR-sequences [38]. It is unclear if similar heating effects would occur in humans. However, we believe that – with an average scanning time of 33 min for whole body staging from which half of the scan time is spaced by low SAR sequences (e.g. DWI and gradient echo T1-weighted sequences) – the WB-DWI/MRI protocol lies within the proposed scan limitations. No adverse foetal effects could be directly linked to imaging in our patient group.

Contrary to using MRI itself, the American College of Radiology (ACR) manual on safe MR practices advises for extreme caution in using gadolinium [39]. Previous studies in non-pregnant cancer patients have shown that DWI reaches at least similar accuracy for detecting primary tumours, nodal and distant metastases as would be achieved by gadolinium-enhanced MRI [10,11,12, 31]. Taking into account prior studies and our data, we believe that the use of DWI can reliably obviate the need for gadolinium contrast for staging during pregnancy [40].

The certainty of covering the entire body in a single examination likely contributed to the higher accuracy of WB-DWI/MRI over conventional staging. However, the improved detection rate of metastases may come at the cost of decreased specificity [7]. We could largely overcome this problem by correlating DWI to the co-registered anatomical T2- and T1-weighted sequences, seeing the similar specificity found between WB-DWI/MRI and conventional staging. Moreover, only one additional CT of the ribcage was requested by one reader for a false-positive WB-DWI/MRI reading of a skeletal haemangioma with T2 shine-through [7].

We acknowledge three limitations of this study. First, we could only include a small number of patients, which impacted most on the small number of patients with distant metastases reflected by the large ranges of confidence intervals. This directly results from the single-centre study setting in a patient population with low cancer incidence. However, to the best of our knowledge, the current literature only consists of two case reports. Vermoolen et al. reported the case of a woman with Hodgkin’s lymphoma at 31 w of gestation [41]. WB-MRI, performed with coronal T1- and STIR T2-weighted acquisitions, found cervical and mediastinal lymph node involvement. Montagna et al. reported the case of a pregnant woman with breast cancer; WB-DWI/MRI showed the breast nodule, axillary involvement, and bone metastases [42]. Second, WB-DWI/MRI was performed irrespective of clinical risk of distant metastatic spread. One could argue that WB-DWI/MRI is not indicated for assessing Tis-T1 breast cancer, as was done in this pilot study. However, the purpose of this study was primarily to assess sensitivity and specificity of WB-DWI/MRI, justifying the inclusion of these patients. We acknowledge that further validation of WB-DWI/MRI in a larger patient population and in a multicentre study setting is preferable. Third, even though the negative DWI/MRI findings in the lungs in this study were corroborated by the absence of lung metastases during the entire follow-up period, it is still possible that due to the resolution limitations of DWI/MRI millimetric deposits could have been missed. In case of doubt, a non-contrast chest CT examination could be considered.

In conclusion, WB-DWI/MRI is feasible for single-step non-invasive staging of cancer during pregnancy with good to excellent reader reproducibility and shows additional value for conventional imaging procedures for detecting distant and nodal metastases. If established for staging during pregnancy, WB-DWI/MRI could obviate the need for radiation, contrast-injection or multiple diagnostic tests. This is better in terms of oncological staging, time management, financial costs, prevention of foetal radiation exposure and emotional burden for the patient.