Introduction

Postmortem MRI (pmMRI) has become a valuable tool for non-invasively documenting forensic and pathological findings in the deceased [15]. Soft tissue pathology can be conducted in cross sections thinner than possible in any routine autopsy and in any geometrical plane [612]. However, MR imaging of a deceased body is far different from clinical MRI, and interpreting postmortem, unenhanced MR images requires special expertise [1315].

Certain circumstances challenge postmortem MR image acquisition and interpretation, such as different corpse temperatures at examination, causing changing tissue contrasts within the images. An approach to overcome this problem is to synthetically calculate MR images using individually optimized parameters, such as echo time (TE), repetition time (TR) and inversion time (TI). This approach is based on a recently implemented MR sequence that allows for a rapid quantification of relaxation times and proton density (PD) using a twin-echo saturation-recovery turbo spin echo acquisition [16, 17]. During a multi-slice saturation-recovery turbo spin echo (TSE) sequence, each excitation is followed by two separate echo acquisitions. This allows for a simultaneous detection of T 1 and the T 2 relaxation times independently of each other. As soon as T 1 and T 2 are known, the combination allows for a calculation of the absolute signal intensity at TE = 0 ms and TI = infinite, equal to the PD. All three parameters can be measured in a single acquisition, which avoids rescaling differences that occur in methods that use several images in combination (Fig. 1). Independent determination of T 1 and T 2 prevents error propagation, whereas in methods using consecutive measurements, determining the second parameter based on the first parameter, error propagation is a common problem. Furthermore the maps are inherently co-registered. Additionally, the sequence allows the actual local flip angle to be calculated. This calculation adds robustness compared with other alternatives requiring an additional B1 mapping for local flip-angle correction [17].

Fig. 1
figure 1

Schematic overview of the applied quantification sequence. In order to measure T1 and T2 relaxation and proton density (PD) a saturation pulse (Sat) is applied at each repetition time TR to decrease the longitudinal magnetization Mz. At 4 different saturation delay times the signal is acquired where each acquisition consists of a dual echo read-out (white dots). Hence the acquisition generates a matrix of 4 × 2 images where signal strength is proportional to the magnetization at various saturation delays and echo times. A least-square minimization algorithm finds the optimal T1, T2 and PD values for each pixel

On the basis of this acquisition and determining the relaxation times and PD in a certain region of interest (ROI), any tissue contrast image with a certain combination of TE and TR or pre-pulses may be synthetically reconstructed using an additional post-processing step. The synthetic images aid in radiologically recognizing the imaged anatomy, since they are similar to conventional MR images without relevant loss of image quality. The quantified parameters T1, T2 and PD are not vendor specific and can be used to characterize soft tissues in an objective manner by measuring and comparing three different parameters. MRI is commonly used to assess different image contrasts at different imaging parameter settings. By quantifying the tissues and lesions, tissue alterations of different individuals become comparable which provides a basis for automatic tissue recognition and computer-aided diagnostics [1621]. The purpose of the present study was to investigate and correct for the temperature dependence of postmortem MR quantification used for soft tissue characterization and differentiation in thoraco-abdominal organs, thus providing the basis for contrast optimization efforts, with respect to the broad temperature ranges of corpses.

Material and methods

In forensic cases (n = 35), a quantification sequence was added to a postmortem cardiac 3-T MR examination (Achieva, Philips Healthcare, Best, the Netherlands) [11, 12]. The examinations were performed at the Institute of Forensic Medicine, University of Zürich, Switzerland. The age at death ranged from 11 to 94 years (mean 52 years, SD 18.3). In all cases, a forensic autopsy was ordered by the local authorities. The MR examination was performed prior to autopsy. Board-certified forensic pathologists performed the autopsy directly after imaging or the following morning. The postmortem interval (time from death to imaging) ranged from several hours to 3 days. Relevant putrefaction gas formation was ruled out using whole-body computed tomography prior to pmMRI to avoid image artefacts. The corpses were wrapped in an artefact-free body bag, a linen sheet or plastic foil. All subjects were examined in supine position. The MRI examination time was 2 h, of which the quantification sequence lasted 15 min. Twenty to 24 short axis slices 4 mm in thickness were acquired with a gap of 0.3 mm. The use of the image data for the present study was approved by the local ethics committee.

A commercially available post-processing tool (SyMRI Autopsy, SyntheticMR, Linköping, Sweden) [22] was used to measure the T1 and T2 relaxation times and PD by placing five ROIs in organs (myocardium, liver, spleen) and tissues (subcutaneous fat and pectoralis muscle) covered on cardiac short axis images (Fig. 2). T1 and T2 relaxation times in milliseconds (ms) as well as the PD (%) were plotted in a 3D plot using the ThreeDifyExcelGrapher add-in for Microsoft Excel® to investigate clustering and separation of organs and tissues. Two-way analysis of variance (ANOVA) was carried out to look for significant differences of quantitative T1, T2 and PD values between all organs/tissues. The null hypothesis was rejected when the p value was less than 0.05.

Fig. 2
figure 2

Cardiac short axis synthetically calculated T1w MRI image. Quantitative values (red box) of myocardium, liver, spleen, pectoralis muscle and subcutaneous fat were obtained by placing ROIs (yellow arrow, measurement illustrated for the heart) into organs/tissues

For evaluating the temperature of thoraco-abdominal organs and tissues the rectal temperature was measured prior to and after pmMR examination. The temperatures of corpses ranged from 7 to 32 °C (mean temperature 16.8 °C, SD 7.3). Linear interpolation to the time of acquisition of the quantification sequence provided an estimation of the core body temperature at imaging. The quantitative values for T1, T2 and PD of each investigated organ/tissue were plotted and related to the body core temperature of the corpse at the time of data acquisition.

Linear equations were generated using Microsoft Excel® to assess the T1/temperature, T2/temperature and PD/temperature relationships for each investigated organ. Those equations were used to correct the values of T1, T2 and PD for all investigated organs/tissues to a temperature of 37 °C. The T1, T2 and PD values not corrected for temperature were compared to the values corrected to 37 °C using the F test for equality of variances with Microsoft Excel®.

Results

Quantitative values/temperature relations as obtained in the investigated cases were expressed by linear equations. Table 1 lists the linear equations generated for each investigated organ/tissue. Individual variation of the relation between the change of temperature and change of quantitative values was observed in all organs/tissues. A temperature dependence was observed mainly for the T1 values that increased with rising temperature in all investigated organs/tissues, as illustrated for the heart in Fig. 3. T2 and PD values were only slightly influenced by temperature in all organs/tissues.

Table 1 Linear equations for all investigated organs/tissues generated from the relation of T1 and T2 time as well as PD to core temperature at the time of scanning as obtained in 35 forensic cases
Fig. 3
figure 3

Plots of myocardial T1 and T2 relaxation times and PD vs. body core temperature at the time of scanning. Note that a strong temperature dependence is observed for the T1 values only

Table 2 shows the mean quantitative values (T1 and T2 relaxation times as well as the PD) obtained for all examined organs/tissues in 35 cases that were correlated with body temperature. Quantitative values were plotted in a 3D coordinate system and are depicted as corrected for a temperature of 37 °C (Fig. 4a) and not corrected for temperature (Fig. 4b). In the 3D plot the clusters of myocardium, liver, spleen, pectoralis muscle and subcutaneous fat can be easily discriminated from each other. A significant difference between the quantitative values of all organs/tissues could be observed in the ANOVA. A greater significance between all organs/tissues was observed for the quantitative values that were corrected to 37 °C (Table 3). Comparison of the T1 and T2 values that were corrected to 37 °C with T1 and T2 relaxation times for 3-T applications collected from the literature showed that the ex vivo values from the present study are within the range of the known in vivo values (Table 2). In the current literature there were no quantitative PD values or quantitative T1 and T2 values for pectoral muscles [2327].

Table 2 Mean quantification values (T1 and T2 in ms; PD in %) obtained from different organs/tissues in 35 forensic cases
Fig. 4
figure 4

Three defined views (PD/T2 view, T2/T1 view and PD/T1 view) on a 3D plot of different tissues based on T1 and T2 relaxation times and PD. Values are depicted as corrected for a temperature of 37 °C (a) and not corrected for temperature (b). Note that the clusters of different organs/tissues can be easily differentiated from each other. Values corrected for a temperature of 37 °C depicted slightly tighter clustering with less overlapping (seen best in fat tissue) than those not corrected

Table 3 Results of the two-way analysis of variance (ANOVA) for comparison of quantitative values for T1 (ms), T2 (ms) and PD (%) between all organs in 35 forensic cases

Table 4 shows the results from the F test that was performed to compare the variances of the distributions from the means of T1, T2 and PD values not corrected for temperature and the quantitative values corrected for 37 °C for every single organ/tissue. In every organ/tissue the values corrected to 37 °C exhibit a slightly lower standard deviation yet the variances of distributions from the mean values corrected to 37 °C did not differ significantly compared to those of the uncorrected values within the single organs/tissues.

Table 4 Results of the F test for equality of variances of T1, T2 and PD values in single organs/tissues corrected to 37 °C and not corrected for temperature in 35 forensic cases

Discussion

Synthetic MRI sequences allow for simultaneous determination of T1 and T2 tissue relaxation times as well as for tissue PD [1619]. In the present postmortem study, the synthetic MRI approach was applied for the first time to investigate different thoraco-abdominal organs and tissues. The investigated organs and tissues could be differentiated from each other by their quantitative values of T1, T2 and PD. This provides the basis for additional promising applications of quantitative MRI. Databases for quantitative values of T1, T2 and PD of normal organ tissues as well as of pathologic tissues can be established. As a result, a scale of quantitative values for tissue lesions and normal tissue can be created similar to the Hounsfield scale in CT applications, only with a far greater potential of soft tissue analysis and discrimination [20]. Using quantitative data from synthetic MRI applications also provides the basis for automatic tissue detection and pathologic tissue detection approaches. Synthetic MRI images can be loaded on a personal computer with recently developed software. Computer-aided diagnostics can be implemented on the basis of established ranges of the signal behaviour of normal postmortem tissue and pathological alterations. Pathologic tissue can, for example, automatically be detected and encoded in different colours. On one hand, this can help the image reader to detect pathologic tissue; on the other hand, information regarding the nature of the pathologic area can also be obtained. It can be speculated that the use of synthetic MRI can overcome the shortcoming of the rather insufficient soft tissue analysis in forensic postmortem imaging. The more commonly used postmortem CT is a useful tool, for example, in detecting gas or bone lesions. However, it is unsatisfactory for adequate soft tissue analysis and discrimination [3, 4, 11, 20]. The combined use of postmortem CT and postmortem quantitative MRI in routine forensic casework promises to reveal more substantial information on forensic-relevant soft tissue pathologies than so far possible. Synthetic MRI offers the potential of detecting soft tissue lesions, such as oedema, inflammation, infarction, degeneration and tumour lesions. Jackowski et al. demonstrated that it is also possible to specifically enhance haemorrhages [20]. Warntjes et al. and Vågberg et al. demonstrated using synthetic MRI in vivo that inflammatory multiple sclerosis lesions in the brain can be differentiated from normal healthy brain tissue by their quantitative T1, T2 and PD values [17, 21]. Further studies are required to reveal the factual diagnostic possibilities and limits of postmortem synthetic MRI for soft tissue analysis.

Quantified isotropic MRI data allow not only for measuring quantitative values of relaxation times and change of image contrast but also for reformatting and image plane reconstruction. Upon manuscript submission, there was no available software that combined all visualization possibilities of synthetic MRI. However, the final goal at the publication’s host institute is to establish a forensically optimized workstation for CT and MRI data sets that allows for automatic tissue detection, image plan reconstruction and volume rendering on regular radiological workstations [20, 28, 29].

The temperature dependence of the quantitative values challenges the postmortem synthetic MRI approach. In the present study temperature dependence for the T1 relaxation time in all examined tissues was observed and investigated. Preliminary studies demonstrated a temperature dependence of T1 relaxation times in blood and living tissue [3033]. It was shown that the T1 relaxation time correlated linearly with rising temperature. Our data confirm these examination results. As with Youl et al., we did observe that T1 changes also differed between different tissues [34]. As the proton spin behaviour, especially the longitudinal spin–lattice T1 relaxation, is influenced by tissue temperature, it explains the strong T1 temperature dependence [35]. In forensic practice, corpses are examined with temperatures usually ranging between 0 and 40 °C. To compare synthetic MRI quantitative data from different corpse scans, the temperature dependence of the quantitative values has to be taken into account. A valid comparison of different MRI scans is possible only when the quantitative values gained from the scans are corrected for the temperature. Therefore, the temperature-induced differences between individual corpses can be equalized. Comparison of the T1 and T2 values from the present study that were corrected to 37 °C with in vivo T1 and T2 relaxation times for 3-T applications collected from the literature showed that our ex vivo values are within the range of the known in vivo values [2327]. Therefore, if quantitative values are corrected to 37 °C, diagnostic criteria developed from the data of postmortem quantitative MRI scans may also be applied for detection of pathologic tissue in living patients [1621]. In the present study equations to correct for the temperature dependence of T1, T2 and PD to 37 °C were introduced. As a result of correcting quantitative values to 37 °C, standard deviations especially of the T1 values decreased in all organs. Those decreases were not significant in the F test. This, however, may be explained by the fact that the standard deviations per se also represent the individuals tissue compositions, e.g. in fatty and non-fatty livers. Also correction of quantitative values to 37 °C resulted in a statistically better discrimination between the examined organs and tissues which proves the need for temperature correction.

To the best of our knowledge the proton density (as percentage) has not previously been evaluated in vivo for quantification of thoraco-abdominal organs in 3-T applications. In the present study the PD values were a suitable parameter to quantify and characterize tissues. Thus, use of the PD for tissue characterization in vivo may also be promising.

A 3-T MRI system was used for the present study. Given that the relaxation behaviour of tissues is influenced by the magnitude of the magnetic field, it is to expect that the use of MRI scanners with different magnetic field strengths, such as 1.5 T, will result in different quantitative values for T1 and T2 for the same tissue and temperature [35]. Quantitative values of relaxation times should therefore be adapted not only to different temperatures but also to different magnetic field strengths.

Another aspect that has to be taken into consideration is that different tissue temperatures influence not only the quantitative relaxation times but also the image contrast. In preliminary studies, the scanning sequences used and optimized for the living did not work comparably on human corpses. In particular, the routinely established T1-weighted images were often of low contrast when applied on corpses as a result of the strong temperature dependence of the spin–lattice relaxation [6, 36]. To demonstrate the true potential of MRI in postmortem non-invasive diagnostics, the sequences have to be adapted and optimized in consideration of the corpse core temperature at the time of imaging. The imaging parameters of the widely used T1, T2 and PD weighted images need to be optimized for normal corpse temperatures ranging from 0 to 40 °C.

The present study has several limitations. Data were obtained from only 35 cases. In order to create a database with regular quantitative T1, T2 and PD values for different organs, higher case numbers are required. Another limitation is that the investigated organs and tissues were not examined histologically. Quantitative values of organs might have been influenced by histologically evident pathologies not visible in the synthetically calculated MRI images. For example, liver steatosis that was not visible in the MRI images but histologically evident might have affected the quantitative values of the liver tissue leading to a greater range of the standard deviations. Furthermore only one quantification set was acquired per corpse at one certain temperature. No corpse was examined at different temperatures. This and low case numbers may have affected the accuracy of the equations introduced for temperature/quantitative value relations. Further studies are required to improve the correlation between changing temperatures and quantitative tissue values. Thus, the same objects should be scanned at different temperatures ranging from 0 to 40 °C; then, more accurate equations to correct for the temperature dependences may be introduced.

Conclusion

Postmortem MR quantification using a combination of the relaxation times of T1, T2 and the value for PD allows for soft tissue discrimination. The application promises to open a new horizon in characterizing normal and pathologic soft tissue based on MRI images in an objective manner. Further studies are required to reveal the diagnostic possibilities and limits of postmortem synthetic MRI soft tissue analysis. The temperature dependence of the T1 values and the differing core temperatures of corpses challenge the application of MR quantification. Equations to correct for the temperature dependence at 3 T have been introduced.