Abstract
The author presents his experience with breast augmentation using a next-generation, form-stable, anatomically shaped silicone gel breast implant. Rotation is a potential complication for anatomically shaped breast implants. Anatomically shaped saline implants have been reported to have a rotation rate as high as 14%, while lower rotation rates of 1–2.6% for anatomic cohesive gel silicone implants have been reported. Currently, these implants are limited in the United States to US FDA-approved clinical trials. The author reviews the appropriate surgical techniques to prevent rotation when using these devices. A recent innovation, placement of the superior pole of the implant underneath the superficial fascia of the pectoralis major muscle, is described. Primary and secondary breast augmentations in 241 procedures using the Allergan Style 410 implant resulted in a 0.0% rotation rate. Overall, the anatomic form-stable silicone gel breast implants, when placed subfascially, improve common complications such as capsular contracture and implant rupture with improved aesthetic outcomes and patient satisfaction.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Breast implant technology has evolved in recent years with the introduction of anatomic (or teardrop-shaped) implants. These implants, available in both saline and silicone gel forms, more closely resemble the natural shape of the breast, making them readily accepted by both patients and surgeons.
Silicone gel-filled anatomic implants such as the Style 410 implant (Allergan, Inc., Irvine, CA) and the Contour Profile Gel® (CPG™) (Mentor Corporation, Santa Barbara, CA) implant are marketed in most countries for women seeking breast augmentation, revision, or reconstruction. The Style 410 implant is a form-stable, highly cohesive, silicone gel-filled breast implant constructed with a low-diffusion silicone elastomer shell (Intrasheil™ barrier technology) [2, 31]. The Style 410 implant, which is manufactured with a Biocell® surface texture and is available in various sizes, has been available in Europe and Brazil since 1994 and in Canada for general use since 2006. A premarket approval application for the device was submitted to the US Food and Drug Administration (FDA) in 2004; its current use in the US is limited to clinical trials conducted under FDA-approved study protocols. The CPG anatomic implant is a soft cohesive silicone gel implant [5]. The differences between the CPG implant and its round counterpart are the shape, slightly more crosslinking of the gel, and the slightly more textured outer surface. The CPG implant has been available in Europe and Brazil since 2003 and in Canada since 2006. Current use in the US is limited to FDA-approved clinical trials.
Cohesive gel implants, unlike traditional silicone gel implants, are form-stable, meaning that in the unlikely event of implant rupture, the gel remains confined inside the shell. Form-stable implants are available with three different types of gels: soft cohesive (CPG), highly cohesive (Style 410), and dual-gel soft touch (Style 510). All three types of implants appear to be well accepted by patients and surgeons.
History of Silicone Gel Implants
The history of silicone gel breast implants has been widely reported in the literature, with several advances and many setbacks [35]. In brief, the first generation of silicone gel implants (1963–1972) contained a thick outer shell filled with thick silicone gel. The second-generation implants (1972–1980) were manufactured with both a thin outer shell and a thin silicone gel. The third- (1980s), fourth-, and fifth-generation implants (1992 to present) returned to a thicker outer shell and a barrier layer that reduced the diffusion of the silicone gel through the implant shell. Third-generation implants have a smooth surface, round shape, and more viscous silicone gel. Fourth-generation implants have smooth or textured surfaces, round or anatomical shape, and more viscous (cohesive) silicone gel. Fifth-generation implants are form-stable devices filled with enhanced cohesive silicone gel and have smooth or textured surfaces and round or diverse anatomical shapes [2, 23, 29].
In 1992, amid concerns of safety and effectiveness, the FDA issued a moratorium on the use of silicone gel implants. A report issued in 2000 by the Institute of Medicine found no convincing evidence of connective tissue disease, neurologic disease, breast cancer, or cancer treatment interference in silicone gel breast implant patients, which were the specific concerns that led to the ban [21]. Thus, in 2006, the moratorium was lifted after the FDA accepted the vast body of scientific evidence supporting the performance and safety of these prosthetics [24].
The fifth-generation, form-stable, highly cohesive silicone gel implants such as the Style 410 and newer dual-gel Style 510 implants, which are available outside the US, are a significant improvement over earlier generations [2, 31]. Breast implant manufacturers continue to research and make advances in implant technology and design resulting in lower complication rates, better aesthetic outcomes, and more consistent results. Surgeons can maximize the likelihood of good aesthetic results by focusing on implant selection and surgical technique.
This review of breast implants discusses the aesthetic benefits of anatomic silicone gel implants, focusing on design elements that help to prevent or reduce the complications of breast implants, especially the issue of implant rotation. In addition, the placement of the implant in the subfascial plane rather than in the submuscular or subglandular plane to more effectively stabilize the implant and prevent implant rotation or contouring distortion is discussed.
Rotation of Breast Implants
Round implants were the standard breast implant until the introduction of anatomic implants. Rotation was not a problem with round implants because their symmetrical shape did not allow them to appear malpositioned [3]. However, when anatomic saline or silicone gel implants rotate, they can appear misshaped, which falls into the category of implant malposition [3, 8].
Several clinical studies specifically have evaluated the incidence of rotation of the different types of anatomic implants. Tebbetts [32] reported that no rotation occurred after 5 years among 609 primary augmentation patients with anatomic saline implants with adequate fill (McGhan style 468). Adequate fill of any breast implant, irrespective of shell characteristics, shape, or filler material (saline or silicone), is important to prevent shell wrinkling, folding, or collapse. Any of these consequences could potentially shorten the life of the implant and require reoperation. In addition, adequate fill produces an increased firmness of the breast.
In a retrospective study of mostly Mentor implants, Baeke [3] reported that anatomic saline implants had a 14% rotation rate and stated that the nipple-areola complex is often the telltale visual sign of implant rotation. When the fuller lower pole of the implant rotates 90° medially on the z axis, the nipple-areola complex shifts, pointing laterally. If the implant spins 180°, the nipple-areola complex will point down.
In a German trial among 132 women who received a total of 240 implants within 3 years, one case of implant rotation/mobility occurred following primary augmentation, three cases occurred after secondary augmentation, and two cases occurred after immediate reconstruction following mastectomy with Style 410 cohesive silicone gel implants [19].
A prospective, randomized, controlled, double-blind Swedish study of 80 women (Style 410 implants, n = 40; Eurosilicone Vertex implants, n = 40) reported a 5% rotation rate after 4-6 years (median = 5 years). Both types of implants produced similar instances of rotation [26].
An exemption study required by the FDA for premarketing approval of the Style 410 implant in the United States is ongoing. In this prospective, nonrandomized, 10-year study, 48 plastic surgeons in the US used the highly cohesive anatomic silicone gel Style 410 implant. After a 3-year follow-up of 941 women, 2.6% of primary augmentations, 4.7% of augmentation revision surgeries, 4.9% of primary reconstruction surgeries, and 3.0% of revision reconstruction surgeries resulted in implant malposition [5]. In contrast, no rotations were reported in a large Canadian retrospective study using the Style 410 implant in 235 implantations in 117 women [8]. After 4 years of follow-up, implant malpositions occurred in less than 1% (3/885) of the 467 patients receiving Style 410 implants [5].
Hedén [17] reported nine implant malpositions (1.1%) after seven Swedish plastic surgeons implanted 1676 Style 410 implants for mostly primary augmentation (6 for reconstruction after cancer resection; 6 for congenital breast deformities) in 823 women.
In addition, the soft cohesive anatomic silicone gel implants (CPG) had no rotation reported in 25 reconstruction and 10 augmentation cases followed for about 2 years [14]. A recent study reported rotation rates of 1.1, 2.3, 2.5, and 1.5% in primary augmentation (n = 551), augmentation revision (n = 146), primary reconstruction (n = 251), and revision reconstruction (n = 69) cohorts, respectively, at 2-year follow-up [12].
Causes of Implant Rotation
One theory about the cause of implant rotation is the lack of development of a connective tissue adhesion layer (or stable interface) between the implant and the capsule, thus allowing the implant to move freely in the pocket [16].
Because of its cohesive gel, the anatomic implant maintains its shape and is less likely to fold or collapse, especially in the upper pole of the breast. Implant shape is maintained whether the implant is inserted correctly or is slightly rotated or upside down. A report of 132 cases of primary augmentation found that a surgical pocket that is too large can lead to implant rotation, which occurred in 6/132 cases, and a surgical pocket that is too narrow and too high can lead to upper-pole fullness (10/132) because the gel does not sink [19].
Other factors that contribute to the possibility of implant rotation have been hypothesized in the literature. Capsular fluid, which may persist for weeks following submuscular augmentation mammaplasty, may be a cause of implant rotation; this can be circumvented by routinely using small suction drains postoperatively [3, 7]. Double capsules [27], capsule within a capsule [11], periprosthetic mesh, and prosthetic massage [28] also have been reported as possible contributors to implant rotation. An implant placed in the submuscular plane is susceptible to the dynamic action of the muscle, which may induce the implant to move laterally or vertically.
Implant Design to Prevent Rotation
Texturing of the implant with larger pore sizes encourages the development of a stable interface between the implant and the capsule, thus forming a connective tissue adhesion layer. In one study [13], surface texturing with Biocell, which is used in the Style 410 implant, showed a “mirror image” interface and full integration, as detected by scanning electron microscopy on explanted human capsules and implants. The authors concluded that the adhesive effect of the Biocell texture of the Style 410 implant may have a positive impact on implant stabilization in primary reconstruction and secondary corrections of asymmetry or bad position [13].
A group of 114 patients who received a total of 228 Style 410 implants for cosmetic augmentation were followed clinically and by ultrasonography to detect the development of a stabilizing implant-capsule interface. Two cases (0.9%) of implant rotation were detected ultrasonographically. Ultrasonography demonstrated a stable interface in 171 cases (75%), while palpation alone was able to detect the interface in only 145 cases (64%). Four cases (1.7%) were rated false positive using palpation alone compared with ultrasonography. Based on this study, ultrasonography is recommended as an additional step in routine clinical follow-up [16].
Surgical Techniques to Prevent Rotation
Surgical technique may influence the probability of implant rotation. In breast augmentation, pocket plane selection is a major influential factor in the dynamics between the implant and soft tissue following surgery. Precise pocket dissection is essential to encourage a tight fit of the implant to enhance adhesion [1].
In reoperation following the rotation of an implant, a partial capsulectomy should be performed to reposition the implant and thus enhance prosthetic adherence. Inserting the new anatomic implant into the old capsule could result in a lack of adherence and possibly a rotational relapse [28].
Placement of the implant in the submuscular and subglandular positions leads to a higher incidence of rotation. Of 41 cases of breast augmentation revision, 5 cases of rotation occurred when these positions were used [28].
Subfascial Plane
A relatively new implant position involves the subfascial plane, which is gaining popularity because of its low incidences of both implant rotation and capsular contracture [15]. The subfascial plane provides a stronger supporting system for an implant’s superior pole. This surgical technique offers more accurate control of breast shape and inframammary fold position. It also provides a more rapid postoperative recovery compared with other techniques and avoids distortion when the pectoral muscle is contracted [15].
In this technique the implant is placed in the subfascia position on the superior half (particularly the upper pole). The dissected anterior-pectoral fascia with the curve limit of the upper pole fits to the upper curved edge of the implant, creating a muscle-fascia system that surrounds and adheres to the implant (Fig. 1).
This author has had remarkable success with placing the highly cohesive anatomic Style 410 implant in the subfascial position (Fig. 2). Recently, he has been using this same position with the Style 510 implant. The newest dual-gel Style 510 implants have a concave posterior that conforms and adheres better to the chest wall. In addition, the edging of the Style 510 implant makes this device easier to control during insertion.
For exact implant adaptation and to ensure implant stability, it is important to create a tight pocket. This can be accomplished by using the exact dimensions of the implant’s height and width, which helps to avoid implant movement and/or rotation. The skin envelope should be adjusted to the implant size. If an implant has insufficient volume compared with the quantity of skin available, rotation will occur more frequently. If necessary, the skin may be adjusted with vertical or periareolar resections in order to achieve an appropriate match between the implant and pocket volume. The Style 410 or 510 implant can achieve excellent results without the submuscular placement often customary for older-style round implants. An aspirator drain should be inserted for at least 5 days to avoid liquid collection (seroma or hematoma) and induce the adherence of scar tissue to the implant. Immobilization with an occlusive curative and adhesive bandage should be used for 5 days to facilitate cicatrization around the implant as well as to maintain its ideal positioning.
In the unlikely event that implant exchange is required, a capsulectomy must be performed and internal sutures used to readapt the pocket dimensions to the new implant, following the procedures described above.
It is important to advise patients to curtail certain activities for a period of time following implantation to minimize the risk of implant rotation and other complications (Table 1).
Although a direct comparison of subfascial placement with submuscular and subglandular placements has not been undertaken, anatomic implant rotation using the latter two surgical techniques has been reported in the literature. In this author’s earlier report of 241 primary and secondary breast augmentation procedures using highly cohesive, textured, anatomic silicone gel implants (Style 410), no incidences of rotation occurred using the subfascial plane surgical technique [15]. Thus, use of this technique combined with a tight pocket results in a stable implant.
Other Complications
In addition to implant rotation, other postoperative local complications of concern to surgeons are capsular contracture and rupture. The overall complication rate is lowest for highly cohesive silicone gel implants compared with implants filled with silicone or silicone and saline [22].
The Baker classification of capsular contracture is the most popular and practical method of assessing clinical breast firmness following augmentation mammaplasty and thus determining outcome. Classifications range from Baker class I to IV. A soft visible implant (class IB), an implant with mild firmness (class II), and an implant with moderate firmness (class III) are considered good or excellent outcomes. Only a Baker class IV classification, which defines an excessively firm and symptomatic breast with poor aesthetic result, would be considered a poor outcome [30].
Several animal studies have examined the effects of implant surfaces on overall soft tissue contracture around the implanted prosthesis. In these studies and in human studies, capsular contracture was less of a concern with the newest generation of refined textured surface implants [4, 6, 9, 10].
Capsular contracture rates reported in several large studies of anatomic implants [17, 33, 34] ranged from a low of 0.5% to a high of 5.0%, with most cases classified as Baker class I to III. Capsular contracture rates also were low in the author’s previously published study [15], suggesting that implant placement in the subfascial plane does not compromise capsular contracture rates.
Implant rupture has been identified in the literature as an issue for silicone gel implants. Magnetic resonance imaging is the most accurate way to detect a possible implant rupture and is recommended by the implant manufacturers [20, 25]. Rupture rates with the first third generations of silicone gel implants were 8%, as reported in a study by Hedén et al. [18], while rupture rates with anatomic implants are shown to be low, ranging from less than 1.0–2.2% [31].
Conclusion
The latest fourth- and fifth-generation, form-stable, cohesive, silicone gel breast implants offer plastic surgeons another option to improve both surgical and aesthetic outcomes, thus meeting patient expectations and increasing patient satisfaction. Highly cohesive, Biocell-textured Style 410 and Style 510 silicone gel implants offer optimal aesthetic outcomes in breast augmentation. When implanted in the subfascial plane, these prostheses are stable, resulting in low rotation rates (<1%) over the long-term.
References
Adams WP Jr (2003) Breast deformity caused by anatomical or teardrop implant rotation. Plast Reconstr Surg 111:2110–2111
Adams WP Jr, Potter JK (2005) Breast implants: materials and manufacturing past, present and future. In: Spear S, Willey SC, Robb GL, Hammond DC, Nahabedian MY (eds) Surgery of the breast: principles and art, 2nd edn. Lippincott Williams & Wilkins, Baltimore, MD, pp 424–437
Baeke JL (2002) Breast deformity caused by anatomical or teardrop implant rotation. Plast Reconstr Surg 109:2555–2564
Barone FE, Perry L, Keller T, Maxwell GP (1992) The biomechanical and histopathologic effects of surface texturing with silicone and polyurethane in tissue implantation and expansion. Plast Reconstr Surg 90:77–86
Bengtson BP, Van Natta BW, Murphy DK, Slicton A, Maxwell GP, for the Style 410 US Core Clinical Study Group (2007) Style 410 highly cohesive silicone breast implant core study results at 3 years. Plast Reconstr Surg 120:40S–48S
Bern S, Burd A, May JW Jr (1992) The biophysical and histologic properties of capsules formed by smooth and textured silicone implants in the rabbit. Plast Reconstr Surg 89:1037–1042
Brink RR (1996) Sequestered fluid and breast implant malposition. Plast Reconstr Surg 98:679–684
Brown MH, Shenker R, Silver SA (2005) Cohesive silicone gel breast implants in aesthetic and reconstructive breast surgery. Plast Reconstr Surg 116:768–779
Burkhardt BR, Demas CP (1994) The effect of Siltex texturing and povidone-iodine irrigation on capsular contracture around saline inflatable breast implants. Plast Reconstr Surg 93:123–128
Clugston PA, Perry LC, Hammond DC, Maxwell GP (1994) A rat model for capsular contracture: the effects of surface texturing. Ann Plast Surg 33:595–599
Cross PA (2003) Double capsule or capsule within a capsule: is there a difference? Br J Plast Surg 56:76
Cunningham B (2007) The Mentor study on contour profile gel silicone MemoryGel breast implants. Plast Reconstr Surg 120:33S–39S
Danino AM, Basmacioglu P, Saito S, Rocher F, Blanchet-Bardon C, Revol M, Servant JM (2001) Comparison of the capsular response to the Biocell RTV and Mentor 1600 Siltex breast implant surface texturing: a scanning electron microscopic study. Plast Reconstr Surg 108:2047–2052
Fruhstorfer BH, Hodgson EL, Malata CM (2004) Early experience with an anatomical soft cohesive silicone gel prosthesis in cosmetic and reconstructive breast implant surgery. Ann Plast Surg 53:536–542
Goes JC, Landecker A (2003) Optimizing outcomes in breast augmentation: seven years of experience with the subfascial plane. Aesthetic Plast Surg 27:178–184
Hahn M, Kuner RP, Scheler P, Freidel K, Hoffmann G, Madjar H, Wallwiener D, Krainick-Strobel U (2008) Sonographic criteria for the confirmation of implant rotation and the development of an implant-capsule interaction (“interface”) in anatomically formed textured breast implants with texturised Biocell surface. Ultraschall Med 29(4):399–404
Hedén P, Jernbeck J, Hober M (2001) Breast augmentation with anatomical cohesive gel implants: the world’s largest current experience. Clin Plast Surg 28:531–552
Hedén P, Nava MB, van Tetering J, Magalon G, Fourie le R, Brenner RJ, Lindsey LE, Murphy DK, Walker PS (2006) Prevalence of rupture in Inamed silicone breast implants. Plast Reconstr Surg 118:303–308
Heitmann C, Schreckenberger C, Olbrisch RR (1998) A silicone implant filled with cohesive gel: advantages and disadvantages. Eur J Plast Surg 21:329–332
INAMED silicone-filled breast implants directions for use, smooth & BIOCELL texture. Allergan, Inc., Santa Barbara, CA, 2006
Institute of Medicine (2000) Reoperations and specific local and perioperative complications. In: Bondurant S, Ernster V, Herdman R (eds) Safety of silicone breast implants. National Academy Press, Washington, DC, pp 114–178
Kulmala I, McLaughlin JK, Pakkanen M, Lassila K, Holmich LR, Lipworth L, Boice JD Jr, Raitanen J, Luoto R (2004) Local complications after cosmetic breast implant surgery in Finland. Ann Plast Surg 53:413–419
Maxwell GP, Baker MB (2005) Augmentation mammaplasty: general considerations. In: Spear SL, Willey SC, Robb GL, Hammond DC, Nahabedian MY (eds) Surgery of the breast: principles and art, 2nd edn. Lippincott Williams & Wilkins, Philadelphia, PA, pp 1237–1260
Maxwell GP, Hammond DC (1993) Breast implants: smooth vs. textured. In: Habal MB, Woods JE, Morain WD, Parsons RW (eds) Advances in plastic and reconstructive surgery, vol 9. Mosby-Year Book, Inc., St Louis, MO, pp 209–220
Mentor MemoryGel Silicone Gel-Filled Breast Implants [package insert]. Mentor, Santa Barbara, CA, 2006
Niechajev I, Jurell G, Lohjelm L (2007) Prospective study comparing two brands of cohesive gel breast implants with anatomic shape: 5-year follow-up evaluation. Aesthetic Plast Surg 31:697–710
Pandya AN, Dickson MG (2002) Capsule within a capsule: an unusual entity. Br J Plast Surg 55:455–456
Panettiere P, Marchetti L, Accorsi D (2004) Rotation of anatomic prostheses: a possible cause of breast deformity. Aesthetic Plast Surg 28:348–353
Peters W (2000) Current status of breast implant survival properties and the management of the woman with silicone gel breast implants. Can J Plast Surg 8:54–67
Spear SL, Baker JL Jr (1995) Classification of capsular contracture after prosthetic breast reconstruction. Plast Reconstr Surg 96:1119–1123
Spear SL, Heden P (2007) Allergan’s silicone gel breast implants. Expert Rev Med Devices 4:699–708
Tebbetts JB (2000) Patient acceptance of adequately filled breast implants using the tilt test. Plast Reconstr Surg 106:139–147
Tebbetts JB (2002) Achieving a predictable 24-hour return to normal activities after breast augmentation: part I. Refining practices by using motion and time study principles. Plast Reconstr Surg 109:273–290
Tebbetts JB (2006) Axillary endoscopic breast augmentation: processes derived from a 28-year experience to optimize outcomes. Plast Reconstr Surg 118:53S–80S
Young VL, Watson ME (2001) Breast implant research: where we have been, where we are, where we need to go. Clin Plast Surg 28:451–483, vi
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Sampaio Góes, J.C. Breast Implant Stability in the Subfascial Plane and the New Shaped Silicone Gel Breast Implants. Aesth Plast Surg 34, 23–28 (2010). https://doi.org/10.1007/s00266-009-9429-6