Introduction

Colorectal cancer (CRC) is the second leading cause of cancer death and the third most frequently diagnosed cancer worldwide [1]. Positron emission tomography (PET) integrated with computed tomography (CT) is recommended to assess the resectability of potentially surgically curable CRC according to the NCCN guidelines [2]. The other uses of [18F]-FDG PET/CT in clinical practice for CRC include diagnosing tumors, staging disease, restaging disease, and evaluating treatment efficacy [3,4,5,6].

The axial field of view (AFOV) of conventional commercial PET/CT is approximately 15 to 30 cm. This short AFOV restricts the detection efficiency and sensitivity of the system, leading to limited image quality. To acquire whole-body PET images, conventional PET/CT still needs 5 to 9 bed positions to cover the entire torso and legs, which not only reduces the throughput but also limits the patients’ cooperation. Total-body PET/CT has been introduced into clinical practice [7, 8]. This system has an AFOV of 194 cm and was reported to have a spatial resolution of 2.9 mm, with time-of-flight (TOF) capability for a time resolution of ~ 430 ps and energy resolution of 11.7%, according to the NEMA NU-2 2018 protocol [9]. This system provided an ultra-high sensitivity of 174 kcps/MBq and a peak noise-equivalent count rate (NECR) of approximately 2 Mcps for total-body imaging [9].

With these improvements, total-body PET/CT could provide competitive image quality with ultra-low tracer activity. The first human imaging study with total-body PET/CT incorporated an ultra-low dose of injected [18F]-FDG (0.45 MBq/kg), and the images appeared to be of good quality [10]. After that, several relevant studies were undertaken by our center to test the clinical practicability of total-body PET/CT [11,12,13,14]. In research on fast scanning with a full dose of activity in oncological patients, our team found that a 1-min acquisition time was adequate for clinical diagnosis compared to a 15-min acquisition time, although the image quality was degraded [11]. Another half-dose study of lung cancer revealed that a 2-min acquisition time achieved comparable image quality to the conventional PET/CT protocol [12]. Our recent research also certified that total-body dynamic PET imaging with ultra-low activity could obtain relevant kinetic metrics of [18F]-FDG and comparable image contrast with full-activity imaging [14].

Based on these previous studies, we used an ultra-low dose (0.37 MBq/kg) of [18F]-FDG with a 7–15-min scan duration as one of the routine protocols in our center. Different cancers have different requirements for PET/CT acquisition time and image reconstruction parameters due to differences in the SUV measurements of lesions and background tissues. We were curious whether protocols could be optimized for one specific disease. Although conventional PET/CT with [18F]-FDG is useful in the N-staging and M-staging of CRC, its sensitivity for LN metastases of CRC is low, and its sensitivity drops for detecting liver metastases smaller than 10 mm [15, 16]. The improvement of sensitivity provided by the total-body PET/CT equipment can increase the lesion detectability in patient with malignant tumor. There have been no studies on total-body PET/CT with an ultra-low dose of [18F]-FDG in CRC. Therefore, in this study, we explored the boundaries of acquisition time for ultra-low-dose [18F]-FDG imaging in CRC and shared our initial experience, providing reference for follow-up multicenter study about total-body PET/CT with low-dose [18F]-FDG.

Materials and methods

Patients

This study was approved by the Medical Ethics Committee of Zhongshan Hospital Affiliated to Fudan University (2019-029R), and informed consent was obtained. From April 2020 to June 2021, a total of 91 consecutive patients with suspected or diagnosed colorectal cancer underwent total-body PET/CT with ultra-low dose (0.37 MBq/kg) were analyzed. The inclusive criteria for this study were as follows: (a) colorectal cancer confirmed with a histologic examination, (b) no previous chemotherapy, radiotherapy, or surgery before PET/CT, and (c) no history of other malignancies. Among the 91 patients, 29 were excluded from this study, as shown in Fig. 1. A total of 62 patients fulfilled all the eligibility criteria.

Fig. 1
figure 1

Study flowchart summarizing the patient enrolment process

Total-body PET/CT examination

Patients received a weight-based (0.37 MBq/kg) injection of [18F]-FDG. All patients were required to avoid strenuous exercise within 24 h and fast for at least 6 h before [18F]-FDG imaging. Blood glucose levels were measured at the time of [18F]-FDG injection. Before the [18F]-FDG injection, the fasting blood glucose level need to be less than 11 mmol/L. All patients rested quietly for approximately 60 min post-injection and then underwent PET/CT imaging [17].

3D list-mode PET data were acquired for 15 min on a total-body PET/CT (uEXPLORER, United Imaging Healthcare, Shanghai, China) with a long AFOV of 194 cm. First, the PET images were reconstructed with the entire 15-min dataset, and then the data were split into 13-, 8-, 5-, 4-, 3-, 2-, and 1-min duration groups to simulate fast scanning images. For simplicity, the images reconstructed from the 15- to 1-min datasets were referred to as G15, G13, G8, G5, G4, G3, G2, and G1 in the rest of this paper. All PET images were reconstructed with the ordered subset expectation maximization algorithm (OSEM) with the following parameters: TOF and PSF modeling, 2 iterations and 20 subsets, matrix 192 × 192, slice thickness 1.443 mm, FOV 600 mm (pixel size 3.125 × 3.125 × 2.89 mm3) with a Gaussian postfilter (3 mm), and all necessary corrections, such as attenuation and scatter correction. The details of the CT scan parameters were as follows: tube voltage 120 kV, tube current 140 mAs, pitch 1.0, collimation 0.5 mm, and reconstructed slice thickness 0.5 mm.

Qualitative PET/CT image analysis

The acquired images were analyzed by two nuclear medicine physicians with 10 years of experience. A vendor-provided workstation (uWS-MI, United Imaging Healthcare) was used to process the data. For accuracy, two readers evaluated the images individually. They assessed subjective image quality with 5-point Likert scales [18]. The scoring criteria were as follows: 1 for very poor image quality, 2 for poor image quality, 3 for average image quality, 4 for good image quality, and 5 for very good image quality. The schematic diagram is shown in Fig. 2. Only images with scores of 3 and higher were defined as qualified to meet the need for clinical diagnosis. Images with scores under 3 did not meet the need for clinical diagnosis.

Fig. 2
figure 2

Schematic diagram of the 5-point Likert scales: score of 1 for very poor image quality, score of 2 for poor image quality, score of 3 for average image quality, score of 4 for good image quality, and score of 5 for very good image quality. The schematic diagram of the 5-point Likert scales is shown on a maximum intensity projection (MIP) (a–e) and corresponding axial image of the liver (f–j)

The objective image quality evaluations were performed by an experienced nuclear medicine physician. A 2D circular region of interest (ROI) avoided the vessel walls was placed in the ascending aorta root, which was considered as the blood pool. The maximum standard uptake value (SUVmax), mean standard uptake value (SUVmean), and the standard deviation (SD) of the ROI in the blood pool were recorded. The signal-to-noise ratio (SNR) of the blood pool was calculated by dividing the SUVmean by its SD. Four circular 2D ROIs were placed in the following sites in the liver: (1) hepatic dome, (2) right posterior lobe in portal vein layer, (3) left lobe in portal vein layer, and (4) inferior horn of right lobe of liver. The diameters of these four ROIs were 20 mm, and the ROIs avoided intrahepatic lesions and large blood vessels. The SUVmax, SUVmean, and SD of the ROIs of the four different liver lesions were recorded, and an average was taken to represent the final SUVmax, SUVmean, and SD of the liver. The calculation of the liver SNR was the same as that of the blood pool. To ensure that the size and location of the ROIs were identical for each patient’s images, the ROI measurements were made first on G15, and then the copy-and-paste function was used to measure the lesion on G13, G11, G8, and so on down to G1.

Volumes of interest (VOIs) were placed on lesions in the colorectum, lymph nodes, liver, and metastases to other organs for SUVmax measurements. If multiple lesions of the colorectum and lymph nodes were found, all of them were analyzed in this study. All patients with fewer than 10 liver lesions, all of them were analyzed in this study. However, for cases of more than 10 liver lesions, we selected the largest lesion for analysis in this study. The tumor-to-background ratio (TBR) was calculated by dividing the SUVmax of the lesions by the SUVmean of the liver. To ensure that the size and location of the VOIs were identical across each patient’s images, the VOI measurements were made first on G15, and then the copy-and-paste function was used to measure the lesion on G13, G11, G8, and so on down to G1.

Lesion detectability

In this study, lesion detectability was measured by the lesion detection rate. Two nuclear medicine physicians who assessed the subjective image quality read all PET images in a joint session. The order of the images was randomized to avoid the memory effect. All FDG-avid lesions, including colorectal, lymph node, and liver lesions as well as metastatic lesions in other organs were counted only if they could be confidently identified by the readers. The location of the lesion was recorded. The results of G15 served as the control to assess lesion detectability in the other groups. Lymph nodes were graded as malignant or benign based on functional criteria on PET/CT, and the metastatic lymph nodes were categorized as positive based on their increased [18F]-FDG uptake relative to the surrounding background, independent of their size. A true positive was defined as a match between the location of the metastatic lymph nodes on pathologic examination and the location of the positive lymph nodes on [18F]-FDG PET/CT imaging.

Statistical analysis

Statistical analyses were performed with SPSS 23.0 software for Windows (IBM SPSS Inc., Armonk, NY, USA), and p values < 0.05 were considered significant. The inter-rater agreement of subjective image scores was tested with the weighted kappa test, and a kappa value > 0.81 was considered to indicate excellent agreement. Friedman’s test with post hoc comparisons Bonferroni-adjusted for multiple comparisons was used to compare differences in the subjective image quality score and lesion SUVs among groups.

Results

Patient characteristics

A total of 62 patients (43 men and 19 women) were enrolled in this study, with an average age of 64.0 ± 10.9 years and body mass index (BMI) of 23.8 ± 3.4 kg/m2. The injected FDG dose was 25.3 ± 4.8 MBq. The average blood glucose level before injection was 5.8 ± 1.0 mmol/L. The details are listed in Table 1.

Table 1 Clinicopathological characteristics of the 62 patients who underwent total-body PET/CT with an ultra-low dose of 18F-FDG

There were 51 patients who underwent colorectal cancer surgery performed by experienced colorectal surgeons, and at least 12 LNs were harvested. The diagnoses of the remaining 11 patients were pathologically confirmed through colonoscopy. All 51 patients were pathologically staged according to the 8th American Joint Committee on Cancer (AJCC) staging system. Liver surgery was performed in 3 out of 62 patients. In total, 62 patients had a total of 64 colorectal cancer lesions and 10 low-grade intraepithelial neoplasia (LGIN) lesions certified by pathology. Three patients with colorectal adenocarcinoma were diagnosed with two other LGIN lesions. Two patients with colorectal adenocarcinoma were diagnosed with another tubular adenoma with LGIN. One patient had two adenocarcinoma lesions of the transverse colon and colonic hepatic flexure. One patient with two adenocarcinoma lesions in the colonic splenic flexure and sigmoid colon was diagnosed with another two LGIN lesions.

Subjective image quality scores

The inter-reader agreement of subjective image quality was excellent (kappa = 0.875). Table 2 shows the values of the subjective image quality scores. The average subjective score was highest for G15 (4.5 ± 0.5) and then decreased from G13 (4.3 ± 0.4) to G8 (3.7 ± 0.5), G5 (3.2 ± 0.4), G4 (3.0 ± 0.3), G3 (2.3 ± 0.4), G2 (1.8 ± 0.3), and G1 (1.0 ± 0.0). For G5, G8, G13, and G15, all of the image quality scores were equal to or greater than 3 points, thus meeting the needs of clinical diagnosis. The frequency of image quality scores of 4–5 points was higher for G15 than for G8. The average scores for G8 and G15 were 3.7 ± 0.5 and 4.5 ± 0.5, respectively, with a significant difference (p = 0.012).

Table 2 Subjective image quality (number)

Objective evaluation of PET/CT image quality

As shown in Table 3, the liver SNR increased with the extension of acquisition time from G1 (7.1 ± 1.2) to G2 (9.8 ± 2.0), G3 (11.6 ± 2.2), G4 (13.1 ± 2.2), G5 (14.6 ± 2.5), G8 (17.2 ± 2.8), G13 (20.6 ± 3.4), and G15 (21.9 ± 3.4). The liver SNR on G8 was not significantly different from that on G13 (p = 0.15) and was significantly different from that on G15 (p = 0.001). In addition, the liver SUVmax decreased from G8 (3.1 ± 0.5) to G13 (3.0 ± 0.5) and G15 (3.0 ± 0.5), with a significant difference between G8 and G13 (p = 0.03). All the liver SUVmean measurements on G8, G13, and G15 were 2.7 ± 0.5, with no significant difference between G8 and G13 (p = 0.19) or between G13 and G15 (p = 0.25). The SNRs of the blood pool on G8 and G15 were 18.1 ± 4.8 and 21.9 ± 5.6, respectively, with no significant difference (p = 0.22). The SUVmax of the blood pool on G8 and G15 were 2.2 ± 0.4 and 2.1 ± 0.4, respectively, which was a significant difference (p = 0.015). The SUVmean of the blood pool on G8 (2.0 ± 0.4) was higher than that on G15 (1.9 ± 0.4), with p < 0.05. The details are listed in Table 3.

Table 3 The objective image quality values (mean and SD)

Image analysis of the lesions

Table 4 shows the SUVmax of the colorectal lesions on G1 (21.1 ± 12.6), G2 (24.0 ± 13.8), G3 (25.1 ± 14.2), G4 (25.3 ± 14.1), G5 (24.9 ± 13.8), G8 (24.5 ± 13.1), G13 (24.1 ± 12.8), and G15 (24.0 ± 12.9), with no significant difference between G8 and G15 (p = 1.00). The TBRs of the colorectal lesions on G8 (9.2 ± 4.9) and G15 (9.1 ± 4.9) were not significantly different (p = 1.00). The SUVmax and TBR of lymph nodes were not significantly different between G8 (8.4 ± 6.7 and 3.3 ± 2.5, respectively) and G15 (8.6 ± 6.7 and 3.5 ± 2.5, respectively), with p = 1.00 and p = 0.32. The SUVmax and TBR of liver lesions were not significantly different between G8 (8.4 ± 4.8 and 3.6 ± 2.4, respectively) and G15 (8.7 ± 4.7 and 3.8 ± 2.4, respectively), with p = 1.00 for both. Other areas with metastasis included the lung, peritoneum and ovaries in this study. The SUVmax and TBR of other lesions were not significantly different between G8 (8.4 ± 5.9 and 3.5 ± 2.4, respectively) and G15 (8.7 ± 6.2 and 3.7 ± 2.6, respectively), with p = 1.00 for both. The details are listed in Table 4.

Table 4 The SUVmax and TBR measurements of the lesions (mean and SD)

Lesion detectability

The results of G15 served as the control to assess lesion detectability in the other groups. In the 74 colorectal lesions (10 LGIN lesions, 64 CRC lesions), all 64 CRC lesions could be identified on all images, even G1. One of ten LGIN lesions was missed on G1, G2, and G3 and could be identified on G4, G5, G8, G13, and G15, and one of ten LGIN lesions was missed on G1, G2, G3, and G4 and could be identified on G5, G8, G13, and G15. The other eight LGIN lesions were avid on all images, even G1. The images of the patient with missed LGIN lesions are shown in Fig. 3.

Fig. 3
figure 3

A 58-year-old male with adenocarcinoma of the splenic flexure of the colon diagnosed with one LGIN lesion and two liver metastases by pathology. No positive peri-intestinal lymph nodes were found after surgery. (A) The MIP of the groups with different reconstruction times. The reviewers had the same score, and the subjective image quality scores were 1 (G1 and G2), 2 (G3), 3 (G4 and G5), 4 (G8 and G13), and 5 (G15). (B) The corresponding axial images showed LGIN lesion in colon descendens. The visualization of LGIN lesion was negative on G1, G2, G3, G4, and G5 and positive on G8, G13, and G15

Lymph node metastasis was identified by postoperative pathology in 51 patients. G15 served as the control, on which 48 FDG-positive lymph nodes were identified. On G13 and G8, 100% (48/48) lymph nodes were found, compared to 93.8% (45/48) lymph nodes on G5 and G4, 85.4% (41/48) lymph nodes on G3, 81.3% (39/48) lymph nodes on G2, and 77.1% (37/48) lymph nodes on G1. The short diameters of the missed lymph nodes were less than 9 mm. Four lymph nodes were false positives, which was consistent across groups. A total of 53 lymph nodes turned out to be false negatives, which was consistent across groups. A typical patient with extensive regional lymph node metastasis is shown in Fig. 4.

Fig. 4
figure 4

A 61-year-old female with adenocarcinoma of the sigmoid colon diagnosed by pathology. A total of 13 peri-intestinal lymph nodes were resected, all of which were negative, and one cancerous nodule was found by pathology. The last column shows the 15-min PET/CT fusion images, which showed thickening of the sigmoid colon with abnormal glucose metabolism and enlargement of part of the lymph node next to the left iliac artery and para-aortic region with increased glucose metabolism (shown by the white arrow). (A) The lesion of the sigmoid colon was FDG-avid in each duration group. (B) The lymph nodes next to the left iliac artery were FDG-avid in each duration group, despite the limited image quality of G1 and G2. (C) The enlarged lymph nodes of the para-aortic region were FDG-negative on G1 and G2 but FDG-avid on G3, G4, G5, G8, G13, and G15. (D) The para-aortic region lymph node was FDG-negative on G1, G2, and G3 but FDG-avid on G4, G5, G8, G13, and G15. (E) The para-aortic region lymph node was FDG-negative on G1, G2, G3, G4, and G5 but FDG-avid on G8, G13, and G15

Liver metastases were identified by PET/CT in 10 of 62 patients. There were 4 patients with more than 10 lesions, 2 patients with 7 lesions, and 2 patients with 2 lesions, and the remaining 2 had one lesion. G15 served as the control, and there were no missed liver lesions on G13 and G8. In total, 3, 4, 6, 7, and 9 liver lesions were missed on G5, G4, G3, G2, and G1, respectively. The images of a patient with missed liver lesions are shown in Fig. 5. The diameter of all missed liver lesions was less than 15 mm. Metastasis in other areas, including the lung, peritoneum, and ovaries, was identified by PET/CT in 7 of 62 patients. G15 served as the control, and there were no missed lesions in any group.

Fig. 5
figure 5

A 68-year-old male with adenocarcinoma of the rectum diagnosed by pathology of colonoscopy samples. The last column shows the 15-min PET/CT fusion images, which showed thickening and stenosis of the rectum with abnormal glucose metabolism and multiple liver metastases with abnormal glucose metabolism (shown by the white arrow). (A) The rectal lesion was FDG-avid in each duration group. (B, C) The liver lesions were FDG-avid in each duration group, despite the limited image quality of G1 and G2. (D) The lesion in the right posterior lobe of the liver was FDG-negative on G1, G2, G3, G4, and G5 but FDG-avid on G8, G13, and G15; the remaining two lesions were FDG-avid in each duration group. (E) The lesion in the lower corner of the right lobe of the liver was FDG-negative on G1, G2, and G3 but FDG-avid on G4, G5 G8, G13, and G15

Discussion

Before the clinical application of total-body PET/CT, the implications brought by a long AFOV and wider acceptance angle were intensively discussed [19]. It was estimated in simulations that gains of up to 40-fold in the effective count rate could be made [20]. Theoretically, an increase in sensitivity could result in a reduction in injected activity and acquisition time in clinical applications [21, 22]. As mentioned in the “Introduction” section, our previous research has suggested that ultra-low-dose imaging is possible in clinical practice. In this study on a single disease, we certified the utility of total-body PET imaging with ultra-low [18F]-FDG activity in the clinical diagnosis of colorectal cancer. We attempted to optimize the scanning protocol for one typical disease by reducing the acquisition time and provided our initial experience for the compilation of future clinical guidelines.

The image quality of G8 met the clinical diagnosis needs for CRC without affecting lesion detectability. Regarding subjective image quality scores, G8 (3.7 ± 0.5) had good quality, with all image scores equal to or greater than 3 points. In terms of objective image quality evaluations, the liver SNR of G8 was not significantly different from the liver SNR of G13 (p = 0.15), but was significantly different from that of G15 (p = 0.001). G15 served as the control. For colorectal lesions, liver lesions and other metastases were taken into consideration, and all of them could be identified on G8.

The SNR of the liver and blood pool gradually declined from G15 to G1. The shorter the acquisition time, the poorer the PET image quality was, which was consistent with previous studies [23]. This is because the number of measured counts is reduced along with the reduced acquisition time. Additionally, we found that the shortened acquisition time was accompanied by an increase in both the mean and SD of SUVmax in the liver and the blood pool. In other words, the shortened acquisition time led to deviations in SUVmax caused by noise, as discussed in the research by Halpern et al. [24]. This phenomenon did not occur for the SUVmax of lesions. Therefore, the TBR was not significantly different among G8, G13, and G15 (all p > 0.05).

Prior studies found that the specificity of traditional [18F]-FDG PET/CT was relatively high in predicting regional lymph node metastasis in colorectal cancer patients [25,26,27]. Post-operative inflammation or post-treatment changes were possible reasons for false-positive lymph node findings [28]. In this study, the patients did not receive any treatment before PET/CT imaging. Therefore, the number of false-positive lymph nodes in this total-body PET/CT study was quite small. We observed 4 false-positive lymph nodes, and 1 of them was attributed to peri-intestinal scattering artifacts that occurred in one patient. In this study, a high false-negative rate for lymph nodes was observed with total-body [18F]-FDG PET/CT, similar to in other studies using conventional PET/CT [29]. The high false-negative rate is mainly due to the following reasons. The partial volume effect causes an underestimation of radioactivity concentration for small lymph nodes. For some fused lymph nodes, the number of lymph nodes was underestimated by PET/CT compared with that by pathology. Metastatic lymph nodes are usually smaller than primary tumors, and adjacent lymph nodes may not be visible as separate structures on PET images.

There were some potential limitations in this study. We excluded patients who were waited time range out of 55–75 min after the [18F]-FDG injection when designing this study. This was because we preferred comparable results that followed that EANM guidelines. However, due to its ultra-high sensitivity, total-body PET/CT could have provided images with good quality even if the waiting time was extended. Nonetheless, further research is needed to reveal the recommended waiting time for total-body PET/CT. In this single-center study, selection bias might exist because of the limited population. Obesity affects PET image quality because of high photon attenuation and scatter. There was no further stratified analysis based on body weight in this study because of the relatively small population. Our study applied the fixed reconstruction method, not the newer generation of PET image reconstruction algorithms such as Bayes penalized reconstruction or deep learning techniques, which will also affect the PET image quality. Our study mainly focused on the analysis of PET images and not on CT images. In addition, our study was based on FDG injection protocols, and the findings may not be extrapolated to tracers beyond FDG.

With an ultra-high system sensitivity, total-body PET/CT is bound to benefit patients in various fields. Reducing the injected activity and acquisition time was the most meaningful innovation that could benefit patients. This preliminary experience in our center showed that ultra-low-dose [18F]-FDG PET/CT imaging was feasible in the clinical diagnosis and staging of early colorectal cancer. In terms of the PET reconstruction parameters in this study, total-body PET/CT with an ultra-low dose of [18F]-FDG can maintain satisfactory image quality and lesion detectability in CRC. The 8-min acquisition time could meet the clinical needs of colorectal cancer.